Documente Academic
Documente Profesional
Documente Cultură
com/healthcare
SOMATOM Drive
System Owner Manual – Dosimetry and imaging performance report
Table of contents
Index 47
SOMATOM Drive 3
Print No. C2-047.660.01.02.02
Table of contents
SOMATOM Drive 5
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
CTDI100 locations in dosimetry phantoms with the line of sight towards the front side of the
gantry
The dose is measured in single axial scans, so that the dose chamber measures
the integrated dose profile along the z-direction over 10 cm, in the center of the
scan plane. The measured values should be large compared to the accuracy of
the dosemeter. As a result, it might be necessary to select either a large tube
current and exposure time or to average the dosimeter readings over several
scans. For measurements with the dose chamber in one of the peripheral
positions it is recommended to average the results over several scans.
Otherwise the scan start angle of the individual scans may influence the result.
IEC 61674: 2012
'Medical electrical equipment – Dosimeters with ionization chambers and/or
semiconductor detectors as used in X-ray diagnostic imaging'
ISO 4037-4: 2004
'X and gamma reference radiation for calibrating dosemeters and doserate
meters and for determining their response as a function of photon energy'
To calculate the CTDI100 from the dosimeter readings (usually displayed as dose
integrated over z-direction, i.e. as Dose Length Product (DLP), e.g. in mGy · cm)
the value has to be divided by the nominal beam collimation. For scan modes
without a z-flying focal spot, the nominal beam collimation is equal to the
acquisition displayed at the scanner's console (e.g. 'Acquisition 32×0.6 mm' =
1.92 cm). For scan modes with a z-flying focal spot, the nominal beam
collimation is equal to half of the acquisition displayed on the scanner's console
(e.g. 'Acquisition 128×0.6 mm' · 0.5 = 64×0.6 mm = 3.84 cm).
( Page 40 Nominal tomographic section thicknesses)
The CTDIw is the sum of a third of the CTDI100 measured in the phantom's central
chamber position (CTDI100c) and two third of the average of CTDI100 measured in
the phantom's four peripheral chamber position (CTDI100p).
SOMATOM Drive 7
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
p – Pitch factor
NxT – Nominal collimation
n – Number of scans
vf – Table feed per scan
R – Table range
Application type Typical head Typical head perfusion Typical body Typical cardiac Typical body
Patient type Adult Adult Adult Adult Child
Scan type Sequence Multiscan Spiral Sequence Spiral
Kernel H31s H30s B30f B30f B30f
*) Flex 0.28 s, trigger: scan 70 to 70, pulsing off, 60 bpm, best phase manual, phase start 70 %
Application type Typical Head Typical Head Typical Body Typical Cardiac Typical Body
Perfusion
CTDI phantom Ø 16 cm Ø 16 cm Ø 32 cm Ø 32 cm Ø 32 cm
CTDI100 (mGy) per scan per scan per rotation per scan* per rotation
Chamber position A (central) 62.8 mGy 331 mGy 3.66 mGy 3.76 mGy 2.90 mGy
Chamber position B (Up) 64.0 mGy 347 mGy 6.76 mGy 6.3 mGy** 4.92 mGy
Chamber position C (down) 65.4 mGy 354 mGy 6.41 mGy 6.1 mGy** 4.73 mGy
Chamber position D (left) 63.5 mGy 342 mGy 6.64 mGy 6.2 mGy** 4.85 mGy
Chamber position E (right) 63.3 mGy 341 mGy 6.67 mGy 6.3 mGy** 4.87 mGy
Average peripheral 64.0 mGy 346 mGy 6.6 mGy 6.2 mGy 4.85 mGy
CTDIW 63.6 mGy 341 mGy 5.63 mGy 5.4 mGy 4.20 mGy
CTDIvol 65.5 mGy 341 mGy 9.4 mGy 6.01 mGy 3.00 mGy
CTDIvol and CTDIw are calculated values based on the measured CTDI100,
where the CTDI100 has to be measured in axial scans (according to their
21 CFR 1020.33
SOMATOM Drive 9
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
CTDI100 dose factors for varying tube Values are given for central and peripheral CTDI100 and are expressed as a
voltage and shaped filter proportion of the CTDI100 for the typical mode, shown in bold.
Dose Factors of CTDI100 for variation of tube voltage and shaped filter for central /
peripheral /weighted CTDI100; each related to typical mode (bold)
Dose Factors of CTDI100 for variation of tube voltage and shaped filter for central /
peripheral /weighted CTDI100; each related to typical mode (bold)
Dose factors for varying acquisition Values are given for central and peripheral CTDI100 and are expressed as a
type proportion of the typical mode CTDI100, shown in bold.
SOMATOM Drive 11
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
Dose factors for varying mAs Values are given for central and peripheral CTDI100 and are expressed as a
proportion of the typical mode CTDI100, shown in bold.
Dose factors are valid for deactivated CARE Dose 4D.
Current* Typical Head Typical Head Perfusion Typical Body Typical Cardiac Typical Pediatric Body
mAs factor mAs factor eff. mAs factor mAs/rot factor eff. mAs factor
20 mA 40 0.10 20 0.10 17 0.11 11 0.05 7 0.07
176 mA 352 0.90 176 0.88 147 1.00 100 0.40 63 0.64
195 mA 390 1.00 195 0.98 163 1.11 111 0.44 70 0.71
200 mA 400 1.03 200 1.00 167 1.14 114 0.46 72 0.73
274 mA 548 1.41 274 1.37 229 1.56 156 0.62 98 1.00
439 mA 878 2.25 439 2.20 367 2.49 250 1.00 157 1.60
683 mA 1366 3,50 683 3,42 570 3,88 389 1,56 244 2,49
736 mA 1472 3,77 n.a. n.a. 615 4,18 419 1,68 263 2,69
797 mA n.a. n.a. n.a. n.a. 666 4,53 454 1,82 285 2,91
800 mA n.a. n.a. n.a. n.a. n.a. n.a. 456 1,82 n.a. n.a.
*) per tube
Dose Factors for variation of mAs central and peripheral CTDI100, related to typical mode
(bold)
SOMATOM Drive
Filter (z-sharp) (z-sharp) (z-sharp) (z-sharp) (z-sharp)
Tube Voltage
CTDI100, central 2,57 4,17 4,39 2,31 3,04 2,31 2,96 3,10 2,86 2,59 2,52
CTDI100, peripheral 2,74 4,44 4,64 2,46 3,22 2,46 3,16 3,28 3,04 2,75 2,67
CTDIw 2,69 4,35 4,56 2,41 3,16 2,41 3,09 3,22 2,98 2,70 2,62
70 kV
80 kV
CTDI100, central 6,06 9,81 10,4 5,44 7,19 5,44 6,97 7,33 6,75 6,09 5,97
CTDI100, peripheral 6,26 10,1 10,6 5,62 7,38 5,62 7,21 7,52 6,95 6,29 6,12
CTDIw 6,20 10,0 10,6 5,56 7,32 5,56 7,13 7,46 6,89 6,22 6,07
90 kV
CTDI100, central 8,35 13,5 14,3 7,49 9,91 7,49 9,60 10,1 9,30 8,38 8,22
CTDI100, peripheral 8,57 13,9 14,6 7,69 10,1 7,69 9,86 10,3 9,51 8,61 8,37
CTDIw 8,49 13,8 14,5 7,62 10,0 7,62 9,77 10,2 9,44 8,53 8,32
100 kV
CTDI100, central 11,2 18,1 19,2 10,1 13,3 10,1 12,9 13,5 12,5 11,3 11,0
CTDI100, peripheral 11,4 18,5 19,5 10,3 13,5 10,3 13,2 13,7 12,7 11,5 11,2
CTDIw 11,4 18,4 19,4 10,2 13,4 10,2 13,1 13,7 12,6 11,4 11,1
110 kV
CTDI100, central 14,0 22,7 23,9 12,6 16,6 12,6 16,1 16,9 15,6 14,1 13,8
CTDI100, peripheral 14,3 23,1 24,3 12,8 16,8 12,8 16,4 17,1 15,8 14,3 13,9
CTDIw 14,2 23,0 24,1 12,7 16,7 12,7 16,3 17,1 15,8 14,2 13,9
120 kV
CTDI100, central 16,9 27,4 29,0 15,2 20,1 15,2 19,5 20,5 18,9 17,0 16,7
CTDI100, peripheral 17,3 28,0 29,4 15,5 20,4 15,5 19,9 20,8 19,2 17,4 16,9
CTDIw 17,2 27,8 29,3 15,4 20,3 15,4 19,8 20,7 19,1 17,3 16,8
130 kV
CTDI100, central 20,3 32,9 34,8 18,2 24,1 18,2 23,4 24,6 22,6 20,4 20,0
CTDI100, peripheral 20,7 33,5 35,2 18,5 24,4 18,5 23,8 24,9 23,0 20,8 20,3
CTDIw 20,5 33,3 35,1 18,4 24,3 18,4 23,6 24,8 22,9 20,6 20,2
140 kV
CTDI100, central 1,54 2,50 2,65 1,38 1,84 1,38 1,77 1,87 1,72 1,55 1,52
CTDI100, peripheral 1,53 2,48 2,62 1,38 1,82 1,38 1,76 1,85 1,71 1,54 1,51
CTDIw 1,54 2,49 2,63 1,38 1,82 1,38 1,77 1,86 1,71 1,54 1,51
Sn 100 kV
CTDI100, central 6,49 10,5 11,2 5,82 7,74 5,82 7,47 7,88 7,25 6,52 6,42
CTDI100, peripheral 6,58 10,6 11,3 5,90 7,81 5,90 7,57 7,97 7,33 6,61 6,48
Dosimetry and imaging performance report
CTDIw 6,55 10,6 11,2 5,87 7,79 5,87 7,54 7,94 7,30 6,58 6,46
Sn 140 kV
Phantom: Ø 16 cm; Shaped Filter: Standard; Typical Application: Head
1
13
Collimation [mm] 2x1 6 x 0.6 8 x 0.6 1x5 10 x 0.6 1 x 10 12 x 1.2 20 x 0.6 32 x 0.6 32 x 1.2 64 x 0.6
14
1
Acquisition [mm] 16 x 0.6 20 x 0.6 40 x 0.6 32 x 0.6 mm, 128 x 0.6
64 x 0.6 mm
Filter (z-sharp) (z-sharp) (z-sharp) (z-sharp) (z-sharp)
Tube Voltage
CTDI100, central 2,32 3,75 3,95 2,08 2,74 2,08 2,67 2,79 2,58 2,33 2,27
CTDI100, peripheral 2,28 3,69 3,85 2,05 2,67 2,05 2,62 2,72 2,52 2,29 2,21
CTDIw 2,29 3,71 3,88 2,06 2,69 2,06 2,64 2,74 2,54 2,30 2,23
70 kV
CTDI100, central 3,74 6,05 6,38 3,35 4,43 3,35 4,30 4,51 4,16 3,76 3,67
CTDI100, peripheral 3,63 5,88 6,15 3,26 4,26 3,26 4,18 4,34 4,02 3,65 3,54
CTDIw 3,67 5,94 6,23 3,29 4,32 3,29 4,22 4,40 4,07 3,68 3,58
80 kV
CTDI100, central 5,54 8,97 9,47 4,97 6,57 4,97 6,38 6,69 6,17 5,56 5,45
CTDI100, peripheral 5,36 8,68 9,06 4,81 6,28 4,81 6,17 6,41 5,93 5,38 5,21
CTDIw 5,42 8,77 9,20 4,86 6,38 4,86 6,24 6,50 6,01 5,44 5,29
90 kV
CTDI100, central 7,66 12,4 13,1 6,87 9,08 6,87 8,81 9,25 8,53 7,69 7,53
CTDI100, peripheral 7,39 12,0 12,5 6,63 8,67 6,63 8,50 8,84 8,18 7,42 7,19
CTDIw 7,48 12,1 12,7 6,71 8,81 6,71 8,60 8,98 8,30 7,51 7,31
100 kV
CTDI100, central 10,3 16,7 17,6 9,26 12,2 9,26 11,9 12,5 11,5 10,4 10,1
CTDI100, peripheral 9,94 16,1 16,8 8,92 11,7 8,92 11,4 11,9 11,0 9,99 9,68
CTDIw 10,1 16,3 17,1 9,03 11,9 9,03 11,6 12,1 11,2 10,1 9,83
110 kV
CTDI100, central 12,9 20,9 22,1 11,6 15,3 11,6 14,9 15,6 14,4 13,0 12,7
CTDI100, peripheral 12,5 20,2 21,1 11,2 14,6 11,2 14,3 14,9 13,8 12,5 12,1
CTDIw 12,6 20,4 21,4 11,3 14,9 11,3 14,5 15,1 14,0 12,7 12,3
120 kV
CTDI100, central 15,7 25,4 26,8 14,1 18,6 14,1 18,1 19,0 17,5 15,8 15,4
CTDI100, peripheral 15,2 24,6 25,8 13,6 17,9 13,6 17,5 18,2 16,8 15,3 14,8
CTDIw 15,4 24,9 26,1 13,8 18,1 13,8 17,7 18,5 17,1 15,4 15,0
130 kV
CTDI100, central 18,8 30,5 32,3 16,9 22,4 16,9 21,7 22,8 21,0 18,9 18,5
Dosimetry and imaging performance report
CTDI100, peripheral 18,2 29,5 31,0 16,3 21,5 16,3 21,0 21,9 20,2 18,3 17,8
CTDIw 18,4 29,8 31,4 16,5 21,8 16,5 21,2 22,2 20,5 18,5 18,0
140 kV
CTDI100, central 1,45 2,35 2,49 1,30 1,73 1,30 1,67 1,76 1,62 1,46 1,43
CTDI100, peripheral 1,38 2,23 2,35 1,24 1,63 1,24 1,59 1,66 1,53 1,38 1,35
CTDIw 1,40 2,27 2,40 1,26 1,66 1,26 1,61 1,69 1,56 1,41 1,38
Sn 100 kV
CTDI100, central 6,16 9,97 10,6 5,52 7,35 5,52 7,09 7,49 6,88 6,19 6,09
CTDI100, peripheral 6,00 9,72 10,3 5,39 7,12 5,39 6,91 7,26 6,69 6,03 5,91
CTDIw 6,06 9,80 10,4 5,43 7,20 5,43 6,97 7,34 6,75 6,08 5,97
Sn 140 kV
Phantom: Ø 16 cm; Shaped Filter: Narrow; Typical Application: Inner Ear
Tube Voltage
CTDI100, central 0,58 0,94 0,99 0,52 0,69 0,52 0,66 0,70 0,64 0,58 0,57
SOMATOM Drive
CTDI100, peripheral 1,29 2,08 2,20 1,15 1,52 1,15 1,48 1,55 1,43 1,29 1,26
CTDIw 1,05 1,70 1,79 0,94 1,24 0,94 1,21 1,27 1,17 1,05 1,03
70 kV
CTDI100, central 1,04 1,69 1,79 0,94 1,24 0,94 1,20 1,26 1,16 1,05 1,03
80 kV
CTDI100, central 1,66 2,68 2,84 1,49 1,97 1,49 1,91 2,00 1,85 1,66 1,63
CTDI100, peripheral 3,27 5,30 5,55 2,93 3,85 2,93 3,76 3,92 3,63 3,29 3,19
CTDIw 2,73 4,42 4,64 2,45 3,22 2,45 3,14 3,28 3,03 2,74 2,67
90 kV
CTDI100, central 2,38 3,85 4,07 2,13 2,82 2,13 2,73 2,88 2,65 2,39 2,34
CTDI100, peripheral 4,53 7,34 7,69 4,06 5,33 4,06 5,21 5,43 5,03 4,55 4,42
CTDIw 3,81 6,17 6,48 3,42 4,50 3,42 4,39 4,58 4,23 3,83 3,73
100 kV
CTDI100, central 3,28 5,31 5,62 2,94 3,90 2,94 3,77 3,97 3,66 3,29 3,23
CTDI100, peripheral 6,10 9,88 10,35 5,47 7,17 5,47 7,02 7,31 6,77 6,13 5,95
CTDIw 5,16 8,36 8,77 4,63 6,08 4,63 5,94 6,20 5,73 5,18 5,04
110 kV
CTDI100, central 4,22 6,84 7,23 3,79 5,01 3,79 4,86 5,11 4,70 4,24 4,16
CTDI100, peripheral 7,69 12,4 13,1 6,90 9,05 6,90 8,85 9,22 8,53 7,72 7,50
CTDIw 6,53 10,6 11,1 5,86 7,70 5,86 7,52 7,85 7,25 6,56 6,39
120 kV
CTDI100, central 5,29 8,57 9,07 4,75 6,29 4,75 6,09 6,41 5,90 5,31 5,22
CTDI100, peripheral 9,50 15,4 16,1 8,52 11,2 8,52 10,9 11,4 10,5 9,54 9,28
CTDIw 8,10 13,1 13,8 7,26 9,55 7,26 9,32 9,74 8,99 8,13 7,92
130 kV
CTDI100, central 6,49 10,5 11,1 5,82 7,72 5,82 7,47 7,86 7,24 6,52 6,40
CTDI100, peripheral 11,5 18,6 19,6 10,3 13,6 10,3 13,3 13,8 12,8 11,6 11,2
CTDIw 9,84 15,9 16,7 8,83 11,6 8,83 11,3 11,8 10,9 9,89 9,63
140 kV
CTDI100, central 0,54 0,87 0,92 0,48 0,64 0,48 0,62 0,65 0,60 0,54 0,53
CTDI100, peripheral 0,89 1,44 1,52 0,80 1,05 0,80 1,02 1,07 0,99 0,89 0,87
CTDIw 0,77 1,25 1,32 0,69 0,91 0,69 0,89 0,93 0,86 0,78 0,76
Sn 100 kV
CTDI100, central 2,45 3,97 4,21 2,20 2,92 2,20 2,82 2,97 2,73 2,46 2,42
CTDI100, peripheral 3,98 6,45 6,81 3,57 4,72 3,57 4,59 4,81 4,44 4,00 3,91
CTDIw 3,47 5,62 5,94 3,12 4,12 3,12 4,00 4,20 3,87 3,49 3,42
Sn 140 kV
Dosimetry and imaging performance report
15
Collimation [mm] 2x1 6 x 0.6 8 x 0.6 1x5 10 x 0.6 1 x 10 12 x 1.2 20 x 0.6 32 x 0.6 32 x 1.2 64 x 0.6
16
1
Acquisition [mm] 16 x 0.6 20 x 0.6 40 x 0.6 32 x 0.6 mm, 128 x 0.6
64 x 0.6 mm
Filter (z-sharp) (z-sharp) (z-sharp) (z-sharp) (z-sharp)
Tube Voltage
CTDI100, central 0,50 0,81 0,86 0,45 0,60 0,45 0,58 0,61 0,56 0,50 0,49
CTDI100, peripheral 0,98 1,58 1,66 0,88 1,15 0,88 1,13 1,17 1,08 0,98 0,95
CTDIw 0,82 1,33 1,39 0,74 0,97 0,74 0,94 0,98 0,91 0,82 0,80
70 kV
CTDI100, central 0,91 1,48 1,56 0,82 1,08 0,82 1,05 1,11 1,02 0,92 0,90
CTDI100, peripheral 1,65 2,68 2,79 1,48 1,94 1,48 1,90 1,97 1,83 1,66 1,61
CTDIw 1,41 2,28 2,38 1,26 1,65 1,26 1,62 1,68 1,56 1,41 1,37
80 kV
CTDI100, central 1,46 2,36 2,49 1,31 1,73 1,31 1,68 1,76 1,62 1,47 1,43
CTDI100, peripheral 2,52 4,09 4,26 2,26 2,95 2,26 2,90 3,01 2,79 2,53 2,45
CTDIw 2,17 3,51 3,67 1,95 2,54 1,95 2,50 2,59 2,40 2,18 2,11
90 kV
CTDI100, central 2,10 3,40 3,59 1,89 2,49 1,89 2,42 2,54 2,34 2,11 2,07
CTDI100, peripheral 3,52 5,69 5,93 3,15 4,11 3,15 4,05 4,19 3,89 3,53 3,41
CTDIw 3,04 4,93 5,15 2,73 3,57 2,73 3,50 3,64 3,37 3,06 2,96
100 kV
CTDI100, central 2,91 4,72 4,98 2,61 3,45 2,61 3,35 3,52 3,24 2,93 2,86
CTDI100, peripheral 4,75 7,69 8,02 4,26 5,56 4,26 5,47 5,67 5,26 4,77 4,61
CTDIw 4,14 6,70 7,01 3,71 4,86 3,71 4,76 4,95 4,59 4,16 4,03
110 kV
CTDI100, central 3,76 6,09 6,42 3,37 4,45 3,37 4,33 4,54 4,18 3,78 3,69
CTDI100, peripheral 6,03 9,76 10,2 5,41 7,05 5,41 6,93 7,18 6,66 6,05 5,84
CTDIw 5,27 8,53 8,92 4,73 6,18 4,73 6,06 6,30 5,84 5,29 5,13
120 kV
CTDI100, central 4,73 7,65 8,08 4,24 5,60 4,24 5,44 5,71 5,26 4,75 4,65
CTDI100, peripheral 7,48 12,1 12,6 6,71 8,76 6,71 8,60 8,93 8,28 7,51 7,27
CTDIw 6,56 10,6 11,1 5,88 7,71 5,88 7,55 7,86 7,27 6,59 6,39
130 kV
CTDI100, central 5,81 9,41 9,96 5,21 6,90 5,21 6,69 7,04 6,48 5,84 5,72
Dosimetry and imaging performance report
CTDI100, peripheral 9,10 14,7 15,4 8,16 10,7 8,16 10,5 10,9 10,1 9,14 8,84
CTDIw 8,00 13,0 13,6 7,18 9,40 7,18 9,21 9,58 8,87 8,04 7,80
140 kV
CTDI100, central 0,48 0,78 0,82 0,43 0,57 0,43 0,55 0,58 0,54 0,48 0,47
CTDI100, peripheral 0,71 1,15 1,20 0,64 0,83 0,64 0,82 0,85 0,79 0,71 0,69
CTDIw 0,63 1,02 1,08 0,57 0,75 0,57 0,73 0,76 0,70 0,64 0,62
Sn 100 kV
CTDI100, central 2,23 3,61 3,83 2,00 2,66 2,00 2,57 2,71 2,49 2,24 2,20
CTDI100, peripheral 3,24 5,24 5,50 2,90 3,81 2,90 3,73 3,88 3,59 3,25 3,16
CTDIw 2,90 4,70 4,94 2,60 3,43 2,60 3,34 3,49 3,22 2,92 2,84
Sn 140 kV
Phantom: Ø 32 cm; Shaped Filter: Narrow; Typical Application: Cardio, Pediatric Body, Flash (except Flash Trauma)
SOMATOM Drive
CTDI100, free air 4,48 7,25 7,65 4,02 5,30 4,02 5,16 5,40 4,98 4,50 4,40
CTDI100, free air 6,79 11,0 11,6 6,09 8,06 6,09 7,82 8,21 7,56 6,82 6,68
CTDI100, free air 9,50 15,4 16,3 8,53 11,3 8,53 10,9 11,5 10,6 9,55 9,36
CTDI100, free air 16,3 26,3 27,8 14,6 19,3 14,6 18,7 19,7 18,1 16,3 16,0
CTDI100, free air 20,0 32,4 34,3 18,0 23,8 18,0 23,0 24,3 22,3 20,1 19,7
CTDI100, free air 24,2 39,2 41,5 21,7 28,8 21,7 27,9 29,3 27,0 24,3 23,9
CTDI100, free air 28,9 46,7 49,5 25,9 34,3 25,9 33,2 35,0 32,2 29,0 28,5
CTDI100, free air 2,00 3,23 3,44 1,79 2,39 1,79 2,30 2,43 2,23 2,01 1,98
CTDI100, free air 8,60 13,9 14,8 7,72 10,3 7,72 9,90 10,5 9,62 8,64 8,53
Collimation [mm] 2x1 6 x 0.6 8 x 0.6 1x5 10 x 0.6 1 x 10 12 x 1.2 20 x 0.6 32 x 0.6 32 x 1.2 64 x 0.6
Acquisition [mm] 16 x 0.6 20 x 0.6 40 x 0.6 32 x 0.6 mm, 128 x 0.6
64 x 0.6 mm
Filter (z-sharp) (z-sharp) (z-sharp) (z-sharp) (z-sharp)
CTDI100, free air 4,12 6,68 7,05 3,70 4,89 3,70 4,75 4,98 4,59 4,14 4,06
CTDI100, free air 6,29 10,2 10,8 5,64 7,48 5,64 7,24 7,62 7,01 6,32 6,20
CTDI100, free air 8,87 14,4 15,2 7,96 10,5 7,96 10,2 10,7 9,9 8,91 8,74
CTDI100, free air 11,8 19,0 20,1 10,6 14,0 10,6 13,5 14,2 13,1 11,8 11,6
Dosimetry and imaging performance report
CTDI100, free air 15,3 24,7 26,2 13,7 18,2 13,7 17,6 18,5 17,0 15,3 15,1
17
Collimation [mm] 2x1 6 x 0.6 8 x 0.6 1x5 10 x 0.6 1 x 10 12 x 1.2 20 x 0.6 32 x 0.6 32 x 1.2 64 x 0.6
18
1
Acquisition [mm] 16 x 0.6 20 x 0.6 40 x 0.6 32 x 0.6 mm, 128 x 0.6
64 x 0.6 mm
Filter (z-sharp) (z-sharp) (z-sharp) (z-sharp) (z-sharp)
CTDI100, free air 18,9 30,6 32,4 16,9 22,5 16,9 21,7 22,9 21,1 19,0 18,6
CTDI100, free air 22,9 37,0 39,2 20,5 27,2 20,5 26,3 27,7 25,5 23,0 22,5
CTDI100, free air 27,4 44,3 46,9 24,5 32,5 24,5 31,5 33,2 30,5 27,5 27,0
CTDI100, free air 1,92 3,11 3,31 1,72 2,30 1,72 2,21 2,34 2,15 1,93 1,91
CTDI100, free air 8,34 13,5 14,4 7,48 10,0 7,48 9,60 10,2 9,32 8,38 8,27
128 × 0.6
(typical)
16 × 0.6
40 × 0.6
12 × 1.2
32 × 0.6
64 × 0.6
32 × 1.2
16 x 0.3
20 x 0.6
6x 0.6
1 × 10
2×1
1x5
Total collimation [mm] 2.0 mm 3.6 mm 4.8 mm 5.0 mm 6.0 mm 10.0 mm 12.0 mm 14.4 mm 19.2 mm 19.2 mm 38.4 mm 38.4 mm
70 kV 6,6 10,7 11,2 5,9 7,8 5,9 7,9 7,6 7,3 7,4 6,6 6,5
80 kV 10,0 16,2 17,1 9,0 11,8 9,0 12,1 11,5 11,0 11,2 10,0 9,8
90 kV 14,0 22,6 23,9 12,5 16,6 12,5 16,9 16,1 15,4 15,8 14,0 13,8
100 kV 18,4 29,9 31,6 16,5 21,9 16,5 22,3 21,2 20,3 20,8 18,5 18,2
110 kV 23,9 38,7 40,9 21,4 28,4 21,4 28,9 27,5 26,3 26,9 24,0 23,5
120 kV (typical) 29,4 47,6 50,5 26,4 35,0 26,4 35,7 33,9 32,4 33,2 29,6 29,0
130 kV 35,6 57,6 61,0 31,9 42,3 31,9 43,1 41,0 39,2 40,2 35,7 35,1
Shaped filter:
140 kV 42,4 68,7 72,8 38,1 50,5 38,1 51,4 48,8 46,7 47,9 42,6 41,9
standard
Sn 100 kV 2,9 4,8 5,1 2,6 3,5 2,6 3,6 3,4 3,2 3,3 2,9 2,9
Sn 140 kV 12,6 20,5 21,8 11,3 15,1 11,3 15,4 14,6 13,9 14,4 12,7 12,5
70 kV 6,1 9,8 10,4 5,4 7,2 5,4 7,3 7,0 6,7 6,8 6,1 6,0
80 kV 9,2 15,0 15,9 8,3 11,0 8,3 11,2 10,6 10,2 10,4 9,3 9,1
90 kV 13,0 21,1 22,3 11,7 15,5 11,7 15,8 15,0 14,4 14,7 13,1 12,8
100 kV 17,3 28,0 29,6 15,5 20,5 15,5 20,9 19,9 19,0 19,5 17,4 17,0
110 kV 22,5 36,4 38,5 20,2 26,7 20,2 27,2 25,8 24,7 25,3 22,6 22,1
120 kV 27,7 44,9 47,7 24,9 33,0 24,9 33,7 31,9 30,6 31,4 27,9 27,4
130 kV 33,6 54,4 57,6 30,1 40,0 30,1 40,7 38,7 37,0 38,0 33,8 33,1
Shaped filter:
140 kV 40,2 65,1 69,0 36,1 47,8 36,1 48,7 46,3 44,3 45,4 40,4 39,7
Sn 100 kV 2,8 4,6 4,9 2,5 3,4 2,5 3,4 3,3 3,1 3,2 2,8 2,8
narrow
Sn 140 kV 12,3 19,8 21,1 11,0 14,7 11,0 14,9 14,1 13,5 13,9 12,3 12,2
CTDIfree air in mGy for typical body mode (bold type) with variation of collimation, kV, and
filter setting
SOMATOM Drive 19
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
SOMATOM Drive 21
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
By using the default Siemens scan protocols for patients with a standard patient
size, without changing the default settings of the scan parameters, and without
repeating the scans, the accumulated peripheral CTDI100 will be kept reasonably
below 1 Gy.
The following list gives examples of situations that may lead to an accumulated
peripheral CTDI100 of 1.0 Gy and above (numbers are approximations). The list
refers to default Siemens scan protocols and concentrates on scan modes with
relatively high radiation exposure. This list is not exhaustive.
◾ Use of perfusion protocols with changes in kV, scan times, or mAs, for
example:
– Neuro PCT with tube voltage changed from 80 kV to 120 kV (≈ 1.2 Gy)
– Neuro PCT with mAs increased from 200 mAs to ≈ 580 mAs
◾ Use of interventional protocols, for example:
– Head Intervention i-Fluoro with an accumulated scan time of ≈ 21 s
– Head Intervention i-Fluoro with 10 s scan time and mAs increased from
100 mAs to ≈ 220 mAs
– Head Intervention i-Fluoro with 15 s scan time and tube voltage increased
from 120 kV to 140 kV
◾ Scanning of sequence protocols without table feed, for example:
– Approximately 16 scans of Head Neuro Seq without table feed
◾ Repeating standard sequences or spiral scans within an examination, for
example:
– Repeated application of the default protocol Head Neuro Seq
(approximately 16 times)
– Repeated application of the default protocol Abdomen Seq (approximately
93 times)
– Repeated application of the default protocol Abdomen Routine
(approximately 89 times)
for Dose Notification and Dose Alert according NEMA XR-25 and
IEC 60601-2-44 (Ed. 3.1). The default threshold value for Dose Alert is an
accumulated CTDIvol of 1.0 Gy for adult patients and 1.0 Gy for pediatric
patients.
Ratio of peripheral CTDI100 to In general, the accumulated peripheral CTDI100 for a scan can be derived from
displayed CTDIvol the displayed CTDIvol by multiplying the CTDIvol with a given factor that depends
on the tube voltage and shaped filter used:
Shaped Filter: Narrow Standard
CTDI phantom size: Ø 16 cm Ø 32 cm Ø 16 cm Ø 32 cm
70 kV 1.0 1.2 1.0 1.2
80 kV 1.0 1.2 1.0 1.2
90 KV 1.0 1.2 1.0 1.2
100 kV 1.0 1.2 1.0 1.2
110 kV 1.0 1.1 1.0 1.2
120 kV 1.0 1.1 1.0 1.2
130 kV 1.0 1.1 1.0 1.2
Tube Voltage
21 CFR 1020.33
The actual exposure values, such as CTDI100, CTDIw, CTDIvol and DLP, may deviate
from the values displayed at the scanner and may deviate from the values stated
in this manual within the following tolerances:
SOMATOM Drive 23
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
The table below provides conversion factors to calculate the CTDIvol and DLP for
the 16 cm phantom from the values of a 32 cm phantom. The factors are stated
separately for each setting of shaped filter. Please note that for adult neck
protocols the standard shaped filter is used and for pediatric body and pediatric
neck protocols the narrow shaped filter is used.
70 kV 2.5 2.8
80 kV 2.4 2.6
90 kV 2.3 2.5
Example
For a typical pediatric body protocol at 100 kV, the displayed CTDIvol (related to
the Ø 32 cm CTDI-phantom) is 3.00 mGy × cm; see ( Page 8 Typical CT
conditions of operation). For the same protocol, the CTDIvol of a Ø 16 cm CTDI-
phantom is 2.5 times higher than the CTDIvol of a Ø 32 cm CTDI-phantom,
resulting in 7.50 mGy.
Phantom size Ø 16 cm Ø 32 cm Ø 32 cm
Phantom size Ø 16 cm Ø 32 cm Ø 32 cm
y Dose
z z-axis
(1) Total dose profile width
(2) Dose profile
(3) Detector
(4) Acquisition range
Dose efficiency values for the various Acquisition Total collimation Dose efficiency
collimation types
2 × 1 mm 2.0 mm 79%
1 × 5 mm 5.0 mm 84%
SOMATOM Drive 25
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
1 × 10 mm 10.0 mm 92%
For CT scans with a dose efficiency lower than 70%, a message containing
information about the dose efficiency is displayed after a scan range is loaded.
The user has to confirm the message before starting the scan.
21 CFR 1020.33
Dose profiles for single axial scans have been measured for the narrowest and
widest collimations, as well as for a mid-range collimation. For each of these
collimations, the dose profiles have been measured without any phantom (free
air), at the center of the Ø 16 cm CTDI phantom (head) and at the center of the
Ø 32 cm CTDI phantom (body).
The phantoms were centered in the isocenter and aligned with the scanner
axes. For the measurements, a semiconductor diode sensor was moved through
the phantom.
Dose profile diagrams
In the diagrams below, dose profiles (thick lines) are presented as a percentage
of their maximum value.
Sensitivity profiles for the detector slices used in these collimations are plotted
as thin lines (sum of single-slice sensitivity profiles). The dotted lines indicate
the nominal collimation.
Nominal values and tolerances for acceptance testing
The stated tolerances are determined by the accuracy of the measurement
method.
For acceptance testing, dose profile widths are measured using film
(GAFCHROMIC XR-QA / XR-QA2) and without any phantom. The tube is fixed
and positioned at 12 o'clock. The exposure parameters are chosen to
overexpose the film (120 kV, 160 mA, 10 s). Using this method, no
densitometry is needed to evaluate the film. The width of the blackened range
is a measure of the base width of the profile. Nominal values and tolerances are
defined accordingly.
2 × 1 mm
12 × 1.2 mm
64 × 0.6 mm
Dose profiles
12 × 1.2 mm 18 mm 22 mm 55 mm ± 4.0 mm
64 × 0.6 mm 43 mm 45 mm 88 mm ± 4.0 mm
2 × 1 mm 3.5 mm ± 1.0 mm
SOMATOM Drive 27
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
Beam quality
Minimum filtration, permanent in the useful Quality equivalent filtration ≥ 6.8 mm Al at CARE filter
beam (tube housing assembly) 145 kV
Tube A and tube B
Minimum filtration
◾ X-CARE: Control of the angular tube current to reduce dose for an angular
segment with radiation sensitive organs, such as eye lenses and breast tissue
◾ CARE kV: Optimization of kV to minimize patient dose but maintaining image
quality
The modulation type applied depends on the protocol selected and the
individual setting:
Abdomen XYZ exposure control The tube current is adapted to the patient size.
It is varied along the z-axis according to the
AngioBody (CARE Dose 4D)
patient’s attenuation profile and modulated
Pediatric Head angularily according to the patient's angular
attenuation profile that has been measured
Pediatric Angio Head online.
Neck
Pelvis
Runoff
Shoulder
Spine
Thorax
Adult Head Z exposure control The tube current is adapted to the patient size
and varied along the z-axis according to the
Adult Angio Head (CARE Dose 4D)
patient’s attenuation profile.
Respiratory
X-CARE protocols Angular modulated exposure control The tube current is modulated angularly rela-
tively to the patient's orientation to reduce
(X-CARE)
dose in certain body ranges (eye lenses, breast
(possibly combined with fixed exposure con- tissue).
trol, Z exposure control or XY exposure con-
trol)
CARE Dose 4D CARE Dose 4D automatically adapts the tube current to the patient's body size
and shape.
Using the patient's topogram, Care Dose 4D evaluates two profiles of the
patient's body size in the a.p. and lateral directions, related to the X-ray
attenuation of the patient's body.
Based on these profiles, the mAs value is adapted to the patient during the
subsequent CT scans. The adaptation follows a curve, which determines the
correlation between patient size and tube current. The adaptation curve has
been derived from clinical optimization for constant diagnostic image quality.
SOMATOM Drive 29
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
Adaptation of mAs to patient size for standard protocols (adult and child)
Depending on the modulation type of the protocol, the patient size adapted
tube current can be fixed, modulated along the z-axis according to the
adaptation curve, modulated according to the patient's online measured
angular attenuation profile, or a combination of these.
For certain organ characteristics (obese, runoff, neck), special adaptation curves
are applied.
SOMATOM Drive 31
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
SOMATOM Drive 33
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
ECG Gating The tube current is synchronized to the patient's ECG signal and gated
depending on the user-defined setting for the image time window, which is set
as a percentage range of the RR interval (e.g., 70% to 70%). Depending on the
data range needed for the particular image reconstruction, the width of the
tube current gating varies between a minimum of a 260° rotation angle and a
maximum of full rotation. Outside the tube current gating window, the tube
current is reduced to 20% (or 4% with MinDose) of the value inside the gating
window.
X-CARE The tube current is reduced over an angle segment of 80°, with respect to the
anterior position of the patient, to reduce the organ dose for radiation-sensitive
peripheral organs such as eye lenses or female breast tissue. Depending on the
system's power limits, the tube current reduction is by up to 80%. The reduction
is compensated for by an increased tube current over the rest of the rotation,
keeping the average CTDIvol constant.
CARE kV The kV is optimized to minimize the CTDIvol but maintaining image quality
(contrast-to-noise ratio).
CARE kV evaluates the body diameter using the patient's topogram. Based on
the user-defined image quality and tissue of interest, a kV setting is determined
that minimizes the CTDIvol, keeping the contrast-to-noise ratio constant.
SOMATOM Drive 35
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
◾ Halliburton SS, Abbara S, Chen MY, Gentry R, Mahesh M, Raff GL, Shaw LJ,
Hausleiter J: SCCT guidelines on radiation dose and dose-optimization
strategies in cardiovascular CT. J Cardiovasc Comput Tomogr 2011, 5(4):
198-224.
◾ AAPM Position Statement on the Use of Bismuth Shielding for the Purpose of
Dose Reduction in CT scanning (http://www.aapm.org/publicgeneral/
BismuthShielding.pdf)
CTDIvol (32 cm) for low-contrast Low-contrast detectability in a 20 cm CATPhan CTP 515 for a reconstructed
detectability slicewidth of 10 mm:
5 mm / 3 HU 8 mGy
4 mm / 3 HU 10 mGy
3 mm / 3 HU 13 mGy
1.2.2 CT number
The CT number is specified for air (-1000 ± 4 HU) and water (0 ± 4 HU) only.
The specification is not valid for other materials in the beam. The specified CT
number of air according to the YY 0310 testing method is -1000 ± 10 HU. The
CT number for water is valid for a cylindrical phantom with a diameter ranging
from 20 cm to 30 cm which is centered and aligned with the scanner axis for
both, typical head modes and typical body modes.
The CT numbers for other materials depend on the kV settings and the shaped
filter.
1.2.3 Uniformity
The uniformity of the CT values is specified for a typical scan mode within a
cylindrical 20 cm water phantom that has been centered in the scan plane
without any other objects in the scan field.
21 CFR 1020.33
Image noise
21 CFR 1020.33
SOMATOM Drive 37
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
MTF Typical Head (H30s) Typical Body (B30f) Sharpest without UHR Sharpest with UHR
(H70h) (U95u)
50% 3.5 LP/cm ± 10% 3.5 LP/cm ± 10% 11.4 LP/cm ± 10% 18.7 LP/cm ± 10%
10% 5.9 LP/cm ± 10% 5.9 LP/cm ± 10% 14.0 LP/cm ± 10% 23.3 LP/cm ± 10%
x z [mm]
y SSP [%]
x z [mm]
y SSP [%]
SOMATOM Drive 39
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
x z [mm]
y SSP [%]
21 CFR 1020.33
10 .0
0.75
14.4
15.0
20.0
0.4
0.5
0.6
1.0
1.2
1.5
2.0
2.4
3.0
4.0
4.8
5.0
6.0
7.0
7.2
8.0
Routine Spiral 128 × 0.6 x1) x x x x x x x x x x x x
32 × 1.2 x x x x x x x x x
Sequence 128 × 0.6 x1) x x x x x x x x x x x x
32 × 1.2 x x x x x x x x
1 × 10 x
1×5 x
Flash Sprial 128 × 0.6 x1) x x x x x x x x x x x x
Neuro Spiral 40 × 0.6 x1) x x x x x x x x x x x x
20 × 0.6 x1) x x x x x x x x x x x x
Sequence 12 × 1.2 x x x x x
1 × 10 x
1×5 x
Obese Spiral 32 × 0.6 x1) x x x x x x x x x x x x
Thorax HR Sequence 2×1 x x
UHR Spiral 16 × 0.6 x1) x x x x x x x x x
Sequence 16 × 0.6 x1) x x x x x x x x x x x x
z UHR Spiral 16 × 0.3 x x1) x x x x x x x x x
Chest Pain Spiral 128 × 0.6 x1) x x x x x x x x
32 × 1.2 x x x x x
Cardio Spiral 128 × 0.6 x1) x x x x x x x x
32 × 1.2 x x x x x
Sequence 128 × 0.6 x1) x x x x x x x x
32 × 1.2 x x x x
Cardiac Shuttle Sequence 128 × 0.6 x x x x x x x x x x x x x x
32 × 1.2 x x x x x x x x x x
Respiratory Spiral 128 × 0.6 x1) x x x x x x x x
32 × 1.2 x x x x x
Sequence 128 × 0.6 x1) x x x x x x x x x x x x
32 × 1.2 x x x x x x x x
i-Fluoro Multiscan 12 × 1.2 x x x
1 × 10 x
1×5 x
i-Sequence Sequence 32 × 1.2 x x x x x x
12 × 1.2 x x x x
1 × 10 x
1×5 x
i-spiral Spiral 32 × 1.2 x x x x x x x x x
Perfusion Multiscan 128 × 0.6 x x x x x x x x x x x x x x
32 × 1.2 x x x x x x x x x x
1 × 10 x
Adaptive 4D Spi- 128 × 0.6 x x x x x x x x x x x x
ral
32 × 1.2 x x x x x x x x x
1) With Stellar detector
SOMATOM Drive 41
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
10 .0
0.75
14.4
15.0
20.0
0.4
0.5
0.6
1.0
1.2
1.5
2.0
2.4
3.0
4.0
4.8
5.0
6.0
7.0
7.2
8.0
Dual Energy Spiral 128 × 0.6 x1) x x x x x x x x x x x x
64 × 0.6 x x x x x x x x x x x x
40 × 0.6 x1) x x x x x x x x x x x x
32 × 0.6 x1) x x x x x x x x x x x x
Sequence 128 × 0.6 x1) x x x x x x x x
32 × 0.6 x1) x x x x x x x x
64 × 0.6 x1) x x x x x x x x
Dual Energy Car- Spiral 128 × 0.6 x1) x x x x x x x x
dio
64 × 0.6 x1) x x x x x x x x
Sequence 128 × 0.6 x1) x x x x x x x x
64 × 0.6 x1) x x x x x x x x
1) With Stellar detector
According to:
21 CFR 1020.33
> 2 mm Tolerance: ± 1 mm
The upper arrow indicates the specified tolerance around the nominal value
(black dot) specified for the scanner type concerned. This tolerance is valid for
the initial constancy measurement (Constancy Reference) where the baseline
values for the individual system (black circle) are established and stored for
comparison with the succeeding constancy tests. The lower arrow indicates the
specific tolerance around the baseline value established in the Constancy
Reference measurement. This tolerance is valid for the periodical constancy
measurement (Constancy Normal). It indicates the allowed deviation from the
baseline value cut to the additional valid tolerances for the nominal value
indicated by the upper arrow.
New reference values must be recorded in the following situations:
◾ When a new CT scanner is brought into use
◾ After installation of a new software version, the reference test is required by
the manufacturer
◾ After changing any component relevant to image quality and dose (exchange
of tube, tube collimator or detector) or its accessories (detector module)
◾ Whenever changes in test equipment may cause a variation
After installation or relocation of the CT scanner or after major modifications
(other than normal maintenance or repair), a repetition of partial acceptance
testing according to IEC 61223-3-5 may be necessary.
The test tool offers a routine test plan according to the requirements of the
standard. In addition, constancy tests must be done in the following situations:
◾ Whenever malfunction is expected
◾ Immediately after the CT scanner has undergone maintenance that could
affect the performance parameter under test
◾ Whenever the constancy test fails, in order to confirm the test result
Reports for the quality tests with measured values and their tolerance range can
be extracted as pdf files for the configured language.
SOMATOM Drive 43
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
Test Content
Slice thickness The tomographic section thickness is evaluated by measuring the width of the
image of an inclined aluminium ramp at the intersection of the ramp with the
slice. The width is defined as the Full Width at Half-Maximum (FWHM) of the
baseline corrected CT-value profile. The influence of the reconstruction kernel
and the influence of the cone angle on the measured width are taken into
account. The measurement is performed for two sets of scan parameters, repre-
senting a typical head mode and a typical body mode. The evaluation is per-
formed for at least both outer and one central slice. The measurement and eval-
uation is performed separately for system A and B.
Homogeneity and water value The measurement is performed for three sets of scan parameters, representing
a typical head mode, a typical body mode and a body dual-source mode. For the
dual-source mode the evaluation is performed separately for system A and B.
The water value and homogeneity is evaluated by measuring the mean CT num-
ber in five Regions Of Interest (ROI) in the image of a cylindrical water phantom
with diameter 20 cm. The ROI are located in the center and near the edge of the
water phantom at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock. The water value
is the CT number of the central ROI, the homogeneity is the difference between
the outer ROI and the central ROI each.
Image noise The measurement is performed for three sets of scan parameters, representing
a typical head mode, a typical body mode and a body dual-source mode. For the
dual-source mode the evaluation is performed separately for system A and B.
The image noise is evaluated by measuring the standard deviation of CT num-
bers in a Region Of Interest (ROI) in the image of a cylindrical water phantom
with a diameter of 20 cm. The ROI is located in the phantom center. To suppress
any influences of image inhomogeneity or ring artifacts, the evaluation is per-
formed for the difference image of two succeeding scans.
Contrast scale Contrast scale is evaluated from the mean CT numbers in a central Region Of
Interest (ROI), measured from a scan of a cylindrical water phantom with a
20 cm diameter and a scan without a phantom. Attenuation coefficients of
0 cm-1 for air and 0.192 cm-1 for water are taken for the evaluation. Evaluation
is performed for two scans, representing a typical body-mode and a dual-
source-mode.
Low-contrast resolution The low-contrast resolution is tested for the A-system by visual asssessment of
low contrast rods of 5 mm diameter embedded in a plastic disc providing a con-
trast of approximately 6 HU.
An image with a 10 mm slice thickness, taken with a CTDIfreeAir of approximately
50 mGy, is evaluated.
*) CTDIair and CTDIw are alternative methods for the mandatory dose constancy test. Siemens offers CTDIair by
default; CTDIw can be configured.
Test Content
High-Contrast-Resolution (MTF) The high-contrast resolution is determined by evaluating the image of a thin
tungsten wire placed in a plastic rod, centered in the scan plane and aligned
along the scanner axis. The point spread function is obtained from the images
of the wire and the Modulation Transfer Function (MTF) is calculated as the
Fourier transformation of the point spread function. The 50% and 10% values of
the MTF are evaluated.
The measurement is performed for four to five sets of scan parameters, repre-
senting a typical head mode, a typical body mode, a dual-source mode, the
sharpest mode without UHR comb and, if available, the sharpest mode with
UHR comb. For the dual-source mode the evaluation is performed separately for
system A and B.
CTDIw* The CTDIw is evaluated from the measured Dose-Length-Product (over 10 cm)
for three sets of scan parameters, representing a typical head mode, a typical
body mode and a dual-source mode. For the head mode, a 16 cm PMMA phan-
tom is used and for the body mode and dual-source mode, a 32 cm CTDI phan-
tom is used. The dose is measured in the central and upper borehole only, as
the differences of the four peripheral positions are known as small compared to
the tolerances.
The dual-source-mode is like the typical body mode, but with both tubes contri-
buting half of the dose each.
CTDIfree air* The CTDIfree air is evaluated from the measured Dose-Length-Product (over
10 cm) in a typical head mode, a typical body-mode (A-System) and a scan with
the typical body mode parameter in the B-system.
Tube voltage The tube voltage is measured with an internal voltage divider for scans with all
nominal tube voltages. The test is performed separately for system A and B.
Light marker For the test of the light markers, the system phantom is centered and aligned,
so that the lightmarkers hit the engraved notches on the phantom. From the
image of the phantom, scanned with narrow collimation, the congruence of the
engraved notches with the scan plane, as well as the saggital and coronal
planes, are tested.
Topogram position In the topogram image of the system phantom, the position of an tomogram
scan with narrow collimation is selected for the engraved notches in the phan-
tom. From the recorded tomogram image of the phantom, the congruence of
the engraved notches with the scan plane is inspected.
Table position The patient table is automatically moved forward and backward over 30 cm in
continuous movement and stepwise movement. After each series of move-
ments, the distance is measured with a ruler.
*) CTDIair and CTDIw are alternative methods for the mandatory dose constancy test. Siemens offers CTDIair by
default; CTDIw can be configured.
SOMATOM Drive 45
Print No. C2-047.660.01.02.02
1 Dosimetry and imaging performance report
SOMATOM Drive 47
Print No. C2-047.660.01.02.02
This page has been intentionally left blank.
SOMATOM Drive 49
Print No. C2-047.660.01.02.02
This page has been intentionally left blank.
Manufacturer’s note:
The CE marking applies only to Medical Devices which have been put on
the market according to the above-mentioned EC Directives.
www.siemens.com/healthcare