Sunteți pe pagina 1din 7

Lawrence Berkeley National Laboratory

Lawrence Berkeley National Laboratory

Title
Conceptual design of a compact positron tomograph for prostate imaging

Permalink
https://escholarship.org/uc/item/39w571wf

Authors
Huber, J.S.
Derenzo, S.E.
Qi, J.
et al.

Publication Date
2000-11-04

Peer reviewed

eScholarship.org Powered by the California Digital Library


University of California
1
Accepted to IEEE Transactions on Nuclear Science LBNL-47068

Conceptual Design of a Compact Positron


Tomograph for Prostate Imaging
J. S. Huber, Member, IEEE, S. E. Derenzo, Fellow, IEEE, J. Qi, Member, IEEE, W. W. Moses,
Senior Member, IEEE, R. H. Huesman, Senior Member, IEEE, and T. F. Budinger, Member, IEEE

palpable. Treatment decision is based mainly on biopsy


Abstract—We present a conceptual design of a compact confirmation of prostate cancer, but the diagnostic accuracy of
positron tomograph for prostate imaging using a pair of prostate biopsies is problematic. Treatment usually consists
external curved detector banks, one placed above and one
below the patient. The lower detector bank is fixed below the
of radical prostatectomy, external beam irradiation,
patient bed, and the top bank adjusts vertically for maximum brachytherapy (interstitial implantation of radioactive seeds),
sensitivity and patient access. Each bank is composed of 4 0 androgen ablation (hormone) therapy, or "watchful waiting".
conventional block detectors, forming two arcs (44 cm minor, In some cases, radiation therapy is prescribed to the pelvis
60 cm major axis) that are tilted to minimize attenuation and and/or prostate bed even without confirmation by biopsy or
positioned as close as possible to the patient to improve
sensitivity. The individual detectors are angled to point
imaging [5-7], despite the fact that irradiating a wide field in
towards the prostate to minimize resolution degradation in the pelvis (when disease is only suspected to extend beyond
that region. Inter-plane septa extend 5 cm beyond the the known cancerous prostate) can lead to a higher incidence
scintillator crystals to reduce random and scatter of post-irradiation morbidity than with irradiation of the
backgrounds. A patient is not fully encircled by detector rings prostate bed alone [8]. A major problem with prostatic cancer
in order to minimize cost, causing incomplete sampling due to
the side gaps. Monte Carlo simulation (including randoms and
therapy is the question of when to treat or whether to treat at
scatter) demonstrates the feasibility of detecting and all. This is particularly problematic in the case of an
differentiating partial and whole prostate tumors with a tumor increased PSA level after a prostatectomy. A new imaging
to background ratio of 2:1, utilizing the number of events that technology for sensitive detection of early stage prostate
should be achievable with a 6-minute scan after a 10 mCi cancer is needed to confirm PSA and help guide treatment
injection (e.g., carbon-11 choline or fluorine-18 fluorocholine).
decisions. In addition, a new method is needed to assess
response shortly after treatment intervention.
INTRODUCTION
I.
In order to help meet these needs, we are designing a
E present a conceptual design of a compact positron
W tomograph optimized to image the prostate. This
instrument images radiopharmaceuticals that specifically
compact PET camera optimized to image the prostate.
Functional PET imaging will help detect malignant tumors
in the prostate and/or prostate bed to confirm an increased
localize in the prostate to confirm the presence, absence or
PSA level, as well as possibly help determine tumor
progression of disease. It will have approximately 4 times
less detectors than a conventional whole-body positron "aggressiveness" based on metabolic uptake levels, in order
emission tomograph (PET), which will reduce the cost and to help guide whether to treat suspected prostatic cancer.
increase clinical availability. Although not optimized to detect distant metastatic disease,
Prostate cancer has a prevalence and diagnostic rate similar this compact PET camera should also image local spread
to breast cancer, with 360,000 new cases diagnosed each year beyond the prostate bed to help guide treatment decisions
and 2 million men affected by the disease. Four percent of such as whether a narrow or wide irradiation treatment field is
men over 50 will have clinically significant disease [1]. needed.
Currently, prostate cancer suspicion is typically based on an Many groups have shown that PET imaging with
18
elevated prostate-specific antigen (PSA) level or a suspicious [ F]fluorodeoxyglucose (FDG) is not a good technique for
node found during a digital rectal exam. Serum PSA values prostate cancer diagnosis, because FDG is not very prostate
do not always correlate well with clinical diagnosis or specific (with a standardized uptake value of typically ~2)
outcomes [2-4]. Palpation is subjective, insensitive and primarily due to bladder accumulation. However, more
inexact; more than half of all cancers detected today are not promising PET radiopharmaceuticals have recently
demonstrated outstanding results in the sensitive detection of
Manuscript received November 3, 2000. This work was supported by prostate cancer, inspiring a new interest in using PET for
the Director, Office of Energy Research, Office of Health and
Environmental Research, Medical Applications and Biophysical Research prostate cancer imaging. Consistent with the evidence of
Division of the U.S. Department of Energy under contract No. DE-AC03- increased pool size of choline in prostate cancer [9, 10], Hara
76SF00098. and co-workers demonstrated prostate tumor uptake by
J. S. Huber, S. E. Derenzo, J. Qi, W. W. Moses, R. H. Huesman and T.
F. Budinger are with the Lawrence Berkeley National Laboratory, Mailstop [11C]choline. They find that: [11C]choline clears the blood
55-121, 1 Cyclotron Road, Berkeley, CA 94720 USA (telephone: 510-486- faster than FDG; its uptake in prostate tumors is significantly
6445, e-mail: jshuber@lbl.gov). higher than in normal and surrounding tissues [11, 12],
2
Accepted to IEEE Transactions on Nuclear Science LBNL-47068
(a)

Prostate

(b)
Upper Detector “Ring”
Adjustment Direction

Fig. 2: Proposed positron tomograph for prostate imaging (shielding not


shown). (a) Drawing of a transaxial view through prostate, showing the
Fig. 1: [11C]choline image of prostate cancer (a) before and (b) after patient centered between two detector banks. The individual detector
treatment. These grayscale images indicate a high (white) uptake in the modules are angled to point towards the prostate. (b) Drawing of the
prostate cancer compared with a low (gray) uptake elsewhere. Images are sagittal view. The bottom arc is fixed below the patient bed, whereas the
provided by Hara and co-workers [13]. top arc adjusts vertically for patient access and compactness. Both detector
banks are tilted and positioned as close as possible to the prostate, which
improves sensitivity and minimizes attenuation.
providing excellent tumor/normal contrast; and bladder
accumulation is minimal if the correct time course is chosen
[13] which is a major advantage over FDG. Additionally, II. CAMERA DESIGN
they observe that [11C]choline PET is more sensitive for
detecting bone metastases than bone scintigraphy. Therefore, These new prostate tracers motivate us to design a low
[11C]choline is an attractive PET tracer for imaging primary cost PET camera optimized to image the prostate.
and metastatic tumors of the prostate and potentially for other Coincidence imaging of positron emitters is achieved using a
regions of the body [14-18]. Figure 1 shows a [11C]choline pair of external curved detector banks, one placed above and
image of prostate cancer before and after therapy, one below the patient. Fig. 2 shows the transaxial and
demonstrating the ability to detect prostate carcinoma using sagittal views of the proposed camera design. The bottom
choline and possibly its analogs. Other
11
C bank is fixed below the patient bed, and the top bank moves
radiopharmaceuticals are also under investigation for prostate upward for patient access and downward for maximum
cancer PET imaging, including [11C]acetate and sensitivity. Each bank consists of 40 conventional block
11
[ C]methionine. detectors such as those in the Siemens/CTI ECAT HR+ (8 x
18
F imaging has the advantage of a longer half-life, 8 array of 4.5 mm x 4.5 mm x 30 mm BGO crystals),
increasing commercial viability since an on-site cyclotron forming two arcs with a minor axis of 44 cm and major axis
facility would not be necessary (as it would for 11C imaging). of 60 cm. The proposed prostate camera has one-half the axial
There are several 18F radiopharmaceuticals currently under coverage and uses about one-fourth the number of detectors as
investigation for prostate cancer imaging, including in a conventional PET system. However, since the average
[18F]fluorocholine (FCH) [19, 20]. PET images using distance to the detectors is approximately one-half that of a
[18F]fluorocholine demonstrate high standardized uptake conventional 2D PET system, the solid angle for a central
values (e.g., SUV of 8), indicating that FCH is well source is approximately double.
localized in the prostate cancer and can be imaged with good Individual detector modules are angled to point towards
resolution if a short scan time (~5 minutes) is used. the camera center near the prostate location in order to reduce
Therefore, [18F]fluorocholine is also an attractive PET tracer penetration effects (in the detector) for annihilation photons
for imaging primary and metastatic tumors of the prostate. originating in the prostate, which is an unique feature for a
non-circular camera geometry. Annihilation photons from
other parts of the field of view (FOV) will suffer increased
3
Accepted to IEEE Transactions on Nuclear Science LBNL-47068
Inter-plane −150

Septa −100

−50

50
Lead Shielding
Fig. 3: Cut-out side view of the proposed positron tomograph, showing the 100
shielding and inter-plane septa.
150
−200 −100 0 100 200
penetration effects, but these FOV regions are less important.
Both detector banks are tilted to image the prostate while Fig. 5: Statistical noise-free reconstructed image of five point sources,
demonstrating the local impulse response function and sampling
minimizing attenuation (i.e., above the buttocks and below
sufficiency. Scale in millimeters.
the stomach), but the gantry can also allow zero tilt for thin
patients. The patient is not fully encircled in 2D, which will coincidences and from photons that Compton scatter in the
be discussed further in Section III. patient. Lead shields are also used on the ends to reduce
Standard block detectors with thick gamma ray detectors activity from outside the field of view.
(i.e., 3 attenuation lengths) are needed for good detection
efficiency. Narrow detector elements (i.e., 4.5 x 4.5 x 30 mm III. EXPECTED PERFORMANCE
BGO crystals) are desired to achieve good spatial resolution.
The 511 keV photons of primary interest originate from the A. Simulation
prostate which is roughly a small (e.g., 3 cm diameter) Our camera geometry provides a compact design (i.e., 44
spherical source with a known central location at a reasonable cm minor axis between opposing detector faces) that adjusts
distance from detector modules, so expensive detector vertically to fit the patient. However, a patient of average size
modules with depth of interaction capability are not will not be fully encircled by the detector rings, as mentioned
necessary. A two ring camera is more practical for patient in Section II. Fig. 4 shows the resulting irregular and
positioning. Therefore, our proposed camera design attempts incomplete sampling due to the side gaps. Similar irregular
to maximize performance while minimizing cost, in order to and incomplete sampling patterns have been dealt with in our
produce a PET camera with nearly twice the detection positron emission mammography camera [21]. Despite this
efficiency as a conventional 2D PET camera with incomplete sampling, we are able to reconstruct nearly artifact
approximately one-quarter the detector costs. free images in the region of interest by using an iterative
The camera design also includes shielding, as shown in reconstruction as described below. This iterative
Fig. 3. Inter-plane septa extend 5 cm beyond the scintillator reconstruction algorithm can easily deal with the missing
crystals to reduce background events from random data of the system and can also partially recover the
100 resolution loss (due to the crystal penetration effect) in the
80 off-center FOV region. The gap size will increase for larger
patients and this may impact the image quality, but we have
60
not yet explored the effects of variations in gap size.
40 Our camera design includes inter-plane septa, so the data
will be acquired in 2D. We use Monte Carlo data to simulate
Angle (Degree)

20
the camera performance. Reconstructions are performed with
0 an iterative maximum likelihood algorithm [22]. Self-
−20 attenuation is simulated assuming a constant attenuation
coefficient of 0.0095/mm inside the body, then attenuation is
−40
modeled in the reconstruction. Coincidences between detector
−60 modules within the same bank are allowed.
−80
Fig. 5 shows an image of five point sources that are
reconstructed without statistical noise to demonstrate how
−100 irregular sampling affects the local response function. All
−300 −200 −100 0 100 200 300
Radial Distance (mm) point sources are clearly visible. We observe increased radial
Fig. 4: Incomplete sampling for the field of view due to side gaps between blurring near the outer surface of the patient as expected due
detector banks. Coincidences between detector modules within the same to penetration effects, but observe good resolution in the
bank and imaging plane are allowed, resulting in the ~triangular sampling
regions on the left and right. The field of view of the proposed camera is region near the center (i.e., near the prostate). In short, we are
an elliptical region, so the only missing samples are the “X” region in the optimizing resolution at the center near the prostate at the
center of the sinogram. price of resolution at the edges, which implies that distant
4
Accepted to IEEE Transactions on Nuclear Science LBNL-47068
metastases will suffer from some resolution degradation. Data shows the anticipated spatial resolution obtained in this
are generated by forward projection of 256x256 2 mm x 2 manner as a function of distance from the center of the field
mm phantom voxels to simulate real data, and reconstruction of view (along both the x- and y- axes). Notice that little
is done with 128x128 4.51 mm x 4.51 mm pixels, thus degradation due to penetration artifacts is seen for objects
causing some point source spread among adjacent pixels. within 5 cm of the camera center, and that 6 mm fwhm or
As we are planning to use iterative methods to reconstruct better resolution is obtained within 8 cm of the camera
images, the anticipated spatial resolution will depend on the center. This region is easily large enough to contain the entire
geometrical parameters used for reconstruction (e.g., the prostate.
voxel size and detector model) as well as the number of We also use Monte Carlo data to simulate the performance
events. Thus, it is difficult to derive meaningful estimates of of our proposed camera for imaging extended sources.
the expected spatial resolution from the test image in Fig. 5. Extended “prostate” sources are reconstructed with statistical
Therefore, we estimate the spatial resolution analytically with noise representing 745 kcounts using a preliminary iterative
the help of a simplifying assumption. That assumption is maximum likelihood algorithm. We assume a 2.5 cm
that if two tomographs employ the same scintillation diameter spherical “prostate” (see fig. 7(a)) with a tumor to
material, the contribution of penetration to the spatial background ratio of 2:1, a 30% scatter fraction and 20%
resolution for a source at any position within either camera randoms fraction. The images are post-filtered with a
depends only on the angle θ that the incoming gamma ray Gaussian function having a standard deviation of 1 pixel. The
makes to the axis of the detector crystals (see Fig. 6). Given spherical "prostate" is divided into quadrants. Fig. 7 (b)
this assumption, we can use measured spatial resolutions [23] shows a reconstructed image with a tumor in the upper left
to estimate the resolution for the proposed camera. Fig. 6(b) quadrant of the spherical prostate. Fig. 7 (c) shows a
reconstructed image with a tumor in the right half of the
(a) prostate. Fig 7 (d) shows a reconstructed image with a tumor
in the upper right and lower left quadrants. These "prostate"
tumors are clearly visible in all images, demonstrating that
we can reconstruct nearly artifact free images near the prostate
despite incomplete sampling and that the scanner can
θ differentiate partial and whole prostate tumors. Our
y preliminary rate calculations (described below) indicate that
x 745 kcounts per imaging plane will be achievable with a 6-
minute scan after injecting a clinically acceptable (10 mCi)
patient dose of [11C]choline or [18F]fluorocholine. A relatively
short value of six minutes is chosen to demonstrate that good
images can be obtained even if the bladder fills rapidly. All
published [11C]choline or [18F]fluorocholine studies done to
(b) date have used three minutes or longer (up to 30 minutes)
16 imaging times.
Radial Resolution (mm fwhm)

14 X-Direction (60 cm)


Y-Direction (44 cm) B. Estimated Rates
12 Circular Camera (60 cm) The average distance between detector banks in our
proposed camera is approximately one-half that of a
10
conventional PET system such as the HR, so our sensitivity
8 per slice is nearly 2 times higher than the HR (i.e., our
6
detectors will have nearly twice the rate). However, our
camera also has half the number of imaging planes, resulting
4 in a total system rate that is approximately the same as the
2 HR. Forty five minutes after injecting 15 mCi of FDG into
an average-sized patient (i.e., with ~10 mCi remaining in the
0
0 5 10 15 20 patient), we see a 50 kHz total coincidence rate using a HR
Distance from Tomograph Center (cm) scanner. Therefore, our total system coincidence rate is
expected to be approximately 50 kHz, with a detector live
Fig. 6: (a) Proposed prostate PET camera drawing shown with a source
within the camera. The angle θ is defined to be that which an incoming time of at least 80%.
gamma ray makes to the axis of the detector crystal. (b) The anticipated For our proposed camera geometry, there are 166 total
spatial resolution for the proposed prostate PET camera as a function of cross planes. Assuming a 50 kHz total coincidence rate, this
distance from the center of the field of view along both the x- (square) and
y- (circle) axes, based on measured spatial resolutions with a circular
implies a rate of 300 Hz per cross plane. Summing over 7
tomograph. Little degradation due to penetration artifacts is seen for objects cross plane slices per imaging plane, this corresponds to a 2
within 5 cm of the camera center, and 6 mm fwhm or better resolution is kHz coincidence rate per imaging plane. Therefore, 745
obtained within 8 cm of the camera center. Plot also shows spatial
resolution for the reference circular PET camera as a function of distance
kcounts per imaging plane will be achievable with a 6-minute
from the center of the FOV along the x- (triangle) axis.
5
Accepted to IEEE Transactions on Nuclear Science LBNL-47068
(torso 36 cm major axis, 17 cm minor axis), resulting in an
activity concentration of 0.15 µCi per ml. These calculated
rates are in agreement with patient data within a factor of 2.

IV. DATA CORRECTION


This design does not include hardware for obtaining data
to perform attenuation correction. The proposed design has
sufficient room to incorporate a single photon or positron
emitting source for transmission scans, but it would add to
the camera cost. However, many clinical whole-body PET
scans are done without an attenuation scan so it may not be
necessary for the detection of abnormal activity in the
prostate bed. A practical approach is to correct without
additional hardware using calculated attenuation coefficients
based on body contours (obtained from emission data) and an
uniform attenuation coefficient. Anatomical boundaries are
obtained from the outer edges of emission sinograms
acquired from transverse sections [26]. This method can be
generalized to three dimensions, and a uniform soft tissue
attenuation coefficient can be used to compute the attenuation
correction factors. This method assumes that the distribution
of attenuating medium is uniform, that its external boundary
is convex, that the attenuation coefficient for 511 keV
photons in soft tissue is uniform, and that inter-subject
variability is small. These assumptions should be valid for
the geometry and anatomy being imaged in this work.
Attenuation can also be estimated using X-ray CT images to
identify tissue types (bone, soft tissue, air, etc.), but this
requires a second imaging procedure.
Because of the unusual geometry of our proposed camera,
the sensitivity of each line of response depends on the
distance between the two detectors; the shorter the distance
is, the higher the sensitivity. Normalization will include
geometric correction (sensitivity changes caused by the septa
that are not modeled in the solid angle computation), block
effect correction, individual detector sensitivity correction,
and dead time correction. These factors will be obtained
using a scan of an uniform activity cylindrical source such as
used in conventional PET scanners. However, because of our
specialized geometry and lack of rotational symmetry, the
method for deriving the first two factors will be different
from that of circular systems and may require a longer scan
time to achieve good count statistics.
Fig. 7: Spherical "prostate" (2.5 cm diameter) is divided into quadrants, as
shown in (a). Reconstructed image with a tumor in the (b) upper left
Random events will be estimated using the delayed
quadrant, (c) right half, and (d) upper right and lower left quadrants of the window technique. Instead of pre-subtracting histogrammed
prostate, assuming a tumor to background ratio of 2:1, 30% scatter fraction data, delayed events will be saved in the list mode data
and 20% randoms fraction. All prostate tumors are clearly visible in all
images, demonstrating that we can reconstruct nearly artifact free images
stream with a tag to distinguish them from prompt events.
near the prostate despite incomplete sampling and that the scanner can Such information can be used to obtain a better estimate of
differentiate partial and whole prostate tumors. the mean of the random events [27]. For our statistical
reconstruction algorithm, the estimated randoms will be
scan after injecting a clinically acceptable (10 mCi) patient included in the forward model in reconstruction. Anatomical
dose of [11C]choline or [18F]fluorocholine. registration can be done using ultrasound or a point source.
Rates are also computed using analytic estimates that were
developed in our research group [24] and fit to published V. DISCUSSION
sensitivity and backgrounds of the HR tomograph [25]. They
This project is in its infancy, so there are many remaining
assume a 10 mCi injection of [11C]choline or
uncertainties such as: which radiopharmaceutical will be
[18F]fluorocholine uniformly distributed in a 70 kG subject
used, the specific requirements for the radiopharmaceutical
6
Accepted to IEEE Transactions on Nuclear Science LBNL-47068
and imaging procedure, and the accuracy that physicians serum of prostatic cancer patients,” Anticancer Res, vol. 7, pp. 927-
35, 1987.
would require to perform effective diagnosis and evaluate [9] J. Kurhanewicz, D. B. Vigneron, H. Hricak, et al., “Three-
response to therapy. However, new PET radiopharmaceuticals dimensional H-1 MR spectroscopic imaging of the in situ human
have recently demonstrated promising results in the sensitive prostate with high (0.24-0.7-cm3) spatial resolution,” Radiology, vol.
198, pp. 795-805, 1996.
detection of prostate cancer. Hence, a low cost PET scanner [10] J. Kurhanewicz, D. B. Vigneron, R. G. Males, et al., “The Prostate:
coupled with a 18F-based isotope could make PET diagnosis Magnetic Resonance Imaging and Spectroscopy: Present and Future,”
of prostate cancer more clinically available. This PET camera in Radiologic Clinics of North America, H. Hricak and P. R. Carroll,
Eds. New York, New York: W. B. Saunders Co., 2000, pp. 115-138.
design is also likely to out-perform whole body PET for this [11] T. Hara, N. Kosaka, N. Shinoura, et al., “PET imaging of brain tumor
task, as its 2 times higher sensitivity will help for the short with [methyl-C11]choline,” J Nucl Med, vol. 38, pp. 842-7, 1997.
scans that appear to be necessary with choline-like [12] N. Shinoura, M. Nishijima, T. Hara, et al., “Brain tumors: detection
with C-11 choline PET,” Radiology, vol. 202, pp. 497-503, 1997.
radiopharmaceuticals due to bladder accumulation. We [13] T. Hara, N. Kosaka, and H. Kishi, “PET imaging of prostate cancer
believe that providing improved PET imaging performance using carbon-11-choline,” J Nucl Med, vol. 39, pp. 990-5, 1998.
near the prostate bed at lower cost (than commercial PET) is [14] T. Hara, N. Kosaka, T. Kondo, et al., “Imaging of brain tumor, lung
cancer, esophagus cancer, colon cancer, prostate cancer, and
clinically significant, outweighing the trade-off of some bladder cancer with [C-11]choline,” J Nucl Med, vol. 38 (suppl), pp.
resolution degradation for distant metastases. 250P (abstract), 1997.
[15] T. Hara, K. Inagaki, N. Kosaka, et al., “Sensitive detection of
mediastinal lymph node metastasis of lung cancer with 11C-choline
VI. CONCLUSION PET,” J Nucl Med, vol. 41, pp. 1507-13, 2000.
Promising new PET tracers for prostate cancer, such as [16] I. J. Jong, T. H. Que, J. Pruim, et al., “Imaging of bladder cancer
using carbon-11 choline positron emission tomography,” J. Nucl.
[11C]choline and [18F]fluorocholine, motivate us to build a Med., vol. 41 (5 Suppl), pp. 74, 2000.
low cost PET camera optimized for prostate imaging. Monte [17] J. Kotzerke, J. U. Prang, B. Neumaier, et al., “Carbon-11 choline
Carlo simulations demonstrate the feasibility of detecting and positron emission tomography (PET) of prostate cancer -- first clinical
experience,” J. Nucl. Med., vol. 41 (5 Suppl), pp. 74, 2000.
differentiating partial and whole prostate tumors with a tumor [18] J. Kotzerke, J. Prang, B. Neumaier, et al., “Experience with carbon-
to background ratio of 2:1, using a 6-minute scan after a 10 11 choline positron emission tomography in prostate carcinoma,” Eur
mCi injection. Thus, the proposed camera will image J Nucl Med, vol. 27, pp. 1415-9, 2000.
[19] T. R. Degrado, R. E. Coleman, S. W. Baldwin, et al., “Fluorine-18
prostate tumors with good spatial resolution and image fluorocholine (FCH) as an oncological PET tracer: evaluation in
contrast at low cost relative to commercial PET scanners. murine prostate cancer xenograft model,” J. Nucl. Med., vol. 41 (5
Suppl), pp. 231, 2000.
[20] T. R. Degrado, R. E. Coleman, S. Wang, et al., “Synthesis and
VII. ACKNOWLEDGMENT Evaluation of F18-labeled Choline as an Oncologic Tracer for
We appreciate Dr. Hara from the International Medical Positron Emission Tomography: Initial Findings in Prostate Cancer,”
Cancer Research, vol. 61(1), pp. 110-117, 2001.
Center of Japan for providing the PET images presented in [21] P. R. G. Virador, W. W. Moses, and R. H. Huesman, “Reconstruction
Fig. 1 of this paper. This work was supported by the in PET cameras with irregular sampling and depth of interaction
Director, Office of Science, Office of Biological and capability,” IEEE Trans Nucl Sci, vol. NS-45, pp. 1225–30, 1998.
[22] J. Qi, R. M. Leahy, S. R. Cherry, et al., “High-resolution 3D Bayesian
Environmental Research, Medical Science Division of the image reconstruction using the microPET small-animal scanner,”
U.S. Department of Energy under contract No. DE-AC03- Phys Med Biol, vol. 43, pp. 1001-13, 1998.
76SF00098. Reference to a company or product name does [23] S. E. Derenzo, R. H. Huesman, J. L. Cahoon, et al., “Positron
tomograph with 600 BGO crystals and 2.6 mm resolution,” IEEE
not imply approval or recommendation by the University of Trans Nucl Sci, vol. NS-35, pp. 659-64, 1988.
California or the U.S. Department of Energy to the exclusion [24] S. E. Derenzo, “Method for optimizing side shielding in positron
of others that may be suitable. emission tomographs and for comparing detector materials,” J Nucl
Med, vol. 21, pp. 971-977, 1980.
[25] K. Wienhard, M. Dahlbom, L. Eriksson, et al., “The ECAT EXACT
VIII. REFERENCES HR: performance of a new high resolution positron scanner,” J Comp
[1] http://osp.nci.nih.gov/PRGReports/PPRGReport/toc.htm. Asst Tom, vol. 18, pp. 110-118, 1994.
[2] H. I. Scher and L. W. Chung, “Bone metastases: improving the [26] T. F. Budinger and G. T. Gullberg, “Three-dimensional reconstruction
therapeutic index,” Semin Oncol, vol. 21, pp. 630-56, 1994. in nuclear medicine emission imaging.,” IEEE Trans Nucl Sci,, vol.
[3] H. I. Scher, M. Mazumdar, and W. K. Kelly, “Clinical trials in NS-21, pp. 2-20, 1974.
relapsed prostate cancer: defining the target,” J Natl Cancer Inst, vol. [27] E. Mumcuoglu, R. M. Leahy, and S. R. Cherry, “Bayesian
88, pp. 1623-34, 1996. reconstruction of PET images: methodology and performance
[4] M. A. Eisenberger and W. G. Nelson, “How much can we rely on the analysis,” Phys Med Bio, vol. 41, pp. 1777-1807, 1996.
level of prostate-specific antigen as an end point for evaluation of
clinical trials? A word of caution!,” J Natl Cancer Inst, vol. 88, pp.
779-81, 1996.
[5] S. E. Schild, W. W. Wong, G. L. Grado, et al., “Radiotherapy for
isolated increases in serum prostate-specific antigen levels after
radical prostatectomy,” Mayo Clin Proc, vol. 69, pp. 613-9, 1994.
[6] C. A. Perez and A. Eisbruch, “Role of post-radical prostatectomy
irradiation in carcinoma of the prostate,” Semin Rad Oncol, vol. 3, pp.
198, 1993.
[7] B. E. Shevlin, B. B. Mittal, W. N. Brand, et al., “The role of adjuvant
irradiation following primary prostatectomy, based on histopathologic
extent of tumor,” Int J Rad Oncol Biol Phys, vol. 16, pp. 1425-30,
1989.
[8] J. S. Horoszewicz, E. Kawinski, and G. P. Murphy, “Monoclonal
antibodies to a new antigenic marker in epithelial prostatic cells and

S-ar putea să vă placă și