Sunteți pe pagina 1din 106

5392

Dual Chamber Temporary External Pacemaker

Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a
physician.
The following list includes trademarks or registered trademarks of Medtronic in the
United States and possibly in other countries. All other trademarks are the property
of their respective owners.
Medtronic
Medtronic 5392

Contents

1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.5 Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.6 Compatible accessory components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.7 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2 Warnings, precautions, and adverse events . . . . . . . . . . . . . . . . . . . . . . . . . . . 12


2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.3 Environmental precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4 Adverse effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3 Controls, indicators, and other features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20


3.1 Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.2 Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.3 RATE, A OUTPUT, and V OUTPUT parameters . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4 Lower screen functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

4 Preparation for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32


4.1 Checks prior to use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2 Physical features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.3 Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.4 Battery installation and replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.5 Connector Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.6 Placement during use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

5 Instructions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43


5.1 Basic operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.2 Thresholds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.3 Pacing parameter adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

Technical Manual 3
Medtronic 5392

5.4 Timing violations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69


5.5 Rapid Atrial Pacing (RAP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.6 Automatic responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

6 Cleaning, disinfecting, and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78


6.1 Cleaning and disinfecting the temporary pacemaker . . . . . . . . . . . . . . . . . . . . . 78
6.2 Safety and technical checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
6.3 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
6.4 Product useful life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

7 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
7.1 Device specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
7.2 Pacing information tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

8 Special notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
8.1 Special notice for the temporary pacemaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

9 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
9.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

4 Technical Manual
Medtronic 5392

1 Overview

1.1 About this manual


This manual describes the features and functions of the Medtronic Model 5392 Dual
Chamber Temporary External Pacemaker (referred to as the “temporary pacemaker”).

1.2 Symbols
Explanation of symbols
System meets the applicable Canadian and U.S. electrical safety
standards.

Consult instructions for use

Caution

Conformité Européenne (European Conformity)


This symbol means that the device fully complies with European
Directive 93/42/EEC.
Temporary external pacemaker

Atrium chamber designator

Ventricle chamber designator

For atrial use

For ventricular use

Technical Manual 5
Medtronic 5392

Explanation of symbols
Defibrillation-proof type CF applied part

Do not dispose of this product in the unsorted municipal waste


stream. Dispose of this product according to local regulations. See
http://recycling.Medtronic.com for instructions on proper disposal
of this product.
For US audiences only

Package contents

Dual chamber temporary pacemaker

Product documentation

Accessories

Storage temperature limitation

Humidity limitation

Battery

Reorder number

Authorized representative in the European Community

Manufacturer/Date of manufacture

6 Technical Manual
Medtronic 5392

Explanation of symbols
Date of manufacture

Serial number

Non-ionizing electromagnetic radiation

Notice of proper disposal

This product conforms to IP21. There are no openings that allow the
user to insert a finger or similarly sized objects. The product is
resistant to dripping water or vertically falling drops.
China RoHS

MR Unsafe. An item that is known to pose hazards in all MR envi-


ronments.

1.3 General description


The temporary pacemaker is a battery-powered, dual chamber, temporary pacemaker
designed primarily for temporary antibradycardia pacing therapy. The temporary
pacemaker provides 7 selectable modes of pacing therapy: DDD, DDI, DOO, AAI, AOO,
VVI, and VOO. High-rate, burst pacing therapy up to 800 ppm for atrial tachyarrhythmias is
available in the asynchronous mode.1
The temporary pacemaker typically is connected to temporary transvenous, epicardial, or
myocardial pacing leads in a bipolar configuration, using either Medtronic patient cables,
Medtronic surgical cables, or compatible patient cables (see Section 1.6).
The temporary pacemaker operates using 2 LR6-sized (AA-sized) alkaline batteries (see
Section 7.1). The batteries are installed in the battery drawer at the bottom of the temporary
pacemaker.

1 For atrial use only.

Technical Manual 7
Medtronic 5392

Note: The temporary pacemaker is a constant current device. When it emits a pulse, the
current output is maintained at a constant value. This value is set by the output control and
does not vary.

1.3.1 Safety features


The temporary pacemaker includes the following safety features:
● Self-test function
● Low battery indicator
● Lock feature to prevent accidental change of parameters
● Safe, two-step operation to turn off the temporary pacemaker to avoid unintended
shutdown
● Runaway rate protection
● Protection from defibrillation shock
● Continuous operation during battery replacement (see Section 7.1)
● Electrostatic protection
● Minimized susceptibility to electromagnetic and magnetic interference

1.3.2 Operating features


The temporary pacemaker includes the following operating features:
● Single chamber pacing modes – AAI, AOO, VVI, and VOO
● Dual chamber pacing modes – DDD, DDI, DOO
● No pacing therapy – OOO
● Easy-to-view rate and output settings
● Pacing and sensing status indicators – shows temporary pacemaker interaction with the
heart
● Low battery indicator – indicates when to replace the batteries
● Four-dial operation – provides therapy for most pacing needs
● Rate-dependent parameters – rate adjustment automatically sets Upper Rate, PVARP,
and A-V Interval
● Lock/Unlock key – safeguards against unintentional parameter changes
● Lower screen messages – to aid in device operation

8 Technical Manual
Medtronic 5392

● Menu screens – for adjusting additional parameters, including sensitivity,


rate-dependent parameters; for Rapid Atrial Pacing (RAP); for directly selecting 7
pacing modes (DDD, DDI, DOO, AAI, AOO, VVI, and VOO) or selecting no pacing
therapy (OOO).
● Pause key – to suspend pacing and sensing to view the patient’s intrinsic rhythm
● DOO/Emergency key – starts emergency dual chamber asynchronous (DOO) pacing at
maximum atrial and ventricular outputs
● Automatic sensitivity adjustment
● Automatic mode switch during atrial arrhythmias

1.4 Intended use


The temporary pacemaker is used with a cardiac pacing lead system for temporary single or
dual chamber pacing in a clinical environment by trained personnel. Train clinical personnel
on the functionality and use of the temporary pacemaker prior to initial use of the device, as
needed, and per clinic procedures. Contact your Medtronic representative to schedule
training.
The temporary pacemaker can be used where short-term demand (synchronous) or
asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The
temporary pacemaker must be used in an environment where the patient is monitored
continuously to ensure that it is operating properly and delivering appropriate therapy to the
patient.
Specific indications for temporary cardiac pacing include, but are not limited to, the
following:
● Complete heart block
● Sinus bradycardia
● Sick sinus syndrome
● Bradycardia with congestive heart failure
● Atrial and/or ventricular arrhythmias
● Cardiac arrest
● Support, management, and evaluation of a patient before permanent pacemaker
implantation
● Support during permanent pacemaker replacement
● Cardiac complications during invasive or surgical procedures
● Support following cardiac surgery

Technical Manual 9
Medtronic 5392

● Acute myocardial infarction complicated by heart block


● Atrial tachyarrhythmias that require high-rate burst pacing for treatment

1.5 Package contents


The temporary pacemaker is supplied with the following items:
● Two LR6-sized (AA-sized) alkaline batteries (see Section 7.1)
● Literature

1.6 Compatible accessory components


The following compatible accessory components are available for the temporary
pacemaker:
● Medtronic Model 5409 Disposable Pouch
● Medtronic Model 53922 Disposable Cover
The following reusable compatible cables are available for the temporary pacemaker:
● Medtronic Patient Cables (Model 5433 family)
● Medtronic Surgical Cables (Model 5832 family)
The following disposable compatible cables are available for the temporary pacemaker:
● Medtronic Surgical Cable (Model 5833 family)
● Medtronic Patient Cables (Model 5846 family)
● Medtronic Patient Cables (Model 5487 family)
● Compatible temporary transvenous, epicardial, or myocardial pacing leads
Contact your local Medtronic representative to order them.

1.7 Contraindications
There are no known contraindications for the use of temporary pacing as a means to control
heart rate. However, the patient’s age and medical condition may dictate the type of
temporary pacemaker and lead system that the physician uses.

10 Technical Manual
Medtronic 5392

1.7.1 Atrial sensing


Pacing modes that allow sensing in the atrium to trigger a ventricular response are
contraindicated in the presence of rapid atrial arrhythmias, such as atrial fibrillation or atrial
flutter.

1.7.2 Atrial pacing


Atrial pacing is ineffective in the presence of atrial fibrillation or flutter.
Single chamber atrial pacing is contraindicated in the presence of AV conduction disorders.

1.7.3 Asynchronous pacing


Asynchronous pacing is contraindicated in the presence of intrinsic cardiac rhythms.

1.7.4 Atrial high-rate burst pacing therapy


Atrial high-rate burst pacing therapy is intended for use in the atrium only. High-rate burst
pacing is contraindicated in the ventricle; it may result in life-threatening arrhythmias.

1.7.5 Concomitant pacing


Temporary pacing is contraindicated in the presence of another pacing system.
Do not use the temporary pacemaker to pace a patient while another pacing system is also
actively pacing the patient. Concomitant pacing can occur where both pacing systems
compete to pace the patient.
If concomitant pacing occurs, the temporary pacemaker may not be able to pace the patient,
or may pace the patient asynchronously. Concomitant pacing could cause the temporary
pacemaker to potentially pace into a T-Wave or result in a pacemaker-mediated
tachycardia.

Technical Manual 11
Medtronic 5392

2 Warnings, precautions, and adverse events

2.1 Warnings
Patient monitoring – Monitor the patient continuously while the temporary pacemaker is in
use to ensure it is operating properly and delivering appropriate therapy to the patient.
Equipment modification – Do not modify the temporary pacemaker. Modifications could
impact the temporary pacemaker effectiveness and adversely affect patient safety.
Temporary pacemaker compatibility – Only connect items that have been specified as
part of the temporary pacemaker or that have been specified as being compatible with the
temporary pacemaker.
Temporary pacemaker use – The temporary pacemaker is not suitable for use in the
presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Defibrillation/cardioversion – The temporary pacemaker is protected from damage
caused by internal defibrillation discharges up to 50 J (watt-seconds) and external
defibrillation discharges up to 360 J. However, it is recommended that paddles be placed as
far away from the temporary pacemaker or the lead system as is practical.
Whenever possible, for the safety of the patient, disconnect the temporary pacemaker from
the implanted lead system before defibrillating or cardioverting. Excessive defibrillation
energy can damage the temporary pacemaker. This can result in a large current flowing
through the implanted lead system and temporary pacemaker, which could reduce intended
defibrillation energy delivered to the patient or cause myocardial damage.
If damage to the temporary pacemaker is suspected due to defibrillation, disconnect it from
the patient and return it to Medtronic for service.
Line-powered equipment – An implanted lead or a lead with an extension cable
constitutes a direct, low-resistance current pathway to the myocardium. Due to the danger
of tachyarrhythmias resulting from alternating current leakage, extreme caution must be
taken to properly ground all line-powered equipment used on or in the vicinity of the patient.
Electrosurgical units (cautery) – Electrosurgical units can cause loss of pacing from
oversensing or tachyarrhythmias by inducing current on the leads, and thus should never be
used within 15 cm (6 in) of the pacemaker/lead system.
Ablation (RF ablation or microwave ablation) – Ablation is a surgical technique in which
radio frequency (RF) or microwave energy is used to destroy cells by creating heat. Ablation
used in cardiac device patients may result in, but is not limited to, induced ventricular
tachyarrhythmias, oversensing, unintended tissue damage, device damage, or device
malfunction.

12 Technical Manual
Medtronic 5392

Pulse-modulated ablation systems may pose higher risk for induced ventricular
tachyarrhythmias. Medtronic cardiac devices are designed to withstand exposure to
ablation energy. To mitigate risks, observe the following precautions:
● Ensure that temporary pacing and defibrillation equipment is available.
● Avoid direct contact between the ablation catheter and the temporary leads.
● Position the return electrode patch so that the electrical current pathway does not pass
through or near the device and leads.
● Continuously monitor the patient during ablation with at least two separate methods,
such as arterial pressure display, ECG, manual monitoring of the patient’s rhythm
(taking pulse) or monitor by some other means such as ear or finger pulse oximetry, or
Doppler pulse detection.
To avoid or mitigate the effects of oversensing, if appropriate for the patient, initiate
asynchronous pacing.
Electromagnetic interference (EMI) – Pacemakers operating in the demand mode
respond to intracardiac potentials with magnitudes of a few mV. This level of sensitivity
makes the temporary pacemaker inherently sensitive to some external fields. In the
presence of excessive levels of interference, the temporary pacemaker may inhibit
completely or revert to asynchronous operation, pacing at the rate set by the RATE dial.
It is recommended that the temporary pacemaker be set to an asynchronous pacing mode
at a rate higher than the patient’s intrinsic rate when operated in the presence of strong
electromagnetic interference (EMI).
Sources of excessively strong EMI that may temporarily affect the operation of the
temporary pacemaker include the following:
● Electrosurgical equipment
● Diathermy equipment
● Some medical telemetry equipment (when operated within 1 m [about 3 feet] of the
pacemaker)
● Communication transmitters such as cellular phones, “walkie talkies”, and transmitters
in emergency transport vehicles
● Magnetic resonance imaging (MRI) equipment
Atrial High-Rate Burst Pacing Therapy (Rapid Atrial Pacing) – Use of high rates in the
atrium could result in high-rate conduction to the ventricle. Defibrillation equipment should
be on standby, immediately available during atrial high-rate burst pacing therapy.
There is no ventricular back-up pacing during delivery of atrial high-rate burst pacing
therapy.

Technical Manual 13
Medtronic 5392

Connecting the lead system – The patient cables should be connected to the temporary
pacemaker before the lead(s) is connected to the patient cable(s).
Handling implanted leads – When handling implanted leads (temporary or permanent),
the terminal pins or exposed metal must not be touched nor be allowed to contact electrically
conductive or wet surfaces.
MR unsafe – The temporary pacemaker is MR unsafe. Do not bring the temporary
pacemaker into Zone 4 (magnet room), as defined by the American College of Radiology.

2.2 Precautions
Random failures – The physician should be aware that operational failure of the temporary
pacemaker can occur as the result of battery depletion, mishandling, or random component
failure.
Possible operational failures of the temporary pacemaker can include the following:
● No output or erratic output
● No sensing or erratic sensing
● False indicator light signals
● Inappropriate variance of rate, output pulse width, or output amplitude
● Reversion to asynchronous pacing
● Loss of control of rate, output, sensitivity, or power
If loss of control of rate, output, sensitivity, or power occurs, and it is not due to a low battery,
disconnect the temporary pacemaker from the patient and return it to Medtronic for service.
Temporary pacemaker repair – Do not attempt to repair the temporary pacemaker. Only
a qualified Medtronic Technical Services representative can repair the temporary
pacemaker. Contact Medtronic at the telephone number on the back cover of this manual if
the temporary pacemaker requires service.
Service condition – Before each use, evaluate the temporary pacemaker for damage and
observable defects. Do not use the temporary pacemaker if the case is cracked, the controls
are not functioning, the displays are not working, or if the controls, displays, or connectors
are broken. If the temporary pacemaker has any observable defects, contact Medtronic at
the telephone number on the back cover of this manual for service.
Cleaning, disinfection, and sterilization – Clean and disinfect the temporary pacemaker
before each use for a new patient. Clean and sterilize the reusable cables before each use
for a new patient.
Batteries – Only install the recommended batteries in the temporary pacemaker. Batteries
with different physical dimensions, non-alkaline (e.g., lithium or rechargeable) or batteries

14 Technical Manual
Medtronic 5392

with contamination on the battery terminals, may result in erratic operation of the temporary
pacemaker, no pacing output, or damage to the temporary pacemaker, specifically to the
battery compartment.
Replace the batteries for each new patient and when the low battery indicator flashes during
temporary pacemaker operation.
Only use new, fresh batteries that have not passed their expiration date.
Inspect battery terminals for contamination. Using batteries with contaminated terminals
can result in the temporary pacemaker turning off, decreased battery life, or corrosion to the
battery compartment.
Check the battery status at least twice daily. Replace alkaline batteries at least once every
week when the temporary pacemaker is in continuous use or when the low battery indicator
flashes. Verify that the battery drawer is fully closed and latched in place.
Failure to ensure that the battery drawer is fully latched may result in loss of power.
Continued temporary pacemaker operation is not an indication that the battery drawer is
properly latched.
Battery installation – Ensure that the new, fresh batteries are installed with the correct
battery polarity by verifying that the batteries align with the polarity markings on the inside of
the battery drawer. The temporary pacemaker requires proper battery polarity for operation.
After installing the batteries, ensure that the battery status indicator displays full battery
power and that the low battery indicator is not flashing. The temporary pacemaker may
temporarily continue to pace and sense with weak, dead, or incorrectly installed batteries.
Pacing leads and cables – Improper connection, displacement, or fracture of leads or
cables may result in pacemaker system failure. Inspect leads and cables for damage before
each use.
Pacing system adjustments – Monitor the patient’s ECG and blood pressure. Keep
defibrillation equipment on standby, immediately available for emergency use during
evaluation of stimulation and sensing thresholds, pacemaker and pacing lead connections
and adjustments, and atrial high-rate burst pacing therapy.
Default DDD pacing mode – The default power-up settings for the DDD pacing mode are
not always appropriate for every patient or situation. Select the appropriate pacing mode to
meet the pacing needs of the patient.
Patient monitoring after defibrillation – Monitor the patient after a defibrillation has
occurred to verify that the temporary pacemaker and the cable/lead systems are still
delivering the appropriate therapy.
Bipolar lead systems – Bipolar lead systems are recommended because they are less
susceptible to electromagnetic interference. Separation between the positive (+) electrode
and negative (–) electrode of the same lead system should not exceed 15 mm (0.6 in). Place
the atrial and ventricular lead systems at right angles to each other. Verify that the electrodes

Technical Manual 15
Medtronic 5392

of one system are a minimum of 4 cm (1.5 in) from the electrodes of the other system. Failure
to follow these spacing recommendations could result in oversensing. Clinical risks for not
following these spacing recommendations include, but are not limited to, loss of pacing
output, asynchronous ventricular pacing, and pacemaker mediated tachycardia.
Unipolar lead systems – Unipolar lead systems are not recommended because they are
more susceptible to electromagnetic interference, which may result in inappropriate pacing.
Unipolar lead systems should not be used in the dual-chambered pacing modes because
the current path of one lead system may interfere with the current path of the other.
Dual chamber pacing modes – Do not configure the temporary pacemaker in any of the
dual chamber pacing modes unless both channels are connected to the heart. If the
temporary pacemaker is configured in a dual chamber pacing mode and one of the channels
is not connected, the open channel could pick up unintended noise. The noise can be
interpreted as a sensed event and can lead to events such as, but not limited to,
asynchronous ventricular pacing, unintended pacing inhibition, or ventricular tachycardia.
Avoid use of dual chamber pacing modes without an atrial connection as this could, in rare
circumstances, result in inappropriate delivery of a ventricular pace on an intrinsic T-wave.
Atrial sensing – When programming to a mode that requires atrial sensing, the sensing
threshold should be evaluated for sufficient safety margin.
Place the temporary pacing lead on the right atrial free wall, oriented along the direction of
the myocardial fibers, approximately 1 cm (0.4 in) apart. It is important to achieve a sensing
threshold of at least 1.0 mV. Set atrial sensitivity to a minimum of one-half the measured
threshold. The setting ensures a minimum safety margin of two times the sensing threshold.
Failure to follow this procedure can lead to delivery of asynchronous pulses.
Sensing thresholds – Do not use the temporary pacemaker to determine sensing
thresholds for permanently implanted lead systems. When implanting a permanent
pacemaker, Medtronic recommends the use of a pacing system analyzer (PSA).
Sensitivity settings – Since the sensitivity setting determines the smallest signal that can
be sensed by the pacemaker, set the sensitivity dial to one-half the mV value of the patient’s
sensitivity threshold. This setting will provide a 2x safety margin to ensure proper sensing.
A more sensitive setting may be chosen to provide a greater safety margin. However, be
aware that setting the sensitivity value too low (too sensitive) could result in inappropriate
sensing of far field signals (for example, sensing of R-waves or T-waves on the atrial channel
or P-waves on the ventricular channel), leading to inappropriate inhibition of pacing pulses.
Sensitivity threshold testing – Complete the sensitivity threshold testing to determine the
appropriate settings for sensitivity. Clinical risks for failure to perform this step include, but
are not limited to, asynchronous ventricular pacing and pacemaker mediated tachycardia.
Output threshold testing – Complete the output threshold testing to determine the
appropriate settings for output. Clinical risks for failure to perform this step include, but are

16 Technical Manual
Medtronic 5392

not limited to, loss of capture, retrograde conduction, induced tachycardia, and loss of
hemodynamic support.
High output and maximum sensitivity – Although the temporary pacemaker contains a
safety pacing feature that prevents inappropriate inhibition of ventricular pacing due to
far-field sensing, the simultaneous use of high output and maximum sensitivity (that is, the
lowest mV value) in the presence of atrial flutter or fibrillation can result in fatal arrhythmias.
Electrostatic discharge (ESD) – The pacing lead(s) provides a low-impedance pathway
to the heart. Therefore, it is recommended that attending health care professionals
discharge any static electricity by touching a large metal or conductive, grounded surface
before touching the patient, the cable, the leads, or the temporary pacemaker. Also,
neutralize any static electricity from the patient by touching the patient away from (i.e., distal
to) the leads.
Retrograde conduction – If retrograde P-waves are being sensed outside the
rate-dependent, automatic Post-Ventricular-Atrial-Refractory Period (PVARP) setting,
manually increase the PVARP until the retrograde waves fall inside the PVARP. Failure to
follow this procedure may lead to a pacemaker-mediated tachycardia (PMT).
Termination of pacing – Abrupt termination of pacing stimuli may result in intervals of
asystole before an intrinsic rhythm is re-established. Before terminating pacing, set the
temporary pacemaker to a demand mode; then gradually reduce the pacing rate below the
patient’s intrinsic rate.
Pause key – Use the Pause key with care since the patient receives no pacing support (for
a maximum of 10 s at a time) when the Pause key is pressed and held.
A-V Interval – Programming long A-V intervals may result in pacing the ventricle during the
vulnerable period of ventricular repolarization, thus precipitating ventricular arrhythmias in
unstable patients.
DOO/Emergency key – Use the DOO/Emergency key only when high-output
asynchronous pacing (DOO) is needed. When the DOO/Emergency key is pressed, the
emergency pacing mode is entered and remains in effect until the emergency pacing mode
is deactivated. Press the Enter key to deactivate emergency pacing mode.

Technical Manual 17
Medtronic 5392

2.3 Environmental precautions


The temporary pacemaker has been carefully designed and tested to ensure reliability
during normal use. However, electronic devices are susceptible to many environmental
stresses. To avoid damage to the temporary pacemaker, observe the following precautions:
● Do not drop the temporary pacemaker or handle it in a way that might physically damage
it. The temporary pacemaker may appear to work appropriately immediately after being
dropped or mishandled, but operational damage may have occurred. Perform safety
and technical checks if the temporary pacemaker has been dropped.
● Do not place the temporary pacemaker in any area where a patient may interact with it.
Tampering with programmed parameters may have direct and serious patient health
effects. The temporary pacemaker should be placed in an area that minimizes
tampering with the device by unauthorized personnel (for example, patients or visitors).
Medtronic recommends use of the Model 53922 Disposable Cover to reduce the risk of
tampering with the programmed parameters.
● Avoid spilling fluid on the temporary pacemaker. The temporary pacemaker was
carefully designed to minimize leakage, but fluid incursion may still occur. Medtronic
recommends the use of a protective cover, such as the Model 5409 Disposable Pouch,
to minimize fluid incursion.
● Avoid contaminating the patient cable receptacles with blood or other body fluids.
● Always use safe electrostatic discharge (ESD) procedures; the temporary pacemaker
could be adversely affected by ESD.
● Do not open the temporary pacemaker. The seam joining the unit is designed to
minimize fluid incursion and may not be effective if improperly opened and resealed.
Furthermore, removing the label on the back of the temporary pacemaker may
compromise the ESD barrier. Opening the temporary pacemaker voids the warranty.
● Do not sterilize the temporary pacemaker by gamma irradiation or steam (autoclave).
● Do not store the temporary pacemaker with the batteries in the battery drawer. Remove
the batteries when the temporary pacemaker is not in use.
● Rapid temperature changes may affect proper operation. Always allow the temperature
of the temporary pacemaker to stabilize in the environment in which it will be used before
attachment and operation.
● Prolonged storage or operation of the temporary pacemaker in high humidity may affect
proper operation. Allow the temporary pacemaker to completely dry after exposure to
humidity.
Other environmental factors may impact proper performance of the temporary pacemaker
in the hospital setting. Use of appropriate environmental health and safety practices will help
prevent environmental damage to the temporary pacemaker.

18 Technical Manual
Medtronic 5392

2.4 Adverse effects


Temporary pacemakers – Potential adverse effects related to the use of the temporary
pacemaker include, but are not limited to the following:
● Asystole following abrupt cessation of pacing
● Inhibition or reversion in the presence of strong electromagnetic interference
● Initiation of a tachyarrhythmia or acceleration of an existing tachyarrhythmia
Atrial high-rate burst pacing – Atrial high-rate burst pacing may result in the onset of
tachycardia, acceleration of an existing tachycardia, or fibrillation. Application of temporary
atrial high-rate burst pacing should be performed in a carefully monitored and controlled
patient environment. Monitor the patient’s ECG and blood pressure. Keep defibrillation
equipment on standby and immediately available for emergency use.
Dual chamber modes – In the DDI and DDD pacing modes, the ventricular sense amplifier
may sense the atrial pacing pulse. Reducing the atrial amplitude, the ventricular sensitivity,
and/or repositioning the electrodes may be necessary to avoid this situation.
Safety margins – Determine an adequate safety margin for sensing and pacing in both the
ventricle and atrium. Failure to do so may result in inappropriate pacing.
Lead systems – Potential adverse effects related to the use of pacing lead systems used in
conjunction with the temporary pacemaker include, but are not limited to the following:
● Inappropriate lead connections
● Inadvertent disconnection of the lead system
● Lead fracture or displacement causing intermittent or complete loss of capture and/or
sensing
● Perforation and tamponade
Other potential adverse effects related to the use of any implanted lead system include, but
are not limited to, the following:
● Myocardial irritability resulting in fibrillation
● Infarction
● Pericarditis
● Body rejection phenomena (local tissue reaction)
● Muscle and nerve stimulation
● Infection
Nerve or muscle stimulation can be caused by pacing lead contact with the nerve or muscle
tissue and/or by high-output settings. The stimulation may be controlled by repositioning or
replacing the electrode, or by reducing the output pulse amplitude.

Technical Manual 19
Medtronic 5392

3 Controls, indicators, and other features

3.1 Controls
The dials and keys used to control the functions and parameter settings of the temporary
pacemaker are described in this chapter.

Note: All adjustments to the RATE, A (Atrial) OUTPUT, and V (Ventricular) OUTPUT dials
take effect within the next two pacing cycles.

3.1.1 Temporary pacemaker controls and indicators


The upper screen indicators display the RATE, A OUTPUT, and V OUTPUT values, pacing
and sensing status, the currently selected pacing mode, the battery status, and the lock
indicator. See Figure 1.
The controls next to the upper screen are used for the following:
● To adjust the RATE, A OUTPUT, and V OUTPUT values using the RATE, A OUTPUT,
and V OUTPUT dials
● To select high-output, dual-chamber asynchronous pacing (DOO for emergency) by
pressing the DOO/Emergency key
● To turn on or turn off the temporary pacemaker by pressing the On/Off key

20 Technical Manual
Medtronic 5392

Figure 1. Controls and indicators for the upper screen

A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

1 DOO/Emergency key 8 Lock indicator


2 On/Off key 9 Pacing mode indicator
3 Pacing and sensing status bar indicators 10 Battery indicator
4 RATE dial 11 V (Ventricular) OUTPUT scale
5 A (Atrial) OUTPUT dial 12 A (Atrial) OUTPUT scale
6 V (Ventricular) OUTPUT dial 13 RATE scale
7 Lock/Unlock key 14 Upper screen

The lower screen controls are used to select pacing modes, to adjust pacing parameter
values, to deliver RAP pacing therapy, to resume synchronous pacing from asynchronous
pacing, and to pause pacing therapy (see Figure 2). The lower screen displays the Mode
Selection menu, pacing parameters, the RAP screens, and warnings and information
messages.

Technical Manual 21
Medtronic 5392

Figure 2. Controls and indicators for the lower screen

Mode Selection

Back

1 Enter key 4 Menu Parameter dial


2 Selection indicator 5 Pause key
3 Up/Down arrow keys 6 Lower screen

Note: Upper and lower screens values contained in this manual are presented for reference
only. Actual values may vary, depending on pacing mode, outputs, and parameter value
selections.

3.1.2 DOO/Emergency key


Press the DOO/Emergency key to select high-output, dual-chamber asynchronous pacing
(DOO for emergency) at any time, including when the temporary pacemaker is off or locked
(see Figure 3). Avoid accidentally activating the DOO/Emergency key. See Section 7.1 for
emergency pacing values.
Figure 3. DOO/Emergency key

3.1.3 On/Off key


Use the On/Off key to turn on or turn off the temporary pacemaker (see Figure 4).
Figure 4. On/Off key

22 Technical Manual
Medtronic 5392

3.1.4 Rate dial


Use the RATE dial to set the base rate, in ppm, at which pacing pulses are delivered (see
Section 5.1.4).

3.1.5 A (Atrial) OUTPUT dial


The A (Atrial) OUTPUT dial is used to set the current amplitude, in mA, of the atrial pacing
pulse (see Section 5.1.4).

3.1.6 V (Ventricular) OUTPUT dial


The V (Ventricular) OUTPUT dial is used to set the current amplitude, in mA, of the
ventricular pacing pulse (see Section 5.1.4).

3.1.7 Lock/Unlock key


The Lock/Unlock key locks the temporary pacemaker to prevent inadvertent adjustment of
the pacing parameters, or unlocks the temporary pacemaker when it is locked. See
Figure 5.
Figure 5. Lock/Unlock key

3.1.8 Enter key


The Enter key is used to select pacing modes in the Mode Selection menu, to select the
Rapid Atrial Pacing (RAP) screen or the Mode Selection menu from the Pacing Parameters
menu, to confirm power-off, to deliver Rapid Atrial Pacing (RAP) from the RAP menu, and to
resume synchronous pacing from asynchronous pacing. See Figure 6.
Figure 6. Enter key

Technical Manual 23
Medtronic 5392

3.1.9 Arrow keys


The Up and Down arrow keys are used to move the selection indicator in the lower screen.
Press the Up or Down arrow keys to select pacing modes from the Mode Selection screen
and to select pacing parameter values from a Pacing Parameters menu. See Figure 7.
Figure 7. Up and Down Arrow keys

3.1.10 Menu parameter dial


The Menu Parameter dial is used to adjust the pacing parameters in the selected pacing
mode (see Figure 2).

3.1.11 Pause key


Press and hold the Pause key to suspend pacing and sensing for up to 10 s (see Figure 8).
When the Pause key is released or when the 10 s timer runs out, pacing and sensing
resumes. Use the Pause key also to view the patient’s intrinsic rhythm (see Section 5.1.3).
Figure 8. Pause key

Caution: Use the Pause key with care. The patient receives no pacing support when the
Pause key is pressed and held (for a maximum of 10 s at a time).

3.2 Indicators
The upper screen displays indicators for the pacing and sensing status bar, current pacing
mode, battery power, and lock status. See Figure 9.

24 Technical Manual
Medtronic 5392

Figure 9. Indicators

A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

1 Pacing and sensing status bar indicators 3 Battery status indicator


2 Pacing mode indicator 4 Lock indicator

3.2.1 Pacing and sensing status bar indicators


The pacing and sensing status bar indicators, located at the top of the upper screen, identify
which chambers the temporary pacemaker is currently set to pace and/or sense and indicate
the current pacing mode (see Section 7.2). The indicators are the A (atrium) and V (ventricle)
chamber indicators, the PACE and SENSE indicators, and the PACE and SENSE LEDs.
See Figure 10.
Figure 10. Pacing and sensing status bar indicators

A V
PACE SENSE PACE SENSE

A V
PACE SENSE PACE SENSE

RATE

30 200 ppm

1 A and V chamber indicators 3 PACE and SENSE indicators


2 PACE and SENSE LEDs

Technical Manual 25
Medtronic 5392

The A and V chamber indicators, if present, indicate that the selected pacing mode enables
the A and/or V chambers for pacing and sensing. However, pacing and sensing only occurs
when the PACE and SENSE indicators are present.
The PACE and SENSE indicators, if present, indicate that the selected pacing mode
enables pacing and sensing in the A and/or V chambers.

Note: While the PACE and SENSE indicators identify which chambers the temporary
pacemaker is set to pace and/or sense, they do not indicate actual temporary pacemaker
interaction with the heart.
The PACE and SENSE LEDs next to the A and V chamber indicators indicate delivery of a
pacing pulse or a sensed event. The LEDs flash when the following pacing or sensing events
occur, as follows:
● The green LED to the left of the A chamber indicator flashes each time the temporary
pacemaker delivers a pacing pulse on the atrial channel.
● The green LED to the left of the V chamber indicator flashes each time the temporary
pacemaker delivers a pacing pulse on the ventricular channel.
Note: The green LED flashes indicate delivery of a pacing pulse, but they are not
confirmation that the pacing pulse has initiated cardiac stimulation.

● The blue LED to the right of the A chamber indicator illuminates when events are sensed
on the atrial channel. The A SENSE LED flashes when the temporary pacemaker
detects events inside and outside the Atrial Refractory Period (ARP).
● The blue LED to the right of the V chamber indicator illuminates when events are sensed
on the ventricular channel. The V SENSE LED flashes only when the temporary
pacemaker detects an event outside a ventricular blanking period.
Note: The blue LEDs indicate a sensed event by the temporary pacemaker, but they are
not confirmation of a cardiac contraction.

26 Technical Manual
Medtronic 5392

3.2.2 Status indicators


The status indicators, at the bottom of the upper screen, display the pacing mode, battery
status, and lock indicators.

3.2.2.1 Pacing mode indicator


The pacing mode indicator displays the current pacing mode.

3.2.2.2 Battery status indicator


The battery status indicator displays the amount of available battery power remaining. When
all of the indicator bars are visible, the batteries have full power or have been replaced with
a fresh set of batteries.
The low battery indicator, a red light, flashes behind the battery status indicator when only
one bar is visible. When the low battery indicator begins flashing, the temporary pacemaker
has approximately 24 hours or less of battery life remaining.

Note: When the low battery indicator appears, the temporary pacemaker maintains pacing
at the current settings, for a minimum of 24 hours, if the settings were at nominal values (see
Section 7.1).
If the batteries are removed, no bars are visible in the battery status indicator. The upper
screen is visible, but the lower screen is not displayed. The temporary pacemaker continues
to pace and sense until insufficient power is available (see Section 4.3).
After the batteries are depleted, the temporary pacemaker shuts down.

3.2.2.3 Lock indicator


The Lock indicator appears when the temporary pacemaker is locked.
If any of the dials are turned or keys are pressed while the temporary pacemaker is locked,
the Lock indicator flashes and the Locked message appears in the lower screen (see
Figure 24). When the temporary pacemaker is locked, the RATE, A OUTPUT, and
V OUTPUT settings, pacing mode selections, or pacing parameters cannot be adjusted until
the temporary pacemaker is unlocked. See Section 5.1.2.

Note: The DOO/Emergency key remains active when the temporary pacemaker is locked.
If the DOO/Emergency key is pressed when the temporary pacemaker is locked, the
temporary pacemaker initiates high-output, dual-chamber asynchronous pacing (DOO for
emergency).

Technical Manual 27
Medtronic 5392

3.3 RATE, A OUTPUT, and V OUTPUT parameters


The RATE, A OUTPUT, and V OUTPUT values are displayed both numerically and
graphically. The scale next to each numerical value and dial shows the range available for
that parameter. Scale segments appear, showing where the parameter is set within the
available range. The numerical value for each setting appears to the right of the scale.

3.3.1 RATE
The RATE ranges from 30 to 200 ppm (see Figure 11). Turn the RATE dial clockwise to
increase RATE and counterclockwise to decrease RATE.
When the temporary pacemaker is turned on, the RATE is set to 80 ppm (nominal). See
Section 7.1 for the RATE increments.
Figure 11. RATE scale, value, and dial
A V
PACE SENSE PACE SENSE

RATE

30 200 ppm

Note: The rate-dependent parameters UPPER RATE, PVARP, and A-V INTERVAL are
automatically adjusted each time RATE is adjusted unless they are manually set (see
Section 5.3.2).

3.3.2 A (Atrial) OUTPUT


The atrial output ranges from 0.1 to 20 mA (see Figure 12). Turn the A OUTPUT dial
clockwise to increase A OUTPUT, or counterclockwise to decrease or turn off A OUTPUT.
The A OUTPUT scale and value are blank when A OUTPUT is turned off.
When the temporary pacemaker is turned on, A OUTPUT is set to10 mA (nominal). See
Section 7.1 for the A OUTPUT increments.
Figure 12. A OUTPUT scale, value, and dial
A OUTPUT

0 20 mA

28 Technical Manual
Medtronic 5392

Note: When A OUTPUT is turned off manually, both the atrial output and the atrial sensitivity
are turned off and there is no atrial pacing or sensing. If A OUTPUT is turned back on
manually before the temporary pacemaker locks (see Section 5.1.2), atrial sensitivity is set
to the previously selected value. If the temporary pacemaker locks after turning off
A OUTPUT, atrial sensitivity is set to the nominal value of 0.5 mV when A OUTPUT is turned
back on.

3.3.3 V (Ventricular) OUTPUT


The ventricular output ranges from 0.1 to 25 mA (see Figure 13). Turn the V OUTPUT dial
clockwise to increase V OUTPUT or counterclockwise to decrease or turn off the
V OUTPUT. The V OUTPUT scale and value are blank when V OUTPUT is turned off.
When the temporary pacemaker is turned on, V OUTPUT is set to 10 mA (nominal). See
Section 7.1 for the V OUTPUT range increments.
Figure 13. V OUTPUT scale, value, and dial
V OUTPUT

0 25 mA

Note: When V OUTPUT is turned off manually, both the ventricular output and the
ventricular sensitivity are turned off and there is no ventricular pacing or sensing. If
V OUTPUT is turned back on manually before the temporary pacemaker locks (see
Section 5.1.2), ventricular sensitivity is set to the previously selected value. If the temporary
pacemaker locks after turning off V OUTPUT, ventricular sensitivity is set to the nominal
value of 2.0 mV when V OUTPUT is turned back on.

3.4 Lower screen functions


The lower screen has 3 functions:
● To select a pacing mode from the Mode Selection menu
● To select and adjust pacing parameters from a Pacing Parameters menu
● To display warnings and instructions

3.4.1 Pacing mode selection


The Mode Selection menu is used to select dual chamber pacing modes (DDD, DDI, or
DOO), single chamber pacing modes (AAI, AOO, VVI, or VOO), or no pacing therapy
(OOO). See Section 5.1.6.

Technical Manual 29
Medtronic 5392

When the Mode Selection menu is displayed, use the Up or Down arrow keys to highlight a
pacing mode with the selection indicator (see Figure 14). Press the Enter key to select the
pacing mode.
Figure 14. Mode Selection menu

Access the Mode Selection menu using one of the following methods:
● Turn on the temporary pacemaker using the On/Off key
● Select Mode Selection from a Pacing Parameters menu (see Figure 15)

3.4.2 Pacing Parameters menu


The Pacing Parameters menu is used to access and adjust the pacing parameters for the
selected pacing mode, to select the Rapid Atrial Pacing (RAP) screen, or to navigate to the
Mode Selection menu (see Figure 15).
Figure 15. Pacing Parameters menu

30 Technical Manual
Medtronic 5392

See Section 5.1.6, Section 5.3, and Section 5.5 for additional information about pacing
parameters, RAP, and the Mode Selection menu.

Note: The Pacing Parameters menu only provides access to the parameters for the selected
pacing mode. Not every pacing parameter is accessible for each pacing mode.

Technical Manual 31
Medtronic 5392

4 Preparation for use

4.1 Checks prior to use


4.1.1 Cleaning and disinfection prior to use
During normal use, the temporary pacemaker and cables can be contaminated. Verify that
the temporary pacemaker is cleaned and disinfected before each use for a new patient. See
Section 6.1 for instructions on cleaning and disinfecting the temporary pacemaker.
Verify that the reusable cables are cleaned and sterilized before each use for a new patient.

Note: For information about cleaning and sterilizing the reusable cables, refer to the
applicable technical manual.

Caution: Clean and disinfect the temporary pacemaker before each use for a new patient.
Clean and sterilize the reusable cables before each use for a new patient.

4.1.2 Service condition


Check the temporary pacemaker and reusable cables before each use for a new patient to
verify that there are no observable defects. Do not use the temporary pacemaker or the
reusable cables if there are any observable defects. Verify that the temporary pacemaker
controls function and that the battery drawer closes.
Visually inspect the reusable cables and connectors. Do not use the reusable cables if they
are damaged. Damage includes, but is not limited to, deterioration of the cable insulation,
brittleness, cracking, thinning, or bare spots. Do not use the reusable cables if the
conductive wires are exposed.

Caution: Before each use, evaluate the temporary pacemaker for damage and observable
defects. Do not use the temporary pacemaker if the case is cracked, the controls are not
functioning, the displays are not working, or if the controls, displays, or connectors are
broken. If the temporary pacemaker has any observable defects, contact Medtronic at the
telephone number on the back cover of this manual for service.

4.2 Physical features


4.2.1 Batteries
Battery drawer – The battery drawer, on the bottom of the temporary pacemaker, accepts
two LR6-sized (AA-sized) alkaline batteries (see Section 7.1).

32 Technical Manual
Medtronic 5392

Battery life – The battery life is 7 days minimum with continuous operation for an alkaline
battery when the RATE is set at 80 ppm and all other parameters are at nominal values (see
Section 7.1).

Note: After 6 days, or when the battery status displays 1 bar, the low battery status indicator
flashes. When the low battery indicator flashes, the temporary pacemaker has 24 hours of
battery life remaining.
Battery drawer latch release button – The battery drawer latch release button at the
bottom of the temporary pacemaker opens the battery drawer when it is pressed.
Continued operation after the batteries are removed – If the batteries are removed, the
temporary pacemaker continues to operate for 30 s, minimum (see Section 7.1), under the
following conditions: RATE of 80 ppm or less, A OUTPUT and V OUTPUT of 10 mA or less,
backlight off, and lower screen blank.

Note: The temporary pacemaker may shut down immediately, depending upon the battery
level, if the batteries are removed while it is turned on.

4.2.2 Connector block


The connector block, at the top end of the temporary pacemaker, has sockets that accept
patient and/or surgical cables. Chamber designations are marked A for atrium and V for
ventricle (see Figure 16). The sockets are also color-coded blue for atrium and white for
ventricle.
Figure 16. Temporary pacemaker connector block

A V

4.2.3 IV pole hanger


The Medtronic Model 53921 IV pole hanger is attached to the back of the temporary
pacemaker and is used to hang the temporary pacemaker on an IV pole (see Figure 17). Fold
the IV pole hanger flat against the back of the temporary pacemaker when it is not in use.

Technical Manual 33
Medtronic 5392

Figure 17. IV pole hanger

1 IV pole hanger

Note: If the Medtronic Model 53921 IV pole hanger requires replacement, contact your
Medtronic sales or service representative.

4.3 Batteries
Warning: Properly ground all line-powered equipment used on the patient or in the vicinity
of the patient (see Section 2.1).

Caution: Monitor the patient’s ECG and blood pressure and keep defibrillation equipment
on standby, immediately available for emergency use during pacing lead insertion and
pacemaker connection.
The temporary pacemaker uses two LR6–sized (AA–sized) alkaline batteries for operation
(for example, Duracell MN1500 or Eveready E91 batteries).

Cautions:
● Only install the recommended batteries in the temporary pacemaker. Batteries with
different physical dimensions, non-alkaline (e.g., lithium or rechargeable) or batteries
with contamination on the battery terminals, may result in erratic operation of the
temporary pacemaker, no pacing output, or damage to the temporary pacemaker,
specifically to the battery compartment.
● Inspect battery terminals for contamination. Using batteries with contaminated
terminals can result in the temporary pacemaker turning off, decreased battery life, or
corrosion to the battery compartment.

34 Technical Manual
Medtronic 5392

When installing, replacing, or checking the battery status, verify the following items:
● Check the battery status regularly while the temporary pacemaker is in use.
● Verify that the low battery indicator is not flashing.
Remove the batteries when the temporary pacemaker is not in use.

4.3.1 Battery installation recommendations


Verify the following when installing batteries in the temporary pacemaker:
● Install only the recommended batteries. Using non-recommended batteries may result
in less than 24 hours of battery life after the low battery indicator illuminates, degraded
pacemaker performance, and/or overall reduced battery life.
● Install fresh LR6-sized (AA-sized) alkaline batteries.
● Install the batteries with proper polarity. The temporary pacemaker does not turn on or
provide pacing therapy with incorrect battery polarity.

4.3.2 Battery replacement recommendations


Replace the temporary pacemaker batteries in the following situations:
● Replace the batteries for each new patient.
● Replace the batteries when the low battery indicator flashes during temporary
pacemaker operation (see Section 3.2.2.2).
● Replace the batteries at least once every week when the temporary pacemaker is in
continuous use.

Note: When the low battery indicator flashes, the temporary pacemaker maintains pacing at
the current settings for a minimum of 24 hours, if the settings were at nominal values (see
Section 7.1).

Caution: Medtronic does not recommend replacing the batteries while the temporary
pacemaker is turned on or actively pacing the patient. However, if during an emergency
situation the batteries must be replaced while the temporary pacemaker is in use, ensure
that the temporary pacemaker is locked before replacing the batteries. Pacing is maintained
at the current settings for 30 s, minimum, if the settings are at nominal values (see
Section 7.1).

Technical Manual 35
Medtronic 5392

4.3.3 Battery polarity


The temporary pacemaker requires proper battery polarity for operation. Ensure that the
batteries align with the polarity markings on the inside of the battery drawer.
The temporary pacemaker does not turn on when batteries are installed with incorrect
polarity. If the batteries are replaced while the temporary pacemaker is turned on and the
battery polarity is incorrect, the temporary pacemaker continues to pace and sense until
insufficient power is available. The low battery indicator continues to flash, and pacing is
maintained at the current settings for 30 s minimum, with the settings at nominal values.
When internal reserve power is depleted, the temporary pacemaker shuts down.
When the new, fresh batteries are installed with proper polarity in the temporary pacemaker,
the following occurs:
● The low battery indicator stops flashing.
● The battery status indicator displays full battery power.
When the batteries are installed with incorrect polarity, the following occurs:
● If the temporary pacemaker is turned off, it does not turn on.
● If the temporary pacemaker is turned on, the low battery indicator continues to flash. The
temporary pacemaker continues to pace and sense until internal reserve power is
depleted. When internal reserve power is depleted, the temporary pacemaker shuts
down.

4.4 Battery installation and replacement


To install (or replace) the batteries, do the following:
1. Press the battery drawer latch release button until the battery drawer opens (see
Figure 18).
Figure 18. Battery drawer latch release button

36 Technical Manual
Medtronic 5392

2. Remove the old batteries.


3. Install two new LR6-sized (AA-sized) alkaline batteries. Verify that the batteries align
with the polarity markings on the inside of the battery drawer (see Figure 19).
Figure 19. Installing the batteries with proper polarity

Caution: Ensure that the new, fresh batteries are installed with the correct battery
polarity by verifying that the batteries align with the polarity markings on the inside of the
battery drawer. The temporary pacemaker requires proper battery polarity for
operation. After installing the batteries, ensure that the battery status indicator displays
full battery power and that the low battery indicator is not flashing. The temporary
pacemaker may temporarily continue to pace and sense with weak, dead, or incorrectly
installed batteries.

4. Close the battery drawer firmly until the battery drawer is fully latched.
Note: Failure to close the battery drawer completely can result in the battery drawer
opening and the temporary pacemaker shutting down.

5. Discard the old batteries properly according to local regulations.

Technical Manual 37
Medtronic 5392

4.5 Connector Setup


4.5.1 Using the cables with the temporary pacemaker
Before connecting the cables to the temporary pacemaker, verify the following:
● The reusable cables are supplied non-sterile. Clean and sterilize them before each use.
Refer to the applicable technical manual for cleaning, disinfecting, and sterilizing
instructions.
● Carefully inspect the reusable cables for visible signs of wear or damage before
connecting them to the temporary pacemaker. Do not use the reusable cables if they
appear damaged. Damage includes, but is not limited to, deterioration of the cable
insulation, brittleness, cracking, thinning, or bare spots. Do not use the reusable cables
if the conductive wires are exposed.
● Do not connect the temporary pacemaker to the lead system if it is turned on and is
operating at an output amplitude that could cause capture.

Warnings:
● Before connecting the cables to the temporary pacemaker, verify that it is turned off.
● Connect the cables to the temporary pacemaker before connecting the leads to the
cables.
● To prevent pacing into the vulnerable period of the T-wave, turn on the temporary
pacemaker and turn down A OUTPUT and V OUTPUT to the minimum amplitude before
connecting the temporary pacemaker to the patient’s lead system. Determine sensing
thresholds before turning up A OUTPUT and V OUTPUT to threshold levels.

Cautions:
● When mechanical support is necessary, hang the temporary pacemaker by the IV pole
hanger from an IV pole. Do not hang the temporary pacemaker to an IV pole by the
cables.
● Avoid contaminating areas that are difficult to clean on the temporary pacemaker. Keep
hands and gloves free of blood and body fluids when connecting or disconnecting the
patient cables, surgical cables, and/or pacing leads to the temporary pacemaker.

38 Technical Manual
Medtronic 5392

4.5.2 Connecting the cables to the temporary pacemaker


To connect the cables to the temporary pacemaker, do the following:
1. Verify that the temporary pacemaker is turned off.
2. Plug the patient cables or a pair of surgical cables into appropriate sockets on the
connector block on top of the temporary pacemaker. One socket is marked A (atrium);
the other is marked V (ventricle).
3. Verify that each cable clicks when it is inserted into the temporary pacemaker
connector receptacle (see Figure 20).
Note: The audible click verifies that the plug is completely inserted into the receptacle.

Figure 20. Connecting the cables to the temporary pacemaker

4. To ensure a good connection, pull gently on the cables after insertion.


5. Connect the leads to the appropriate cable. Match positive (+) and negative (–) leads
to positive (+) and negative (–) sockets or clips for the atrium and ventricle (not shown).

Refer to the applicable patient cable or surgical cable technical manual for more information.

4.5.3 Disconnecting the cables from the temporary pacemaker


To disconnect the cables from the temporary pacemaker, do the following:
1. Set the controls of the temporary pacemaker to allow the patient’s intrinsic rhythm to
take over pacing. See Section 5.1.3.
2. Press the connector release button on the cable plug (see Figure 21).
3. Gently pull the plug from the receptacle.

Technical Manual 39
Medtronic 5392

Figure 21. Disconnecting the cable from the temporary pacemaker

Refer to the applicable patient cable or surgical cable technical manual for more information.

4.5.4 Connecting the pacing lead system to the patient cables


Caution: Unipolar lead systems are not recommended because they are more susceptible
to electromagnetic interference, which may result in inappropriate pacing. Unipolar lead
systems should not be used in the dual-chambered pacing modes because the current path
of one lead system may interfere with the current path of the other.
To connect the pacing lead system to the patient cables, do the following:
1. Loosen the patient cable connector knobs by twisting each knob counterclockwise until
resistance is felt.
2. Insert the lead connector pins into the patient cable receptacles as shown (see
Figure 22).
3. Verify that each lead system is connected to the appropriate patient cable (atrial or
ventricular).
4. Rotate each patient cable connector knob clockwise until finger tight.
5. Gently pull on each lead conductor to verify secure connection.

40 Technical Manual
Medtronic 5392

Figure 22. Connecting the pacing lead system to the patient cable receptacles

Refer to the applicable patient cable technical manual for more information.

4.5.4.1 For bipolar systems


Insert each connector pin into the appropriate receptacle (marked + and –). Bipolar lead
systems may exhibit different threshold values depending on the polarity of the lead
connections.
Refer to the applicable patient cable technical manual for more information.

4.5.4.2 For single chamber unipolar systems (1-lead systems)


To connect single chamber unipolar lead systems to the temporary pacemaker, perform the
following:
1. Insert the cardiac lead connector pin into the negative (–) receptacle of the cable.
2. Insert the connector pin of the “indifferent” electrode (or “ground”) into the positive (+)
receptacle of the cable.
Refer to the applicable patient cable technical manual for more information.

Technical Manual 41
Medtronic 5392

4.5.4.3 For dual chamber unipolar systems (2-lead systems)


To connect dual chamber unipolar lead systems to the temporary pacemaker, perform the
following:
1. Insert the connector pin of each cardiac lead into the negative (–) receptacle of the
appropriate cable.
2. Insert the connector pin of the indifferent electrode into the positive (+) receptacle of
one patient cable.
3. Connect the indifferent electrode to the positive receptacle of the second patient cable
with a jumper cable.
Note: The temporary pacemaker will not pace or sense in the chamber if the positive
receptacle of the second patient cable is not connected with a jumper cable.

Refer to the applicable patient cable technical manual for more information.

4.6 Placement during use


When the temporary pacemaker is in use, place it in an area that reduces potential
unauthorized access from patient interaction or tampering by non-medical personnel.
To reduce potential unauthorized access to the temporary pacemaker when it is in use, do
one or more of the following:
● Verify that the temporary pacemaker is directly observable by medical staff.
● Secure the Medtronic Model 53922 Disposable Cover over the temporary pacemaker to
prevent inadvertent adjustment of the pacing modes or pacing parameters.
● Hang the temporary pacemaker by either the IV pole hanger or the attachment panel of
the disposable pouch to an IV pole.

Caution: Tampering with programmed parameters may have direct and serious patient
health effects.

42 Technical Manual
Medtronic 5392

5 Instructions for use

5.1 Basic operation


5.1.1 Turning on or turning off the temporary pacemaker
To turn on the temporary pacemaker, press and hold the On/Off key momentarily.
When the temporary pacemaker turns on, the following occurs: a self-test is initiated. When
the self-test completes successfully, the temporary pacemaker first searches for cardiac
activity (during the first pacing cycle), and then begins sensing and pacing in both chambers
(DDD pacing mode).
● The upper screen and the backlight illuminate.
● A self-test is initiated (see Section 9.1.1).
● The temporary pacemaker first searches for cardiac activity (during the first pacing
cycle), and then begins sensing and pacing in both chambers (DDD pacing mode).

Warning: To prevent pacing into the vulnerable period of the T-wave, turn on the temporary
pacemaker and turn down A OUTPUT and V OUTPUT to the minimum amplitude before
connecting the temporary pacemaker to the patient’s lead system. Determine sensing
thresholds before turning up A OUTPUT and V OUTPUT to threshold levels.
To turn off the temporary pacemaker, do the following:
1. Unlock the temporary pacemaker if it is locked (see Section 5.1.2).
2. Press the On/Off key once. A message is displayed in the lower screen to confirm
temporary pacemaker shutdown (see Figure 23).
3. Press the Enter key once within 30 s to confirm temporary pacemaker shutdown.

Note: If the Enter key is not pressed, the temporary pacemaker remains on and continues
to pace at the currently selected values.

Technical Manual 43
Medtronic 5392

Figure 23. Temporary pacemaker shutdown message

Notes:
● See Section 7.1 for nominal values when the temporary pacemaker is turned on.
● If the temporary pacemaker fails the self-test, it remains on but does not pace.
● If the batteries are nearing depletion, a red backlight begins flashing behind the battery
status indicator, indicating that the batteries have approximately 24 hours or less of
battery life. If the batteries are depleted, the LEDs may flash momentarily (see
Section 3.2.1) when the On/Off key is pressed, but the temporary pacemaker does not
operate.

5.1.2 Lock/Unlock
The Lock/Unlock key locks the temporary pacemaker to prevent inadvertent adjustment of
the parameters, or unlocks the temporary pacemaker when it is locked (see Section 3.1.7).

Note: The DOO/Emergency key does not lock.


The temporary pacemaker locks when one of the following occurs:
● 60 s elapses after the last parameter adjustment is made
● When the Lock/Unlock key is pressed
When the temporary pacemaker locks, the following occurs:
● The RATE, A OUTPUT, and V OUTPUT parameter values lock and cannot be adjusted.

● Pacing therapy continues to be delivered at the currently selected values.


● The Lock indicator appears in the upper screen.

44 Technical Manual
Medtronic 5392

● The lower screen does not appear. The Mode Selection options and pacing parameters
cannot be adjusted.
● The Pause key is locked.

Notes:
● If any parameter dials are adjusted or any keys are pressed while the temporary
pacemaker is locked (other than the DOO/Emergency key), the Lock indicator flashes,
and the lower screen displays the Locked message for approximately 30 s (see
Figure 24).
● If the DOO/Emergency key is pressed while the temporary pacemaker is locked, the
temporary pacemaker begins asynchronous pacing (see Section 5.1.9).
Figure 24. Locked message

Press the Lock/Unlock key to unlock the temporary pacemaker if it is locked.


When the temporary pacemaker unlocks, the following occurs:
● The Lock indicator disappears.
● Pacing therapy continues to be delivered at the currently selected values.
● The RATE, A OUTPUT, and V OUTPUT pacing parameters unlock and can be
adjusted.
● The lower screen appears. The Mode Selection options and pacing parameters can be
adjusted.
● The Pause key is unlocked.

Technical Manual 45
Medtronic 5392

5.1.3 Viewing the patient’s intrinsic rhythm


There are two methods for viewing the patient’s intrinsic rhythm.
Recommended method – Reduce the RATE gradually, while watching the ECG, until the
patient’s intrinsic rhythm takes over pacing.
Pause key method – Press and hold the Pause key and watch the ECG. The patient’s
intrinsic rhythm is viewable on the ECG.

5.1.3.1 Pausing pacing and sensing


Press and hold the Pause key to suspend pacing for up to 10 s (see Section 3.1.11). For the
safety of the patient, when the Pause key is pressed and held, pacing is only suspended for
10 s before timing out.

Caution: Use the Pause key with care. The patient receives no pacing support (for a
maximum of 10 s at a time) when the Pause key is pressed and held.
When the Pause key is pressed and held, the message shown in Figure 25 appears in the
lower screen.
Figure 25. Caution: Pacing is suspended message

When the Pause key is released or pause times out at 10 s, the temporary pacemaker
resumes pacing at the programmed parameters.

Note: To pause again (up to 10 s), release the Pause key, and then press and hold the Pause
key again (this is a safety feature).

46 Technical Manual
Medtronic 5392

5.1.4 RATE and OUTPUT adjustments


Use the dials next to the upper screen to adjust the pacing rate, atrial output, and ventricular
output. The upper screen displays a numerical value and segmented circular scale that
reflects the current setting for each dial. See Figure 26.
Figure 26. RATE and OUTPUT
A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

To adjust RATE, A (Atrial) OUTPUT, or V (Ventricular) OUTPUT, turn the RATE,


A OUTPUT, or V OUTPUT dials clockwise to increase their values; turn the dials
counterclockwise to decrease their values, or to set the outputs to off.
See Section 7.1 for RATE, A OUTPUT, or V OUTPUT ranges.

5.1.5 Pacing setup


The temporary pacemaker can be set to a single chamber pacing mode (AOO, VOO, AAI,
VVI), a dual chamber pacing mode (DDD, DDI, DOO), or no pacing mode (OOO).
Select the appropriate pacing mode from the Mode Selection menu. The pacing mode can
also be set by adjusting A OUTPUT or V OUTPUT from the upper screen, and A Sensitivity,
V Sensitivity and/or A. Tracking from the Pacing Parameters menu.
The pacing information tables (see Section 7.2) provide a quick reference for selecting a
pacing mode. Refer to “Basic operation” (see Section 5.1) and “Pacing parameter
adjustments” (see Section 5.3) for step-by-step instructions on how to adjust output and
sensitivity.

5.1.6 Mode Selection menu


The Mode Selection menu (see Figure 27) is used to select pacing modes or to navigate to
the Pacing Parameters menu (see Section 5.3). Use the Up and Down arrow keys to select
dual chamber pacing modes (DDD, DDI, and DOO), single chamber pacing modes (AAI,
AOO, VVI, and VOO), no pacing mode (OOO), or the Pacing Parameters menu (Back).

Technical Manual 47
Medtronic 5392

Figure 27. Mode Selection menu

The Pacing Mode indicator in the upper screen displays the selected mode. The new mode
is activated within the next two cardiac cycles.
If no timing violations occur, the new pacing mode retains the current setting of all applicable
parameters of the previous pacing mode. If timing violations occur, the rate-dependent
parameters are set to automatic values (see Section 5.4).

Caution: Do not configure the temporary pacemaker in dual chamber mode unless both
channels are connected to the heart. If the temporary pacemaker is configured in a dual
chamber mode and one of the channels is not connected, the open channel could pick up
unintended noise. The noise can be interpreted as a sensed event and can lead to events
such as, but not limited to, asynchronous ventricular pacing, pacemaker mediated
tachycardia, or ventricular tachycardia.

5.1.6.1 Pacing mode selection


To select a pacing mode, perform these steps:
1. Navigate to the Mode Selection menu.
2. Press the Up or Down Arrow keys to highlight a pacing mode.
3. Press the Enter key to select the pacing mode.
RATE, OUTPUT values, and Sensitivity values are set to the nominal values when a pacing
mode is selected unless they have been manually adjusted before the pacing mode was
selected. If they have been manually adjusted before the pacing mode was selected, the
new pacing mode retains these values.
For example, if you change from AAI to DDD pacing mode, the value for A OUTPUT is
retained; V OUTPUT is set to the nominal value.

48 Technical Manual
Medtronic 5392

Note: Manually set pacing parameter values are not retained when the temporary
pacemaker is turned off and then turned back on.

5.1.7 Synchronous (demand) pacing


During synchronous (demand) pacing, output is inhibited when the pacemaker senses
intrinsic activity to minimize competition between the paced rhythm and the intrinsic activity
of the heart.

Note: Determine sensitivity and stimulation thresholds (see Section 5.2); otherwise
asynchronous pacing, and/or loss of heart capture may occur.
Select one of the pacing modes in Table 1 from the Mode Selection menu to initiate demand
pacing.
Table 1. Pacing modes with demand pacing
Synchronous (demand) pac-
ing type Pacing mode Result
Dual chamber DDD, DDI Pacing and sensing occurs in
both chambers.
Atrium AAI Pacing and sensing occur only
in the atrium. No pacing or
sensing occurs in the ventricle.
Ventricle VVI Pacing and sensing occur only
in the ventricle. No pacing or
sensing occurs in the atrium.

5.1.8 Asynchronous pacing


Patients best suited for asynchronous (non-sensing) modes have one of the following
issues:
● An intrinsic rate consistently below the pacing rate.
● No intrinsic activity.

Caution: Because it may compete with the intrinsic activity of the heart, asynchronous
pacing may result in tachyarrhythmia. Use caution when setting the device to asynchronous
modes.
Initiate asynchronous pacing by pressing the DOO/Emergency key or selecting an
asynchronous pacing mode from the Mode Selection menu.
Select one of the pacing modes in Table 2 from the Mode Selection menu to initiate
asynchronous pacing.

Technical Manual 49
Medtronic 5392

Table 2. Asynchronous pacing types and modes


Asynchronous pacing type Pacing mode Result
Dual chamber DOO Pacing occurs in the atrium and ventricle. No
sensing occurs.
Adjust A OUTPUT and V OUTPUT to provide
adequate safety margins to ensure capture
(see Section 5.2.3.2).
Atrium AOO Pacing occurs only in the atrium. No sensing
occurs.
Adjust A OUTPUT to provide an adequate
safety margin (see Section 5.2.3.2).
Ventricle VOO Pacing occurs only in the ventricle. No sens-
ing occurs.
Adjust V OUTPUT to provide an adequate
safety margin (see Section 5.2.3.2).

Note: Determine the patient’s stimulation threshold; otherwise loss of heart capture may
occur (see Section 5.2.3.1).

5.1.8.1 Terminating asynchronous pacing


To terminate asynchronous pacing and to return to synchronous (demand) pacing, perform
one of the following actions:
● Press the Enter key, if the DOO/Emergency key was pressed (see Section 5.1.9.2).
● Select a pacing mode with synchronous (demand) pacing from the Mode Selection
menu.

5.1.9 Emergency pacing


The emergency pacing feature is used to select high-output, dual-chamber asynchronous
pacing (DOO for emergency). See Section 7.1 for emergency pacing values.

5.1.9.1 Initiating emergency pacing


Press the DOO/Emergency key to initiate emergency pacing (e.g., dual chamber
asynchronous pacing) immediately at maximum output levels (see Section 3.1.2).
Emergency pacing is initiated whether the temporary pacemaker is turned on or off, locked,
or unlocked.
When the DOO/Emergency key is pressed, the Asynchronous pacing message is displayed
in the lower screen (see Figure 28).

50 Technical Manual
Medtronic 5392

Figure 28. Asynchronous pacing message

Adjust the RATE, A OUTPUT, and V OUTPUT by turning the 3 upper dials clockwise or
counterclockwise.

5.1.9.2 Terminating emergency pacing


Press the Enter key to terminate emergency pacing and to resume demand pacing (e.g.,
dual chamber synchronous pacing).
When emergency pacing is terminated and synchronous pacing resumes, the following
occurs:
● The RATE value remains at the current setting, or at 80 ppm if the temporary pacemaker
was turned off before the DOO/Emergency key was pressed.
● A OUTPUT, and V OUTPUT values remain at the emergency pacing values or at the
values set during emergency pacing (if the values were manually adjusted).
● A SENSITIVITY and V SENSITIVITY return to their nominal values.
● The temporary pacemaker is set to DDD pacing mode, unless A OUTPUT or V OUTPUT
have been turned off during emergency pacing. If A OUTPUT has been turned off during
emergency pacing, the temporary pacemaker is set to VVI pacing mode after
emergency pacing is terminated. If V OUTPUT has been turned off during emergency
pacing, the temporary pacemaker is set to AAI pacing mode after emergency pacing is
terminated.

Technical Manual 51
Medtronic 5392

5.2 Thresholds
Threshold values are needed to determine the appropriate settings for output and
sensitivity. Procedures for finding atrial and ventricular sensing and stimulation thresholds
are described in this section.

Note: To reduce the risk of competitive pacing, determine the sensing thresholds first, if the
patient’s intrinsic rhythm is adequate.

5.2.1 Sensing definitions


The ECG in Figure 29 shows the intrinsic beats mixed with paced beats. The temporary
pacemaker detects the heart’s own beat and does not deliver a pacing stimulus.
Figure 29. Sensing

200 ms

The ECG in Figure 30 shows one example of undersensing. The temporary pacemaker does
not detect intrinsic activity and thus paces on or between beats.
Figure 30. Atrial undersensing

200 ms

5.2.1.1 Sensing threshold


The sensing threshold is the least sensitive mV setting at which the temporary pacemaker
can detect a heartbeat. Monitor the patient’s ECG and blood pressure as you follow the
procedure to determine the atrial and ventricular sensing thresholds.

52 Technical Manual
Medtronic 5392

5.2.1.2 Automatic adjustment of sensitivity threshold


The temporary pacemaker automatically adjusts the sensing thresholds after certain paced
and sensed events to help reduce the oversensing of T-waves, cross-chamber events, and
pacing. The threshold adjustment depends on the type of event that precedes the
adjustment. During an automatic adjustment, the sensing threshold automatically
increases, but it gradually decreases toward the programmed sensitivity value, which is the
minimum amplitude that can be sensed. The threshold decrease is intended to be rapid
enough to allow subsequent signals to be sensed. An illustration of the sensing threshold
adjustment is shown in Figure 31 for nominal values.
Figure 31. Adjustment of sensing thresholds

A. Sense EGM
AS AP AS

Marker Channel
VS VS VP

V. Sense EGM

Sensing threshold

1 After an atrial sensed event, the temporary pacemaker is temporarily less sensitive to atrial
events.
2 After a ventricular sensed event, the temporary pacemaker is temporarily less sensitive to
ventricular events.
3 After an atrial paced event, the temporary pacemaker is temporarily less sensitive to ventricular
events, but the sensitivity to atrial events remains the same.
4 After a ventricular paced event, the temporary pacemaker is temporarily less sensitive to atrial
events.
5 After the post-pace blanking period, the device is temporarily less sensitive to ventricular events.
6 After the post-pace blanking period, the temporary pacemaker is temporarily less sensitive to
atrial events.

Note: Automatic adjustment of sensitivity thresholds is limited to certain sensitivity settings.

Technical Manual 53
Medtronic 5392

5.2.1.3 Sensing threshold safety margin


Lead maturation and drug therapy can affect the sensing threshold. To ensure sensing and
accommodate a changing threshold, it is important to provide at least a 2:1 safety margin.
Set A Sensitivity and V Sensitivity to values that are at least one-half to one-third of the
sensing threshold values. For example, an appropriate setting for a patient with a 5.0 mV
sensing threshold is 2.5 mV or less.

Caution: Avoid selecting a pacing mode that requires sensing if adequate sensing margins
cannot be established. A 2:1 safety margin cannot always be achieved due to a low sensing
amplitude and/or a high pacing amplitude. If a 2:1 safety margin cannot be achieved,
increase monitoring of the patient to verify that the expected therapy is being delivered.

5.2.2 Determining atrial or ventricular sensing thresholds


Caution: Pacing-dependent patients have limited or no intrinsic rate or rhythm. Only use this
procedure on patients with adequate intrinsic rhythm.
To determine atrial or ventricular sensing threshold, perform the following steps:
1. Turn on the temporary pacemaker without connecting it to the patient lead system.
Caution: Do not connect the temporary pacemaker to the patient lead system until step
4.

2. Set RATE to at least 10 ppm under the patient’s intrinsic rate. If necessary, continue to
reduce the RATE until the temporary pacemaker is not pacing the patient (see
Figure 32).
3. Adjust the atrial or ventricular output to prevent the risk of competitive pacing (see
Figure 32), as follows:
● Atrial: Set A OUTPUT to 0.1 mA.
● Ventricular: Set V OUTPUT to 0.1 mA.
Figure 32. Decrease RATE, A OUTPUT, and V OUTPUT
A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

54 Technical Manual
Medtronic 5392

4. Connect the temporary pacemaker to the patient lead system.


5. Navigate to the Mode Selection menu and select the appropriate pacing mode for the
patient leads that are connected to the patient.
● Select DDD pacing mode if both channels are connected.
● Select AAI pacing mode if only the atrial channel is connected.
● Select VVI pacing mode if only the ventricular channel is connected.
6. Navigate to the Sensitivity settings.
a. If the atrial channel is connected, complete steps 7 thru 9 for the A Sensitivity
setting.
b. If the ventricular channel is connected, complete steps 7 thru 9 for the V Sensitivity
setting.
7. Decrease Sensitivity: Slowly turn the Menu Parameter dial counterclockwise
(increase mV value) until the SENSE indicator stops flashing (see Figure 33).
The PACE indicator flashes continuously, but capture is not likely because the
OUTPUT value is set to minimum.
Figure 33. Decrease sensitivity

8. Increase Sensitivity: Slowly turn the Menu Parameter dial clockwise (decrease mV
value) until the SENSE indicator starts flashing (see Figure 34). The following occurs:
● The PACE indicator stops flashing.
● This value is the sensing threshold.

Technical Manual 55
Medtronic 5392

Figure 34. Increase sensitivity

9. Set Sensitivity to half (or less) the threshold value. This setting provides at least a 2:1
safety margin.
10. Restore RATE, A OUTPUT, or V OUTPUT to previous values.

Note: Determine the atrial or ventricular stimulation thresholds after determining sensing
thresholds.

5.2.3 Capture definitions


When a pacing pulse captures the heart, it causes the heart to beat — that is, contract and
pump blood. The ECG shows a P-wave or QRS complex after the pulse, as in the example
shown in Figure 35.
Figure 35. Capture

200 ms

When capture is lost, the ECG shows no heart response after the pulse, as in the example
shown in Figure 36.

56 Technical Manual
Medtronic 5392

Figure 36. Loss of ventricular capture

200 ms

5.2.3.1 Stimulation threshold


The stimulation threshold is the minimum output (mA) needed to consistently capture the
heart. Monitor the patient’s ECG and blood pressure as you follow the procedure to find the
atrial and ventricular stimulation thresholds.

5.2.3.2 Stimulation threshold safety margin


Lead maturation and drug therapy can affect the stimulation threshold. To achieve
consistent capture and accommodate a changing threshold, it is important to provide at least
a 2:1 safety margin. Set A OUTPUT and V OUTPUT to a value at least 2 to 3 times greater
than the stimulation threshold value. For example, the appropriate output setting for a patient
with a 1.0 mA threshold is 2.0 mA or greater.

Caution: A 2:1 safety margin cannot always be achieved due to a very high pacing
amplitude. If a 2:1 safety margin cannot be achieved, increase monitoring of the patient to
verify that the expected therapy is being delivered.

5.2.4 Determining atrial or ventricular stimulation thresholds


To determine atrial or ventricular stimulation thresholds, perform these steps:
1. Verify that the patient is connected to the temporary pacemaker and is being monitored
on the ECG.
2. Set RATE at least 10 ppm above the patient’s intrinsic rate (see Figure 37).
If necessary, continue to increase the RATE until the temporary pacemaker is pacing
the patient. The PACE indicator flashes.
3. Decrease OUTPUT: Slowly turn the OUTPUT dial counterclockwise until the ECG
shows loss of capture (see Figure 37).
The PACE and SENSE indicators flash intermittently.

Technical Manual 57
Medtronic 5392

Figure 37. Increase RATE, decrease A OUTPUT or V OUTPUT


A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

4. Increase OUTPUT: Slowly turn the output dial clockwise until ECG shows consistent
capture (see Figure 38). The following occurs:
● The PACE indicator flashes continuously; the SENSE indicator stops flashing.
● This value is the stimulation threshold.
Figure 38. Increase OUTPUT
A V
PACE SENSE PACE SENSE

RATE

30 200 ppm
A OUTPUT

0 20 mA
V OUTPUT

0 25 mA

5. Set OUTPUT to a value at least 2 to 3 times greater than the stimulation threshold value.
This setting provides at least a 2:1 safety margin.
6. Restore RATE to the previous value.

5.3 Pacing parameter adjustments


The Mode Selection menu is used to select different pacing modes. After a pacing mode is
selected, the Pacing Parameters menu for that pacing mode is displayed and the
parameters can be adjusted (see Figure 39).

58 Technical Manual
Medtronic 5392

The Pacing Parameters menu displays the parameters appropriate for the current pacing
mode and allows you to change their values. Depending on the pacing mode, a subset of
pacing parameters is displayed. Parameters that do not apply to the current chambers being
paced and sensed are not displayed. Parameters that did not apply in the previous mode are
set to nominal values in the new mode (see Section 5.6.7).
The parameters displayed in a pacing mode are based on the currently programmed pacing
mode and rate.
To adjust pacing and sensing parameters, perform the following steps:
1. Select a pacing mode from the Mode Selection menu by pressing the Up or Down
Arrow key.
2. Press the Enter key. The pacing parameters for the selected pacing mode are
displayed.
3. Press the Up or Down Arrow key to scroll through and select a pacing or sensing
parameter.
4. Turn the Menu Parameter dial clockwise or counterclockwise to adjust the value of the
selected parameter.

Note: Any value changes to the pacing parameters are not retained when the temporary
pacemaker is turned off. All pacing parameters are set to default values when the temporary
pacemaker is turned on.

Technical Manual 59
Medtronic 5392

Figure 39. Lower screen pacing parameters

1 Mode Selection menu 3 V Sensitivity selected on the DDD Pacing


2 Upper Rate selected on the DDD Pacing Parameters menu
Parameters menu 4 RAP screen

5.3.1 Sensitivity
5.3.1.1 A (Atrial) Sensitivity
Unless manually adjusted, the atrial sensitivity (A Sensitivity) is set to the nominal value of
0.5 mV. A Sensitivity can be adjusted between 0.4 and 10 mV by turning the Menu
Parameter dial, as follows:
● To increase A Sensitivity, turn the dial clockwise (the mV value decreases).
● To decrease A Sensitivity, turn the dial counterclockwise (the mV value increases).
The change takes effect within the next two pacing cycles.

60 Technical Manual
Medtronic 5392

To turn off A Sensitivity to allow the temporary pacemaker to pace asynchronously in the
atrium, turn the Menu Parameter dial counterclockwise until the term ASYNC appears (see
Figure 40).

Note: When A Sensitivity is turned off, a pacing mode change occurs.


Figure 40. A Sensitivity field

The pacing and sensing status bar indicators at the top of the upper screen reflect this
change. The SENSE indicator under A SENSE LED is not displayed, and the A SENSE LED
no longer flashes.
A Sensitivity is automatically set to ASYNC when the following occur:
● The DOO/Emergency key is pressed.
● RAP is delivered.
● DOO pacing mode or AOO pacing mode is selected from the Mode Selection menu.

Note: A Sensitivity is inaccessible when A OUTPUT is turned off or in VVI or VOO pacing
modes.

5.3.1.2 V (Ventricular) Sensitivity


Unless manually adjusted, ventricular sensitivity (V Sensitivity) is set to the nominal value
of 2.0 mV.
When selected, the sensitivity can be adjusted between 0.8 and 20 mV by turning the Menu
Parameter dial as follows:
● To increase V Sensitivity (the mV value decreases), turn the dial clockwise.
● To decrease V Sensitivity (the mV value increases), turn the dial counterclockwise.

Technical Manual 61
Medtronic 5392

The change takes effect within the next two pacing cycles.
To turn off V Sensitivity to allow the temporary pacemaker to pace asynchronously in the
ventricle, turn the Menu Parameter dial counterclockwise until the term ASYNC appears
(see Figure 41).

Note: When V Sensitivity is turned off, a pacing mode change occurs.


Figure 41. V Sensitivity field

The pacing and sensing status bar indicators at the top of the upper screen reflect the
change. The SENSE indicator under the V SENSE LED is not displayed, and the V SENSE
LED no longer flashes.
V Sensitivity is automatically set to ASYNC when the following occur:
● The DOO/Emergency key is pressed.
● DOO pacing mode or VOO pacing mode is selected from the Mode Selection menu.

Note: V Sensitivity is inaccessible when V OUTPUT is turned off or in AAI or AOO pacing
modes.

5.3.2 Rate-dependent parameters


A-V Interval, Upper Rate, and PVARP are rate-dependent parameters. When not
manually adjusted, these parameters are automatically set to a factor of the RATE (see
Section 7.1).

62 Technical Manual
Medtronic 5392

When manually set, the rate-dependent parameters do not change with adjustments to the
RATE. However, an increase in the RATE can cause a timing violation with the Upper Rate.
Also, a decrease of the RATE can cause a timing violation with the A-V Interval or PVARP.
If adjustments to RATE cause a timing violation, a warning message appears on the lower
screen. The RATE cannot be increased until Upper Rate is increased or A-V Interval
and/or PVARP are adjusted (see Section 5.4).

5.3.2.1 A-V interval


A-V Interval may be adjusted from 20 to 300 ms in increments of 10 ms.
The A-V Interval after an atrial pace (that is, paced atrioventricular interval [PAV], or
A-V Interval) is the amount of time, in ms, that the temporary pacemaker waits between the
delivery of an atrial pacing pulse and delivery of the corresponding ventricular pacing pulse.

Notes:
● The A-V Interval after an atrial sensed event [that is, sensed atrioventricular interval
(SAV)] is not programmable. The SAV is automatically set to a value 30 ms less than the
A-V Interval in DDD pacing mode. In DDI pacing mode, SAV = PAV.
● Unless manually adjusted, A-V Interval is set to a value determined by the RATE
setting. It can never be shorter than 50 ms or longer than 250 ms.

Warning: If A-V Interval is set shorter than 50 ms, ventricular events may not be sensed
during that interval, due to ventricular blanking after an atrial event.

5.3.2.2 Adjusting A-V interval


To adjust A-V Interval, perform the following steps:
1. Navigate to the Pacing Parameters menu.
2. Press the Up or Down Arrow key to select A-V Interval (see Figure 42).
3. Turn the Menu Parameter dial clockwise to lengthen the A-V Interval, or
counterclockwise to shorten the A-V Interval.

Technical Manual 63
Medtronic 5392

Figure 42. A-V Interval field

5.3.2.3 Upper Rate


The Upper Rate parameter sets the maximum ventricular pacing rate allowed while tracking
the atrium. At sensed atrial rates above the Upper Rate, a high rate response results (see
Section 5.6.6).

Note: The Upper Rate parameter is only adjustable in DDD pacing mode.
Unless manually adjusted, this parameter is limited to the range of 110 to 230 ppm. When
selected on the Pacing Parameters menu, however, the Upper Rate can be adjusted to
values from 80 to 230 ppm. See Section 7.1.

5.3.2.4 Adjusting Upper Rate


To adjust the Upper Rate, perform the following steps:
1. Navigate to the Pacing Parameters menu.
2. Press the Up or Down Arrow key to select Upper Rate (see Figure 43).
3. Turn the Menu Parameter dial clockwise to increase the Upper Rate, or
counterclockwise to decrease the Upper Rate.

64 Technical Manual
Medtronic 5392

Figure 43. Upper Rate field

5.3.2.5 Post Ventricular Atrial Refractory Period (PVARP)


The PVARP parameter sets the length of time following a ventricular event during which
atrial sensing does not affect pacemaker timing. PVARP is designed to prevent the
temporary pacemaker from responding to atrial sensing of ventricular activities such as
far-field R-waves, premature ventricular contractions (PVCs), and retrograde conduction.
Unless manually adjusted, this parameter is determined by the RATE. See Section 7.1.
The PVARP parameter can be manually adjusted from a minimum value of 150 ms to a
maximum value of 500 ms in 10 ms increments so long as the value does not cause a timing
violation (see Section 5.4).

Note: If PVARP is set to the minimum value of 150 ms, atrial events may not be sensed due
to the tolerance allowed for blanking after a paced event (see Section 7.1).

5.3.2.6 Adjusting PVARP


To adjust PVARP, perform the following steps:
1. Navigate to the Pacing Parameters menu.
2. Press the Up or Down Arrow key to select PVARP (see Figure 44).
3. Turn the Menu Parameter dial clockwise to increase the PVARP, or counterclockwise
to decrease the PVARP.

Technical Manual 65
Medtronic 5392

Figure 44. PVARP field

5.3.3 A. Tracking
A. Tracking (atrial tracking) is only accessible or applicable when the temporary pacemaker
is set to sense and pace in both chambers. When A. Tracking is turned On, the temporary
pacemaker paces the ventricle in synchrony with intrinsic atrial depolarizations.
When A. Tracking is turned on (DDD pacing mode), each sensed event on the atrial lead
not only inhibits the scheduled atrial pacing pulse, but also triggers an A-V Interval.

Warning: If a patient is prone to atrial arrhythmias, atrial tracking could lead to the
development of ventricular arrhythmias (see Section 1.7).
When A. Tracking is off (DDI pacing mode), an atrial sense will inhibit an atrial pace, but it
does not trigger an A-V Interval (see Figure 45). The ventricle is paced at the selected
RATE.

66 Technical Manual
Medtronic 5392

Figure 45. A. Tracking field

5.3.3.1 Turning on or turning off A. Tracking


To turn A. TrackingOff or On, perform the following steps:
1. Navigate to the Pacing Parameters menu.
2. Press the Up or Down Arrow key until A. Tracking is selected.
3. Turn the Menu Parameter dial to change A. Tracking from On to Off. When
A. Tracking is changed from On to Off, the following occurs:
● DDI pacing and sensing begin.
● The DDI pacing mode indicator appears.
4. Turn the Menu Parameter dial to change A. Tracking from Off to On. When
A. Tracking is changed from Off to On, the following occurs:
● DDD pacing and sensing begin.
● The DDD pacing mode indicator appears.

Note: A. Tracking can be set to Off only from DDD pacing mode and set to On only from DDI
pacing mode.

Technical Manual 67
Medtronic 5392

5.3.4 Settings
The Settings parameter sets the A-V Interval, Upper Rate, and PVARP to automatic,
rate-dependent settings (see Section 7.1).
If A-V Interval, Upper Rate, or PVARP pacing parameters are manually adjusted (in a dual
chamber pacing mode), the Pacing Parameters menu displays the following changes:
● Manual* appears to the right of Settings (see Figure 46).
● An asterisk (*) appears next to the value of each setting that is manually adjusted.
Figure 46. Automatic/Manual Settings field

To change A-V Interval, Upper Rate, and PVARP back to automatic, rate-dependent
settings, perform these steps:
1. Navigate to the Pacing Parameters menu.
2. Select Settings.
3. Turn the Menu Parameter dial either clockwise or counterclockwise until Automatic
replaces Manual.
4. Use the Menu Parameter dial to alternate between Automatic and Manual parameter
settings, as long as Settings remains selected.
Previous Manual settings are lost when Automatic is selected.

Note: The rate-dependent parameter values are immediately updated on the screen and
take effect at the next appropriate event when switching Settings from Automatic to
Manual, or Manual to Automatic.

68 Technical Manual
Medtronic 5392

5.4 Timing violations


If the relationship between two or more parameters reaches a point where improper pacing
could occur, further change in the conflicting direction is prevented.
The following sections describe the timing violations.

5.4.1 Mode changes


During a pacing mode change, if one or more of the rate-dependent parameters was
manually set in the previous pacing mode to a value that causes improper pacing to occur in
the new mode, the temporary pacemaker changes the parameter to its automatic,
rate-dependent setting.

5.4.2 Parameter adjustments


During adjustments of the rate-dependent parameters, if the parameters are set to values
that violate timing rules (e.g., a block point), a warning message is displayed in the lower
screen for approximately 30 s, or until the Enter key is pressed. See Figure 47.
Figure 47. Pacemaker Block Point warning message

If the temporary pacemaker reaches a block point, press the Enter key and adjust one of the
following pacing parameters:
● RATE
● A-V Interval
● Upper Rate

Technical Manual 69
Medtronic 5392

● PVARP
● Select Settings and adjust the value to Automatic (see Section 5.3.4).
If no adjustments are made, the message disappears after approximately 30 s, and the
temporary pacemaker continues to operate at the value selected prior to reaching the block
point.

5.4.2.1 Upper Rate versus SAV and PVARP


The total atrial refractory period (TARP), which is SAV Interval + PVARP, cannot be longer
than the Upper Rate Interval, or the temporary pacemaker reaches the 2:1 block point before
being limited by the Upper Rate. This condition is prevented by the following formula:
SAV Interval + PVARP < Upper Rate Interval

Notes:
● DDD pacing mode: SAV Interval = AV Interval - 30 ms
● DDI pacing mode: SAV = PAV
During manual adjustment of RATE or the rate-dependent parameters, the temporary
pacemaker limits adjustment at the block point and displays the warning message shown in
Figure 47.

5.4.2.2 Minimum V-A interval


The V-A interval is the time period elapsing from a ventricular event (sensed or paced) to the
next scheduled atrial pace. The minimum V-A interval required by the temporary pacemaker
when operating in DOO pacing mode is 70 ms, as defined in the following formula:
A-V Interval + 70 ms ≤ RATE interval

Note: RATE interval = A-V Interval + V-A interval.


In DOO pacing mode, the temporary pacemaker limits adjustment of RATE, or the
rate-dependent parameters, at the block point. If the block point is reached, the lower screen
displays a warning message (see Figure 47).

Note: This limit only occurs in DOO pacing mode. In DDD and DDI pacing modes, the RATE
versus A-V Interval and PVARP block point maintains a 180 ms V-A interval (where V-A
interval = 150 ms minimum PVARP + 30 ms).

70 Technical Manual
Medtronic 5392

5.4.2.3 RATE versus A-V Interval and PVARP


Mode timing restrictions for atrial sensing are defined with the following formula:
A-V Interval + PVARP + 30 ms ≤ RATE Interval
If the RATE is increased after the rate-dependent parameters are adjusted manually, the
temporary pacemaker limits the increase of the RATE at the block point. The lower screen
displays a warning message (see Figure 47).

5.4.2.4 RATE versus Upper Rate


The RATE and Upper Rate can be set to the same value. However, the RATE cannot be set
higher than the Upper Rate, the Upper Rate cannot be set lower than the RATE, and the
Upper Rate cannot exceed the temporary pacemaker block point.
If any of these adjustments are attempted, the lower screen displays the warning message in
Figure 47.

5.5 Rapid Atrial Pacing (RAP)


RAP can be used to interrupt some types of atrial tachycardias or to induce an atrial
tachycardia.

Caution: RAP is for atrial use only. Before enabling RAP, be sure that the atrial leads are
connected to the atrium, not the ventricle.

5.5.1 Rapid Atrial Pacing (RAP)


To access the RAP screen, select Rapid Atrial Pacing from the Pacing Parameters menu.
The RAP screen displays the RAP rate. See Figure 48.

Technical Manual 71
Medtronic 5392

Figure 48. Rapid Atrial Pacing screen

When the Enter key is pressed and held, the temporary pacemaker waits a maximum of two
pacing cycles, and then begins pacing asynchronously in the atrium (AOO pacing mode) at
the selected RAP rate. See Figure 49.
Figure 49. Delivering rapid atrial pacing

Note: The temporary pacemaker does not deliver RAP until the Enter key is pressed and
held. RATE, A OUTPUT, V OUTPUT remain at their selected values for the current pacing
mode until the Enter key is pressed and held.

5.5.2 Delivering RAP


Caution: RAP may result in tachycardia, acceleration of existing tachycardia, or fibrillation.
Apply high rates under careful patient monitoring and control. Monitor the patient’s ECG and
blood pressure, and ensure that defibrillation equipment is immediately available.
To deliver RAP, perform the following steps:
1. Navigate to the Pacing Parameters menu.
2. Press the Up or Down Arrow key to highlight Rapid Atrial Pacing (RAP).

72 Technical Manual
Medtronic 5392

3. Verify that the leads are in contact with the atrium and are connected to the atrial
channel of the temporary pacemaker through a patient or surgical cable.
4. Press the Enter key to open the RAP screen.
The RAP screen displays the RAP rate (initially the rate of 250 ppm). Pacing continues
at currently displayed settings.
5. Adjust RAP rate as needed. Turn the Menu Parameter dial clockwise to increase rate,
or counterclockwise to decrease rate.
Note: The range for RAP is 80 ppm to 800 ppm.

6. Press and hold the Enter key to deliver RAP burst.


AOO pacing begins at displayed RAP rate and current A OUTPUT. The A PACE LED
flashes during delivery of RAP pulses.
Note: RAP delivery stops when either the Enter key is released, or after 2 min have
passed.

To exit the RAP screen, use the Up or Down Arrow key to highlight Back, and then press the
Enter key.
During the delivery of RAP, the following value settings occur:
● A OUTPUT does not lock and may be adjusted during RAP delivery.
● V OUTPUT is off and not accessible during RAP delivery. V OUTPUT returns to its
previous value when the Enter key is released.
● There is no ventricular support during RAP.
● If the Enter key is pressed while A OUTPUT is turned off, RAP is delivered at 10 mA.
A OUTPUT returns to off as soon as the Enter key is released, even if A OUTPUT is
adjusted during RAP delivery.

5.5.2.1 Adjusting rate or atrial output during RAP delivery


The RAP rate and A OUTPUT can be adjusted during RAP delivery by turning the
Menu Parameter dial. To adjust the RAP rate and A OUTPUT, do the following:
1. Continue to press and hold the Enter key.
2. Turn the Menu Parameter dial clockwise or counterclockwise to adjust RAP rate.
3. Turn the A OUTPUT dial clockwise or counterclockwise to adjust atrial output.

Technical Manual 73
Medtronic 5392

5.5.2.2 Resuming pacing at previous settings


Release the Enter key to resume pacing at the previous settings. The temporary pacemaker
stops delivering RAP and resumes operation at the non-RAP settings, within 3 s.
If the A OUTPUT is adjusted during RAP, the new setting is retained when RAP is
terminated.

Note: When resuming operation in a demand mode, the temporary pacemaker first
searches for cardiac activity (during the first pacing cycle), and then begins sensing and
pacing in both chambers (DDD pacing mode).

Caution: If the temporary pacemaker continues to deliver RAP after the Enter key is
released, press the On/Off key or the DOO/Emergency key to stop RAP. If RAP continues to
be delivered, remove the batteries from the temporary pacemaker. Return the temporary
pacemaker for service.

5.6 Automatic responses


5.6.1 Blanking periods
Blanking periods follow paced and sensed events and help to prevent the following from
occurring:
● Sensing pacing pulses
● Post-pacing depolarization
● T-waves
● Oversensing the same event
The blanking periods after paced events are longer than or equal to those after sensed
events to avoid sensing the atrial and ventricular depolarizations (see the specifications in
Section 7.1).

5.6.2 Refractory periods


A pacing refractory period is the interval during which a sensed event does not affect pacing
timing. During a pacing refractory period, the temporary pacemaker senses normally but
classifies sensed events as refractory and limits its response to these events.
A pacing refractory period prevents inappropriately sensed signals, such as far-field
R-waves or electrical noise, from triggering certain pacing timing intervals.

74 Technical Manual
Medtronic 5392

The temporary pacemaker has an Atrial Refractory Period (ARP), but does not have a
Ventricular Refractory Period (VRP). The initiation of these two types of pacing refractory
periods are described below:
Atrial Refractory Period (ARP) – The temporary pacemaker has two types of atrial
refractory periods:
● ARP, which is initiated by an atrial sense or pace
● Post Ventricular Atrial Refractory Period (PVARP), which is initiated by a ventricular
sense or pace
Ventricular Refractory Period (VRP) – A VRP is initiated by a ventricular pace or sense.
The temporary pacemaker does not use a VRP.

5.6.3 Noise response


The temporary pacemaker paces asynchronously at the programmed RATE in the presence
of continuous noise (such as 50 Hz or 60 Hz interference) with cycle lengths of 40 ms or less
until the noise terminates. This noise response applies to the atrial or ventricular channel or
both, depending on the channel for which the noise is sensed.

5.6.4 Wenckebach response


The Wenckebach response occurs during DDD pacing mode. When the intrinsic atrial rate
of the patient increases beyond the UPPER RATE, the temporary pacemaker continues to
lengthen the SAV interval until a P-wave falls within the PVARP. When the P-wave falls within
the PVARP, it is not sensed.
Because the P-wave is not sensed, the SAV interval is not started, and a ventricular pace is
not issued.

5.6.5 Ventricular safety pacing


The Ventricular Safety Pacing feature is intended to prevent improper inhibition of
ventricular pacing pulses if a non-ventricular event (such as cross-talk or noise) is sensed by
the ventricular lead. This feature functions when the temporary pacemaker is operating in a
mode that senses and paces in the ventricle, and senses and paces in the atrium (DDI and
DDD pacing modes).

Technical Manual 75
Medtronic 5392

If a ventricular sensed event is detected outside the 30 ms blanking window, but within
110 ms after an atrial paced event, a ventricular safety pacing (VSP) pulse is delivered at
110 ms after the atrial pace if the A-V INTERVAL is greater than 110 ms. The ventricular
pacing pulse is delivered at the A-V INTERVAL if the A-V INTERVAL is less than 110 ms.
If the RATE is set to 86 ppm or higher, the values of 110 ms change to 70 ms (see
Section 7.1). The VPACE LED and the VSENSE LED flash within 15 ms of each other.

5.6.6 High atrial rate response


The Mode Switch feature temporarily switches from DDD pacing mode to a nontracking DDI
pacing mode upon detection of an atrial tachyarrhythmia (see Figure 50). The programmed
DDD pacing mode is restored when the atrial tachyarrhythmia ends. By operating in a
nontracking pacing mode, the temporary pacemaker prevents rapid ventricular pacing that
may result from a high atrial rate.

Note: Mode Switch is only available when the temporary pacemaker is operating in DDD
pacing mode.
Mode Switch operation starts when the temporary pacemaker detects the onset of an atrial
tachyarrhythmia episode. The detection of AT/AF onset is based on the timing of atrial
events within the ventricular intervals.
After the temporary pacemaker detects the onset of an atrial tachyarrhythmia, Mode Switch
temporarily changes the pacing mode from DDD to a non-atrial tracking DDI pacing mode.
The ventricular pacing rate gradually changes from the tracking rate to the base rate. This
prevents an abrupt drop in the ventricular rate.
When the atrial tachyarrhythmia ends and the atrial rate decreases below the programmed
Upper Rate, Mode Switch changes the pacing mode back to DDD pacing mode.
Figure 50. Temporary DDI pacing mode screen

76 Technical Manual
Medtronic 5392

5.6.7 Pacing mode transitions


The temporary pacemaker switches pacing modes in accordance with the following
guidelines:
● When switching from an atrial-only pacing mode to a pacing mode involving an
A-V Interval (which could be as short as 20 ms), the first ventricular pacing pulse will be
inhibited to avoid pacing into a T-wave.
● A complete cycle of the current mode is allowed to expire before implementing a new
mode.
● When switching from an asynchronous pacing mode to a synchronous pacing mode,
the temporary pacemaker first searches for cardiac activity (during the first pacing
cycle), and then begins sensing and pacing in both chambers (DDD pacing mode).
● Upper Rate is implemented immediately after switching to DDD pacing mode.
● Parameters that did not apply in the previous mode will be set to nominal (or
rate-dependent) values in the new mode.

Technical Manual 77
Medtronic 5392

6 Cleaning, disinfecting, and maintenance

6.1 Cleaning and disinfecting the temporary pacemaker


Cautions:
● Clean and disinfect the temporary pacemaker before each use for a new patient.
● Do not immerse the temporary pacemaker in water or cleaning agents. Severe damage
to the temporary pacemaker may occur. Do not use automated machine washers. Do
not sterilize the temporary pacemaker by ethylene oxide, gamma radiation or
steam-sterilization (autoclave). Damage to the temporary pacemaker may occur using
these methods.
Clean and disinfect the temporary pacemaker before each use for a new patient.
● Remove the batteries prior to cleaning and disinfecting the temporary pacemaker.
Cleaning – Before disinfection, clean the temporary pacemaker thoroughly using a 70%
isopropyl alcohol prep pad. Wipe to remove all visible soil or blood. Allow the temporary
pacemaker to air dry approximately 5 min or until it is dry.
Disinfecting – Disinfect the temporary pacemaker by using a 70% isopropyl alcohol and a
sterile prep pad, gauze pad or sponge. Wipe down all the external surfaces of the temporary
pacemaker. Maintain an exposure time (wet or damp time) of 15 min. Allow the temporary
pacemaker to air dry for approximately 5 min or until it is dry.
Maintenance – During use, the temporary pacemaker can become too contaminated for
effective cleaning or disinfection by the clinic. If the temporary pacemaker has blood or soil
ingress in its battery compartment, cable ports, or under the knobs, return it to Medtronic for
cleaning and disinfecting. When blood or soil enters into these areas, the clinic cannot
effectively clean or disinfect the temporary pacemaker.

Note: Do not expose the temporary pacemaker to ethers, acetone, chlorinated solvents, or
disinfectants. These solvents may damage the case, labels, or metal components.

6.2 Safety and technical checks


Perform safety and technical checks on the temporary pacemaker at a minimum of every
12 months, and after any malfunction or accident. Medtronic recommends that qualified
engineers and technicians trained in the service of Medtronic products perform the checks.
Contact your Medtronic sales or service representative for service or training.

78 Technical Manual
Medtronic 5392

Note: Understanding the warnings listed in this manual is necessary to perform safety and
technical checks successfully.

6.2.1 Visual inspection


Perform the following visual inspections each time the temporary pacemaker is used:
● Check that there is no mechanical or physical damage to the temporary pacemaker.
● Inspect the battery compartment and battery connection for corrosion and other
contamination.

6.2.2 Functional inspection


Perform the following functional inspections each time the temporary pacemaker is used:
● Verify that the temporary pacemaker passes the self-test at power-up.
● Verify that the front panel dials, keys, and displays function and work properly.
● Inspect all connections and cables. Verify that the patient cables are properly connected
and are not damaged.

6.2.3 Practical measurements


Verify the following practical measurements during scheduled safety and technical checks:
● Rate Test
● Rapid Atrial Pacing
● Output
● Sensitivity
● Battery drain test
● Off Current Drain (10 µA maximum)
● On Current Drain with backlight and lower screen off (5.9 mA maximum when measured
away from pacing peaks)
● Patient Leakage Current Measurements per IEC 60601-1 and Patient Auxiliary Current
Measurements per IEC 60601-2-31.

Caution: Do not open the external case of the temporary pacemaker. Opening the external
case of the temporary pacemaker voids the warranty.
Medtronic does not recommend field repair of the device. Contact your local Medtronic sales
or service representative for service or repair.

Technical Manual 79
Medtronic 5392

6.3 Service
Medtronic employs highly trained representatives and engineers located throughout the
world to serve you and, upon request, to provide training to qualified hospital personnel in the
use of Medtronic products. Medtronic also maintains a professional staff to provide technical
consultation to product users. For medical consultation, Medtronic can often refer product
users to outside medical consultants with appropriate expertise. For more information,
contact your local Medtronic representative.
Should service or repair be necessary, contact your local Medtronic sales or service
representative.
A serial number identifying each individual temporary pacemaker is printed on the back
surface of the device. Reference this serial number in any correspondence regarding this
device.

6.4 Product useful life


Long-term reliability of the temporary pacemaker is subject to the actual use conditions of
the device. The routine testing and preventative maintenance recommended in this manual
will help provide reliable operation of the temporary pacemaker.
The expected service life of the temporary pacemaker is five years. Medtronic will not
service or repair the temporary pacemaker after five years. Contact your Medtronic
representative to replace your temporary pacemaker after it has been in service for five
years.

80 Technical Manual
Medtronic 5392

7 Specifications

7.1 Device specifications


Table 3. Temporary pacemaker specifications
Pacing DDD, DDI, DOO, AAI, AOO, VVI, VOO
modes
RATE Range (in ppm) Increments (in ppm) Tolerance
30 – 50 5 30 – 200 ±2%
50 – 100 2
100 – 170 5
170 – 200 6
RAP rate Range (in ppm) Increments (in ppm) Tolerance
80 – 180 20 80 – 360 ±2%
180 – 250 5 380 – 800 ±4%
250 – 360 10
360 – 800 20
Output amplitude
Atrial Range (in mA) Increments (in mA) Tolerance
0.1 – 0.4 0.1 0.1 – 20
0.4 – 1.0 0.2 Greater of ±0.1 mA or ±10%
1.0 – 5.0 0.5 (200-1000 Ω)
5.0 – 20 1.0
Ventricular Range (in mA) Increments (in mA) Tolerance
0.1 – 0.4 0.1 0.1 – 20
0.4 – 1.0 0.2 Greater of ±0.1 mA or ±10%
1.0 – 5.0 0.5 (200-1000 Ω)
5.0 – 25 1.0 20 – 25
±10%
(200–500 Ω)
Pulse width (fixed)
Atrial 1.0 ms ±10%
Ventricular 1.5 ms ±10%
Sensitivity a

Atrial Range (in mV) Increments (in mV) Tolerance


0.4 – 0.8 0.1 < 0.8 mV ±60%
0.8 – 2.0 0.2 ≥ 0.8 mV ±40%
2.0 – 3.0 0.5
3.0 – 10 1.0

Technical Manual 81
Medtronic 5392

Table 3. Temporary pacemaker specifications (continued)


Ventricular Range (in mV) Increments (in mV) Tolerance
0.8 – 1.0 0.2 ±55%
1.0 – 3.0 0.5
3.0 – 10 1.0
10 – 20 2.0
A-V Interval
Paced A-V (PAV) Sensed A-V (SAV)
Formula 300 – (1.67 x RATE = PAV – 30 within range in DDD pacing mode
in ppm) within range = PAV within range in DDI pacing mode
Range 50 – 250 – Auto 50 – 250
(in ms) 20 – 300 – Manual
Increments 10 10
(in ms)
Tolerance Greater of ±5 ms or ±5% Greater of ±15 ms or ±15%
Refractory period
Atrial
At atrial SAV or PAV, whichever is in effect
event 310 ms +5/-30 ms or 75% of the base rate, whichever is lower (AAI pacing mode only)
At ven- Auto (all values Rate range (in ppm) PVARP (in ms)
tricular are +5/-30 ms)
event
(PVARP)
≤ 100 300
> 100 and ≤ 150 250
> 150 and ≤ 180 230
> 180 200
Manual (all values Range (in ms) Increments (in ms)
are +5/-30 ms)
150 – 500 10
Upper Rate
Auto RATE + 30 ppm ±10% Minimum of 110 ppm
Manual Range (in ppm) Increments (in ppm) Tolerance
80 – 100 2 ±10%
100 – 170 5
170 – 200 6
200 – 214 7
214 – 230 8

82 Technical Manual
Medtronic 5392

Table 3. Temporary pacemaker specifications (continued)


Safety RATE < 86 ppm Occurs 110 ms after atrial pace, if A-V interval is set to
pace more than 110 ms, or at programmed A-V interval if A-V
interval is set to less than 110 ms
RATE ≥ 86 ppm Occurs 70 ms after atrial pace, if A-V interval is set to
more than 70 ms, or at programmed A-V interval if A-V
interval is set to less than 70 ms
Mode = 171 ppm, if the Upper Rate is < 165 ppm
switch = Upper Rate + 10 ppm, if the Upper Rate is ≥ 165 ppm
detection
rate
Blankingb
Atrial
200 ms +5/-30 ms – after atrial pace
100 ms +2/-30 ms – after atrial sense
100 ms +2/-15 ms – after ventricular pace/sense
Ventricular
30 ms +2/-15 ms – after atrial pace
200 ms +5/-30 ms – after ventricular pace
120 ms +2/-30 ms – after ventricular sense
RATE limit 230 ppm If a non-RAP rate exceeds 230 ppm, pacing is termina-
(non-RAP) ted. A recoverable error message is displayed in the
lower screen.
Nominal values - DDD pacing mode Nominal values - DOO for Emergency
Pacing DDD DOO for Emergency
mode
RATE 80 ppm When the DOO/Emergency key is pressed, current set-
ting (or 80 ppm if the temporary pacemaker was off
before the DOO/Emergency key was pressed).
Output amplitude
Atrial 10 mA 20 mA for Emergency
Ventricu- 10 mA 25 mA for Emergency
lar
Pulse width (fixed)
Atrial 1.0 ms
Ventricu- 1.5 ms
lar
Sensitivity
Atrial 0.5 mV Asynchronous for Emergency
Ventricu- 2.0 mV Asynchronous for Emergency
lar

Technical Manual 83
Medtronic 5392

Table 3. Temporary pacemaker specifications (continued)


AV Interval
Sensed 140 ms
Paced 170 ms When the DOO/Emergency key is pressed, current man-
ual setting (or 170 ms if temporary pacemaker is off, or
automatic rate-dependent when the temporary pace-
maker is on)
PVARP 300 ms Not applicable
Upper Rate 110 ppm Not applicable
RAP rate 250 ppm
Dimensions
Height 20.3 cm (8.0 in) ±15%
Width 8.6 cm (3.375 in) ±15%
Depth 4.45 cm (1.75 in) ±15%
(without
dials)
Weight 680 g (24 ounces) maximum
(with bat-
tery)
Temperature
Operat- 15 °C to 35 °C (59 °F to 95 °F)c
ing
Storage -40 °C to 70 °C (-40 °F to 158 °F)
(without
battery)
Humidity > 80% and ≤ 95% at 35 °C (95 °F), use after 48 hours dry time
(storage) ≥ 10% and ≤ 80% at 35 °C (95 °F), for immediate use
Battery Two IEC type LR6-sized (AA-sized) 1.5 V alkaline batteries (Duracell MN1500, Ever-
type eady E91 or equivalent)
Battery life 7 days minimum, when the RATE is 80 ppm, and all other parameters are at the
nominal values. Higher amplitudes and rates decrease battery life.
Operation 30 s (typical) under the following conditions: RATE of 80 ppm or less, A OUTPUT and
after bat- V OUTPUT of 10 mA or less, backlight off, and lower screen blank.d
tery
removal
a When sensing 40 ms-wide Haversine waveform for ventricular inputs, 20 ms-wide Haversine waveform for atrial
inputs.
b When tested with a 1 ms square pulse with sufficient amplitude.
c Within the ranges of 10 °C to 15 °C (50 °F to 59 °F) and 35 °C to
43 °C (95 °F to 110 °F), the specification for
OUTPUT is derated an additional ± 5%, the specification for SENSITIVITY is derated an additional ± 7%; and the
specification for RATE is not derated.
d Medtronic does not recommend replacing the batteries while the pacemaker is turned on or actively pacing the
patient.

84 Technical Manual
Medtronic 5392

7.2 Pacing information tables


Table 4. Temporary pacemaker single chamber pacing setup table
Pacing mode AOO VOO AAI VVI
A (Atrial) and V (Ventricle) A V A V
Indicators
PACE and SENSE Indicators PACE PACE PACE + SENSE PACE + SENSE
Instructions
1. Set OUTPUT
A OUTPUT On Off On Off
V OUTPUT Off On Off On
2. Set Sensitivity
A Sensitivity ASYNC NA On NA
V Sensitivity NA ASYNC NA On
3. Set A Tracking NA NA NA NA

Table 5. Temporary pacemaker dual chamber pacing setup table


Pacing mode DOO DDD DDI
A (Atrial) and V (Ventricle) A+V A+V A+V
Indicators
PACE and SENSE Indica- PACE (A) + PACE PACE + SENSE (A) PACE + SENSE
tors (V) and (A)
PACE + SENSE (V) and
PACE + SENSE
(V)
Instructions
1. Set OUTPUT
A OUTPUT On On On
V OUTPUT On On On
2. Set Sensitivity
A Sensitivity ASYNC On On
V Sensitivity ASYNC On On
3. Set A Tracking NA On Off

Table 6. Rate and interval conversion chart for RATE and RAP
Rate RAP
Rate Interval Rate Interval
30 ppm 2000 ms 80 ppm 750 ms
35 ppm 1714 ms 100 ppm 600 ms

Technical Manual 85
Medtronic 5392

Table 6. Rate and interval conversion chart for RATE and RAP (continued)
Rate RAP
40 ppm 1500 ms 120 ppm 500 ms
45 ppm 1333 ms 140 ppm 429 ms
50 ppm 1200 ms 160 ppm 375 ms
52 ppm 1154 ms 180 ppm 333 ms
54 ppm 1111 ms 185 ppm 324 ms
56 ppm 1071 ms 190 ppm 316 ms
58 ppm 1034 ms 195 ppm 308 ms
60 ppm 1000 ms 200 ppm 300 ms
62 ppm 968 ms 205 ppm 293 ms
64 ppm 938 ms 210 ppm 286 ms
66 ppm 909 ms 215 ppm 279 ms
68 ppm 882 ms 220 ppm 273 ms
70 ppm 857 ms 225 ppm 267 ms
72 ppm 833 ms 230 ppm 261 ms
74 ppm 811 ms 235 ppm 255 ms
76 ppm 789 ms 240 ppm 250 ms
78 ppm 769 ms 245 ppm 245 ms
80 ppm 750 ms 250 ppm 240 ms
82 ppm 732 ms 260 ppm 231 ms
84 ppm 714 ms 270 ppm 222 ms
86 ppm 698 ms 280 ppm 214 ms
88 ppm 682 ms 290 ppm 207 ms
90 ppm 667 ms 300 ppm 200 ms
92 ppm 652 ms 310 ppm 194 ms
94 ppm 638 ms 320 ppm 188 ms
96 ppm 625 ms 330 ppm 182 ms
98 ppm 612 ms 340 ppm 176 ms
100 ppm 600 ms 350 ppm 171 ms
105 ppm 571 ms 360 ppm 167 ms
110 ppm 545 ms 380 ppm 158 ms
115 ppm 522 ms 400 ppm 150 ms
120 ppm 500 ms 420 ppm 143 ms
125 ppm 480 ms 440 ppm 136 ms
130 ppm 462 ms 460 ppm 130 ms
135 ppm 444 ms 480 ppm 125 ms

86 Technical Manual
Medtronic 5392

Table 6. Rate and interval conversion chart for RATE and RAP (continued)
Rate RAP
140 ppm 429 ms 500 ppm 120 ms
145 ppm 414 ms 520 ppm 115 ms
150 ppm 400 ms 540 ppm 111 ms
155 ppm 387 ms 560 ppm 107 ms
160 ppm 375 ms 580 ppm 103 ms
165 ppm 364 ms 600 ppm 100 ms
170 ppm 353 ms 620 ppm 97 ms
176 ppm 341 ms 640 ppm 94 ms
182 ppm 330 ms 660 ppm 91 ms
188 ppm 319 ms 680 ppm 88 ms
194 ppm 309 ms 700 ppm 86 ms
200 ppm 300 ms 720 ppm 83 ms
740 ppm 81 ms
760 ppm 79 ms
780 ppm 77 ms
800 ppm 75 ms

Technical Manual 87
Medtronic 5392

8 Special notice

8.1 Special notice for the temporary pacemaker


Use of prior Medtronic temporary pacemakers has met with some success in the treatment
of certain heart disorders, including heart block and heart arrhythmias. However Medtronic
makes no warranty that the Model 5392 Dual Chamber Temporary External Pacemaker will
efficiently restore adequate cardiac function for all patients. For information regarding
common causes of pacing difficulty, consult other portions of the manual.

88 Technical Manual
Medtronic 5392

9 Troubleshooting

9.1 Troubleshooting
9.1.1 Self-test
When the temporary pacemaker is turned on, a startup message appears in the lower
screen. A self-test is executed while the temporary pacemaker is turning on. The self-test
includes a check of all keys and critical internal circuits. The PACE and SENSE LEDs
illuminate to indicate progress during the self-test.

Note: Pressing any key while the self-test is in process can cause the temporary pacemaker
to fail the self-test. The temporary pacemaker interprets the pressed key as being “stuck”
and, therefore, malfunctioning. If a key is pressed during the self-test, causing a self-test
failure, the lower screen displays an error message until the key is released.
The upper and lower screens initialize when the temporary pacemaker turns on. All
indicators display, including the low battery indicator, for 2 s. The low battery indicator
displaying during the upper screen initialization does not indicate that the batteries are low.
If the batteries are low, the low battery indicator remains visible during temporary pacemaker
operation.
After successful completion of the self-test, the temporary pacemaker first searches for
cardiac activity (during the first pacing cycle), and then begins sensing and pacing in both
chambers (DDD pacing mode).
If the temporary pacemaker fails the self-test, the PACE and SENSE LEDs remain on in a
pattern indicating the failed test, and no output pulses are issued. Failure codes may be
displayed on the lower screen. Follow the instructions that appear on the screen to restart
the temporary pacemaker, or to return the temporary pacemaker for service (contact your
Medtronic representative). When returning the temporary pacemaker, remove the batteries,
and return both the temporary pacemaker and the batteries.

9.1.2 Loss of sensing


Figure 51. Loss of atrial sensing (DDD pacing mode)

200 ms

Technical Manual 89
Medtronic 5392

Figure 52. Loss of ventricular sensing (DDD pacing mode)

200 ms

9.1.2.1 Keys to identifying loss of sensing


If a P-wave or R-wave is present, the temporary pacemaker does not detect intrinsic activity
and delivers a pacing pulse. The following occurs:
● Pacing artifacts are seen asynchronously on ECG.
● Sense indicator does not flash, though ECG shows depolarization.

9.1.2.2 Potential causes of loss of sensing


Temporary pacemaker rela-
Heart/patient related Patient cable/lead related ted
Inadequate cardiac signal ● Loose connection at the ● Inappropriate sensitivity
connector block setting
● Inappropriate lead place- ● Inappropriate pacing mode
ment selection
● Insulation break or wire ● Small sensing window
fracture ● Electrical noise

9.1.2.3 Potential solutions for loss of sensing


Check one or more of the following when troubleshooting loss of sensing:
● Check cable connections for loose wires.
● Verify that the apparent loss of sensing is not due to blanking periods.
● Verify that the event not being sensed is not occurring during the refractory period. If
necessary, re-evaluate the refractory period.
Note: An intrinsic event during the refractory period is not indicated by the SENSE LED.

● Perform sensing threshold test for the affected chamber. Provide at least a 2:1 safety
margin.
Note: If the patient does not have adequate intrinsic rhythm, consult a physician before
the test.

90 Technical Manual
Medtronic 5392

● Increase the sensitivity for the appropriate chamber. Select A Sensitivity or


V Sensitivity, then slowly turn the Menu Parameter dial clockwise (decrease mV until
the ECG shows intrinsic activity).
Note: Use caution when adjusting the sensitivity for patients with a history of sustained
ventricular tachycardia.

9.1.3 Oversensing
Figure 53. Atrial oversensing

200 ms

Figure 54. Ventricular oversensing

200 ms

9.1.3.1 Keys to identifying oversensing


Verify the following when identifying if oversensing has occurred:
● ECG shows erratic prolonging or shortening of pacing interval.
● Persistent oversensing may result in asynchronous pacing or no pacing output.

Technical Manual 91
Medtronic 5392

9.1.3.2 Potential causes of oversensing


One (or more) of the following issues can cause oversensing:
Heart or patient related Patient cable or lead related Temporary pacemaker related
● T-wave sensing ● Insulation break ● Inappropriate sensitivity set-
ting(s)
● Far-field sensing (sens- ● Conductor wire fracture
ing, P- waves, or ● Cross-sensing (DDD or DDI
R-waves) pacing modes). ECG shows a
paced A-V Interval of 110 ms
● Muscle sensing
or 70 ms occurring on a regu-
● EMI lar basis.

9.1.3.3 Potential solutions for oversensing


When troubleshooting oversensing, verify one or more of the following:
● Check the cable connections for loose wires.
● Replace the cable if there is an insulation break or wire fracture.
● Perform the sensing threshold test for the affected chamber. Provide at least a 2:1 safety
margin.
Note: If the patient does not have adequate intrinsic rhythm, consult a physician before
the test.

● If the source is muscle activity, EMI, T-wave, or far-field sensing. Decrease the
sensitivity for the affected chamber. Select A Sensitivity or V Sensitivity, and then
slowly turn the Menu Parameter dial counterclockwise (increase mV value).
Note: When adjusting sensitivity or output, verify the appropriate safety margin.

● If the source is cross sensing, try one or both of the following:


– Decrease the ventricular sensitivity for the affected chamber. Select V Sensitivity,
and then slowly turn the Menu Parameter dial counterclockwise (increase mV value).
– Decrease the A OUTPUT by slowly turning the A OUTPUT dial counterclockwise.

92 Technical Manual
Medtronic 5392

9.1.4 Loss of capture


Figure 55. Loss of atrial capture

200 ms

Figure 56. Loss of ventricular capture

200 ms

9.1.4.1 Keys to identifying loss of capture


If the ECG shows no depolarization after delivery of a pacing stimulus, loss of capture has
occurred.

Note: If the intrinsic event occurs just before pacing, see Section 9.1.2.

9.1.4.2 Potential causes of loss of capture


Temporary pacemaker rela-
Heart or patient related Patient cable or lead related ted
● Increased stimulation ● Loose connection at the ● Inadequate output set-
threshold connector block ting – mA
● Exit block ● Heartwire dislodgement ● Inappropriate pacing mode
selection
● Tissue changes ● Inappropriate lead place-
ment
● General metabolic imbal-
ances ● Insulation break or wire
fracture
● Drug effects

Technical Manual 93
Medtronic 5392

9.1.4.3 Possible misdiagnosis of loss of capture


Verify the following when determining if loss of capture has been misdiagnosed:
● Bipolar artifacts are too small to be seen.
● Digital ECG monitor or recorder is not sensitive to pacing spikes.

9.1.4.4 Possible solutions for loss of capture


When troubleshooting loss of capture, verify one or more of the following:
● Check the cable connections for loose wires.
● Replace the cable if there is an insulation break or wire fracture.
● Verify that the correct chamber is being paced.
● Verify the pacing mode.
● Perform the stimulation threshold test for the affected chamber. Provide at least a 2:1
safety margin.
● If necessary, increase the output for the appropriate chamber. Slowly turn the
A OUTPUT or V OUTPUT dial clockwise until the ECG shows capture.
Note: Use caution when adjusting the output for patients with a history of sustained
ventricular tachycardia.

9.1.5 Stimulation of chest wall or diaphragm


9.1.5.1 Potential causes of stimulation of chest wall or diaphragm
Check the following when determining the causes of stimulation of chest wall or diaphragm:
Patient cable or lead related Temporary pacemaker-related
● Improper lead placement ● Excessive output settings – mA
● Loose connection at the connector block ● Inappropriate pacing mode selection
● Conductor wire fracture

94 Technical Manual
Medtronic 5392

9.1.5.2 Possible solutions for stimulation of chest wall or diaphragm


Verify the following when determining the causes of stimulation of the chest wall or
diaphragm:
● Determine if the atrial or ventricular lead is the cause of the problem. First, reduce the
output to one lead and then to the other lead. When the affected lead is identified, reduce
the output for that lead.
● Perform the stimulation threshold test for the affected chamber. Set the output to a value
more than 2 times the threshold value, but below the point of stimulating the chest or
diaphragm.
● If the outcome does not improve, correct the placement of the leads.

9.1.6 No output pulse


9.1.6.1 Key to identifying no output pulse
The ECG does not show the pacing spikes when there is no output.

9.1.6.2 Possible misdiagnosis of no output pulse


The following can cause misdiagnosis of no output pulse:
● Pacing is inhibited because of sensed noise or intrinsic activity.
● Digital ECG monitor or recorder is not sensitive to pacing spikes.

9.1.6.3 Possible solutions for no output pulse


When troubleshooting no output pulse, do one or more of the following:
● Observe the pace indicators.
● Re-insert the patient cables.
● Replace the batteries.
● Change the ECG monitor or recorder.
● Re-select the pacing mode.
● Set output to the appropriate safety margin.

Technical Manual 95
Medtronic 5392

9.1.7 Hemodynamic changes


9.1.7.1 Keys to identifying hemodynamic changes
Check the following items when identifying possible causes of hemodynamic changes:
● Decreased blood pressure
● Altered brachial pulse rate

9.1.7.2 Possible solutions to hemodynamic changes


When troubleshooting hemodynamic changes, check one or more of the following:
● Check the pacing mode. If necessary, change to DDD pacing mode.
● Verify that sensing or capture has not been lost. Correct if necessary.
● Adjust the pacing rate (increase it in most cases).
● Adjust the A-V Interval.
● Consult a physician.

9.1.8 Pacemaker-mediated tachycardia


9.1.8.1 Keys to identifying pacemaker-mediated tachycardia
Check one or more of the following items when identifying pacemaker-mediated
tachycardia:
● Occurs only in DDD pacing mode.
● Extended periods of ventricular pacing at or close to the Upper Rate setting.
● Patient may be symptomatic.
● The ECG shows retrograde P-waves after each paced ventricular event.

9.1.8.2 Potential causes of pacemaker-mediated tachycardia


One or more of the following items can cause pacemaker-mediated tachycardia:
● Retrograde conduction
● Loss of AV synchrony due to the effect of the Upper Rate
● PVC, sensed by the atrium, which re-triggers the A-V Interval
● PAC, blocked by the Upper Rate

96 Technical Manual
Medtronic 5392

9.1.8.3 Possible misdiagnosis of pacemaker-mediated tachycardia


Sinus tachycardia can cause a misdiagnosis of pacemaker-mediated tachycardia.

9.1.8.4 Possible solutions of pacemaker-mediated tachycardia


When troubleshooting pacemaker-mediated tachycardia, do one or more of the following:
● Consult a physician to determine the presence of sinus tachycardia.
● Restore the AV synchrony with one of the following adjustments:
– Shorten the A-V Interval.
– Increase the PVARP.
– Adjust the Upper Rate in the presence of fast intrinsic atrial rates.
● Set the PVARP to a value longer than the retrograde conduction time. Use the ECG to
determine the retrograde conduction time.
● Test sensing and stimulation thresholds.

9.1.9 Recoverable error message


If the temporary pacemaker displays the recoverable error message (see Figure 57) in the
lower screen, press the Enter key. The temporary pacemaker restarts and performs the
self-test.
Figure 57. Recoverable error message

Technical Manual 97
Medtronic 5392

9.1.10 Non-recoverable error message


If the temporary pacemaker displays the non-recoverable error message (see Figure 58) in
the lower screen, contact Medtronic for service.
Figure 58. Non-recoverable error message

9.1.11 Button press detected error message


If the temporary pacemaker displays the Button Press Detected error message in the lower
screen, a button has been pressed while the temporary pacemaker is turning on. Release
any pressed key and the temporary pacemaker continues to power on normally.
Figure 59. Error – Button Press Detected

98 Technical Manual
Medtronic 5392

Index
Numerics Atrial pacing
blanking specifications . . . . . . . . . . . . . . . . . 83
2:1 safety margin
contraindicated . . . . . . . . . . . . . . . . . . . . . 11
sensing threshold . . . . . . . . . . . . . . . . . . . . 54
mode transition . . . . . . . . . . . . . . . . . . . . . 77
stimulation threshold . . . . . . . . . . . . . . . . . . 57
output dial . . . . . . . . . . . . . . . . . . . . . . . . 23
A Atrial sensing
AAI contraindication . . . . . . . . . . . . . . . . . . . . . 11
Mode selection . . . . . . . . . . . . . . . . . . . . . 29 Automatic adjustment
AAI mode sensitivity threshold . . . . . . . . . . . . . . . . . . . 53
pacing setup table . . . . . . . . . . . . . . . . . . . 85 Automatic/Manual(*) . . . . . . . . . . . . . . . . . . . . . 68
A (atrial) OUTPUT A-V Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
during RAP . . . . . . . . . . . . . . . . . . . . . . . . 73 A-V Interval (PAV) . . . . . . . . . . . . . . . . . . . . . . 63
nominal value . . . . . . . . . . . . . . . . . . . . . . 83 timing violations vs. RATE, PVARP . . . . . . . . . 71
A (atrial) Sensitivity . . . . . . . . . . . . . . . . . . . . . . 60 A-V INTERVAL (PAV)
vs. A OUTPUT . . . . . . . . . . . . . . . . . . . . . . 61 nominal value . . . . . . . . . . . . . . . . . . . . . . 84
A (atrial) SENSITIVITY ranges, automatic and manual . . . . . . . . . . . . 82
nominal value . . . . . . . . . . . . . . . . . . . . . . 83 safety pacing . . . . . . . . . . . . . . . . . . . . . . . 76
range, increments . . . . . . . . . . . . . . . . . . . . 81
ablation
B
Basic operation . . . . . . . . . . . . . . . . . . . . . . . . 43
microwave . . . . . . . . . . . . . . . . . . . . . . . . 12
off . . . . . . . . . . . . . . . . . . . . . . . . . . . 43, 44
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
on . . . . . . . . . . . . . . . . . . . . . . . . . . . 43, 44
Accessories
Batteries
temporary pacemaker . . . . . . . . . . . . . . . . . 10
installation . . . . . . . . . . . . . . . . . . . . . . . . 35
Adverse effects . . . . . . . . . . . . . . . . . . . . . . . . 19
polarity . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Amplitude, inappropriate variance . . . . . . . . . . . . . 14
recommendations . . . . . . . . . . . . . . . . . . . . 35
AOO
recommended battery type . . . . . . . . . . . . . . 34
Mode selection . . . . . . . . . . . . . . . . . . . . . 29
replacement . . . . . . . . . . . . . . . . . . . . . . . 35
AOO mode
Battery
during RAP . . . . . . . . . . . . . . . . . . . . . . . . 72
drawer . . . . . . . . . . . . . . . . . . . . . . . . . . 32
pacing setup table . . . . . . . . . . . . . . . . . . . 85
drawer release button . . . . . . . . . . . . . . . . . 33
Arrow keys
life . . . . . . . . . . . . . . . . . . . . . . . . . . . 33, 84
selecting parameter values, modes . . . . . . . . . 24
precaution . . . . . . . . . . . . . . . . . . . . . . . . 14
ASYNC . . . . . . . . . . . . . . . . . . . . . . . . . . . 61, 62
type . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Asynchronous pacing . . . . . . . . . . . . . . . . . . . . 49
Battery indicator
caution . . . . . . . . . . . . . . . . . . . . . . . . . . 49
at power on . . . . . . . . . . . . . . . . . . . . . . . 44
contraindication . . . . . . . . . . . . . . . . . . . . . 11
Battery replacement
DOO/Emergency key . . . . . . . . . . . . . . . . . . 50
continued operation during . . . . . . . . . . . . 33, 84
to resume demand pacing . . . . . . . . . . . . . . . 50
Bipolar lead system . . . . . . . . . . . . . . . . . . . . . . . 7
transition rules . . . . . . . . . . . . . . . . . . . . . . 77
connecting to patient cable . . . . . . . . . . . . . . 41
A Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
precautions . . . . . . . . . . . . . . . . . . . . . . . 15
A Tracking (atrial tracking) . . . . . . . . . . . . . . . . . . 66
Blanking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Atrial
arrhythmias . . . . . . . . . . . . . . . . . . . . . . . 11 C
mapping . . . . . . . . . . . . . . . . . . . . . . . . . 16 Cable
connecting to temporary pacemaker . . . . . . . . . 38
connecting to temporary pacemaker, fig . . . . . . 39

Technical Manual 99
Medtronic 5392

Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Disinfection
precautions . . . . . . . . . . . . . . . . . . . . . . . 15 temporary pacemaker . . . . . . . . . . . . . . . . . 78
sockets for . . . . . . . . . . . . . . . . . . . . . . . . 33 Disposable pouch . . . . . . . . . . . . . . . . . . . . . . . 18
Cable to lead system connections . . . . . . . . . . . . . 41 DOO/Emergency key
Capture, definition . . . . . . . . . . . . . . . . . . . . . . 56 function . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Cardioversion . . . . . . . . . . . . . . . . . . . . . . . . . 12 if upper screen is locked . . . . . . . . . . . . . . . . 45
Cautery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 DOO mode
Cellular phones . . . . . . . . . . . . . . . . . . . . . . . . 13 Mode selection . . . . . . . . . . . . . . . . . . . . . 29
Chart pacing setup table . . . . . . . . . . . . . . . . . . . 85
mode vs parameter settings . . . . . . . . . . . . . . 85 Dual chamber pacing
Checks, safety and technical . . . . . . . . . . . . . . . . 78 definition . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cleaning Dual chamber pacing mode
temporary pacemaker . . . . . . . . . . . . . . . . . 78 DDD, DDI, DOO, . . . . . . . . . . . . . . . . . . . . 29
Communication transmitters . . . . . . . . . . . . . . . . 13 Dual chamber pacing modes
Concomitant pacing precaution . . . . . . . . . . . . . . . . . . . . . . . . 16
contraindication . . . . . . . . . . . . . . . . . . . . . 11
Condition use
E
Electromagnetic interference (EMI) . . . . . . . . . . . . 13
prior to use . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrostatic discharge (ESD) . . . . . . . . . . . . . . 17, 18
Conduction, retrograde . . . . . . . . . . . . . . . . . . . 17
Electrosurgical equipment . . . . . . . . . . . . . . . . . . 13
Connector block, temporary pacemaker . . . . . . . . . 33
Emergency pacing . . . . . . . . . . . . . . . . . . . . . . 50
Connector setup
initiating . . . . . . . . . . . . . . . . . . . . . . . . . . 50
cables and the temporary pacemaker . . . . . . . . 38
nominal values . . . . . . . . . . . . . . . . . . . . . 51
lead system to patient cables . . . . . . . . . . . . . 40
terminating . . . . . . . . . . . . . . . . . . . . . . . . 51
overview . . . . . . . . . . . . . . . . . . . . . . . . . 38
Enter key
precaution . . . . . . . . . . . . . . . . . . . . . . . . 14
Mode selection menu . . . . . . . . . . . . . . . . . 23
Contraindications . . . . . . . . . . . . . . . . . . . . . . . 10
Rapid atrial pacing . . . . . . . . . . . . . . . . . . . 72
Controls . . . . . . . . . . . . . . . . . . . . . . . . . . 20, 24
timing violations . . . . . . . . . . . . . . . . . . . . . 70
upper screen . . . . . . . . . . . . . . . . . . . . . . . 23
to resume demand pacing . . . . . . . . . . . . . . . 50
Conversion chart, rate and interval . . . . . . . . . . . . . 85
Environmental precautions . . . . . . . . . . . . . . . . . 18
D Equipment
DDD, DDI, DOO, AAI, AOO, VVI, VOO, OOO . . . . . . 47 diathermy . . . . . . . . . . . . . . . . . . . . . . . . . 13
DDD mode electrosurgical . . . . . . . . . . . . . . . . . . . . 12, 13
Mode selection . . . . . . . . . . . . . . . . . . . . . 29 line-powered . . . . . . . . . . . . . . . . . . . . . . . 12
pacing setup table . . . . . . . . . . . . . . . . . . . 85 magnetic resonance imaging (MRI) . . . . . . . . . 13
DDI medical telemetry . . . . . . . . . . . . . . . . . . . . 13
Mode selection . . . . . . . . . . . . . . . . . . . . . 29 Erratic output, sensing . . . . . . . . . . . . . . . . . . . . 14
DDI mode
pacing setup table . . . . . . . . . . . . . . . . . . . 85
F
Failures
Defibrillation . . . . . . . . . . . . . . . . . . . . . . . . . . 12
random . . . . . . . . . . . . . . . . . . . . . . . . . . 14
precaution . . . . . . . . . . . . . . . . . . . . . . . . 15
self-test . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Demand pacing
Features
resume . . . . . . . . . . . . . . . . . . . . . . . . . . 51
automatic . . . . . . . . . . . . . . . . . . . . . . . . . 74
Demand (synchronous) pacing
physical . . . . . . . . . . . . . . . . . . . . . . . . . . 32
to resume from asynchronous . . . . . . . . . . . . 51
safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Description, general . . . . . . . . . . . . . . . . . . . . . . 7
Fluid incursion, precautions . . . . . . . . . . . . . . . . . 18
Device maintenance . . . . . . . . . . . . . . . . . . . . . 80
temporary pacemaker . . . . . . . . . . . . . . . . . 78 G
Dials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 General description . . . . . . . . . . . . . . . . . . . . . . . 7
Diathermy equipment . . . . . . . . . . . . . . . . . . . . 13
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 84

100 Technical Manual


Medtronic 5392

H Lock feature . . . . . . . . . . . . . . . . . . . . . . . . . . 44
lock/unlock key . . . . . . . . . . . . . . . . . . . . . 23
High atrial rate response
Lower bar indicator
mode switch . . . . . . . . . . . . . . . . . . . . . . . 76
battery indicator . . . . . . . . . . . . . . . . . . . . . 27
HIGH OUTPUT
lock indicator . . . . . . . . . . . . . . . . . . . . . . . 27
adverse effect . . . . . . . . . . . . . . . . . . . . . . 19
Lower screen . . . . . . . . . . . . . . . . . . . . . . . . . 59
precaution . . . . . . . . . . . . . . . . . . . . . . . . 17
controls . . . . . . . . . . . . . . . . . . . . . . . . . . 58
High-rate pacing therapy
menus . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
contraindicated in ventricle . . . . . . . . . . . . . . 11
messages . . . . . . . . . . . . . . . . . . . 44, 70, 71
no ventricular back-up . . . . . . . . . . . . . . . . . 13
Mode selection menu . . . . . . . . . . . . . . . . . 29
precaution . . . . . . . . . . . . . . . . . . . . . . 13, 15
Pacing parameters menu . . . . . . . . . . . . . . . 30
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
RAP . . . . . . . . . . . . . . . . . . . . . . . . . . 71, 73
I Rate-dependent parameters . . . . . . . . . . . 62, 68
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
M
Magnetic resonance imaging (MRI) . . . . . . . . . . . . 13
false signals . . . . . . . . . . . . . . . . . . . . . . . 14
Manual/Automatic . . . . . . . . . . . . . . . . . . . . . . . 68
lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Mapping, atrial . . . . . . . . . . . . . . . . . . . . . . . . . 16
pace and sense LEDs . . . . . . . . . . . . . . . . . 25
Medical telemetry equipment . . . . . . . . . . . . . . . . 13
Indifferent electrode . . . . . . . . . . . . . . . . . . . . . 42
Menu parameter dial . . . . . . . . . . . . . . . . . . . . . 24
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
IV pole hanger . . . . . . . . . . . . . . . . . . . . . . . . . 33
Messages
fig. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2:1 block point . . . . . . . . . . . . . . . . . . . . . . 70
Physical features . . . . . . . . . . . . . . . . . . . . 34
locked message . . . . . . . . . . . . . . . . . . . . . 45
use of the IV pole hanger . . . . . . . . . . . . . . . 38
off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
J Upper Rate violation . . . . . . . . . . . . . . . . . . 71
Jumper cable . . . . . . . . . . . . . . . . . . . . . . . . . 42 microwave ablation . . . . . . . . . . . . . . . . . . . . . . 12
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
K nominal . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Mode selection . . . . . . . . . . . . . . . . . . . . . . 47, 48
during lock . . . . . . . . . . . . . . . . . . . . . . . . 45
DDD, DDI, DOO, AAI, AOO, VVI, VOO, OOO . . . 47
L timing violations . . . . . . . . . . . . . . . . . . . 48, 69
Lead system transition rules . . . . . . . . . . . . . . . . . . . . . . 77
connection to cables . . . . . . . . . . . . . . . . . . 40 Mode switch
Lead systems DDD mode . . . . . . . . . . . . . . . . . . . . . . . . 76
adverse effects . . . . . . . . . . . . . . . . . . . . . 19 DDD pacing mode . . . . . . . . . . . . . . . . . . . 76
bipolar . . . . . . . . . . . . . . . . . . . . . . 7, 15, 41 detection rate . . . . . . . . . . . . . . . . . . . . . . 83
connecting to patient cables . . . . . . . . . . . . . 40 high atrial rate response . . . . . . . . . . . . . . . . 76
connecting to patient cables, fig. . . . . . . . . . . . 41 N
inappropriate connections . . . . . . . . . . . . . . . 19
Noise response . . . . . . . . . . . . . . . . . . . . . . . . 75
precautions . . . . . . . . . . . . . . . . . . . . . . . 15
asynchronous pacing . . . . . . . . . . . . . . . . . . 75
reposition . . . . . . . . . . . . . . . . . . . . . . . . . 19
Nominal values
unipolar . . . . . . . . . . . . . . . . . . . . . . . . 16, 41
DDD pacing mode . . . . . . . . . . . . . . . . . . . 83
warnings . . . . . . . . . . . . . . . . . . . . . . . . . 14
DOO for emergency . . . . . . . . . . . . . . . . . . 83
LED indicators
emergency pacing . . . . . . . . . . . . . . . . . . . 51
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . 62
No pacing mode
LEDs (light-emitting diodes)
OOO . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
during self-test . . . . . . . . . . . . . . . . . . . . . . 89
Line-powered equipment . . . . . . . . . . . . . . . . . . 12 O
Locked-up, self-test failure . . . . . . . . . . . . . . . . . 89 On/Off key . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
function . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Technical Manual 101


Medtronic 5392

OOO Periods
Mode selection . . . . . . . . . . . . . . . . . . . . . 29 blanking . . . . . . . . . . . . . . . . . . . . . . . . . 74
Operation after battery removal . . . . . . . . . . . . . . . 84 refractory . . . . . . . . . . . . . . . . . . . . . . . 74, 75
OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . 28, 29 Physical features . . . . . . . . . . . . . . . . . . . . . . . 32
high, adverse effect . . . . . . . . . . . . . . . . . . . 19 Pouch, disposable . . . . . . . . . . . . . . . . . . . . . . 18
inappropriate variance . . . . . . . . . . . . . . . . . 14 Power, loss of control . . . . . . . . . . . . . . . . . . . . 14
loss of control . . . . . . . . . . . . . . . . . . . . . . 14 precaution . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
nominal values . . . . . . . . . . . . . . . . . . . . . 83 Precautions
ranges and increments . . . . . . . . . . . . . . . . . 81 environmental . . . . . . . . . . . . . . . . . . . . . . 18
vs. Sensitivity . . . . . . . . . . . . . . . . . . . . 61, 62 Pulse width . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
atrial . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
P inappropriate variance . . . . . . . . . . . . . . . . . 14
PACE LEDs
ventricular . . . . . . . . . . . . . . . . . . . . . . . . 29
during self-test . . . . . . . . . . . . . . . . . . . . . . 89
PVARP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Pacing
manual adjustments . . . . . . . . . . . . . . . . . . 65
asynchronous, contraindication . . . . . . . . . . . 11
nominal value . . . . . . . . . . . . . . . . . . . . . . 84
atrial . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
range, increments . . . . . . . . . . . . . . . . . . . . 82
atrial, contraindication . . . . . . . . . . . . . . . . . 11
timing violations vs. RATE, A-V Interval . . . . . . . 71
high-rate therapy . . . . . . . . . . . . . . . . . . 11, 13
timing violations vs. SAV, Upper Rate . . . . . . . . 70
pacemaker mediated tachycardia . . . . . . . . . . 11
setup table . . . . . . . . . . . . . . . . . . . . . . . . 85 R
termination of . . . . . . . . . . . . . . . . . . . . . . 17 radio frequency ablation . . . . . . . . . . . . . . . . . . . 12
T-wave . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Random failures . . . . . . . . . . . . . . . . . . . . . . . . 14
Pacing mode . . . . . . . . . . . . . . . . . . . . . . . . . . 81 RAP . . . . . . . . . . . . . . . . . . . . . . . . . . 71, 72, 73
nominal . . . . . . . . . . . . . . . . . . . . . . . . . . 83 see also Rapid atrial pacing . . . . . . . . . . . . . . 71
selecting . . . . . . . . . . . . . . . . . . . . . . . . . 47 Rapid atrial pacing
selection . . . . . . . . . . . . . . . . . . . . . . . 47, 48 deliver . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
transitions . . . . . . . . . . . . . . . . . . . . . . . . 77 Rapid atrial pacing (RAP) . . . . . . . . . . . . . 71, 72, 73
Pacing mode at power-up A OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . 73
DDD mode . . . . . . . . . . . . . . . . . . . . . . . . 22 caution . . . . . . . . . . . . . . . . . . . . . . . . 71, 72
Pacing mode selection . . . . . . . . . . . . . . . . . . . . 47 rate, nominal . . . . . . . . . . . . . . . . . . . . . . . 84
Pacing parameters menu . . . . . . . . . . . . . . . . . . 30 rate, range, increments . . . . . . . . . . . . . . . . 81
A-V Interval . . . . . . . . . . . . . . . . . . . . . . . . 63 to resume upper screen pacing . . . . . . . . . . . . 74
Parameters RATE
A Tracking . . . . . . . . . . . . . . . . . . . . . . . . 66 dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
A-V Interval . . . . . . . . . . . . . . . . . . . . . . . . 63 inappropriate variance . . . . . . . . . . . . . . . . . 14
locked parameters . . . . . . . . . . . . . . . . . . . 44 limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
PVARP . . . . . . . . . . . . . . . . . . . . . . . . . . 65 loss of control . . . . . . . . . . . . . . . . . . . . . . 14
RATE . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 nominal value . . . . . . . . . . . . . . . . . . . . 28, 83
SENSITIVITY . . . . . . . . . . . . . . . . . . . . . . 60 range, increments . . . . . . . . . . . . . . . . . . . . 81
Sensitivity . . . . . . . . . . . . . . . . . . . . . . 60, 61 timing violations . . . . . . . . . . . . . . . . . . . . . 70
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 68 timing violations vs. A-V Interval, PVARP . . . . . . 71
timing violations . . . . . . . . . . . . . . . . . . . . . 69 timing violations vs A-V Interval, PVARP . . . . . . 71
Upper Rate . . . . . . . . . . . . . . . . . . . . . . . . 62 timing violations vs.Upper Rate . . . . . . . . . . . . 71
Upper Rate . . . . . . . . . . . . . . . . . . . . . . . . 64 Rate-dependent parameters . . . . . . . . . . . . . . . . 62
Patient cable A-V Interval . . . . . . . . . . . . . . . . . . . . . . . . 63
connecting to lead system . . . . . . . . . . . . . . . 40 Lower screen . . . . . . . . . . . . . . . . . . . . . . 68
sockets for . . . . . . . . . . . . . . . . . . . . . . . . 33 manual settings . . . . . . . . . . . . . . . . . . . . . 62
Pause key mode transition . . . . . . . . . . . . . . . . . . . . . 77
precaution . . . . . . . . . . . . . . . . . . . . . . 17, 24 specifications . . . . . . . . . . . . . . . . . . . . . . 82
timing violations . . . . . . . . . . . . . . . . . . . . . 69

102 Technical Manual


Medtronic 5392

Rate dependent parameters Single chamber pacing mode


Upper Rate . . . . . . . . . . . . . . . . . . . . . . . . 64 AAI, AOO,VVI, VOO . . . . . . . . . . . . . . . . . . 29
Rate to interval conversion chart . . . . . . . . . . . . . . 85 Sockets
Receptacles patient cables . . . . . . . . . . . . . . . . . . . . . . 33
safety cable . . . . . . . . . . . . . . . . . . . . . . . 18 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 81
Refractory period Status scale
refractory sense . . . . . . . . . . . . . . . . . . . . . 74 upper screen . . . . . . . . . . . . . . . . . . . . . . . 28
Refractory periods . . . . . . . . . . . . . . . . . . . . 74, 75 Sterilization
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 environmental precaution . . . . . . . . . . . . . . . 18
Retrograde conduction . . . . . . . . . . . . . . . . . . . . 17 Stimulation threshold
Reversion capture definition . . . . . . . . . . . . . . . . . . . . 56
random failure . . . . . . . . . . . . . . . . . . . . . . 14 precaution . . . . . . . . . . . . . . . . . . . . . . . . 15
RF ablation . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 procedure . . . . . . . . . . . . . . . . . . . . . . . . 57
safety margin . . . . . . . . . . . . . . . . . . . . . . 57
S Storage
Safety
remove batteries . . . . . . . . . . . . . . . . . . . . 18
checks . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Surgical cables
features . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
sockets for . . . . . . . . . . . . . . . . . . . . . . . . 33
margin, sensing threshold . . . . . . . . . . . . . . . 52
Synchronous (demand) pacing
margin, stimulation threshold . . . . . . . . . . . . . 57
definition . . . . . . . . . . . . . . . . . . . . . . . . . 49
pacing . . . . . . . . . . . . . . . . . . . . . . . . . 75, 83
to resume . . . . . . . . . . . . . . . . . . . . . . . . . 51
SAV (sensed A-V interval) . . . . . . . . . . . . . . . . . . 63
transition rules . . . . . . . . . . . . . . . . . . . . . . 77
nominal value . . . . . . . . . . . . . . . . . . . . . . 84
range . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 T
vs. Upper Rate, PVARP . . . . . . . . . . . . . . . . 70 Table
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 pacing setup . . . . . . . . . . . . . . . . . . . . . . . 85
SENSE LEDs Tables
during self-test . . . . . . . . . . . . . . . . . . . . . . 89 rate and interval conversion . . . . . . . . . . . . . . 85
Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Tachyarrhythmia . . . . . . . . . . . . . . . . . . . . . . . 19
atrial, contraindication . . . . . . . . . . . . . . . . . 11 Technical checks . . . . . . . . . . . . . . . . . . . . . . . 78
Sensing threshold . . . . . . . . . . . . . . . . . . . . . . . 52 Technical service . . . . . . . . . . . . . . . . . . . . . . . 80
precaution . . . . . . . . . . . . . . . . . . . . . . . . 15 Temperature
procedure . . . . . . . . . . . . . . . . . . . . . . 54, 56 operating . . . . . . . . . . . . . . . . . . . . . . . . . 84
safety margin . . . . . . . . . . . . . . . . . . . . . . 54 stabilization . . . . . . . . . . . . . . . . . . . . . . . 18
Sensing thresholds storage . . . . . . . . . . . . . . . . . . . . . . . . . . 84
precaution . . . . . . . . . . . . . . . . . . . . . . . . 16 Terminating pacing . . . . . . . . . . . . . . . . . . . . . . 17
SENSITIVITY . . . . . . . . . . . . . . . . . . . . . . . . . 60 Threshold evaluations . . . . . . . . . . . . . . . . . . . . 52
loss of control . . . . . . . . . . . . . . . . . . . . . . 14 precaution . . . . . . . . . . . . . . . . . . . . . . . . 15
nominal values . . . . . . . . . . . . . . . . . . . . . 83 sensing threshold . . . . . . . . . . . . . . . . . . 52, 56
precaution . . . . . . . . . . . . . . . . . . . . . . . . 16 stimulation threshold . . . . . . . . . . . . . . . . 57, 58
ranges, increments . . . . . . . . . . . . . . . . . . . 81 Timing violations . . . . . . . . . . . . . . . . . . . . . . . 69
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 14, 80 Transition rules . . . . . . . . . . . . . . . . . . . . . . . . 77
Service condition
precaution . . . . . . . . . . . . . . . . . . . . . . . . 14
U
Unipolar lead systems
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
connecting to patient cables . . . . . . . . . . . . . 41
Settings,rate dependent parameters . . . . . . . . . . . . 68
dual chamber . . . . . . . . . . . . . . . . . . . . . . 42
Setup indicators
precaution . . . . . . . . . . . . . . . . . . . . . . . . 16
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . 61
Unlocking temporary pacemaker
Shut-down message . . . . . . . . . . . . . . . . . . . . . 44
Lock/Unlock key . . . . . . . . . . . . . . . . . . . . . 45
Single chamber pacing
unipolar lead systems . . . . . . . . . . . . . . . . . 41

Technical Manual 103


Medtronic 5392

Upper Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 VOO


timing violations vs. RATE . . . . . . . . . . . . . . . 71 Mode selection . . . . . . . . . . . . . . . . . . . . . 29
timing violations vs. SAV, PVARP . . . . . . . . . . 70 VOO mode
Upper Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 pacing setup table . . . . . . . . . . . . . . . . . . . 85
manual adjustments . . . . . . . . . . . . . . . . . . 64 VSP (ventricular safety pace)
UPPER RATE specification . . . . . . . . . . . . . . . . . . . . . . . 83
mode transition . . . . . . . . . . . . . . . . . . . . . 77 V (ventricular) OUTPUT
nominal value . . . . . . . . . . . . . . . . . . . . . . 84 dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
range, increments . . . . . . . . . . . . . . . . . . . . 82 nominal value . . . . . . . . . . . . . . . . . . . . 29, 83
Upper screen range, increments . . . . . . . . . . . . . . . . . . . . 29
A OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . 28 V (ventricular) Sensitivity . . . . . . . . . . . . . . . . . . 61
RATE . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 nominal value . . . . . . . . . . . . . . . . . . . . . . 61
status scale . . . . . . . . . . . . . . . . . . . . . . . 28 vs. V OUTPUT . . . . . . . . . . . . . . . . . . . . . . 62
V OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . 29 V (ventricular) SENSITIVITY
V output . . . . . . . . . . . . . . . . . . . . . . . . . . 29 nominal value . . . . . . . . . . . . . . . . . . . . . . 83
Use, intended . . . . . . . . . . . . . . . . . . . . . . . . . . 9 range, increments . . . . . . . . . . . . . . . . . . . . 81
VVI mode
V Mode selection . . . . . . . . . . . . . . . . . . . . . 29
V-A interval
pacing setup table . . . . . . . . . . . . . . . . . . . 85
timing violations . . . . . . . . . . . . . . . . . . . . . 70
Ventricular pacing W
blanking specifications . . . . . . . . . . . . . . . . . 83 WARNING messages . . . . . . . . . . . . . . . . . . . . 63
refractory . . . . . . . . . . . . . . . . . . . . . . . . . 75 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . 12, 14
safety pacing . . . . . . . . . . . . . . . . . . . . . . . 75 Wenckebach response . . . . . . . . . . . . . . . . . . . 75
Viewing intrinsic rhythm
view using PAUSE . . . . . . . . . . . . . . . . . . . 46

104 Technical Manual


Medtronic, Inc. Europe/Africa/Middle East
710 Medtronic Parkway Medtronic International Trading Sàrl
Minneapolis, MN 55432 Route du Molliau 31
USA Case Postale 84
www.medtronic.com CH-1131 Tolochenaz
+1 763 514 4000 Switzerland
+41 21 802 7000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical Technical manuals
consultation for physicians and medical www.medtronic.com/manuals
professionals)
Bradycardia: +1 800 505 4636
Tachycardia: +1 800 723 4636

© Medtronic, Inc. 2013


M953303A001B
2013-06-05

*M953303A001*

S-ar putea să vă placă și