diagnostic ultrasound system

Product Data
No. MPDUS0017EA V1.0

outline
NemioTM MX (SSA-590A) is a compact full-digital color
diagnostic ultrasound system. This next-generation sys-
tem has achieved high image quality and high sensitivity
using advanced technologies.

• Superior imaging performance


Nemio MX includes THI (Tissue Harmonic Imaging) in the
standard configuration and generates high-quality B-mode
images. Advanced DYNAMIC FLOWTM, which produces
high-resolution blood flow images and is a feature of top-
of-the-line systems, and ApliPureTM, which is compounding
technology for high contrast resolution, are supported.
Thus, Nemio MX provides high-end imaging performance.

• Sophisticated ergonomics
A 15-inch LCD monitor whose brightness has been adjust-
ed appropriately for the display of ultrasound images is
used in Nemio MX. The displayed images are easy-to-
view and their excellent visibility is relatively insensitive to
the brightness of the room.
A flexible multijoint arm supports the LCD monitor, allowing
appropriate positioning for operations in the standing or
sitting posture to be achieved easily.
A lightweight body has been achieved through the use of
the LCD monitor, making it much easier to move the sys-
tem around in the hospital.

Functions can be assigned freely to buttons on the operat- • Efficient data management
ing panel to suit individual preferences. Also, operation A large-capacity HDD is provided in the standard configu-
protocols can be registered in advance, allowing examina- ration, facilitating efficient management of acquired images.
tions to be performed smoothly with simple operations Filed images can be output via the USB port or stored on
(Sonoset). CD/DVD.
In addition, the B-mode image quality can be adjusted DICOM is supported by the optional Advanced Data
appropriately with a single operation using the IP dial or Management board. In addition, Stress Echo and the
QuickScan. reporting function can be installed as options.
system matrix for SSA-590A
No. Unit Model name Remarks
Including color Doppler, DVD drive, THI, PW, Sector, 15-inch LCD monitor,
1 Main unit SSA-590A
QuickScan, Sonoset
<Options>
2 B-PSP & CDI-PSP/QSP kit UIQS-590A PSP for 2D/CDI, QSP for CDI

3 CW kit UICW-590A Enables CWD capabilities. Pencil transducer connector is included.

UJUR-590A Reference signal display unit + ECG cable


4 Reference signal kit
UJUR-592A PCG sensor
5 Transducer selector UIPS-590A Adds a 3rd transducer port.
6 M-TEE hanger kit UAEH-590A
7 Footswitch kit UZFS-590A Foot SW + connector kit for Main unit.
8 Gel warmer UZGW-590A
9 Data Management unit UIDM-590A Data Management unit with filing, DICOM storage, and report functions
Add QSP/PSP, 3rd transducer port, CW, Reference signal unit, Dynamic Flow
10 Premium kit UIUG-590A
and Directional Color Angio.
11 CV kit UICV-590A ECG, CWD (without pencil transducer support).
12 Cable for printer UZCA-590A For digital printers
Enables OB 4D scan using PVM-375MV.
13 Mecha 4D kit UIMV-590A
UIDM-590A, UIPS-590A, and UIQS-590A are required.
Mounting kit for peripheral
14 UZRI-590A For mounting a color printer or DVD recorder (with cables)
unit (for a single unit)
Mounting kit for peripheral
15 UZRI-591A For mounting a color printer and a DVD recorder (with cables)
unit (for 2 units)
<Optional software>
16 DCA kit USDA-590A Enables directional display in Color Angio mode (Directional Color Angio).
17 Trapezoid scan kit USTS-590A Enables trapezoid scan function.
18 HPRF kit USHP-590A Enables High PRF PW Doppler.
19 TDI kit USTD-590A
20 CHI/FEI kit USHI-590A
Enables Time-Intensity Curve generation.
21 TIC kit USTI-590A
USHI-590A is required.
22 ACM kit USAC-590A Automated Cardiac Flow Measurement
23 Dynamic Flow kit USDF-590A
<Options for UIDM-590A>
24 DICOM system kit UDDS-590A
25 Stress Echo kit UDSE-590A
26 Panoramic View kit UDPA-590A

2
MPDUS0017EA

transducer options
Freq. Multi-freq. (MHz)
Model name Scan type Application Remarks
(MHz) B-mode THI
Transcranial
PSM-20CT Sector 90° 2.0 3.0 2.5 2.0 3.8 3.6
Doppler
PSM-50AT Sector 90° 5.0 6.0 4.5 3.7 6.0 5.6 (Pediatric) Heart
PSM-70AT Sector 90° 7.0 10.0 7.5 5.0 – – Pediatric heart
PSM-30BT Sector 90° 3.0 3.7 3.0 2.0 4.0 3.6 Adult heart
PSM-37CT Sector 90° 3.75 5.0 4.2 3.2 4.8 4.6 Abdominal
PEF-510MA Sector 90° 5.0 6.0 5.0 3.7 6.0 5.6 Multiplane TEE
PEF-704LA Linear 34 mm 7.0 10.0 7.5 5.0 – – Laparoscopic *4
PVM-375AT Convex 70° 3.75 6.0 4.2 3.0 5.0 4.6 Abdomen
PVM-672AT Convex 62° 6.0 8.0 7.0 5.7 – – Neonatal head
PVM-662AT Convex 62° 6.0 8.0 7.0 5.7 – – Neonatal head *4
PVM-651VT Convex 143° 6.0 7.5 6.0 5.0 8.0 6.5 Endovaginal *1
PVM-740RT Convex 144° 7.0 10.0 7.5 5.0 – – Endorectal
PVM-787LA Convex 85° 7.0 10.0 7.5 5.0 10.0 8.0 Laparoscopic *1
PVM-620ST Convex 144° 6.0 7.5 6.0 4.0 – – Endocavitary
PVF-620ST Convex 144° 6.0 7.5 6.0 4.0 – – Endocavitary *4
Adult heart,
PVM-381AT Convex 75° 3.75 6.0 3.7 3.0 4.8 4.6
abdomen
Adult heart,
PVF-381MT Convex 75° 3.75 6.0 3.7 3.0 4.8 4.6 *4
abdomen
PVF-738H Convex 32° 7.0 10.0 7.5 5.0 – – Intraoperative
Abdominal, 
PVF-738F Convex 32° 7.0 10.0 7.5 5.0 – –
small parts
PVF-745V Convex 38° 7.0 10.0 7.5 5.0 – – Intraoperative
Linear 53 mm 7.0 12.0 9.0 6.0
PVL-715RT Convex Endorectal (Biplane)
7.0 10.0 7.5 5.0 – –
R53 143°
PLM-503AT Linear 32 mm 5.0 8.5 7.2 6.0 9.0 7.0 Vascular *1
PLM-703AT Linear 32 mm 7.5 11.0 9.0 6.0 11.0 9.0 Carotid *1
PLM-805AT Linear 56 mm 8.0 12.0 9.0 6.0 12.0 10.0 Small parts *1
Small parts
PLM-1202S Linear 25 mm 12.0 14.0 12.0 8.0 – –
Intraoperative
PLM-1204AT Linear 32 mm 12.0 14.0 12.0 8.0 – – Small parts
PLF-308P Linear 85 mm 3.75 5.0 4.0 3.0 – – Abdominal biopsy
PVM-375MV Convex 3.75 6.0 4.2 3.0 5.0 4.6 OB fetus *2
PC-19M Pencil 1.9 – – – – – Heart Doppler *3
PC-50V Pencil 5.0 – – – – – Peripheral Vascular *3

*1: UIQS-590A is required for THI.


*2: UIDM-590A, UIPS-590A, UIMV-590A, and UIQS-590A are required. Only for 3rd port.
*3 Premium kit and CV kit (UICV-590A) or UICW-590A is required.
*4 These transducers are no longer available for purchase. The transducers at the customer sites can still be used.

3
block chart systems
Nemio MX

Main unit
SSA-590A

B-PSP & CDI-PSP/QSP kit


UIQS-590A

CW kit
UICW-590A

Trapezoid scan kit


USTS-590A

Reference signal kit


UJUR-590A

Reference signal kit


UJUR-592A

Transducer selector
UIPS-590A A

M-TEE hanger kit


UAEH-590A DCA kit
USDA-590A

Footswitch kit
HPRF kit
UZFS-590A USHP-590A

Gel warmer
TDI kit
UZGW-590A
USTD-590A
Data Management unit
ACM kit
UIDM-590A
USAC-590A
DICOM system kit
UDDS-590A CHI/FEI kit
USHI-590A
Stress Echo kit
TIC kit
UDSE-590A
USTI-590A
Panoramic View kit
UDPA-590A Dynamic Flow kit
USDF-590A
Mecha 4D kit
UIMV-590A*
Mounting kit for Mounting kit for
4D transducer peripheral unit peripheral unit
PVM-375MV (for a single unit) (for 2 units)
UZRI-590A UZRI-591A

A * UIQS-590A and UIPS-590A are required.

4
MPDUS0017EA

SPECIFICATIONS • Image processing: 5 IP types are available.


The following parameters can be
System controlled.
• Transducer: M-series with Echo Booster
– Dynamic range: 30 to 100 dB (can be set inde-
(Chip-in-the-Tip)
pendently for 2D and M)
• Focusing: Autofocusing with Digital
– Edge enhancement: OFF, 1, 2, 3
Continuous Beam Former of
– Automatic gain
ROI-tracing type
control: 4 types
• Protocol Registration
– Postprocessing: 2D, 8 types
(SonoSet): Available
– Temporal smoothing: OFF, 1 to 5
– Spatial smoothing: 4 types
2D- and M-modes – Contrast
• Scanning methods: Electronic convex scan
enhancement: OFF, ON (can be set independ-
Electronic linear scan
ently for 2D and THI)
Electronic sector scan
• Frame rate: Max. 391 frames/s
Electronic trapezoid scan*
• Compounding Scan
* Available with PLM-703AT, (ApliPure): Available with all convex and lin-
PLM-805AT, PLM-1204AT,
ear transducers
PLM-503AT, PLM-1202S,
PLF-308P, PEF-704LA, PVL-715RT • Auto STC/GAIN
adjustment
• Scan width: WIDE (100%)/WIDE-MIDDLE
(QuickScan): Available
(80%)/MEDIUM (70%)/
NARROW (50%)/FREE (Min.
4D (UIDM-590A, UIPS-590A, UIMV-590A,
20°)
UIQS-590A, and PVT-375MV are required)
• 2D
• Acquisition modes: 4D Live, Single Sweep
– Depth: 2 to 24 cm, 1-cm steps
• Display modes: Single display of a 2D image,
– Line density: L, H
Single diplay of a 4D image,
– Independent
Single display of a planar image
steering angle: ±10°
of a 4D image, Dual display of a
– Multiple frequencies: Max. 5 (including THI)
4D image and a planar image,
• M-mode
4-frame display of a 4D image
– Channel: Single
and three planar images,
– Sweep speed: 1, 2, 4, 8 s/screen
Multiframe display of planar
– Display method: Moving bar
images of a 4D image
• Focusing
• 3D display modes: Real-time, Freeze
– Transmission: 8 steps max.
• Review modes: 4D volume cine playback, movie
– Reception: Continuous
play back, frame-by-frame play-
• Gain adjustment
back
– Palm controller: 62 to 100 dB for B-mode
• Measurements: Distance, angle, area
52 to 110 dB for M-mode
• Image storage: 4D volume, movie (dynamic
– STC (DGC): 8-step sliding VR’s.
image), still image
Corresponds to the viewing
• Control method: Operation from the main panel
depth
(switches, paddle switches,
rotary encoders)

5
Spectrum Doppler (SD) • Velocity range (scale)
• Mode: Pulsed-Wave Doppler (PWD), The following parameters can be changed with one dial.
Continuous-Wave Doppler with – Alternative step: 16
Pencil Transducer, Full Steerable – PRF: 1.3 to 12.5 kHz
Continuous-Wave Doppler – Color filter: 4 types
(SCWD) – Motion eliminator: 4 types
• Pulse repetition • Gain: 1 to 16 dB
frequency (PRF): 1.3 to 15.2 kHz, PWD 0.5-dB steps
5.8 to 15.2 kHz, HPRF • Display format: CDI, CDI/CDI
(USHP-580A is required) CDI+MCDI
4.0 to 47.6 kHz, CWD MCDI
• Gain: 31 dB (1-dB steps) • Image processing: Programmable image quality
• Doppler filter: PWD, 7 steps adjustment function; for
CWD, 7 steps Time Smooth (3 types),
• Sampling width: 13 steps (1 to 15 mm) Spatial Smooth (5 types)
• Display control: Reverse (toward/away flow) • Display control: Base shift, Color contrast,
display, Base shift, B-refresh B/W and color balance, CDI
(Manual, Time, ECG-Sync), ROI, Reverse
D-expand, 2D/D-expand, Angle • Independent steering
mark angle: ± 20° max (2 steps)
• Independent steering • Color density: L/M/H/SH
angle: ±20° max. (2 steps) • Frame rate: Max. 371 frames/s
• Audio stereo output (UIQS-590A is required)
• Display format: 2D+SD (vertical layout, horizon-
tal layout) Advanced Dynamic Flow (USDF-590A is
SD required)
• Auto adjustment: Base shift, Velocity Range • Applicable transducers: PVM-375AT, PLM-1204AT
• Real time Auto trace: Available PLM-703AT, PSM-20CT,
• Maximum measurable PSM-37CT, PSM-30BT
velocity PW: 6.75 m/s (±3.37 m/s) PLM-805AT
HPRF: 6.75 m/s (±3.37 m/s)
CW: 19.2 m/s (±9.65 m/s) CHI/FEI kit (USHI-590A is required)
• Auto Velocity Range and Zero Shift • Injection Timer: Available
Control (iDoppler): Available • Mode: 2D, Power, Dynamic Flow kit
(USDF-590A is required)
Color Doppler (CDI) • Flash Trigger: Manual, Internal, ECG
• Display mode: Velocity-Variance (3 types) • Multiple Flash: 1–15
Power (4 types) • Monitor Mode: Available
Directional Power (4 types) • Applicable Transducers: PSM-20CT, PSM-37CT,
(USDA-590A is required) PSM-30BT, PVM-375AT
Power variance (1 type)
Variance (1 type) Data Management unit (UIDM-590A is required)
Velocity HUE (4 types) • Acquiring and saving
Velocity SATURATION (3 types) ultrasound images: Acquires dynamic and still imag-
TDI HUE (2 types, option) es output from diagnostic ultra-
QUANT (1 type) sound systems used for routine
• Color-reversed: Available examinations and/or detailed
• Color contrast: 4 steps examinations in hospitals or
examination rooms and saves
such images.

6
MPDUS0017EA

• Acquisition methods: Acquisition of images into the • Output to the Printer


ultrasound workstation is per- – Output to the printer provided with the diagnostic ultra-
formed using either of two tech- sound system
niques: one is Pre-Trigger/Post- – Output to the network printer in DICOM format
Trigger acquisition, for standard – Output to local or network printers that are Windows
real-time images; and the other XP compatible
is IM-Transfer acquisition, in • Report
which the images stored in – WMS (Wall Motion
image memory of the diagnostic Scoring): Conforms to ASE.
ultrasound system are acquired – Import and display of various measurement results
one at a time in order to acquire – Display of data measured by the diagnostic ultrasound
images with a higher frame rate. system
• Displaying and editing – Paste of image
images: Displays on the monitor the – Entry of comment
images acquired as described
above and permits the user to Acoustic output power adjustment
display these images for four- • B/M, Doppler and Color
frame display, playback in syn- – Doppler output: 10% to 100%, 5% steps
chronization with the R wave of (varies according to transducer)
the ECG, etc. – Auto gain control: Automatic gain compensation
• Support of DICOM for lower acoustic output power
setting
functions (UDDS-590A
is required): Supports the following DICOM
Display
functions. • Gray scale level: 256
– Storage SCU/SCP* • Format
– Query/Retrieve SCU – B/M scan: 2D (dual 2D), 2D/M, M
– Print SCU* – Doppler (D-mode): 2D, 2D/M/D, 2D/D, M/D, D
– MWM SCU (Modality Worklist Management)* – Color Doppler: CDI (dual CDI),
– Storage Commitment SCU CDI/MCDI CDI/MCDI/D,
– Modality Performed Procedure Step SCU MCDI, MCDI/D, 2D/CDI
– Structured Reporting SCU • Observation monitor: 15-inch LCD color monitor
– Scheduled Workflow • Interface: DVI-D single link
– Patient Information Reconciliation • PAN/ZOOM (EXPAND)
– Veterans Affairs – ZOOM (EXPAND)
– Evidence Document method: Real-time/ROI specified (selected
– Measurement TAG using the preset only)
* UDDS-590A is not required. – Images for ZOOM
• Stress echo (EXPAND): Real-time images and FREEZE
(UDSE-590A is required): Provides the acquisition/display/ images
scoring functions which support – Magnification factor: Up to a viewing field of 2 cm
stress echo examinations in the – PAN method: Real-time images, FREEZE
hospital or laboratory. images and IM images
• Integrated operability: Can be operated from the main – Images for PAN: Real-time images, FREEZE
panel of the diagnostic ultra- images, and IM images
sound system and supports – Other types of control: The transmission focus is opti-
transmission/reception of data mized.
including patient IDs. EXPAND for real-time images
• Panoramic View optimizes the number of rasters.
(UDPA-590A is required): This function enables a continu- • Language: English, German, French,
ous image for a wide area to be American English, Italian,
displayed by moving the trans- Spanish, Portuguese,
ducer along the body surface. Norwegian, Finnish, Swedish
• Others: Display reversal: Left-Right,
Top-Bottom, 90°-step rotation.

7
Reference signal kit • Doppler:
This function provides reference signal display and ECG VTI measurement (Conti. Equation)
trigger mode in single mode. Flow volume measurement (F Vol VTI)
This mode requires the reference signal kit UJUR-590A. Velocity ratio measurement (RATIO)
• Signal: ECG (Lead II) Velocity ratio measurement (RI)
AUX Velocity measurement (F Vol Vel Trace)
PCG Mitral valve measurement
• ECG Trigger: 1 channel Mitral valve regurgitation measurement (MR)
• PCG Filter: 3 steps available Aortic valve function measurement
• ECG syncronization Aortic valve regurgitation measurement (AR)
setting: 0 to 1.94 s Tricuspid valve measurement
Tricuspid valve regurgitation measurement (TR)
Image memory Pulmonary valve measurement
• No. of images that can Pulmonary vein measurement
be recorded • Obstetric calculations:
– 2D mode: Up to 512 images – Up to 16 items can be measured, including fetal
– CDI mode: Up to 512 images growth, gestational age, estimated delivery date, etc.
– M mode – Trend graph function
MCDI mode • Report items: Patient ID, application measure-
PWD/CWD mode: Up to 120 s ment data, comment

Measurement Security functions


• B mode: Distance (8 ch), angle (8 ch), • Log on
area (8 ch), circumference (8 ch), – Maximum number of characters for user name:
volume (1 ch), cross distance (8 64 alphanumeric characters
ch), parallel distance (8 ch), hip – Maximum number of characters for password:
joint (8 ch), meanIMT (8 ch), and 64 alphanumeric characters
curve distance (8 ch) – Maximum number of users that can be registered:
• M mode: Distance (8 ch), time (8 ch), 64 users
slope (8 ch), heart rate (1 ch) – Entry of user names and passwords is case-sensitive.
• Doppler: Velocity (8 ch), time (8 ch), • Log off
acceleration (8 ch), heart rate (1 • Emergency Log-on
ch), velocity trace (8 ch)
• CDI Velocity histogram (2 ch), Image Export via USB
velocity (8 ch), profile (2 ch) • Memory Device
(Operation with all memory devices is not guaranteed.)
Application software
(calculation, measurement, etc.)
• Calculation of left ventricular (LV) function
Mode Method
2D mode Single-plane, Biplane, Bullet, Simpson,
Modified Simpson, Cube
M mode Cube, Teichholz, Gibson
• Measurement of mitral
valve (MV) function: M mode
• Measurement of aortic
valve (AV) function: M mode
• % Stenosis: 2D mode, M mode
• Prostate: 2D mode

8
MPDUS0017EA

Input/Output B/W Printer


• Transducer connectors: Max. 3 for standard electronic • Recommended B/W printer models
transducers – Sony: UP-897, UP-D897*
1 for pencil transducer – Mitsubishi: P93W, P93DW*, P95DW*
• Composite output signals * UZCA-590A is required
– Color: 1
– B/W: 1 Color Printer
• S-VIDEO input/output • Recommended color printer models
– Input: 2 – Sony: UP-25MD, UP-D23MD*
– Output: 2 UP-D25MD*
• RGB input/output – Mitsubishi: CP30W, CP30DW*
– Output: 1
* UZCA-590A is required
• DVI
– Output: 1
DVD Recorder
• Printer control signal • Recommended DVD recorder model
– Output: 2 – Victor BD-X201ME (PAL)
• Audio input/output BD-X201M (NTSC/PAL)
– Input: R+L
– Output: R+L
• RS-232C: 1 port
VCR/DVD recorder control
• Ethernet: 100Base-T 1
(UIDM-590A is required)
• USB: 1 (front)
4 (rear)
• Footswitch: 2 (front)
• Reference signal
– ECG: 1
– AUX: 1
– PCG: 1

Recording
• CD/DVD drive (supports DVD-R, DVD-RW, DVD+R,
DVD+RW, CD-R, CD-RW, and DVD-RAM)
• HDD image filing (80 GB)
• Color video printer
• B/W video printer
• DVD recorder

9
Operating conditions conformance standards
• Power supply requirements EU and other regions requiring compliance with European
– Line voltage: 100 VAC to 127 VAC (50 Hz) Directive 93/42/EEC and subsequent amendments
220 VAC to 240 VAC • General: EN60601-1 (2006)
– Line frequency: 50 Hz to 60 Hz • Particular: EN60601-2-37 (2008)
– Power capacity: 1030 VA (max.) • Acoustic power: Information for Manufacturers
• Environmental conditions Seeking Marketing Clearance of
– Operating conditions Diagnostic Ultrasound Systems
· Ambient and Transducers (2008) (FDA
temperature: 10°C to 35°C guidance)
· Relative humidity: 35% to 85% (no condensation) • EMC: EN60601-1-2 (2007)
· Atmospheric
Other regions
pressure: 700 hPa to 1060 hPa
• General: IEC60601-1 (1988), Amd. 1
– Storage and transportation conditions
(1991), Amd. 2 (1995)
· Ambient
• Particular: IEC60601-2-37 (2001), Amd. 1
temperature: -10°C to 55°C
(2004), Amd. 2 (2005)
· Relative humidity: 30% to 85% (no condensation)
• Acoustic power: Information for Manufacturers
· Atmospheric
Seeking Marketing Clearance of
pressure: 700 hPa to 1060 hPa
Diagnostic Ultrasound Systems
• Safety classification
and Transducers (2008) (FDA
– According to the type of protection against electric
guidance)
shock:
• EMC: IEC60601-1-2 (2001), Amd. 1
CLASS I EQUIPMENT
(2004)
– According to the degree of protection against electric
shock:
EQUIPMENT WITH TYPE-BF APPLIED PARTS dimensions and mass
– According to the degree of protection against harmful • External dimensions: Color systems (including cardio-
ingress of water: vascular color systems):
Ordinary EQUIPMENT (enclosed EQUIPMENT without 451 mm (W) × 785 mm (D) ×
protection against ingress or water) 1,276 to 1,488 mm (H)
– According to the degree of safety of application in the • Mass: Approximately 83 kg (peripheral
presence of a FLAMMABLE ANESTHETIC MIXTURE units and TEE stand not includ-
WITH AIR or WITH OXYGEN OR NITROUS OXIDE: ed)
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANESTHETIC MIXTURE WITH AIR or
WITH OXYGEN OR NITROUS OXIDE
– According to the mode of operation:
CONTINUOUS OPERATION

10
MPDUS0017EA

MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit Transducers
SSA-590A Main unit 83 (183.0) PC-19M Pencil transducer 0.11 (0.2)
Options/Accessories for main unit PC-50V Pencil transducer 0.08 (0.2)
BD-X201M DVD 4.8 (10.6) PEF-510MA Transesophageal transducer 1.5 (3.3)
BD-X201ME DVD 4.8 (10.6) PEF-704LA Laparoscopic transducer 0.94 (2.1)
CP30W Color printer 7.3 (16.1) PLF-308P Biopsy transducer 0.74 (1.6)
CP30DW Color printer 7.3 (16.1) PLM-503AT Linear transducer 0.7 (1.5)
P93W B/W printer 2.8 (6.2) PLM-703AT Linear transducer 0.7 (1.5)
P93DW B/W printer 2.8 (6.2) PLM-805AT Linear transducer 0.7 (1.5)
P95DW B/W printer 2.6 (5.7) PLM-1202S Linear transducer 0.56 (1.2)
UAEH-590A M-TEE hanger kit 1.6 (3.5) PLM-1204AT Linear transducer 0.7 (1.5)
UDDS-590A DICOM system kit 0.5 (1.1) PSM-20CT Sector transducer 0.6 (1.3)
UDPA-590A Panoramic View kit 0.5 (1.1) PSM-30BT Sector transducer 0.5 (1.1)
UDSE-590A Stress Echo kit 0.5 (1.1) PSM-37CT Sector transducer 0.6 (1.3)
UICV-590A CV kit 1.4 (3.1) PSM-50AT Sector transducer 0.6 (1.3)
UICW-590A CW kit 0.5 (1.1) PSM-70AT Sector transducer 0.6 (1.3)
UIDM-590A Data management unit 1.4 (3.1) PVF-381MT Convex transducer 0.55 (1.2)
UIMV-590A Mecha 4D kit 0.7 (1.5) PVF-620ST Endocavitary transducer 0.73 (1.6)
UIPS-590A Transducer selector 0.3 (0.7) PVF-738F Micro transducer 0.45 (1.0)
UIQS-590A B-PSP & CDI-PSP/QSP kit 0.1 (0.2) PVF-738H Micro transducer 0.45 (1.0)
UIUG-590A Premium kit 2.0 (4.4) PVF-745V Micro transducer 0.45 (1.0)
UJUR-590A Reference signal kit 0.9 (2.0) PVL-715RT Endocavitary transducer 1.4 (3.1)
UJUR-592A Reference signal kit 0.9 (2.0) PVM-375AT Convex transducer 0.8 (1.8)
UP-25MD Color printer 5.7 (12.6) PVM-375MV Convex transducer 0.9 (2.0)
UP-D23MD Color printer 6.5 (14.3) PVM-381AT Convex transducer 0.55 (1.2)
UP-D25MD Color printer 5.5 (12.1) PVM-620ST Endocavitary transducer 0.75 (1.7)
UP-897 B/W printer 2.6 (5.7) PVM-651VT Endocavitary transducer 0.9 (2.0)
USAC-590A ACM kit 0.1 (0.2) PVM-662AT Convex transducer 0.7 (1.5)
USDA-590A DCA kit 0.1 (0.2) PVM-672AT Convex transducer 0.55 (1.2)
USDF-590A Dynamic Flow kit 0.1 (0.2) PVM-740RT Endocavitary transducer 1.4 (3.1)
USHI-590A CHI/FEI kit 0.1 (0.2) PVM-787LA Laparoscopic transducer 1 (2.2)
USHP-590A HPRF kit 0.1 (0.2)
USTD-590A TDI kit 0.1 (0.2)
USTI-590A TIC kit 0.1 (0.2)
USTS-590A Trapezoid scan kit 0.1 (0.2)
UZCA-590A Cable for printer 0.5 (1.1)
UZFS-590A Footswitch kit 1.4 (3.1)
UZGW-590A Gel warmer 1 (2.2)
Mounting kit for peripheral unit
UZRI-590A 1.2 (2.6)
(for a single unit)
Mounting kit for peripheral unit
UZRI-591A 2.1 (4.6)
(for 2 units)

11
MPDUS0017EA

Toshiba Medical Systems Corporation meets internationally recognized


standards for Quality Management System ISO 9001, ISO 13485.

Toshiba Medical Systems Corporation Nasu Operations meets the


©Toshiba Medical Systems Corporation 2011 all rights reserved. Environmental Management System standard, ISO 14001.
Design and specifications subject to change without notice. Made for Life, Nemio, DYNAMIC FLOW, and ApliPure are trademarks of
2011-02 TME/KI Toshiba Medical Systems Corporation.