Sunteți pe pagina 1din 10

Biotribology 14 (2018) 1–10

Contents lists available at ScienceDirect

Biotribology
journal homepage: www.elsevier.com/locate/biotri

The development of a small-scale wear test for CoCrMo specimens with T


human synovial fluid

Stevenson H.a, , Parkes M.b, Austin L.a,1, Jaggard M.c, Akhbari P.c, Vaghela U.c, Williams H.R.T.c,
Gupte C.c, Cann P.a
a
Tribology Group Department of Mechanical Engineering, Imperial College London, UK
b
Biomechanics Group Department of Mechanical Engineering, Imperial College London, UK
c
Musculoskeletal Laboratory, Imperial College London, UK

A R T I C LE I N FO A B S T R A C T

Keywords: A new test was developed to measure friction and wear of hip implant materials under reciprocating sliding
Synovial fluid conditions. The method requires a very small amount of lubricant (< 3 ml) which allows testing of human
CoCrMo synovial fluid. Friction and wear of Cobalt Chromium Molybdenum (CoCrMo) material pairs were measured for
Friction a range of model and human synovial fluid samples. The initial development of the test assessed the effect of
Wear
fluid volume and bovine calf serum (BCS) concentration on friction and wear. In a second series of tests human
Artificial hip joints
synovial fluid (HSF) was used. The wear scar size (depth and volume) on the disc was dependent on protein
content and reduced significantly for increasing BCS concentration. The results showed that fluid volumes
of < 1.5 ml were affected by evaporative loss effectively increasing the protein concentration resulting in
anomalously lower wear. At the end of the test thick deposits were observed in and around the wear scars on the
disc and ball; these were analysed by Infrared Reflection-Absorption Spectroscopy. The deposits were composed
primarily of denatured proteins and similar IR spectra were obtained from the BCS and HSF tests. The analysis
confirmed the importance of SF proteins in determining wear of CoCrMo couples.

1. Synovial Fluid Lubrication and Wear of Prosthetic Joints from Cobalt-Chromium-Molybdenum (CoCrMo) Alloys [3]. Although
the volumetric wear of the joint was reduced there was an increase in
In 2015 there were over 180,000 [1] primary hip and knee re- the number of wear particles causing metallosis damaging the peri-
placements recorded in the United Kingdom. These devices relieve pain articular bone and soft tissue eventually requiring implant revision
and restore function in articular joints affected by destructive arthro- [4,5]. High failure rates meant MoM joint implantation had reduced
pathy resulting from disease, trauma or infection. Prosthetic joints are to < 0.1% of all primary hip procedures by 2014 [1] although this
tribological devices with bearing surfaces articulating under load, and combination (1–2%) is still used in resurfacing hip replacement [1].
thus are susceptible to high friction, wear, corrosion and fatigue. These The wear of implant materials are usually evaluated in bench top
issues can contribute to early failure necessitating implant revision and and joint simulator tests [6–18]. Bench-top tests [6–12] focus upon
in 2015 nearly 10% of hip replacements required revision [1]. Pros- material response to rubbing rather than the joint design factors. The
theses are increasingly being implanted into younger patients in- tests have a simple geometric configuration, usually pin-on-disc [6–8],
creasing the performance demands on joint due to higher activity levels pin-on-plate [9, 10] or pin/flat-on-sphere [11,12] which simulates in
and longer operational life. To achieve this, new implant materials and vivo contact pressures. The loading (usually constant) and kinematics
joint designs have been developed [2]. (unidirectional/bidirectional sliding) are chosen to provide a simple
The first generation hip joint was the Metal-on-Polymer (MoP) simulation of implant contact conditions. Wear of the pin or ball after a
however they gave poor longevity due to polyethylene wear and par- certain sliding distance and specific number of cycles [7–11] is usually
ticle debris resulting in osteolysis [3]. Implant lifespans of only evaluated by simple weight loss [7–11].
12–15 years made the implant suitable solely for older patients. Designs Hip and knee joint simulators replicate the complex loading and
switched to the MoM prosthesis where both bearing surfaces are made kinematics experienced in a physiological gait cycle [19] and are used


Corresponding author.
E-mail address: harriet.stevenson10@imperial.ac.uk (H. Stevenson).
1
Now at UK AWE.

https://doi.org/10.1016/j.biotri.2018.04.001
Received 23 November 2017; Received in revised form 29 March 2018; Accepted 6 April 2018
Available online 11 April 2018
2352-5738/ © 2018 The Authors. Published by Elsevier Ltd. This is an open access article under the CC BY license (http://creativecommons.org/licenses/BY/4.0/).
H. Stevenson et al. Biotribology 14 (2018) 1–10

Nomenclature HSF Human synovial fluid


i-EHL Isoviscous Elastohydrodynamic Lubrication
BCS Bovine Calf Serum Micro-IRRAS Micro - Infrared Reflection-Absorption Spectroscopy
CoC Ceramic on Ceramic MoM Metal on Metal
CoCrMo Cobalt Chromium Molybdenum MoP Metal on Polymer
CoF Coefficient of friction PP Periprosthetic synovial fluid
FTIR Fourier Transform Infrared Spectroscopy OA Osteoarthritis
HA Hyaluronic Acid RA Rheumatoid arthritis
HeSF Healthy synovial fluid SF Synovial Fluid

to evaluate the effect of implant design or material combinations. The performed. The protein concentration of PP (~30 mg/ml) was found to
test usually achieves millions of cycles before the cumulative wear is be significantly higher than that of bovine SF and dilution of BCS to
measured [14–18,20]. The measurement of wear on either the head or 40% would give similar total protein concentrations.
cup components is based on small changes in mass. This measurement The foregoing review has identified significant deficiencies in our
can also be affected by the environmental conditions over the lifetime understanding of synovial fluid lubrication of prosthetic joints. It raises
of the test [14–18,20] and so often has poor repeatability [20]. The two important questions:
joint specimens are usually submerged in Bovine Calf Serum (BCS)
which is changed regularly during the test, typically every 5 × 105 1. What is the effect of HSF composition on implant wear? Are some
cycles [17]. Thus completing a 5 million-cycle test requires very large patients/diseases/implant material combinations at a higher risk of
amounts (~ 5 l) of fluid [15,18]. implant failure as a result?
The focus of the majority of implant tribology testing is the effect of 2. Is 25 wt% BCS a suitable model lubricant to replace human synovial
implant design and material wear. The effect of synovial fluid chemistry fluid? If not can we define a better one?
on implant wear is rarely studied. Synovial fluid is the natural lubricant
found in articular joints. It is rarely available and only attainable in The overall aim of the study was to investigate how the friction and
small quantities, certainly less than the large volumes required for si- wear of implant materials are affected by SF chemistry including HSF.
mulator tests. Therefore, the lubricant model most communally used is The focus of this initial work was on the protein content as these are the
BCS diluted concentration of 25% w/w. It has been shown to have a most abundant molecular species in SF.
similar total protein content (18–21 mg/ml) when compared to healthy One of the practical problems is the very small volumes (< 5 ml) of
synovial fluid (HeSF) [21–23]. However, the overall chemical compo- HSF available from clinical procedures, typically joint effusion. The first
sition is different [21–23]. In addition diseased joints show large in- requirement therefore was to develop a test procedure which used very
creases in total protein content not mirrored by the model [23–27]. small volumes of fluid (< 3 ml) which would allow for some repeat
The chemical composition and physical properties of human syno- testing of samples from a single patient. We decided to concentrate
vial fluid varies between patients and healthy or diseased fluids initially on CoCrMo material pairs to understand if the fluid chemistry
[23–27]. The most abundant components of SF reported in the litera- might be a contributing factor to why this MoM produced such high
ture [22–26,28] are summarised in Table 1. The primary proteins are levels of wear for some patients but works very well for others. The
identified as albumin and γ-globulin and their relative and total con- MHRA [26] recently recommended routine monitoring of asympto-
centration can vary widely depending on the fluid pathology matic patients, highlighting a significant concern about implant life-
[23,24,26]. span and patient health. Thus a long term aim of the research is to
The influence of model SF composition upon implant wear (parti- develop a SF screening test identifying patients at risk of excess wear
cularly MoM), in both simulation tests and in vivo is unclear. This is problems with MoM and other implants.
driven, in part, by the idea [4,5] that some MoM hip designs operate The aims of the research program was as follows:
predominately in the fluid film (i-EHL) regime, this would reduce wear
rates to avoid clinical wear problems. In this regime lubricant film 1. Develop of simple bench test to measure friction and wear of im-
formation will not be affected by the biochemical properties of SF, such plant materials under simulated physiological conditions with small
as protein content. The lubricant film thickness is predicted to scale fluid volumes (~1–3 ml) as HSF volumes were restricted.
with the high-shear rate viscosity of the fluid [4,5] and the viscosity is 2. Investigate the effect of fluid volume, protein concentration and
fairly constant over the protein concentration range [29]. Therefore oxygen content on friction and wear using model SF.
protein concentration will have little effect on predicted i-EHL film 3. Investigate the use of HSF in the small volume wear test to examine
thickness and thus, by implication, wear. In contrast MoP hips are the validity of model SF for use in wear testing.
considered to operate in the boundary lubrication regime where the 4. Identify and characterise organic deposits in and around the wear
changing protein content and thus the boundary lubrication behaviour scar using an optical microscope and Micro InfraRed Reflection
might have an effect on wear [23]. Proteins (albumin and globulin) are
thought to adsorb at the implant surfaces and act as boundary additives Table 1
[6,30,31]. Thus most studies of the effect of SF chemistry on implant Comparison of model and human synovial fluid protein concentration and pH
friction and wear have been for UHMWPE components [6,31–35]. [21–26].
The vast majority of implant tribology testing uses BCS as the model
Albumin (mg/ Globulin (mg/ Approx. total pH
SF lubricant: There are very few instances of human synovial fluid ml) ml) protein (mg/ml)
(HSF) being used. Yao et al. [36] published one of the few papers to
report friction tests with HSF using a pin-on-disk tribometer. They HeSF 7–18 0.5–2.4 7–20 7.3–7.4
PP 20–39 4.6–15.4 43 25–55 7.5–8.5
compared friction measurements for different polymer pin materials
OA 17.8–18.5 12.9–13.4 30–34 7.4–8.1
rubbing against a CoCrMo disc with an array of lubricants: DI water, RA 18.6–28.0 4.8–23.5 22–52 6.6–7.6
varying concentrations of BCS diluted with DI water, bovine SF and 100 wt% 35.00 22.00 57 7.0–8.1
periprosthetic (PP) HSF. The PP HSF gave the lowest friction coefficient BCS
for all material combinations however no wear measurements were 25 wt% BCS 8.75 5.5 14 7.0–8.1

2
H. Stevenson et al. Biotribology 14 (2018) 1–10

Absorption Spectroscopy (Micro IRRAS). and end of the test. This method was chosen as it only requires a very
small amount of fluid and there are no cleaning/decontamination issues
2. Materials and Methods when using HSF.

2.1. Friction and Wear Tests 2.2. Optical and Micro IRRAS Examination of Wear Scar Deposits

A High Frequency Reciprocating Rig (HFRR) (PCS Instruments At the end of the test low power microscope images were taken from
London) was used to measure friction and wear of CoCrMo specimens the disc before cleaning with detergent in order to carry out the wear
with BCS and HSF. The HFRR has been used extensively to study fric- measurement. The ball was carefully rinsed with water and then air
tion and wear of boundary lubricants [37]. The test parameters can be dried. An example of the disc wear scar is shown in Fig. 2.
adjusted to provide a slow speed reciprocating contact where hydro- The ball was retained for further analysis by Micro-IRRAS. This
dynamic effects are minimised. A similar test configuration has been technique provides information on the organic composition of films
used in earlier studies [13]. deposited on metal surfaces [43] and is used extensively to study the
A schematic diagram and photograph of the experimental set up is conformation and structure of proteins [44–47].
shown in Fig. 1. The upper specimen consists of a CoCrMo ball oscil- The sampling method is shown in Fig. 3. A Perkin Elmer Multiscope
lating against a stationary lower specimen (CoCrMo disc) using an Infrared Microscope coupled to a Spectrum One FTIR was used to take
electromagnetic vibrator. The upper specimen is attached to a pivoted Micro-IRRAS spectra from a small area of organic deposits in and
ridged shaft and loaded via a dead weight and a test fluid is injected around the wear scar on the ball. The sample diameter was 100 μm, 100
into the lower specimen holder to lubricate the contact. The fluid bath scans were taken and the resulting spectrum was baseline corrected,
was made of nylon rather than stainless steel to prevent contamination smoothed (11 point) and normalised (to absorbance 0.1 at 1650 cm−1)
of the test fluid by metal ions. A temperature-controlled plate beneath using the standard Perkin Elmer software. Before spectra were taken the
the lower specimen was used to heat the lubricant bath to 37 °C which ball was lightly rinsed with deionised water and allowed to dry, this
was monitored by a thermocouple immersed in the fluid. The test removed excess test fluid but left the tightly bound deposits.
conditions are summarised in Table 2. Ball and disc CoCrMo test spe-
cimens (19.05 mm diameter ball, 10 mm diameter disc) were polished 2.3. Wear Scar Analysis
to a mirror finish (as obtained from PCS Instruments, UK) with ar-
ithmetic average surface roughness 39.6 ± 1.5 nm and 8.8 ± 0.8 nm A White Light Interferometer (Wyko NT9100) was used to measure
respectively. the volume and depth of the wear scar formed on the disc during
Before testing the metal specimens were cleaned in 1% detergent rubbing tests. Images were obtained using a 20× magnification ob-
solution and rinsed (3 times) in distilled water. The specimens were jective (at 0.55 FOV). On each disc multiple images were taken and
then stored under isopropanol. Just before testing they were air-dried stitched together to enable examination of the full length of the wear
and cleaned in an oxygen plasma (Diener) which removes organic scar. An example of the final image is shown in Fig. 4. The wear volume
contamination and leaves a hydrophilic surface [38]. was obtained from the negative volume computed by the Wyko analysis
The CoCrMo ball was loaded and reciprocated against the disc to software. In addition wear depth readings at 30%, 50% and 70% along
simulate joint articulation. A frequency of 20 Hz and stroke length of the length of the scar were taken and averaged. The section and loca-
1 mm were used. The sliding speed varied from 0 (at the end of the tions of these are also shown in Fig. 4. Before scans were performed any
stroke) to approximately 42 mm/s at the mid stroke which covers the surface deposits on the discs were removed by soaking in 2% detergent
normal speed range of the hip gait cycle [39]. A load of 1 N was used, solution followed by rinsing with water (3×) and isopropanol and air
giving a mean Hertzian pressure of 216 MPa, this is high for a walking
gait cycle (MoM ~ 99.8 MPa max). However it has been calculated that
during running a maximum load of 10.3BW [40] (MoM ~ 141 MPa
max) can be experienced in the hip joint. Whilst the initial test contact
pressure is high this will rapidly decrease as the contact area increases
due to wear. The test was run for 72,000 cycles (1 h). Tests with the BCS
fluids were repeated at least twice, this was not possible for all the HSF
tests (due to the limited volumes of human samples) but repeats were
run where feasible.
The contact was immersed in a small volume (1–2 ml) of SF. The
depth of fluid above the sample was approximately ~5 mm for the 2 ml
volume. A lightweight cover placed over the fluid bath was used to
reduce evaporation during the test. The test device was housed in a
plastic box which ensured a stable temperature environment as one of
the practical challenges was to establish an equilibrium temperature
without significant loss of the fluid through evaporation. The ball and
disc were assembled and the fluid bath heated to the desired tem-
perature. Once this was reached the test fluid was introduced. The box
was then purged with N2 for 20 min to reduce the atmospheric oxygen
content to physiological levels (24 mmHg, GOX 100T Greisinger)
[41,42] The disadvantage of this method was it extended the total test
time and the continued flow of gas also contributed to increased eva-
poration. In later work the effect of this was investigated by running
tests without the purge but heating for the 20 min and omitting the
20 min delay but running the test once the fluid was introduced.
Friction was recorded throughout the test. At the end of the test the
samples were drained of excess fluid. The fluid pH was measured by Fig. 1. Schematic diagram of HFRR device and photograph of the loaded ball
narrow range litmus paper (pH 6.5–10 VWR MColorHast) at the start and fluid bath.

3
H. Stevenson et al. Biotribology 14 (2018) 1–10

Table 2 was used as the reference test fluid. On receipt the 500 ml sample was
HFRR test conditions. divided into 7 ml aliquots and then refrozen and stored at −40 °C. An
Test conditions aliquot was removed the day before testing, defrosted overnight in a
Max Hertzian pressure 200 MPa refrigerator and then diluted (deionised water) to the required con-
Frequency 20 Hz centration before testing.
Stroke length 1 mm HSF samples were collected by MJ and his team during surgical
Temperature 37 °C
Duration 72,000 cycles
procedures. Sample SF from patients with joint effusions were not in-
Test specimens cluded. Immediately after collection the sample is centrifuged for
Material CoCrMo 15 min at 10000G to remove large debris and cellular material. Samples
Ball 19.05 mm diameter Ra are divided into ~1 ml aliquots and frozen (−80 °C) until required for
Disc 10 mm diameter Ra
testing. Ethical approval was obtained and subjects gave informed
consent for all human sample collection and testing. Results are shown
for SF samples from four patients, sample data is shown in Table 3.

3. Results and Discussion

3.1. Validation of Tribology Test Method: Effect of Fluid Volume for BCS
Reference Fluids

One of the aims of the work was to develop a wear test which only
required a small volume of fluid. To validate the test method a series of
initial tests were run and these included the effect of fluid volume on
wear. Fig. 5 shows example friction traces for BCS 25 wt% for different
fluid volumes (1 and 2 ml). The BCS fluids give low, stable friction
traces over the test period, typically COF was in the range
μ = 0.22–0.26. However sample volume appeared to affect friction as
CoF1ml < CoF2ml. For volumes of < 1.5 ml friction coefficient was
Fig. 2. Low power microscope image of wear scar on the disc specimen for
found to decrease compared to tests of > 1.5 ml, for example with BCS
25 wt% BCS after cleaning.
25 wt% the CoF averages for 1 and 2 ml tests are 0.223 ( ± 0.005) and
0.248 ( ± 0.007) respectively.
drying. WLI images of typical wear scars on the disc are shown in Fig. 6.
Results are shown for 25 wt% 1 and 2 ml. The wear scar is wider and
2.4. Test Fluids deeper for the 2 ml samples compared to the 1 ml sample. A series of
BCS 25 wt% tests were run for a range of fluid volumes between 1 and
Bovine calf serum (Sigma Aldrich Poole, UK12133C sterile-filtered) 2 ml and the wear results are shown in Fig. 7. We surmise this is due to

Fig. 3. Micro-IRRAS analysis of organics deposits on the ball.

4
H. Stevenson et al. Biotribology 14 (2018) 1–10

Fig. 4. WLI wear scar measurement on the disc.

Table 3 Evaporation of the test fluid appeared to influence measured wear


HSF samples. for some test fluids suggesting that a reduction in overall test time
Test fluid Patient gender Age Disease pH SF appearance
which includes pre-heating and rubbing time would be advantageous.
Up to now tests have been run with a N2 purge which gave the longest
Start End test time, two further series of tests were run to examine the influence
of reducing this. Wear results for (2 ml) BCS 25 wt% were compared for
HSF77 Male 69 OA 8.3 9.0 Light yellow
HSF41 Female 59 OA + RA 8.3 8.8 Light yellow
three test protocols:
HSF145 Female 78 OA + RA 8.5 9.5 Light yellow
HSF 135 NA NA NA 7.7 8.5 Light yellow 1. N2 purge “Nitrogen” (standard test used so far)
BCS25wt% 7.9 8.5 2. 20 min delay with heating/no purge: “20 min hold” which mimics
the N2 heating profile but without the gas flow
3. No delay/no purge: “Final”

Fig. 9b compares results from these protocols. The “nitrogen” pro-


tocol gave slightly higher wear values but there was little difference
between the other protocols. Therefore future tests (“Final” protocol)
will be started immediately after test fluid is added with no nitrogen
purge to reduce evaporation effects.
An increase in pH of each of the fluids was recorded however this
will be examined in more detail in a future paper.
In conclusion a bench-test, which uses a small volume of fluid, has
been successfully developed to measure wear of CoCrMo specimens.
The minimum fluid volume which can be used is 2 ml, for volumes less
than this evaporation increases the effective protein concentration
which affects both the wear and friction. The new test will now be used
to evaluate the wear on implant materials with human synovial fluid.

Fig. 5. Friction traces for BCS 25 wt% for different fluid volumes. 3.2. Friction and Wear Results: Human Synovial Fluid

In a second series of tests the friction and wear using HSF samples
evaporation effectively changing the protein concentration in the low
were investigated using the final configuration of 2 ml sample without
volume/low BCS concentration tests. Evaporation occurs preferentially
N2 purge. The average friction coefficient was similar for all tests and
from the upper surface of the fluid thus creating locally higher con-
was in the range μ = 0.24–0.27. Average wear depth results are shown
centrations of proteins close to the rubbing contact, which resulted in
in Fig. 10, error bars are shown were possible for repeat tests. The HSF
lower wear volumes for the 1 ml samples but had a negligible effect on
wear (both average depth and volume) was > 25 wt% BCS for all
the 2 ml tests. To further investigate the effect of protein concentration
samples.
a series of tests for 1 and 2 ml fluid volumes were run for a range of BCS
concentrations (10, 25, 50, 100 wt%).
Average CoF results are summarised in Fig. 8 for different BCS 3.3. Examination of Organic Deposits in and around the Wear Scar
concentrations. For low BCS concentrations (10, 25 wt%)
CoF1ml < CoF2ml but for the higher concentrations (50 and 100 wt%) Microscope images from the ball are shown in Fig. 11 for 25 wt%
CoF was similar for both volumes. Wear depth and wear volume are BCS and HSF77 samples. Surface deposits were clearly visible around
plotted against BCS concentration for different fluid volumes in Fig. 9a. the wear scar on the ball, these were generally seen as smooth, co-
The wear results are similar for the high concentrations tests (50, loured, “halo” films. The optical interference colours indicated film
100 wt%) but diverge significantly at low concentrations (10, 25 wt%). thickness to be in the range 100–400 nm. Chromatic optical inter-
This result is in line with the CoF values shown in Fig. 8 assuming these ferometry has been used extensively to measure EHL films in lubricated
can be used as an indicator of wear, in this case higher average friction contacts [42] The HSF “halo” films appeared to be thinner and less
equates to higher wear. Generally CoF cannot be used as an ersatz wear extensive than the BCS films. Beyond the “halo” films more extensive
measurement, particularly for very different chemical systems, however black, particulate deposits were observed. These images were taken
for these BCS results there does appear to be a CoF/wear correlation. after light rinsing to remove excess fluid. The wear scar is clearly seen
Error margins shown are the standard deviation of reported results. in the centre of the photograph. At either end of the wear scar smooth
Indications are given on the wear depth graph for approximate range of “halo” films are observed, this can be seen more clearly in the right
protein concentration for healthy and diseased SF. hand side image in Fig. 11.
Micro-IRRAS spectra were taken from the surface deposits, both the

5
H. Stevenson et al. Biotribology 14 (2018) 1–10

Fig. 6. WLI images of wear scars on discs: 25% BCS 1 and 2 ml.

“halo” and black particulate films, on the ball after rinsing. Similar absorbing frequencies in the range of 1510–1580 cm−1 [44,45,48].
spectra were obtained from both these regions however the black par- Amide I and II vibrations are less affected by the nature of the side chain
ticulate films were generally too thick to give good results. The “halo” however they are influenced by the secondary structure of the backbone
deposit spectra were compared to FTIR spectra from the untested fluids. [44–46,48]. Therefore it follows that Amide I and II vibrational bands
These were sampled as dried films on a clean CoCrMo surface and ex- are used for secondary structure analysis, although interpretation of the
amples are shown in Fig. 12a for the HSF and BCS 25 wt% fluids. The Amide II band information is less straightforward than Amide I [44].
spectral range is limited to the Amide I and II peak region Generally HSF Amide I bands were above 1655 cm−1 and for BCS below
(1850–1450 cm−1). In this region the infrared absorption spectrum of 1645 cm−1, seen in Fig. 12, which is line with the greater γ-globulin
proteins is dominated by bonds on the polypeptide backbone of the content in BCS compared to HSF.
amino acid structure (44). Spectra from fresh fluid and deposited “halo” films are compared in
The Amide I peak is the most prominent absorbing at frequencies in Fig. 13a for BCS 25 wt% and Fig. 13b for HSF041. In all the BCS tests
the range of 1600–1700 cm−1 and arises mainly due to the C]O the Amide I main peak position decreases from ~1640 cm−1 for the
(~80%) stretching vibration [45] with contributions from the out of fresh sample to ~ 1630 cm−1 for the wear scar deposits. The Amide II
phase CN stretching vibration [44]. The Amide II peak is a combination band also broadens and the maximum peak position shifts to a lower
of the NH bending (~60%) and CN stretching vibration (~40%) wavenumber (1542 to 1530 cm−1). IR spectra for HS041 films also

Fig. 7. Average wear depth plotted against fluid volume for BCS 25 wt%.

6
H. Stevenson et al. Biotribology 14 (2018) 1–10

relatively short time which permits testing of small fluid volumes. At


first sight the test does not appear to closely simulate the tribology
conditions in a hip implant as it uses a stroke length of 1 mm and fre-
quency of 20 Hz. However over the stroke cycle the speed varies from
zero at turn round to approximately 42 mm/s which covers the speed
range experienced in the walking cycle [39]. The contact region is
subject to bidirectional (reciprocation) shear but there is no cross-shear
which occurs in full hip joint motion [49]. The importance of cross-
shear in determining the wear of UHMWPE is well known [50], but is
unclear for MoM joints. Pin-on-disc testers usually provide a con-
tinuous, unidirectional sliding motion [6–8] where an excess of fluid is
present. In contrast the HFRR test provides a reciprocating motion
where the Hertzian contact diameter is relatively large (76 μm) com-
pared to the stroke length (1 mm). However this is comparable to the
contact conditions in hip simulators which typically have a similarly
high contact diameter/stroke length ratio. In addition the ball comes to
Fig. 8. Summary of average friction values for different BCS concentrations 1
and 2 ml fluid volumes (with N2 purge).
a halt at either end of the stroke simulating stance or an interruption of
the gait cycle. Hadley [20] in her review of MoM hip simulator studies
noted that the wear rate measured in vitro were often much lower than
in vivo and suggested this was due to the continuous nature of the si-
mulator motion rather than active and inactive periods typical of
human behaviour. The increase of wear in vivo was attributed to the
collapse of the i-EHL film at zero speed so that minimal surface pro-
tection was present due to inactivity. In her PhD Hadley [20] reported
the development of simulator test profiles which incorporated a period
of stance (stop-dwell-motion) under load which was found to increase
the wear due to a collapse of the i-EHL film. The HFRR introduces a
“stop” for each stroke and as such a collapse of the i-EHL film occurs
before re-entering into the cavity unlike a uni-directional test. This
potentially provides a more severe wear condition compared to the
continuous motion tests.
To obtain measureable wear rates it was necessary to test for a large
number of cycles, in this case 72,000. One of the concerns was the effect
of fluid evaporation on wear as this would be expected to change the
protein composition. For the 2 ml volume tests measured wear in-
creased with decreasing protein concentrations below BCS 50 wt%,
above this value wear did not vary with concentration. The wear de-
pendence on protein concentration meant the lower BCS content/lower
fluid volume tests were susceptible to evaporation effects. This artifi-
cially increased the BCS content resulting in anomalously low wear
values and was associated with the low volume tests (< 1.5 ml). The
final test condition uses 2 ml sample volume (without N2 purge) which
was not susceptible to evaporation effects.
The work with the BCS fluids (2 ml samples) showed wear de-
creasing with increasing protein concentration levelling off at > 50 wt
% BCS (~28 mg/ml protein concentration). The role of proteins in
forming a thicker and more protective lubricant film has been reported
in a number of papers [6,51–53]. As the HSF samples tested in this
study were all OA/RA pathologies with estimated total protein con-
centrations in the range 22–52 mg/ml they would be expected to give
lower wear than the BCS 25 wt% reference fluid (20 ml/mg). However
Fig. 9. Effect of test parameters: (a) fluid volume and BCS concentration on
in all tests the wear (volume and average depth) was much greater than
wear scar depth (b) N2 purge “Nitrogen”), 20 min delay with heating/no purge
(“20 min hold”) and no delay/no purge (“Final”). BCS 25 wt%, in some cases 70% higher. Optical imaging of the films on
the ball showed far fewer deposits for the HSF samples compared to
BCS, which can be seen in Fig. 11.
shows changes in the Amide I and II peak positions and shapes after The formation of thick protein deposits at either end of the wear
testing. The main Amide I peak shifts to a lower wavenumber scar suggests that high viscosity material is formed during sliding, it
(1644 cm−1) and other peaks are seen at 1650, 1632 and 1612 cm−1. collects in the inlet region [30,51,52,54] and is deposited at the turn-
The Amide II band becomes broader and more complex with distinct round. The formation of solid [11,55] or high viscosity [54] deposits
peaks at 1559, 1542, 1521 and 1508 cm−1. has been widely reported and these are thought to contribute to in-
creased surface separation [53] and wear reduction [11]. FTIR analysis
3.4. Discussion of the deposits showed they were predominately proteinaceous which is
to be expected with the BCS fluids but was also evident for the HSF
The paper has reported the development of a wear test which re- samples. It is suggested that SF proteins are denatured during the
quires very small volumes of fluid. The HFRR test uses a sliding, re- rubbing process forming insoluble agglomerations which are deposited.
ciprocating motion which provides a high number of cycles in a IR spectra were taken of dried fluid films before testing and compared

7
H. Stevenson et al. Biotribology 14 (2018) 1–10

Fig. 10. Average wear depth results for human synovial fluid and 25 wt% BCS (no N2 purge).

to post-test deposits and important differences were observed. Different role in the lubrication process? These aspects will be studied further in
protein profiles were observed for the fresh HSF fluids and BCS; this is a future papers.
fairly trivial observation but still needs to be emphasised, even from the The author's recognise that there are several limitations in the
simplest analysis these fluids have very different protein chemistries. present study. First of all HSF contains many components not only
BCS is characterised by broad Amide I and II bands at 1644 cm−1 and proteins. Therefore to better simulate in vivo conditions the effect of
1543 cm−1 respectively. The HSF samples gave different spectra both in phospholipids and hyaluronic acid on friction and wear should be in-
the position and shape of the Amide I and II bands; this can be seen in vestigated and will be included in future work. Secondly at present
Fig. 12. This difference in protein chemistries may change the ability of exact protein concentration and composition values are not available
the fluid to from a protective lubricant film explaining why higher wear for the HSF fluids. This is important for direct comparisons to model
was seen for HSF despite the higher protein content. The long-term goal fluids. Thirdly a full interpretation of FTIR spectra for both HSF and
is to link HSF chemistry to implant wear and thus risk of failure and BCS. Many studies [58–60] have suggested that FTIR spectra can be
FTIR spectroscopy would provide a convenient and rapid clinical ana- used to diagnose different pathologies and in future work deconvolu-
lysis tool. tion and curve fitting techniques will be used to provide a more detailed
IRRAS analysis also showed that there were significant differences interpretation of the spectra. Finally it should be noted that although
when compared post-test deposits and dried untested fluids. The most only MoM material combinations are currently reported the test pro-
obvious change was a shift in the Amide I band position to lower wa- cedure can be extended to include ceramic bearings allowing the study
venumber. Typically this was from 1641 to 1630 cm−1 for 25 wt% BCS of CoC material combinations.
and is associated with the formation of β-sheet structures (44–46). In
nature β-sheet proteins are associated with animal “glues” for example
barnacles sticking to metal surfaces [56] where this material is typically 4. Conclusions
very strong high viscosity, insoluble and adherent [57]. The formation
of this material in a sliding contact raises a number of questions: is this A bench test was developed to measure implant material wear
behaviour representative of in vivo MoM implant lubrication? What is which uses small volumes of fluid which allowed HSF to be tested. Our
the mechanism of formation? And does this material play a significant findings are summarised below:

Fig. 11. Microscope optical images from ball wear scar (scale marker 0.2 mm) showing different amounts of protein deposits: HSF77 (LHS), BCS 25 wt% (RHS).

8
H. Stevenson et al. Biotribology 14 (2018) 1–10

Fig. 12. Micro IRRAS spectra from untested human synovial fluid samples and BCS.

3. Significantly higher wear was recorded for the HSF samples com-
pared to 25 wt% BCS fluid. All the HSF samples tested would be
expected to have higher protein concentrations and thus give lower
wear.
4. IRRAS analysis of deposits formed during the wear test indicated
they were predominately composed of proteins. Significant differ-
ences were observed between fresh films (untested) and post-test SF
deposits indicating denaturing of the proteins.
5. The protein deposits had a higher β-sheet content and appeared to
be adherent and solid when dried. It is suggested proteins denatured
during rubbing and contributed to reduced wear of the disc.

Acknowledgements

The authors wish to thank Professor J. Lindon and Claire L Boulangé


(Computational and Systems Medicine Imperial College London) for
their contribution to the research project and to the UK EPSRC for fi-
nancial support of Ms. Stevenson (EP/M506345/1) and Dr. Parkes (EP/
N025059/1).

Conflict of Interest

The authors have no conflicts of interest to declare.

Human Ethics and Consent

The work was carried out under human ethics REC reference 15/
LO/0388 and patient consent was obtained for all human sample
testing.

References

[1] 14th Annual Report 2017 national Joint Registry for England, Wales and Northern
Ireland, 2016.
Fig. 13. Micro-IRRAS spectra from 25 wt% BCS and HSF041 films on the ball: [2] S. Santavirta, Y.T. Konttinen, R. Lappalainen, A. Anttila, S.B. Goodman, M. Lind,
comparison of untested and post-test deposits. et al., Materials in total joint replacement, Curr. Orthop. 12 (1) (1998) 51–57.
[3] G. Isaac, J. Thompson, S. Williams, J. Fisher, Metal-on-metal bearings surfaces:
materials, manufacture, design, optimization, and alternatives, Proc. Inst. Mech.
1. With the new test configuration a fluid volume of 2 ml gave re- Eng. H J. Eng. Med. 220 (2) (2006) 119–133.
peatable wear results which were not affected by evaporation. [4] D. Dowson, Z.M. Jin, Metal-on-metal hip joint tribology, Proc. Inst. Mech. Eng. H
220 (2) (2006) 107–118.
2. Wear on the disc decreased with increasing BCS concentration le- [5] D. Dowson, Tribological principles in metal-on-metal hip joint design, Proc. Inst.
velling off at ~50 wt% BCS which is approximately 28 mg/ml total Mech. Eng. H J. Eng. Med. 220 (2) (2006) 161–171.
protein concentration. [6] M. Heuberger, M. Widmer, E. Zobeley, R. Glockshuber, N. Spencer, Protein-medi-
ated boundary lubrication in arthroplasty, Biomaterials 26 (10) (2005) 1165–1173.

9
H. Stevenson et al. Biotribology 14 (2018) 1–10

[7] V. Saikko, A multidirectional motion pin-on-disk wear test method for prosthetic (1) (2006) 149–158.
joint materials, J. Biomed. Mater. Res. 41 (1) (1998) 58–64. [34] Y.S. Liao, P.D. Benya, H.A. McKellop, Effect of protein lubrication on the wear
[8] J. Cooper, D. Dowson, J. Fisher, Macroscopic and microscopic wear mechanisms in properties of materials for prosthetic joints, J. Biomed. Mater. Res. A 48 (4) (1999)
ultra-high molecular weight polyethylene, Wear 162 (1993) 378–384. 465–473.
[9] J. Tipper, P. Firkins, E. Ingham, J. Fisher, M. Stone, R. Farrar, Quantitative analysis [35] J.-M. Brandt, K. Charron, L. Zhao, S.J. MacDonald, J.B. Medley, Calf serum con-
of the wear and wear debris from low and high carbon content cobalt chrome alloys stituent fractions influence polyethylene wear and microbial growth in knee si-
used in metal on metal total hip replacements, J. Mater. Sci. Mater. Med. 10 (6) mulator testing, Proc. Inst. Mech. Eng. H J. Eng. Med. 226 (6) (2012) 427–440.
(1999) 353–362. [36] J. Yao, M. Laurent, T. Johnson, C. Blanchard, R. Crowninshield, The influences of
[10] R. Varano, J. Bobyn, J. Medley, S. Yue, The effect of microstructure on the wear of lubricant and material on polymer/CoCr sliding friction, Wear 255 (1) (2003)
cobalt-based alloys used in metal-on-metal hip implants, Proc. Inst. Mech. Eng. H J. 780–784.
Eng. Med. 220 (2) (2006) 145–159. [37] D. Wei, H. Spikes, S. Korcek, The lubricity of gasoline, Tribol. Trans. 42 (4) (1999)
[11] M. Wimmer, J. Loos, R. Nassutt, M. Heitkemper, A. Fischer, The acting wear me- 813–823.
chanisms on metal-on-metal hip joint bearings: in vitro results, Wear 250 (1) (2001) [38] S. Tang, O.-J. Kwon, N. Lu, H.-S. Choi, Surface characteristics of AISI 304L stainless
129–139. steel after an atmospheric pressure plasma treatment, Surf. Coat. Technol. 195 (2)
[12] M.A. Wimmer, C. Sprecher, R. Hauert, G. Täger, A. Fischer, Tribochemical reaction (2005) 298–306.
on metal-on-metal hip joint bearings: a comparison between in-vitro and in-vivo [39] O. Calonius, V. Saikko, Analysis of relative motion between femoral head and
results, Wear 255 (7–12) (2003) 1007–1014. acetabular cup and advances in computation of the wear factor for the prosthetic
[13] Y. Yan, A. Neville, D. Dowson, Biotribocorrosion of CoCrMo orthopaedic implant hip joint, Acta Polytech. 43 (4) (2003).
materials—assessing the formation and effect of the biofilm, Tribol. Int. 40 (10) [40] G. Bergmann, F. Graichen, A. Rohlmann, Hip joint loading during walking and
(2007) 1492–1499. running, measured in two patients, J. Biomech. 26 (8) (1993) 969–990.
[14] S.L. Smith, D. Dowson, A.A. Goldsmith, The effect of femoral head diameter upon [41] P.S. Treuhaft, D.J. McCarty, Synovial fluid pH, lactate, oxygen and carbon dioxide
lubrication and wear of metal-on-metal total hip replacements, Proc. Inst. Mech. partial pressure in various joint diseases, Arthritis Rheumatol. 14 (4) (1971)
Eng. H. 215 (2) (2001) 161–170. 475–484.
[15] D. Dowson, C. Hardaker, M. Flett, G.H. Isaac, A hip joint simulator study of the [42] P.M. Cann, H.A. Spikes, J. Hutchinson, The development of a spacer layer imaging
performance of metal-on-metal joints: part II: design, J. Arthroplast. 19 (8) (2004) method (SLIM) for mapping elastohydrodynamic contacts, Tribol. Trans. 39 (4)
124–130. (1996) 915–921.
[16] J. Medley, J. Krygier, J. Bobyn, F. Chan, A. Lippincott, M. Tanzer, Kinematics of the [43] P. Cann, The “leaves on the line” problem—a study of leaf residue film formation
MATCO™ hip simulator and issues related to wear testing of metal-metal implants, and lubricity under laboratory test conditions, Tribol. Lett. 24 (2) (2006) 151–158.
Proc. Inst. Mech. Eng. H J. Eng. Med. 211 (1) (1997) 89–99. [44] A. Barth, Infrared spectroscopy of proteins, Biochim. Biophys. Acta Bioenerg. 1767
[17] J. Bowsher, A. Hussain, P. Williams, J. Shelton, Metal-on-metal hip simulator study (9) (2007) 1073–1101.
of increased wear particle surface area due to ‘severe'patient activity, Proc. Inst. [45] M. Jackson, H.H. Mantsch, The use and misuse of FTIR spectroscopy in the de-
Mech. Eng. H J. Eng. Med. 220 (2) (2006) 279–287. termination of protein structure, Crit. Rev. Biochem. Mol. Biol. 30 (2) (1995)
[18] D. Dowson, C. Hardaker, M. Flett, G.H. Isaac, A hip joint simulator study of the 95–120.
performance of metal-on-metal joints: Part I: The role of materials, J. Arthroplast. [46] J. Kong, S. Yu, Fourier transform infrared spectroscopic analysis of protein sec-
19 (8, Supplement) (2004) 118–123. ondary structures, Acta Biochim. Biophys. Sin. 39 (8) (2007) 549–559.
[19] ISO, Implants for surgery - wear of total hip-joint prostheses Part 1: Loading and [47] V.P. Gupta, T.A. Keiderling, Vibrational CD of the amide II band in some model
displacement parameters for wear-testing machines and corresponding enviro- polypeptides and proteins, Biopolymers 32 (3) (1992) 239–248.
mental conditions for tests, (2014). [48] L.K. Tamm, S.A. Tatulian, Infrared spectroscopy of proteins and peptides in lipid
[20] M.J.F. Hadley, Long Term Functional Simualtion of Large Diameter Metal on Metal bilayers, Q. Rev. Biophys. 30 (04) (1997) 365–429.
Hip Implants, The University of Leeds, 2012. [49] B.S. Ramamurti, C.R. Bragdon, D.O. O'Connor, J.D. Lowenstein, M. Jasty,
[21] Aldrich S. Certifivate of Analysis, Catalog Number 12133C, Lot Number 11N238. In: D.M. Estok, et al., Loci of movement of selected points on the femoral head during
SAFC, editor. normal gait: three-dimensional computer simulation, J. Arthroplast. 11 (7) (1996)
[22] K.M.N. Oates, W.E. Krause, R.L. Jones, R.H. Colby, Rheopexy of synovial fluid and 845–852.
protein aggregation, J. R. Soc. Interface 3 (6) (2006) 167–174. [50] L. Kang, A.L. Galvin, T.D. Brown, Z. Jin, J. Fisher, Quantification of the effect of
[23] A. Mavraki, P.M. Cann, Friction and lubricant film thickness measurements on si- cross-shear on the wear of conventional and highly cross-linked UHMWPE, J.
mulated synovial fluids, Proc. Inst. Mech. Eng. J. J. Eng. Tribol. 223 (3) (2009) Biomech. 41 (2) (2008) 340–346.
325–335. [51] J. Fan, C. Myant, R. Underwood, P. Cann, Synovial fluid lubrication of artificial
[24] S. Ghosh, D. Choudhury, N.S. Das, B. Pingguan-Murphy, Tribological role of sy- joints: protein film formation and composition, Faraday Discuss. 156 (0) (2012)
novial fluid compositions on artificial joints — a systematic review of the last 69–85.
10 years, Lubr. Sci. 26 (6) (2014) 387–410. [52] C. Myant, R. Underwood, J. Fan, P.M. Cann, Lubrication of metal-on-metal hip
[25] H. Fam, J.T. Bryant, M. Kontopoulou, Rheological properties of synovial fluids, joints: the effect of protein content and load on film formation and wear, J. Mech.
Biorheology 44 (2) (2007) 59–74. Behav. Biomed. Mater. 6 (0) (2012) 30–40.
[26] G.E. Perlmann, M.W. Ropes, D. Kaufman, W. Bauer, The electrophoretic patterns of [53] A. Mavraki, P.M. Cann, Lubricating film thickness measurements with bovine
proteins in synovial fluid and serum in rheumatoid arthritis, J. Clin. Invest. 33 (3) serum, Tribol. Int. 44 (5) (2011) 550–556.
(1954) 319–325. [54] J. Fan, C.W. Myant, R. Underwood, P.M. Cann, A. Hart, Inlet protein aggregation: a
[27] D. Mazzucco, R. Scott, M. Spector, Composition of joint fluid in patients undergoing new mechanism for lubricating film formation with model synovial fluids, Proc.
total knee replacement and revision arthroplasty: correlation with flow properties, Inst. Mech. Eng. H J. Eng. Med. 225 (7) (2011) 696–709.
Biomaterials 25 (18) (2004) 4433–4445. [55] M.R. Widmer, M. Heuberger, J. Vörös, N.D. Spencer, Influence of polymer surface
[28] A. Galandáková, J. Ulrichová, K. Langová, A. Hanáková, M. Vrbka, M. Hartl, et al., chemistry on frictional properties under protein-lubrication conditions: implica-
Characteristics of synovial fluid required for optimization of lubrication fluid for tions for hip-implant design, Tribol. Lett. 10 (1–2) (2001) 111–116.
biotribological experiments, J. Biomed. Mater. Res. B Appl. Biomater. 105 (6) [56] D.E. Barlow, G.H. Dickinson, B. Orihuela, J.L. Kulp III, D. Rittschof, K.J. Wahl,
(2017) 1422–1431. Characterization of the adhesive plaque of the barnacle Balanus amphitrite: amy-
[29] V. Sharma, A. Jaishankar, Y.-C. Wang, G.H. McKinley, Rheology of globular pro- loid-like nanofibrils are a major component, Langmuir 26 (9) (2010) 6549–6556.
teins: apparent yield stress, high shear rate viscosity and interfacial viscoelasticity [57] P.J. Cheung, G.D. Ruggieri, R.F. Nigrelli, A new method for obtaining barnacle
of bovine serum albumin solutions, Soft Matter 7 (11) (2011) 5150–5160. cement in the liquid state for polymerization studies, Mar. Biol. 43 (2) (1977)
[30] Myant C, Cann P. In contact observation of model synovial fluid lubricating me- 157–163.
chanisms. Tribol. Int. 2013;63(0):97–104. [58] D.I. Ellis, R. Goodacre, Metabolic fingerprinting in disease diagnosis: biomedical
[31] J.-M. Brandt, K. Mahmoud, S. Koval, S. MacDonald, J. Medley, Antimicrobial agents applications of infrared and Raman spectroscopy, Analyst 131 (8) (2006) 875–885.
and low-molecular weight polypeptides affect polyethylene wear in knee simulator [59] H. Eysel, M. Jackson, A. Nikulin, R. Somorjai, G. Thomson, H. Mantsch, A novel
testing, Tribol. Int. 65 (2013) 97–104. diagnostic test for arthritis: multivariate analysis of infrared spectra of synovial
[32] A. Wang, A. Essner, G. Schmidig, The effects of lubricant composition on in vitro fluid, Biospectroscopy 3 (2) (1997) 161–167.
wear testing of polymeric acetabular components, J. Biomed. Mater. Res. B Appl. [60] R. Shaw, S. Kotowich, H. Eysel, M. Jackson, G. Thomson, H. Mantsch, Arthritis
Biomater. 68 (1) (2004) 45–52. diagnosis based upon the near-infrared spectrum of synovial fluid, Rheumatol. Int.
[33] M. Gispert, A. Serro, R. Colaco, B. Saramago, Friction and wear mechanisms in hip 15 (4) (1995) 159–165.
prosthesis: comparison of joint materials behaviour in several lubricants, Wear 260

10

S-ar putea să vă placă și