Sunteți pe pagina 1din 6

OKP 421598

OKTIS - 2

OCTANE NUMBER METER


FOR MOTOR GASOLINES

User Manual

Version 2.1 - 2016

WWW.OKTIS.RU
This translation was made by the Chamber of Commerce and Industry of
the Vladimir Region. 2016
11
USER INFORMATION
REPAIR REQUEST FORM
The meter is easy to use, and you will easily see it for yourself. for warranty repair of OKTIS-2
The unique feature of the meter is that you may check
Please make warranty repair of OKTIS-2 meter
octane number in the moment of filling fuel at the gas station. Factory number ___________________
After filling 2 - 5 liters in the tank through the metering pipe of the Fault description:
meter, you may decide if you should fill more or not if, when buying AI-95, ___________________________________________________________
your meter shows that its octane number is, for example, 88,0. Most ___________________________________________________________
probably, this gas station is selling fuel with octane number artificially ___________________________________________________________
increased with octane-increasing additives. Practice shows that acceptable ___________________________________________________________
deviation of octane number is within ±1.5 units of the number claimed by ___________________________________________________________
the gas station.
In case you always buy gasoline at the same gas station, periodically Information on meter owner:
checking your fuel helps you be sure that its quality remains the same, or Family name
reveal a sudden change of gas characteristics. First name
Patronymic
Postal code
Region
Rayon
City, settlement
Street , Building Apart.
Phone
Thank you for your interest to our meter; before use, please E-mail address
read first of all sections 1, 4, 6 and 7.
Meter delivery kit is as per with Section 3 of this manual.
Owner _____________ (signature) Date _____________
Please mail your meter with filled request form at the address:
OOO KAMELIT, p.o. Box 12, Vladimir, 600028
NOTICE: Always install batteries with ВКЛ/ВЫКЛ. (ON/OFF) Please by email to info@oktis.ru your notification on mailing your meter
button 5 in ВЫКЛ. ( OFF) position. and Russian Postal Service tracking number (14 digits).

11 10
ADDITIONAL INFORMATION 3
1. PURPOSE
To set the meter in expanded mode, after pressing button ‘5’ and the
appearance of a ticker bar in the middle of the display, using a thin object, press OKTIS-2 is intended for the determination of motor gasoline octane
button 8 “CORR”. number in Russian gasoline standards AI-80, AI-92, AI-95 and AI-98.
Ticker tape messages will appear in the lower display. Further, with a 2-3 sec Principle of meter operation is based on the dependence between dielectric
interval
permeability of gasoline and its research octane number (RON).
the following data will be displayed:
• measured value in nominal units;
The meter can operate in 2 modes: in continuous-flow mode while
• measured value in nominal units with correction for gas temperature; filling car tank at the gas station and in sensor immersion mode.
• gasoline temperature; The meter is used as an indicator permitting the car owner to obtain
• gasoline research (octane) number (RON) with correction for gas temperature.
information on the main parameter of gasoline quality.
If in course of meter operation after the ticker bar a fixed 3-element bar will
appear in the middle of the display, it means that electrical conductivity of the The meter IS NOT INTENDED for use as a measuring instrument
gasoline is above the limit due to the presence of impurities (e.g. water). We do not when resolving disputes concerning gasoline quality at the gas station.
recommend filling such gas in the tank, especially, in winter.
In Section 11 you can see a figure placed in brackets. It characterizes the 2. TECHNICAL SPECIFICATIONS
condition of dry sensor in the air and is entered in meter memory prior to its
calibration. This figure is specific for each meter. If you want to check it, please
Range of measured octane levels 70 to 110
switch on a dry meter in the air and set it into expanded mode. The meter will
check sender condition. Measured value in nominal units with correction for gas Minimum quantity of gas required for octane number
temperature shall correspond to the figure in brackets. If the two figures are not the determination in continuous-flow mode 2-5l
alike, wash the sensor and /or use measurement program correction mode. Time required to prepare the meter for operation, sec not more than 15
Minimum quantity of gas required for octane number
NOTE: To remove from sensor deposits formed due to gasoline determination in sensor immersion mode 80 ml
evaporation, it is recommended to clean the sensor (see p.8) at least once in Power supply voltage (two AA batteries) 3V
two months or more often, depending on the frequency of meter use.
Consumed current not exceeding 100 mА
Operating conditions:
According to Letter of 23.03.09 № 101-кс/616 from Russian - temperature of checked gasoline 0°С to +50°С
Research Institute of Certification the meter does not require - incumbent temperature -10°С to +40°С
compulsory certification in GOST R Certification System. - humidity at 30°С, % 30 to 80
Mass not exceeding 0,36 kg

Telephone of Technical Support Service


4
8-4922-44-73-81.
( from 10.00 to 18.00, Moscow Time)
3. DELIVERY KIT
In case a workmanship defect is found in the meter within warranty
• Carton 1 pc period, Manufacturer shall repair the defects free of charge, provided the
• Octane number meter 1 pc following conditions are met:
• User manual 1 pc • the meter shall be used only in accordance with this Manual;
• Instructions on the removal of deposits 1 pc • the warranty shall not be extended for a meter damaged by fire, due to misuse, and
aggressive liquids or water getting inside the meter.
• Set of spare seals 1 pc
The warranty shall be lost and warranty repair shall not be made if
• Case 1 pc
there are visible signs of intrusion, violation of factory assembly, any
• AA Battery 2 pcs
improvement or modification.
Faulty parts which have been replaced are the property of
4. SAFETY MEASURES
the Manufacturer.
Use the meter in the open air or in a well ventilated room preventing
gasoline vapor concentration.
Warranty and post-warranty maintenance and replacement of faulty
When working with the meter, IT IS FORBIDDEN TO:
meter shall be made within warranty period. To repair the meter under
• supply power to the meter from any outside source.
warranty, customer shall fill the order form (see page 11). After the
• change batteries when the meter is in contact with gasoline;
warranty period is expired, the meter may be repaired on a paid basis.
• use the meter near a source of open fire.
“I have read and understood maintenance rules, and I have no claims as
After the use, please shake the meter several times to shake off drops
of the appearance of the meter”.
of gasoline from sensor surface in measuring tube and replace the meter in
Buyer signature __________________________
its case.
11. SALE REGISTRATION.
Never use acetone or other solvents for washing the metering tube.
OKTIS-2 Octane number meter for motor gasoline, Manuf. №
________________ fully meets technical specifications listed in this
5. DESIGN AND PRINCIPLE OF OPERATION.
Manual, performs its functions, has been checked by Manufacturer, and
The meter is designed as a one-piece device (see fig.1) comprising three
produces no harmful effect on the environment or people.
main units. The body includes compartment 1 containing electrical parts and
(________) _________________________________________________
transition compartment 2 with seal rings. Metering tube 2 with sensor 3 are
secured in the transition compartment. Transition compartment 2 and Date of manufacture _____________ Signature of responsible person_____________
metering tube 3 form a channel either containing or passing through a flow
of gas, and compartment 1 contains electronic unit with numeric display 4 Sale Date _________ 20____. Signature of Salesperson _____________________
and
batteries. Seal of Retailer company (seal)

9 8
8. TROUBLESHOOTING. When gasoline reaches sensor in tube 3, sensor parameters are changing;
microprocessor compares them with data stored in memory, and digital
If, after the meter is switched on, ticker bar instead of '00' displays a display 4 shows figures corresponding to gasoline octane number. Meter is
stationary three-element message, it means that meter sensor in tube 3 is not switched ON by pressing button 5 once, until it locks. Repeated pressure
fit for work (metal chips, dust, drops of water, or other foreign matter). In on the button switches the meter off.
this case, wash the sensor in metering tube 3 in gasoline, shake off the Microprocessor program supports basic mode, expanded mode, reading
remaining drops, purge with air, and dry. For the complete removal of correction for gas temperature mode and reading correction for sensor
contamination after washing, you may use paper strips. Sensor may be contamination mode.
washed in rectified rubber solvent gasoline (NEFRAS-S2-20\120
TU38.401-67-108-92) KALOSHA by repeated dipping of the tube in a
vessel containing gasoline; gasoline level in the vessel should be above
mark 7. After '00' appears in the display, the meter is ready for work.

9. STORAGE AND TRANSPORTATION


The meter shall be stored and transported in factory case with closed
cover.
No dust, acid and alkali vapors, aggressive gases or other corrosive
impurities should be present in meter storage places.
The meter may be transported by any mode of transport.
When the meter is not used for a long time (over 2 months), please
remove batteries.
Do not leave the meter in direct sunlight for more than 1 hour.
10. WARRANTY LIABILITY AND Fig. 1
WARRANTY MAINTENANCE RULES 1 – electrical compartment
Manufacturer guarantees replacement of a meter which gets out of order 2 – transition compartment with seal rings
within 24 months after the meter was sold, provided the meter was operated 3 – metering tube
in compliance with this User manual. When buying the meter, ask the 4 – digital display
salesperson to fill Section 11 (Sale Registration). If this section is not filled 5 – ON/OFF button
or filled incorrectly, no claim for meter quality is accepted and no warranty 6 - cover
repair is made. 7 – gasoline level mark
5 8 – correction and mode switchover button
9 – calibration button (no access)
6
6. PREPARING FOR OPERATION
Press button 5 at compartment 1. Make sure there is a ticker bar in the
• Push button 5 ON/OFF in OFF position. middle of the display and take the readings. Ignore the first reading and
• Remove cover 6. wait for the next 2-3. Then switch off the meter, remove it from gasoline
• Observing the polarity, install two batteries (type АА). and shake off the remaining fuel from the meter.
• Set the cover in place and fasten it with screws. In course of measurements, two-or three-digit octane number values
• Press button 5 at compartment 1 until it locks. are displayed with increment of one. To obtain more accurate information
A ticker bar will appear in the middle of the digital display 4, and in a on measurement, you can press button 8 KOPP (CORR) (see p.10), and
few moments figures ‘00’ will be displayed. The meter is ready for work. values with increment of 0.1 will be displayed.
“РАЗ” (RAZ) message on the digital display indicates that batteries are Measurement cycles are repeated 10 times; then, the display dims, and
discharged. They should be replaced. an intermittent sound signal is heard, reminding you that you should
switch the meter off.
7. METER OPERATION. Often gasoline contains dissolved foreign matter (in particular, (motor
oil, fuel oil, diesel oil and other poorly evaporating substances), remaining
The meter may work in two main modes: after gasoline is evaporated, and accumulating at sensor plates. Such
• continuous-flow mode with gasoline flowing along metering tube 3; contamination affects initial properties of the sensor and results in too
• sensor immersion mode with metering tube immersion in vessel with high meter readings.
gasoline. To eliminate this problem, the meter has measurement program
correction mode, taking into account sensor contamination. To make a
For operation in continuous-flow mode, install the gas station fuel gun correction, press button 8 “CORR”, using a thin object; press button 5 to
as far as it would go into transition compartment 2 through rubber seals. switch the meter on, and release button 8. After a double sound signal,
Install metering pipe 3 in fuel tank filler, press button 5 at compartment 1, lower part of the display will show:
achieve even flow of gas through tube 3 by fuel flow regulator on the fuel • two ticker bars,
gun without actuating fuel gun shutoff valve. In a few moments ticker bar in • absolute value of sensor resistance deviation from factory setting in
the middle of display 4 will be replaced by figures indicating octane number standard units,
of gasoline flowing through the meter. Ignore the first reading and wait for • ЗАП (ZAP) message
the next 2-3. Then stop filling, switch off the meter, remove it from fuel Deviation will be stored in meter memory and will be factored in while
gun and shake off the remaining fuel from the tube. displaying octane number value.
This mode allows the possibility of monitoring gasoline quality in Deviation value “000” means that the sensor is clean.
course of filling. If there are air bubbles in flowing gas, meter readings will Each subsequent correction eliminates the previous one. Only the last
be too low. correction matters in measurement.
For operation in sensor immersion mode, place metering pipe 3 ………………………………………………………………………….