Documente Academic
Documente Profesional
Documente Cultură
WHAT IS MEASURABLE
AND MAKE MEASURABLE
THAT WHICH IS NOT
'ALILEO 'ALILEI
Synthos 3000
Microwave Synthesis
Reference Manual
Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.
Nor is any liability assumed for damages resulting from information contained in the document.
Anton Paar GmbH reserves the right to content changes. This also extends to changes to delivery
volumes or any features of delivered parts.
All rights reserved (including translation). This document, or any part of it, may not be reproduced,
changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm
or any other process) without prior written permission by Anton Paar GmbH.
Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked
as such. They are the property of their respective owner.
E-Mail: info@anton-paar.com
Web: www.anton-paar.com
Date: 12.2006
C49IB08A 1
6.6.1 Configuration ......................................................................................................... 75
6.6.2 Printer Setup ......................................................................................................... 76
6.6.3 Serial Port Setup ................................................................................................... 78
6.6.4 Safety Cooling ....................................................................................................... 79
6.6.5 Rotors .................................................................................................................... 80
6.6.6 Sensors ................................................................................................................. 81
6.6.7 Accessories ........................................................................................................... 82
6.7 Service .................................................................................................................................. 83
6.7.1 Software Update .................................................................................................... 84
6.7.2 Power Calibration .................................................................................................. 84
6.7.3 Power Check ......................................................................................................... 85
6.7.4 Temperature Sensor Calibration with the Calibration Accessory .......................... 86
6.7.5 Temperature Sensor Calibration with the Calibration Unit ..................................... 94
6.7.6 Temperature Sensor Check with the Calibration Accessory ............................... 109
6.7.7 Temperature Sensor Check with the Calibration Unit ...........................................110
6.7.8 Temperature Sensor Settings ...............................................................................112
6.7.9 IR Sensor Calibration with the Calibration Accessory ..........................................114
6.7.10 IR Sensor Calibration with the Calibration Unit ....................................................118
6.7.11 IR Sensor Check with the Calibration Accessory ................................................ 122
6.7.12 IR Sensor Check with the Calibration Unit .......................................................... 124
6.7.13 Solvent Sensor Settings ...................................................................................... 125
6.7.14 Exhaust Unit Installation ...................................................................................... 127
6.7.15 Library Maintenance Tool .................................................................................... 127
6.7.16 Advanced Service Routines ............................................................................... 129
6.8 Info ...................................................................................................................................... 130
6.9 AP Softprint ......................................................................................................................... 131
7 Troubleshooting .......................................................................................................................... 132
7.1 Software .............................................................................................................................. 132
7.1.1 Error Message Menu - List of Error Codes .......................................................... 132
7.1.2 Pop-up Error Messages in the Case of an "Alarm" ............................................. 134
7.1.3 Pop-up Messages During System Check ............................................................ 138
7.1.4 Pop-up Error Messages During System Check ................................................... 139
7.1.5 Pop-up Error Messages Under Various Conditions ............................................. 143
7.1.6 Pop-up Error Messages During "Power Calibration" ........................................... 147
7.1.7 Pop-up Error Messages During "Temp. Sensor Calibration" and "Temp. Sensor
Check" ................................................................................................................. 149
7.2 Rotor 16 and Rotor 8S ........................................................................................................ 151
7.3 Sensors .............................................................................................................................. 152
8 Rotor 48MF50 .............................................................................................................................. 153
8.1 Technical Description of Rotor 48 ....................................................................................... 153
8.1.1 Vessels for Rotor 48 ............................................................................................ 155
8.2 Limitations of Operation ...................................................................................................... 156
8.2.1 Number of Vessels .............................................................................................. 156
8.2.2 Temperature profiles ............................................................................................ 157
8.2.3 Power Profiles ..................................................................................................... 157
8.2.4 Fan Settings ........................................................................................................ 157
8.3 Prior to the First Reaction ................................................................................................... 158
8.4 Performing a Reaction ........................................................................................................ 159
8.4.1 Adding Reagents and Solvents ........................................................................... 159
8.4.2 Closing the Vessels ............................................................................................. 160
8.4.3 Loading Rotor 48 ................................................................................................. 163
8.4.4 Performing a Synthesis Run ................................................................................ 165
8.4.5 Unloading Rotor 48 ............................................................................................. 167
8.5 Cleaning and Maintenance ................................................................................................. 170
C49IB08A 2
8.5.1 Cleaning of the Reaction Vessels ........................................................................ 170
8.5.2 Cleaning After Every Run .................................................................................... 171
8.5.3 Checking After Every Run ................................................................................... 174
8.5.4 Standard Cleaning and Maintenance Procedures ............................................... 175
8.5.5 Procedures After the Rupture of a Safety Disk ................................................... 179
8.5.6 Procedures After a Vessel Explosion .................................................................. 185
8.6 Maintenance of the Pressure / Temperature Sensor M50 .................................................. 195
8.6.1 Replacing Seal and Safety Disk .......................................................................... 195
8.6.2 Replacing the Thermometer Capillary ................................................................. 200
8.6.3 Replacing the Immersion Tube ............................................................................ 202
9 Rotor 16MF100 / 16HF100 .......................................................................................................... 205
9.1 Technical Description of Rotor 16 ....................................................................................... 205
9.1.1 Vessels for Rotor 16 ............................................................................................ 207
9.2 Prior to the First Reaction ................................................................................................... 209
9.2.1 To Prepare the First Run ..................................................................................... 209
9.3 Performing a Synthesis ....................................................................................................... 210
9.3.1 Adding Reagents and Solvents ........................................................................... 210
9.3.2 Assembling the Vessels .......................................................................................211
9.3.3 Loading Rotor 16 ................................................................................................. 220
9.3.4 Performing a Synthesis Run ................................................................................ 221
9.3.5 Unloading Rotor 16 ............................................................................................. 223
9.4 Cleaning and Maintenance ................................................................................................. 228
9.4.1 Cleaning of the Reaction Vessels ........................................................................ 228
9.4.2 Cleaning After Every Run .................................................................................... 229
9.4.3 Checking After Every Run ................................................................................... 231
9.4.4 Standard Cleaning and Maintenance Procedures ............................................... 234
9.4.5 Procedures After the Breakage of a Safety Disk ................................................. 244
9.4.6 Procedures After a Vessel Explosion .................................................................. 247
9.4.7 Intensive Cleaning of Ceramic Vessels ............................................................... 253
9.5 Maintenance of the Pressure / Temperature Sensor Accessory ......................................... 255
9.5.1 Replacing Seal and Safety Disk .......................................................................... 255
9.5.2 Replacing the Thermometer Capillary ................................................................. 260
9.5.3 Replacing the Immersion Tube ............................................................................ 263
10 Rotor 8SXF100 / 8SXQ80 ............................................................................................................ 265
10.1 Technical Description of Rotor 8S ....................................................................................... 265
10.1.1 Vessels for Rotor 8S ............................................................................................ 267
10.2 Prior to the First Reaction ................................................................................................... 270
10.2.1 To Prepare the First Run ..................................................................................... 270
10.3 Performing a Synthesis ....................................................................................................... 271
10.3.1 Adding Reagents and Solvents ........................................................................... 271
10.3.2 Assembling the Vessels ...................................................................................... 272
10.3.3 Loading Rotor 8S ................................................................................................ 281
10.3.4 Performing a Synthesis Run ................................................................................ 285
10.3.5 Unloading Rotor 8S ............................................................................................. 287
10.4 Cleaning and Maintenance ................................................................................................. 294
10.4.1 Cleaning of the Reaction Vessels ........................................................................ 294
10.4.2 Cleaning After Every Run .................................................................................... 295
10.4.3 Checking After Every Run ................................................................................... 297
10.4.4 Standard Cleaning and Maintenance Procedures ............................................... 300
10.4.5 Procedures After the Rupture of a Safety Disk ................................................... 312
10.4.6 Procedures After a Vessel Explosion .................................................................. 316
10.4.7 Intensive Cleaning of Ceramic Vessels ............................................................... 323
10.5 Maintenance of the T-Probe and the Immersion Tube with Seal XF / XQ .......................... 324
10.5.1 Replacing the Seal and Safety Disk .................................................................... 324
C49IB08A 3
10.5.2 Replacing the Thermometer Capillary ................................................................. 328
11 Applications ................................................................................................................................ 332
11.1 General Information ............................................................................................................ 332
11.1.1 Basics of Microwave Heating .............................................................................. 333
11.1.2 Hazards During Microwave Synthesis ................................................................. 335
11.1.3 Reagents ............................................................................................................. 338
11.2 Performing Synthesis .......................................................................................................... 347
11.2.1 General ................................................................................................................ 347
11.2.2 Operation Without an Internal Temperature Sensor ............................................ 348
11.2.3 Operating With Sensors ...................................................................................... 348
11.2.4 Interpretation of Results ...................................................................................... 351
11.2.5 Power Reduction by the Instrument ................................................................... 351
12 Appendix ...................................................................................................................................... 353
12.1 Addresses ........................................................................................................................... 353
12.2 Packing the Synthos 3000 and the Rotors ......................................................................... 354
12.2.1 Packing the Instrument ........................................................................................ 354
12.2.2 Packing the Rotors .............................................................................................. 358
12.3 Technical Data of the Instrument ........................................................................................ 360
12.3.1 Synthos 3000 Microwave Oven ........................................................................... 360
12.3.2 Options ................................................................................................................ 362
12.3.3 Rotor 48MF50 ..................................................................................................... 362
12.3.4 Rotor 16 MF100 .................................................................................................. 363
12.3.5 Rotor 16 HF100 ................................................................................................... 363
12.3.6 Rotor 8SXF100 .................................................................................................... 364
12.3.7 Rotor 8SXQ80 ..................................................................................................... 364
12.4 List of Spare Parts .............................................................................................................. 365
12.5 Safety Certification .............................................................................................................. 366
12.6 CE Declaration of Conformity ............................................................................................. 368
12.7 Safety Declaration for Repair and Malfunction Report ....................................................... 369
C49IB08A 4
1 Introduction
1 Introduction
The components are highly compatible with each other, making the system
suitable for a variety of different applications.
Synthos 3000 has been developed by Anton Paar GmbH as a result of its long-
standing experience and in close cooperation with the Karl-Franzens University
in Graz, Austria.
C49IB08A 5
2 Symbols
2 Symbols
Warning:
The "Warning" sign indicates a hazard.
It calls attention to an operating procedure, practice, etc. which could result in
injury or loss of life, if not correctly performed or adhered to.
Do not proceed beyond a "Warning" sign until the indicated conditions are
fully understood and met.
Important:
The "Important" sign indicates a hazard.
It calls attention to an operating procedure, practice, etc. which could result in
damage or destruction of the instrument or parts of it, if not correctly
performed or adhered to.
Do not proceed beyond an "Important" sign until the indicated conditions are
fully understood and met.
Hint:
The "Hint" sign calls attention to additional information, which might be
useful to the operator.
C49IB08A 6
2 Symbols
Warning screens located on the front side and on the right side panel of the
instrument.
C49IB08A 7
3 Safety Instructions
3 Safety Instructions
Synthos 3000 has been tested and is listed for compliance with US, Canadian
and European standards by Intertek Testing Services, which is an accredited
testing lab. The instrument was found to be fully compliant with the Equipment
Safety Law. This indicates the high safety standard of the Synthos 3000 system.
Hint:
This instruction manual is supplied with Synthos 3000 and contains important
information and warnings that must be followed by the user to ensure safe
operation and to maintain the instrument in safe condition. This manual is
intended to supplement, not supersede, the normal safety code of conduct of
each individual country of operation.
The provided information does not cover every safety procedure that should be
practiced. Ultimately, the operator's organization bears responsibility for the
adherence to safety rules. Possible hazards that could harm the user or result in
damage to the instrument are clearly stated at appropriate places throughout this
instruction manual as well as in the reference manual.
C49IB08A 8
3 Safety Instructions
Important:
Read this reference manual carefully before the first use of your instrument!
Make sure that all instructions are fully understood and met to ensure safe
and efficient operation of Synthos 3000.
Warning:
The instrument cavity is equipped with light emitting diodes (LED). Do not look
directly into the light with optical instruments as the LEDs are class 1M laser
products.
• All operators have to be sufficiently trained in the correct and safe use of this
instrument. Operators must be able to judge dangerous situations and take
the right measures to prevent accidents, injuries and damages.
• Do not use Synthos 3000 for any purpose other than described in this
instruction manual or the reference manual, respectively. Anton Paar GmbH
accepts no responsibility for injuries or damages caused by unauthorized
use of the instrument.
• Follow the installation instructions provided with this manual. The installation
has to be carried out by qualified personnel who are able to read, understand
and apply these instructions.
C49IB08A 9
3 Safety Instructions
Warning:
Do not lean against the instrument or the door, respectively.
In case of a sudden pressure build-up inside the cavity (opening of a safety
disk, vessel breakage) the door will slightly open for less than a second and
noxious reaction gases may be released.
Warning:
This instrument features strong magnets to close the door.
People with pacemakers should keep a distance of at least 50 cm to the
instrument.
During operation (microwave energy "ON") an electromagnetic field is
generated by two transformers. Keep a safe distance of approx. 1 m to ensure
maximum safety.
Warning:
- Never open the door during a running program!
- If you abort a reaction procedure, wait until the safety cooling has been
automatically stopped, i.e. the vessel temperatures are lower than the
above mentioned limits.
- After a mains power failure, wait for the vessels to cool down sufficiently
before taking the rotor out. If a failure does not allow you to activate the
safety cooling, it may take up to six hours for the vessels to cool down!
• In case of a safety disk breakage turn off the instrument after the cooling has
been finished, and remove the rotor. Clean the instrument and rotor parts
carefully and inspect them for visual damage. If the door shows deformations
or can no longer be closed tightly, an increased risk of microwave radiation
C49IB08A 10
3 Safety Instructions
• Never clean the rotor body and lid with organic solvents, especially
halogenated hydrocarbons or acetone.
• Never leave a rotor inside the Synthos 3000 oven when transporting the
instrument. This could damage the microwave cavity or the door.
Warning:
Do not open the instrument casing, as this would uncover live parts. Repair
work has to be carried out by authorized field service engineers or by
Anton Paar GmbH only.
• Anton Paar GmbH warrants proper function and safe operation only if no
unauthorized modifications have been made to mechanical or electronic
parts or software.
• When passing this instrument to other users, also hand over all
documentation and manuals.
C49IB08A 11
3 Safety Instructions
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling chemical reagents. Follow any special national safety
regulations.
• When performing a new synthesis for the first time, begin with small amounts
and diluted solutions.
Warning:
Do not use any sensors, accessories or consumption parts other than those
supplied by Anton Paar GmbH, as this might result in serious damage to the
instrument or injuries of users.
• Never operate Synthos 3000 without sufficient load. Never use less than 4
vessels, as surplus microwave energy may interact with rotor parts and
damage them.
• Vent the reaction vessels either inside the microwave cavity of Synthos 3000
with the exhaust unit activated or under a working fume hood. Noxious
gaseous reaction products may be released.
C49IB08A 12
3 Safety Instructions
• If there is a fault (e.g. opening of the safety disk) let the safety cooling run for
at least one hour then switch of the instrument. Refer to the reference
manual. Clean the instrument carefully and inspect it for visible damage. If
the door can no longer be closed tightly, there is a danger of increased
microwave radiation. Do not continue operating Synthos 3000.
C49IB08A 13
3 Safety Instructions
Synthos 3000 has been designed for safe operation even in case of incidents. In
case of safety disk- or vessel breakage, the microwave power is shut down
immediately due to the sudden overpressure release inside the cavity and the
safety door opening. Normally, the safety cooling is activated simultaneously.
Let the safety cooling run for at least one hour, then switch off the instrument and
proceed as described in the reference manual.
Warning:
In case of unexpected operation behavior, NEVER open the door of
Synthos 3000!
Hot and pressurized vessels inside!
The opening of the door will deactivate a vital safety feature, which is
indispensable to ensure user safety.
• Switch on the instrument again. If the safety cooling does not start
automatically, gently pull the door (Fig. 3 - 1). The door will still be held by its
safety hooks, while the door switches will trigger the safety cooling.
Do not use the door opener! Switch off the instrument after one hour of
safety cooling, as the cooling will not stop automatically.
• If the instrument cannot be switched on again, wait for the vessels to cool
down. Depending on the reached temperature, the type of vessels as well as
the number and loading of the vessels, it may take up to six hours for the
vessels to cool down!
Do not open the door earlier than after these six hours!
C49IB08A 14
3 Safety Instructions
Warning:
Before removing a rotor from a switched off instrument, make sure that rotor
and vessels have been cooled down completely! To do so, you can switch on
the instrument and check if the last run was finished completely (Review
menu).
If you are not sure whether rotor and vessels have cooled down, wait at least
six hours before removing the rotor.
• Wait for the vessels to cool down. If the exhaust unit fails during the run,
terminate the run by pressing the <Stop> button twice. Depending on the
reached temperature, the type of vessels as well as the number and loading
of the vessels, it may take up to six hours for the vessels to cool down!
Do not open the door earlier than after these six hours!
• In very rare cases the instrument might reset during a run. After the
automatic restart the safety cooling is usually activated by default. If the
safety cooling does not start automatically, gently pull the door (Fig. 3 - 1).
The door will still be held by its safety hooks, while the door switches will
trigger the safety cooling. Do not use the door opener! Switch off the
instrument after one hour of safety cooling, as the cooling will not stop
automatically.
• If the safety cooling can not be started, wait for the vessels to cool down.
Depending on the reached temperature, the type of vessels as well as
number and loading of the vessels, it may take up to six hours for the vessels
to cool down! Do not open the door earlier than after these six hours!
• Switch off the instrument. Never open the door to extinguish the fire! The
emitted flue gas will cause the fire to die out rapidly. Thus do not activate
the safety cooling!
Warning:
Hot pressure vessels!
Overheating may result in rupture of safety disks or vessels! The design of the
oven cavity ensures user safety even in case of unexpected operation
behavior. This safeguard is only provided if the door is kept shut!
C49IB08A 15
3 Safety Instructions
• After the fire is extinguished wait at least six hours before opening the door.
Waiting six hours ensures that rotor and vessels are cooled down sufficiently
and overpressure is diminished.
Important:
Always check rotor and vessels carefully for damage if a run has been
interrupted unexpectedly.
C49IB08A 16
4 Technical Description of Synthos 3000
The following chapters contain a short description of the working principle, fields
of application and the technical description of the instrument as well as of
available sensor accessories.
The reagents are weighed into the reaction vessels made of quartz or
fluoropolymers. The vessels are closed by a screw cap (Rotor 16 and Rotor 48)
or by a seal holder (Rotor 8S). The vessels are sealed either with a self-sealing
lip-type seal (Rotor 8S and Rotor 16) or a conical seal (Rotor 48). The pressure
vessels are inserted into the rotor, the closed rotor is placed into the Synthos 3000
oven and the door is closed.
Depending on the rotor type and accessories used, the operating pressure and
temperature can be recorded simultaneously and displayed.
The measured data is transmitted wireless from the rotor to the built-in
microprocessor. As soon as one of the predefined limits is reached, the
microwave energy is reduced or completely switched off. If necessary, cooling
intensity is increased.
During the process, the integrated cooling system generates a slight cooling air
flow along reaction vessels and therefore protects rotor components against
excessive thermal stress of rotor and vessels. This has a positive effect on the
service life of the components.
C49IB08A 17
4 Technical Description of Synthos 3000
The main field of application for Synthos 3000 is the synthesis of organic
compounds in the pharmaceutical, biomedical and agrochemical industry in the
50-250 g range using optimized methods.
The numerous accessories for Synthos 3000 also make it ideal for special
applications such as reactions under different atmospheres, extreme pressure
and temperature conditions and for carrying out cleaning and processes steps.
Compared with other microwave synthesis systems, Synthos 3000 provides the
highest safety standards, high quality accessories and excellent reproducibility of
the experiments. This is due to:
• Protein hydrolysis
• Solid phase synthesis including cleavage and filtration in the same vessel
C49IB08A 18
4 Technical Description of Synthos 3000
The microwave cavity is sealed by a safety door, which can be opened easily by
pressing the door opener button.
Above the door is a graphic display and an acid- and water-repellent keypad.
To the right of the keypad is space to place the alphanumeric keyboard.
The main switch for the instrument is located at the front lower right corner of the
instrument.
Door
Mains switch
C49IB08A 19
4 Technical Description of Synthos 3000
Turntable
Exhaust unit
Mains supply
The exhaust unit and the appropriate electrical connections are located on the
rear side of the instrument together with the mains supply socket.
Ports for optional accessories, interfaces and the view angle control knob for the
graphic display are located on the right side of the instrument:
C49IB08A 20
4 Technical Description of Synthos 3000
3 1 4 5 6
2 7
7 PRINTER
Parallel (Centronics) interface to connect printers.
C49IB08A 21
4 Technical Description of Synthos 3000
These properties are the key to fast and efficient heating time of aqueous
solutions and other polar substances. In contrast to the conventional heating in
heating blocks or ovens (where the heat has to be transferred to the medium via
metal parts, air or vessels), microwave radiation heats up the medium directly,
which results in significant saving of time in many fields of application.
C49IB08A 22
4 Technical Description of Synthos 3000
Synthos 3000 supplies unpulsed microwave energy over the entire power
range.
Safety features:
By reducing the limit of the pressure increase rate, spontaneous reactions can be
controlled very quickly by automatically reducing the microwave power in an early
stage of the reaction.
C49IB08A 23
4 Technical Description of Synthos 3000
In addition, the sensor unit includes a second electronic protection circuit that
reacts as soon as the maximum pressure increase rate of 3.0 bar/s is exceeded.
In this case, microwave power is completely shut down (temporarily) and safety
cooling is activated.
Each pressure vessel is protected by a metal safety disk. If the pressure exceeds
the permitted pressure limit (due to spontaneous chemical reactions or thermal
runaways), the safety disk breaks and the pressure is released. This usually
prevents the pressure vessels from rupture.
Safety disk
Screw cap
Fig. 4 - 6 Screw cap (Rotor 16), safety disk and safety disk holder
The rotor surrounds the vessels and has been designed to ensure that the
microwave cavity is well protected in the case of safety disk or vessel rupture.
C49IB08A 24
4 Technical Description of Synthos 3000
Safety hooks
The safety cooling cools down the hot, pressurized vessels and exhausts any
reaction products that have escaped after the rupture of a safety disk or vessel.
C49IB08A 25
4 Technical Description of Synthos 3000
This temperature sensor is directly built into the exhaust unit and protects it
against thermal overload.
This sensor detects the temperature of the exhaust air. If the temperature of the
exhaust air exceeds 130°C both microwave energy and cooling are switched off
and a fire alarm is shown on the display.
Due to the permanent cooling during the process, a temperature gradient is built
up in the vessel which leads to strong convection behavior and constant
exchange between the liquid and the gaseous phase.
Any gases which may escape during the reaction are withdrawn by the exhaust
system.
After completion of reaction the vessels are automatically cooled down below an
appropriate temperature limit.
C49IB08A 26
4 Technical Description of Synthos 3000
The standard version of Synthos 3000 includes optical sensors which are located
in the microwave cavity of the instrument:
A sensor located in the base of the oven cavity sends an infrared light beam onto
the bottom plate of the rotor. The bottom plate reflects the beam to the sensor,
which is electronically analyzed. The following information and/or functions are
processed:
C49IB08A 27
4 Technical Description of Synthos 3000
IR sensor
C49IB08A 28
4 Technical Description of Synthos 3000
Microwave cavity
Sensor Evaluation
Electrical Signal
C49IB08A 29
4 Technical Description of Synthos 3000
The infrared sensor measures the temperature at the base surface of each
reaction vessel. Measured temperature values are assigned and recorded to the
corresponding vessel via the automatic position recognition. The temperature
measurement protects against overheating.
As the measurements are taken through a Fresnel lens on the bottom of the
microwave cavity, make sure that this Fresnel lens is always kept clean.
Reaction vessel
For checking and calibrating the IR sensor see Chapter 6.7.9 or Chapter 6.7.10
and Chapter 6.7.11 or Chapter 6.7.12.
Important:
The IR sensor has to be checked (and recalibrated, if necessary) at least
once a year.
C49IB08A 30
4 Technical Description of Synthos 3000
This description applies for both, p/T sensor accessory for Rotor 16 and p/T
sensor accessory M50 for Rotor 48. These tools are optional, but recommended
to be used for optimum reaction control.
Electronics Receiver
Knurled nut
C49IB08A 31
4 Technical Description of Synthos 3000
Thermometer Capillary
Immersion Tube
Vent Pipe
Seal
Seal Husk
Syringe connector
with O-ring
Venting screw
Immersion tube
Seal holder
Seal
C49IB08A 32
4 Technical Description of Synthos 3000
A hydraulic piston is installed in the sensor head above the seal of the reference
vessel. This piston takes up the reaction pressure via the seal and the seal holder.
The hydraulic system transfers the pressure of the reference vessel to the sensor
module, where it is measured continuously during the reaction and transferred
wireless (via IR) to the control electronics for the following purposes:
• If the pressure increase rate is 0.1 - 0.2 bar/s below the limit, the microwave
power is reduced.
• If the pressure increase rate limit is reached, the microwave power will be
cut off until the pressure rate is below the selected limit value; thereafter it
will gradually increase again.
• If, after cutting off the microwave power, the pressure limit has NOT fallen
below the limit value within 2 seconds, the intensity of the exhaust unit will
be increased to its highest level (level 3).
C49IB08A 33
4 Technical Description of Synthos 3000
• If the pressure exceeds the limit (20 bar for MF50 and HF100 and 40 bar
for HF100) by more than 5 bar as a result of a spontaneous reaction, the
experiment will be terminated. In this case the vessels and seals must be
cleaned and checked very carefully for signs of damage.
Important:
After (accidentally or on purpose) opening of the conical seal of the capillary
tube (knurled nut on the pressure sensor unit), the temperature sensor has to
be re-calibrated.
A hermetically sealed glass bulb is filled with gas. The gas in the bulb expands
with increased temperature during the reaction. This gas expansion causes an
increased pressure which is transferred to the pressure sensor via a capillary. The
pressure increase is converted to a corresponding temperature value. This
temperature value of the reference vessel is transmitted to the electronics module
and from there wireless (via IR) to the microwave oven. The temperature is
displayed graphically on the LCD as a “temperature over time” graph or as a
numeric value.
Using the optional calibration unit the immersion thermometer can be calibrated
to an accuracy of ± 1% within a measuring range of 0 to 280°C. (Refer to chapter
6.7.4, "Temperature Sensor Calibration with the Calibration Accessory" or chapter
6.7.5, "Temperature Sensor Calibration with the Calibration Unit".
Hint:
Calibration has to be done once a year, although it is recommended to
calibrate the temperature sensor at least twice a year.
Each p/T Sensor Accessory is shipped with a unique serial number that is printed
on an adhesive label and stored in the sensor electronics.
Synthos 3000 uses this information to identify the appropriate p/T Sensor
Accessory and to assign the correct calibration factors.
C49IB08A 34
4 Technical Description of Synthos 3000
• If the temperature exceeds the limit (200 °C for MF50, 190 °C for MF100
and 240 °C for HF100) by more than 5 °C as a result of a spontaneous
reaction, the reaction run will be terminated. In this case the vessels and
seals must be cleaned and checked very carefully for signs of damage.
C49IB08A 35
4 Technical Description of Synthos 3000
Hint:
The T-probe unit of Rotor 8S can be used for both pressure vessel types,
XF100 and XQ80 and is used together with immersion tube with seal XF or
immersion tube with seal XQ.
This optional accessory can be easily installed by the customer and it is
recommended to use for optimum reaction control
Seal XQ or XF
Seal husk
Immersion tube
Knurled nut
Fixing screw
Capillary
T-probe
C49IB08A 36
4 Technical Description of Synthos 3000
Important:
After (accidentally or on purpose) opening of the conical seal of the capillary
tube (knurled nut on the pressure sensor unit), the temperature sensor has to
be re-calibrated.
A hermetically sealed glass bulb is filled with gas. The gas in the bulb expands
with increased temperature during the reaction. This gas expansion causes an
increased pressure, which is transferred to the pressure sensor via a capillary.
The pressure increase is converted to a corresponding temperature value. This
temperature value of the reference vessel is transmitted to the electronics module
and from there wireless (via IR) to the microwave oven. The temperature is
displayed graphically on the LCD as a “temperature over time” graph or as a
numeric value.
Using the optional calibration unit the immersion thermometer can be calibrated
to an accuracy of ± 1% within a measuring range of 0 to 300 °C. Refer to chapter
6.7.4, "Temperature Sensor Calibration with the Calibration Accessory" or chapter
6.7.5, "Temperature Sensor Calibration with the Calibration Unit".
Hint:
Calibration has to be done once a year, although it is recommended to
calibrate the temperature sensor at least twice a year.
Each T-probe accessory is shipped with a unique serial number that is printed on
an adhesive label and stored in the sensor electronics.
Synthos 3000 uses this information to identify the appropriate T-probe accessory
and to assign the correct calibration factors.
• If the temperature exceeds the limit (260 °C for XF100 and 300 °C for
XQ80) by more than 5 °C as a result of a spontaneous reaction, the
reaction run will be terminated. In this case the vessels and seals must be
cleaned and checked very carefully for signs of damage.
C49IB08A 37
4 Technical Description of Synthos 3000
The safety module SOLV is a hardware accessory for Synthos 3000, which is
mandatory for safe operation using organic solvents.
It consists of mainly two membrane switches, which monitor the exhaust airflow
through the cavity in order to prevent the formation of dangerous concentrations
of solvent vapors inside the system. In case that the airflow is interrupted (e.g.
failure of the exhaust fan, blocking of the exhaust system), microwave power is
switched off immediately and the current run is interrupted.
The safety module SOLV is installed underneath the cavity, therefore later
upgrades in the field have to be done by authorized service engineers only.
C49IB08A 38
4 Technical Description of Synthos 3000
The optional solvent sensor monitors the concentration of organic solvents in the
exhaust air channel of Synthos 3000. The sensor element is a metal oxide
semiconductor detector that changes resistance in correlation to the gas
concentration.
4.8.1 Installation
Important:
Take care that Synthos 3000 is switched off before installing the Solvent
Sensor Accessory!
The Solvent Sensor Accessory measures the solvent concentration inside the
cavity, but it is not required for safe use of the instrument like the obligatory
Safety Module SOLV (SMS).
1. Remove the safety bracket and the cover from the exhaust channel at the
rear of Synthos 3000 by loosening the two nuts.
2. Plug the sensor into the exhaust channel. The cable has to point to the left.
C49IB08A 39
4 Technical Description of Synthos 3000
3. Reassemble the safety bracket on the exhaust channel and fasten it with the
two nuts.
4. Attach the connector to the “AUX” port of the instrument and fix it with the
two screws for proper connection.
Sensor
cable
C49IB08A 40
4 Technical Description of Synthos 3000
The function verification is only used for testing the sensor. For the function
verification you will need acetone and a 1 or 1.5 L bottle (any material), which can
be closed with a screw cap.
2. Switch on the instrument and wait until the Solvent Sensor Accessory has
reached a minimum temperature of 36 °C.
8. Press <Fan 1 F4 > for the test mode. The exhaust unit is set to level 1.
9. Fill acetone into the bottle (0.5 mL acetone per 1L bottle volume). Close the
bottle with the screw cap and shake it for at least 30 seconds.
10. Hold the bottle upside down onto the exhaust holes in the cavity and open
the screw cap.
11. The signal on the display should reach a minimum value of 1000-1500. If
not, contact your local customer service.
C49IB08A 41
4 Technical Description of Synthos 3000
Important:
Calibration of the Solvent Sensor Accessory has to be done regularly.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents and observe the relevant national regulations.
3. Connect the Solvent Sensor Accessory to the AUX port of Synthos 3000.
5. Use the arrow keys <Ï F4> and <Ð F5> to select the solvent sensor
settings menu in the service menu. Confirm by pressing <Enter F1>.
C49IB08A 42
4 Technical Description of Synthos 3000
8. Use an adhesive tape for sealing. Take care for proper sealing.
9. Depending on your application inject one of the following solvents into the
measuring cylinder (through the adhesive tape):
C49IB08A 43
4 Technical Description of Synthos 3000
n-Hexan or
Methanol/acetone
10. Shake the measuring cylinder and read out the highest signal value on the
display of the instrument.
Signal value
11. Remove the sensor from the measuring cylinder and vent the flask by
purging with an air stream.
12. Repeat these steps 2 times and enter the average value into the instrument
as your new "Calib. Constant".
13. A variation of the signal of +/- 200 between the different trials is within the
specified operation limits. If the limits excess the variation of +/- 200, contact
your local customer service.
C49IB08A 44
4 Technical Description of Synthos 3000
The magnetic stirrer device is a useful tool for mixing the contents of the closed
reaction vessels inside Synthos 3000. Depending on the application, stirring
helps to obtain better recoveries or to attenuate spontaneous reactions by
avoiding local inhomogeneity inside the vessels.
The magnetic stirrer device is installed underneath the cavity and consists of
three plates with strong permanent magnets, which drive the magnetic stir bars
inside the reaction vessels.
Different rotation speeds (200, 400 and 600 rpm) are selected and controlled via
the software.
C49IB08A 45
5 Installation
5 Installation
The following chapter explains the requirements and the steps to be performed
for a successful installation of Synthos 3000.
Important:
A reliable exhaust system with sufficient capacity (min. 300 m³/h) must be
provided when operating Synthos 3000.
Hazardous and noxious gases may build up if an appropriate and working
exhaust system is not provided at the place of installation. Synthos 3000 must
not be set in operation in this case.
• Provide sufficient clearance (10 cm) between the instruments side panel and
rear panel, respectively, and the laboratory walls for satisfactory circulation
of the cooling air.
• The utilized fume hood must be operating and switched on during the
experiments to withdraw vapors and gases, especially after venting actions.
C49IB08A 46
5 Installation
Warning:
Synthos 3000 must be connected ONLY to an approved standard socket with
protective conductor!
Before plugging in the cable, always make sure that the local mains supply
and frequency correspond to the values on the type plate of the instrument.
If your laboratory is equipped with another plug system, you have to provide a
suitable and approved connection cable.
C49IB08A 47
5 Installation
For detailed information follow the instructions given on the video manual on CD.
1. Unpack Synthos 3000 and inspect it for possible signs of damage in transit.
Important:
Keep the packing material for transporting the instrument at a later time
(change of location, repair, upgrade etc).
2. Make sure that the door can be closed properly and the sealing surfaces of
the door are not damaged.
Warning:
Never operate Synthos 3000 if the door is damaged!
In this case the function of the safety features can no longer be guaranteed
and harmful microwave radiation may escape.
4. Check if the voltage and frequency are in accordance with the ratings on the
instrument label.
Important:
When lifting Synthos 3000, only hold it by the outer edges of the base.
Never lift Synthos 3000 by the instrument door.
Never transport Synthos 3000 with a rotor inserted.
5. Check the proper installation of the foam seal between tube and exhaust
unit.
C49IB08A 48
5 Installation
7. Lead the open end of the exhaust hose into an exhaust system or fume
hood. Make sure that there is a bypass between the hose of Synthos 3000
and the exhaust system.
Warning:
Never run the system without the exhaust hose.
8. Check if the plug of the exhaust unit has been inserted properly into the
appropriate socket.
10. Insert the mains cable plug into the wall socket.
Synthos 3000 has to be connected to an approved standard wall socket with
a ground conductor.
C49IB08A 49
5 Installation
12. Adjust the view angle of the LCD after the software has started automatically.
The knob is marked "VIEW ANGLE" and located on the upper right-hand
side of Synthos 3000.
By turning the knob you can adjust the LCD for optimum readability.
C49IB08A 50
5 Installation
5.3.1 Printer
Reaction data can be printed for documentation. All printer models with an
installed ASCII character set are suitable for printing text documents (without
graphs).
If you wish to print text and graphs, e.g. pressure / temperature curves, you may
connect any compatible printer. The printer port is located on the right side of the
instrument and marked "PRINTER".
5.3.2 Keyboard
Reaction data can be transferred to an external PC via a serial interface. The port
"COM1" is provided on the right side of the instrument. This RS232 interface is an
SELV (safety extra-low voltage) circuit that can only be connected to other SELV
or SELV-E circuits. Only use shielded cables with a length of less than 3 m for this
connection.
A standard terminal program can be used to store the run data as text file on an
optionally computer.
The free Excel Add-in "AP-Softprint for Synthos 3000" (available at AP-Website)
can be used to import the run data directly into a MS-Excel sheet, which simplifies
both, data collection and diagram generation.
C49IB08A 51
5 Installation
Prior to the first reaction with Synthos 3000, proper instrument settings have to be
checked and the appropriate rotor types and accessories have to be activated in
the setup menu.
Hint:
All rotors, sensors and accessories for Synthos 3000 which are added or put
into operation must be activated and configured before first use.
You can also deactivate rotors, sensors or accessories which are no longer
required.
Starting from the main menu, press <Setup F3> to reach the setup menu.
The setup menu is used to change instrument settings. The arrow keys <Ï F4>
and <Ð F5> are used to select the desired options from the setup menu, which
are subsequently confirmed with <Enter F1>.
C49IB08A 52
5 Installation
Hint:
Microwave power must be calibrated before the first reaction and can be
repeated at periodical intervals.
Calibrate the microwave power at 800 W maximum as described below.
Required material:
The calibration is based on the knowledge of the specific heat capacity of water.
By heating up a certain amount of water with microwave energy, the apparent
power can be calculated from the water volume, the temperature of the water
before and after heating.
The power calibration consists of the following steps which are carried out
automatically by the software:
3. Press <Ð F5> to select the power calibration option. Confirm by pressing
<Enter F1>.
C49IB08A 53
5 Installation
4. Fill 1000 g of tap water (approx. 20 °C) into a beaker (diameter approx.
180 mm).
5. Check if the amount of water in the beaker corresponds to the value shown
on the LCD. Possible deviations can be corrected by pressing <+ F2> or <-
F3>. Press <Next F4>.
6. Select the target microwave power for the calibration. The recommended
power is 800 W. Changes can be entered via <+ F2> or <- F3>. Press <Next
F4>.
C49IB08A 54
5 Installation
7. Measure the temperature of the water. Use the <+ F2> or <- F3> key to enter
the measured temperature value, then press <Next F4>.
8. Open the door of Synthos 3000 and place the beaker on the turntable. Close
the door.
C49IB08A 55
5 Installation
9. Press the green <START> button and wait until the heating time (80 seconds
for 800 W) has elapsed. You can terminate the procedure at any time by
pressing the red <STOP> key.
10. After the time has elapsed and the red LED on the front panel has switched
off, the next menu step is displayed. Open the instrument door and remove
the beaker. Measure the water temperature with an immersion thermometer
while stirring (homogenizing) the water thoroughly. Enter the final water
temperature via <+ F2> or <- F3> and press the <Next F4> key.
C49IB08A 56
5 Installation
11. Press <Save F1> to store the new power factor. Press <Repeat F4> to
repeat the power calibration once again. Press <Print F5> if you want to
print out the calculation for documentation purposes.
If the calculated power ratio is out of range (v1.52 or lower: 0.8 - 1.2; v2.0 or
higher: 0.7 - 1.3) saving is not possible.
In this case enter either 0.8 (if the power ratio was below 0.8) or 1.2 (if the
power ratio was above 1.2), press <Save F1> and repeat the calibration
procedure with the new settings (<Repeat F4>).
Press <ESC> to exit without any changes.
C49IB08A 57
6 Software
6 Software
Hint:
Periodically check the Anton Paar website (www.anton-paar.com) for software
updates or contact our local distributor.
Install the updated software version according to chapter 6.7.1, "Software
Update".
C49IB08A 58
6 Software
• ESC key
The <ESC> key is used to terminate a process and/or move back one step in
the menu hierarchy.
• Start button
The green <Start> button is used to start a selected method.
• Stop button
The red <Stop> button is used to interrupt or stop a microwave program.
Softkeys Keypad with signal lamp
C49IB08A 59
6 Software
START
C49IB08A 60
6 Software
The main menu appears automatically after switching on the instrument. The
installed software version is shown in the lower left corner of the graphic display.
The following sections of the program or submenus are accessible via the
function keys:
• F1 Library
Allows you to select and modify methods or to create new methods. New or
modified methods can be saved for future use. Comments can be added to
the methods.
• F2 Review
For viewing and/or printing the stored process data as tables and diagrams
(power vs. time, temperature vs. time and pressure vs. time). The available
diagrams depend on the sensors used. The stored data can also be sent to
an external PC.
• F3 Setup
This key accesses a submenu containing different instrument settings.
• F4 Service
This key accesses a submenu containing maintenance and service
programs.
• F5 Info
Shows telephone number, fax number and contact e-mail address of
Anton Paar GmbH as well as the serial number of the instrument and the
usage information (v2.0 or higher).
C49IB08A 61
6 Software
Starting from the main menu, press <Library F1> to reach the library menu.
The library menu is used to load stored methods and to create or edit methods
using the method editor.
To jump directly between these areas, use the arrow keys <Í F2> or <ÎF3>.
For various sample materials there may exist several methods using different
rotor types. Make sure to select the appropriate method for the rotor used. After
selecting a method, the sample menu will be opened.
C49IB08A 62
6 Software
• Start
Starts the selected method displayed in the sample menu.
• Edit
Edits the selected method displayed in the sample menu.
• Notes
This option is used to enter a comment or additional information. A flashing
notes field indicates information stored for this method.
C49IB08A 63
6 Software
6.4.1.1 Start
Biginelli
By pressing <Start F1> a startup routine is activated, which checks presence and
type of rotor in the microwave cavity and verifies whether the rotor has been
closed with the protection lid or not.
If optional sensors have been installed (e.g. a p/T sensor) and properly configured
in the instrument setup, the function of these sensors will also be checked.
If no errors have been detected, the method can be started by pressing the green
<Start> button.
6.4.1.2 Edit
Depending on the installed rotors and accessories (sensor etc.), the actual
sample menu might look different.
C49IB08A 64
6 Software
• Press <+ F2> and <– F3> to create or change information or parameters for
a method. You can use them to enter characters, select options and
increase / decrease numerical values.
• The arrow keys <Í F4> and <Î F5> are used to jump between the entry
fields.
• Press <Exit F1> to exit the editing dialogue and select one of the following
options from the pop-up window:
- Save:
Press <Save F1> to save your changes by overwriting the (existing)
version. This function can be deactivated by locking the method in the
library maintenance tool (refer to chapter 6.7.15, "Library Maintenance
Tool").
- New:
Press <New F2> to store as a new method, the existing one remains
unchanged.
- Discard:
Press <Discard F3> to cancel your changes.
- Delete:
Press <Delete F4> if you wish to delete the method. This function can be
deactivated by locking the method in the library maintenance tool (refer to
chapter 6.7.15, "Library Maintenance Tool").
• Sample
This field is used to enter a sample name consisting of max. 23 characters
but only 18 will be shown in the library.
• Rotor
This field is used to select the rotor type you are going to work with.
Deactivated rotors are not displayed and therefore not selectable.
Hint:
After selecting the appropriate rotor type, the pressure and temperature limits
will be changed automatically.
Important:
The HEAT rotor type enables an operating mode that is exclusively for
testing purposes (e.g. for service).
Misuse of these instructions may result in damage to the Synthos 3000.
C49IB08A 65
6 Software
• Vessels
This field is used to enter the number of vessels applied in the rotor.
Please follow the recommended loading patterns according to the respective
rotor type. Never use less than 4 vessels!
• Weight
This field is used to enter the weight of your sample material.
• P-Program / T-Program
Depending on the type of rotors and sensors, this field can be used to select
a power or temperature program:
- P-Program:
When using a P-program (power program), the instrument follows a
defined power profile and only deviates from this profile when limits such
as pressure, pressure increase rate or temperature are exceeded.
- T-Program:
When using a T-program (temperature program), the instrument follows a
defined temperature profile and only deviates from this profile when limits
such as pressure, pressure increase rate or temperature are exceeded.
Hint:
It is highly recommended always to use an immersing temperature sensor,
even when a power program is performed.
• Ph
This field indicates the "Phase“ or step of the power / temperature profile.
• Power / Temp
Power / temperature setpoint at the end of the ramp time.
• Ramp
Time period in which the previously defined power / temperature setpoint has
to be reached. Possible range 00:00 to 99:59 [minutes: seconds].
• Hold
Time period wherein the defined power / temperature setpoint has to be
held. Possible range: 00:00 to 99:59 [minutes: seconds].
Fan
This field is used for setting the intensity of the exhaust unit:
- Level 0: minimum intensity (special applications)
- Level 1: low intensity (used during heating)
- Level 2: medium intensity (for cooling)
- Level 3: highest intensity (for cooling)
C49IB08A 66
6 Software
• p-rate
The limit of the pressure increase rate is entered in this field. The possible
range is between 0.1 - 3.0 bar/s.
If the pressure rate exceeds the predefined limit, the microwave generation
will be interrupted for a short period of time.
• IR
This field is used to enter the IR temperature limit.
• p
This field is used to enter the pressure limit for power controlled runs.
• P
This field is used to enter the power limit for temperature controlled runs.
• Drive
This field is used to chose the rotor drive mode.
Rot.: Constant rotation
Osc: Oscillation
• Stirrer
This field is used for setting the intensity of the stirrer unit in a range of 0 - 3.
Off: The magnetic stirrer is switched off.
1, 2 or 3: The magnetic stirrer rotates at varying speeds.
6.4.1.3 Notes
You can enter additional information or notes related to the method in the Notes
Editor. The following functions are available:
• Start F1
This function is used to start the method directly without going back to the
primary menu.
• Edit F2
This function is used to change existing information or to enter new ones
with the alphanumeric keyboard.
Leave the Edit mode by pressing the <Exit F1>.
• Sample F3
The <Sample F3> key will bring you back to the Sample menu.
C49IB08A 67
6 Software
The sample documentation editor can be used to enter additional information with
the alphanumeric keyboard.
• Start F1
This function is used to start the method directly without going back to the
primary menu.
• Sample F4
The <Sample F4> key brings you back to the sample menu.
C49IB08A 68
6 Software
6.5 Review
Starting from the main menu, press <Review F2> to reach the review menu. Run
data such as temperature and pressure values are stored automatically during
each process. Depending on the number of stored methods, the data of the last
4-8 hours of operation will be available under the review menu option.
The list is sorted by date / time and offers the following options:
• Enter F1
Press <Enter F1> to view the data of the run on the display. Depending on
the type of sensors used, you can select different displays.
• Send
Press <Send F2> to send the data of the selected run to an external PC
connected to Synthos 3000 via the serial interface.
• Print
Press <Print F3> to print the selected data on a connected printer.
Use the arrow keys <Ï F4>, <Ð F5> to select a stored run.
C49IB08A 69
6 Software
This screen provides an overview (text based) of all maximum values reached
during a run.
QQ
T$5DgF#ODU
ODU
%ODU&f
• Power
The max. microwave power [W] which was applied during the run. The
applied power may deviate from the maximum power set point defined in the
method e.g. if the temperature or pressure limit was reached or the permitted
pressure rate was exceeded.
• Temp
Maximum inner temperature which was reached in the reference vessel.
• Pres
Maximum pressure which was reached during the reaction.
• IR
Maximum IR temperatures of reaction vessels and empty positions,
respectively.
• IR Lim
Temperature limit defined in the method for the IR temperature.
• T Lim
Temperature limit defined in the method for the internal temperature in the
reference vessel.
C49IB08A 70
6 Software
• p Lim
Pressure limit defined in the method.
• p Rate
Limit of pressure increase rate defined in the method.
Hint:
Pressure and temperature values will only be available if the reaction was
carried out with activated and installed sensors.
The review P-graph displays the power and pressure vs. the reaction time.
Programmed power
profile
Pressure curve
Actually applied microwave Time and reason of the first power reduction
power
• Pressure
This curve indicates the pressure course during the reaction.
• P-reduction time
For explanation of the herein stated "control parameters" refer to chapter
11.2.5, "Power Reduction by the Instrument".
C49IB08A 71
6 Software
The review T-graph displays the internal temperature and pressure vs. the
reaction time.
• (Internal) temperature
Shows the temperature course in the reference vessel.
• Pressure
Shows the pressure course during the run.
The review IR-graph displays the individual IR temperatures and the reaction
pressure vs. the reaction time.
IR Temperatures
IR Temperature of the empty positions Pressure
C49IB08A 72
6 Software
• IR temperature
Shows the IR temperature curves during a reaction. These data mainly
provide information about uniformity of heating.
Hint:
IR values may differ if varying mixtures are applied in the individual vessels.
• Pressure
Shows the pressure course during the run.
This menu item allows the entry of additional sample information. Changes can
be entered as necessary. The printout includes date and time of changes, if the
changes were done after the run has been completed.
C49IB08A 73
6 Software
Starting from the main menu, press <Setup F3> to reach the setup menu.
The setup menu is used to change instrument settings. Use the arrow keys
<Ï F4> and <Ð F5> to select the desired options from the setup menu. Confirm
by pressing <Enter F1>.
• Configuration
Use this option to set the language (v2.0 or higher), date, time and "device
name" and if necessary, make the appropriate layout settings for an external
keyboard.
• Printer Setup
This option is used to set the printer type, printer interface and character
settings.
• Safety Cooling
Set parameters for safety cooling like cooling intensity, cooling time or
temperature limits.
• Rotors
Activate / deactivate the appropriate rotor types.
• Sensors
Activate / deactivate the appropriate sensor types.
• Accessories
Activate / deactivate accessories.
C49IB08A 74
6 Software
6.6.1 Configuration
Use the arrow keys <Ï F4> and <Ð F5> to select the configuration menu.
Confirm by pressing <Enter F1>.
Using the keys <+ F2> and <– F3> you may create or change any data or
parameters in the entry or option fields, i.e. you may select letters or change
between different options.
The arrow keys <Í F4> and <Î F5> are used to move to the next or return to
the previous entry field. Press <Exit F1> to store your entries / modifications and
revert to the setup menu.
Press <ESC> to exit without saving.
• Language
Multilingual user interface (v2.0 or higher).
• Ext. Keyboard
Select between the German or English keyboard layout.
• Device Name
Enter an instrument name consisting of max. 12 characters. This name will
be stored in the run data and will be printed out.
• Date
Set the current date ("YYYY-MM-DD").
• Time
Set the internal clock.
C49IB08A 75
6 Software
Use the arrow keys <Ï F4> and <Ð F5> to select the printer setup menu. Confirm
by pressing <Enter F1>.
The setup printer menu is used to select the connected printer. It consists of four
option fields.
• Printer Type
The system distinguishes between the following printer types:
- Plain Text
This setting is for printer models with an internal "character set", i.e. printers
that are capable of printing ASCII sequences. This setting can only be used
to print text - you will not be able to print any graphs using this setting.
- Epson 9 Pin
This setting is for Epson 9-pin printers or Epson-compatible printers. Used
printer protocol: ESC/P Standard; printing text and graphics.
- Epson 24 Pin
This setting is for Epson 24-pin printers or Epson-compatible printers. Used
printer protocol: ESC/P Standard; printing text and graphics.
- PostScript
This requires a PostScript printer (like special laser printers): e.g. HP
Laserjet 4 M-plus or a newer model, such as HP 4050n. Used printer
protocol: PostScript Language Level 3; printing text and graphics.
- HP DeskJet
This setting is for HP DeskJet printers or HP DeskJet-compatible printers.
Used printer protocol: PCL 5; printing text and graphics.
- HP LaserJet
This setting is for HP LaserJet printers or HP LaserJet-compatible printers.
Used printer protocol: PCL 5; printing text and graphics.
• Printer Port
Select Centronics if a printer is connected to the parallel port. Select Serial
for sending data via the serial port (RS232) to a connected PC.
C49IB08A 76
6 Software
• Decimal Separator
Use this option to select between a period (1.123) or comma (4,567) as
decimal separator.
• Character Set
Use this option to select between the ASCII MS-DOS CP437 and ISO 8859-
1 Latin 1 character set, if some characters are not printed correctly.
The <+ F2> and <– F3> keys are used to select the different options. The arrow
keys <Í F4> and <Î F5> are used to move between the entry fields. Press
<Exit F1> to save your changes and revert to the Setup menu. Press <ESC> to
discard your entries.
C49IB08A 77
6 Software
The serial port setup menu is used to set the parameter for data transfer.
Use the arrow keys <Ï F4> and <Ð F5> to select the serial port setup menu.
Confirm by pressing <Enter F1>.
The following parameters are available in the serial port setup menu:
• Text Compression
Text compression reduces the amount of data for faster transfer rates. Data
can be imported into a MS Excel irrespective of the used compression mode.
- None:
No compression, slowest data transfer. Best format for processing with a
text editor.
- Medium:
Character spaces are removed, transfer is accelerated.
- High:
Data duplications are removed, transfer needs only a quarter of the time
without compression.
• Transfer Delay
Depending on the CPU-type and operation system, data transfer without
delay might be too fast for the computer. Increase the transfer delay if you
experience problems with the data transfer.
- For data transfer to external PCs via a terminal program a delay is usually
not necessary.
- For data transfer with AP-Softprint we recommend a transfer delay of at
least 30 ms. (Possible transfer delays: None, 10, 20, 30, 50, 100, 250 and
500 ms;)
C49IB08A 78
6 Software
The safety cooling menu can be used to adjust the intensity of cooling fan level,
cooling time, temperature limit and IR limit.
Use the arrow keys <Ï F4> and <Ð F5> to select the safety cooling menu.
Confirm by pressing <Enter F1>.
Hint:
After an overpressure release inside the cavity (e.g. safety disk or vessel
breakage), the safety cooling will always run on fan level 3.
• Temperature Limits
Depending on the used sensors, you can choose the temperature values,
which have to be reached during the cooling process. Default settings for
synthesis applications are 60 °C for internal temperature and 50 °C for IR
temperature, both limits can be reduced down to 40 °C.
Please note that reducing these values will result in increased cooling times.
C49IB08A 79
6 Software
6.6.5 Rotors
Move the black cursor to the desired rotor type using the arrow keys <Ï F4> and
<Ð F5>.
Activate or deactivate the selected rotor type by pressing <+ F2> or <– F3>.
Press <Exit F1> to confirm the changes.
Press <ESC> to exit without saving the changes.
The current status is displayed by a symbol, which is shown to the left of the
selected rotor type:
C49IB08A 80
6 Software
6.6.6 Sensors
Move the black cursor to the desired sensor using the arrow keys <Ï F4> and
<Ð F5>. Activate or deactivate the selected sensor by pressing <+ F2> or <– F3>.
Press <Exit F1> to confirm the changes.
Press <ESC> to exit without saving the changes.
The current status is displayed by a symbol, which is shown to the left of the
selected sensor:
C49IB08A 81
6 Software
6.6.7 Accessories
Move the black cursor to the desired accessory using the arrow keys <Ï F4> and
<Ð F5>. Activate or deactivate the selected accessory by pressing <+ F2> or
<– F3>.
Press <Exit F1> to confirm the changes.
Press <ESC> to exit without saving the changes.
The current status is displayed by a symbol, which is shown to the left of the
selected accessory:
C49IB08A 82
6 Software
6.7 Service
• Software Update
For updating the software (requires an external PC)
• Power Calibration
For adjusting the microwave power output
• Power Check
For checking the microwave power output
• IR Sensor Calibration
For adjusting the IR sensor
• IR Sensor Check
For checking the IR sensor
C49IB08A 83
6 Software
Hint:
We recommend to check regularly for new software versions (http://
www.anton-paar.com) and to install if appropriate.
- External PC with a free serial interface, Windows 98 or higher with NET 2.0
(v2.00 or higher)
(Alternatively: External PC with USB port, serial-to-USB converter cable)
- Serial link (0-Modem) cable, this cable is also available from Anton Paar
GmbH or its sales representatives
- Flash update program for Synthos 3000 (available from Anton Paar GmbH
free of charge)
Synthos 3000 software installation file(s) (available from Anton Paar GmbH
free of charge)
Press <Enter F1> to enter the software update menu. Connect the host
computer´s serial port and run the program Flash Update to activate the
bootloader. Follow the instructions given therein to proceed with the update.
The calibration is based on the knowledge of the specific heat capacity of water.
By heating up a certain amount of water with microwave energy, the apparent
power can be calculated from the water volume, the temperature of the water
before and after heating.
The power calibration consists of the following steps which are carried out
automatically by the software:
C49IB08A 84
6 Software
Use the arrow keys <Ï F4> and <Ð F5> to select the power calibration menu.
Confirm by pressing <Enter F1>.
The power check function is used to verify the microwave power calibration. The
procedure is the same as described for the power calibration, but at a target
power of 1400 W.
Hint:
If the apparent microwave power deviates from the nominal value by more than
10% a microwave power calibration should be performed.
C49IB08A 85
6 Software
Hint:
Description only valid for calibration accessory up to version 4.
If you are using a calibration unit, please follow the instructions in chapter
6.7.5, "Temperature Sensor Calibration with the Calibration Unit".
Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor
calibration menu. Confirm by pressing <Enter F1>.
Version 4
1. Carefully insert the p/T sensor into the calibration accessory and close the
screw cap.
C49IB08A 86
6 Software
2. Place the calibration accessory with the temperature sensor into position 1 of
your rotor and place the rotor without lid into the microwave cavity.
Important:
Do not close the instrument door as this will damage the cable of the
calibration accessory.
C49IB08A 87
6 Software
3. Connect the cable of the calibration accessory to the CAL port on the right-
hand side of the instrument. Tighten the screws of the plug to ensure good
contact.
4. Press the <Calib F1> key to start the calibration. The serial number of the
temperature sensor is read out first. The calibration accessory starts to heat
and the temperature values are recorded automatically.
C49IB08A 88
6 Software
5. At the end of the heating phase (approx. 15 - 20 minutes) the result of the
calibration is displayed. Press <Save F1> to store the calculated calibration
factors. You might press <Print F5> to print out the factors. By pressing
<Repeat F4> you can repeat the calibration. Press <ESC> to exit without
any changes.
Warning:
The surface of the calibration accessory is hot. Do not touch the calibration
accessory for at least 20 min after finishing the calibration.
1. Put the calibration accessory into the protective casing. Make sure you lead
the cable of the calibration accessory to the left side of the protective casing
strap.
C49IB08A 89
6 Software
2. The immersion tube with seal (XF/XQ) can now be put into the calibration
accessory. The venting screw has to point to the same direction like the
cable of the calibration accessory (left of the protective casing strap).
Fig. 6 - 31 Putting the immersion tube with seal into the calibration accessory
3. Put the protective cap onto the protective casing so that the cable and the
venting screw are led through the opening of the cap.
C49IB08A 90
6 Software
5. Use the glass rod to make sure that the vessel in position 1 is positioned
correctly. If there is any resistance when inserting the rod, adjust the vessel
and try again. Repeat the procedure until you feel no resistance when
inserting the rod.
Important:
Never insert the T-probe unless the vessel is correctly positioned and the rotor
is correctly closed, otherwise the glass bulb and the capillary may break!
6. Insert the T-probe very carefully through the hole on the upper plate of the
rotor into the vessel at position 1. The torus-shaped part of the T-probe is
placed over the sensor module. Adjust the thumb wheels so that the T-probe
is attached completely on the rotor upper plate.
C49IB08A 91
6 Software
7. Place the rotor without protection lid into the microwave cavity.
Important:
Do not close the door of the instrument as this might damage the cable of the
calibration accessory.
8. Connect the cable of the calibration accessory to the CAL port on the right-
hand side of the instrument. Tighten the screws of the cable plug to ensure
good contact.
9. Press the <Calib F1> key to start the calibration. The serial number of the
T-probe is read out first. The calibration accessory starts to heat and the
temperature values are recorded automatically.
C49IB08A 92
6 Software
10. At the end of the heating phase (approx. 15 - 20 minutes) the result of the
calibration is displayed on the display. Press <Save F1> to store the
calculated calibration factors. You might press <Print F5> to print out the
factors. By pressing <Repeat F4> you can repeat the calibration. Press
<ESC> to exit without any changes.
Warning:
The surface of the calibration accessory is hot. Do not touch the calibration
accessory for at least 20 min after finishing the calibration.
11. To avoid entering wrong calibration constants, remove the adhesive label
from the sensor arm, if new calibration factors have been stored in the
instrument.
C49IB08A 93
6 Software
The calibration unit consists of a vessel jacket with an integrated heating block, a
Pt-100 temperature sensor for correct measurement of the temperature of the
heating block and a metal support.
During the calibration the heating block is heated up to two measuring points
(40 °C and 180 °C).
The difference between the measured temperature values of the heating block
and the p/T-sensor is used to calculate two calibration factors.
Important:
After carrying out any maintenance or repair work on the temperature sensor,
it has to be recalibrated.
C49IB08A 94
6 Software
Metal support
Lock I
Front side
Cable
Calibration insert
Fig. 6 - 41 Metal support for the calibration insert and calibration insert
C49IB08A 95
6 Software
Metal support
Fig. 6 - 42 All parts, which are needed for calibration of the p/T sensor for
Rotor 48 or Rotor 16
Back side
Lock I
Front side
1 2
3 4
C49IB08A 96
6 Software
3. Attach the calibration insert to the ring on the front side of the metal support
(see Fig. 6 - 44 /1). Be aware that the two screws have to be at the side of
the lock II (see Fig. 6 - 44 /2). Lift the lock II, hold it and push the calibration
insert back to the fixed pendant (see Fig. 6 - 44 /3). Release the lock II, the
two screws have to fit into the two holes and then the calibration insert itself
is fixed (see Fig. 6 - 44 /4).
Screws
Lock II
with
two
holes
1 2
3 4
4. Put the corresponding p/T sensor into the calibration insert. Be aware that
the arm of the p/T sensor rests between the two screws on the upper side of
the metal support.
C49IB08A 97
6 Software
5. Place the assembled calibration unit carefully (hold the sensor arm with one
hand) into the microwave cavity to avoid scratches to the coating. Take care
that the metal support fits as accurate as possible on the turntable. Then
connect the cable of the calibration unit to the CAL-port on the right-hand
side of the instrument. Tighten the screws of the cable plug to ensure good
contact.
Important:
Do not close the instrument door as this might damage the cable of the
calibration insert.
C49IB08A 98
6 Software
6. Select the service menu in the main menu by pressing <Service F4>.
7. Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor
calibration menu. Confirm by pressing <Enter F1>.
8. Press the <Calib F1> key to start the calibration. The serial number of the
temperature sensor is read out first. The calibration insert starts to heat and
the temperature values are recorded automatically.
C49IB08A 99
6 Software
9. At the end of the heating phase (approx. 15 - 20 minutes) the result of the
calibration is shown on the display. Press <Save F1> to store the calculated
calibration factors. You might press <Print F5> to print out the factors. By
pressing <Repeat F4> you can repeat the calibration. Press <ESC> to exit
without any changes.
Warning:
Hot surface!
The surface of the calibration insert is hot.
Do not touch the calibration insert for at least 20 min after finishing the
calibration.
10. To avoid entering wrong calibration constants, remove the adhesive label
from the sensor arm, if new calibration factors have been stored in the
instrument.
11. After complete cooling, clean the calibration insert with a damp cloth, if
necessary. Let the calibration insert dry completely before next use.
C49IB08A 100
6 Software
Rotor 8
(without
protection lid)
Centering
jacket
Protective
cap
Spacer
Protective cap
Spacer
Calibration insert
Centering jacket
C49IB08A 101
6 Software
2. Put the calibration insert into the centering jacket. Make sure to lead the
cable of the calibration insert through the slot of the casing.
3. Add the spacer directly on the top of the calibration insert with the two pits on
the bottom side. The two screws of the calibration unit have to fit into the two
pits.
4. The immersion tube with seal (XF / XQ) can now be put into the calibration
insert. The venting screw has to point to the same direction as the cable of
the calibration insert.
C49IB08A 102
6 Software
Fig. 6 - 54 Putting the immersion tube with seal into the calibration insert
C49IB08A 103
6 Software
7. Use the glass rod to make sure that the vessel (in position 1) is positioned
correctly. If there is any resistance when inserting the rod, adjust the vessel
and try again. Repeat the procedure until you feel no resistance when
inserting the rod
8. Afterwards close the rotor thumb wheels completely (no movement of the
vessels).
C49IB08A 104
6 Software
Fig. 6 - 58 Closing the rotor thumb wheels with inserted glass rod
Important:
Never insert the T-probe unless the vessel is correctly positioned and the rotor
is correctly closed. Otherwise the glass bulb and the capillary may break!
9. Remove the glass rod and insert the T-probe very carefully through the hole
on the upper plate of the rotor into the vessel at position 1. The torus-shaped
part of the T-probe is placed over the sensor module. Adjust the thumb
wheels so that the T-probe is attached completely on the rotor upper plate.
10. Place the rotor without protection lid into the microwave cavity. Then connect
the cable of the calibration insert to the CAL-port on the right-hand side of
C49IB08A 105
6 Software
the instrument. Tighten the screws of the cable plug to ensure good contact.
Important:
Do not close the instrument door as this might damage the cable of the
calibration insert.
Fig. 6 - 60 Placing the rotor into the cavity and connecting the cable of the unit
11. Select the service menu in the main menu by pressing <Service F4>.
12. Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor
calibration menu. Confirm by pressing <Enter F1>.
C49IB08A 106
6 Software
13. Press the <Calib F1> key to start the calibration. The serial number of the T-
probe is read out first. The calibration insert starts to heat and the
temperature values are recorded automatically.
14. At the end of the heating phase (approx. 15 - 20 minutes) the result of the
calibration is shown on the display. Press <Save F1> to store the calculated
calibration factors. You might press <Print F5> to print out the factors. By
pressing <Repeat F4> you can repeat the calibration. Press <ESC> to exit
without any changes.
C49IB08A 107
6 Software
Warning:
The surface of the calibration insert is hot. Do not touch the calibration insert
for at least 20 min after finishing the calibration.
15. To avoid entering wrong calibration constants, remove the adhesive label
from the sensor arm, if new calibration factors have been stored in the
instrument.
16. After complete cooling, clean the calibration insert with a damp cloth, if
necessary. Let the calibration insert dry completely before next use.
C49IB08A 108
6 Software
You can use the temperature sensor check menu to test the accuracy of the
temperature sensor between 40 °C and 220 °C, by use of the calibration
accessory unit.
This procedure cannot be used for calibration purposes.
Preparing the calibration accessory is the same as described for temperature
sensor calibration, refer to chapter 6.7.4, "Temperature Sensor Calibration with
the Calibration Accessory".
Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor
calibration menu. Confirm by pressing <Enter F1>.
1. Insert the rotor and the calibration accessory into the oven cavity as
described in chapter 6.7.4.1, "To Calibrate the p/T Sensor" and chapter
6.7.4.2, "To Calibrate the T-Probe". Connect the calibration accessory and
press <Check F1> to continue.
2. Enter the desired temperature value via the external keyboard or use the
<+ F2> or <– F3> keys. The test temperature can be between 40 °C and
220 °C. Recommended values are:
MF100 190
4. At the end of the heating phase the results are shown on the display. Using
the <Print F5> key you can print out the results or press <Repeat F4> to
repeat the check. (If the difference between the set value and the actual
value is more than 2 - 3, we recommend to repeat the check.)
C49IB08A 109
6 Software
Warning:
The surface of the calibration accessory is hot. Do not touch the calibration
accessory for at least 20 min after finishing the calibration.
You can perform a temperature sensor check to test the accuracy of the
temperature sensor between 40 °C and 220 °C, by use of the calibration
accessory unit.´
This procedure cannot be used for calibration purposes.
Preparing the calibration unit is the same as described for temperature sensor
calibration, refer to chapter 6.7.5, "Temperature Sensor Calibration with the
Calibration Unit".
Important:
Instead of the rotor (when using the calibration accessory) you have always to
use the assembled calibration unit.
Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor
calibration menu. Confirm by pressing <Enter F1>.
1. Insert the calibration unit into the oven cavity as described in chapter 6.7.5.1,
"To Calibrate the p/T Sensor". Connect the cable of the calibration unit and
press <Check F1> to continue.
2. Enter the desired temperature value via the external keyboard or use the
<+ F2> or <– F3> keys. The test temperature can be between 40 °C and
C49IB08A 110
6 Software
MF50 200
MF100 190
HF100 220
4. At the end of the heating phase the results are shown on the display. Using
the <Print F5> key you can print out the results or press <Repeat F4> to
repeat the check.
Important:
If the difference between the set value and the actual value is too big (more
than 1%), perform a calibration!
Warning:
The surface of the calibration unit is hot. Do not touch the calibration unit for at
least 20 min after finishing the calibration.
5. After complete cooling, clean the calibration insert with a damp cloth, if
necessary. Let the calibration insert dry completely before next use.
C49IB08A 111
6 Software
Temperature sensors are calibrated at the factory before shipping. The correct
factors are stated on an adhesive label attached to the temperature sensor and
on the supplied calibration certificate. The factors are assigned according to the
serial number of the sensor.
Before using a new sensor for the first time, the calibration factors have to be
entered into the Synthos 3000 software. You can enter the factors using the
function keys. It is recommended to connect an external keyboard for easy entry
of numbers and letters.
Use the arrow keys <Ï F4> and <Ð F5> to select the temperature sensor settings
menu. Confirm by pressing <Enter F1>.
- If you want to enter factors for a new sensor, use <Ï F4> or <Ð F5> to
select "New Sensor“.
- For editing factors of an installed sensor, use <Ï F4> or <Ð F5> to select
an installed sensor.
- Press <Exit F1> to exit without any changes.
- For printing press <Print F3>.
Important:
Entering wrong values will result in incorrect measurement!
C49IB08A 112
6 Software
2. Use the arrow keys <Í F4> and <Î F5> to move the cursor.
3. Enter or edit the values for the sensor via the function keys <+ F2> or <– F3>
or use an external keyboard:
- Serial number of the sensor (Sensor ID)
- Calib. Factor A
- Calib. Factor B
4. Press <Save F1> to store the entries. You might press <Discard F3> to
cancel the entries / modifications. By pressing <Delete F4> you can remove
an entry completely from the list. Press <ESC> to exit without any changes.
C49IB08A 113
6 Software
Hint:
For checking the accuracy of the IR sensor you have to perform an IR check
instead of a calibration (refer to chapter 6.7.11, "IR Sensor Check with the
Calibration Accessory"). Possible deviations of the IR measurement are not
shown during the calibration procedure.
Important:
For calibration of the IR sensor with the calibration accessory a calibration
accessory version 4 is required.
Indicates version 4
44
Use the arrow keys <Ï F4> and <Ð F5> to select the IR sensor calibration menu.
Confirm by pressing <Enter F1>.
Important:
• Make sure you have entered the calibrator constant correctly.
- The calibrator constant is important for correct calibration.
Calibration without entering this constant or with an incorrect constant
leads to incorrect IR values and may leads to vessel overheating.
• Make sure that the vessel holder is correctly aligned.
- The correct alignment of the vessel holder is important for correct
calibration. Incorrectly aligned vessel holder leads to vessel
overheating.
C49IB08A 114
6 Software
This procedure works with both, Rotor 16 as well as Rotor 8S. For proper setup
with Rotor 8S refer to chapter 6.7.4.2, "To Calibrate the T-Probe".
1. Place the rotor in the oven cavity so that the calibration accessory is posi-
tioned exactly above the IR sensor. You can use the sensor for the rotor lid
detection as an orientation guide.
Important:
Do not close the door of the instrument as this will damage the cable of the
calibration accessory.
2. Connect the cable of the calibration accessory to the CAL port on the right-
hand side of the instrument. Tighten the fastening screws of the plug to
ensure good contact.
C49IB08A 115
6 Software
which takes the emission value of the surface of the calibration accessory
into consideration. This calibrator constant is given on the supplied
calibration certificate.
5. Set the calibration temperature (range: 100 - 220 °C) depending on your
main operation range. Recommended values are:
MF100 160
HF100 210
C49IB08A 116
6 Software
8. When pressing <Calib F1> the calibrator is heated. After a short period for
equilibration the turntable starts to rotate, indicating the beginning of the
calibration.
9. After completing the calibration you can either save the new transmission via
<Save F1>, print the calibration data via <Print F5> and/or repeat the
calibration via <Repeat F4>. Press <ESC> to exit without any changes.
%
Warning:
The surface of the calibration accessory is hot. Do not touch the calibration
accessory for at least 20 min after finishing the calibration.
C49IB08A 117
6 Software
Hint:
For checking the accuracy of the IR sensor you have to perform an IR check
instead of a calibration (refer to chapter 6.7.12, "IR Sensor Check with the
Calibration Unit"). Possible deviations of the IR measurement are not shown
during the calibration procedure.
Before calibrating the IR sensor, the calibrator constant has to be entered, which
takes the emission value of the surface of the calibration unit into consideration.
This calibrator constant is given on the supplied calibration certificate.
Use the arrow keys <Ï F4> and <Ð F5> to select the IR sensor calibration menu.
Confirm by pressing <Enter F1>.
2. Place the assembled unit carefully into the microwave cavity to avoid
scratches in the oven. Make sure that the calibration insert is opposite to the
sensor for the protection lid detection (on the right hand side of the oven
cavity).
Take care that the metal support fits as accurately on the turntable. Then
connect the cable of the calibration insert to the CAL-port on the right-hand
side of the instrument. Tighten the screws of the cable plug to ensure good
contact.
C49IB08A 118
6 Software
Important:
Do not close the door of the instrument as this will damage the cable of the
calibration unit.
C49IB08A 119
6 Software
Important:
• Make sure you have entered the calibrator constant correctly.
- The calibrator constant is important for correct calibration.
Calibration without entering this constant or with an incorrect constant
leads to incorrect IR values and may leads to vessel overheating.
• Make sure that the vessel holder is correctly aligned.
- The correct alignment of the vessel holder is important for correct
calibration. Incorrectly aligned vessel holder leads to vessel
overheating.
5. Set the calibration temperature (range: 100 - 220 °C) depending on your
main operation range.
Recommended values are:
MF50 140
MF100 160
HF100 210
XF100/XQ80 220
C49IB08A 120
6 Software
6. Press <Next F4> and start the calibration while pressing <Calib F1>.
7. When pressing <Calib F1> the calibrator is heated. After a short period for
equilibration the turntable starts to rotate, indicating the beginning of the
calibration.
8. After completing the calibration you can either save the new transmission via
<Save F1>, print the calibration data via <Print F5> and/or repeat the
calibration via <Repeat F4>. Press <ESC> to exit without any changes.
C49IB08A 121
6 Software
%
Warning:
The surface of the calibration unit is hot. Do not touch the calibration unit for at
least 20 min after finishing the calibration.
By performing the IR sensor check you can observe a possible deviation of the IR
Sensor. This procedure cannot be used for calibration purposes.
Indicates version 4
Use the arrow keys <Ï F4> and <Ð F5> to select the IR sensor check menu.
Confirm by pressing <Enter F1>.
C49IB08A 122
6 Software
Warning:
The surface of the calibration accessory is hot. Do not touch the calibration
accessory for at least 20 min after finishing the calibration.
C49IB08A 123
6 Software
The accuracy of the IR sensor can be tested by using the IR sensor check.
By performing the IR sensor check you can observe a possible deviation of the IR
Sensor. This procedure cannot be used for calibration purposes.
Use the arrow keys <Ï F4> and <Ð F5> to select the IR sensor check menu.
Confirm by pressing <Enter F1>.
C49IB08A 124
6 Software
Warning:
The surface of the calibration unit is hot. Do not touch the calibration unit for at
least 20 min after finishing the calibration.
The solvent sensor is calibrated at the factory before shipping. The correct
calibration constant is stated on an adhesive label attached to the solvent sensor
and on the supplied calibration certificate.
Before using a new sensor for the first time, the calibration factors have to be
entered into the Synthos 3000 software.
The settings of the solvent sensor can be checked by using the solvent sensor
settings.
Use the arrow keys <Ï F4> and <Ð F5> to select the solvent sensor settings
menu. Confirm by pressing <Enter F1>.
C49IB08A 125
6 Software
2. Enter the calibration value noted on the calibration certificate of your sensor.
4. Press <New F2> to add a new setup. You might press <Discard F3> to
cancel the entry. Press <ESC> to exit without any changes.
C49IB08A 126
6 Software
Hint:
This function is only needed if the whole exhaust unit has been replaced.
Use the arrow keys <Ï F4> and <Ð F5> to select the exhaust unit installation
menu. Confirm by pressing <Enter F1>.
To install an exhaust unit press <Edit> and select the type. Confirm by pressing
<Install>.
The library maintenance tool is a software function for the maintenance and
management of the method library. After opening the "Library Maintenance Tool“
menu, a list of all stored methods will be displayed.
Hint:
To simplify matters you can jump to an appropriate method by pressing the
initial letter.
Use the arrow keys <Ï F4> and <Ð F5> to select the library maintenance tool
menu. Confirm by pressing <Enter F1>.
%
%
1. Use the external keyboard (with page up and page down) to select the
sample. The active entry is highlighted.
C49IB08A 127
6 Software
- View:
Allows you to view the methods on the display. In the "view" mode you may
add notes but you cannot modify the method.
- Lock:
Use this feature if you want to prevent a method from being modified.
Locked methods can be unlocked by repeating this command. A locked
method is marked by an "L" on the left side of the method name. When
editing a locked method in the library menu, it will be stored as a new
method.
- Hide:
Allows you to "hide" a method. The method is not visible in the library menu
but not deleted. Hidden methods are marked with "H“.
- Delete:
Allows you to delete a method. You must confirm this step by answering the
prompt with <Yes F1> or you can abort it by pressing <No F5>.
- Send:
Allows you to send all methods within an encrypted txt-file to an external
PC.
You can transfer the encrypted method file from Synthos 3000 to your PC for
backup purposes. It is not possible to edit the methods within this file.
Interconnect the external PC with COM1 of Synthos 3000 using a serial link cable
(0-Modem) and switch on Synthos 3000.
C49IB08A 128
6 Software
1. To store the file select "Capture Text- Stop" from the "Transfer" menu.
Hint:
For security reasons, the transferred method file is encrypted, thus it cannot
be edited on the PC.
The menu "Advanced Service Routines" is for authorized service engineers only.
For this reason it is secured with a password.
C49IB08A 129
6 Software
6.8 Info
Pressing <Info F5> on the main menu provides the Info screen giving the
following information about the instrument and Anton Paar GmbH.
6j\dS]c
• Instrument:
- Postal address
- Telephone and fax number
- Email and URL address
When pressing <Next F1>, the usage history is shown (v2.0 or higher).
C49IB08A 130
6 Software
6.9 AP Softprint
AP Softprint is used to transfer run data from Synthos 3000 directly into a
Microsoft Excel sheet. This Excel Add-In supports the following functions:
The program is available free of charge for users of Anton Paar's Synthos 3000
at our homepage (www.anton-paar.com) and already included on the Manual CD.
Hint:
For detailed information refer to the instruction manual for AP Softprint on the
Manual CD (Doc. No. B83IB39) or download this document from our
homepage.
RTP
C49IB08A 131
7 Troubleshooting
7 Troubleshooting
This section includes the most important error messages that might be given
during the operation of Synthos 3000. The tables below give an explanation of
error causes and indicate possible remedies.
If you have any problems which are not described below, do not hesitate to
contact our local customer service.
7.1 Software
The most important error messages appear as pop-up error messages. This is a
highlighted text on the instrument display which gives a description of the error
situation and recommends ways to remedy these errors.
More detailed error information is frequently offered by Synthos 3000 in the form
of error codes. In contrast to "pop-up error messages" these error codes have to
be accessed and viewed manually in an error message menu. The error menu
(list) includes a print option, i.e. you can print the listed error codes and messages
on a connected printer or send them to an external computer.
• Via several selected "pop-up error messages". In such cases the <F5>
function key is re-assigned for details, i.e. will give you access to the error
menu by pressing <Details F5>.
C49IB08A 132
7 Troubleshooting
>?-DUPoDUF._]g]D[]qDg]`_+
>
?.5$6F_f`U._]g+D][FP
>
?.5$6F_f`U5FNFTg]`_7]^F
>
?.5$6F_f`U5FNFTg]`_7]^F
>
?4`oFU(`DUP8_W_`o_5FfT`
C49IB08A 133
7 Troubleshooting
Error message
Explosion!
Rotor may be polluted with reagents / sample! In case of vessel breakage, do
not continue with your work! Danger of microwave leakage! (see F4 Info)
Safety cooling will not stop automatically!
Power off in: xx:xx min.
Fire or Overtemperature!
Danger of hot vessels and reagents! Let cool down for at least 6 hours before
opening! (see Info F4)
C49IB08A 134
7 Troubleshooting
Error message
Overtemperature Alarm!
The magnetron´s temperature limit was exceeded!
C49IB08A 135
7 Troubleshooting
Error message
• The exhaust unit was switched • Make sure that the cooling air
off by the thermal protector. inlet and outlet of the exhaust
• The exhaust unit is defective / unit are not blocked.
blocked. • Check proper function of the
• The exhaust unit is activated exhaust unit.
but the safety module SOLV • Reconnect any loose sensor
causes an error. hoses.
• The sensor hoses are loose. • If further maintenance is
required, contact our local
customer service.
C49IB08A 136
7 Troubleshooting
Error message
Stirrer blocked!
The power supply of the power unit failed.
Warning!
Pressure increase exceeded 3.0 bar/s after xx:xx min. Process continues
automatically. Change profile or sample weight to optimize the method!
C49IB08A 137
7 Troubleshooting
Error message
STOP!
The IR sensor is not activated! For safe operation you must use a slightly
increasing power profile. Do you want to continue?
Start temperature lower than 10 °C! Please check the temperature sensor.
The starting pressure is not within the range of 2 to 25 bar! Please close the
rotor in the proper way and restart the process.
C49IB08A 138
7 Troubleshooting
Error message
• You are using the wrong rotor • Use the correct rotor type.
type for the selected method. • Clean the position recognition
• The optical sensor for position sensors with a damp cloth (or
recognition is contaminated or with a cotton bud).
defective. • Check and clean the reflective
• The rotor does not turn. stripes (vessel position marks)
• The reflective stripes on the on the rotor bottom.
rotor are missing or dirty. • If the problem still persists,
contact our local customer
service.
Invalid rotor type. Please change method according to the used rotor.
• You are using the wrong rotor • Use the correct rotor type or
type for the selected method. change the method.
Please install p/T sensor to run temperature controlled reaction or change your
method according to the used sensors!
• You have started a method that • Insert the p/T sensor or change
requires the optional p/T the method.
sensor but the sensor was not • If the p/T sensor is installed,
detected. check the correct alignment of
the sensor. If it is aligned
properly but the failure still
persists, contact our local
customer service.
C49IB08A 139
7 Troubleshooting
Error message
• The power plug of the exhaust • Insert the plug of the exhaust
unit is not connected or is not unit into the appropriate socket
connected properly. at the rear side of the
• The exhaust unit is defective. instrument.
• Check if it is connected
properly.
• If it is connected properly but
the problem still persists,
contact our local customer
service.
The power unit reports an alarm and is not ready for operation!
Please power up the device again.
C49IB08A 140
7 Troubleshooting
Error message
The Temperature Sensor was not calibrated! Please refer to menu service for
calibration.
C49IB08A 141
7 Troubleshooting
Error message
C49IB08A 142
7 Troubleshooting
Error message
• The door opened during the • Let the instrument cool down
run due to thermal influences. and contact our local customer
• Failure of power supply. service.
C49IB08A 143
7 Troubleshooting
Error message
C49IB08A 144
7 Troubleshooting
Error message
Turntable Error!
The turntable is not rotating.
• The rotor has stopped during a • Check the turntable and the
run or the rotor does not rotate reflective stickers.
at the defined speed. • If the error still persists, contact
our local customer service.
C49IB08A 145
7 Troubleshooting
Error message
C49IB08A 146
7 Troubleshooting
Error message
The power ratio exceeds the range 0.7 and 1.3 and cannot be saved. Please
refer to the user manual!
C49IB08A 147
7 Troubleshooting
Error message
The required microwave power cannot be achieved due to the power supply
limitations. The process has to be aborted.
C49IB08A 148
7 Troubleshooting
Error message
C49IB08A 149
7 Troubleshooting
Error message
C49IB08A 150
7 Troubleshooting
Malfunction
• The rotor lid is not positioned • Turn the lid to the correct
correctly. position
• The venting screw is broken. • Rotor 16:
• The venting system of the seal Allow the vessel to cool down
has been clogged. completely and remove it from
the rotor. Carefully and slowly
open the screw cap. Venting
will occur via the safety
grooves in the liner.
• Rotor 8S:
Allow the vessels to cool down
completely. Slowly open the
thumb wheels of Rotor 8S
under a fume hood. Venting will
occur via the safety grooves in
the liner.
• Purge the seal (refer to chapter
9.4, "Cleaning and
Maintenance" for Rotor 16 and
chapter 10.4, "Cleaning and
Maintenance" for Rotor 8S).
C49IB08A 151
7 Troubleshooting
Malfunction
• The seal was not expanded • Expand the seal for at least
properly. 3 seconds or use the seal
• The sealing lip is damaged. forming device. Expand a new
seal for 10 seconds.
• Replace the seal and/or liner.
7.3 Sensors
Malfunction
Traces of oil on the p/T sensor of Rotor 16 resp. pressure sensor of Rotor 8S
C49IB08A 152
8 Rotor 48MF50
8 Rotor 48MF50
The following chapters contain a description of Rotor 48MF50 and provide
information on handling, cleaning and maintenance.
Rotor 48 has been designed specially for high sample throughput and is used for
a wide range of standard applications with Synthos 3000, which require pressures
and temperatures of up to 20 bar and 200 °C, such as library generation and
functionalization. It can be operated with up to 48 pressure vessels of the type
MF50.
The rotor lid with a bayonet locking closes the rotor and ensures protection in
case of a sudden overpressure release (e.g. breakage of a safety disk).
C49IB08A 153
8 Rotor 48MF50
Important:
Never attempt to change the rotor geometry or to bring metal parts into the
microwave field. This will cause serious damage.
Rotor 48 Lid
Interlock Bolt 48
Rotor 48 Tray
Rotor 48 Body
C49IB08A 154
8 Rotor 48MF50
MF50
Liner Material PFA
Vessel Jacket Fiber-reinforced PEEK
Max. Operating Pressure 20 bar (290 psi)
Opening Pressure 30 bar (435 psi)
a 200 °C, for 30 min
Max. TINTERNAL
b 140 °C
Max. TIR
Nominal Volume 50 mL
Min. Filling Volume 6 mL
cMax. Filling Volume 25 mL
Typical Applications Library generation, functionalization/
derivatisation, small scale method
optimisation, solid-phase synthesis;
a. Temperature inside the reference vessel, measured via the p/T
sensor
b. Surface temperature at the base of the vessel, measured via the IR
sensor
c. The max. filling volume may exceed 25 mL for special methods
after consultation of Anton Paar application specialists
Important:
Do not fill in less than the minimum volume of 6 mL.
When using a volume of less than 6 mL, the amount of liquid phase in the
reaction vessel will be too low. As a result, microwave energy will couple
directly to parts of the vessel, the rotor or the sample. This may cause
damage to rotor components and pressure vessels.
C49IB08A 155
8 Rotor 48MF50
For uniform heating, the vessels have to be loaded symmetrically into the rotor. If
the number of used vessels does not allow symmetrical loading it is possible to
substitute the missing vessel positions with blanks. (Refer to Chapter 8.4.3,
"Loading Rotor 48")
C49IB08A 156
8 Rotor 48MF50
At max. temperature of 200 °C the hold time of 30 min. must not be exceeded!
At temperatures lower than 180 °C hold times of 60 minutes and longer may be
applied without compromising the vessel stability. Heat-up times (ramp times) are
not critical concerning material stability. However, a warning message will be
displayed when entering programs longer than 60 minutes.
It is also possible to run power controlled experiments but power profiles must be
carefully chosen in order to avoid material stress to the vessel. The stored library
contains only tested temperature controlled methods as we highly recommend to
use the temperature/pressure sensor in any experiment.
Registered users can download the latest version of the method library from the
Anton Paar website (www.anton-paar.com).
In order to avoid thermal overload, the fan must be set to level 2 (medium
intensity) after 30 minutes if a combination of following conditions applies:
The cooling time at fan level 3 must be set to 30 minutes. Even longer cooling
times may be required for fully loaded rotors at high ambient temperature.
C49IB08A 157
8 Rotor 48MF50
Before you start your first experiment, it is recommended to get familiar with the
Synthos 3000 Microwave Reaction System. The best way is to start with either a
cleaning or a performance run from the method library.
If you are more familiar with microwave reactions, you should also perform a
blank run to clean the liner and seal (refer to chapter 8.5.1.1, "To Perform a
Cleaning Run").
Both ways enable you to operate the instrument without the risk of spontaneous
reactions and to learn about:
2. Prepare 10 mL of acidified water (1% of conc. nitric acid) per vessel. Using
this mixture will allow you also to check the functioning of the pressure and
temperature sensors.
Important:
Do not use deonised water due to the possible super-boiling effect. Using
pure water may result in a vessel explosion.
4. When doing a blank run use either 6 mL of aceton, isopropanol, or the same
reagent mixture which will be used for synthesis.
C49IB08A 158
8 Rotor 48MF50
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents. Follow any special national safety regulations.
Important:
Before the first experiment read chapter 11, "Applications" very
carefully.
Guidelines for frequently used standard procedures are given in the method
library of Synthos 3000. Depending on a possibly different reaction behavior of
your reagents, the suggested mixture may have to be modified. The user is
responsible for this decision.
Important:
The use of the p/T sensor is highly recommended to thoroughly observe the
reaction progress.
C49IB08A 159
8 Rotor 48MF50
2. Add solvent.
Rinse the liner by adding the solvent to remove material sticking on the top of
the liner wall. Make sure that no deposits remain on the wall of the vessel.
The vessel is closed with a screw cap containing a conical seal made of PTFE-
TFM. Contrary to a lip-type seal, there is no need to expand this conical seal. This
facilitates vessel handling and saves time.
Sealing area
Important:
To avoid damage of the seal, do not close the screw cap without a liner
inserted into the vessel jacket.
Replace damaged seals immediately, as tightness of the vessel is not
ensured any longer.
Be aware that the O-Ring, which is positioned on the liner, fits properly to the
corresponding groove on the top of the liner. If the O-Ring is worn out, replace it
(refer to chapter 8.5.4.2, "To Replace the O-Ring").
C49IB08A 160
8 Rotor 48MF50
O-ring
1. Remove any solvents or water droplets from the vessel jacket and the outer
surface of the liner. Liquid residues between liner and vessel jacket may
cause overheating and damage to the reaction vessel!
Important:
Do not use any tools for closing the screw cap.
For appropriate reaction control the use of the optional p/T sensor M50 is
recommended.
Important:
The p/T sensor is equipped with a lip type seal instead of a conical seal. For
that reason it is important to expand the seal before use to ensure
tightness.
C49IB08A 161
8 Rotor 48MF50
1. Remove any solvents or water droplets from the vessel jacket and the outer
surface of the liner. Liquid residues between liner and vessel jacket may
cause overheating and damage to the reaction vessel!
3. Press the screw cap completely onto the seal forming tool for at least 3
seconds to expand the sealing lip. (New seals have to be expanded for at
least 10 seconds.)
Important:
Make sure not to damage the immersion tube of the p/T sensor when
handling the seal forming tool.
Important:
Be careful when inserting the immersion tube of the p/T sensor, especially if
higher amounts of bulky samples or a magnetic stir bar is used. The
immersion tube may break!
6. The seal of the p/T sensor is connected to the hydraulic system and thus
slightly movable. For correct pressure measurement in the reference vessel,
the screw cap of the reference vessel has to be reopened from the stop
position for approx. 60° - 90°. If the reference vessel with the p/T sensor is
not reopened properly, pressure will be applied already by the vessel jacket.
As a consequence, the pressure readout will not show the correct reaction
pressure inside the reference vessel.
C49IB08A 162
8 Rotor 48MF50
90°
90°
The following procedures describe how to load the reaction vessels into the rotor
and how to close and insert the rotor into Synthos 3000.
1. Place the vessels into the rotor following the recommended loading patterns.
To ensure uniform temperature distribution during the reaction process, the
rotor has to be loaded symmetrically. The recommended load patterns are
shown in Fig. 8 - 11.
2. If applicable: Insert the reference vessel with p/T sensor into position 1.
C49IB08A 163
8 Rotor 48MF50
Important:
Never use less than 4 vessels to avoid overheating of rotor or vessel parts!
3. Reduce the max. power in the method when using less than 48 vessels. As a
guideline, limit the max. power output to 800 W (8 vessels) or 400-600 W
(4 vessels), respectively.
4. If you have other numbers than the recommended ones to be processed, fill
the vacant positions with pure solvent to obtain a symmetrical load pattern.
C49IB08A 164
8 Rotor 48MF50
After the rotor and vessels have been prepared, the loaded rotor is put into the
microwave cavity. The experiment is started from the software menu.
Synthos 3000 controls all parameters automatically and usually does not require
user intervention.
At the end of a run, the cooling starts automatically. The safety cooling is also
activated automatically if an experiment run is aborted. The cooling is stopped as
soon as all vessels have cooled down to an internal temperature below 60 °C and
an IR temperature below 50 °C, depending on the installed sensors.
Warning:
Do not attempt to remove the rotor and vessels from the instrument before the
safety cooling has finished! Danger of hot and pressurized vessels!
Refer to chapter 3, "Safety Instructions".
2. To put the loaded rotor into the microwave cavity, hold it with both hands at
the rotor bottom plate. Place it on the turntable inside the oven cavity until it
engages onto the rotor drive.
C49IB08A 165
8 Rotor 48MF50
5. Start the experiment by pressing the green <Start> button after the system
check was successful. During the run the display informs you about the
reaction progress.
1. To interrupt (pause) the program, press the <Stop> button once to shut
down the microwave energy.
The program can be aborted at any time, pressure and temperature data
acquisition will be continued. To continue with the run, press the <Start> but-
ton.
1. After the run has been finished, select from the menu on the display one of
the following options:
• Print: Print out a summary of the last run, including pressure and
temperature graphs, on a connected printer.
• Send: Transfer all relevant data of the last run to an external
computer via the serial interface.
• Review: View all relevant data of the last experiments on the graphic
display and allows to print them on a connected printer or send them
to an external computer.
• Next Run: Return to the sample menu to modify and/or restart the
current method.
2. Take the rotor out of the cavity and unload it as soon as practicable.
C49IB08A 166
8 Rotor 48MF50
After the cooling has finished vent the pressure vessels under a fume hood prior
to opening. Follow these instructions to ensure safe and proper handling of the
vessels.
Warning:
Take safety precautions: Use goggles, protective gloves and lab coat.
Depending on the application, noxious gases may be released during the
venting procedure.
1. Remove the rotor from the oven cavity and place it into a fume hood.
2. Turn the bayonet lid of the rotor counterclockwise to the stop position.
3. Carefully lift off the lid and place it with the top facing upwards on a clean
surface.
4. Remove the reference vessel with the p/T sensor from the rotor.
5. Unscrew the p/T sensor carefully in front of the supplied protection shield,
which is placed into a beaker. The beaker has to be filled with water to
absorb potential solvent drops.
The reaction gases are vented via venting grooves in the liner through the
vent pipe.
Warning:
Make sure that the vent pipe points away from your face!
Vent pipe
C49IB08A 167
8 Rotor 48MF50
Important:
Take care not to damage the immersion tube.
8. Place the sensor with the liner in a rack. If applicable, rinse the immersion
tube while removing it from the liner.
1. Vent each pressure vessel by opening the screw cap in front of the supplied
protection shield, which should be placed into a beaker.
Warning:
Make sure that the vessel points away from your face!
Vent pipe
correct incorrect
C49IB08A 168
8 Rotor 48MF50
Warning:
Do not use any tool for opening the pressure vessel, even if the vessel cannot
be opened by hand due to a high residual pressure.
If so, let the vessel stand overnight and try to open it at the next day.
Do not force cooling of the vessel.
Reduce the sample amount during subsequent runs.
Hint:
If the reaction gases were not released via the venting system, the gases can
also escape safely via the venting grooves in the liner.
Venting grooves
2. Remove the vessel jacket and store it in a safe place or leave them in the
Rotor tray.
Warning:
Always open the vessel in a fume hood as noxious gases may be released.
Leakage may lead to chemical deposits on vessel or rotor parts.
3. Pop off the cap by pressing down on one side with your thumb to remove it
from the liner.
C49IB08A 169
8 Rotor 48MF50
In this chapter you will find information on how to clean and inspect the instrument
and the vessels after every reaction run, but also how to proceed after breakage
of a safety disk or vessel.
At the end of this chapter, you will find appropriate maintenance procedures for
the p/T sensor.
Important:
Proper cleaning and maintenance of all instrument parts, especially reaction
vessels and rotors, is vital to extend the service life and to ensure proper
functioning of the instrument.
Disregarding the instructions given in this chapter will lead to premature
failure and cost-intensive repairs.
The most effective cleaning procedure for reaction vessels is to perform a blank
reaction in Synthos 3000.
1. Fill each vessel with at least 6 mL HNO3 (10%) or any cleaning solvent (ace-
tone, isopropanol etc.).
2. Select a cleaning program from the method library stored in Synthos 3000.
Instructions for handling and preparing the vessels and rotor are given in
chapter 8.4.4, "Performing a Synthesis Run".
4. If there are vacant positions during a reaction, these positions can be filled
up with “solvent blanks” for cleaning the vessels and seals.
C49IB08A 170
8 Rotor 48MF50
Carefully clean the reaction vessels after every run to protect subsequent
samples against contamination, but also to avoid corrosion and overheating of the
vessel parts and rotor due to contamination with residues of reagents. This will
help to increase the service life of these parts.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
2. Rinse the inner side of the seal thoroughly with deionized water and/or
acetone.
3. Remove any residues of reagents from the screw cap and the vessel jacket
with deionized water.
C49IB08A 171
8 Rotor 48MF50
Hint:
You can alternatively clean the liners in a dishwasher.
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Rinse the inner side of the seal thoroughly with deionized water and/or ace-
tone with the immersion tube facing downwards.
2. Carefully rinse the immersion tube with deionized water and/or acteone.
C49IB08A 172
8 Rotor 48MF50
3. If necessary, carefully remove solid residues from the immersion tube with a
acteone-damp cloth. Do not use sharp tools! Stubborn residues are removed
best during a run.
1. Wipe the interior of the microwave cavity and the safety door with a damp
cloth.
2. Clean the outer surfaces of Synthos 3000 with a damp cloth. Do not use
organic solvents!
Important:
If you detect any reagent residues at the bottom of the cavity, proceed with a
thorough inspection of your reaction vessels for signs of leakage.
C49IB08A 173
8 Rotor 48MF50
Perform a visual check of the reaction vessels, rotor and instrument after every
experiment.
Important:
Replace damaged parts immediately! If not detected, even slight damages
may develop into cost-intensive repairs.
1. Check the liner for deformations and scratches in the sealing area.
Replace liners with a scratched sealing area (approx. 1 cm from the top), as
tightness during subsequent experiments is not ensured anymore.
2. Check the O-ring of the liner for damage. Replace corroded or worn O-rings.
3. Check the PEEK vessel jacket for any sign of cracks and corrosion.
4. Check the conical seal for any sign of damage, like deformation or cuts.
2. Check the sealing lip of the lip-type seal for any sign of damage, e.g.
deformation or cuts.
3. Check the p/T sensor for oil leakage of the hydraulic system.
An oil leakage requires repair at Anton Paar, contact our local distributor.
Fig. 8 - 22 Potential oil leakage from the hydraulic system on the p/T sensor
C49IB08A 174
8 Rotor 48MF50
1. Check both the rotor body and the rotor tray for any obvious signs of dam-
age, e.g. melting of the rotor tray.
Perform these combined cleaning and maintenance procedures after one or two
working weeks, depending on the operation frequency of Synthos 3000. If you
have to handle unknown sample material, perform these procedures daily.
Following these procedures will help to increase the service life of rotor and
vessel components, but will also reduce the risk of downtimes due to equipment
failure.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
C49IB08A 175
8 Rotor 48MF50
3. Check the liner for any sign of damage, like deformations, cracks or
scratches of the sealing area.
Replace damaged liners immediately, as tightness during subsequent
experiments is not ensured anymore.
4. Perform a cleaning run to remove any detergent residues from the liner and
seal (refer to chapter 8.5.1.1, "To Perform a Cleaning Run").
Alternatively: Steam the liner with nitric acid vapors overnight using a
steaming apparatus. When doing so, remove the black O-ring from the
liner prior to steaming.
Important:
Do not store the liner in the vessel jacket.
1. Squeeze the top of the liner and push the O-ring slightly forwards.
C49IB08A 176
8 Rotor 48MF50
3. Slip the new O-ring over the top of the liner until it fits into the groove of the
liner.
1. Clean the vessel jacket with water, a mild detergent and a soft brush.
3. Let the vessel jacket dry completely before the next use.
4. Check the vessel jacket for any sign of damage, like cracks or chemical
attack (corrosion).
Replace PEEK vessel jackets immediately if you detect cracks or severe
material degradation.
1. Follow the instruction given in Chapter 8.5.2.2, "To Clean the p/T Sensor
After Every Run" and Chapter 8.5.3.2, "To Check the p/T Sensor After Every
Run".
C49IB08A 177
8 Rotor 48MF50
1. Wipe the interior of the microwave cavity and the safety door with a damp
cloth.
2. Clean the outer surfaces of Synthos 3000 with a damp cloth. Do not use
organic solvents!
3. Clean the Fresnel lens of the IR sensor with a damp cotton bud.
Sensor for the protection lid detection Sensor for the position recognition
4. Clean the sensors for the position recognition (at the bottom of the cavity)
and for the protection lid detection (at the right hand side of the cavity) with a
damp cotton bud.
5. Check the microwave cavity and safety door for any sign of damage or
deformation.
Warning:
Do not use the instrument if you detect any deformation on the microwave
cavity or the door. Danger of microwave leakage!
Contact an authorized service engineer.
C49IB08A 178
8 Rotor 48MF50
If a safety disk breaks during the experiment (e.g. due to spontaneous reactions),
reagents usually escape from the reaction vessels. This leads to contamination of
the rotor, the outer surfaces of the vessels and the microwave cavity.
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to Chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
3. Carefully remove the rotor, place it under a fume hood, remove the rotor lid
and vent the vessels (refer to chapter 8.4.5.2, "To Vent and Open the
Pressure Vessel").
5. Take out and open the reaction vessels (refer to chapter 8.4.5.2, "To Vent
and Open the Pressure Vessel"). It is most probable that the reaction was
incomplete, so you may discard the solutions.
Important:
After the breakage of a safety disk, both safety disk and conical seal have to
be replaced.
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
and a soft brush if necessary.
C49IB08A 179
8 Rotor 48MF50
For replacement of seal you have to disassemble the screw cap. Therefore you
will need the unlock disk and the seal forming device.
1. Thoroughly rinse the complete screw cap assembly under running water to
remove any reagent residues.
2. Place the unlock disk on the pressure vessel jacket (without liner).
3. Compress the conical seal by closing the pressure vessel with the unlock
disk in place. Close it a half turn further than hand-tight.
4. Open the screw cap again and put in on a clean working bench.
5. Use the seal forming tool to push the seal out of the screw cap with the two
pins.
6. Replace the defective parts (safety disk, seal and vent pipe).
C49IB08A 180
8 Rotor 48MF50
7. Clean the screw cap with water, a mild detergent and a soft brush. Remove
any deposits, especially from the inner side of the screw cap.
1. Insert the safety disk in the center of the screw cap. Take care that the safety
disk is positioned correctly.
Safety Disk
3. Put the assembled seal with the vent pipe on a liner with an O-ring.
C49IB08A 181
8 Rotor 48MF50
4. Turn the liner upside down and push the seal into the screw cap until you
reach the locking position. Take care to immerse the vent pipe into the
appropriate borehole of the screw cap.
5. Use a pressure jacket to fix the seal by tightening the screw cap.
6. Unscrew the vessel and check the correct (centered) position of the safety
disk.
C49IB08A 182
8 Rotor 48MF50
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Remove the turntable, the drive star and the drive base from the microwave
cavity.
2. Clean these parts under running water, use a mild detergent and a soft brush
if necessary.
C49IB08A 183
8 Rotor 48MF50
3. Use a damp cloth to thoroughly clean the microwave cavity and the white
PTFE covers.
4. Clean the door in the same way, also the contacting surfaces between door
and instrument.
5. Clean the Fresnel lens of the IR sensor with a damp cotton bud.
6. Clean the sensors for the position recognition and for the protection lid
detection in the same way.
Sensor for the protection lid detection Sensor for the position recognition
8. Check the microwave cavity and door for signs of damage or deformation.
Warning:
Do not use the instrument if you detect any deformation on the microwave
cavity or the door. Danger of microwave leakage!
Contact an authorized service engineer.
9. Check all other parts for signs of damage, like cracks or fissures. Replace
damaged parts.
C49IB08A 184
8 Rotor 48MF50
Although very unlikely, it may happen that a reaction vessel breaks during a run
instead of the safety disk. This may be caused by spontaneous exothermal
reactions, which are too fast for the safety disk, or by previously damaged vessel
jackets.
In case of a vessel breakage, solvent and reagents will be released together with
vessel debris inside the microwave cavity. This leads to contamination of the rotor,
the outer surfaces of the vessels as well as the microwave cavity.
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
Take care of splinters and reagent residues.
Warning:
After a vessel breakage you have to contact an authorized service engineer
who has to perform a microwave leakage test. Do not use the instrument
before it has passed this test successfully.
Due to the explosion either the door or the microwave cavity might have been
deformed. Danger of microwave radiation leakage!
4. Place the rotor under a fume hood, remove the rotor lid and vent the vessels
(refer to chapter 8.4.5.2, "To Vent and Open the Pressure Vessel").
6. Take out and open the reaction vessels (refer to chapter 8.4.5.2, "To Vent
and Open the Pressure Vessel"). It is most probable that the synthesis was
incomplete, so you may discard the solutions.
C49IB08A 185
8 Rotor 48MF50
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
and a soft brush if necessary.
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Follow the instruction given in chapter 8.5.2.2, "To Clean the p/T Sensor
After Every Run" and chapter 8.5.3.2, "To Check the p/T Sensor After Every
Run".
Carefully remove splinters and debris from the microwave cavity and follow the
instructions given in chapter 8.5.4.6, "To Clean and Check the Microwave Cavity".
Important:
After every vessel explosion contact an authorized service technician to
perform a microwave leakage test.
C49IB08A 186
8 Rotor 48MF50
1. Remove the safety bracket and the cover from the exhaust channel at the
rear of Synthos 3000 by loosening the two nuts.
1 2
3 4
Foam part
3. Clean the inside of the exhaust channel with a damp cloth and remove all
debris and reagent residues.
C49IB08A 187
8 Rotor 48MF50
Foam Part
Hint:
The foam part has to be installed distorted approximately 10 cm into the
exhaust channel and is for noise reduction.
5. Reinstall the cover and the safety bracket at the exhaust channel.
6. Check the proper installation of the foam seal between tube and exhaust
unit.
C49IB08A 188
8 Rotor 48MF50
Hint:
When disassembling the three interlock bolts the interlock ring drops down at
the bottom side of the rotor.
Interlock bolts
Interlock ring
3. Remove the vessel holders from the rotor inner base ring of the rotor.
4. Pull off the 16 silicone seals, which are mounted on the bottom of the vessel
holders in the center circle.
C49IB08A 189
8 Rotor 48MF50
Silicone seal
Important:
Do not dismount the 8 white reflective stripes at the bottom of the rotor body.
Do not dismount the black rotor center ring at the base of the rotor body.
1. Clean all parts under running water. Use a mild detergent and a soft brush if
necessary.
2. Remove splinters.
5. Check all parts for damages, especially vessel holders, silicone seals and
the rotor tray. Immediately replace damaged or broken parts!
1. Plug the 16 silicone seals onto the bottom of the vessel holders and check
for proper installation. Note that the 16 silicone seals are only fixed on those
vessel holders, which have to be installed in the center circle of the rotor
inner base ring.
C49IB08A 190
8 Rotor 48MF50
Silicone seal
2. Start the assembly by fixing the first vessel holder (with silicone seal) in the
center circle. The two feet of the vessel holders have to fit exactly into the
appropriate boreholes.
Take care that the bigger opening of the vessel holder faces in clockwise
direction.
Important:
If the vessel holders are not positioned correctly, heating and cooling behavior
of the rotor may be adversely affected.
3. Continue the assembly by fixing two vessel holders (without silicone seal) in
the outer and inner base ring to form a triple formation.
Fig. 8 - 44 Assembling the vessel holder in the rotor inner base ring (1)
C49IB08A 191
8 Rotor 48MF50
center circle
outer ring
inner ring
"triple formation"
Fig. 8 - 45 Assembling the vessel holder in the rotor inner base ring (2)
Hint:
The bigger openings of the three vessel holders belonging to one triple
formation have to show towards each other.
4. After completion of this triple formation, continue with the assembly of the
remaining vessel holders in the same sequence.
5. Make sure that all vessel holders are engaged properly in their positions.
6. Put the rotor tray onto the vessel holders. Take care that the position one on
the tray fits to the reflector mark on the rotor body.
C49IB08A 192
8 Rotor 48MF50
8. When all vessel holders are positioned correctly, fix the rotor tray by pressing
it down with your fingers and screw in the 8 Allen screws. Do not overtighten
the Allen screws!
C49IB08A 193
8 Rotor 48MF50
9. Finally screw in the 3 interlock bolts. For that reason tilt the rotor body a little
bit and insert the interlock ring into the recess until it rests on the rotor base.
Now turn the interlock ring until the three bore holes on the interlock ring
correspond with the three bore holes on the rotor tray.
Interlock Ring 48
Interlock Bolt
Interlock Ring
C49IB08A 194
8 Rotor 48MF50
In the following chapters you will find information on how to replace the seal and
safety disk, capillary and immersion tube of the p/T sensor.
After rupture of the safety disk you have to replace seal, seal husk and safety disk.
Losses of reagents or visible mechanical damage also indicate the necessity of
replacement.
Important:
When you replace the seal or remove it for cleaning, always replace the seal
husk! When removing and inserting the seal husk, do not use sharp objects,
which could damage it.
1. Put the unlock disk for seal onto the pressure jacket.
2. Compress the conical seal by closing the pressure vessel with the unlock
disk in place. Close it a half turn further than hand-tight.
C49IB08A 195
8 Rotor 48MF50
3. Open the p/T sensor again and put in on a clean working bench.
4. Use the seal ejector to push the seal out of the p/T sensor via the two pins
(through the two boreholes).
Press the seal ejector carefully down, do not bend the PEEK tube.
Seal ejector
Fig. 8 - 55 Using the seal ejector pin to push out the seal for p/T sensor
6. Remove the seal carefully by sliding to the end of the immersion tube. Avoid
any scratches on the immersion tube.
C49IB08A 196
8 Rotor 48MF50
Seal ejector
PEEK tube
PEEK tube,
do not bend!
Vent pipe
1. Stick the vent pipe with the corresponding O-ring into the seal.
O-ring
C49IB08A 197
8 Rotor 48MF50
3. Take care that the vent pipe fits exactly into the appropriate borehole of the
seal holder. Do not twist the seal to avoid bending of the vent pipe!
Push the seal further upwards to the seal holder by the help of a liner
(without O-ring).
Fig. 8 - 60 Fixing the seal for p/T sensor with the liner
4. Install the new seal husk using the supplied husk assembly tool.
C49IB08A 198
8 Rotor 48MF50
Seal husk
Fig. 8 - 61 Pushing the seal husk onto the immersion tube using the husk
assembly tool
Important:
The two ends of the seal husk have different diameters. Put the new seal husk
onto the immersion tube so that the wider end faces to the screw cap. To
provide good sealing properties, the dimensions of the seal husk are chosen
to fit tightly. Use the special husk assembly tool to slide the seal husk up to the
stop position.
5. Make sure that the lip of the seal is not damaged during the assembly.
Important:
After exchanging the seal, always carry out a performance run before the first
reaction to check the tightness of the seal. To do this, weigh the solution
before and after a performance run and check for losses. If the losses are
higher than 2% check the installation steps of the seal husk once again.
C49IB08A 199
8 Rotor 48MF50
Important:
After replacing the thermometer capillary you have to calibrate the
temperature sensor! Refer to chapter 6.7.5, "Temperature Sensor
Calibration with the Calibration Unit".
Carry out the steps described below only if you are able to perform a
temperature sensor calibration!
The capillary is very fragile. The most critical part is the connection between
the capillary and glass bulb, which can easily break.
If you are not completely sure how to replace the thermometer capillary
consult our local customer service.
2. Pull the knurled nut back (together with the PEEK sleeve of the capillary
tube) and take it out of the screw-type fitting.
3. The PEEK sleeve has to be separated from the immersion tube. Pull the
PEEK tube out of the metal sleeve which is positioned at the top of the
immersion tube. The use of a vice or pliers is recommended.
Fix the PEEK tube in the vice as shown below. Squeeze with the help of a
C49IB08A 200
8 Rotor 48MF50
screw driver the sensor from the PEEK sleeve. Take care not to damage the
immersion tube.
Vice
Screwdriver
Glass bulb
6. Insert the end of the PEEK tube into the metal sleeve up to the mechanical
stop position (3 - 4 mm) (with the help of a vice). The tube is automatically
fastened with a snap-in connector. Be very careful not to break the capillary.!
C49IB08A 201
8 Rotor 48MF50
7. Mount the new knurled nut (delivered with the new capillary) onto the metal
end of the capillary tube so that the metal end protrudes from the nut for
2 mm.
2 mm
8. Place the p/T sensor on the edge of a table to keep one hand free for
manipulation of the capillary.
9. Pull back the knurled nut a bit and insert it into the fitting of the p/T sensor
unit without changing the protruding length of the metal end.
10. Close the knurled nut hand-tight and perform a temperature sensor
calibration with the calibration unit. (Refer to chapter 6.7.5, "Temperature
Sensor Calibration with the Calibration Unit".)
If the capillary is not damaged and should be used together with a new immersion
tube, take care not to break the capillary while disassembling.
Remove the capillary tube as described in Chapter 8.6.2.1, "To Replace the
Thermometer Capillary".
C49IB08A 202
8 Rotor 48MF50
For replacing the immersion tube, follow the instructions below to ensure safe and
proper operation.
Important:
After replacing the immersion tube you have to calibrate the temperature
sensor!
When you replace the immersion tube, you also have to replace the seal
husk. When removing and inserting the seal husk, do not use sharp objects,
which could damage it.
C49IB08A 203
8 Rotor 48MF50
3. Check the surfaces of the p/T sensor for splinters and contaminated parts
and clean the surfaces with a moist cloth, if necessary.
Important:
Never clean the p/T sensor under running water as this may damage the
sensor electronics.
4. Insert a new immersion tube into the p/T sensor and tighten the immersion
tube by turning it clockwise. Take care not to damage the sealing area of the
immersion tube.
sealing area
5. Install the new seal husk. Refer to chapter 8.6.1.2, "To Install the Seal for p/T
Sensor".
6. Take care when inserting the capillary with the glass bulb through the metal
sleeve into the immersion tube. Refer to chapter 8.6.2.1, "To Replace the
Thermometer Capillary".
glass bulb
C49IB08A 204
9 Rotor 16MF100 / 16HF100
The rotor lid with a bayonet locking closes the rotor and provides protection in
case of a sudden overpressure release (e.g. rupture of a safety disk). Additionally,
the rotor lid serves as a rack to hold the liners before and after the reaction. The
positions are numbered to avoid mixing up of samples.
Openings on the side of the rotor lid facilitate venting of the reaction vessels after
reaction directly inside Synthos 3000, without the need of handling pressurized
reaction vessels.
C49IB08A 205
9 Rotor 16MF100 / 16HF100
Important:
Never attempt to change the rotor geometry or to bring metal parts into the
microwave field. This will cause serious damage.
Rotor lid
Reaction vessels
Rotor tray
Rotor body
Fig. 9 - 2 Rotor 16
C49IB08A 206
9 Rotor 16MF100 / 16HF100
• Fluoropolymers for the reaction vessel and for the self-sealing lip-type seal
• Vessel jacket for providing mechanical stability
• An easy-to-handle screw cap with venting system
• Metal safety disk for overpressure protection
MF100 HF100
Liner Material Fluoropolymer Fluoropolymer
Vessel Jacket PEEK Ceramic
Max. Operating Pressure 20 bar (290 psi) 40 bar (580 psi)
Opening Pressure 70 bar (1000 psi) 70 bar (1000 psi)
a 190 °C, for 30 min 240 °C
Max. TINTERNAL
b 160 °C 210 °C
Max. TIR
Nominal Volume 100 mL 100 mL
Min. Filling Volume 6 mL 6 mL
c 60 mL 60 mL
Max. Filling Volume
Typical Applications Parallel reactions, Parallel reactions,
multi-gram library gen- multi-gram library gen-
eration, solid-phase eration, solid-phase
synthesis synthesis
a. Temperature inside the reference vessel, measured via the p/T sensor
b. Surface temperature at the base of the vessel, measured via the IR
sensor
c. The max. filling volume may exceed 60 mL for special methods after
consulting Anton Paar GmbH
Important:
Do not fill in less than the minimum volume of 6 mL.
When using a total volume of less than 6 mL, the amount of liquid phase in the
reaction vessel will be too low to absorb microwave energy sufficiently. As a
result, microwave energy will couple directly to parts of the vessel, the rotor or
the sample. This may cause damage to components or may lead to self-
ignition and damage of the pressure vessel.
C49IB08A 207
9 Rotor 16MF100 / 16HF100
Screw cap
Seal
Liner fluoropolymer
Vessel jacket:
MF100: PEEK
HF100: ceramic
Important:
The ceramic vessel jackets (HF100) have to be handled with care. Avoid any
mechanical impact on these vessel jackets, otherwise crack growth will
increasingly occur, thus drastically raising the probability of a vessel
breakage.
C49IB08A 208
9 Rotor 16MF100 / 16HF100
Before you start your first experiment, it is recommended to get familiar with the
Synthos 3000 Microwave Reaction System. The best way is to start with either a
cleaning or a performance run from the method library.
If you are more familiar with microwave synthesis, you should also perform a
blank run to clean the liner and seal (refer to Chapter 9.4.1.1, "To Perform a
Cleaning Run").
Both ways enable you to operate the instrument without the risk of spontaneous
reactions and to learn about:
Important:
Always take care to select the appropriate vessel type in the method.
The maximum operating pressure and temperature values are defined within
the method.
Do not use both types of pressure vessel (MF100 and HF100) together!
Hint:
Before the first reaction remove the test certificates from the ceramic vessel
jackets. Keep these test certificates with the other instrument documents.
Important:
Do not use deionized water due to the possible super-boiling effect, which
may result in a vessel explosion.
C49IB08A 209
9 Rotor 16MF100 / 16HF100
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents. Follow any special national safety regulations.
Important:
Before the first experiment read chapter 11, "Applications" very carefully.
Do not fill less than 6 mL of sample and reagent into the vessel.
Guidelines for reagent mixtures are given in the method library of Synthos 3000.
Depending on possibly different reaction behavior of your reagent, the suggested
reagent mixture may have to be modified. The user is responsible for this
decision.
Important:
The use of the p/T sensor is highly recommended to thoroughly observe the
reaction.
C49IB08A 210
9 Rotor 16MF100 / 16HF100
2. Add solvent.
Rinse the liner by adding the solvent to remove any material sticking on the
top of the liner wall. Make sure that no deposits remain on the wall of the
vessel.
The vessel is closed with a screw cap containing a self-sealing lip-type seal made
of PTFE-TFM. Due to the pressure build-up during the heating phase, the sealing
lip is pressed against the liner wall, thus providing tight closure of the vessel.
The seal is a sensitive part. To ensure proper functioning and to avoid losses of
reagents or analytes, the seal has to be handled carefully as described below.
Important:
Take care when handling and storing the seal! The sealing lip can be
damaged easily.
Replace damaged seals immediately, as tightness of the vessel is not
ensured any longer. As a consequence, damage to the instrument, rotor and
vessels may occur due to escaping reagents.
Never store the seal standing on its sealing lip!
C49IB08A 211
9 Rotor 16MF100 / 16HF100
incorrect correct
The sealing lip has to be expanded shortly before use. To do so, you can use
either the seal forming tool or the optional seal forming device.
Important:
Do not start a run without having the seals properly expanded. In such a case
tightness of the vessel is not ensured any longer. As a consequence, damage
to the instrument, rotor and vessels may occur due to escaping reagents.
The seal forming device is used for the simultaneous expansion of 8 seals and for
safe storage. The sensitive sealing lip is protected and the working life of the seal
is extended. The design of the seal forming device allows you to expand and store
all types of seals for Rotor 16 and Rotor 8S.
C49IB08A 212
9 Rotor 16MF100 / 16HF100
1. Press the screw cap completely onto the seal forming tool for at least 3 sec-
onds to expand the sealing lip. Expand new seals for at least 10 seconds.
2. Immediately close the pressure vessel (refer to chapter 9.3.2.3, "To Close
the Pressure Vessel").
Note that the seal will regain its original shape after some time, therefore you
have to start the reaction within approx. 15 minutes. Otherwise, re-expand
the seal.
Important:
Make sure not to damage the immersion tube of the p/T sensor when
handling the seal forming tool.
C49IB08A 213
9 Rotor 16MF100 / 16HF100
1. Ensure that the cones are mounted properly on both the intermediate panel
and the upper panel, depending on the seal type.
To form MF and HF seals, the longer, smaller cone must protrude upwards
from the panel. The seals are expanded by the shorter, bigger cone.
2. Place four screw caps into the recesses of the lower panel with the seal
facing upwards.
Fig. 9 - 10 Placing the first screw caps into the seal forming device
3. Take the intermediate panel and carefully place it onto the seals.
C49IB08A 214
9 Rotor 16MF100 / 16HF100
Important:
The cones have to fit exactly into the seals. Take care not to place the cones
onto the sealing lips, as this will damage the seals.
Check for correct positioning of the seals.
4. Place another four screw caps into the recesses of the intermediate panel.
This side up
correct
incorrect
C49IB08A 215
9 Rotor 16MF100 / 16HF100
Correct Incorrect
1. Remove any solvent or reagent droplets from the vessel jacket and the outer
surface of the liner.
Important:
Take care that all vessel parts (liner, vessel jackets, screw caps) are
completely dry. Any residues of liquid between the liner and vessel jacket may
cause damage to the vessel.
C49IB08A 216
9 Rotor 16MF100 / 16HF100
Important:
Close the venting screw finger-tight only without using a tool.
The supplied key is used for opening only. The tight closure during reaction is
ensured via the front side of the white PTFE-plug of the venting screw.
Fig. 9 - 16 Venting screw in closed position: The front side of the PTFE-plug
seals the outlet opening
5. Close the screw cap by hand in clockwise direction until you reach the stop
position.
C49IB08A 217
9 Rotor 16MF100 / 16HF100
1. Remove any solvent or reagent droplets from the vessel jacket and the outer
surface of the liner. Liquid residues between liner and vessel jacket may
cause overheating and damage to the reaction vessel!
Important:
Close the venting screw finger-tight only without using a tool.
The supplied key is used for opening only. The tight closure during reaction is
ensured via the front side of the white PTFE-plug of the venting screw.
Important:
Make sure not to damage the immersion tube of the p/T sensor when
handling the seal forming tool.
5. Close the screw cap by hand in clockwise direction until you reach the stop
position.
C49IB08A 218
9 Rotor 16MF100 / 16HF100
Important:
Be careful when inserting the immersion tube of the p/T sensor, especially if a
magnetic stir bar and/or a heating element is used. The immersion tube may
break!
6. For correct pressure measurement, reopen the screw cap for approx. 45°!
45°
Hint:
The seal of the p/T sensor is connected to the hydraulic system and thus
slightly movable. For correct pressure measurement in the reference vessel,
its screw cap has to be reopened from the stop position for approx. 45°. This
is equivalent to 1.5 ribs on the screw cap.
If the screw cap is not reopened properly, pressure will be applied already by
the vessel jacket. As a consequence, the pressure readout will not show the
correct reaction pressure inside the reference vessel.
C49IB08A 219
9 Rotor 16MF100 / 16HF100
The following procedures describe how to load the reaction vessels into the rotor
and how to close and insert the rotor into Synthos 3000.
2. Place the other vessels into the rotor with the venting screws facing outward,
following the recommended loading patterns.
To ensure uniform temperature distribution during the reaction, the rotor has
to be loaded symmetrically. The recommended load patterns ensure uniform
heating of all vessels.
Venting screw
3. Reduce the max. power in the method when using less than 16 vessels. As a
guideline, limit the max. power output to 800 W (8 vessels) or 400-600 W
(4 vessels), respectively.
C49IB08A 220
9 Rotor 16MF100 / 16HF100
4. If you have other numbers than the recommended ones to be processed, fill
the vacant positions pure solvent to obtain a symmetrical load pattern.
Important:
Never use less than 4 vessels to avoid overheating of rotor or vessel parts!
close
2. Rotate the lid clockwise until the arrow points to the word "CLOSE".
The bayonet locking is now engaged.
After the rotor and vessels have been prepared, the loaded rotor is put into the
microwave cavity. The synthesis run is started from the software menu.
Synthos 3000 controls all parameters automatically and usually does not require
user intervention.
At the end of an experiment, the cooling starts automatically. The safety cooling
is also activated automatically if a run is aborted. The cooling is stopped as soon
as all vessels have cooled down to an internal temperature below 60 °C and an
IR temperature below 50 °C, depending on the installed sensors.
Warning:
Do not attempt to remove the rotor and vessels from the instrument before the
safety cooling has finished! Danger of hot and pressurized vessels!
Refer to chapter 3, "Safety Instructions".
C49IB08A 221
9 Rotor 16MF100 / 16HF100
2. To put the loaded rotor into the microwave cavity, hold it with both hands at
the rotor bottom plate. Place it on the turntable inside the oven cavity until it
engages onto the rotor drive.
5. Start the experiment by pressing the green <Start> button after the system
check was successful. During the run the display informs you about the
reaction progress.
1. To interrupt (pause) the experiment, press the <Stop> button once to shut
down the microwave energy.
The program can be aborted at any time, pressure and temperature data
acquisition will be continued. To continue with the run, press the <Start> but-
ton.
1. After the run has been finished, select from the menu on the display one of
the following options:
• Print: Print out a summary of the last run, including pressure and
temperature graphs, on a connected printer.
C49IB08A 222
9 Rotor 16MF100 / 16HF100
• Next Run: Return to the sample menu to modify and/or restart the
current method.
2. Take the rotor out of the cavity and unload it as soon as practicable.
After the cooling has finished either vent the pressure vessels inside the
instrument or remove the rotor and vent the vessels under a fume hood.
Warning:
Take safety precautions: Use goggles, protective gloves and lab coat.
Depending on the application, noxious gases may be released during the
venting procedure.
C49IB08A 223
9 Rotor 16MF100 / 16HF100
Important:
You must not open the venting screw completely, one turn counter clockwise is
sufficient!
Fig. 9 - 25 Correct maximum opening position: The O-ring seals the exhaust
pathway
When you open the venting screw for more than one turn the O-ring does not
seal anymore. The gases are additionally released via spaces between O-ring
and the thread of the venting screw. This may result in corrosion of the venting
screw and other parts of the seal holder.
Fig. 9 - 26 Incorrect opening position: The O-ring cannot seal the exhaust
pathway anymore
C49IB08A 224
9 Rotor 16MF100 / 16HF100
Warning:
Do not attempt to remove the rotor and vessels from the instrument before the
safety cooling has finished! Danger of hot and pressurized vessels!
Refer to chapter 3, "Safety Instructions".
1. Open the door of Synthos 3000 after the cooling has been finished. A pop-up
window will appear on the graphic display.
3. Slowly open each venting screw through the rotor lid with the supplied key
until you hear the noise of releasing gases. Do not open the venting screw
for more than 1 turn!
Supplied key
4. Close the door and wait a few minutes for the released gases to be drawn
off.
5. Switch off the exhaust unit, remove the rotor from the oven cavity and place
it into a fume hood.
6. Turn the bayonet lid of the rotor counterclockwise to the stop position.
7. Lift off the lid and place it with the top facing upwards on a clean surface. You
may use the holes in the lid as a rack for the liners and screw caps later on.
C49IB08A 225
9 Rotor 16MF100 / 16HF100
Warning:
Do not attempt to remove the rotor and vessels from the instrument before the
safety cooling has finished! Danger of hot and pressurized vessels!
Refer to chapter 3, "Safety Instructions".
1. Remove the rotor from the oven cavity and place it into a fume hood.
2. Slowly open each venting screw through the rotor lid with the supplied key
until you hear the noise of releasing gases. Do not open the venting screw
for more than 1 turn!
Alternatively, you can vent the vessels after removing the rotor lid.
Supplied key
3. Turn the bayonet lid of the rotor counterclockwise to the stop position.
4. Lift off the lid and place it with the top facing upwards on a clean surface. You
may use the holes in the lid as a rack for the liners and screw caps later on.
Hint:
If the reaction gases were not released via the venting system, the gases can
also escape safely via the venting grooves in the liner.
C49IB08A 226
9 Rotor 16MF100 / 16HF100
Warning:
Always open the vessel in a fume hood as noxious gases may be released.
Leakage may lead to chemical deposits on vessel or rotor parts.
3. Put the liners (with the screw caps) into an appropriate rack (like Rack 16) or
into the corresponding openings of the rotor lid. The openings are numbered
to avoid mixing up the samples.
4. If the liner is stuck in the vessel jacket, use the liner extractor bolt on the rotor
tray to push the liner out of the vessel jacket.
Liner
extractor bolt
Important:
The ceramic vessel jackets have to be handled with care. Avoid any
mechanical impact on these parts.
C49IB08A 227
9 Rotor 16MF100 / 16HF100
In this chapter you will find information on how to clean and inspect the instrument
and the vessels after every run, but also how to proceed after breakage of a safety
disk or vessel.
Additionally, you will find the description of an intensive cleaning procedure for
ceramic vessels, which should be performed on a regular basis.
At the end of this chapter, you will find appropriate maintenance procedures for
the p/T sensor.
Important:
Proper cleaning and maintenance of all instrument parts, especially reaction
vessels and rotors, is vital to extend the service life and to ensure proper
functioning of the instrument.
Disregarding the instructions given in this chapter will lead to premature
failure and cost-intensive repairs.
The most effective cleaning procedure for reaction vessels is to perform the
corresponding cleaning run in Synthos 3000.
1. Fill each vessel with at least 6 mL HNO3 (10%) or any cleaning solvent (ace-
tone, isopropanol).
2. Select a cleaning program from the method library stored in Synthos 3000.
Instructions for handling and preparing the vessels and rotor are given in
chapter 9.3.4, "Performing a Synthesis Run".
4. If there are vacant positions during a synthesis, these positions can be filled
up with “solvent blanks”.
C49IB08A 228
9 Rotor 16MF100 / 16HF100
Carefully clean the reaction vessels after every experiment to protect subsequent
samples against contamination, but also to avoid corrosion and overheating of the
vessel parts and rotor due to contamination with residues of reagents. This will
help to increase the service life of these parts.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
1. Rinse the liner thoroughly with deionized water and /or acetone.
2. Rinse the inner side of the seal thoroughly with deionized water and /or
acetone.
3. Remove any residues of reagents from the screw cap and if necessary rinse
the vessel jacket with deionized water and /or acetone. If required,
disassemble the screw cap completely (refer to chapter 9.4.4.2, "To
Disassemble the Screw Cap Assembly").
C49IB08A 229
9 Rotor 16MF100 / 16HF100
Hint:
You can alternatively clean the liners in a dishwasher.
6. Keep the venting screws open (max. 1 turn) during storage to increase the
service life of both the seal and the venting screw.
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Rinse the inner side of the seal thoroughly with deionized water and/or ace-
tone.
2. Carefully rinse the immersion tube with deionized water and/or acetone.
3. If necessary, carefully remove solid residues from the immersion tube with a
damp cloth. Do not use sharp tools! Stubborn residues are removed best
during a cleaning run.
C49IB08A 230
9 Rotor 16MF100 / 16HF100
6. Keep the venting screw of the p/T sensor open (max. 1 turn) during storage
to increase the service life of both the seal and the venting screw.
1. Wipe the interior of the microwave cavity and the safety door with a damp
cloth.
2. Clean the outer surfaces of Synthos 3000 with a damp cloth. Do not use
organic solvents.
Important:
If you detect any reagent residues at the bottom of the cavity, proceed with a
thorough inspection of your reaction vessels for signs of leakage.
Perform a visual check of the reaction vessels, rotor and instrument after every
reaction run.
Important:
Replace damaged parts immediately! If not detected, even slight damages
may develop into cost-intensive repairs.
C49IB08A 231
9 Rotor 16MF100 / 16HF100
3. Check the sealing lip of the lip-type seal for any sign of damage, like
deformation or cuts.
C49IB08A 232
9 Rotor 16MF100 / 16HF100
2. Check the sealing lip of the lip-type seal for any sign of damage, e.g.
deformation or cuts.
3. Check the p/T sensor for oil leakage of the hydraulic system.
An oil leakage requires repair at Anton Paar, contact our local distributor.
Fig. 9 - 35 Potential oil leakage from the hydraulic system on the p/T sensor
1. Check both the rotor body and the rotor tray for any obvious sign of damage,
e.g. melting of the rotor tray.
correct incorrect
C49IB08A 233
9 Rotor 16MF100 / 16HF100
Perform these combined cleaning and maintenance procedures after one or two
working weeks, depending on the operation frequency of Synthos 3000. If you
have to handle unknown sample material, perform these procedures daily.
Following these procedures will help to increase the service life of rotor and
vessel components, but will also reduce the risk of downtimes due to equipment
failure.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
3. Check the liner for any sign of damage, like deformations, cracks or
scratches of the sealing area.
Replace damaged liners immediately, as tightness during subsequent
experiment is not ensured anymore.
4. Perform a cleaning run to remove any detergent residues from the liner and
seal. Refer to chapter 9.4.1.1, "To Perform a Cleaning Run".
Alternatively: Steam the liner with nitric acid vapors overnight using a
steaming apparatus.
C49IB08A 234
9 Rotor 16MF100 / 16HF100
Important:
Do not store the liner in the vessel jacket. Acid residues may diffuse from the
liner material and lead to corrosion and subsequent damage of the vessel
jackets.
1. Thoroughly rinse the complete screw cap assembly under running water to
remove any reagent residues.
Venting screw
3. Use the reverse side of the venting screw to unscrew the syringe connector.
C49IB08A 235
9 Rotor 16MF100 / 16HF100
Syringe connector
4. Remove the seal and place it with the sealing lip facing upwards.
Important:
Do not mix up the seals and venting screws. Always keep the seal and its
venting screw in pairs.
C49IB08A 236
9 Rotor 16MF100 / 16HF100
1. Clean the screw cap with water, a mild detergent and a soft brush. Remove
any deposits, especially from the inner side of the screw cap and from the
thread for the venting screw.
2. Thoroughly clean the thread of the venting screw. Take care not to damage
the O-ring.
3. Carefully remove any deposits from the seal, then rinse thoroughly with
deionized water.
Important:
Avoid any damage to the lip-type seal.
4. Clean all other parts (syringe connector, safety disk and safety disk holder)
thoroughly under running water, then rinse with deionized water. You may
use a mild detergent and a soft brush if necessary.
C49IB08A 237
9 Rotor 16MF100 / 16HF100
Corroded thread
Crack
REPLACE!
Corrosion!
Replace!
Hint:
Before replacing the venting screw, cleaning the thread using a screw cap is
recommended. An appropriate screw tap is available on request.
C49IB08A 238
9 Rotor 16MF100 / 16HF100
Corroded thread
2. Check the thread of the venting screw. Replace the venting screw if the
thread is severely corroded or if it is not running smoothly.
1. Check the sealing lip for any sign of damages, like deformation or cuts.
A defective sealing lip does not ensure proper tightening of the reaction ves-
sels anymore. Replace damaged seals immediately to avoid cost-intensive
repairs.
Hint:
Avoid damage of the lip-type seal during storage by using the optional seal
forming device.
C49IB08A 239
9 Rotor 16MF100 / 16HF100
Sealing lip ok
Damaged seals
Hint:
When replacing the seal, also replace the venting screw as well, otherwise
tightness of the reaction vessels may not be ensured.
Do not use new seals with old venting screws!
C49IB08A 240
9 Rotor 16MF100 / 16HF100
1. Insert the safety disk into the screw cap with the label "M&H" facing out-
wards. Take care that the safety disk is positioned correctly. The holes in the
safety disk have to form a "C".
Important:
After the breakage of a safety disk, both safety disk and seal have to be
replaced. Also replace the venting screw. Do not use an old venting screw
together with a new seal to avoid leakiness.
C49IB08A 241
9 Rotor 16MF100 / 16HF100
Fig. 9 - 48 Fixing the safety disk with the safety disk holder
3. Insert the seal into the screw cap and fix it with the syringe connector and the
venting screw.
Venting screw
Syringe connector
C49IB08A 242
9 Rotor 16MF100 / 16HF100
1. Clean the vessel jacket with water, a mild detergent and a soft brush.
3. Let the vessel jacket dry completely before the next use.
4. Check the vessel jacket for any sign of damage, like small cracks or
chemical attack.
A yellowish discoloration does not necessarily affect proper functioning, but it
is a strong indication for attack by acid vapors and therefore requires a
thorough inspection. Replace PEEK vessel jackets immediately if you detect
cracks or severe material degradation.
1. Rinse the vessel jacket thoroughly with deionized water. Do not use any
detergents, as they might affect the heating behavior of these parts!
2. Let the vessel jacket dry completely before the next use.
3. Check the vessel jacket for any sign of damage, like cracks or chipping.
Discoloration or chipping are indications for improper usage and for
significantly decreased mechanical strength of the vessel jacket. Do not use
such a vessel jacket unless it has been tested by Anton Paar.
C49IB08A 243
9 Rotor 16MF100 / 16HF100
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Follow the instruction given in chapter 9.4.2.2, "To Clean the p/T Sensor
After Every Run" and chapter 9.4.3.2, "To Check the p/T Sensor After Every
Run".
• Follow the instructions given in chapter 9.4.2.3, "To Clean the Rotor After
Every Run" and chapter 9.4.3.3, "To Check the Rotor After Every Reaction".
To clean and check the rotor follow the instructions given in chapter 9.4.5.4, "To
Clean and Check the Microwave Cavity".
If a safety disk breaks during the experiment (e.g. due to spontaneous reactions),
reagents usually escape from the reaction vessels. This leads to contamination of
the rotor, the outer surfaces of the vessels and the microwave cavity.
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
C49IB08A 244
9 Rotor 16MF100 / 16HF100
3. Carefully remove the rotor, place it under a fume hood and vent the vessels
(refer to chapter 9.3.5.2, "To Vent Under a Fume Hood").
4. Remove the rotor lid. Take care for splinters and reagent residues.
5. Take out and open the reaction vessels (refer to chapter 9.3.5.3, "To Open
the Pressure Vessel"). It is most probable that the synthesis was incomplete,
so you may discard the solutions.
Important:
After the breakage of a safety disk, both safety disk and seal have to be
replaced.
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
(except for the ceramic vessel jackets) and a soft brush if necessary.
Hint:
After breakage of a safety disk the liner might be deformed due to the rapid
pressure loss.
If the sealing area (approx. 2 cm from the top) is not affected, such a
deformed liner can be reshaped by performing a cleaning run.
Replace the liner if the deformation has affected also the sealing area.
Deformation
C49IB08A 245
9 Rotor 16MF100 / 16HF100
If the safety disk of the p/T sensor breaks during the experiment, follow the
instructions given in chapter 9.5.1, "Replacing Seal and Safety Disk".
Important:
Do not wet the electronic module of the p/T sensor when cleaning the sensor.
1. Follow the instruction given in chapter 9.4.2.2, "To Clean the p/T Sensor
After Every Run" and chapter 9.4.3.2, "To Check the p/T Sensor After Every
Run".
1. Remove the turntable, the drive star and the drive base from the microwave
cavity.
2. Clean these parts under running water, use a mild detergent and a soft brush
if necessary.
3. Use a damp cloth to thoroughly clean the microwave cavity and the white
PTFE covers.
4. Clean the door in the same way, also the contacting surfaces between door
and instrument.
5. Clean the Fresnel lens of the IR sensor with a damp cotton bud.
C49IB08A 246
9 Rotor 16MF100 / 16HF100
6. Clean the sensors for the position recognition and for the protection lid
detection in the same way.
Sensor for the protection lid detection Sensor for the position recognition
8. Check the microwave cavity and door for signs of damage or deformation.
Warning:
Do not use the instrument if you detect any deformation on the microwave
cavity or the door. Danger of microwave leakage!
Contact an authorized service engineer.
9. Check all other parts for signs of damage, like cracks or fissures. Replace
damaged parts.
Although very unlikely, it may happen that a reaction vessel breaks during a run
instead of the safety disk. This may be caused by spontaneous exothermal
reactions, which are too fast for the safety disk, or by previously damaged vessel
jackets.
Synthos 3000 is equipped with numerous safety features, which ensure that the
laboratory personnel is safely protected in case of such an incident.
In case of a vessel breakage, solvent and reagents will be released together with
vessel debris inside the microwave cavity. This leads to contamination of the rotor,
the outer surfaces of the vessels as well as the microwave cavity.
C49IB08A 247
9 Rotor 16MF100 / 16HF100
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
Take care of splinters and reagent residues.
Warning:
After a vessel breakage you have to contact an authorized service engineer
who has to perform a microwave leakage test. Do not use the instrument
before it has passed this test successfully.
Due to the explosion either the door or the microwave cavity might have been
deformed. Danger of microwave radiation leakage!
3. Carefully remove the rotor. Take care of splinters, especially when using
ceramic vessels!
4. Place the rotor under a fume hood and vent the vessels (refer to chapter
9.3.5.2, "To Vent Under a Fume Hood").
5. Remove the rotor lid. Take care for splinters and reagent residues.
6. Take out and open the reaction vessels (refer to chapter 9.3.5.3, "To Open
the Pressure Vessel"). It is most probable that the synthesis was incomplete,
so you may discard the solutions.
C49IB08A 248
9 Rotor 16MF100 / 16HF100
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
(except for the ceramic vessel jackets) and a soft brush if necessary.
Hint:
If you are not sure if the ceramic vessel jackets (Rotor 16HF) have been
damaged during the explosion, you may send them to Anton Paar for
inspection.
1. Follow the instruction given in chapter 9.4.2.2, "To Clean the p/T Sensor
After Every Run" and chapter 9.4.3.2, "To Check the p/T Sensor After Every
Run".
To clean and check the microwave cavity, follow the instruction given in chapter
9.4.5.4, "To Clean and Check the Microwave Cavity".
To clean the exhaust unit, follow the instructions given in chapter 8.5.6.5, "To
Clean the Exhaust Channel".
C49IB08A 249
9 Rotor 16MF100 / 16HF100
1. After all vessels have been removed and cleaned, remove the rotor tray. Pay
attention to any splinters and reagent deposits.
2. Pull off the silicone seals from the lower end of the vessel holders.
Silicone seal
4. Remove the vessel holders if the rotor tray or vessel holders have been
damaged.
To do so, grip the vessel holder by the three tongues, which protrude from
the rotor tray and press them together. Push the vessel holder down and
remove it from the rotor tray.
C49IB08A 250
9 Rotor 16MF100 / 16HF100
Important:
Do not dismount the black rotor center ring at the base of the rotor body.
1. Clean all parts under running water. Use a mild detergent and a soft brush if
necessary.
2. If necessary wipe off any traces of non-digested sample material with a piece
of cloth moistened with isopropyl alcohol.
3. Let all parts dry completely. Take special care that the space between the
black center ring and the rotor body is dried completely. It is recommended to
use pressurized air for drying.
C49IB08A 251
9 Rotor 16MF100 / 16HF100
3. Assemble the positioning ring on the lower side of the rotor tray.
4. Plug the 16 silicone seals onto the base of the vessel holders and check for
proper installation. The silicone seal has to snap in completely. Defective
silicone seals can lead to incorrect IR measurement.
Silicone seal
C49IB08A 252
9 Rotor 16MF100 / 16HF100
Important:
If the reaction vessels are improperly aligned, IR measurement may provide
erroneous data.
correct incorrect
The essential part of this procedure consists of a heating step. After a few pre-
cleaning steps, the ceramic vessel jackets and liners are heated to 250 °C for
several hours to reduce the amount of contaminations and to restore the original
operation behavior.
C49IB08A 253
9 Rotor 16MF100 / 16HF100
1. Remove any visible impurities from the ceramic vessel jacket and the liner
with water, a mild detergent and a soft brush.
Hint:
Any sample residues or visible impurities, which are not removed thoroughly
before the heating step, will leave unsightly residues, which can be hardly
removed afterwards.
Alternatively: You may use a dishwasher for cleaning the liner, but not for
cleaning the ceramic vessel jacket.
3. If necessary, perform a cleaning run to remove any sample residues from the
liner. Take care that all parts have been dried completely before the run.
4. Place the ceramic vessel jacket and the liner into a drying furnace and heat
all parts for at least 4 hours at 250 °C.
Hint:
If you have to run this procedure at a reduced temperature (min. 200 °C),
extend the heating time (e.g. overnight). The efficiency of this procedure
decreases with lower temperature.
5. After the heating step is completed, let the vessel jacket and liner SLOWLY
cool down inside the oven (max. 100 °C/h).
Important:
Do not try to speed up the cooling process. This will cause severe damage to
the ceramic vessel jackets.
C49IB08A 254
9 Rotor 16MF100 / 16HF100
In the following chapters you will find information on how to replace the seal and
safety disk, capillary and immersion tube of the p/T sensor.
After rupture of the safety disk you have to replace seal, seal husk and safety disk.
Losses of reagents or visible mechanical damage also indicate the necessity of
replacement.
Important:
When you replace the seal or remove it for cleaning, always replace the seal
husk! When removing and inserting the seal husk, do not use sharp objects,
which could damage it.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents. Follow any special national safety regulations.
2. Use the reverse side of the venting screw as tool for unscrewing the syringe
connector.
C49IB08A 255
9 Rotor 16MF100 / 16HF100
Syringe connector
3. Insert a flat screwdriver into the aperture of the syringe connector and
carefully lever the seal holder with the seal out of the screw cap as far as
possible.
Important:
Be careful not to scratch the immersion tube!
Seal holder
4. Take the seal carefully without damaging its lip and slide the seal to the end
of the immersion tube. During this process do not tilt the seal to avoid
breakage of the immersion tube.
C49IB08A 256
9 Rotor 16MF100 / 16HF100
Seal holder
5. Check the threads of the seal holder (especially the thread for the venting
screw) for any sign of corrosion or cracks. A damaged seal holder has to be
replaced immediately!
6. Check the inner side of the screw cap for any sign of corrosion or cracks. A
damaged screw cap has to be replaced immediately!
7. Remove the seal husk, which is located at the middle of the seal.
Seal husk
1. Remove the seal holder as described in chapter 9.5.1.1, "To Remove the
Seal".
2. Screw out the safety disk holder with an appropriate screwdriver or a similar
tool.
C49IB08A 257
9 Rotor 16MF100 / 16HF100
4. Insert the safety disk into the screw cap with the label "M&H" facing
outwards. Take care that the safety disk is positioned correctly. The holes in
the safety disk have to form a "C".
Important:
After the breakage of a safety disk, both safety disk and seal have to be
replaced. Also replace the venting screw. Do not use an old venting screw
together with a new seal to avoid leakiness.
1. Mount the seal holder with the assembled safety disk and the seal onto the
immersion tube and push it into the screw cap to the stop position. Turn the
seal until the chamfered side fits to the inner side of the seal holder. Take
care not to damage the lip of the seal. It is recommended to use a liner to
push the seal into the seal holder.
2. Check that the openings for the syringe connector and venting screw
correspond to the openings in the screw cap. If necessary, turn the seal
holder until the boreholes face to the openings of the screw cap.
C49IB08A 258
9 Rotor 16MF100 / 16HF100
3. Fix the seal in the screw cap with the syringe connector and the venting
screw. Completely close the venting screw, then reopen it for one turn during
storage.
4. Push the new seal husk onto the immersion tube using the husk assembly
tool.
Important:
The two ends of the seal husk have different diameters. Put the new seal husk
onto the immersion tube so that the wider end faces the screw cap. To provide
good sealing properties, the dimensions of the seal husk are chosen to fit
tightly. Use the special husk assembly tool (plastic tube) to slide the seal husk
up to the final stop position.
Fig. 9 - 71 Pushing the seal husk onto the immersion tube using the husk
assembly tool
Important:
Never use sharp objects to push the seal husk onto the immersion tube.
Make sure that the seal is not damaged during assembly.
Important:
After replacing the seal, always carry out a performance run before the next
reaction to check the tightness of the seal. Weigh the solution before and after
the performance run and check for losses. If the losses are higher than 2%,
check the tightness of the seal and seal husk once again.
C49IB08A 259
9 Rotor 16MF100 / 16HF100
Important:
After replacing the thermometer capillary you have to calibrate the
temperature sensor! Refer to chapter 6.7.4, "Temperature Sensor
Calibration with the Calibration Accessory" or to chapter 6.7.5,
"Temperature Sensor Calibration with the Calibration Unit".
Carry out the steps described below only if you are able to perform a
temperature sensor calibration!
The capillary is very fragile. The most critical part is the connection between
the capillary and glass bulb, which can easily break.
If you are not completely sure how to replace the thermometer capillary
consult our local customer service.
2. Pull the knurled nut back (together with the PEEK sleeve of the capillary
tube) and take it out of the screw-type fitting.
knurled nut
3. The PEEK sleeve has to be separated from the immersion tube. Pull the
PEEK tube out of the metal sleeve which is positioned at the top of the
immersion tube. The use of a vice or pliers is recommended.
C49IB08A 260
9 Rotor 16MF100 / 16HF100
Fig. 9 - 73 Pulling the PEEK tube with capillary out of the metal sleeve
5. Insert the new capillary very carefully into the immersion tube.
Take care when inserting the glass bulb. This is a sensitive part!
6. Insert the end of the PEEK tube into the metal sleeve up to the mechanical
stop position (3 - 4mm) (with the help of a vice). The sleeve is automatically
fastened with a snap-in connector. The tube is automatically fastened with a
snap-in connector. Be very careful not to break the capillary!
C49IB08A 261
9 Rotor 16MF100 / 16HF100
7. Mount the new knurled nut (delivered with the new capillary) onto the metal
end of the capillary tube so that the metal end protrudes from the nut for
2 mm.
2 mm
Knurled nut
Fig. 9 - 76 Inserting the knurled nut
8. Place the p/T sensor on the edge of a table to keep one hand free for
manipulation of the capillary.
9. Pull back the knurled nut a bit and insert it into the fitting of the p/T sensor
unit without changing the protruding length of the metal end.
10. Close the knurled nut hand-tight and perform a temperature sensor
calibration with the calibration accessory (refer to chapter 6.7.4,
"Temperature Sensor Calibration with the Calibration Accessory") or with the
calibration unit (refer to chapter 6.7.5, "Temperature Sensor Calibration with
the Calibration Unit").
C49IB08A 262
9 Rotor 16MF100 / 16HF100
If the capillary is not damaged and should be used together with a new immersion
tube, take care not to break the capillary while disassembling.
Remove the capillary tube as described in Chapter 9.5.2.1, "To Replace the
Thermometer Capillary".
For replacing the immersion tube, follow the instructions below to ensure safe and
proper operation.
Important:
After replacing the immersion tube you have to calibrate the temperature
sensor!
When you replace the immersion tube, you also have to replace the seal
husk. When removing and inserting the seal husk, do not use sharp objects,
which could damage it.
C49IB08A 263
9 Rotor 16MF100 / 16HF100
3. Check the surfaces of the p/T sensor for splinters and contaminated parts
and clean the surface with a damp cloth, if necessary.
Important:
Never clean the p/T sensor under running water as this may damage the
sensor electronics.
4. Insert a new immersion tube into the p/T sensor and tighten the immersion
tube by turning it clockwise. Take care not to damage the sealing area of the
immersion tube.
sealing area
5. Install the new seal husk. Refer to chapter 9.5.1.3, "To Install the Seal".
6. Take care when inserting the capillary with the glass bulb through the metal
sleeve into the immersion tube. Refer to chapter 9.5.2.1, "To Replace the
Thermometer Capillary".
C49IB08A 264
10 Rotor 8SXF100 / 8SXQ80
The rotor upper plate made of aluminum contains a unique, patented hydraulic
pressure sensor for simultaneous measurement of all 8 vessels with wireless data
transmission to the oven.
The rotor can be closed easily by the fastening screws and a multifunctional
bayonet lid combined with a safety closure by a bolt. The lid serves as a protection
shield in case of a sudden overpressure release (e.g. rupture of a safety disk).
Additionally, the rotor lid serves as a rack for up to 8 vessels after the reaction.
The positions are numbered to avoid mixing up of samples.
Openings on the side of the rotor lid facilitate venting of the reaction vessels after
reaction directly inside Synthos 3000, without the need of handling pressurized
reaction vessels.
C49IB08A 265
10 Rotor 8SXF100 / 8SXQ80
Important:
Never attempt to change the rotor geometry or to bring other metal parts into
the microwave field. This will cause serious damage.
Rotor 8S lid
Fastening screw
assembled
Bottom guard
Fig. 10 - 2 Rotor 8S
C49IB08A 266
10 Rotor 8SXF100 / 8SXQ80
Important:
The quartz vessels and the ceramic vessel jackets have to be handled with
care. Avoid any mechanical impact on these parts, as this might raise the
probability of a vessel breakage.
XF100 XQ80
Liner Material Fluoropolymer ---------
Vessel Jacket Ceramic Quartz glass
Max. Operating Pressure 60 bar (870 psi) 80 bar (1160 psi)
Opening Pressure 120 bar (1740 psi) 120 bar (1740 psi)
a 260 °C 300 °C
Max. TINTERNAL
b 240 °C 280 °C
Max. TIR
Nominal Volume 100 mL 80 mL
Min. Filling Volume 6 mL 6 mL
cMax. Filling Volume 60 mL 60 mL
Typical Applications Parallel scale-up, prepressurized
prepressurized reactions, "Green
reactions Chemistry" approach,
near critical water
chemistry
a. Temperature inside the reference vessel, measured via the T-probe
b. Surface temperature at the base of the vessel, measured via the IR
sensor
c. The max. filling volume may exceed 60 mL for special methods after
consulting Anton Paar GmbH
C49IB08A 267
10 Rotor 8SXF100 / 8SXQ80
Important:
Do not fill in less than the minimum volume of 6 mL.
When using a volume of less than 6 mL, the amount of liquid phase in the
reaction vessel will be too low to absorb microwave energy sufficiently. As a
result, microwave energy will couple directly to parts of the vessel, the rotor or
the sample. This may cause damage to rotor components or pressure
vessels.
Protective cap
Syringe connector with
O-ring
Safety disk holder "X"
Venting screw
Safety disk "X"
Seal holder X
Liner fluoropolymer
Center ring
C49IB08A 268
10 Rotor 8SXF100 / 8SXQ80
Protective cap
Syringe connector with
O-ring
Safety disk holder "X"
Venting screw
Safety disk "X"
Seal holder X
Seal for XQ
Center ring
C49IB08A 269
10 Rotor 8SXF100 / 8SXQ80
Before you start your first experiment, it is recommended to get familiar with the
Synthos 3000 Microwave Reaction System. The best way is to start with either a
cleaning or a performance run from the method library.
If you are more familiar with microwave synthesis, you should also perform a
blank run to clean the liner and seal (refer to chapter 10.4.1.1, "To Perform a
Cleaning Run").
Both ways enable you to operate the instrument without the risk of spontaneous
reactions and to learn about:
Important:
Always take care to select the appropriate vessel type in the method.
The maximum operating pressure and temperature values are defined within
the method.
Do not use both types of pressure vessel (XF100 and XQ80) together!
Hint:
Before the first reaction remove the test certificates from the ceramic vessel
jackets or the quartz glass vessels, respectively. Keep these test certificates
with the other instrument documents.
Important:
Do not use deionized water due to the possible super-boiling effect, which
may result in a vessel explosion.
C49IB08A 270
10 Rotor 8SXF100 / 8SXQ80
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents. Follow any special national safety regulations.
Important:
Before the first experiment read chapter 11, "Applications" very carefully.
Guidelines for reagent mixtures are given in the method library of Synthos 3000.
Depending on a possibly different reaction behavior of your reagents, the
suggested mixture may have to be modified. The user is responsible for this
decision.
C49IB08A 271
10 Rotor 8SXF100 / 8SXQ80
2. Add solvent.
Rinse the liner / quartz glass vessel by adding the solvent to remove sample
material sticking on the top of the liner wall / quartz glass vessel wall. Make
sure that no deposits remain on the wall of the vessel.
The vessel is closed with a seal holder containing a self-sealing lip-type seal
made of PTFE-TFM. Due to the pressure build-up during the heating phase, the
sealing lip is pressed against the liner wall / quartz glass vessel wall, thus
providing tight closure of the vessel.
The seal is a sensitive part. To ensure proper functioning and to avoid losses of
reagents, the seal has to be handled carefully as described below.
Important:
Take care when handling and storing the seal! The sealing lip can be
damaged easily.
Replace damaged seals immediately, as tightness of the vessel is not
ensured any longer. As a consequence, damage to the instrument, rotor and
vessels may occur due to escaping solvents.
Never store the seal standing on its sealing lip!
Starting a run without properly expanded seals may damage the seals.
C49IB08A 272
10 Rotor 8SXF100 / 8SXQ80
incorrect
incorrect correct
correct
The sealing lip has to be expanded shortly before use. To do so, you can use
either the seal forming tool or the optional seal forming device.
The seal forming device is used for the simultaneous expansion of 8 seals and for
safe storage. The sensitive sealing lip is protected and the working life of the seal
is extended. The design of the seal forming device allows you to expand and store
all types of seals for Rotor 16, Rotor 16 SOLV and Rotor 8S.
1. Press the seal holder with seal completely onto the seal forming tool for at
least 3 seconds to expand the sealing lip. Expand new seals for at least 10
seconds.
C49IB08A 273
10 Rotor 8SXF100 / 8SXQ80
2. Immediately close the pressure vessel (refer to chapter 10.3.2.3, "To Close
the Pressure Vessel").
Note that the seal will regain its original shape after some time, therefore you
have to start the reaction within approx. 15 minutes. Otherwise, re-expand
the seal.
3. If applicable: proceed with the immersion tube in the same way (see also
chapter 10.3.3.2, "To Apply the T-Probe").
Fig. 10 - 9 Expanding the seal with the immersion tube (XF100 / XQ80)
Important:
Make sure not to damage the immersion tube when handling the seal forming
tool.
1. Ensure that the cones are mounted properly on both the intermediate panel
and the upper panel, depending on the seal type.
To expand XF seals, the longer, smaller cone must protrude upwards from
the panel. To expand XQ seals, the shorter, bigger cone must protrude
upwards from the panel.
C49IB08A 274
10 Rotor 8SXF100 / 8SXQ80
2. Place four seal holders with seal into the recesses of the lower panel with the
seal facing upwards.
Fig. 10 - 11 Placing the first four seal holders into the seal forming device
Hint:
After a reaction, rinse the seals thoroughly with acetone and allow them to dry
before placing them into the seal forming device.
3. Take the intermediate panel and carefully place it onto the seals.
Important:
The cones have to fit exactly into the seals. Take care not to place the cones
onto the sealing lips, as this will damage the seals.
Check for correct positioning of the seals.
C49IB08A 275
10 Rotor 8SXF100 / 8SXQ80
4. Place another four seal holders into the recesses of the intermediate panel.
Fig. 10 - 13 Placing the next four seal holders into the seal forming device
This side up
C49IB08A 276
10 Rotor 8SXF100 / 8SXQ80
This side up
Correct Incorrect
Important:
The seal XQ is expanded too much when closing the seal forming device
hand tight. For this reason it is necessary to tighten the handwheel clockwise
until you feel resistance but then reopen it approximately 180° (1/2 turn)
otherwise the XQ seals will be deformed and damaged.
C49IB08A 277
10 Rotor 8SXF100 / 8SXQ80
1. Remove any droplets from the vessel jacket and the outer surface of the liner
or from the quartz glass vessel.
Important:
Take care that all vessel parts (liner, vessel jackets, quartz vessels, seal
holder assemblies) are completely dry. Any residues of liquid between the
liner and vessel jacket may cause damage to the vessel.
Important:
Close the venting screw finger-tight only without using a tool.
The supplied key is used for opening only. The tight closure during reaction is
ensured via the front side of the white PTFE-plug of the venting screw.
Fig. 10 - 18 Venting screw in closed position: The front side of the PTFE-plug
seals the outlet opening
3. Close the liner or the quartz glass vessel with the seal holder.
Make sure that the seal of the seal holder has been expanded shortly before
and is inserted exactly into the liner or the quartz glass vessel.
C49IB08A 278
10 Rotor 8SXF100 / 8SXQ80
4. XF: Put the liner with the seal holder into the vessel jacket and place it into
the protective casing so that the venting screw rests on the strap of the
protective casing.
XQ: Place the quartz glass vessel with the seal holder into the protective
casing so that the venting screw rests on the strap of the protective casing.
5. Put the protective cap onto the protective casing. The distance between the
cap and the casing should not exceed 2 mm.
Seal holder
Liner
Ceramic vessel
Protective casing
Distance between
cap and vessel holder
less than 2 mm
1. Remove any droplets from the vessel jacket and the outer surface of the liner
or from the quartz glass vessel. Liquid residues between liner and vessel
jacket or on the quartz glass vessel may cause overheating and damage to
the reaction vessel!
Important:
Close the venting screw finger-tight only without using a tool.
The supplied key is used for opening only. The tight closure during reaction is
ensured via the front side of the white PTFE-plug of the venting screw.
C49IB08A 279
10 Rotor 8SXF100 / 8SXQ80
3. Close the liner or the quartz glass vessel with the immersion tube with seal.
Make sure that the seal of the immersion tube has been expanded shortly
before and is inserted exactly into the liner or the quartz glass vessel.
4. XF: Put the liner with the seal holder into the vessel jacket and place it into
the protective casing so that the venting screw rests on the strap of the
protective casing.
XQ: Place the quartz glass vessel into the protective casing so that the
venting screw rests on the strap of the protective casing.
5. Put the protective cap onto the protective casing. The distance between the
cap and the casing should not exceed 2 mm.
C49IB08A 280
10 Rotor 8SXF100 / 8SXQ80
The following procedures describe how to load the reaction vessels into the rotor,
the use of the T-probe and how to close and insert the rotor into Synthos 3000.
1. Put the vessels into the assigned rotor positions with the venting screws fac-
ing outwards. The vessels have to engage into position.
Important:
Take care that the vessels are positioned properly, otherwise there could
arise serious damages to the vessels and rotor during reaction. Additionally
incorrect positioning may lead to erroneous pressure values.
Important:
Load Rotor 8S with 4 or 8 vessels only for proper functioning.
To ensure accurate pressure measurement during the process the rotor has
to be loaded symmetrically.
Reduce the max. power in the method when using less than 8 vessels. As a
guideline, limit the max. power output to 800 W for 4 vessels.
If you have less than 4 or 8 samples to be processed, fill the vacant positions
with blank solvent.
C49IB08A 281
10 Rotor 8SXF100 / 8SXQ80
1400 W 800 W
glass rod
3. To close the rotor, screw down the rotor upper plate (with the glass rod, if
applicable) until the vessels are fixed.
Important:
The initial pressure of Rotor 8S has to be between 2 and 10 bar. If the
pressure is out of this range reassamble the rotor.
C49IB08A 282
10 Rotor 8SXF100 / 8SXQ80
Correct! Incorrect!
1. Before inserting the T-probe into the reference vessel, use the glass rod to
make sure that the vessel is positioned correctly. If there is any resistance
when inserting the rod, adjust the vessel positioning and try it again. Repeat
the procedure until you feel no resistance when inserting the rod.
Important:
Do not introduce the T-probe until the above test was successful, otherwise
the capillary of the T-probe may break when inserted into the vessel.
3. Insert the T-probe very carefully through the hole of the upper plate of the
rotor into the vessel at position 1.
C49IB08A 283
10 Rotor 8SXF100 / 8SXQ80
Important:
The torus-shaped part of the T-probe is placed on the central rotor sensor
module.
The arm of the T-probe has to fit exactly into the recesses of the fastening
screws of the rotor upper plate.
Never insert the T-probe unless the rotor top plate is correctly fixed. The glass
bulb and the capillary may break!
1. Put the rotor lid on the rotor so that the top of the fastening screws protrude
through the holes in the lid.
2. Close the rotor by turning the lid clockwise until you reach the stop position.
C49IB08A 284
10 Rotor 8SXF100 / 8SXQ80
Important:
The locking button near the “Warning” symbol must engage.
"Warning" symbol
Locking button
After the rotor and vessels have been prepared, the loaded rotor is put into the
microwave cavity. The run is started from the software menu. Synthos 3000
controls all parameters automatically and usually does not require user
intervention.
At the end of a run, the cooling starts automatically. The safety cooling is also
activated automatically if a run is aborted. The cooling is stopped as soon as all
vessels have cooled down to an internal temperature below 60 °C and an IR
temperature below 50 °C, depending on the installed sensors.
Warning:
Do not attempt to remove the rotor and vessels from the instrument before the
safety cooling has finished! Danger of hot and pressurized vessels!
Refer to chapter 3, "Safety Instructions".
C49IB08A 285
10 Rotor 8SXF100 / 8SXQ80
5. Start the program by pressing the green <Start> button after the system
check was successful. During the run the display informs you about the
reaction progress.
1. To interrupt (pause) the program, press the <Stop> button once to shut
down the microwave energy.
The program can be aborted at any time, pressure and temperature data
acquisition will be continued. To continue with the run, press the <Start> but-
ton.
1. After the run has been finished, select from the menu on the display one of
the following options:
• Print: Print out a summary of the last run, including pressure and
temperature graphs, on a connected printer.
C49IB08A 286
10 Rotor 8SXF100 / 8SXQ80
• Next Run: Return to the sample menu to modify and/or restart the
current method.
2. Take the rotor out of the cavity and unload it as soon as practicable.
After the cooling has finished either vent the pressure vessels inside the
instrument or remove the rotor and vent the vessels under a fume hood.
Warning:
Take safety precautions: Use goggles, protective gloves and lab coat.
Depending on the application, noxious gases may be released during the
venting procedure.
C49IB08A 287
10 Rotor 8SXF100 / 8SXQ80
Important:
You must not open the venting screw completely, one turn counter clockwise is
sufficient!
Fig. 10 - 31 Correct maximum opening position: The O-ring seals the exhaust
pathway
When you open the venting screw for more than one turn the O-ring does not
seal anymore. The gases are additionally released via spaces between O-ring
and the thread of the venting screw. This may result in corrosion of the venting
screw and other parts of the seal holder.
Fig. 10 - 32 Incorrect opening position: The O-ring cannot seal the exhaust
pathway anymore
C49IB08A 288
10 Rotor 8SXF100 / 8SXQ80
1. Open the door of Synthos 3000 after the cooling has been finished. A pop-up
window will appear on the graphic display.
3. Slowly open each venting screw through the rotor lid with the supplied key
until you hear the noise of releasing gases. Do not open the venting screw
for more than 1 turn!
4. Close the door and wait a few minutes for the released gases to be drawn
off.
5. Switch off the exhaust unit, remove the rotor from the oven cavity and place
it into a fume hood.
6. Turn the bayonet lid of the rotor counterclockwise to the stop position.
7. Lift off the lid and place it with the top facing upwards on a clean surface. You
may use the holes in the lid as a rack for the liners and seal holders.
Warning:
Take care of hot and pressurized vessels after removing the rotor from the
oven cavity!
1. Remove the rotor from the oven cavity and place it into a fume hood.
2. Slowly open each venting screw through the rotor lid with the supplied key
until you hear the noise of releasing gases. Do not open the venting screw
for more than 1 turn!
Alternatively, you can vent the vessels after removing the rotor lid.
C49IB08A 289
10 Rotor 8SXF100 / 8SXQ80
3. Press the locking button and turn the bayonet lid of the rotor
counterclockwise to the stop position.
4. Carefully lift off the lid and place it with the top facing upwards on a clean
surface. You may use the holes in the lid as a rack for the liners and seal
holders later on.
1. If a T-probe was used, carefully remove it before opening the rotor fastening
screws. Hold the T-probe by the sensor head and sensor arm and pull the
unit carefully straight upwards out of the vessel.
Important:
Do not take out the reference vessel until the T-probe has been removed.
Otherwise the capillary of the T-probe will be damaged.
C49IB08A 290
10 Rotor 8SXF100 / 8SXQ80
Hint:
If the reaction gases were not released via the venting system, the gases can
also escape safely via the venting grooves in the liner of XF vessels.
Warning:
Always open the vessels in a fume hood as noxious gases may be released.
2. Open the rotor fastening screws slowly until all vessels can be removed
easily!
3. Take the vessels out of the rotor and place them nearby.
C49IB08A 291
10 Rotor 8SXF100 / 8SXQ80
8. Put the liners (with the seal holder) into an appropriate rack (like Rack 16) or
into the corresponding openings of the rotor lid. The openings are numbered
to avoid mixing up the samples.
C49IB08A 292
10 Rotor 8SXF100 / 8SXQ80
Important:
Both the ceramic vessel jackets and the quartz glass vessels have to be
handled with care. Avoid any mechanical impact on these parts, like hitting
them together.
Inappropriate storage of ceramic and quartz vessels, like putting them loosely
into a drawer, will drastically decrease the service life of these parts and
increase the probability of vessel breakage.
C49IB08A 293
10 Rotor 8SXF100 / 8SXQ80
In this chapter you will find information on how to clean and inspect the instrument
and the vessels after every run, but also how to proceed after breakage of a safety
disk or vessel.
Additionally, you will find the description of an intensive cleaning procedure for
ceramic vessels, which should be be performed on a regular basis.
At the end of this chapter, you will find appropriate maintenance procedures for
the T-probe and the immersion tube with seal.
Important:
Proper cleaning and maintenance of all instrument parts, especially reaction
vessels and rotors, is vital to extend the service life and to ensure proper
functioning of the instrument.
Disregarding the instructions given in this chapter will lead to premature
failure and cost-intensive repairs.
Important:
Only clean the quartz glass vessels with a soft brush, never use metal parts or
tools, respectively. Try to avoid using mechanical force as this could damage
the vessels by causing scratches. Never use damaged or scratched vessels
for reactions, as the pressure resistance is not longer guaranteed.
The most effective cleaning procedure for reaction vessels is to perform a blank
run in Synthos 3000.
1. Fill each vessel with at least 6 mL HNO3 (10%) or any cleaning solvent (ace-
tone, isopropanol etc.).
2. Select a cleaning program from the method library stored in Synthos 3000.
Refer to chapter 6.4, "Library Menu".
Instructions for handling and preparing the vessels and rotor are given in
chapter 10.3.4, "Performing a Synthesis Run".
4. If there are vacant positions during a run, these positions can be filled up
with blank solvent.
C49IB08A 294
10 Rotor 8SXF100 / 8SXQ80
Carefully clean the reaction vessels after every experiment to protect subsequent
samples against contamination, but also to avoid corrosion and overheating of the
vessel parts and rotor due to contamination with residues reagents. This will help
to increase the service life of these parts.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
1. Rinse both liner and quartz glass vessel thoroughly with deionized water
and/or acetone.
2. Rinse the inner side of the seal thoroughly with deionized water and/or
acetone.
3. Remove any residues of reagents from the seal holder and the vessel jacket
with deionized water. If necessary, disassemble the seal holder completely
(refer to chapter 10.4.4.4, "To Disassemble the Seal Holder Assembly").
C49IB08A 295
10 Rotor 8SXF100 / 8SXQ80
Hint:
You can alternatively clean the liners in a dishwasher.
6. Keep the venting screws open (max. 1 turn) during storage to increase the
service life of both the seal and the venting screw.
10.4.2.2 To Clean the Immersion Tube with Seal After Every Run
1. Rinse the inner side of the seal thoroughly with deionized water and/or ace-
tone.
2. Carefully rinse the immersion tube with deionized water and/or acetone.
C49IB08A 296
10 Rotor 8SXF100 / 8SXQ80
3. If necessary, carefully remove solid residues from the immersion tube with a
damp cloth. Do not use sharp tools! Stubborn residues are removed best
during a cleaning run.
Important:
If the immersion tube is removed from the seal by mistake, the seal husk has
to be replaced.
1. Remove visible contaminations (e.g reagent spills) with a damp cloth. Never
rinse Rotor 8S with running water. Check the inner side of the rotor lid for
reagent residues, which might be due to the venting procedure.
1. Wipe the interior of the microwave cavity and the safety door with a damp
cloth.
2. Clean the outer surfaces of Synthos 3000 with a damp cloth. Do not use
organic solvents!
Important:
If you detect any reagent residues at the bottom of the cavity, proceed with a
thorough inspection of your reaction vessels for signs of leakage.
Perform a visual check of the reaction vessels, rotor and instrument after every
experiment.
Important:
Replace damaged parts immediately! If not detected, even slight damages
may develop into cost-intensive repairs.
C49IB08A 297
10 Rotor 8SXF100 / 8SXQ80
2. Check the quartz glass vessel for any sign of damage, like cracks, scratches
or blurred surfaces. Check for any sign of devitrification. Replace damaged
parts, alternatively send them to our local customer service for inspection.
Chipped-off quartz
glass vessel
Replace
immediately!
4. Check the sealing lip of the lip-type seal for any sign of damage, like
deformation or cuts.
C49IB08A 298
10 Rotor 8SXF100 / 8SXQ80
10.4.3.2 To Check the Immersion Tube with Seal After Every Run
2. Check the sealing lip of the lip-type seal for any sign of damage, e.g.
deformation or cuts.
3. Check the seal husk for any sign of damage and for correct positioning.
4. Proceed with chapter 10.5, "Maintenance of the T-Probe and the Immersion
Tube with Seal XF / XQ" if the T-probe or immersion tube shows any
damages.
1. Check both the rotor and the rotor lid for any obvious sign of damage.
C49IB08A 299
10 Rotor 8SXF100 / 8SXQ80
Perform these combined cleaning and maintenance procedures after one or two
working weeks, depending on the operation frequency of Synthos 3000. If you
have to handle unknown sample material, perform these procedures daily.
Following these procedures will help to increase the service life of rotor and
vessel components, but will also reduce the risk of downtimes due to equipment
failure.
Important:
Always let the parts dry completely before the next use. Do not use wet parts.
The evaporating water may lead to deformation of parts or may initiate local
overheating.
If you use a drying oven, do not exceed 80 °C.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
while cleaning the rotor, vessels and instrument.
3. Check the liner for any sign of damage, like deformations, cracks or
scratches of the sealing area.
5. Perform a cleaning run to remove any detergent residues from the liner and
the seal. Refer to chapter 10.4.1.1, "To Perform a Cleaning Run".
Alternatively: Steam the liner with nitric acid vapors overnight using a
steaming apparatus.
C49IB08A 300
10 Rotor 8SXF100 / 8SXQ80
Hint:
Do not store the liner in the vessel jacket. Acid residues may diffuse from the
liner material and lead to corrosion and subsequent damage of the vessel
jackets
2. Rinse the vessel jacket thoroughly with deionized water. Do not use any
detergents, as they might affect the heating behavior of these parts!
3. Let the vessel jacket dry completely before the next use.
5. Check the vessel jacket for any sign of damage, like cracks or chipping.
Discoloration or chipping are indications for improper usage and for
significantly decreased mechanical strength of the vessel jacket. Do not use
such a vessel jacket unless it has been tested by Anton Paar.
1. To remove stubborn residues on the quartz glass vessel, fill some 2% hydrof-
luoric acid into the vessel and place it into an ultrasonic bath for approx. 10-
15 minutes. Do not exceed the maximum time of 15 minutes as the hydroflu-
oric acid may attack the quartz glass.
3. Check the quartz glass vessel for any sign of damage, like deformations,
cracks or scratches and check for any sign of devitrification.
C49IB08A 301
10 Rotor 8SXF100 / 8SXQ80
Chipped-off
quartz glass
vessel
Replace
immediately!
5. Perform a cleaning run to remove any detergent residues from the quartz
glass vessel and the seal. Refer to chapter 10.4.1.1, "To Perform a Cleaning
Run".
Alternatively: Steam the quartz glass vessel with nitric acid vapors overnight
using a steaming apparatus.
1. Thoroughly rinse the complete seal holder assembly under running water to
remove any reagent residues.
Venting screw
3. Use the reverse side of the venting screw to unscrew the syringe connector.
C49IB08A 302
10 Rotor 8SXF100 / 8SXQ80
Syringe connector
4. Remove the seal and place it with the sealing lip facing upwards.
Important:
Do not mix up the seals and venting screws. Always keep the seal and its
venting screw in pairs.
If applicable: Carefully pull out the immersion tube from the seal and remove the
seal husk from the immersion tube.
C49IB08A 303
10 Rotor 8SXF100 / 8SXQ80
Seal husk
Safety disk
1. Clean the seal holder with water, a mild detergent and a soft brush. Remove
any deposits, especially from the inner side of the seal holder and from the
thread for the venting screw.
2. Thoroughly clean the thread of the venting screw. Take care not to damage
the O-ring.
3. Carefully remove any deposits from the seal, then rinse thoroughly with
deionized water and/or acteone.
C49IB08A 304
10 Rotor 8SXF100 / 8SXQ80
4. Clean all other parts (syringe connector, safety disk and safety disk holder)
thoroughly with deionized water and/or acetone. You may use a mild
detergent and a soft brush if necessary.
Crack
C49IB08A 305
10 Rotor 8SXF100 / 8SXQ80
Corroded thread
Hint:
Before replacing the venting screw, cleaning the thread using a screw tap is
recommended. An appropriate screw tap is available on request.
Damaged O-ring
Corroded thread
C49IB08A 306
10 Rotor 8SXF100 / 8SXQ80
1. Check the sealing lip for any sign of damages, like deformation or cuts.
A defective sealing lip does not ensure proper tightening of the reaction ves-
sels anymore. Replace damaged seals immediately to avoid cost-intensive
repairs due to escaping acid vapors during subsequent runs.
Hint:
Avoid damage of the lip-type seal during storage by using the optional seal
forming device.
Sealing lip ok
Hint:
When replacing the seal, replace the venting screw as well, otherwise
tightness of the reaction vessels may not be ensured.
Do not use new seals with old venting screws!
C49IB08A 307
10 Rotor 8SXF100 / 8SXQ80
1. Insert the safety disk into the seal holder with the label "X" facing outwards.
Take care that the safety disk is positioned correctly. The holes in the safety
disk have to form a "C".
Important:
After the breakage of a safety disk, both safety disk and seal have to be
replaced. Also replace the venting screw. Do not use an old venting screw
together with a new seal to avoid leakiness.
C49IB08A 308
10 Rotor 8SXF100 / 8SXQ80
Fig. 10 - 61 Fixing the safety disk with the safety disk holder
3. Insert the seal into the seal holder and fix it with the syringe connector and
the venting screw.
1. Before inserting the seal in the seal holder, take the immersion tube and
push it into the new lip-type seal from the top downwards until approx. 5 mm
of the immersion tube sticks out of the top of the seal. Take care not to dam-
age the sealing lip.
5 mm
2. Fix the seal by screwing in the venting screw and the syringe connector.
Reopen the venting screw for one turn during storage.
C49IB08A 309
10 Rotor 8SXF100 / 8SXQ80
3. Push a new seal husk provided with the seal onto the immersion tube.
Important:
The two ends of the seal husk have different diameters. Put the new seal husk
onto the immersion tube so that the wider end faces the seal holder. To
provide good sealing properties, the dimensions of the seal husk are chosen
to fit tightly. Use the special husk assembly too to slide the seal husk up to the
final stop position.
Seal husk
Husk
assembly
tool
Fig. 10 - 64 Pushing the seal husk onto the immersion tube using the husk
assembly tool
4. Fix the immersion tube while carefully removing the husk assembly tool.
Important:
After replacing the seal, always carry out a performance run before the next
experiment to check the tightness of the seal. Weigh the solution before and
after the performance run and check for losses. If the losses are higher than
2%, check the tightness of the seal and seal husk once again.
2. Remove all fastening screws and remove the rotor upper plate.
C49IB08A 310
10 Rotor 8SXF100 / 8SXQ80
Important:
Do not operate Rotor 8S with deformed or without bushings!
4. Remove deposits from the PEEK bushings either with a soft cloth, moistened
with acetone, under running water or in an ultrasonic bath.
5. Remove any metallic residue from the fastening screw. The best way is to
knock the fastening screw with the thread facing down against a hard
surface (e.g. table).
8. Remove the 8 silicon protection hoses from the tension bars and clean them
under running water with a mild detergent.
9. Wipe the tension bars and the rotor bottom plate with a damp cloth. Remove
any metallic residues, which may come from the fastening screws.
C49IB08A 311
10 Rotor 8SXF100 / 8SXQ80
1. Check both the rotor and the rotor lid for any obvious sign of damage.
3. Check the correct position of the position marker labels underneath the
bottom load ring.
To clean and check the microwave cavity, follow the instructions given in chapter
10.4.5.5, "To Clean and Check the Microwave Cavity".
If a safety disk breaks during the experiment (e.g. due to spontaneous reactions),
reagents usually escape from the reaction vessels. This leads to contamination of
the rotor, the outer surfaces of the vessels and the microwave cavity.
For this reason, extensive cleaning and checking procedures are required.
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
C49IB08A 312
10 Rotor 8SXF100 / 8SXQ80
3. Carefully remove the rotor, place it under a fume hood and vent the vessels
(refer to chapter 10.3.5.2, "To Vent Under a Fume Hood").
4. Remove the rotor lid. Take care for splinters and reagent residues.
6. Open the rotor and remove the reaction vessels (refer to chapter 10.3.5.4,
"To Open the Pressure Vessel"). It is most probable that the synthesis was
incomplete, so you may discard the solutions.
Important:
After the breakage of a safety disk, both safety disk and seal have to be
replaced.
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
(except for the ceramic vessel jackets) and a soft brush if necessary.
Hint:
After breakage of a safety disk the liner might be deformed due to the rapid
pressure loss.
If the sealing area (approx. 2 cm from the top) is not affected, such a
deformed liner can be reshaped by performing a cleaning run.
Replace the liner if the deformation has affected also the sealing area.
C49IB08A 313
10 Rotor 8SXF100 / 8SXQ80
Deformation
10.4.5.3 To Clean and Check the Immersion Tube with Seal and the T-
Probe
1. Follow the instruction given in chapter 10.4.3.2, "To Check the Immersion
Tube with Seal After Every Run" and chapter 1.7, "Immersion Tube with Seal
XF/XQ-Maintenance".
Important:
Do not wet the electronic module of the T-probe when cleaning the sensor.
Do not clean the capillary, as this part might break easily.
• Refer to chapter 10.4.4.12, "To Clean Rotor 8S" and chapter 10.4.4.13, "To
Check the Rotor".
1. Remove the turntable, the drive star and the drive base from the microwave
cavity.
C49IB08A 314
10 Rotor 8SXF100 / 8SXQ80
2. Clean these parts under running water, use a mild detergent and a soft brush
if necessary.
3. Use a damp cloth to thoroughly clean the microwave cavity and the white
PTFE covers.
4. Clean the door in the same way, also the contacting surfaces between door
and instrument.
5. Clean the Fresnel lens of the IR sensor with a damp cotton bud.
6. Clean the sensors for the position recognition and for the protection lid
detection in the same way.
Sensor for the protection lid detection Sensor for the position recognition
8. Check the microwave cavity and door for signs of damage or deformation.
Warning:
Do not use the instrument if you detect any deformation on the microwave
cavity or the door. Danger of microwave leakage!
Contact an authorized service engineer.
9. Check all other parts for signs of damage, like cracks or fissures. Replace
damaged parts.
C49IB08A 315
10 Rotor 8SXF100 / 8SXQ80
Although very unlikely, it may happen that a reaction vessel breaks during a run
instead of the safety disk. This may be caused by spontaneous exothermal
reactions, which are too fast for the safety disk or by previously damaged vessel
jackets.
Synthos 3000 is equipped with numerous safety features, which ensure that the
laboratory personnel is safely protected in case of such an incident.
For this reason, extensive cleaning and checking procedures are required.
Warning:
Let the safety cooling run for at least 60 minutes before removing the rotor
from the instrument. Refer to chapter 3, "Safety Instructions".
Always wear goggles, protective gloves and appropriate protective clothing
while removing the rotor and cleaning the instrument.
Warning:
After a vessel breakage you have to contact an authorized service engineer
who has to perform a microwave leakage test. Do not use the instrument
before it has passed this test successfully.
Due to the explosion either the door or the microwave cavity might have been
deformed. Danger of microwave radiation leakage!
C49IB08A 316
10 Rotor 8SXF100 / 8SXQ80
3. Carefully remove the rotor. Take care of splinters, especially when using
ceramic vessels!
4. Place the rotor under a fume hood and vent the vessels (refer to chapter
10.3.5.2, "To Vent Under a Fume Hood").
5. Remove the rotor lid. Take care for splinters and reagent residues.
7. Open the rotor and remove the reaction vessels (refer to chapter 10.3.5.4,
"To Open the Pressure Vessel"). It is most probable that the synthesis was
incomplete, so you may discard the solutions.
1. Clean all vessel parts thoroughly under running water. Use a mild detergent
(except for the ceramic vessel jackets) and a soft brush if necessary.
Hint:
If you are not sure if the ceramic vessel jackets have been damaged during
the explosion, you may send them to Anton Paar for inspection.
10.4.6.3 To Clean and Check the Immersion Tube with Seal and the T-
Probe
1. Follow the instruction given in chapter 10.4.3.2, "To Check the Immersion
Tube with Seal After Every Run" and chapter 10.5, "Maintenance of the T-
Probe and the Immersion Tube with Seal XF / XQ".
C49IB08A 317
10 Rotor 8SXF100 / 8SXQ80
Important:
Do not wet the electronic module of the T-probe when cleaning the sensor.
Do not clean the capillary, as this part might break easily.
To clean and check the microwave cavity, follow the instructions given in chapter
10.4.5.5, "To Clean and Check the Microwave Cavity".
To clean the exhaust channel, follow the instructions given in chapter 8.5.6.5, "To
Clean the Exhaust Channel".
1. Refer to chapter 10.4.4.12, "To Clean Rotor 8S" and chapter 10.4.4.13, "To
Check the Rotor".
2. In case of heavy contamination, also disassemble and clean the rotor bottom
plate. Proceed with chapter 10.4.6.7, "To Disassemble the Rotor".
1. After all vessels have been removed and cleaned, remove the fastening
screws and lift off the rotor upper plate. Do not disassemble the rotor upper
plate. In case of a hydraulic oil leakage, the rotor has to be sent to
Anton Paar GmbH for service.
C49IB08A 318
10 Rotor 8SXF100 / 8SXQ80
Fig. 10 - 72 Removing the fastening screws and lifting off the rotor upper plate
3. Unscrew the bottom guard from the bottom load ring (with an Allen screw
driver, 4 mm).
C49IB08A 319
10 Rotor 8SXF100 / 8SXQ80
Bottom guard
Base ring
1. Clean all parts under running water or use a damp cloth. Use a mild deter-
gent and a soft brush if necessary.
2. Remove splinters.
3. Wipe off any traces of non-decomposed sample material with a piece of cloth
moistened with isopropyl alcohol.
5. Check the correct position of the position marker labels on the bottom side of
the rotor.
C49IB08A 320
10 Rotor 8SXF100 / 8SXQ80
Position 1
2. Start the reassembling by putting the clean base rings onto the bottom load
ring.
Fig. 10 - 77 Putting the base rings onto the bottom load ring
3. Put the bottom guard onto the bottom load ring and fix it with the screws.
Take care of the right positioning. Position 1 of the bottom guard has to
correspond with the position 1 of the bottom load ring (marked with the larger
position marker label).
C49IB08A 321
10 Rotor 8SXF100 / 8SXQ80
Bottom guard
5. Put the rotor upper plate on the tension bars and fix it with the fastening
screws. Take care of the right position of the rotor upper plate. The white
opening in the upper plate has to correspond with the position 1 in the
bottom guard.
C49IB08A 322
10 Rotor 8SXF100 / 8SXQ80
Position 1
Fig. 10 - 80 Putting the rotor upper plate on the tension bars and fixing it with the
rotor fastening screws
For proper maintenance of the ceramic vessels refer to chapter 9.4.7, "Intensive
Cleaning of Ceramic Vessels".
Important:
This cleaning procedure is intended only for users of ceramic vessel jackets
(8SXF100) and does not apply to quartz glass vessels!
C49IB08A 323
10 Rotor 8SXF100 / 8SXQ80
After rupture of the safety disk you have to replace seal, seal husk and safety disk.
Losses of reagents or visible mechanical damage also indicate the necessity of
replacement.
Important:
When you replace the seal or immersion tube you have to replace the seal
husk as well! When removing and inserting the seal husk, do not use sharp
objects, which could damage it.
Warning:
Always wear goggles, protective gloves and appropriate protective clothing
when handling solvents. Follow any special national safety regulations.
2. Use the reverse side of the venting screw as tool for unscrewing the syringe
connector.
C49IB08A 324
10 Rotor 8SXF100 / 8SXQ80
3. Check the venting screw, the O-ring, the threads of the venting screw and
syringe connector and the syringe connector carefully. If you observe any
signs of deformation or corrosion, replace these components.
Seal husk
1. Unscrew the safety disk holder with a large screwdriver or a similar tool.
Important:
Take care not to damage the seal when unscrewing the safety disk holder!
C49IB08A 325
10 Rotor 8SXF100 / 8SXQ80
2. Insert the safety disk into the screw cap with the label "X" facing outwards.
Take care that the safety disk is positioned correctly. The holes in the safety
disk have to form a "C".
1. Before inserting the seal in the seal holder, take the immersion tube and
push it into the new lip-type seal from the top downwards until approx. 5 mm
of the immersion tube sticks out of the top of the seal. Take care not to dam-
age the sealing lip.
5 mm
C49IB08A 326
10 Rotor 8SXF100 / 8SXQ80
2. Fix the seal by screwing in the venting screw and the syringe connector.
Reopen the venting screw for one turn during storage.
3. Push a new seal husk provided with the seal onto the immersion tube.
Important:
The two ends of the seal husk have different diameters. Put the new seal husk
onto the immersion tube so that the wider end faces the seal holder. To
provide good sealing properties, the dimensions of the seal husk are chosen
to fit tightly. Use the special husk assembly tool to slide the seal husk up to the
final stop position.
Seal husk
Husk
assembly
tool
Fig. 10 - 87 Pushing the seal husk onto the immersion tube using the husk
assembly tool
4. Fix the immersion tube while carefully removing the husk assembly tool.
Important:
Never use sharp objects to push the seal husk onto the immersion tube.
Make sure that the seal is not damaged during assembly.
Important:
After replacing the seal, always carry out a performance run before the next
reaction to check the tightness of the seal. Weigh the solution before and after
the performance run and check for losses. If the losses are higher than 2%,
check the tightness of the seal and seal husk once again.
C49IB08A 327
10 Rotor 8SXF100 / 8SXQ80
Important:
After replacing the thermometer capillary you have to calibrate the
temperature sensor! Refer to chapter 6.7.4, "Temperature Sensor
Calibration with the Calibration Accessory" or to chapter 6.7.5, "Temperature
Sensor Calibration with the Calibration Unit". Carry out the steps described
below only if you are able to perform a temperature sensor calibration!
The capillary is very fragile. The most critical part is the connection between
the capillary and glass bulb, which can easily break.
If you are not completely sure how to replace the thermometer capillary
consult our local customer service.
1. Remove the PEEK screw (fixing screw), which holds the capillary onto the
arm of the T-Probe.
2. Unscrew the knurled nut of the capillary tube, pull it back and take it out of
the screw-type fitting.
Remove the knurled nut from the capillary.
C49IB08A 328
10 Rotor 8SXF100 / 8SXQ80
Knurled nut
4. Remove the small ferrule by tapping it out of the fitting on the electronics
module. The complete sealing part consists of two parts: knurled nut and
ferrule.
C49IB08A 329
10 Rotor 8SXF100 / 8SXQ80
7. Mount the knurled nut (delivered with the new capillary) and ferrule on the
capillary so that the metal end protrudes from the nut for 2 mm.
9. Screw on the PEEK screw (fixing screw), which holds the capillary.
Important:
The thermometer capillary must not be longer than 215 mm! (measured from
the sensor arm to the glass bulb)
2 mm
C49IB08A 330
10 Rotor 8SXF100 / 8SXQ80
C49IB08A 331
11 Applications
11 Applications
Synthos 3000 is delivered with a selected method library stored in the instrument.
The examples represent model reaction types, which are frequently performed in
synthesis laboratories. The programs listed in the method library are tested and
contain the following information:
- Rotor type
- Vessel type
- Number of vessels used
- Method name
- Batch size
- Required reagents
- Temperature or power program
When carrying out similar experiments, those are useful standard conditions to
start with. Any existing method from other microwave devices, which has been
optimized, can be entered in the database to be used with Synthos 3000. The
user is responsible for the consequences when carrying out reactions, which
have not been tested by Anton Paar.
Important:
The reaction behavior of sample materials may differ significantly from the
examples listed in the library, even if the sample designation is the same.
For optimal synthesis of your samples, you should generate and store your
own methods.
Although Synthos 3000 is verified for direct scalability of small batch sizes up to
large batch sizes, you should first test your new synthesis with small amounts to
document the reaction behavior in the microwave field.
The syntheses which are listed in the method library describe reaction types
which are frequently performed in synthesis laboratories. If you want to carry out
similar reactions, these are useful standard conditions for investigating reactions.
If you already have a method which has been optimized for small amounts, you
can enter this protocol in the database and use it with whichever amount you
require.
The user is responsible for the consequences of carrying out reactions which
have not been tested by Anton Paar.
C49IB08A 332
11 Applications
Ionic Conduction: Dissolved charged particles oscillate under the influence of the
microwave field causing continual collisions with neighboring molecules. Similar
to the dipolar polarization, this extensive motion causes heat.
The ability of solvents to convert the introduced electromagnetic energy into heat
is determined by the loss factor (tan d), which is expressed as the following
quotient:
C49IB08A 333
11 Applications
Ref. Microwaves in Organic and Medicinal Chemistry, C.O. Kappe, A. Stadler, Wiley-VCh, 2005
C49IB08A 334
11 Applications
The immediate and rapid heating of the sample using microwave irradiation
considerably accelerates the reaction. However, this direct heating method
involves possible dangers:
• Metals
Hint:
Never carry out synthesis reactions without stirring. Stirring increases the
efficiency by mixing the reaction partners thoroughly. It also prevents
undesired super heating effects (when the solvent is heated above its boiling
point although it does not show signs of boiling). When solutions are in a
super heated state they can boil explosively and destroy the vessels.
• Stir bars
The metallic core of a stir bar can act as an antenna and may intensively interact
with the microwave field. This can lead to hotspots and sparks. Only use stir bars
<3 cm (1/4 of wavelength) to ensure safe performance of microwave reactions.
• Operating error
Only trained personnel with sufficient experience should work with chemicals
under microwave conditions. It is the responsibility of the laboratory manager to
make sure that the relevant safety regulations are observed and that personnel
wear appropriate protective clothing.
Noxious reagents and substances which tend to form explosive reaction products,
require more strict safety regulations.
• Insufficient load
This is particularly problematic when using weak coupling substances. If you try
to heat a small amount of sample to high temperatures so that the magnetrons
continuously operate at high power, the microwave energy will be taken up by the
instrument and accessories. This can cause sparks, thermal deformation of the
rotor lid and other damage to the instrument. The magnetrons are effectively
protected against overheating so that no damage can occur to the magnetrons.
To avoid these effects, never use less than 4 vessels. If the reaction mixtures are
completely non-polar, we recommend using passive heating elements (refer to
chapter 11.1.3.1, "Solvents").
C49IB08A 335
11 Applications
• Explosive mixtures
If the sealing lip is not formed sufficiently or the the screw caps are not sealed
tightly enough, solvent may leak into the microwave cavity. As many organic
solvents can form explosive mixtures and microwave irradiation represents a
permanent source of ignition, it is essential that the vessels are leak-tight.
• Overloading
If the filling volulme is too large or the concentration too high, the safety limit may
be exceeded. If extreme pressure peaks or violent exothermal reactions occur,
this can lead to rupture of the safety disks and/or vessel explosion.
Hint:
• Pressure
Very reactive reagents or rapid heating rates can lead to the formation of high
pressures during the reaction which exceed the permitted limit. Thus, Synthos
3000 is equipped with numerous safety systems such as monitoring of the
pressure increase rate, patented cooling system, specially designed rotor
geometry and safety door. In normal use, any overpressure build-up is controlled
those safety measures.
• Temperature
The integrated security cooling always runs at a low level during a reaction. This
prevents overheating by exhausting the reaction heat. Most microwave-assisted
organic syntheses operate in the 100 to 220 °C range, which poses no problem
for Synthos 3000.
Hint:
For uniform heating distribution, the vessels should be loaded symmetrically
into the rotor. If the number of vessels used does not allow symmetrical
loading, we recommend charging the missing positions with blanks (pure
solvent). Never use less than 4 vessels in any of the rotors. When
programming a power profile, always take care to adapt the programmed
power to the number of used vessels.
C49IB08A 336
11 Applications
• Superheating
Microwave energy can heat solvents far above their classical boiling point.
Solvents in a superheated state can suddenly and explosively boil up and
therefore cause the reaction vessels to rupture. For this reason, never carry out
organic synthesis without a magnetic stirrer.
Solvent Kp Kp
(classical) (superheating)
Dichloromethane 40 °C 55 °C
Methanol 65 ° C 98 °C
Ethanol 78 °C 104 °C
Acetonitrile 81 °C 109 °C
• Local overheating
Important:
After every metal catalyzed reaction, check for deposits on the vessel surface
and clean the quartz vessels thoroughly. Do not carry out metal catalyzed
reactions in PTFE vessels, otherwise the liner may be permanently
contaminated due to adsorption.
C49IB08A 337
11 Applications
11.1.3 Reagents
Important:
Always wear protective glasses, gloves and the appropriate protective
clothing when handling chemicals. Observe the relevant national regulations.
11.1.3.1 Solvents
If this should not be the case, Anton Paar offers re-usable passive heating
elements made of silicon carbide (SiC), to be used as additives in low-absorbing
mixtures. Silicon carbide converts microwave energy very effectively into heat
and, being an excellent conductor, provides to achieve appropriate heat transfer
within the mixture. Thus, heating rates are highly comparable with direct
microwave heating. Furthermore, SiC is chemically inert and does not interfere
with the reagents nor contaminate the products. After the experiment, the heating
elements can be easily removed with tweezers and cleaned by simply rinsing with
any appropriate solvent.
C49IB08A 338
11 Applications
Table 11.1: Heating of low-absorbing solvents (one heating element per vessel)
Hint:
The microwave characteristics of solvents change at high temperatures.
Usually, their polarity decreases and their coupling properties deteriorate.
Take this into consideration when developing new methods.
The properties of water become comparable with those of organic solvents. It
is therefore possible to carry out high-temperature syntheses (>250 °C) in
water, which would not be possible or would not run well in organic solvents.
Whereas low boiling solvents like THF, ethanol or acetonitrile work nicely under
elevated microwave conditions some other organic solvents, such as DMF,
DMSO and NMP, decompose at high temperatures. This can be detected by
some brownish color and an unpleasant smell generated. For this reason, only
open and vent vessels containing these solvents under a well-functioning fume
hood. However, the generally short reaction times for microwave synthesis
prevent the decomposition of these substances progressing too far, so that it has
no negative effects on the reactions.
Important:
Utilizing low-boiling solvents may result in a significant pressure increase at
elevated temperatures. Therefore it is crucial to observe the operation limits of
the equipment used.
C49IB08A 339
11 Applications
Whereas neat reactions employ only the reagents without additional solvent
(maximized concentration) dry-media reactions utilize reagents pre-absorbed on
a solid support. Frequently used absorbing materials are silica gel, alumina, clays
(microwave transparent), graphite or charcoal (strongly absorbing).
Important:
Dry-media reactions can easily be locally overheated as stirring and accurate
temperature measurement can be rather difficult. Thus, degradation or
decomposition of reagents may likely be observed performing this kind of
reactions
Hint:
As polystyrene resins swell in organic solvents, do not add more than 5 g dry
resin per vessel (max. 1 g in MF50 vessels). This guarantees adequate
mixing and effective yield.
Other solid phase materials, such as lanterns, crowns or tubes can only be used
in small amounts due to their geometric form and volume. Problems may occur
when stirring in the reference vessel containing the internal temperature sensor.
The above-mentioned solid supports are not as temperature-stable as
polystyrene resins and therefore should not be used under the same conditions
as applied with polystyrene resins.
Large amounts of corrosive reagents (e.g. mineral acids and alkalis) are seldom
used in organic synthesis. Usually those chemicals are only used in small
amounts as catalysts or mediators. Synthos 3000 and all the accessories are
also designed for use with such reagents. If you observe the relevant cleaning
and maintenance instructions, the parts are ensured a long service life.
C49IB08A 340
11 Applications
Hydrofluoric acid HF
Important:
Never use strong bases such as concentrated potassium hydroxide and
sodium hydroxide in quartz vessels as those reagents will harm the glass
surface at high temperatures. Also hydrofluoric acid in high concentration will
attack quartz glass severely.
Some substrates can also be corrosive but they are usually converted to other
products, with other properties, during the reaction. These substances therefore
pose no danger with normal operation. If a safety disk ruptures, follow the
appropriate safety measures.
11.1.3.5 Gases
Hint:
For detailed information on the gas loading equipment refer to the
corresponding product information (B83IE36B).
This feature can be easily achieved with the aid of the Anton Paar Gas Loading
accessory. The assembly consists of the special Rotor Lid Gas for Rotor 8S and
a 4-way valve station for gas supply with a corresponding gas loading nozzle.
C49IB08A 341
11 Applications
Loading Nozzle
4-way valve
Fig. 11 - 3 Seal holder with gas loading adapter, reaction vessel closed with
protective cap gas and vessel arranged in Rotor 8S
The 4-way valve station enables simultaneous connection to inert gas, reactive
gas and even a vacuum pump. This allows removing air from the vessels and
protecting sensitive reagents effectively.
1. Charge each vessel with the reagents and solvents and seal properly with
the modified seal holder (Fig. 11 - 3/1).
2. Close the reaction container with the protection lid (Fig. 11 - 3/2) and place
the vessels in the Rotor (Fig. 11 - 3/3)
3. Put on the protection lid gas and place the rotor inside the Synthos 3000.
4. Attach the 4-way valve to the gas supply and connect the Loading nozzle
with the first reaction vessel via the appropriate opening of the rotor lid.
C49IB08A 342
11 Applications
Click
Twist 90°
5. Use the venting key to open the venting screw (Fig. 11 - 4 b). Open the gas
supply and flush the vessel for a few seconds.
6. Close the venting screw with the key and apply the required amount of gas
(max. 20 bar). Use the pressure controller of your gas supply to adjust the
appropriate pressure level.
Important:
Do not use too much force when closing the venting screw with the supplied
key. Fingertight closure is sufficient. Any more force applied may strain the
sealing O-ring.
7. When the required pressure level is reached, close the gas supply and
disconnect the bayonet link.
8. Move the rotor manually to get access to the next vessel and proceed
according to steps 4-7. When done with all vessel start your experimental
procedure.
Hint:
To have the maximum of 20 bar pre-pressurizing available software v.2.0 or
higher is required.
Flushing with inert gas (nitrogen, argon) is also possible for reaction vessels of
Rotor 16. Therefore, the corresponding loading adaptors Gas 16 are required
(Fig. 11 - 5). This is achieved outside the cavity before closing the rotor lid,
following step 5 in the above description. Flush the vessel for approximately 90
sec or perform 3 cycles applying vacuum and inert gas alternately for a proper
inert gas atmosphere.
C49IB08A 343
11 Applications
Observe the general safety regulations when using reactive gases such as
ethylene or oxygen and take into consideration the maximum pressure load of the
vessels.
Warning:
Explosive mixtures always pose an operational risk in the microwave field. To
avoid such conditions, we do not recommend to use hydrogen gas for
microwave-mediated reactions. Extreme care has to be taken! The user is
responsible for occuring consequences when utilizing dangerous reagents.
C49IB08A 344
11 Applications
The following vessel types are available with the Synthos 3000:
These vessels are suitable for standard synthesis, in which the pressure and
temperature does not rise into extreme ranges. We also recommend these
vessels for solid phase synthesis as the solid phase components do not adhere
to the vessel surface and can therefore be handled easier.
Quartz vessels:
These vessels were specially designed for applications under extreme conditions.
Pressures up to 80 bar at temperatures up to 300 °C can be reached easily and
safely. However, when using weak coupling solvents the operating time should
not be longer than 60 minutes at 300 °C due to the thermal stress for rotor parts.
To avoid any damage, the use of passice heating elements is recommended in
this case.
Important:
Do not use stir bars when applying SiC heating elements with quartz vessels.
The glass surface may be scratched and pressure stability can not be
generated anymore.
The quartz vessels are also suitable for metal catalyzed reactions as any
elementary metal which is produced can be easily removed from the surface and
there is no vessel contamination due to adsorption.
C49IB08A 345
11 Applications
Hint:
Only clean the quartz vessels with a soft brush; never use a metal spatula. Try
to avoid using mechanical force as this could damage the vessels by causing
scratches. Never use damaged or scratched vessels for reactions as the
pressure resistance is no longer guaranteed. Very dirty vessels can be
cleaned in Synthos 3000 in a cleaning run (refer to chapter 10.4.4, "Standard
Cleaning and Maintenance Procedures").
Rotor load:
For uniform heating distribution, the vessels should be loaded symmetrically into
the rotor. Thus, always adhere to the recommended loading paterns of the
respective rotor type (refer to chapter 8.4.3, "Loading Rotor 48", chapter 9.3.3,
"Loading Rotor 16" and chapter 10.3.3, "Loading Rotor 8S"). If the number of
used vessels does not allow symmetrical loading we recommend loading the
missing vessel positions with blanks (pure solvent).
Hint:
We always recommend to perform synthesis applications with inner
temperature measurement.
C49IB08A 346
11 Applications
11.2.1 General
• There are different sensor types available for Synthos 3000. These allow
optimal control of the reactions. Rotor 16 and Rotor 48, respectively enable
you to use a combined immersing pressure/temperature sensor in one
reference vessel. Rotor 8S has a hydraulic sensor incorporated, which
measures the pressure at every position. An additional temperature probe is
also available which measures the inside temperature of the reference
vessel. Furthermore, Synthos 3000 is equipped with an IR temperature
sensor, which measures the external vessel temperature of each vessel.
Hint:
Many solvents and reagents used in organic synthesis are highly
inflammable and require special safety measures.
C49IB08A 347
11 Applications
• Please note that, due to the continuous flow of air from the cooling system,
the measured surface temperature is always lower than the reaction
temperature inside the vessel. Take this into consideration when
programming a method without internal temperature sensor.
Hint:
Each Synthos 3000 is delivered with an IR sensor as standard. We
recommend additionally using the optional internal temperature sensor.
However, it is not essential for achieving good synthesis results. The decision
should be made by the user.
• The creation of new methods requires only a few steps (refer to chapter 6,
"Software"). Even with unknown, reactive reactions, the automatic system
control prevents vessel limits being exceeded. The optimal conditions are
therefore achieved easily and safely.
• When using sensors, you can choose between two different profiles: power
profile and temperature profile.
Hint:
As synthesis reactions are highly dependent on the temperature, we
recommend using an internal temperature sensor to ensure optimal reaction
conditions during large scale operations.
• Both the power profile and temperature profile allow you to adapt methods
from other manufacturers and from the small scale range. When using a
power profile, always set the programmed power according to the number of
vessels used (the lower the number of vessels used, the lower the power).
C49IB08A 348
11 Applications
Hint:
Be aware that power values of monomode instruments can not be directly
applied similarily in multimode equipments. Due to the different filed density
the heating rates will differ.
• Due to the pressure control at every position, Rotor 8S enables you to use
different reagent mixtures in one run. This allows you to determine the
suitable proportions of reaction partners. The minimum volume of 12 mL also
helps keep the consumption of chemicals as low as possible. Once you have
found the ideal ratio, you can optimize the time/temperature or pressure
profile. The optimized method can then be stored in the method library for
future use.
C49IB08A 349
11 Applications
• The minimum filling volume of the reaction vessels is 6 mL unless you use
the internal T-probe in Rotor 8S - then the minimum filling volume is 12 mL in
the reference vessel. However, to make sure that the weight is evenly
distributed, all vessels should be filled with the same volume.
• If the reaction results in precipitating product upon cooling, the batch size
should be selected to form no more than 20 g of product per vessel.
C49IB08A 350
11 Applications
• If the result of the reaction fails to meet expectations, this can be due to any
of the following reasons:
- if the conversion is incomplete, the reaction time is too short and/or the
temperature is too low
- if the substrate is destroyed, the temperature is too high
- the reaction mixture is unsuitable. Consider using catalysts, passive
heating elements or changing the solvent
Hint:
Do not use mechanical force when transferring the sample. Only use a soft
plastic spatula in order not to damage the liner or quartz vessels.
Damaged or scratched vessels should not be used as the pressure safety
limit is no longer guaranteed.
C49IB08A 351
11 Applications
• M: Magnetron overtemperature
C49IB08A 352
12 Appendix
12 Appendix
12.1 Addresses
Anton-Paar-Str. 20
info@anton-paar.com
www.anton-paar.com
C49IB08A 353
12 Appendix
If you need to transport the instrument, use the original packaging. This is the best
protection against transport damages.
1. Place the accessory parts such as the exhaust hose, mains cable and the
seal forming tool into the supplied accessory carton. Close the carton and
seal it using adhesive tape.
2. Do not use metal clamps to avoid scratches on the PTFE surface of the
instrument!
3. Open the door of the instrument and place the carton inside the microwave
cavity of Synthos 3000.
4. Apply the protective sheet to the inner side of the door and close the door.
C49IB08A 354
12 Appendix
protective sheet
5. Insert the transport lock (white L-shaped foam part) of the exhaust unit at the
rear of Synthos 3000. Insert the white foam piece between the top of the
exhaust unit and the top cover.
6. Place the wooden pallet on the floor and position the bottom carton on the
pallet. Insert the bottom foam piece into the bottom carton. The recess for
the exhaust unit helps you identify the front and rear of the foam piece.
C49IB08A 355
12 Appendix
rear
front
Warning:
Do not use the door for lifting or carrying the instrument!
Lift and carry the instrument by gripping the base at the right and left side (2
people are required).
7. Insert the instrument into the carton. Do not lift or carry the instrument by the
door.
8. Put the carton sleeve on the instrument. Ensure that the bottom corners of
the carton sleeve fit inside the carton.
C49IB08A 356
12 Appendix
10. Close the carton and seal it using adhesive tape. Use belts to secure the
carton on the wooden pallet.
C49IB08A 357
12 Appendix
The rotors are delivered in a separate shipping cartons. All rotors are packed
similarly. The instructions below are shown with Rotor 16 but can be applied to all
other rotor types.
1. Open the rotor by removing the lid. Put the foam piece (transport lock) on the
rotor and close the rotor with the rotor lid. (Hint: Rotor 8S is delivered without
this foam piece)
C49IB08A 358
12 Appendix
C49IB08A 359
12 Appendix
Mass: 74 kg
Dimensions (WxDxH): 60 x 72 x 74 cm
(23.6 x 28.3 x 29 inch)
Cavity volume: 66 L
C49IB08A 360
12 Appendix
Ambient temperature: 15 to 35 °C
Altitude: 0-2000 m
Overvoltage class: II
Pollution degree: 2
C49IB08A 361
12 Appendix
12.3.2 Options
Number of vessels: 48
Volume: 50 mL
Mass: 13.1 kg
C49IB08A 362
12 Appendix
Number of vessels: 16
Volume: 100 mL
Mass: 10.5 kg
Number of vessels: 16
Volume: 100 mL
Mass: 13.8 kg
C49IB08A 363
12 Appendix
Number of vessels: 8
Volume: 100 mL
Mass: 14.9 kg
Number of vessels: 8
Volume: 80 mL
Mass: 14.8 kg
C49IB08A 364
12 Appendix
Detailed information about the Spare Parts for Synthos 3000 can be found in the
Product Description List which is additionally stored on the CD.
C49IB08A 365
12 Appendix
Fig. 12 - 16 ETL-Certification
C49IB08A 366
12 Appendix
Fig. 12 - 17 GS-Certification
C49IB08A 367
12 Appendix
C49IB08A 368
12 Appendix
If you experience technical problems with Synthos 3000, Anton Paar GmbH
provides quick and expert support. For technical support, contact our local
representative or Anton Paar GmbH directly.
If you have a technical question or complaint, please fill in both, the Safety
Declaration for Repair form and the Malfunction Report as detailed as possible
and send it (with the defective part, if possible) to our local representative or to
Anton Paar GmbH directly.
C49IB08A 369
12 Appendix
C49IB08A 370
12 Appendix
C49IB08A 371