Documente Academic
Documente Profesional
Documente Cultură
The Association for the Advancement of Medical Instrumentation the maximum dose as high as possible allows the greatest
(AAMI) generates numerous standards used by professionals in flexibility in processing schedules when product is ready for
the medical device industry. Occasionally, the AAMI Standards routine sterilization. Accelerated aging and package testing are
board provides additional guidance to specific standards in the additional tests to be considered for product irradiated at the
form of a Technical Information Report (TIR). These TIR’s reflect maximum dose.
common industry practices that evolve from an accumulated
Why is the verification dose experiment performed at a
process knowledge base.
lower SAL than the sterilization dose?
When gamma irradiation is selected for product sterilization, the
In order to test a dose for SAL 10-6, one million products would
dose at which the product is irradiated must be established and
need to be irradiated and sterility tested. SAL 10-2 (Method 1)
validated in accordance with appropriate AAMI standards. ANSI/
or SAL 10-1 (VDmax) is used because only 100 or 10 products,
AAMI/ISO 11137: 2006 Sterilization of health care products-
respectively, must be used for the experiment.
Radiation and ISO TS 13004. Sterilization of Healthcare
Products Radiation - substantiation of a selected sterilization What if my product sample is dosed at less than 90% of the
dose method VDmax SD Sterilization of health care products- target verification (sub-lethal) dose?
Radiation -Substantiation of a selected sterilization dose- Method If the sterility test exhibits a failing number of positive tests,
VDmax25 kGy as a sterilization dose-Method VDmax, provide are the verification dose experiment can be performed again and
established methods for completing a validation process. samples re-tested.
FAQs: What if my product sample is dosed greater than the target
Which validation is right for me? verification (sub-lethal) dose plus 10%?
The VDmax25 option (formerly TIR27, now in 11137: 2006) This is considered an overdose. During a verification (sublethal)
is convenient when a company wants several product lines dose experiment, it is not permissible to irradiate over 10%
sterilized at the same minimum dose, when product is expensive above the target. Do not sterility test the samples. Send new
to make, or for companies with markets where a 25 kGy dose samples for irradiation prior to sterility testing.
is the accepted standard. Furthermore, the validation is less
expensive because fewer tests are necessary. Bioburden counts Do I need a Biological Indicator?
must be 1000 CFU or less. Bacillus pumilus, a spore-forming microorganism, served for
Method 1 (from 11137-2) determines the lowest sterilization dose many years as a biological indicator to test for sterility. Its
necessary for the determined bioburden population. This Method use today has been discontinued. BIs also do not accurately
should be used when the lowest possible sterilization dose is represent natural form of bioburden on a product ( spore strip
desired due to cost considerations, use of gamma sensitive vs. actually in or on product as a result of manufacture). Detailed
materials, or when the bioburden count is above 1000 CFU. methodology has been developed and publicized in ISO 11137-
2 to determine minimum radiation dose to achieve a sterile
If one of these validations establishes my minimum dose, product.
how do I establish a maximum dose?
What are the basic steps?
Performed early in product qualification, materials can
be screened for compatibility with irradiation. Pre- and For both types of validations, the first step is identical: Have
postirradiation properties related to functionality and appearance bioburden testing performed on 10 products from three different
must be evaluated to determine maximum dose. Irradiating batches, for a total of 30 products.
Are there other options for dose setting besides VDmax25 required for this testing should be confirmed with the laboratory
and Method 1? performing the testing (usually 3-6).
Bioburden 10
products from each
of 3 lots
Send 13 samples
Find verification to STERIS for
dose (SAL 10-2) on irradiation at
Table 5 of AAMI verification
11137 Method One dose +/- 10%
Send 10 samples
Lab performs B/F Were there
to STERIS for
testing on 3 samples, more than 2
irradiation at
sterility testing on positives?
verification dose
100 samples +/- 10%
Were there This method of dose- Find alternate dose Lab performs
more than 2 setting is not valid. setting method sterility testing on
positive tests? An alternate method 10 samples
should be used
Substantiated for
25 kGy
For more information, please contact:
STERIS Applied Sterilization Technologies
5960 Heisley Road
Mentor, OH 44077
877.783.7479
www.steris-ast.com TechTip 03, 02/16, Rev 2