::: Application Report

Thermal Curing of 1-Component Impregnating


Resins

Relevant for: DMTA reaction kinetics, curing resins Passendes Bild!


Natural and synthetic resins like the tree resin shown on the picture are
used for a broad range of applications, like for the production of varnishes
or adhesives. This report shows the thermal curing of a 1-component
impregnating resin measured on a rheometer. cc by: copyright/friscomix

1 Introduction • disposable aluminum plate(s) D-PP12


used for samples with G’ modulus not significantly
Impregnating resins are key components of modern above 150 MPa
production. The advantages are excellent protection from
• disposable aluminum plate(s) D-PP08
a wide variety of environmental factors as well as improved used for samples with G’ modulus not significantly
heat transfer and wire insulation. above 1500 MPa.

Conventional dip & bake technology is a time consuming


and energy intensive manufacturing process which
requires large-scale equipment. It also leads to high
emissions during the curing process and it requires
frequent cleaning.

To reduce costs and improve product quality, it is essential


to evaluate the material properties during the entire
production process.

Modern rheometers in combination with modular DMTA


accessories allow measurements from the liquid raw Fig. 1 Lower Peltier temperature control device with disposable
plates (upper and lower)
product to the finally cured solid.

2.2 Default
2 Experimental Setup
In most cases, a D-PP15 is a good compromise for a
2.1 Sample and Measuring Device reasonable measuring sensitivity from the liquid to the solid
form.
A 1-component impregnating resin was measured. MCR
series rheometers are suitable for this type of
measurement. 2.3 More Sensitivity until Reaching the Gel Point
It is recommended to use the following plate(s):
For a better sensitivity at the lower end of the measuring
• a special shaft D-CP/PP7 curve the diameter should be increased (25 mm or above)
for holding the disposable measuring plates with the disadvantage of measuring only relative
information of the solid material.
• disposable aluminum plate(s) D-PP25
used for samples with G’ modulus not significantly
above 10 MPa 2.4 More Sensitivity Exceeding the Curing Point
• disposable aluminum plate(s) D-PP15
used for samples with G’ modulus not significantly Whenever the absolute modulus of the solid-like behavior
above 60 MPa is of highest priority, a D-PP08 is the best choice.

C92IA008EN-A 1 www.anton-paar.com
::: Application Report

2.5 Best Choice for Measuring on Pure Solids 2.8 Standards


(Bars)
The rheometer in combination with the parallel plate
An even more precise technique is offered by the solid measuring system fulfils the following standards:
rectangular fixture SRF which allows high precision DMTA DIN53018, ISO6721-10
measurements on solid bars. Disadvantage: the sample
must be cut into the right shape e.g. (10 x 2 x 50) mm
3 Results and Discussion
2.6 Temperature Control Device
The kinetics of the reactive resin has been investigated by
A Peltier oven with a Peltier controlled lower plate presetting a temperature ramp followed by an isothermal
P-PTD 200 and an actively Peltier controlled upper hood step at the curing temperature.
H-PTD 200 was used. The system ensures a virtually The 1-component impregnating resin presented in figure 3
gradient free measurement, offering rapid cooling and was heated from 20 °C to 100 °C within 3 minutes. While
heating based on the thermo-electrical effect of the Peltier increasing the temperature the resin undergoes a viscosity
system (lower & upper one). reduction.
http://en.wikipedia.org/wiki/Thermoelectric_cooling
In order to protect the lower Peltier plate from destruction,
a special holder for disposable dishes was setup (see
3.1 Softening Point or Viscosity Minimum
figure 1).
From interval 2 onwards, the temperature is kept constant
2.7 Measuring Profile and the viscosity reaches a minimum value which is the so
called softening point (red dot at 56 mPas appearing
In accordance to the processing conditions the following 6 minutes after the maximum temperature was reached).
measuring profile was designed. Until reaching the softening point the sample is liquid like
and shows ideal dip-coating or leveling properties.
All intervals: Due to the permanent exposure to a high temperature the
normal force compensation 0 N cross-linking reaction starts. With the occurrence of initial
hysteresis ±0.5 N cross-links the sample turns into a non-Newtonian liquid
oscillation with frequency 1 Hz and as a consequence, the viscosity increases by 7
decades in a logarithmically scaled diagram. The sample is
Interval 1 (3 minutes): found to be solid at the end of the test.
heating with 25 K/min from 20° C to 100 °C
strain = 5 %: low viscosity -> set high strain value

Interval 2 (12 minutes):


isothermal 100 °C
simulation of oven temperature
strain reduced from 5 % to 0.1 %:
while sample becomes increasingly solid-like

Interval 3 (30 minutes):


isothermal 100 °C
simulation of oven temperature
strain=0.1 %:
sample is solid like -> low strain value

Interval 4: Setback to initial temperature 20 °C.


Fig. 3 Complex viscosity versus time of a 1- component curing
resin.

For any further investigations it is recommended to have a


closer look at the moduli (G', G'') versus time in order to
obtain significant changes of the mechanical material
Fig. 2 Measuring profile with 4 intervals properties.

C92IA008EN-A 2 www.anton-paar.com
::: Application Report

3.2 Pot Life

With the occurrence of initial cross-links a storage modulus


G' can be measured.
At this point the end of the almost horizontal plateau with
its constant complex viscosity is reached (figure 3), and the
complex viscosity starts to increase. Until reaching its pot
life (blue dot) the sample is still soft and liquid-like.

3.3 Sol-Gel Point

The value of the cross-point of G' and G'' (figure 4)


indicates the sol-gel transition from where on the value of
the storage modulus predominates over the value of the
Fig. 5 Relative rate of cross-linking X(t) and storage modulus G'
loss modulus. At the sol-gel point (green dot), the resin
versus time in a linear scaled diagram
reaches a gel-like state with an increasing degree of
cross-linkage. During the long lasting after-curing period the sample
reaches its final rigidity and the rate of cross-linking
becomes constant showing the value of almost zero.

Table 1: Evaluated Points

Softening point (red) 9.0 min

Pot life (blue) 10.1 min

Sol-gel point (green) 10.8 min

Curing point (yellow) 18.0 min

4 Summary
Fig. 4 Storage modulus G' and loss modulus G'' versus time of a It could be shown that the EC-Twist or MCR rheometers
1-component curing resin can be used to characterize curing resins beginning from
the liquid state at ambient temperature until reaching the
3.4 Rate of Cross-LInkage solid state after curing finally.
The entire process of curing is measured with respect to
The slope X(t) of the storage modulus in the diagram the sample's viscosity and other material properties. In
(linear scale) below (figure 5) is an indirect measure of the addition, some significant points can be analyzed such as
the softening point, the sol-gel point and the curing point.
rate of cross-linking.
Further, the rheometer software allows the calculation of
the relative rate of cross-linking based on the slope of G'(t).
3.5 Curing Point One more benefit of this method is the ability of having a
simulation which almost meets the true processing
The maximum slope X(18 min) is reached after about conditions with respect of time and temperature.
18 min (yellow dot). From that point onwards the reaction
Text: Klaus Wollny
is slowing down. This maximum slope of G' (linear scale)
can also be interpreted as the curing point. While looking
Contact: Anton Paar GmbH
at the logarithmically scaled diagram in figure 4 it is obvious Tel:+49 711 720 91 0
that the sample has almost reached its final material rheo-application@anton-paar.com
properties after about 18 min. http://www.anton-paar.com

C92IA008EN-A 3 www.anton-paar.com