Ultraformer

ULTRAFORMERUSER’S MANUAL

NO:CTI-V-0900
Warning & Caution

Warning.

The patient must wear an eye shield, during procedure. Doctor


must wear latex gloves. Products do not expose to rain or
moisture. To prevent electric shock, plug into an extension cord,
do not use an outlet and This product must be earth ground.

Caution.

Always turn off the system when it is not in use.


Never leave the system in ready mode unattended.
Never allow untrained personnel to operate the system.

Ultraformer has the ultrasound frequency of 7Mhz, 4Mhz,


Which can cause human eye burns.

Patient must wear eye goggles for all facial treatment.


Ultrasound without gel coating on film cause serious damage to
the cartridge.

2
Contents

Warning and Caution


Components
Installation
Caution
Operation
Maintenance
Trouble shooting
Treatment guide

3
Components

Ultraformer components

1 Ultra former main body.


2 Handpiece (A)
3 Handpiece (B)
4 AC Power Cord ( 5m )

4
Introduction

Ultraformer Front view

1 Knob
2 Handpiece Holder
3 POWER KEY SWITCH
4 EMERGENCY SWITCH
5 10.4“ LCD SCREEN

5
Introduction

Side view

1 Handpiece Jack
2 Caster

6
Introduction

Rear View

1
POWER inlet : AC100V~240V

7
Introduction

Handpiece

1 SHOT : Ultrasound emission Red LED lights.


2 TIP: when Cartridge loaded Blue LED
3 READY: Ready for emission Green LED lights.
4 Handpiece nonslip knob
5 SHOOTING SWITCH : Ultrasound emission switch.

8
Introduction

Cartridge

L4-4.5
Frequency :
4Mhz
Depth : 4.5mm

L7-3
Frequency : 7Mhz
Depth : 3mm

9
Installation

UltraFormer

1. The ultraformer system is designed for installation in an


office or a clinic and requires minimal site preparation.
2. System transportation and installation is carried out by
authorized technical Personnel who trained.

10
Installation
Power cable connecting

POWER inlet :
AC100V~240V

11
Installation

Handpiece cable connecting L4-4.5

Plug your handpiece cable into proper color


4.5mm =>Blue plug
12
Installation

Handpiece cable connecting L7-3

Plug your handpiece cable into proper color


3.0mm =>Green plug

13
Installation

How to replace cartridge

1 Open cover

2 Push cartridge into the inside of cartridge

Push cover down


3

14
caution

Do not shoot without applying ultrasound gel.


caution

Applying ultrasound gel and shooting toward the bottom


caution

Must Be using alcohol

Cleaning the window film


Operation

1. Turn on the “key switch


2. Check skin with other ultrasound image viewer.
3. Set the proper parameter.
4. Apply the ultrasonic transmission get to skin
5. Contact the cartridge of handpiece to skin
6. Press the shot switch of handpiece.
7. Move and confirm contact and next shot.
8. Check the patient and repeat treatment.

Caution
1. Do not shot in the air without applying ultrasound
transmission gel.

2. Do not shot to skin without applying ultrasound


transmission gel.

3. Before shot, check the status of cartridge which is


contacted to skin

18
Operation

1. MAIN MENU SCREEN DISPLAY

12

3 11

4
10

8 9

1 Adjust and Change Value

SIGN FUCTION

Adjust and Change the value of ‘POWER”,


“PITCH”, “LENGTH” values.

19
Operation

2 Ultrasound “POWER “value setting

SIGN FUCTION

Ready for Ultrasound power output


(From 0.1J ~ to1.0J)

Standby for Ultrasound power output

3 Ultrasound “PITCH” value setting


s
SIGN FUCTION

Set interval between ultrasound output


(1.0mm, 1.5mm, 2.0mm)

Standby interval between ultrasound


output

4 Ultrasound “Length” value setting

SIGN FUCTION
Ready for Ultrasound length value
adjust
(From5mm~to 25mm, Setting value can
be adjusted by 1mm)

Standby Ultrasound length value


adjust

5 Adjust and Value change

SIGN FUCTION

Adjust and Change the value of


‘POWER”, “PITCH”, “LENGTH” values.

20
Operation

6 Continuous shooting Ultrasound power


SIGN FUCTION
Continuous shooting Ultrasound power when shooting
switch “ON”

Single shooting Ultrasound power when


shooting switch “ ON”

7 Total Ultrasound shot display


SIGN FUCTION
Current : Total shot from “Power on” to Power
off”
Total : Total accumulated shot

8 Ultrasound selecting mode

SIGN FUCTION

Connecting cartridge status

Ultrasound selecting status

9 Setting value “READY” and “STANDBY”


SIGN FUCTION
7
READY MODE.

STANDBY MODE

21
사용 방법 Operation

10 User setting value save

SIGN FUCTION

Save user setting value

11 Preset value setting

SIGN FUCTION

Recall the saved value

12 Total use of cumulative use of “A”,B” cartridge

SIGN FUCTION

Total use of cumulative use of “A”,B”


cartridge from turn on to turn off.

22
Maintenance

Warning
Always turn the system off and unplug the power cable before performing
Maintenance procedures. And interior of the system or its components may be
Serviced only by local authorized technical personnel.

Main body cleansing


The outer surface of the system may be wiped clean with a soft dry cotton clot
h once every 10days.
Be careful not for water flowing in the inside of the system.

Handpiece cleaning
Cartridge be wiped clean with alcohol .

Storage and Operating.


1. Authorized personnel to operate the system keep and put the start –key on
Special place.
2. Keep the system in the clean and dry area for a long-term use.
3. Clean the outer cover with soft cotton cloth regularly.
4. Do not twist and distort the handpiece cable unnecessarily.

23
Troubleshooting

Symptom Probable Cause Corrective action

1.Check if the power cable is


Properly plugged into the
1.Bad connection of power wall outlet and main body's
Display does cable socket.
not work when 2.Emergency shut-off knob is 2.Turn off the system, turn
the keys witch engaged the emergency shut-off knob
is turned on 3.Fuse is short-circuited clockwise and restart the
4.LCD is malfunctioned. system.
3.4. Call your local
representative agency

1 ,2
Cartridge dose
1. No display frequency Check connecting of
not working
value on main screen handpiece cable and
when the
2. No “READY “ Setting cartridge reload.
loaded
3. Handpiece is defective 3,4 . Call your local
handpiece.
4. Cartridge is defective representative agency

No emission of 1. Main body defective


1,2 ,3 Call your local
Ultrasound 2. Cartridge defective
representative agency.
energy 3. Shot switch defective

24
Ultraformer treatment guide.

Section Type Joule Shot

Forehead 3mm 0.3~0.4j 70~100

Cheek 4.5mm 0.6~0.8j 80~100

Cheek 3mm 0.6~0.8j 150~190

Total 300~400
Specification

Energy Type High intensity focused Ultrasonic surgical unit

Fluency 0.1~1J cm2

Length 5~25mm

Spacing 1~2mm

Frequency mode 4Mhz, 7Mhz

Display LCD 10.4inch touch screen

Electrical
220 VAC, 50/60H
Requirement
Electric power
1200VA
consumption

Dimensions 312(D)×537(W)×902(H)

Weight 34kg

26