Service Manual
REF: M30003.UK
03.11.2016
Congratulations on your purchase of this quality ANDROMEDA product. Please read this
user’s manual through carefully so you can service your equipment correctly. Keep it in a
safe place so you can refer to it in the future.
The author has carefully striven to make it as complete and accurate as possible. How-
ever, ANDROMEDA accepts no responsibility for its accuracy.
Table of Contents
Table of Contents ..............................................................................................................1
Information and warning advice ......................................................................................3
Applicable guidelines ...........................................................................................................3
Warning advice ....................................................................................................................3
Electromagnetic and other influences .................................................................................3
Transportation and storage .................................................................................................4
Liability limitations................................................................................................................4
Disposal ...............................................................................................................................4
Safety...................................................................................................................................4
Maintenance ........................................................................................................................5
Cleaning and disinfection.....................................................................................................5
Key to the utilized symbols ..................................................................................................5
Electrical safety....................................................................................................................6
Arrangement and set-up.................................................................................................6
Safety inspection according to IEC 60601-1 ..................................................................7
Assembling Instructions...................................................................................................8
Stand....................................................................................................................................8
Basic unit .............................................................................................................................9
Adaption module for EMG, Uroflow and Puller..................................................................10
Pump module.....................................................................................................................10
Pressure module................................................................................................................11
Puller..................................................................................................................................11
Uroflow...............................................................................................................................12
EMG...................................................................................................................................12
Technical Details .............................................................................................................13
Basic Unit (E31003)...........................................................................................................13
Parts List.......................................................................................................................13
Basic Unit Lower Assembly (HGE00041)..........................................................................15
Parts List.......................................................................................................................15
Display Assembly (HGE00039) .........................................................................................17
Parts List.......................................................................................................................17
Printer Assembly (HGE00014) ..........................................................................................19
Parts List.......................................................................................................................19
Basic Unit PCB ..................................................................................................................21
Stand (E30502)..................................................................................................................29
Parts List.......................................................................................................................29
Puller/Flow/EMG Interface Module (E30521) ....................................................................31
Parts List.......................................................................................................................31
Puller/Flow/EMG Interface PCB ........................................................................................33
H2O Pump Module (E30802) ............................................................................................36
Parts List.......................................................................................................................36
Adjustments ..................................................................................................................37
Pump Module PCB ............................................................................................................40
Volume transducer PCB ...............................................................................................40
Adjustment....................................................................................................................40
Functional Test .............................................................................................................40
ANDROMEDA 03.11.2016 1
Table of Contents HELIX Service Manual
2 03.11.2016 ANDROMEDA
HELIX – Service Manual Information and warning advice
Applicable guidelines
Helix has been developed and tested in accordance with the European Safety Standard
IEC 60601-1. edition “Medical Electrical Equipment”. This User Manual contains informa-
0123 tion and warning advice which must be heeded by the individual user to ensure safe
operation and to safeguard against damage to the Helix equipment.
The operation of this device is only permitted within buildings with a room temperature
range of between +10°C (+50°F) and +40°C (+104°F).
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (e.g. IEC 60950 for data processing equipment
and IEC 601-1 for medical equipment). Furthermore, all configurations shall comply with
the valid version of the system standard IEC 60601-1-1. Anybody who connects addi-
tional equipment to the signal input part or signal output part is deemed to be configuring
a medical system and therefore to be responsible for the system complying with the
requirements of the valid version of the system standard IEC 60601-1-1. If in doubt, the
technical service department or a local representative should be consulted.
Warning advice
Only when the mains plug has been withdrawn from the mains supply socket is discon-
nection from the mains supply ensured. Opening a connected device leaves live compo-
nents open to possible bodily contact.
The Helix must be disconnected from all sources of electrical mains power before main-
tenance or repair work is carried-out.
Apart from the applications described in this User Manual, service and maintenance work
may be carried-out by authorized service personnel only.
ANDROMEDA 03.11.2016 3
Information and warning advice HELIX Service Manual
Liability limitations
Attention! HELIX may only be operated according to the provisions of the Instructions for Use. Only
peripheral equipment or expendable materials approved by ANDROMEDA may be used
in combination with the HELIX device. ANDROMEDA accepts no liability for damage
based on improper usage.
All repairs and maintenance work on the equipment may be carried out by authorized
specialists. Unauthorized opening of the unit and/or the intervention of an unauthorized
person will result in the immediate cessation of any liability on the part of ANDROMEDA
for the equipment as well as for all consequent damage arising there from.
Disposal
The equipment packaging should be submitted for recycling; the metal parts in the
equipment as scrap-metal; plastics, electrical components and circuit boards for disposal
as electronic scrap.
The respective national laws and regulations applying to waste disposal and recycling
must be adhered to. It may be necessary to employ specialist waste-disposal companies
which can be found by inquiring through local authority services departments.
Safety
As soon as safety deficiencies become apparent, the equipment must be switched off
and disconnected from all sources of mains electrical power and be secured against
unintentional operation. In such a case, authorized service personnel are to be consulted
to carry-out functional testing and a safety examination including a residual current test
according to the regulations in EN60601-1.
Safety deficiencies could occur if:
visible damage is apparent,
the functionality of the equipment is diminished,
the equipment has been stored over an extended period under unsuitable conditions.
4 03.11.2016 ANDROMEDA
HELIX – Service Manual Information and warning advice
Maintenance
Maintenance work which can be carried out by the user:
Cleaning of the peripherals and the control unit.
Work which may only be carried out by authorised service personal:
Calibration of the device, maintenance work inside the Uroflow measuring device and the
power supply.
Equipotential Terminal.
Attention! The Ellipse must only be connected to a power connector with
protective earth.
Caution! Risk of injury to fingers. Do not touch the roller pump when in opera-
tion.
ANDROMEDA 03.11.2016 5
Information and warning advice HELIX Service Manual
Electrical safety
The Helix is insulated according to electrical class 1 BF, thus permitting the connection of
proprietary PCs and printers without risk to patient safety. The casing of the control unit is
grounded and the transducer, which is situated near the patient, is electrically isolated.
6 03.11.2016 ANDROMEDA
HELIX – Service Manual Information and warning advice
ANDROMEDA 03.11.2016 7
Assembling Instructions HELIX Service Manual
Assembling Instructions
Stand
Please notice! Two of the wheels are
equipped with brakes. We recommend to
position these in the front (the bevelling of
the vertical ears points to the rear of the
stand).
5
Insert leg into stand core and fix with Rear
screw and nut - DO NOT TIGHTEN.
3
2
Angle the leg until hole and thread are
aligned and fix with screw . Now tighten
screw and nut . 1
4
Glue the cable guides on the bottom 6
side of the leg. We recommend to use the
rear legs to guide the cables behind the
Helix system.
Rear
8 03.11.2016 ANDROMEDA
HELIX – Service Manual Assembling Instructions
Basic unit
Optional - only when the delivery includes
the PC interface box:
Mount the PC interface box on basic unit 2
1
and fix with curled screws .
ANDROMEDA 03.11.2016 9
Assembling Instructions HELIX Service Manual
Pump module
Connect flat cable from the adaption
module for EMG, Uroflow and Puller and
the flat cable to the pump module .
Guide both cables up through the 3
2
stand.
1
Fix module to the stand with screws .
Connect cables to the basic unit.
4
10 03.11.2016 ANDROMEDA
HELIX – Service Manual Assembling Instructions
Pressure module
Connect flat cable to the module and
basic unit.
Fix module to the stand with screws . 1
2
Puller
Fix puller on stand with screws .
Pull cable through the stand core from
the bottom into the adapter module and
connect.
Prevent the cable sliding on the floor by
pulling the cable into the adapter module. 1
Notice!
Be sure to attach the plug to connector
named “Puller”. Attaching it to other
connectors may damage the puller.
2 3
ANDROMEDA 03.11.2016 11
Assembling Instructions HELIX Service Manual
Uroflow
Open door.
Connect the uroflow cable to the uroflow
stand and the adapter module.
EMG
Place EMG preamplifier in the holder left or
right of the basic unit.
Connect the EMG cable to the the adap-
ter module.
Close door.
12 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Technical Details
Parts List
# Art. No. Description Qty.
ANDROMEDA 03.11.2016 13
Technical Details HELIX – Service Manual
8
3
10 9
1
11
5
7
14 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
ANDROMEDA 03.11.2016 15
Technical Details HELIX – Service Manual
15
14 13
12
11
10
9 7
16
8
4
5
1
16 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
ANDROMEDA 03.11.2016 17
Technical Details HELIX – Service Manual
10
7
6
2
1
18 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
8 HMB00345 Lever 1
9 HMB00346 Arm 1
ANDROMEDA 03.11.2016 19
Technical Details HELIX – Service Manual
17
18
20
15
16
14
19
4
3
6 1
10
13
11
12
9
20 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
ANDROMEDA 03.11.2016 21
Technical Details HELIX – Service Manual
22 03.11.2016 ANDROMEDA
HELIX – Service Manual Basic Unit PCB Technical Details
23 03.11.2016 ANDROMEDA
Technical Details HELIX – Service Manual
24 03.11.2016 ANDROMEDA
HELIX – Service Manual Basic Unit PCB Technical Details
25 03.11.2016 ANDROMEDA
Technical Details HELIX – Service Manual
26 03.11.2016 ANDROMEDA
HELIX – Service Manual Basic Unit PCB Technical Details
27 03.11.2016 ANDROMEDA
Technical Details HELIX – Service Manual
28 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Stand (E30502)
Parts List
# Art. No. Description Qty.
6 HMB00396 Foot 4
11 HMB00431 Bed-plate 1
ANDROMEDA 03.11.2016 29
Technical Details HELIX – Service Manual
12
14
15 13
11
6 4
5
10
8
16
9
7
30 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
4 HMB00401 Door 1
ANDROMEDA 03.11.2016 31
Technical Details HELIX – Service Manual
14
13
8
7
9
9
10
9
11 12
17
16
4
15
5
6
2
3
18
1
32 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
ANDROMEDA 03.11.2016 33
Technical Details HELIX – Service Manual
34 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
ANDROMEDA 03.11.2016 35
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
36 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
46 HMB00420 Housing 1
Adjustments
The over-bend protection plate should be adjusted to block the load cell at 6kg load.
ANDROMEDA 03.11.2016 37
HELIX – Service Manual Technical Details
21
20
18
19
17
23 24
22
14
15
13
16
12
11
10
9
8 6
5
4
7
38 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
47
46
45
44
43
42
41
41
37 39
38
36
34
35
52
25
27
30
26
31 30
48
28
51
29
50 33
49
ANDROMEDA 03.11.2016 39
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
Adjustment
Adjust potentiometer to an output voltage (on soldering bridge) off around -1V at zero
load (0kg).
Functional Test
The output voltage should increase 0.5V / Kg load:
1kg: 0.5V above adjusted zero point.
2kg: 1.0V above adjusted zero point.
3kg: 1.5V above adjusted zero point.
4kg: 2.0V above adjusted zero point.
5kg: 2.5V above adjusted zero point.
40 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
41 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
42 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
43 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
16 HMB00213 Housing 1
44 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
17
16
15
14 13
12
11
10
5
19
3
6
2
4
1
18
ANDROMEDA 03.11.2016 45
HELIX – Service Manual Technical Details
This part is not sensitive to electrostatic discharges and may be handled without ESD
protection!
46 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
ANDROMEDA 03.11.2016 47
HELIX – Service Manual Technical Details
48 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
7 HMB00240 Receptacle 1
8 HMB00242 Funnel 1
10 HGE00027 Arm 1
ANDROMEDA 03.11.2016 49
HELIX – Service Manual Technical Details
32 HMB00297 Brake 1
33 HMN00144 Spring DIN17223, ends grinded and applied, d1, D7,3, L10 1
35 HMB00234 Housing 1
50 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
4
9
7
1
2
3
ANDROMEDA 03.11.2016 51
HELIX – Service Manual Technical Details
34
16
33
32
18
17
19
29
20 28
22 27
21 31
11
30
25
35
36
10 26
14
12
23 15
13
24
52 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
The soldering bridge “FLO” must be short-cut. Do not short-cut “VOL”.
Adjustment
Adjust potentiometer to an output voltage (on soldering bridge) off around -1V at zero
load (0kg).
Functional Test
The output voltage should increase 1.2V / Kg load:
1kg: 1.2V above adjusted zero point.
2kg: 2.4V above adjusted zero point.
ANDROMEDA 03.11.2016 53
HELIX – Service Manual Technical Details
54 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
ANDROMEDA 03.11.2016 55
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
6 HMB00317 Arm 1
56 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
2
20 19
1
4
7
8
17
18
16
9
11
15 10 12
13
14
ANDROMEDA 03.11.2016 57
HELIX – Service Manual Technical Details
Parts List
# Art. No. Description Qty.
1 HMB00355 Housing 1
13 HMB00356 Cover 1
58 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
14
13
12
10
11 9
8
6 7
2
B
G
R
ANDROMEDA 03.11.2016 59
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
Attention! Do not adjust potentiometers P1 and P2 for common mode rejection without
using the proper equipment. The common mode rejection adjustment re-
quires special equipment only owned by ANDROMEDA.
If the PCB or the internal cables are exchanged or manipulated the Amplifier
must be common mode adjusted by ANDROMEDA.
Notice The amplifier PCB contains an isolation barrier for patient safety. Do not
short-cut primary and secondary sides.
60 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
61 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
62 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
PC Interface (HGE00045)
Parts List
# Art. No. Description Qty.
2 HEB00149 PCB 1
6 HMN00151 Threaded bolt UNC4-40 / UNC 4-40 with nut and wascher 2
ANDROMEDA 03.11.2016 63
HELIX – Service Manual Technical Details
64 03.11.2016 ANDROMEDA
7
HELIX – Service Manual Technical Details
Attention! This part is high sensitive to electrostatic discharges and may only be han-
dled using proper ESD protection!
Notice The interface PCB contains an isolation barrier for patient safety. Do not
short-cut primary and secondary sides.
ANDROMEDA 03.11.2016 65
HELIX – Service Manual Technical Details
66 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Details
67 03.11.2016 ANDROMEDA
HELIX – Service Manual Calibration and test
Fig. 3: Calibration
Uroflow transducer
Configuration
HELIX
Uroflow transducer with stand
1 kg weight.
Calibration
68 03.11.2016 ANDROMEDA
HELIX – Service Manual Calibration and test
Test
Start an uroflow measurement.
The display must indicate 1000 ml. The error should be less than 1 %.
In order to measure the linearity, repeat the test procedure (from zero setting) with
500 g or 500 ml.
The deviation should be less than 1 %.
Note: ANDROMEDA states a relative error of less than 1 %. A properly calibrated device,
however, offers much greater accuracy (rel. error less than 0.2 %).
ANDROMEDA 03.11.2016 69
HELIX – Service Manual Calibration and test
Volume transducer
Configuration
HELIX
Volume transducer at the stand, connected to Vinf
1-kg weight with wire eye
Calibration
Select Vinf.
70 03.11.2016 ANDROMEDA
HELIX – Service Manual Calibration and test
Test
Start a cystometry measurement.
Set Vinf at zero with a load of 1 kg or 1000 ml.
Remove the 1 kg weight.
The error should be less than 1 % (at 1000 ml ±10 ml).
To measure the linearity, repeat the test procedure (from zero setting) with 3 kg or
3000 ml.
The error should be less than 1 % (at 3000 ml less than ±30 ml).
Note: ANDROMEDA states a relative error of less than 1 %. A properly calibrated device,
however, offers much greater accuracy (rel. error less than 0.2 %).
Pump
Configuration
HELIX
A bottle or a bag with water connected to a pump tube set.
A measurement beaker; for calibration without a volume transducer the beaker should
have very accurate level marks.
ANDROMEDA 03.11.2016 71
HELIX – Service Manual Calibration and test
if you
1. calibrate with the most-used infusion rate,
2. calibrate using the same water bottle (or bag) that you always use.
Note: You will achieve the greatest accuracy for the infusion rate if you
1. calibrate with the most-used infusion rate,
2. calibrate using the same water bottle (or bag) that you always use and
3. if the markings on the measurement beaker are very precise.
72 03.11.2016 ANDROMEDA
HELIX – Service Manual Calibration and test
Test
Start a cystometry measurement.
Set Vinf at zero with the bottle or bag attached.
Set the infusion rate at 100 ml.
Start the pump and let it run for 2 minutes. Then stop the pump.
Check the Vinf- value in the display.
The relative error should be less than 5 % (200 ml ±10 ml).
Note: ANDROMEDA states a relative error of less than 5 %. A properly calibrated device,
however, offers much greater accuracy (rel. error less than 0.2 %).
Pressure transducer
Configuration
HELIX
A bottle or bag with water, connected to a transducer tube set
A pressure transducer connected to the transducer tube set
A pressure transmission tube connected to the pressure transducer
A ruler (20 cm) to measure the height of the water column
ANDROMEDA 03.11.2016 73
HELIX – Service Manual Calibration and test
Calibration
The calibration procedure is shown for one pressure channel (Pves). All of the other
pressure channels are calibrated in the same way. The following pressure channels are
allocated to the measurement channels in the calibration menu:
74 03.11.2016 ANDROMEDA
HELIX – Service Manual Calibration and test
Test
An example of the test procedure is given here for one channel (Pves). The other pres-
sure channels are tested in the same way. The following table shows which pressure
channels are used in which measurement programs.
Note: ANDROMEDA states a relative error of less than 1 %. A properly calibrated device,
however, offers much greater accuracy (rel. error less than 0.2 %).
ANDROMEDA 03.11.2016 75
HELIX – Service Manual Maintenance Work Specification
• Receptacle, Housing
Uroflow stand:
• Funnel, Connectors and flow cable, Mechanical high adjustment and fixation
EMG Preamplifier:
• Test measurement - if the actual value is more than 500g inaccurate the trans-
ducer might be defective and must be serviced.
• Test measurement - if the actual value is more than 2g inaccurate the transducer
might be defective and must be serviced.
EMG pre-amplifier:
• Short-cut all three electrode inputs – if the ground noise is higher than 5µV the
pre-amplifier might be defective and must be serviced.
• Generate test signal using the hand - if the signal peak is less than 100µV the
pre-amplifier might be defective and must be serviced.
76 03.11.2016 ANDROMEDA
HELIX – Service Manual Maintenance Work Specification
• Housing, Connectors
Pump and Volume Module:
• Receptacle, Housing
Uroflow stand (optional):
• Funnel, Connectors and flow cable, Mechanical high adjustment and fixation
EMG Preamplifier (optional):
• Test measurement - if the actual value is more than 500g inaccurate the trans-
ducer might be defective and must be serviced.
• Test measurement - if the actual value is more than 2g inaccurate the transducer
might be defective and must be serviced.
Infused Volume transducer:
• Test measurement using 3kg - if the actual value is more than 1000g inaccurate
the transducer might be defective and must be serviced.
ANDROMEDA 03.11.2016 77
HELIX – Service Manual Maintenance Work Specification
• Test measurement using 3kg - if the actual value is more than 10g inaccurate the
transducer might be defective and must be serviced.
Filling Pump:
Attention! To test the pump, use always a new pump tube.
• Test output pressure using a pump tube connected to a manometer. If the pres-
sure is higher than 2.5 bars or lower than 1.5 bar then the pump mechanics might
be defective.
• Test measurement – let the pump run for 2 minutes at 50 ml/min. If the actual
value is more 40ml inaccurate the pump mechanics might be defective and must
be serviced.
• Calibrate letting the pump run for 2 minutes and the “Vinf automatic” function for
most accurate result.
• Test measurement - if the actual value is more than 5ml inaccurate the pump
mechanics might be defective and must be serviced.
Pressures:
Attention! To test the pressures, use always new pressure transducers.
• Test measurement - if the actual value is more than 10 cm/H2O inaccurate the
device might be defective and must be serviced.
• Calibrate using 20cm water column for accurate result or 100cm for even better.
• Test measurement - if the actual value is more than 3 cm/H2O inaccurate the de-
vice might be defective and must be serviced.
EMG pre-amplifier (optional):
• Short-cut all three electrode inputs – if the ground noise is higher than 5µV the
pre-amplifier might be defective and must be serviced.
• Generate test signal using the hand - if the signal peak is less than 100µV the
pre-amplifier might be defective and must be serviced.
• Test measurement – let the puller run for 1 minute at 1 mm/sec. If the actual
value is more 10mm inaccurate the pump mechanics might be defective and
must be serviced.
• Calibrate by letting the puller run at a speed of 1mm/sec. for 100mm for most ac-
curate result.
• Test measurement - if the actual value is more than 10mm inaccurate the puller
mechanics might be defective and must be serviced.
78 03.11.2016 ANDROMEDA
HELIX – Service Manual Electrical Safety Test
Needed tools
• Insulation pad (only EMG test)
• ANDROMEDA – Test cable HKB30020
• ANDROMEDA – Test cable HKB30030
• Safety tester according to EN 62353
Preparation
EMG pre-amplifiers with serial numbers less than 08110401 must be opened before
testing.
In the case that the two red marked circuits are not present, then the pre-amplifier must
be modified. This is done by ANDROMEDA free of charge. Please continue the test
without EMG or suspend the test until the modified pre-amplifier is returned from
ANDROMEDA.
ANDROMEDA 03.11.2016 79
HELIX – Service Manual Electrical Safety Test
Test configuration
1. Attach the Helix mains power cable to the safety tester.
2. Attach the two test cables HKB30020 and HKB30030 to the safety tester. The
EMG pre-amplifier must during the entire test stay attached to the Helix. Attach
the test cable to the electrode inputs and the applied part connector of the safety
tester.
Attention:
3. Attach the test cable to the transducer inputs of the Helix and the applied part
connector of the safety tester.
80 03.11.2016 ANDROMEDA
HELIX – Service Manual Electrical Safety Test
4. Attach the safety tester protective earth pin to the housing of the Helix. Use a
mounting screw under the Helix Basic Unit.
Attention! The Helix must be ON during the leakage current tests. If the safety
tester turns OFF the device then press the ON button on the Helix key pad.
ANDROMEDA 03.11.2016 81
HELIX – Service Manual Technical Specifications
Technical Specifications
Signal processing 16 Bit, digital filters, artefact detection
Measurement pro- User-defined, unlimited quantity
grams
Analyses Uroflow, cystometry, rest and stress profilometry, pressure/flow,
ICS nomograph, chess, linear PURR, CLP, VLPP
Measurement unit
Safety
82 03.11.2016 ANDROMEDA
HELIX – Service Manual Technical Specifications
Power supply
Voltage 100–240 V
Mains frequency 50–60 Hz
Power consumption 30 W
Fuses 2 × T1AH/250 V
Stand-by mode <1 W
Dimensions
ANDROMEDA 03.11.2016 83
HELIX – Service Manual Technical Specifications
84 03.11.2016 ANDROMEDA