TECHNICAL INFORMATION

PATIENT MONITOR

Type : PHILIPS INTELLIVUE MX700 (STD+IBP+EtCO2+PiCCO)


Parameter : ECG, NIBP, SpO2, Temp, HR/RR, IBP, etCO2 PICCO
Brand : Philips
Country of Origin : Germany

Physical
Weight : 11 kg (24.3 lb)
Size : 393 x 321 x 233 mm

Power Supply Specification


Power Consumption : <200 W average
Line Voltage : 100 to 240 V
Current : 1.9 to 0.9 A
Frequency : 50/60 Hz

Monitor Performance
Display : 15” color LCD (TFT)
Resolution : 1280 x 768 (WXGA)
Pixel size : 0.261 mm x 0.261 mm
Waveforms : standard 8, optional 12

Indicators/Sounds
Audible feedback for user input
Prompt tone
QRS tone, or SpO2 modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired

Trends
Resolution : MX700 supports up to 48h for up to 24 or 32 numerics
Multiple choices of number of numerics, resolution and duration depending on trend option
Informations :
and application area.

Events
Information :
Trigger condition and time, event classification and associated detailed view of episode data
Capacity : 25 or 50 events for either 8 or 24 hours

IntelliVue X2 Module (ECG, HR, RR, NIBP, SpO2, T/IBP)


Display : Touchscreen, 3.5 in TFT flat panel display with QVGA
Weight : 1.2 kg (2.6 lb)

Power Specification
Power consumption <12 W average
<30 W while battery is loading
Operating Voltage 36 to 60 V DC floating
Battery Specifications
Operating Time : Basic monitor configuration: 2.5 hours (with new, fully charged battery at 25°C)
Charge Time When X2 is off: 2 h.
When X2 is in use and connected to monitor,
Without extensions: 12 h approx.
When X2 is in use and connected to the external power: 12 h approx.

SpO2 Performance
Sensor : Wavelength range: 500 – 1000 nm
Emitted Light Energy: ≤15 mW
Pulse rate range : 30 to 300 bpm
Resolution : 1 bpm
Accuracy : ±2% or 1 bpm, whichever is greater

NIBP performance
Measurement Ranges : Systolic --> Adult: 30 to 270 mmHg (4 to 36 kPa)
Diastolic --> Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Mean --> Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pulse Rate --> Adult: 40 to 300 bpm
Accuracy : Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Measurement Accuracy : 40 to 100 bpm: ±5 bpm
Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
Typical for neonatal cuff: Less than 2 seconds
Auto Mode Repetition Times : 1 min, 2 min, 2.5 min, 3 min, 10 min, 15 min, 20 min, 30 min, 45 min, 1h, 2h,
4h, 8h, 12h, 24h
ECG performance
Range : Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Note: for rates equal to or less than 15 bpm, the displayed heart rate is 0
Accuracy : ±1% of range
Resolution : 1 bpm
Sensitivity : ≥200 μVpeak

Respiration performance
Range : Adult/pedi: 0 to 120 rpm
Neo: 0 to 170 rpm
Accuracy : at 0 to 120 rpm ±1 rpm
at 120 to 170 rpm ±2 rpm
Resolution : 1 rpm
Accuracy : ±1 rpm to ±2 rpm
Bandwith : 0.3 to 2.5 Hz (-6 dB)
Noise : Less than 25 mΩ (rms) referred to the input

Temperature performance
Range : -1 to 45 C ( 30 to 113 F )
Resolution : 0.1 C
Accuracy : 0.1 C
Average time constant : less than 10 second
IBP Performance
Range : -40 to 360 mmHg
Pulse Rate : Range : 25 to 350 bpm
Accuracy : ±1% Full Range
Resolution : 1 bpm
Sensitivity : 5 μV/V/mmHg (37.5 μV/V/kPa)
Frequency Response : dc to 12.5 Hz or 40 Hz
Zero Adjustment : Range : ±200 mmHg (±26 kPa)
Accuracy : ±1 mmHg (±0.1 kPa)
Drift : Less than 0.1 mmHg/°C (0.013 kPa/°C)
Transducer : Load Impedance : 200 to 2000 Ω (resistive)
Output Impedance : ≤3000 Ω (resistive)

etCO2 – Mainstream Performance


Measurement Range : 0 to 150 mmHg (0 to 20.0 kPa)
Accuracy : Up to 5 minutes during warmup: ±4 mmHg or 12%, whichever is greater
After 5 minutes warmup:
0 to 40 mmHg (0 to 5.3 kPa): ±2.2 mmHg (±0.3 kPa)
Above 40 mmHg (5.3 kPa): ±(5.5% + {0.08% per mmHg above 40 mmHg}) of reading
Resolution : Numeric: 1.0 mmHg (0.1 kPa)
: Wave: 0.1 mmHg (0.01 kPa)
Stability Included in Accuracy specifications

PiCCO Performance
Cardiac Output – Right Heart
Range : 0.1 to 20.0 l/min
C.O. instrument accuracy : +/- 3% or 0.1 l/min
C.O. repeatability accuracy : +/- 2% or 0.1 l/min

Cardiac Output – Transpulmonary


C.O. Range : 0.1 to 25.0 l/min
C.O. instrument accuracy : + 4% or 0.15 l/min
C.O. repeatability accuracy : + 2% or 0.1 l/min

EVLW (Extra-vascular Lung Water)


Range : 10 to 5000 ml
Standard Deviation : 10% or 1 ml/kg

ITBV (Intrathoracic Blood Volume)


Range : 50 to 6000 ml
Accuracy : + 10% or 30 ml
Repeatability : + 5% or 20 ml

CCO (Continuos Cardiac Output)


Range : 0.1 to 25.0 l/min
Standard Deviation : 10% or 0.3 l/min
Display Update : 2 seconds nominal

Accessories Included
ECGTrunk cable 5 leads : 1 pcs
ECG electrode disposable : 1 pack
NIBP Tubing : 1 pcs
NIBP Cuff Adult : 1 pcs
Reusable Spo2 sensor adult : 1 pcs
Temperature probe : 1 pcs
IBP Adapter Cable : 2 pcs
IBP Transducer : 1 pack
Transducer Mount, reusable : 1 pcs
Transducer Holder for IV pole, reusable : 1 pcs
etCO2 Patient Adapter : 1 pack
Long Cardiac Output Trunk Cable : 1 pcs
PiCCO Temperature Probe (1.5m) : 1 pcs
Short Baxter CO-Set Temperature Probe : 1 pcs