Sunteți pe pagina 1din 6

NDT&E International 46 (2012) 122–127

Contents lists available at SciVerse ScienceDirect

NDT&E International
journal homepage: www.elsevier.com/locate/ndteint

Characterisation of voids in fibre reinforced composite materials


John Eric Little a, Xiaowen Yuan a,n, Mark Ian Jones b
a
Department of Mechanical Engineering, The University of Auckland, Auckland, New Zealand
b
Department of Chemical and Materials Engineering, The University of Auckland, Auckland, New Zealand

a r t i c l e i n f o a b s t r a c t

Article history: Voids and porosity are critical imperfections in fibre reinforced composite materials. The aim of this
Received 27 July 2011 study is to assess the comparative accuracy and reliability of conventional and novel void character-
Received in revised form isation techniques for analysing voids within carbon fibre reinforced composite (CFRC) laminates.
29 November 2011
While microscopy and Archimedes density measurements can only give results of limited accuracy and
Accepted 30 November 2011
reliability, micro-computed tomography (micro-CT) analysis was found to have no significant inherent
Available online 8 December 2011
errors and is able to characterise the three-dimensional size, shape and distribution of voids in CFRC
Keywords: laminates. The accuracy of micro-CT was also verified by analysing artificial voids of a known size and
Voids shape disbursed throughout a CFRC laminate sample.
Porosity
& 2011 Elsevier Ltd. All rights reserved.
Micro-CT
Fibre reinforced composites

1. Introduction the laminate. It is not uncommon for this method to calculate a


negative void fraction, which is theoretically impossible [4].
Advanced composite materials are becoming increasingly used Matrix burn-off: the Matrix burn-off method utilises a furnace
in the critical structures of both civilian and military aircraft, as to degrade and remove the laminate’s matrix from around the
they often have a higher strength-to-weight ratio compared to fibres, which allows for the fibre weight fraction to be deter-
conventional materials, are corrosion resistant and are able to be mined. Fibre degradation and convection currents affect the
tailored to meet specific mechanical criteria [1]. Due to the non- accuracy of this technique, as fibre mass can be degraded or
homogeneous nature of these materials, problems can arise removed. This error is especially prevalent for CFRC laminates, as
during their manufacture and handling which result in defects, the temperatures reached when achieving matrix burn are above
some of which can be hidden or uncharacterised [1]. Void carbon fibre oxidation [5]. Archimedes density testing is initially
content/porosity, delamination, matrix cracks and impact damage required to measure the samples actual density [3], with accurate
are just some of the common composite material defects. Identi- knowledge of the constituent densities still required to calculate
fication and characterisation of these defects are important for the samples void volume fraction.
determining the ‘quality’ of a composite structure, as these Matrix digestion: similar to matrix burn-off, but instead of
parameters have a significant effect on the mechanical properties using thermal degradation, acid is used to digest and remove the
of the material and can act as failure initiation points [2]. matrix from around the fibres. Toxic fumes are produced during
Conventional composite void characterisation techniques provide the chemical break down of the matrix [6], making this process
results of limited accuracy and/or reliability due to inherent dangerous.
testing errors. Microscopy: this is a visual and, if used to characterise the
Conventional void characterisation techniques include: interior of the samples, a destructive analysis technique, which
Archimedes Theoretical versus Actual Density: this method allows the void size, shape and distribution to be analysed [7].
uses the Archimedes principle to calculate the samples actual Microscopy is widely used in industry due to its visual character-
density, and then by using the theoretical density of the sample isation advantage, but also has an inherent section-bias error
with no voids present, the void volume fraction is able to be associated with analysing two dimensional (2D) cross-sections of
calculated [3]. This technique however relies on accurate knowl- the sample.
edge of the fibre and matrix density and weight fraction values of The conventional void analysis techniques mentioned are not
only limited in their accuracy and reliability, but also have
limitations with regards to what their data can conclude. The
non-visual techniques (theoretical density vs. actual density,
n
Corresponding author. Tel.: þ64 9 923 9535; fax: þ64 9 3737479. matrix burn-off and matrix digestion) can only measure total
E-mail address: xw.yuan@auckland.ac.nz (X.W. Yuan). void fraction, and cannot analyse individual void size, shape and

0963-8695/$ - see front matter & 2011 Elsevier Ltd. All rights reserved.
doi:10.1016/j.ndteint.2011.11.011
J.E. Little et al. / NDT&E International 46 (2012) 122–127 123

distribution. It has been proven that two laminates that show the W coupon
rcoupon ¼ ð2Þ
same void content can have a varying mechanical response [8], V coupon
which highlights the important of visual analysis. Whilst micro-
The void volume fraction (VV) is able to be calculated using
scopy can detect this important visual information, it is however
Eq. (3), with the mass fraction (%m) and density (r) of the primary
limited by the information being 2D and non-representative of
constituents required. It is critical that the material properties are
the whole sample. Unless multiple images are obtained on very
accurately known, as minor variations can significantly affect the
thin slices, it also has potential for missing any pore intercon-
accuracy of the calculated void volume fraction.
nectvity between adjacent slices.
!
X-ray computed tomography is a novel non-destructive and %mmatrix %mf ibre
visual technique that can be used for three dimensional (3D) V v ¼ 100rcoupon þ ð3Þ
rmatrix rf ibre
analysis of internal structures [1,9]. Computed tomography was
conceived in the early 1960s and has been predominately used for In order to gain this visual information a sample cross-section
medical diagnosis [1]. Over the past 20 years, however, micro- is required, making this technique destructive. The testing pre-
computed tomography (micro-CT) has been developed and allows paration required for microscopy analysis of composite materials
for wider application, including the qualitative and quantitative is a time consuming and sometimes difficult process due to the
non-destructive evaluation of composite materials [1]. non-homogenous composition of composites. Microscopy pre-
Micro-CT imaging works by recording the change in attenua- paration requires the sample to be set in resin and then a flat
tion along the path of an X-ray beam fired through a sample [10]. cross-section prepared to a mirror like finish, with no visible
A cross-sectional image is then mathematically reconstructed preparation damage, such as scratches and void distortion. The
from attenuation measurements made at different angles by the high quality surface finish is achieved by sequentially grinding
detectors [10]. Each 2D slice is of a known thickness, which allows the surface with smaller sized particle paper, from 35 to 2.5 mm.
the images to be virtually stacked to create a 3D model. The image This process is difficult, as the brittle fibres regularly break free
stack can then be converted to binary (black and white), allowing and gouge large scratches in the specimen surface. It is very
for the voids to be separated from the fibre and matrix which important that the finished surface has little or no scratches,
make up the laminate. as these will be counted as a void when using image analysis
This research paper compares the aforementioned conven- software.
tional techniques to micro-CT analysis for analysing voids in CFRC The numerical analysis of microscopy images is carried out
materials. Matrix digestion and matrix burn-off were not tested, using image analysis software, which works by analysing the
as they have significant errors and/or difficulties when dealing individual pixels that make up the image cross-section. The image
with carbon fibre reinforced plastic laminates. analysis process requires the microscopy cross-section image to
be converted to binary, with the voids remaining black and the
matter (fibre and matrix) converted to white. The image is then
2. Experimental procedures ‘cleaned’, to remove unwanted noise and outliers [12]. This
conversion and cleaning process is susceptible to error only if
To compare the various composite characterisation techniques, care and common sense are not applied. The image is then able to
two 24-ply oven cured carbon fibre reinforced composite (CFRC) be analysed using built in analysis tools within the software,
laminates were manufactured. One panel was made from unidirec- which is able to calculate the total number of voids, their overall
tional (UD) preimpregnated (prepreg) carbon fibre, while the other area fraction, plus analyse individual void size, shape and two
was a twill weave carbon fibre prepreg. The laminates were oven dimensional location.
cured under vacuum to ensure that voids were present within the The micro-CT scanning and image reconstruction was conducted
laminates for analysis. Samples measuring 15 by 7 by 5 mm were at the Auckland Bioengineering Institute. The micro-CT scanning
cut from the middle of the panels for testing. The size of the reconstruction parameters were kept constant for all samples, while
specimens was a compromise between the different techniques’ the alignment parameter was tailored for increased accuracy. Micro-
requirements, with micro-CT requiring small samples to reduce data CT scanning requires little specimen preparation, as this technique is
file size. Carbon fibre epoxy laminates were used for the samples as non-destructive with respect to the sample specimen. The scanning,
characterisation interest primarily lies with this material. image reconstruction and the data acquisition processes can be time
The determination of the void volume fraction using the consuming steps and vary with respect to the sample size and the
Archimedes Theoretical versus Actual Density technique was micro-CT scan settings, most notably the scanning resolution. A
carried out in accordance with ASTM D3171. Due to size restric- constant scanning resolution of 7 mm was used for each sample and
tions of the micro-CT technique, the samples were smaller than represents the physical dimension of the voxels (3D pixel) making
the recommended testing standard sample volume of 1 cm3. This up the image slices.
variation is necessary and is assumed to have negligible impact The micro-CT data acquisition process was completed using
on the accuracy of the Archimedes results. software provided by Skyscan, the micro-CT scanner’s manufac-
The testing procedure requires the density of the sample to be turer. This software has the ability to be able to calculate overall
measure using Archimedes’s Principle, where the specimen is void fraction and the individual void size, shape and distribution
weighed in air (Wspecimen) and then weighed while submerged in three dimensions (3D), while also being able to produce 3D
(Wsubmerged). It is critical that no air bubbles are present on the rendered models of the voids.
coupon while submerged, as this will decrease the measured
mass and thus inaccurately increase the measured void volume
[11]. The coupon volume (Vcoupon) can be calculated in conjunc- 3. Results and discussion
tion with the submerging liquid density (Eq. (1)), which in this
case is distilled water (rwater). The actual density of the sample The following sections concentrate on the characterisation
specimen is then calculated using Eq. (2). techniques themselves, while using the sample laminates as an
example of the capabilities and comparative accuracy of each
W coupon W submerged technique. The overall void volume fraction calculated for each
V coupon ¼ ð1Þ
rwater technique is compared in a later section.
124 J.E. Little et al. / NDT&E International 46 (2012) 122–127

3.1. Archimedes Theoretical versus Actual Density 3.2. Microscopy

The Archimedes Theoretical versus Actual Density testing There are various kinds of microscopes that can be used for
technique is a non-destructive and non-visual technique that imaging material cross-sections, with optical and environmental
is able to calculate the overall void volume fraction within a scanning electron microscopy (ESEM) the most common for void
specimen, providing accurate material properties and mass frac- analysis [13].
tion values are known and the testing procedure is accurately Fig. 1 shows the same area of a twill weave laminate cross-
performed. Since this technique is non-visual, the only quantita- section under both optical and ESEM microscopy. The optical image
tive data ascertained is the total void volume fraction of the (Fig. 1a) shows the fibres as an illuminated white, the matrix as grey
sample. and the voids black. While the ESEM image (Fig. 1b) shows the fibres
This Archimedes based characterisation technique requires as a light grey, the matrix as a slightly darker grey and the voids
accurate knowledge of the material’s physical properties, includ- black. The most notable and critical difference between the two
ing both constituents’ densities and mass fractions. The necessity microscopy images is the superior contrast between void and matter
of having to know physical properties and make-up of the (fibre and matrix) of the ESEM image in comparison with the optical
laminate is a considerable weakness of the Archimedes based image. When utilising image analysis software, it is important that
testing procedures, and limits it to samples of known material the contrast between void and matter is distinguishable so that an
properties. Small variations in densities or the mass fractions can accurate binary conversion can occur for image analysis purposes.
lead to noticeable variations in the overall result, which can lead An example of this contrast difference can be seen around the large
to calculating the impossibility of a negative void fraction. circular void in Fig. 1, where the optical image shows a dark and less
The material properties, fibre and matrix density, of both well defined area to the right of the void. Optical microscopy,
laminates were acquired from the carbon fibre prepreg manufac- however, has a greater contrast between fibre and matrix, which
turer. Due to cure variables the exact matrix density is unknown, is beneficial if the primary constituents are being examined. Another
with a provided range between 1.15 and 1.18 g/cm3. Due to this benefit of optical microscopy is the financial investment required, as
variation only a range of the sample’s void volume fraction could optical microscopes are generally much cheaper to own, operate or
be calculated (Eq. (3)), with the exact value remaining unknown. lease. In this study, an ESEM microscope was solely used for the
The pre-cure weight fractions were acquired from the material cross-sectional analysis, since it is the preeminent microscopy
data sheets that are produced for each batch of prepreg material technique for the imaging of voids in CFRC laminates.
by the manufacturer, with the final values taking into considera- An advantage of ESEM is the capability to easily and relatively
tion the resin bleed during cure. The complications of having to efficiently image the entire specimen cross-section. Since an ESEM
accurately know the material properties is an inherent error of detector is computerized, it is able to incrementally change location
Archimedes based characterisation techniques (including matrix during imaging, which allows the whole cross-sectional surface to
digestion and burn-off), which limit their effectiveness and use to be systematically imaged. The sequential images are able to be
known materials. meshed together to create a complete image of the specimen cross-
In some cases, however, the Archimedes Theoretical versus section (Fig. 2). The cross-sections for this study, 15 by 5 mm,
Actual Density technique is more suitable than others, as visual required twenty one images (Fig. 2) at 100  magnification.
information is not always, at least initially, required. This techni- Fig. 3 is an example of the numerical analysis that is able to be
que is suited for quality control purposes, as it is able to handle performed using image analysis software. This graph shows the
relatively large samples and is less time consuming than other
techniques. An example application of this technique is the
inspection of autoclaved cure products for manufacturing defects,
such as bag rupture and corner bridging. These particular man-
ufacturing defects generally cause significant increases in the
overall void volume fraction, so providing the material properties
and constituent volume fractions are within reasonable accuracy
these faults are able to be identified using this technique. If a fault
is found the cause of the increased void volume fraction is then
able to be investigated further, using more time consuming and
destructive visual techniques, or simply, the specimen can be
discarded. Fig. 2. An example of ESEM microscopy cross-section.

Fig. 1. (a) Optical image and (b) SEM image of the same area of a twill weave CFRC laminate cross-section.
J.E. Little et al. / NDT&E International 46 (2012) 122–127 125

cumulative number of voids and their cumulative area fraction Microscopy analysis allows important individual void charac-
plotted against void projected area diameter for the unidirec- teristics to be assessed. Fig. 3 for example shows that E45% of
tional laminate cross-section shown (Fig. 2). the total number of voids with a projected area diameter between
To limit the effects of section-bias three cross-sections were 15 and 200 mm have a projected area diameter of less than 50 mm,
analysed per sample, making a total of fifteen cross-sections per while only accounting for E6% of the total void area fraction. On
laminate. This large number makes the analysis process very time the other hand, this graph shows that voids with a projected area
consuming, with each surface requiring careful preparation to a diameter greater than or equal to 200 mm account for only 1.5% of
mirror like finish. the total number of voids greater than 15 mm, but contribute to
E24% of the overall void area fraction. This example highlights
the benefits of visual analysis, where the individual character-
istics of the voids are able to be evaluated against each other.
As can be seen in Fig. 2, the voids observed within a single
cross section of a laminate are not consistent, but have variable
shape and distribution due to the section cutting through pores at
different orientations. Due to this fact, microscopic characterisa-
tion of composite laminates has an inherent section bias, where
one cross-section is not necessarily an accurate representation of
the complete sample. In order to limit the effects of section bias, it
is important that multiple cross-sections are analysed, which is a
time consuming and costly process. It is also difficult when
obtaining multiple sections of the same sample to know exactly
where those sections are being taken from within the sample.

3.3. Micro-CT

Fig. 3. The void number and voidarea cumulative distribution curves from the The limitation of current micro-CT technology means that only
ESEM cross-section in Fig. 2. small samples are able to be scanned. Small increases in the

Fig. 4. Orthographic snapshots of a 3D unidirectional model from micro-CT analysis, with (a) top, (b) 3D model and (c) front and (d) side.
126 J.E. Little et al. / NDT&E International 46 (2012) 122–127

sample size or scanning resolution significantly increases the


required scanning time, data file size and the required data
acquisition computer power. The samples for this study required
a scanning time of 1 h at a scanning resolution of 7 mm, which
produced a file size of 10 gigabytes.
An advantage of micro-CT analysis is the ability to view 3D
rendered models of the internal structure of a sample. Fig. 4(acd)
shows an orthographic projection of the 3D unidirectional lami-
nate model (Fig. 4b). Three dimensional visual analysis of the
internal laminate structure provides the user with a basic under-
standing of what is present within the laminate. In these images,
it is possible to see large elongated voids, while the small rounder
voids are scatted throughout the volume. This non-analytical
Fig. 6. (a) Micro-spheres under optical microscopy and (b) 3D micro-CT render of
understanding of the voids present within the sample can then scanned microsphere.
be used to help direct the analytical analysis.
Three dimensional analytical analysis of micro-CT scans can
also be carried out using the supplied software. This software uncontrollable and independently unknown. In order to validate
works in a similar manner to the image analysis software used the accuracy of the micro-CT analysis of voids in CFRC laminates a
during microscopy, but takes into account the through slice verification laminate was manufactured with artificial voids of a
thickness of the pixels and the interconnectivity of the voids known size and shape. The artificial voids were manufactured
between cross-sectional images. into a 12-ply CFRC twill weave laminate using hollow glass micro-
A scanning resolution of 7 mm was deemed to suitable for the spheres manufactured by 3 M. These micro-spheres were classi-
analysis of voids in CFRC laminates. Although there is likely to be fied using sieves into a size range 75–80 mm. Under microscopic
a very large number of voids within the laminate that have a analysis the spheres were observed to have an average outer
projected area smaller than 7 mm, microscopy (Fig. 3), however, diameter of 68.5–83 mm (Fig. 6a), which includes a 6% sieve error
has shown that their combined area fraction and influence on the and takes into consideration an average shell thickness of 1 mm.
quality of the laminate is negligible, whereby around 20% of the The micro-CT verification laminate was scanned and recon-
voids had a projected area diameter below 20 mm but the area of structed with the same settings used for the sample laminates.
all of these pores contributed less than 0.5% to the total pore area. Fig. 6b shows a 3D rendered model of a glass micro-sphere, which
Fig. 5 shows the average sphericity of individual voids with is shown to be approximately spherical in shape. The volume and
respect to their projected area diameter for the unidirectional surface area of fifty micro-spheres were individually calculated
laminate sample from analytical micro-CT analysis. This graph using micro-CT analysis software. This analysis calculated an
shows a linear decrease in sphericity as the projected area average projected area diameter for the fifty micro-spheres to
diameter increases, which indicates that a void becomes more be 70 and 75 mm with respect to the measured volume and
elongated as its volume increases. This analytical trend matches surface area, respectively , with both having a standard deviation
the non-analytical visual observations made from the 3D ren- of E5 mm. These results are within the known inner glass-bubble
dered images in Fig. 4, highlighting the effectiveness of non- diameter range 68.5–83 mm, while the standard deviation is less
analytical visual analysis. Analytical micro-CT analysis is also able than the 7 mm voxel size. Micro-CT accuracy verification testing
to analyse the overall volumetric void fraction and the total has shown that analytical micro-CT analysis is accurate to within
number of voids within a sample, while also calculating indivi- one voxel, which infers accuracy is dependent on scanning
dual void’s 3D size and location. The individual 3D void analysis resolution.
capabilities of micro-CT which allows for individual voids to be
analysed in a similar manner to microscopy (Fig. 3), however 3.4. Comparison of characterisation techniques
fundamentally micro-CT has no inherent section bias that affects
accuracy. This section compares the accuracy and reliability of the void
Determining the accuracy of the micro-CT analysis using characterisation techniques discussed and tested within this
naturally occurring voids is difficult since their size and shape is study. It is not possible to comment on the accuracy of each
method, as the actual void content is unknown, but the compara-
tive variation is able to be assessed. The void volume fraction
results are obtained from testing five samples per technique. Due
to section-bias errors, three cross-sections per sample were
analysed using microscopy, making a total of fifteen measure-
ments for this technique.
Fig. 7 shows the average calculated void volume fraction and
standard deviation of the five unidirectional laminate samples
tested using each of the analysed techniques. The average void
volume fraction between the three techniques is between 1% and
2%, which can be considered small when taking into account the
whole specimen volume and the inherent errors of each techni-
que. The small discrepancy between the void characterisation
techniques is significant if dealing with high quality, high perfor-
mance products, such as in the aerospace industry. A notable
difference between the techniques is the larger standard devia-
tion of the Archimedes void fraction. This standard deviation
Fig. 5. Unidirectional laminate sphericity distribution with respect to projected (0.74%) is nearly twice as large as that of microscopy (0.41%) and
area diameter from micro-CT analysis. over three times that of micro-CT (0.23%).
J.E. Little et al. / NDT&E International 46 (2012) 122–127 127

the individual void analysis capabilities of both visual techniques,


with both able to assess void size, shape and distribution.
The characterisation techniques analysed have all shown
limitations and accuracy constraints. Microscopy for example is
a destructive technique and micro-CT currently requires a small
sample size, whilst Archimedes Density requires accurate knowl-
edge of the material’s physical properties and constituent mass
fractions. Due to the inherent error with the Archimedes method
it has been proven to have a relatively large standard deviation,
whilst also lacking in ability to obtain characteristic data about
the individual voids. Although microscopy is able to measure this
critical information, its accuracy and reliability is limited due to
the errors of section-bias. The use of micro-CT has shown good
reliability, a low standard deviation and adequate void character-
isation tools. The current limitations of micro-CT analysis, most
notably the small sample size, are due to limitations of technol-
Fig. 7. Comparison of the unidirectional laminate void fraction calculated by ogy and are not inherent within the technique.
Archimedes, microscopy and micro-CT.

4. Conclusion

This work researched and tested void characterisation techni-


ques for composite materials to assess their accuracy and relia-
bility. The study has proven that each technique has its own
strengths and weaknesses, with Archimedes and microscopy
having critical inherent limitations. Taking into consideration
the required preparation and testing processes, the characterisa-
tion information available, cost and the overall accuracy and
reliability, it is concluded that micro-CT analysis is the most
accurate and reliable technique for characterising voids in com-
posite materials.

Acknowledgement

This research was supported by the New Zealand Foundation


Fig. 8. Comparison of the twill weave laminate void fraction calculated by
Archimedes, microscopy and micro-CT.
for Research Science and Technology (FRST) for Advanced Com-
posites Structures.
Fig. 8 shows the void volume fraction calculated by the three
References
characterisation techniques for the twill weave laminate samples.
The void fraction calculated by the Archimedes based technique
[1] Bossi RH, Georgeson GE, Rempt RD. X-ray computed tomography for emer-
averaged  0.22%, with a standard deviation of 0.81. The Archi-
ging aerospace materials and processes development. Boeing Defense and
medes standard deviation for the twill weave samples was Space Group; 1993.
significantly higher than that of the visual techniques (0.07– [2] Ghiorse S. Effect of void content on the mechanical properties of carbon/
0.08), which were similar to the values obtained for the unidirec- epoxy laminate. SAMPE Q 1993;24(2):54–9.
[3] Standard test methods for void content of reinforced plastics. Designation D
tional laminate, and highlights the inaccuracy of the Archimedes 2734–09. American Society for Testing of Materials (ASTM).
method when the material properties are not accurately known. [4] Kelly A, Zweben C. Comprehensive composite materials—test methods, non-
The negative portion of this result is not shown in Fig. 8 as this is a destructive evaluation, and smart materials. Elsevier Science Ltd; 2000. 5.
[5] Carlsson A, Pipes RB. Experimental characterization of advanced composite
physically impossible result. materials. 2nd ed. Technomic Publishing Company Inc; 1997.
There is a good comparison between microscopy (0.1270.07%) [6] Standard test methods for constituent content of composite materials.
and micro-CT (0.2670.08%) in Fig. 8, with the upper range of the Designation D 3171–09. American Society for Testing of Materials (ASTM).
[7] Liu L, Zhang B, Wang D, Wu Z. Effects of cure cycles on void content and
microscopy void fraction over lapping with the lower end of the mechanical properties on composite laminates. Compos. Struct. 2006;73(3):
micro-CT results. The variation in the average void fraction between 303–9.
the two visual techniques is also minimal, with a difference of 0.14%. [8] Oliver P, Cottu JP. Ferret B. Effects of cure cycle pressure and voids on some
mechanical properties of carbon/epoxy laminates. Composites 1994;30(6):
The individual void characterisation carried out by microscopy 505–15.
and micro-CT showed comparatively similar trends. Both techni- [9] Desplentere F, Lomov SV, Woerdeman DL, Verpoest I, Wevers M, Bogdanovich
ques showed that the majority of voids account for a small A. Micro-CT characterization of variability in 3D textile architecture. Compos
Sci Technol 2005;65(13):1920–30.
minority of the total void fraction, while as a void’s volume
[10] Michael G. X-ray computed tomography. Phys. Educ. 2000;36(6):442–51.
increases it becomes increasingly elongated. Although the trends [11] Pratten NA. Review: The precise measurement of the density of small
were similar, microscopy had a predominately higher standard samples. J Mater Sci 1737-47;198:16.
deviation, which can be attributed to the inherent section-bias [12] Cann MT, Adams DO, Schneider CL. Characterization of fiber volume fraction
gradients in composite laminates. J Compos Mater 2008;42(5):447–66.
and highlights the variability and potential inaccuracy when [13] Purslow D. On the optical assessment of the void content in composite
using microscopy. It is noted that Figs. 2 and 3 are examples of materials. Composites 1984;15(3):207–10.

S-ar putea să vă placă și