Sunteți pe pagina 1din 11

                           Tech Note 2009‐09 

 
Tech Note for Air Data Calibration 
Tech Note 2009‐09 Rev B  January 7, 2011 

Description:
The EFD1000 display contains an air data type computer with associated computing systems, 
and incorporates air data correction internally.  This Tech Note contains the manner of testing 
and the specifications for the EFD1000 system to permit altitude calibration of the equipment 
and validation under 14CFR 91.411, and also airspeed calibration. 

NOTE:  Contact an Aspen Field Service Representative prior to calibrating airspeed. 

1. SUBJECT
The EFD1000 display altitude and airspeed can be tested and field calibrated using the 
procedures identified below.  This Tech Note describes the means to acquire the calibration 
tool and instructions to test and calibrate the EFD1000.  These data can be used to permit an 
appropriately rated repair facility to add the EFD1000 to the Operations Specifications, and 
plan, supervise, perform, and approve for return to service the maintenance, or preventive 
maintenance described. 

2. EFFECTIVITY
This Tech Note is applicable to aircraft with the EFD1000 PFD System Part Number A‐05‐110‐00 
with Software version 1.1 or subsequent and hardware modification status of ‘A’ or 
subsequent.  It is also applicable to hardware version 910‐00001‐001, 910‐00001‐002, 910‐
00001‐004, 910‐00001‐101, 910‐00001‐104 or 910‐00001‐105 with Software version 2.0 or 
subsequent.   

3. COMPLIANCE
The test portion of this tech note can be used to fulfill the altimeter instrument portion of the 
14CFR Part 91.411 Altimeter system and altitude reporting equipment tests and inspections (14 
CFR Part 43 Appendix E paragraph (b)(2)).    

All activities described in this Tech note must be accomplished with a pitot static test set that is 
properly calibrated to a standard acceptable to the FAA or other Civil Aviation Regulatory 
entity.  If the airspeed part of the test set is not calibrated, then the airspeed calibration cannot 
be accomplished.  See 14CFR Part 145.109. 

DOCUMENT # 991-00023-001 PAGE 1-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

4. APPROVAL
These data are developed by the manufacturer which are acceptable to the Administrator, and 
constitutes the corrective maintenance procedures for the EFD1000 system. 

5. BACKGROUND
The altitude sensor and airspeed sensor in the EFD1000 can be subject to drift over time.  This 
procedure permits recalibration.  As noted below, if recalibration is necessary in an interval of 
less than six months, return the unit to the factory for repair.   
NOTE:  The purpose of the calibration procedures is to address drift.  If the indications are incorrect 
at initial installation, then it is possible that the sensor is damaged.  Caution must be 
exercised to avoid over‐pressuring the sensors.  It is good practice to connect the air data 
lines after all other work on the EFD1000 system is completed. 

6. CHANGE SUMMARY
No change is accomplished.  These are test and calibration procedures. 

7. WARRANTY INFORMATION
Unless authorized, this Tech Note activity is not covered by warranty. 

8. LABOR
It is estimated that 2 hours of labor will be required to perform the actions described in this 
Tech Note, not including documentation.   

9. REFERENCE
This Tech Note, 14CFR Part 43 Appendix E. 

DOCUMENT # 991-00023-001 PAGE 2-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

10. EQUIPMENT AND MATERIAL REQUIRED


Qty  Part Number  Description (refer to the Installation 
Manual for additional source information) 
1   Kingston® brand Micro SD card  Air Data Field Calibration Tool SD Card.  The 
per PFD  or equivalent, 2GB or less.  file is downloaded from the Aspen website. 
1  377‐4A‐50A  Castleberry Model 377‐4A‐50A or 
equivalent (capable of 30,000 feet for 
altitude testing and 450 knots for airspeed).  
The equipment must be properly calibrated.
1  A-06-507-00 or Pitot Quick Connector (Aspen Avionics) 
275-00001-002
1  A-06-505-00 or Static Quick Connector (Aspen Avionics)  
275-00001-001
4  Aero Seal 6604 or equivalent Hose clamps (Avionics Distributor) 
Convenient  Imperial Eastman 44PN or 1/4” pitot and static tubing (Avionics 
Length  equivalent Distributor) 

11. Obtaining the Air Data Field Calibration Tool SD Card


1. NOTE: These instructions show how to download the correct “sdboot.bin” file.   
2. Obtain a MicroSD card.  MicroSD cards can be purchased from local electronics, office supply or 
camera stores.   
3. Navigate your computer browser to the Aspen Avionics Dealer Ramp, Technical Page. 
4. Under Evolution Flight Display Tech Notes click the link for Air Data Field Calibration Tool.  
5. Click the “Download the Air Data Field Calibration Tool” link.  A File Download dialog box will 
open. 
6. Click the Save button.  A “Save As” dialog box will open. 
7. Select the drive and folder where you want to save the file and click the Save button.   
8. Copy the file (the name of the file is “sdboot.bin”) to the root directory of the MicroSD card 
using a flash card reader/writer.  

12. EFD1000 Altimeter Leak Test


1. Disconnect the EFD1000 system from the aircraft pitot and static systems.  Connect the EFD1000 
static and pitot connections to the static port and pitot port (respectively) of the pitot static test 
set.   
2. Turn on the EFD1000. 
3. Set the barometric pressure adjustment on the EFD1000 to 29.92 inHg 
4. The pitot port should be open to the atmosphere at the pitot‐static test set 1 . 
                                                            
1
 The valves in the EFD1000 are sealed to the atmosphere when the fittings are disconnected.  Therefore opening 
to the atmosphere at the EFD1000 is not practical. 

DOCUMENT # 991-00023-001 PAGE 3-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

5. Set the altitude in the test set to a pressure of 15,000 feet plus the field elevation.   
6. Note any change in altitude after one minute.   
7. To pass the Altimeter leak test, the altitude indication on the EFD1000 must not change more 
than 100 feet in one minute.  If it does not pass the leak test, proceed to Section 18. 

13. EFD1000 Airspeed Leak Test


NOTE:  When the air data is exercised with no associated acceleration, the attitude indication may be
affected. This is normal. 

1. Disconnect the EFD1000 system from the aircraft pitot and static systems.  If not already 
accomplished, connect the EFD1000 static and pitot connections to the static port and pitot port 
(respectively) of the pitot static test set.  Verify the test equipment connections do not leak. 
2. Turn on the EFD1000. 
3. The static port should be open to the atmosphere at the pitot static test set. 
4. Set the airspeed in the test set to 450 knots and seal the system at that pressure.   
5. Wait one minute. 
6. Note any change in airspeed on the EFD1000 airspeed indications.   
7. To pass the Airspeed leak test, the airspeed indication on the EFD1000 must not change more 
than 1 knot in one minute.  If it does not pass the leak test, proceed to Section 18. 

14. Altimeter Test Procedure


1. Verify the Leak Test (Section 12) was accomplished. 
2. Verify the temperature of the EFD1000 system and the outside air temperature is 25 ± 10°C.   
3. If not previously accomplished, connect the EFD1000 pitot connection directly to the pitot port 
of the Pitot Static Test Set. 
4. If not previously accomplished, connect the EFD1000 static connection directly to the static port 
of the Pitot Static Test Set. 
5. Turn on the EFD1000. 
6. Using the test set, perform an altitude test as follows:   
a. With the barometric pressure scale at 29.92 inches of mercury, the altimeter shall be 
subjected successively to pressures corresponding to the altitude specified in Table 1 up 
to 30,000 feet or the maximum normally expected operating altitude of the airplane in 
which the EFD1000 system is installed, whichever is greater.  The reduction in pressure 
shall be made at a rate not in excess of 20,000 feet per minute to within approximately 
2,000 feet of the test point. The test point shall be approached at a rate compatible with 
the test equipment. The altimeter shall be kept at the pressure corresponding to each 
test point for at least 1 minute, but not more than 10 minutes, before a reading is taken. 
The error at all test points must not exceed the tolerances specified in Table 1. 

NOTE:  Setting the Barometric pressure scale on the EFD1000 and the test set is essential for correct 
operation.  

DOCUMENT # 991-00023-001 PAGE 4-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

 In addition, the altitude readings on the test set must be precisely at the proper altitudes 
shown in the table. 

7. Record the calibrated altitudes and the EFD1000 indicated altitudes, particularly at 500 feet, 
10,000 feet and 30,000 feet (these are the calibration points for the Altimeter Calibration 
Procedure) 
8. If the test passes, then no recalibration is required.  Complete the remainder of the static testing 
as required by regulation and update the Aircraft Records, identifying the altitudes tested.  The 
test is accomplished in accordance with 14CFR Part 43 Appendix E paragraph (b) (2). 
9. If the altitude is in error at any altitude by more than three times the tolerance, or a 
recalibration is necessary within six months of a previous calibration, proceed to Section 18. 
10. If the altitude requires calibration, proceed to Section 15. 

15. Altimeter Calibration Procedure

 
Figure 1 PFD “key” Functions 

1. Verify the Altimeter Leak Test was completed (see Section 12 above). 
2. Verify the PFD has software version 1.1 or subsequent. 
3. Verify the temperature of the EFD1000 system and the outside air temperature is 25 ± 10°C.   
4. Obtain the Air Data Field Calibration Tool SD Card (see Section 11 above). 
5. Power on the EFD1000 in the aircraft for a minimum of 10 minutes using a ground power unit 
prior to beginning the calibration procedure. 
6. Insert the SD card loaded with the Air Data Field Calibration Tool and enter the Installation 
Menu by pressing the “MENU” key and turning the right knob clockwise until the last menu page 

DOCUMENT # 991-00023-001 PAGE 5-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

is shown.  Then press MENU and the two line select keys directly below “MENU” simultaneously 
for 3 seconds.  See Figure 1.   
7. Rotate the right knob to turn to the last page of the Installation Menu.  Then boot from the SD 
card by pressing the two buttons closest to the knobs for three seconds.  See Figure 1.  When 
the display responds to the button press, release the buttons. 
8. The EFD will proceed directly into the Air Data Field Calibration Tool.  
9. There are four options available next to the top four line keys (starting below the MENU key)  
a. Adjust Altitude  
b. Adjust IAS  
c. Load Factory Default Cal  
d. Load Other Cal  
10. “Adjust Altitude” is used to calibrate the EFD1000 altitude indications.  When pressed, the title 
“Enter Altitude (ft)” is presented and you must enter the three values recorded earlier for 500, 
10000 and 30000 feet (pressing MENU returns you to the Main Menu).  
a. The first line key is the 500 ft key.  Press the key, and then use the right knob to adjust 
the value to the EFD1000 value recorded when the calibrated source was set at 500 ft.  
b. The second line key is the 10000 ft key. Press this key, and then use the right knob to 
adjust the value to the EFD1000 value recorded when the calibrated source was set at 
10000 ft.  
c. The third line key is the 30000 ft key. Press this key, and then use the right knob to 
adjust the value to the EFD1000 value recorded when the calibrated source was set at 
30000 ft.  
d. The fourth key is the ACCEPT CAL key. Once the values are entered, press this key to 
accept the calibration or press MENU to reject the calibration and return to the main 
menu.  
e. The software tool will determine whether or not the values entered will produce a valid 
calibration2 .  If not, a Calibration Rejected message will be displayed and you must press 
the MENU key to begin the process again.  
f. If the values are accepted, you will see “Calibration Accepted” on this screen. Then you 
must name the calibration by entering the date and time of the calibration (use this 
format: Day Month Year‐Hour Minute; example: 12NOV2009‐0845).  This is 
accomplished by selecting the name by pressing the first line key and using the right 
cursor to scroll through letters (numbers and symbols) use the left knob to select each 
letter in turn.  Up to 16 letters may be used.  
g. If you are satisfied with everything, press the second line key (Write Calibration). 
Writing Calibration will be shown. It takes about 20 seconds while the calibration writes 
to the system.  

                                                            
2
 A valid calibration simply means a linear error distribution.  A valid calibration must rely on accurate correction 
observations and correct inputs performed by the technician. A valid calibration may be an incorrect calibration.  

DOCUMENT # 991-00023-001 PAGE 6-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

h. After the calibration is written, the status will return to Ready and the new name will 
appear on the Main Menu as the Current Calibration. 
11. Once complete, reboot the system by pressing the REV button.  
12. Verify proper calibration by completing the Altimeter Test Procedure, Section 11. 
13. Update the Aircraft Records, identifying the altitudes tested. 

NOTE:  If the System fails to accept the Calibration Procedures refer to Section 18. 

16. Airspeed Test Procedure


1. Verify the Airspeed Leak Test (Section 13) was accomplished. 
2. Verify the temperature of the EFD1000 system and the outside air temperature is 25 ± 10°C. 
3. Unless previously accomplished, connect the EFD1000 pitot connection directly to the pitot port 
of the Pitot Static Test Set. 
4. Unless previously accomplished, connect the EFD1000 static connection directly to the static 
port of the Pitot Static Test Set. 
5. Turn on the EFD1000. 
6. The airspeed indicator shall be subjected successively to pressures corresponding to the 
airspeeds specified in Table 2 up to 200 knots or the maximum airspeed of the airplane in which 
the airspeed indicator is installed, whichever is greater.  The test point shall be approached at a 
rate compatible with the test equipment. The airspeed shall be kept at the pressure 
corresponding to each test point for at least 1 minute, but not more than 10 minutes, before a 
reading is taken. The error at all test points must not exceed the tolerances specified in Table 2. 

NOTE: The airspeeds on the test set must be precisely at the proper airspeeds shown in the table. 

7. Record the calibrated airspeeds and the EFD1000 indicated airspeeds, particularly at 50, 85 and 
200 knots (these are the calibration points for the Airspeed Calibration Procedure). 
8. If the test passes, then no recalibration is required.  Complete the remainder of the pitot static 
testing as required by regulation and update the Aircraft Records. 
9. If the airspeed is in error at any airspeed by more than two times the tolerance, or a 
recalibration is necessary within six months of a previous calibration, proceed to Section 18. 
10. If the airspeed requires calibration, proceed to Section 17. 

17. Airspeed Calibration Procedure


NOTE:  Contact an Aspen Field Service Representative prior to calibrating airspeed. 

1. Verify the airspeed portion of the Pitot Static Test Set is in calibration.  If not, the calibration 
cannot be accomplished. 
2. Verify the temperature of the EFD1000 system and the outside air temperature is 25 ± 10°C.   
3. Verify the leak test was completed (see Section 12 above). 
4. Verify the PFD has software version 1.1 or subsequent. 
5. Obtain the Air Data Field Calibration Tool SD Card (see Section 11 above). 

DOCUMENT # 991-00023-001 PAGE 7-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

6. Power on the EFD1000 in the aircraft for a minimum of 10 minutes using a ground power 
unit prior to beginning the calibration procedure. 
7. Insert the SD card and enter the Installation Menu by pressing the “MENU” key and turning 
the right knob clockwise until the last menu page is shown.  Then press MENU and the two 
line select keys directly below “MENU” simultaneously for 3 seconds.  See Figure 1.   
8. Rotate the right knob to turn to the last page of the Installation Menu (11 of 11).  Then boot 
from the SD card by pressing the two buttons closest to the knobs for three seconds.  See 
Figure 1.  When the display responds to the button press, release the buttons. 
9. The EFD will proceed directly into the Air Data Field Calibration Tool.   
10. There are four options available next to the top four line keys (starting below the MENU 
key): 
a. Adjust Altitude  
b. Adjust IAS  
c. Load Factory Default Cal  
d. Load Other Cal  
11. “Adjust IAS” permits calibration of the airspeed indications. You will see the title: Adjust IAS 
(kts). When pressed, the title “Enter Airspeed (kts)” is presented and you must enter the 
three values recorded earlier for 50, 85 and 200 knots (pressing MENU returns you to the 
Main Menu).  
a. The first line key is the 50 knots key.  Press the key, and then use the right knob to 
adjust the value to the EFD1000 value recorded when the calibrated source was set at 
50 kias.  
b. The second line key is the 85 knots key. Press this key, and then use the right knob to 
adjust the value to the EFD1000 value recorded when the calibrated source was set at 
85 kias.  
c. The third line key is the 200 knots key. Press this key, and then use the right knob to 
adjust the value to the EFD1000 value recorded when the calibrated source was set at 
200 kias.  
d. The fourth key is the ACCEPT CAL key. Once the values are entered, press this key to 
accept the calibration or press MENU to reject the calibration and return to the main 
menu.  
e. The software tool will determine whether or not the values entered will produce a valid 
calibration3 .  If not, a Calibration Rejected message will be displayed and you must press 
the MENU key to begin the process again.  
f. If the values are accepted, you will see Calibration Accepted on this screen. Then you 
must name the calibration by entering the date and time of the calibration (use this 
format: Day Month Year‐Hour Minute; example: 12NOV2009‐0845).  This is 

                                                            
3
 A valid calibration simply means a linear error distribution.  A valid calibration must rely on accurate correction 
observations and correct inputs performed by the technician.  A valid calibration may not be a correct calibration. 

DOCUMENT # 991-00023-001 PAGE 8-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

accomplished by selecting the name by pressing the first line key and using the right 
cursor to scroll through letters (numbers and symbols) use the left knob to select each 
letter in turn.  Up to 16 letters may be used.  
g. If you are satisfied with everything, press the second line key (Write Calibration). 
“Writing Calibration” will be shown. It takes about 20 seconds while the calibration 
writes to the system.  
12. Once complete, reboot the system by pressing the REV button.  
13. Verify that the adjustment worked correctly and verify proper calibration by completing the 
Airspeed Test Procedure, Section 16 above.  
14. Update the aircraft records.  

18. Actions to be Taken in the Event of an Unsuccessful Calibration 
or Unsuccessful Test. 
If the leak test was unsuccessful, contact your Aspen Field Service Representative to arrange for 
returning the equipment to the factory. 

If the altitude test results were in error of more than three times the tolerance at any point, or a 
recalibration is needed within 6 months of a previous calibration, contact your Aspen Field Service 
Representative to arrange for returning the equipment to the factory. 

If the airspeed test results were in error of more than two times the tolerance at any point, or a 
recalibration is needed within 6 months of a previous calibration, contact your Aspen Field Service 
Representative to arrange for returning the equipment to the factory. 

DOCUMENT # 991-00023-001 PAGE 9-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

19. Altitude Tolerance Table 
The following table identifies the acceptable altitude tolerances for the EFD1000 display. 

Altitude in Feet  Tolerance 
0  ±20.0 feet 
500  ±20.0 feet 
1,000  ±20.0 feet 
2,000  ±25.0 feet 
3,000  ±25.0 feet 
4,000  ±25.0 feet 
5,000  ±25.0 feet 
8,000  ±30.0 feet 
10,000  ±33.2 feet 
11,000  ±35.0 feet 
14,000  ±40.0 feet 
17,000  ±45.0 feet 
20,000  ±50.0 feet 
30,000  ±75.0 feet 
Table 1 Altitude Tolerance Table 

20. Airspeed Tolerance Table 
The following table identifies the acceptable airspeed tolerances for the EFD1000 display. 

Caution: Do not exceed the capability of the test set. 

Airspeed in Knots  Tolerance 
50  ±5.0 knots 
80  ±3.0 knots 
100  ±2.0 knots 
120  ±2.0 knots 
150  ±2.0 knots 
200  ±2.0 knots 
250  ±2.4 knots 
300  ±2.8 knots 
350  ±3.2 knots 
400  ±3.6 knots 
450  ±4.0 knots 
Table 2 Airspeed Tolerance Table 

DOCUMENT # 991-00023-001 PAGE 10-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 
                           Tech Note 2009‐09 

21. Air Data Calibration Display Nomenclature 
The following tables show the air data display elements and comments on the meaning. 

Shown on the EFD1000 Display:  Comment: 
Air Data Field Calibration  This is the title of the utility. 
PN: 309‐00005‐001  This is the part number of the tool.   
Current Cal: XQLMEK  This is the title of the current calibration. 
This describes what is about to be done and 
READY  asks for the technician to be certain that this is 
the desired procedure. 
MAIN MENU 
If you wish to exit with no changes, press the 
‘REV’ button.  Otherwise press MENU or a knob 
to continue. 
Press this Line Select Key to go to the altitude 
Adjust Altitude
calibration procedure. 
Press this Line Select Key to go to the airspeed 
Adjust IAS
calibration procedure. 
Press this Line Select Key to load the factory 
Load Factory Default Cal default calibrations for airspeed and altitude.  
See section 22. 
Press this Line Select Key to load a titled 
Load Other Cal calibration for airspeed and altitude.  See 
section 22. 
 

22. Factory Default Cal and Load Other Cal 
Load Factory Default Cal will reload and save the factory default calibrations.  If the pitot or 
static sensor has drifted, this calibration is not satisfactory.  Pressing this line select key will load 
the factory calibration values for both airspeed and altitude.  This is a useful function to restart 
adjustments after an improperly done calibration.  The default cal operation will take about 20 
seconds.  If the sensors have drifted, a Default Cal is most likely an incorrect calibration. 

Load Other Cal will take you to another screen with the title Select Calibration. This screen will 
have up to the last 5 calibrations with titles (as entered during the calibration process). Select 
any one of these and that calibration will be loaded and saved.  The title will change to reflect 
what was loaded.  Do not use this function unless directed by a Field Service Representative. 

DOCUMENT # 991-00023-001 PAGE 11-11 REVISION B


NOTICE: This document contains PROPRIETARY information of Aspen Avionics Inc. No reproductions or disclosure of this information is
permitted in any form without prior written consent.
 

S-ar putea să vă placă și