Documente Academic
Documente Profesional
Documente Cultură
50346, 2013
2975
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
velocity
antinode
pressure
antinode
Figure 1. DARS setup and the pressure field in the cavity. A cylindrical cavity with open ends is immersed in a tank filled
with silicon oil. Source and receiver are connected to a lock-in amplifier through a power amplifier and a preamplifier,
respectively. A computer-controlled stepper motor is used to control sample positioning. There are two particular locations
in the cavity, pressure antinode (velocity node) and velocity antinode (pressure node) in the cavity.
a relationship between rock acoustic properties and porosity in inside the sample to that of the acoustic pressure field inside
the same experiment, which will be a great gain in understand- the cavity when resonance occurs, reflecting the acoustic
ing the properties of fluid-saturated rocks. properties of the measurement system. As shown in Figure 1,
the acoustic pressure field inside the cavity has a spatially
2. DARS Perturbation Equation varying but harmonic distribution when the fundamental
mode resonance occurs. Theoretically, the acoustic pressure
[4] Based on the derivation of Wang et al. [2012], the has a cosine function distribution along the longitudinal direc-
DARS perturbation equation is tion. There are two particular locations at which acoustic pres-
ks k0 Vs r r0 Vs sure or velocity reach their maxima. For the fundamental
o2s o20 ¼ o2s A o20 s B: (1) mode, an acoustic pressure node occurs where sonic velocity
k0 Vc r s Vc
is at its maximum, and a velocity node with acoustic pressure
where Z Z at its maximum. At these locations, either A or B is zero.
Vc [6] Equation (1) shows that the compressibility contrast,
A¼ p1 p2 dV = p1 p2 dV ;
Vs (ks k0)/k0, and the density contrast, (rs r0)/rs, between
Vs Z Vc Z a test sample and the fluid inside the cavity, both contribute
Vc 1
B¼ rp1 :rp2 dV = p1 p2 dV to the resonance frequency shift. Thus, the perturbation
Vs k 2
Vs Vc equation (1), in conjunction with DARS measurements,
k ¼ o0 =c: can be used to estimate the acoustic properties of a sample.
[5] In equation (1), Vc and Vs are the volumes of the cavity 3. Algorithms to Determine Acoustic Properties
and test sample, respectively. We use o0 and os to denote
the resonant frequencies of the cavity with and without 3.1. One-Data-Point-Fitting Method
sample, k0 and ks to denote the compressibility parameters [7] To measure compressibility, a sample is positioned at
of the cavity fluid and test sample, and r0 and rs to denote a velocity node, where h r p1. r p2i vanishes, and equation
the densities of the cavity fluid and test sample, respectively. (1) can be reduced to
The parameter c is the sonic velocity in the cavity fluid, and os o0 ks k0
k = o0/c is the wave number. The acoustic pressure fields p1 ¼ C’Vs ; (2)
o0 k0
and p2 are those for cavity conditions with and without a
sample. The coefficients A and B give the ratio of the energy using approximation os o0. The calibration coefficient
density of the acoustic pressure and particle velocity field C ’, which is related to the cavity geometry and experimental
2976
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
conditions (i.e., temperature and ambient pressure), can be the fundamental mode resonance (with L as the cavity
obtained from the measured resonance frequency shift and length).
known compressibility and volume of a reference sample. [12] When a sample is measured at a given location, z ¼ z,
[8] The calculation of compressibility based on equation as shown in Figure 1, the coefficients A and B can be
(2) is called the one-data-point method, in which only the obtained by calculating the volume integrals in equation
resonance frequency measured at the velocity node is used. (1). The results are
This inevitably causes some limitations because measure- sinð2k0 z2 Þ sinð2k0 z1 Þ
ments at other positions have been made redundant, and Az¼z ¼ 1 þ ;
2k0 ðz2 z1 Þ
errors may arise due to individual measurement uncer-
tainties or inaccurate positioning of the sample. So far, the (6)
sinð2k0 z2 Þ sinð2k0 z1 Þ
one-data-point approach has performed better with samples
Bz¼z ¼ 1 :
of highcompressibility than with those of low and medium 2k0 ðz2 z1 Þ
compressibility, in terms of the estimation accuracy of
where Z1 and z2 denote the bottom and top ends of the
acoustic properties [Wang et al., 2012]. We expect that the
sample, respectively. The determination of A and B with this
DARS measurements with a sample at multiple locations
approach is referred to as Method 1.
inside the cavity will effectively suppress the errors caused
[13] The approximation, p1 p2 cos(k0z), is reasonable
by fluctuations in experiment conditions or measurement
to a great extent, especially if DARS measurements are
uncertainty, thus contributing to more accurate estimates
under stable conditions (e.g., consistent room temperature)
for samples of low compressibility. Moreover, an important
and with samples of high compressibility. The whole-
feature of multiple location-based DARS measurements is
curve-fitting inversion can then be implemented with these
that both compressibility and density parameters can be
known coefficients A and B. However, for samples of low
obtained through a whole-curve-fitting inversion.
compressibility (e.g., fluid-saturated rocks), and/or under
variable temperature and pressure conditions, this approxi-
3.2. Whole-Curve-Fitting Inversion mation is more open to question.
[9] The whole-curve-fitting inversion involves a nonlinear [14] Therefore, we propose an alternative method to
least-squares fitting technique, and it provides a solution to obtain the coefficients, A and B, at z ¼ z by using two refer-
both compressibility and density of a sample. In this tech- ence samples instead of calculating the integrals in equation
nique, the second term of the right side of equation (1) is (1). In this method (hereafter referred to as Method 2),
no longer negligible, as the measurements at other positions theresonance frequencies corresponding to two reference
along the cavity axis (See Figures 1 and 2a) will be used for samples of known acoustic properties and volumes can be
the estimation of the acoustic properties. Accordingly, the measured at any location, thus, the calibration coefficients,
coefficients A and B should be determined for each measure- A and B, can be obtained by simply solving a binary linear
ment location. At any location (zi, i = 1, 2 ⋯ n), an expression equation system. Using the whole-curve-fitting inversion
relating the measured resonance frequency to the contrast in with the calibration coefficients obtained by Method 2, we
acoustic properties between the sample and the fluid can be conjecture that the estimation of both compressibility
obtained anddensity of samples can be more accurately achieved.
Particularly, we anticipate that the estimation of the acoustic
o20 o20 ð1 r0 rÞBi properties of samples with low compressibility may be
o2si ¼ ; i ¼ 1; 2⋯n; (4)
1 þ k10 ks 1 Ai significantly improved.
2977
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
0.25m
1.4
Inverted compressibility
True compressibility
1.2
Compressibility (GPa−1)
0.07m
0.028m
1
0.8
0.05m
0.20m
0.75m
0.38m
0.6
0.09m
0.4
0.2
0
0 2 4 6 8 10 12 14 16 18
Sample No.
(a) (b)
1.4
Inverted compressibility 3000 Inverted density
True compressbility True density
1.2 2800
Compressbility (GPa−1)
2600
1
2400
Density (g/m3)
2200
0.8
2000
0.6 1800
1600
0.4
1400
1200
0.2
1000
0 800
0 2 4 6 8 10 12 14 16 18 0 2 4 6 8 10 12 14 16 18
Sample No. (c) Sample No.
2800
1.4
Inverted compressibility Inverted density
2600 True density
True compressibility
1.2
2400
Compressibility (GPa−1)
1 2200
Density (g/m3)
2000
0.8
1800
0.6 1600
0.4 1400
1200
0.2
1000
0 800
0 2 4 6 8 10 12 14 16 18 0 2 4 6 8 10 12 14 16 18
Sample No. Sample No.
(d)
Figure 2. The three-dimensional simulation of DARS measurements and estimation of acoustic properties based on the
whole-curve-fitting technique and the one-data-point fitting technique. The sonic velocity for the cavity fluid is 960 m/s,
and the density is 908 kg/m3. Eighteen synthetic samples, listed in Table 1, are used in the numerical study. (a) Each sample
is moved along the axis inside the cavity, and the resonance frequency is obtained through the DARS simulation at each
location. The resonance frequencies at all locations are used to estimate both compressibility and density parameters of a test
sample through the whole-curve-fitting technique; (b) the density of 18 synthetic samples inverted using the one-data-point
fitting technique; (c) and (d) show the inversion results of acoustic properties of the 18 synthetic samples. The results in
Figure 2c were obtained with the calibration coefficients A and B determined using equation (6), while those in Figure 2d
employed values for A and B from two reference samples (Sample 9 and 16).
(0.028 m) and length (0.05 m). Resonance frequencies were and 2d show the estimated compressibility and density of the
simulated with the sample positioned at the same locations synthetic samples, using the two different methods outlined
as in the real experiments. These resonance frequencies were above. For comparison, Figure 2b shows the sample com-
then used to invert for both compressibility and density of a pressibility based on the one-data-point fitting method.
test sample using the whole-curve-fitting technique. Figures 2c The compressibility estimations obtained with the whole-
2978
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
Table 1. Eighteen Synthetic Samples With Their Known Compressibility and Density Valuesa
Inversion Results Inversion Results
3 3 1 1
Sample No. Velocity (m/s) Density (kg/m ) Density (kg/m ) Error k(Gpa ) k (Gpa ) Error
1 960 960 943.960 1.67% 1.13028 1.13700 0.59%
2 980 1000 979.495 2.05% 1.04123 1.04680 0.53%
3 1000 1040 1015.200 2.38% 0.96154 0.96610 0.47%
4 1040 1080 1055.000 2.31% 0.85607 0.85960 0.41%
5 1080 1120 1094.600 2.27% 0.76548 0.76810 0.34%
6 1200 1180 1163.200 1.42% 0.58851 0.58990 0.24%
7 1400 1250 1246.400 0.29% 0.40816 0.40870 0.13%
8 1600 1350 1349.600 0.03% 0.28935 0.28950 0.05%
9 1800 1450 1450.000 0.00% 0.21286 0.21280 0.03%
10 2000 1540 1539.900 0.01% 0.16234 0.16230 0.02%
11 2400 1600 1608.200 0.51% 0.10851 0.10850 0.01%
12 2800 1800 1800.400 0.02% 0.07086 0.07080 0.09%
13 3000 1900 1897.200 0.02% 0.05848 0.05840 0.14%
14 3500 2000 1998.600 0.02% 0.04082 0.04080 0.04%
15 4000 2100 2099.000 0.02% 0.02976 0.02970 0.21%
16 5000 2300 2300.100 0.02% 0.01739 0.01740 0.05%
17 6000 2500 2503.700 0.02% 0.01111 0.01110 0.10%
18 7000 2700 2711.300 0.02% 0.00756 0.00760 0.55%
a
The estimated acoustic properties inverted from the simulated resonance frequencies using the whole-curve-fitting technique (see text) are also listed.
Samples No. 9 and 16 are used to obtain the calibration coefficients Ai and Bi in equation (4).
curve-fitting method are comparable to those inverted using 1, D13-2, and D36-1 are four sandstone samples from drilled
the one-data-point method. Figure 2d, which shows the results cores. Their dimensions, density, and compressibility (ultra-
of the whole-curve-fitting method using the calibration param- sonic measurements) in the saturated condition are listed in
eters obtained from Method 2, demonstrates that the latter Table 2 (also see the auxiliary material for measurements
method can provide good density estimation for all samples. under the dry condition). The compressibility of these samples
However, for samples of low compressibility, the densities es- was calculated from the P-wave and S-wave velocities
timated with this method using the calibration parameters de- obtained by ultrasonic measurements under room temperature
termined by Method 1 deviate slightly from the true values and atmospheric pressure, and the density obtained directly
as seen in Figure 2c. This is anticipated, since these coeffi- from mass and volume measurements.
cients are calculated based on the assumption that the pressure [17] Each sample was measured at 75 positions along the
field obeys the cosine function distribution, which may be in- cavity axis at intervals of 12 mm. The frequency range for
accurate when the sample has relatively low compressibility searching the resonance frequency at each sample position
and high density, and when the boundary conditions do not ap- was from 565 Hz to 665 Hz. Such measurements were
ply [Wang et al., 2012]. Nevertheless, this assumption is use- conducted 10 times for each sample to further assess the
ful when we do not have reference samples. measurement repeatability. Figure 3a shows the recorded
resonance curves for the empty cavity and the two standard
4.2. Laboratory Measurements and Discussion samples (Al-1 and Lu-1), for measurements at a fixed loca-
[16] Eight samples from different sources have been tion. As shown in Figure 3a, the resonance frequency shift
measured for their acoustic properties. A detailed description from the resonance frequency of the empty cavity caused
of these samples is given in the auxiliary material. Of these by A1-1 was larger than that caused by Lu-1. Because
samples, Al-1 and Lu-1 are two standard (reference) thecompressibility and density of the other six rock samples
samples, consisting of aluminum and Lucite, with compress- are between those of aluminum and Lucite, it can be antici-
ibilities of 0.01314 GPa1 and 0.17410 GPa1, and densities pated that their resonance curves (omitted in Figure 3a)
of 2.684 g/cm3 and 1.181 g/cm3, respectively; S10-1 and should also lie between those of Al-1 and Lu-1. Using
S10-2 are two artificial sandstone samples; and D7-1, D11- the Lorentzian curve-fitting technique [Xu, 2007], the
Table 2. The Dimensions and Densities of Two Reference Samples (Aluminum and Lucite), Two Artificial Sandstone Samples and Four
Drilled Rock Samplesa
Diameter Length Mass Density Vp Vs k
Sample Material (mm) (mm) (g) (g/cm3) (m/s) (m/s) (GPa1)
Al-1 Aluminum 25.020 40.000 53.060 2.698 6399.70 3074.29 0.01307
Lu-1 Lucite 25.180 40.180 23.630 1.181 2773.25 1456.21 0.17410
S10-1 Artificial sandstone 25.181 41.502 45.698 2.211 3705.54 1900.11 0.05072
S10-2 Artificial sandstone 25.192 40.430 43.952 2.181 3485.35 1772.99 0.05763
D7-1 Claystone 25.412 41.126 53.231 2.552 4031.96 2531.48 0.05081
D11-1 Shaly sand 25.319 39.550 51.295 2.576 4189.62 2541.52 0.04342
D13-2 Shaly sand 25.258 40.684 53.388 2.619 4101.41 2395.30 0.04163
D36-1 Claystone 25.380 39.718 52.666 2.621 4279.96 2658.15 0.04288
a
Ultrasonic transmission measurements were carried out for these samples.
2979
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
(a) (b)
(c) (d)
Figure 3. Implementation of the whole-curve-fitting inversion for two standard samples, Al-1 and Lu-1. (a) Normalized
pressure amplitude with scanning frequency, (b) resonance frequencies with the sample at the measurement locations along
the cavity axis, and the whole-curve-fitting inversions for (c) Al-1, and (d) Lu-1.
resonance frequencies for each sample at different locations Method 2, and the whole-curve-fitting inversions were carried
can be obtained. Figure 3b shows the resonance frequencies out for six rock samples. The rock samples were saturated with
for the two standard samples at different positions. silicon oil, put in a vacuum for 2 days to extract air inside them,
[18] We carried out the whole-curve-fitting inversions for and in a container with 5 MPa pore pressure for two more days,
the acoustic properties of the samples in two steps. First, we before they were sealed on the outer wall (with both ends
obtained the calibration coefficients A and B with Method 1 open). The resonance frequency measurements for the samples
for the reference samples (Al-1 and Lu-1), which were were made under normal room conditions. For each rock sam-
then used for the whole-curve-fitting inversion. The ple, we obtained the estimated compressibility and density
resulting inverted compressibility and density values are when the objective function Y in equation (5) was minimized.
listed in the auxiliary material. For Lu-1, we obtained a The estimated compressibility and density values of the six
resonance frequency curve perfectly fitted with the DARS sandstone samples, and the relative errors between different
measurements, as shown in Figure 3d. The inversion re- techniques-estimated compressibilities and densities, are given
sults, k = 0.17523 (GPa1), and r = 1.179 (g/cm3), agree in the auxiliary material. Figure 4 shows a comparison between
well with the true values: kreal = 0.17410 (GPa1), and the compressibility and density inverted with the whole-curve-
rreal = 1.181 (g/cm3). For Al-1, Figure 3c shows that the fitting technique and measured by the ultrasonic technique.
inversion curve matches the DARS measurements well, Low standard deviations for both the estimated density and
with some deviation at the extremes. The inverted density compressibility indicate that the DARS-based measurements,
obtained with this approach, r = 3.043 (g/cm3), is higher in conjunction with the whole-curve-fitting inversion
than its true value of 2.698 (g/cm3), whereas the inverted approach, can give consistent and reliable estimates of the
compressibility, 0.01355 (GPa1), is comparable to the true acoustic properties. In Figures 4a and 4b (also see auxiliary
value, 0.01307 (GPa1). Generally, the inversion results material), the relative error between the estimated and true den-
conform to the observation in the numerical study. For sam- sities is seen to be the largest for D7-1. The DARS-estimated
ples with extremely low compressibility, the whole-curve- compressibility for six samples has largerelative errors in
fitting inversion approach with the calibration coefficients contrast to the ultrasonic measurements, but relatively low
A and B obtained with Method 1 appeared to cause a larger standard deviations, which also give us confidence that the
error for density than for compressibility. DARS-estimated compressibility is reasonable considering
[19] Next, Al-1 and Lu-1 were used as two reference the large frequency contrast between the ultrasonic and DARS
samples to obtain the calibration coefficients A and B through measurements. Figure 4a shows the compressibility values
2980
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
2.8
2.7
2.6
Density (g/cm3)
2.5
2.4
2.3
2.2
2.1
2
0 1 2 3 4 5 6 7
Sample No.
(a) (b)
(c)
Figure 4. Comparison of compressibility and density parameters between the ultrasonic and DARS measurements for six
rock samples. Standard deviations of DARS-based compressibility and density parameters for the six samples (red bar) are
calculated from the whole-curve-fitting inversions of repeated measurements (see text). Results employ Figures 4a and 4b
calculated by Method 2 using the two reference samples.
obtained using the DARS inversion, the Gassmann fluid sub- seismic frequency would not be in the low-frequency regime.
stitution (see auxiliary material for input parameters), and the Under the circumstances, the Gassmann equation would lose
ultrasonic measurements. The frequency dependence of the some validity, and the direct low-frequency measurement is
compressibility is plotted in Figure 4(c). For the artificial sand- essential.
stones (S10-1 and S10-2) with simple pore shapes, their
compressibilities only show weak frequency dependency.
This reflects that the wave-induced pore pressure can reach 5. Conclusion
equilibrium at the measured frequency of DARS and even ul- [20] A DARS laboratory device, operating in a low
trasonic frequency, which satisfies the basic assumptions of frequency range, has been developed to estimate the acoustic
the Gassmann equation [Gassmann, 1951]. Thus, for this kind properties of rock samples. In comparison with other important
of rock samples, even ultrasonic frequency can be considered low-frequency measurement techniques, the DARS-based
in the low-frequency regime, which implies that the Gassmann measurement is easy and less time-consuming to operate.
equation is applicable in a wide frequency range. For the res- [21] The proposed whole-curve-fitting technique, which
ervoir rocks, obvious discrepancy between the ultrasonic and makes full use of all resonance frequencies measured for sam-
DARS measurements can be observed, which is mainly attrib- ples at different positions in the cavity, can be used to estimate
uted to complex pore shapes and the presence of clay. This both compressibility and density of the sample, in contrast to
complexity gives rise to a variety of dispersion mechanisms the one-data-point approach that only estimates compressibil-
[King and Marsden, 2002; Batzle et al., 2006], such as squirt ity. A critical aspect of this technique is the improved accuracy
flow, which contributes to the apparent compressibility varia- of the calibration coefficients, which yields more reliable
tion. Similar to the artificial sandstones, the DARS-estimated estimation of the acoustic properties of samples, particularly
compressibilities for the four reservoir rock samples are closer those with relatively low compressibility. The numerical and
to the values obtained using the Gassmann equation. However, experimental studies in this work show that the whole-curve-
for the specimens D11-1 and D13-2, the Gassmann equation fitting technique, in conjunction with the calibration coeffi-
gives errors of about 10%, which are not negligible or simply cients A and B obtained by measuring two reference samples,
attributed to measurement errors. Therefore, for some sedi- can achieve fairly good accuracy for both compressibility and
mentary rocks, the DARS frequency (about 600 Hz) and even density in the frequency range below 700 Hz. The DARS
2981
ZHAO ET AL.: A DARS-BASED LOW-FREQUENCY TECHNIQUE
measurements have been applied to compressibility and den- Cadoret, T., D. Marion, and B. Zinszner (1995), Influence of frequency
and fluid distribution on elastic wave velocities in partially saturated
sity estimates of fluid-saturated rock samples at the first time, limestones, J. Geophys. Res., 100(B6), 9789–9803, doi:10.1029/
and the differences in measured compressibility between the 95JB00757.
DARS method, ultra-sonic measurements, and the Gassmann Dvorkin, J., and A. Nur (1993), Dynamic poroelasticity: A unified model
with the squirt and the biot mechanisms, Geophysics, 58, 524–533,
estimation reflect substantial frequency dependence of the doi:10.1190/1.1443435.
compressibility. Therefore, we conclude that the laboratory Gassmann, F. (1951), Über die elastizität porösermedien:Vierteljahrsschrift
device, in conjunction with the whole-curve-fitting inversion, der Naturforschenden Gesellschaft in Zurich, 96, 1–23.
is a useful tool to estimate the properties of acoustically small Gribb, T., and R. Cooper (1998), A high-temperature torsion apparatus for
the high-resolution characterization of internal friction and creep in
rock samples in the low frequency range. Further, the DARS refractory metals and ceramics: Application to the seismic-frequency,
device and the inversion method could potentially be used to dynamic response of Earth’s upper mantle, Rev. Sci. Instrum., 69(2),
estimate the porosity of rock samples, hence establishing a 559–564, doi:10.1063/1.1148694.
Harris, J. M., Y. L. Quan, and C. T. Xu (2005), Differential acoustic
relationship between porosity and other petrophysical proper- resonance spectroscopy: An experimental method for estimating acous-
ties of reservoir rocks. An improved version of DARS is being tic attenuation in porous media, paper presented at 75th Annual Interna-
developed to apply high pressure and temperature in order to tional Meeting, SEG, Expanded Abstracts.
simulate the in situ response of reservoir rocks. Jackson, I., S. Heather, R. S. Douglas, M. Junjie, and D. Alison (2011), A
versatile facility for laboratory studies of viscoelastic and poreelastic
behaviour of rocks, Rev. Sci. Instrum., 82, 064501(1–8).
[22] Acknowledgments. This work is sponsored by 973 Program King, M. S., and R. J. Marsden (2002), Velocity dispersion between ultra-
“Fundamental Study on the Geophysical Prospecting of the Deep-layered sonic and seismic frequencies in brine-saturated reservoir sandstones,
Oil and Gas Reservoirs” (Grant No. 2013CB228600), CNPC 125 program Geophysics, 67, 254–258, doi:10.1190/1.1451700.
“Multiple-frequency-band analysis of seismic rock physics” (2011A-3606), Spencer, J. W. (1981), Stress relaxation at low frequencies in fluid satu-
the Program for Changjiang Scholars and Innovative Research Team in Uni- rated rocks: Attenuation and modulus dispersion, J. Geophys. Res., 86,
versity (PCSIRT), National Science & Technology Major Project (Grant No. 1803–1812, doi:10.1029/JB086iB03p01803.
2011ZX05019-008), the Science Foundation of China University of Petro- Wang, S.-X., J.-G. Zhao, Z.-H. Li, J. M. Harris, and Y. Quan (2012),
leum, Beijing (KYJJ2012–05-02), and the Nation Natural Science Founda- Differential Acoustic Resonance Spectroscopy for the acoustic
tion of China Research (Grant No. 41274138). measurement of small and irregular samples in the low frequency range,
J. Geophys. Res., 117, B06203, doi:10.1029/2011JB008808.
Xu, C. T. (2007), Estimation of effective compressibility and permeability
Reference of porous materials with Differential Acoustic Resonance Spectroscopy,
Batzle, M. L., D. H. Han, and R. Hofmann (2006), Fluid mobility and Ph.D. thesis, Department of Geophysics, Stanford University, Palo
frequency-dependent seismic velocity-direct measurements, Geophysics, Alto, CA.
71, 1–9, doi:10.1190/1.2159053. Yin, C. S., M. L. Batzle, and B. J. Smith (1992), Effects of partial liquid/gas
Biot, M. A. (1956), Theory of propagation of elastic waves in a fluid saturated saturation on extensional wave attenuation in Berea Sandstone, Geophys.
porous solid: II. Higher frequency range, J. Acous. Soc. Am., 28, 179–191. Res. Lett., 19, 1399–1402, doi:10.1029/92GL01159.
2982