OPM-GFD165-EN-112009-R1.doc 1/68
GFD165 & Control unit
1 Introduction
1.1. Message to Customer
We are proud that you have selected and bought one of our products for your Industrial
application. This product has been built according to our highest quality standard and material
quality. We have designed it in a way that is the most friendly and intuitive to the user as we
always care about your safety and comfort.
Feel free in commenting our job and in bringing back your input to help us building the next
generation of products, improving existing ones or their conformance to your industrial needs.
In this Operating Manual, we have compiled all data available about the product, its service
and main uses. We have also added images, schematics and graphs to ease the reading and
facilitate the learning.
Please read the following pages to fully understand all aspects of the driving software,
operations and cares to your unit as this will assure you having an extended service life. On the
safety side, a careful read is absolutely necessary because of the importance of operating the unit
safely for yourself but also for neighbouring activities.
Copy requests, modifications or additional Commercial data about this manual may be
directed to the Marketing Department (marketing@balteau-ndt.com).
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 2/68
GFD165 & Control unit
Additional information about the manufacturing range and products can be obtained
through the Sales and Marketing Department (sales@balteau-ndt.com or marketing@balteau-
ndt.com)
Your equipment is generally covered by a limited 1 year guarantee which exact terms are
part of the Sales conditions of your purchase. Terms and conditions specified in our specific Sales
contract are supplementing this general rule.
To get benefice of your guarantee, conditions are applying and a specific procedure has to
be followed when communicating with Balteau or your After Sales service Center if such a Service
has been officially appointed in your region.
Any works operated through unauthorized service centers, local repairs attempts without
precise and written indications from BNDT will void the guarantee conditions.
Service operations and warranty applications may be delayed by failing to send required
information of material when requested or in diligent time. IN no cases warranty will be extended
because of such problems encountered (see General Sales and Warranty conditions)
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 3/68
GFD165 & Control unit
We packed your equipment with adequate material and cares to ensure proper protection
for the designed transport.
When receiving your package, carefully inspect the external aspect and shape of the box.
Should this box be damaged, If despite our cares, your unit would be damaged or warped,
If this is not the case, Do not open it before having taken some pictures of the box: if
necessary exert all reserves to the carrier, if possible in writings with a copy to Balteau regarding
the reception of the goods.
When opening the box, check that all elements are correctly packed and that seals if any
are intact.
If the equipment is found defective or broken, advise both the carrier and Balteau and wait
for instructions. If your material was covered by an Insurance, inform the latter that some reserves
have to be taken.
In all cases, pictures of the problem should be taken and a copy sent to Balteau ideally by
Email (shipping@balteau-ndt.com).
When calling BNDT for warranty issues, service or calibration requests you will be asked to
provide several information among these are:
You'll then be given a reference number which will become your unique identification
number throughout the service process for efficient and performing tracking of operations
performed on your equipment.
Unless otherwise indicated, shipment is always at your expenses. In order to reduce all
costs related to transport and transfer to our company, we strongly recommend you to follow our
"ShippingAir" procedure that will be sent to you on request if you contact our Shipping Department
by phone, fax or mail (shipping@balteau-ndt.com)
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 4/68
GFD165 & Control unit
1.6. Calibration
During the manufacturing process your equipment has been calibrated to correspond to the
manufacturing standards and/or other applicable standards.
These operations have been performed with adequate personnel and calibrated material.
We recommend a calibration of X Ray devices every two years at least or 1 year if your
quality requirements are stating so.
Calibration consists in checking several important functions of your equipment like:
· Focal spot check
· HV value check and adjustments
· Dose output check and adjustment
Balteau is offering Calibration services through it's Network (only through Calibration
authorized Service centers) or directly from the factory.
Feel free in consulting us for rates, delays and additional information (service@balteau-
ndt.com).
Balteau is providing a Disposal Service at reasonable costs as part of its global Sales
program. Destructions can only be operated at Balteau's facility.
Any X Ray emitting device sold by Balteau can be destroyed on demand if, after
information, approval and payment of corresponding fees, it is sent free of charges to our main
manufacturing place in Belgium.
Under these conditions, after destruction operations, we will provide the user in return with
a Certificate of destruction mentioning the Serial number of the Unit, the Type and Date of
Destruction or Service removal.
This Certificate may be used for further reference to the National or International Atomic
agencies in the Country where the equipment was imported from.
The necessary documents and information for equipment identification and shipment may
be asked at Balteau (xray.disposal@balteau-ndt.com) .
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 5/68
GFD165 & Control unit
2 Preliminary Information
2.1 Hazardous Material
Industrial equipments may contain hazardous substances and materials that may require
adequate procedure for disposal.
Elements hereafter described are concerning general X Ray equipment and may not all be
applicable to your equipment. Please read the operating manual in chapter "General Description"
or the "Technical datasheet" to know what is applicable to your equipment.
In case of doubts, the manufacturer is offering you a disposal service where you can obtain
updated and more precise data about how to proceed to dispose of your old equipment.
Solid insulation
Solid insulation is non toxic as such: insulating material can be modified epoxy, silicon or
PU. Insulated blocs may contain Copper, Steel, Silver and or Lead. It is not allowed to burn
insulated blocs to remove metallic materials inside as fumes can be toxic. All insulated blocs must
be disposed according to local regulation.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 6/68
GFD165 & Control unit
Explosion risks
When used in pressurized can (gas insulated units) the tube may become pressurized
(Called Gaseous tube) if punctured by an electric arc during X ray operations and therefore must
be processed with extreme cares if removed from the pressurized can. It is not allowed to remove
an insert from its can but if made by an Authorized and duly qualified Service center.
All glass or ceramic inserts must be destroyed inside a specific container to avoid
explosion and projection of glass or ceramic particles. Please ask your closest authorized dealer to
proceed to this operation for you.
Beryllium Window
Some X Ray tubes are equipped with a Beryllium (Be) widow which is toxic by
ingestion. It is not allowed to touch the Be windows bare hands: always wear gloves. Be Insert
must be disposed according to local regulations and considered as toxic material. See chapter
"Disposal" for further instructions and help.
I Ingestion of beryllium can be fatal. In case of human skin contact, thoroughly
wash the concerned area in running water and soap.
I If in contact with eyes, wash them with sterile or clean running water and seek
medical attention immediately
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 7/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 8/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 9/68
GFD165 & Control unit
3 General Description
Equipment presentation
The control panel, CF2000 or HAND-X, uses a microprocessor to control the operating
parameters of the equipment thus giving greater reliability of operation. The panel is designed for
use with all BALTEAU NDT tubeheads of the series. No specific settings are necessary because
the µ-processor is able to recognize the connected equipment when you initiate the system.
Please note that no change of unit can be done without completely stopping the control panel to
reset the original settings and because of safety issues.
Equipment Identification
All connected parts (except the cables) are identified by a silver metallic label which indicates at
least:
· Manufacturer's name and phone number + Email
· Model and serial number of the part
· Manufacturing date
The control panel label is also indicating:
· Mains power and power consumption
The generator label is also indicating:
· Tube type
· Focal spot
· Max kV and mA
Plugs receptacle and connections are clearly identified regarding the function and the wiring
diagram mentioned in the operating manual.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 10/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 11/68
GFD165 & Control unit
Main Specifications
a b c d e f a b g
GFD165 mm 538 269 442 160 138 628 20° 27°30' 55°
GFD165
HV range kV 50-160
tube current range mA 0,1-5
beam angle (°) 55°
focal spot size (IEC336) mm² 1,5x1,5
inherent filtration (al) mm 6 Al
duty cycle* % 100
max operating temp (c°) 70
Insulation gas sf6
weight without rings kg 12
anode cooling type air
*at a temperature of 86 °F (30 °C) reduced mA
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 12/68
GFD165 & Control unit
Baltospot CF2000
Main Specifications
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 13/68
GFD165 & Control unit
Baltospot HAND-X
Main Specifications
HAND X characteristics
Technology Radio data communication error free technology < 10 mW emitter
Distance range m Up to 100
YES, radio communication is automatically deactivated when the HAND
Onboard use
X is placed on board
Generator recognition FULL AUTO with unique ID. attribution
Irradiation alarm unit * YES (GM type *) adjustable level
kV selection 1 kV/step
mA setting 0.1mA/step
Time Step of 1 second, up to 90 min.
Intelligent preheating YES (90 sec. first day + 60 sec/each additional day)
24 touch keys with embossed pads, anti-scratch and long life surface
Keypad
treatment
Display 4 lines, crystal clear, green back lighting, adjustable contrast
Internal accumulator (2,5 A) – 8 hrs charge capacity
Mains Intelligent charging unit is included in the Power Control
Optional table and car charger available
*This unit is not intended to replace the personal dose monitoring which is a legal obligation acc
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 14/68
GFD165 & Control unit
3.3 Generator
CERAM range
Beam
The CERAM generators have directional beam. The x-ray tube is made of metal ceramic with
outside anode. The beam goes through a Beryllium window located in the center of the centering
accessories. One aluminum removable disk protects the Beryllium window. One lead stopper can
be mounted into the window to stop x-ray emission during preheating. One collimator can be
assembled into the window to collimate the output beam. The beam angle is 40°.
Emission point
The tube is centered in the can and the emission point is identified by a small point located outside
of the can. A black line around the unit I is also sometimes present to show the axis of the emitting
point.
Manometer
Located on the top side of the x-ray generator, the manometer indicates the internal SF6 gas
pressure. This pressure is supposed to range from 4,5 and 5 kg/cm² at 20°C.
Ventilator
The x-ray tube anode is cooled by forced air. Any disturbance to the air flow will reduce the
performance of the unit. Air access should always be free from any obstruction.
Interconnecting plugs
All plugs are protected by plastic caps. Before interconnecting elements together, check that all the
contacts are absolutely clean and that no connector pins are bent or damaged.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 15/68
GFD165 & Control unit
3.3.1 Description
BALTOSPOT GFD165
Options:
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 16/68
GFD165 & Control unit
Main Characteristics:
Focal Spot
Exposure performance
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 17/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 18/68
GFD165 & Control unit
3.4.1 Description
CF2000
Interconnecting plugs
1 to the X-Ray generator
2 to the safety accessories
3 to the ground1
4 to the power supply
1 2 3 4
8
7
6
1
The n°4 earth point may be used in replacement or in combination with the one in the main plug. In fact our units have been
designed to be used merely on field where power generators are also widely used. These power gens are not always offering a good
earthing point and one can use a separate grounding "pig tail" consisting in an helicoidal rod that is put in the ground to provide
earthing.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 19/68
GFD165 & Control unit
11
12
10 22 17 19
13 16
14 15
23 23
18 21 20
Keys Description
10 LCD display 4 lines display for requested messages
11 Connection Pilot LED LED ON means OK, LED OFF means connection is lost
12 X-Ray ON PILOT LED ON during pre-alarm time and X-Ray emitting period
13 Display back light ON/OFF allows to turn ON/OFF the backlight
14 Escape key allows to quit the different menu or escape some selected value
15 Menu selection key allows to access to different menus.
16 Exposure chart calculation access to a program to help the calculation of exposure time.
17 Exposure kV selection allows to select kV value according to the possibility of each X-Ray tube.
18 Exposure mA selection allows to select the mA values according to the X-Ray tube capabilities
19 Exposure time selection allows to introduce the T from the exposure.
20 Validation key enter and valid the selected parameters.
21 Stop Key stop X-Ray emission
22 Scrolling keys scroll information up and down, activated scrolling sign is displayed
23 Double start buttons pressed at the same time at least 1 sec it start ON X-RAY
24 Internal buzzer activated during the pre-alarm time before each exposure.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 20/68
GFD165 & Control unit
Beep sound
Sounds if a wrong parameter is selected.
Flashing lamp:
flashing lamp 24V assembled into red cap
cable length 20 m, 10 cores
possibility to interconnect the door switch, an external buzzer, a pre-alarm lamp (24V 5W max),
a power ON lamp (24V, 5W max)
the equipment should not be used without the flashing lamp, if defective refer to maintenance.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 21/68
GFD165 & Control unit
Hand-X Power
The Hand-X Power contains the electronical devices for security and regulation of the X-Ray
generator parameter.
1 2 3 4
8
7
6
5
9
Interconnecting plugs
1 to the X-Ray generator
2 to the safety accessories
3 to the ground2
4 to the power supply
2
The n°4 earth point may be used in replacement or in combination with the one in the main plug. In fact our units have been
designed to be used merely on field where power generators are also widely used. These power gens are not always offering a good
earthing point and one can use a separate grounding "pig tail" consisting in an helicoidal rod that is put in the ground to provide
earthing.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 22/68
GFD165 & Control unit
The remote control HAND-X is the link between the user and the control panel. It is a light weight
box containing the keyboard to access to any parameter of the x-ray unit. Furthermore, it includes
a Geiger counting device warning the operator about any accidental irradiation.
The HAND-X can be operated in two modes : radio mode using antenna and high frequency
transmission, or direct mode without radio frequency.
The right mode is automatically selected following the HAND-X position.
Radio mode
Battery
In radio mode the HAND-X uses his internal battery, having autonomy of + 8 hours when fully
charged.
The battery can either be charged with an external charger or when working on direct mode
through the HAND-X POWER unit. Messages about battery will be displayed on the fourth line, or
when requested through the selection menu.
You place the Hand-X remote control in the housing of the Hand-X power and you turn the key
either on the stand-by position or the X position. The Hand-X power is connected on the main
supply. In other words the Hand-X power needs to be turned on in order to recharge the Hand-
X remote control.
You connect the Hand-X remote control with the battery charger that was supply with the
equipment.
For a complete recharge of the Hand-X remote control you must count on +/- 1 hour.
There is an electronic control of the charge that regulates the in and out of the tension so there is
no problem if the remote control recharge for a longer time. However we do not recommend to
leave the Hand-X recharge during a night or a day.
Direct mode
The radio mode is automatically stopped when the HAND-X is clamped on the HAND-X POWER
unit.
The HAND-X battery is also connected to an internal battery charger charging and controlling the
charge.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 23/68
GFD165 & Control unit
15 18
2
1
4 5
6 7
8 10
12
9 11
16 13
14
17
Keys Description
1 LCD display 4 lines display for requested messages
2 Connection Pilot LED LED ON means connected, LED OFF means disconnected
3 X-Ray ON PILOT LED ON during pre-alarm time and X-Ray emitting period
4 Escape key allows to quit the different menu or escape some selected value
5 Display back light ON/OFF allows to turn ON/OFF the backlight
6 Menu selection key allows to access to different menus
3
7 Exposure chart calculation Automatic parameter calculation if installed
8 Exposure kV selection allows to select kV value according to the possibility of each X-Ray tube
9 Exposure mA selection allows to select the mA values according to the X-Ray tube capabilities
10 Exposure time selection allows to introduce the T from the exposure
11 Validation key enter and valid the selected parameters
12 Stop Key stop X-Ray emission and if pressed 5sec totally switch off the Hand-X
13 Scrolling keys scroll information up and down, activated scrolling sign is displayed
14 Double start buttons pressed at the same time at least 1 sec it start ON X-RAY
15 Battery charger connector allows connection to an external 12 VDC negative
16 Numeric key pad
When the Hand X is placed on the power unit it allows to work without
17 Connection with power unit radio frequency, to take care about the battery charge and proceed to the
recognition to an exchanged sub assembly
18 Computer connection Only works with X com software
3
Basic exposure calculator
Enter the steel thickness value in millimeters and press "ok"
The calculator will give as a result the corresponding kV, mA and time.
The values are calculated for steel, for an exposure with a Kodak AA film and at a focus to film distance of 700mm.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 24/68
GFD165 & Control unit
Control unit
Timer
Exposure time selection through key pad from 10 sec to 99 min, second by second. The timer is
starting value kV are at 75 % from the selected kV value.
CF2000
The control unit contains the electronic devices for security and regulation of the X-Ray generator
parameter.
All field mains from 170 to 260V. (optional 110V) – 45Hz to 66Hz
Power consumption less than 10 amps
Can drive all GF and CERAM Series
Rough, Ergonomic, Waterproof keypad with color tagging for friendly use
4 lines clear display with backlight
Operators most used functions only
Enhanced full automatic preheating function
Memory of 100 last exposures and additional parameters
Manual or automatic time and exposure calculation (optional)
Built in RS 232 interface
Compact sized
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 25/68
GFD165 & Control unit
Hand-X only
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 26/68
GFD165 & Control unit
Control Unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 27/68
GFD165 & Control unit
4 Options
Extra external buzzer
An external buzzer can be interconnected to the security plug from the CF2000 or inside the
flashing light box (refer to drawing 4811167).
Interconnecting cable
In standard the unit is delivered with 20 m length cable. Other length can be available on request.
We advise you, this cable is a special one designed to get a free distribution communication
between the CF2000 and the X-Ray generator. Any other cable used is forbidden and may destroy
or damage the unit.
The max length is 75 m.
The performance of the unit is automatically adjusted if cable length is modified.
Door switch
The door switch should have one NO contact with contact impedance better than 10 Ohm when
closed. This contact has to be voltage free connection possible on the Hand X security or in the
flashing light box.
Tripod support
Support to handle and position the X-Ray generator.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 28/68
GFD165 & Control unit
5 Operations
5.1 Preliminary Operations
5.1.2 Interconnection
3 Cables are to be connected to the control unit:
1 is connected to Clignotix & Security
1 is connected to the generator
1 is connected to the mains
Hand-X only : If the control unit is a Hand-X, the Hand-X remote control has to be placed in its
cradle (Direct mode) before starting the unit.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 29/68
GFD165 & Control unit
5.1.3 Start
Before starting the unit, all elements are independent one to the other. To make them operating
together, a specific "pairing" needs to be done. This is achieved both by interconnecting all
elements together and by switching the elements ON following next steps:
· Turn the 3 position key switch from “OFF” (position 0) to “ON” (position 1)
· You can now hear the warning fan
· If all the interconnections have been done correctly and a security (door switch) is
not opened, a welcome message will be displaying
o the Name of the product,
o the Software version,
o the Name of the Manufacturer
After about 1 to 3 seconds with the Welcome message, the right connection and pairing status will
be indicated by the two green leds (one on each side of the display) that are now ON (Lit). The
display will then show name of the generator connected and corresponding parameters
IThe communication protocol is patented and secured for the maximum safety. Should
the communication be interrupted for any reasons, the unit will go in Stand by mode and
the display will show "link lost"
IPairing assures the user that only interconnected elements at the time of starting will be
operating together. This operation called “Pairing” must be repeated each time one of
the components changes (X-ray generator, control unit or remote control) or power
supply is interrupted.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 30/68
GFD165 & Control unit
Hand-X only : During the starting up, the control unit makes the full recognition of the X-ray
generator and of all of its parameters, and establish one unique Radio Frequency protocol
between the control unit and the HDP to get a real safe transmission.
.
IThis pairing cannot be done correctly if the remote control is not plugged
correctly into the control unit or if a security switch is open (door).
· Right after the operation the display will show up with the data ready to be set up (kV,
mA, time) and also the “Ready to start” message or “Remote link lost” if the
interconnections are not well done.
· You can now enter your settings. (See chapter “Principle of operations)
Hand-X only : The Hand-X can be shut off by pressing the stop button during 5 seconds. The
Hand-X shut off automatically if any key is used during 30 minutes.
5.1.5 Note
The external clignotix has to be connected in order to start the exposure (safety rule).
If the clignotix is in failure or not connected, in order to work you have to deactivate the
clignotix watching safety system by going into the menu (see specific chapter) to disable it. After
any reset of the control unit the watching safety system has to be again disabled.
Clignotix®: this is a Balteau patented name for designing "warning flash light". In fact it is
not a simple light that we use. We use a Xenon flash light which emits flashes every ½ second
around. This device is also acting like a safety device as the µchip inside the HDP (control unit
Hand X or CF2000) is checking frequency and power consumption of the Clignotix® to determine
whether it is or not functional. This Clignotix® must then be only replaced by an original spare part.
IAlthough we used specific vinyl clear plastic that is known to be robust and lightweight,
the first cause of a Clignotix failure is not the bulb but a breaking due to shock or harsh
treatment. In that case replace the clignotix as soon as possible. You can temporarily
deactivate the function on the keypad. You’ll get then a blinking warning message
mentioning “Clignotix Off” until you replace it by an original spare part.
Inside the Clignotix® may be found few straps to enable or disable the door security switch,
beacon and other optional devices.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 31/68
GFD165 & Control unit
Direct mode is a mode in which the Hand-X is used on its cradle and there are no RF
emission. This mode will be preferred in areas where either RF emission is not allowed or where it
can be heavily perturbed or instable;
IIn Direct mode the GM is active. Therefore if the dose at the HDP location exceeds the
preset value, the unit will not be able to be started.
Remote mode is the working mode in which the Hand-X is separated of its cradle and
operates the unit remotely. This mode is useful for field operation where the countering of the area
is necessary. It also allows the user to stay on the safe side regarding exposition to dose.
RF connection
The Hand-X is RF connected and speaking in bi-directional mode all the time the Hand-X is
in RF mode.
The distance from Hand-X and Hand-X Power is between 20 and 100 m. The RF
connection is perturbed or lost, the message REMOTE LINK LOST will appear on the display, then
the X-Ray emission are stopped. Start button may restart the X-Ray emission providing the RF
connection is OK. If not, go nearer the Hand-X Power or work in direct mode.
Battery
The Hand-X is powered with an internal battery having life time of +/- 8 hours when fully
charged. An external battery charger can be used for charging the battery. When the Hand-X is
clamped on the Hand-X Power unit, the battery is also automatically recharged.
The RF mode
Allows you to have max 3 units working in the same time in the same area, and thus
without any action from one to the other one.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 32/68
GFD165 & Control unit
5.2.1 Preheating
Principle
Every day, the first starting of X-Ray must be done with gradually kV increase. This time or
operation is called preheating. This curve has been memorized into the Hand-X software for each
X-Ray generator and rest time. This preheating will be requested automatically after a rest time
specific for each X-Ray tube.
Preheating
If requested, the control unit will ask you to proceed to the preheating. The control unit will
show you the needed time.
By pressing the two start buttons, you will start the preheating sequence, the display shows
the remaining time and the kV reached. If you stop the preheating process, it will only be possible
to work below the reached level (90% of this value after full cut off of the unit). If you select higher
kV value, preheating will again be requested.
Preheating can also be requested manually through the menu, this manual preheating will
never be shorter than the automatic one. This point needs to be addressed carefully as it is mainly
an after sales service point of menu. Manual preheating is never recommended nor has to be
used. Indeed the Control unit is able to calculate accurately the necessary preheating time for the
tube that is connected on it. The only exception for performing a manual preheating can be to
bypass encountered difficulties with a "capricious" generator or when the µChip component of the
HDG board (under the Generator's cap) has been changed (after sales or servicing).
IDuring the preheating, the X-Ray generator is producing X-Ray, always position the
output window in a way that prevents any hazardous exposure or place an appropriate
shutter to keep the dose at an acceptable level.
Preheating duration
For the CERAM235 and 35, the minimum duration will be 1 minute after 10 hours rest.
This time will be increased of 90 seconds for each rest day with a maximum of 45 minutes.
For the GFD165 - GFC165 – GFC205 – GFC305 the minimum duration time will be 5 minutes.
This time will be increased by 30 seconds for each rest day, with a maximum of 30 minutes.
End of preheating
At the preheating end, the display will tell the control unit is ready.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 33/68
GFD165 & Control unit
Parameter selection
· kV selection
Press the kV key, the kV value on the display is blinking
Introduce the requested kV through the key pad and press OK
The "mA" value will then be blinking
· mA selection
introduce the requested mA value.
press OK
the "T" will then be blinking.
· T selection
introduce the requested time (minute, point, second)
Press OK
The "Esc" key allows you to cancel and re-enter a new value.
kV – mA – Time– Limitations
· kV - mA:
In the graphic here below, you find the kV and mA limits for the X-Ray generator. If you select
value located outside this graphic, they will automatically be adjusted to the nearest value
with kV priority.
Before X-Ray start, a pre-alarm time with flashing red light, buzzer activated can be adjusted.
GFD165
6
5
4
mA
3
2
1
0
0 50 100 150 200
kV
· Time:
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 34/68
GFD165 & Control unit
X-ray start
I When power is applied to this equipment, it generates ionizing radiations that can
cause health damage. Operators should therefore fully be acquainted with all
operating conditions before switching on the equipment.
I This equipment should be used only as specified in the operating manual
no person other than a suitably qualified competent operator should use this
equipment
I Persons using this equipment should be registered as "classified" under the
"ionizing radiation’s regulations" and carry a personal radiation dosimeter at all
times when using it.
I Leakage radiation of equipment should be measured at least once a year
The start of the X-Ray emission is done by pressing simultaneously the TWO GREEN KEYS.
If the start cannot be done, a message will appear on the fourth line from the display (refer to
message in chapter Maintenance).
The timer will show the remaining time and at zero will stop the X-Ray emission. The selected
parameters are coming back on the display at the end of the X-Ray emission.
To change those parameters, press escape key and enter new values.
X-ray stop
At any time, you may stop the X-Ray by pressing the STOP key from the control unit (or the
emergency push button on the remote control).
After the X-Ray emission, if the equipment has to be disconnected, we advise you to let the fan
running 2 or 3 minutes before turning the key to 0 and disconnecting the unit.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 35/68
GFD165 & Control unit
Comfort functions
· MENU
The menu key gives you access to the menu and sub-menus.
To access to one of those menu, select the number and enter. To get out press escape.
This mode is automatically selected when you start the control unit
It gives access to the exposure parameters and to the production of X-rays.
Sub menu 1 Display the 100 last exposure parameters by scrolling the table line
by line.
Sub menu 2 Allows to store or restore one of the exposure with all parameters.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 36/68
GFD165 & Control unit
SOUS-
HAND-X CF2000
MENU
"Force Preheating"
"Battery Autonomy"
SUB-MENU 1 Allows to ask manual preheating by
Inform you on the battery charge level
introducing the requested time
"Set clock"
"Language"
SUB-MENU 4 Allows you to introduce actual date and
Language selection for the messages
time
"Display contrast"
"Prewarning time" allows to adjust the LCD display contrast.
SUB-MENU 5 pre alarm time adjustment from 3 to 30 Position 110 is a median position used in most
sec circumstances. At low temperature, this level
can be adjusted.
"Radiation detection"
alarm set up ON or OFF and if activated
SUB-MENU 7 to fix the levels of irradiation read with -----------------------------------------------------------
the Geiger.M
"Display contrast"
allows to adjust the LCD display contrast.
SUB-MENU 8 Position 110 is a median position used in -----------------------------------------------------------
most circumstances. At low temperature,
this level can be adjusted.
"Main power detection"
SUB-MENU 9 shows detected power supply (110 AC -----------------------------------------------------------
or 220 AC).
I The option (clignotix disabling) is not available in some countries where the
legislation is not authorizing it.
After each reset of the hand-x power (Hand-X or CF2000), the safety will be re-enabled.
When disabled, the message "CLIGNOTIX UNACTIVE" appears on the 4th line of the LCD display
to inform the user.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 37/68
GFD165 & Control unit
· MENU 4 Maintenance mode
This mode shows the software version and the working hours of each unit
(Hand-x power, Hand-x and x-ray generator).
Hand-X only
This chapter describes the alarm given from the Geiger Hand-X which can be activated
through the menu 3 – GM. ((Menu 3 - Item 7)).
IThe Geiger Mueller tube placed in the Hand-x remote control is not a to be considered
as a calibrated measurement device. It is not replacing the personal monitoring unit but
has to be used as an indicative media for dose survey and monitoring.
IGM tube linearity varies with Energy and recombination in the chamber.
IGM tubes are not intended to be used as a monitoring device below 60 kEV.
The Geiger control is made of an ionization chamber containing a given air volume. When
an ionization particle is crossing this chamber the ionizing capacity of the particle delivers a charge
collected by a cathode converted into an impulsion (count) by a specific circuitry. The number of
counts can be converted into a dose rate at a given energy. Those pulses are pushed into a
calculator which is reset every 10 sec.
If activated (Alert Active set to "1") in the menu, 2 independent levels can be adjusted: one
is Alert and the other is Stop. Alert comes before a Stop and obviously must then always be set
below the Stop level. The first level is a "Warning only then Stop" level, the 2nd level is "Immediate
Stop".
Each scale can be adjusted from 2 to 255.
- The first "level", when reached by X-Ray beam, will show a message RADIATION
DETECTED on the display. The buzzer will be activated (soft ³ V1.6). After 5 seconds,
if the radiation level didn't lower, the HDX will switch off the Generator to preserve the
operator from getting a too high exposure.
- The second "level", when reached will immediately stop the X-Ray emission and display
the message RADIATION HIGH.
Indicative values
IWhen the Dosimeter has been deactivated on purpose from menu 3-7, the 4 line on the th
operation display is flashing during operations with the message "Alert Unactive"
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 38/68
GFD165 & Control unit
6 Maintenance
6.1 Preventive maintenance
Fan
Power unit and X-ray generator are equipped with 24 VDC big fans. The heat exchanger and Fans
have to be regularly cleaned.
Plugs-cables
The plugs contacts should remain clean dry and without arcing trace.
The cable should be free for many mechanical damage. Any damage has to be repaired or cable
replaced.
Flashing lamp
If detective, replace the complete module 24 V 5 W.
6.2 Fuses
F1 to F4 fuses are all placed into the control panel on PCB120
F1 fuse
Placed on PCB120 - value 2A F
This fuse is in the 24volts and feeds the HDG board and the 2 fans from the X-ray generator.
F2 fuse
Placed on PCB120 - value 2A F
This fuse protects the 24volt used for all external accessories.
F3 fuse
Placed on PCB120 - value 0,5 A F
This fuse protect the clignotix (flashing light)
F4 fuse
Placed on PCB120 – value 0,25 A F
This fuse is in the door switch circuit and protects the control panel from any voltage coming from
the door switcher.
External fuses
The complete control panel is protected by 2 fuses 10AT. They are located near the mains
connector.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 39/68
GFD165 & Control unit
HAND-X POWER
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 40/68
GFD165 & Control unit
Old version (bulb) 0 460195 0
New version LED
Red light
Old version (bulb) 0 460015 0
New version LED 0 460194 0
Green light
Old version (bulb) 0 460016 0
New version LED 0 460196 0
Bulb for light 0 460077 0
Emergency switch 0 420145 0
Key switch 0 420008 0
24 V relay 0 422004 0
Fuse holder 0 427003 0
Fuse support 0 427004 0
Fuse 16AT Dim: 6,3x32 0 427002 0
12 V fan 0 462012 0
PCB HDP 120 0 820912 0
Fuse 0,5AF Dim : 5x20 0 427064 0
Fuse 2AF Dim : 5x20 0 427061 0
Power Module 0 820200 0
R.F. box (HDR) 0 820187 0
HAND-X
3.6 V battery 0 451020 0
Fuse 1,6AT Dim : 5x20 0 427059 0
CABLES
Power cable (10m) 0 820177 0
20m connection Cable 0 820179 0
HX/baltospot
Clignotix & cable 0 820178 0
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 41/68
GFD165 & Control unit
8 Trouble Shooting
RADIATION DETECTED
The Geiger has detected radiation
the radiation level can be adjusted by the menu.
RADIATION HIGH
The Geiger has detected high level or radiation and has stopped the X-Ray.
the radiation level can be adjusted but be careful of radiation. See menu.
REMOTE STOP
The X-Ray emission has been stopped by stop button.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 42/68
GFD165 & Control unit
CHOPPER FAIL
Internal Hand-X Power fault
test fuses
contact service center
defective x-ray tube insert or faulty power block
DEMAG FAIL
Internal demagnetization Hand-X Power fault
test fuses
contact service center
X-RAY KEY
The 3 position key is not in X position
turn the key in X position
EXTERNAL SECURITY
Door switch activated or not connected
-test the fuse 2A inside the panel and located on PCB120
EMERGENCY STOP
Emergency button activated
release the emergency button
12 V LOW
24 V LOW
12 or 24 Volts Hand-X Power voltage low
test fuses inside the panel on power supply
contact service center
MAINS LOW
The main voltage lower than 170 V
check voltage supplies
24 V I MAX
Too much current in the 24 V going to the external uses
disconnect extra security
Extra security are all wires, loops and external securities belonging to a fixed installation that can
be connected to the unit to interrupt X Ray production in case of activation. All these external
devices must comply to the requirements written otherwise there are serious hazard of internal
PCB damages.
INTERLOCK
Hand-X Power internal bad contact
check internal interconnection
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 43/68
GFD165 & Control unit
mA HIGH
Tube current too high
contact service center
mA LOW
Tube current too low or such
check cable and plug contact
contact service center
kV HIGH
kV to high regarding to normal value
contact service center
kV LOW
kV to low regarding to normal value
check cable and plug contact
check mains power supply
contact service center
VCC SECURITY
Measured KV different from the requested KV
Contact service center.
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 44/68
GFD165 & Control unit
The Hand-X gives the message REMOTE LINK LOST as soon you remove from the Hand-X
Power
detective antenna
defective RF module
too much disturbances on the working area
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 45/68
GFD165 & Control unit
9 Annexes
9.1 Block diagram / Electrical drawing
Overview 5.811.686
GFD165 & GFC165 schema 4.811.317
General schema Power Unit 4.811.298
Flashing light and safety interconnections 5.811.246
Schematic bloc HDX 4.811.276
Schematic bloc HDX 4.811.278
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 46/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 47/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 48/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 49/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 50/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 51/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 52/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 53/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 54/68
GFD165 & Control unit
9.3 Revision
Modification notes
Issue Reference Date Comments
0 06302150 October 2007 Original version
1 06302150 November 2009 Preheating
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 55/68
GFD165 & Control unit
9.4 Certification
- CEE conformity
- Controlatom certificate
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 56/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 57/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 58/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 59/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 60/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 61/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 62/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 63/68
GFD165 & Control unit
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 64/68
GFD165 & Control unit
Table of content (1/2)
1 Introduction .................................................................................................................... 2
1.1. Message to Customer ...................................................................................................... 2
1.2. Manufacturing program................................................................................................... 3
1.3. Warranty & after sales..................................................................................................... 3
1.4. Damages during transit .................................................................................................... 4
1.5. Sending an equipment to Balteau NDT............................................................................ 4
1.6. Calibration ...................................................................................................................... 5
1.7. Equipment disposal ......................................................................................................... 5
2 Preliminary Information................................................................................................. 6
2.1 Hazardous Material ......................................................................................................... 6
2.1.1 Insulation material ................................................................................................... 6
2.1.2 X Ray tube............................................................................................................... 7
2.1.3 Canister, housings and electronic components ......................................................... 7
2.2 Safety & Health............................................................................................................... 8
2.2.1 Radiation emitting material...................................................................................... 8
2.2.2 Usage of equipment ................................................................................................. 8
2.3 Personal qualification ...................................................................................................... 8
2.3.1 Overview of a minimum training course for operating an X ray tube........................ 9
3 General Description ..................................................................................................... 10
3.1 Box Content .................................................................................................................. 11
3.2 Data Sheet ..................................................................................................................... 12
3.3 Generator ...................................................................................................................... 15
3.3.1 Description ............................................................................................................ 16
3.3.2 Technical features.................................................................................................. 17
3.3.3 Lay out & pictures ................................................................................................. 18
3.4 Control unit ................................................................................................................... 19
3.4.1 Description ............................................................................................................ 19
3.4.2 Technical features.................................................................................................. 25
3.4.3 Lay out & pictures ................................................................................................. 27
4 Options ......................................................................................................................... 28
5 Operations .................................................................................................................... 29
5.1 Preliminary Operations.................................................................................................. 29
5.1.1 Advise before interconnection................................................................................ 29
5.1.2 Interconnection...................................................................................................... 29
5.1.3 Start....................................................................................................................... 30
5.1.4 Shut off.................................................................................................................. 31
5.1.5 Note....................................................................................................................... 31
5.1.6 Hand-X only.......................................................................................................... 32
5.2 Principle of operations................................................................................................... 33
5.2.1 Preheating.............................................................................................................. 33
5.2.2 Operating mode ..................................................................................................... 34
6 Maintenance ................................................................................................................. 39
6.1 Preventive maintenance................................................................................................. 39
6.2 Fuses ............................................................................................................................. 39
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 65/68
GFD165 & Control unit
Table of content (2/2)
7 Spare parts List ............................................................................................................ 40
8 Trouble Shooting ......................................................................................................... 42
8.1 Error messages .............................................................................................................. 42
8.2 Fault trouble shooting level 1 ........................................................................................ 45
9 Annexes ........................................................................................................................ 46
9.1 Block diagram / Electrical drawing................................................................................ 46
9.2 Exposure charts ............................................................................................................. 53
9.3 Revision ........................................................................................................................ 55
9.4 Certification .................................................................................................................. 56
9.5 Service and Repairs Flow Chart..................................................................................... 64
10 Notes ............................................................................................................................. 67
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 66/68
GFD165 & Control unit
10 Notes
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 67/68
GFD165 & Control unit
Balteau in words…
BALTEAU NDT is a company based and established in Belgium since 1906. The activity of
the power transformers successfully turned into the X Ray business in 1932. Years after years
BALTEAU has affirmed its leadership leading the technology and setting the foundation of modern
X Ray technologies. The first company ever to reach 200, 300 and 400 kV in the past, Balteau has
today brought to the industry new standards that are consciously Safety oriented to allow users to
operate equipment safer and easier. Balteau is also greatly caring about environment and always
uses top grade and specification materials to ensure that the efficiency of the equipment will be at
the top of what is technologically achievable.
On the service and spares side, BALTEAU is providing the buyer with an exceptional and
unequalled commitment of continuous service all through the life cycle of the product.
Qualified engineers and an extensive Worldwide network dedicated to serve our Customers
is actively participating to the success of our brand and to your possibility of always getting, from
the selection of your equipment to its follow up in service, a qualified and professional support.
Balteau in numbers…
BALTEAU has a very long established network in more than 40 countries, some of our
agencies are of second generation proving they rely on the quality of our products and that they
are totally committed to their business in NDT.
BALTEAU has produced more than 13.000 generators spread in more than 60 countries.
We have 3 ranges of products and are the only manufacturer in the world to provide from
Crawlers to Systems built into the same factory.
Balteau in names…
Producer Distributor
Balteau NDT sa
Voie de Liège, 12
B-4681 Hermalle Sous Argenteau
BELGIUM
All specifications are non contractual and are subject to change without prior notice – Exact at the date of edition
OPM-GFD165-EN-112009-R1.doc 68/68