C = Co(1 - B [ 1; - ~])
For a Cannon-FenskeOpaque viscometer,the following equationis used to obtain the
viscometerconstantC.
c Co(1 + B [ T, - ~ J)
Pleasenote that the two equationsare identical exceptfor the sign beforeB. In theseequations,
If you are measuringthe kinematic viscosity of liquids other than mineral oils, the following
equationpermits the calculation of the temperaturedependencefactor for the liquid:
SampleCalculationfor a Cannon-Fenske
Routine Viscometer
A typical calculation, using data from a calibration certificate for a size 450 Cannon-Fenske
Routine viscometer, is given here. The certificate itself is included at the end of this article.
First Co is calculatedfrom the actual value of C determinedby calibrating the viscometerat
40°C. Cannondeterminesthe viscometerconstantsusing two different standards,then takes
the average of the two constants. One standard is normally sufficient whenever a user
recalibrateshis or her viscometer. In this particular case,a constantof 2.540 mm2!s2was found.
The value ofB for this viscometeris 81 x lO-6/oC. The room temperature(fuling temperature)
was 23°C. Plugging thesevaluesinto the equationfor the Cannon-FenskeRoutine, we obtain
the following:
The calibration constantfor the viscometer at 100°C can now be calculatedusing the same
equation.
C1OO = 2.544[1 - «81 x 1Q-6)(100 - 23»]
A typical calculation, using data from a calibration certificate for a size 100 Cannon-Fenske
Opaqueviscometer, is given here. The certificate itself is included at the end of this article.
First Cois calculatedfrom the actual value of C determinedby calibrating the viscometer at
40°C. Cannondeterminesthe viscometerconstantsusingtwo different standards,then takesthe
averageof the two constants. One standardis normally sufficient whenevera user recalibrates
his or her viscometer. In this particular case,a constantof 0.01446 mm2js2was found for bulb
C. The value of B for bulb C of this viscometer is 121 x 10"6jOC. The room temperature
(filling temperature)was 23°C. Plugging thesevaluesinto the equationfor the Cannon-Fenske
Opaque,we obtain the following:
The calibration constant for the viscometer at 100°C can now be calculated using the same
equation.
The aboveconstantsassumea value for the coefficientof thermal expansiontypical to that for mineral oil, and that the
viscometerwas fllled wilh test sampleat room temperature. If the filling temperatureT p is substantiallydifferent than
- -
room temperature, the viscometer constant at test ten1perature T'( is Co (1 B (TT TpD. The valuesof Co and B
shownbelow are basedon a coefficientof thermal expansiontypical to that for a mineral oil.
Kinematic viscositiesof the standardsused in calibratingwere establishedin Master Viscometersas describedin Ind.
.Eng. Chern.Anal. Ed. 16,708(1944), ASTM D 2162,and the Journal of Researchof the National Bureau of Standards,
Vol. 52, No.3, March 1954,ResearchPaper2479.
Kinematic viscositiesare basedon the value for water adoptedby the National Institute for Standardsand Technology
and Thc American Socicly for Tesling Matcrials July 1, 1953. Thc kinematicviscositybasisis 1.0038mm2js( cSt) for
water at 2Q°C(ITS-90). The gravitational constant,g, is 980.1cm/sec2at the Cannon Instrument Company. The
gravitationalconstantvariesup to 0.1% in the United States. To make this small correctionin the viscometerconstant,
multiply the above viscometerconstantby the factor [g(at your laboratory) /980.1]. The calibration data below are
traceableto the National Institute for Standardsand Technology. Temperaturemeasurementtraceableto NIST (Test
No. 246089). CALIBRATION DATA AT 40°C
The S.l. unit of kinematicviscosityis 1 meter squaredper second,and is equalto 104stokes.The S.l. unit of viscosityis
1 pascalsecond,and is equal to 10 poises.One centistokcsis equalto one millimeter squaredper second. 991
[e
V Iscome
. t er N O. AOOO
100
The S.l. unit of kinematic viscosity is 1 meter squared per second,and is equal to 104stokes.The S.l. unit of viscosityis
1 pascal second,and is equal to 10 poises.One centistokesis equal to one millimeter squared per second. 691