Documente Academic
Documente Profesional
Documente Cultură
User Manual
2
Contents
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2 Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
2.1 Functional Components of the CT Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.1.2 Mains Connection Unit and Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.1.3 Compact Flash Card Slot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.1.4 Inputs and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.1.5 I/0 Key with Status LEDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.1.6 Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2.2 Block Diagram (Simplified) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3
CT Analyzer
4
Contents
5
CT Analyzer
6
Contents
7
CT Analyzer
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .231
8
Preface
Preface
The purpose of this user manual is to familiarize users with the CT Analyzer and
its application fields. It contains helpful instructions on how to use the
CT Analyzer safely, properly, and efficiently.
Following these instructions will help you to prevent danger, repair costs and
possible down time due to incorrect operation. Furthermore it ensures the
reliability and life-cycle of your CT Analyzer.
Glossary of Symbols
In this manual, different symbols are used to highlight text of special safety and/
or operational relevance. These symbols are listed below.
) Note
Indicates notes with special meaning, i.e., additional important
information.
Caution
Indicates sections with special safety-relevant meaning.
9
CT Analyzer
10
Safety Instructions for CT Analyzer
Safe Operation
) If working in other than laboratory environment only use the CT Analyzer with a
solid connection to ground of at least 6mm2. Use a ground point as close as
possible to the test object.
• When disconnecting cables, always start at the device feeding the power.
• Never connect or disconnect a test object while the outputs are active. Lethal
voltages can occur due to the high energy stored in external inductors.
• During the test always connect one terminal of the transformer’s primary side
to protective earth.
• Do not insert objects (e.g. screwdrivers, etc.) into the ventilation slots or any
input/output sockets.
• Before putting the CT Analyzer into operation, check the test set for visible
damages.
• Do not operate the CT Analyzer under wet or moist conditions
(condensation).
• Do not operate the CT Analyzer when explosive gas or vapors are present.
• When taking the CT Analyzer into operation, make sure that the air slots, the
power switch and the power supply plug at the test set remain unobstructed.
• Voltages of up to 400V can occur inside the CT Analyzer! Therefore, only
qualified personnel at the factory is permitted to open the CT Analyzer.
• If the CT Analyzer seems to be functioning improperly, please call the
OMICRON Hotline (refer to chapter "Contact Information / Technical
Support" on page 45).
• For protection against parasitic currents or voltages, always connect the
equipotential connector on the CT Analyzer’s side panel (see section 2.1.2
on page 20) to protective ground (PE). Only use the cable set listed in section
14.2 on page 223.
11
CT Analyzer
• Make sure that the terminals of the test object to be connected to the
CT Analyzer do not carry any voltage potential. During a test, the
CT Analyzer is the only permitted power source for the test object.
• Never connect or disconnect a test object while the outputs are active.
Note: Even if you switched off the CT Analyzer, wait until all LED lights
are fully extinguished.
As long as an LED lights or the display is on, there may still be
voltage and/or current potential on one or more of the CT Analyzer
outputs.
• When measuring the ratio of transformers make sure that the test voltage is
connected to the corresponding secondary winding, and the primary winding
is connected to the according measurement input. Accidentally mixing up the
windings can cause life-threatening voltages within the transformer and/or
destroy the connected CT or the CT Analyzer!
• Only use wires with 4mm safety "banana" connectors and plastic housing for
connection to the front panel input/output sockets.
• Do not stand right next to or directly underneath a connection point because
the clamps may fall off and touch you.
• Flashing lights on the front panel indicate that the CT Analyzer output is
active and hazardous voltages can be present on the input or output
terminals.
• If the CT Analyzer or any add-on device or accessory does not seem to
function properly, do not use it anymore. Please call the OMICRON hotline.
Power Supply
• Supply the CT Analyzer only from a power outlet that is equipped with
protective ground (PE).
• Instead of supplying the CT Analyzer from phase-neutral (L1-N, A-N), it may
also be supplied from phase-phase (e.g. L1-L2, A-B). However, the voltage
must not exceed 240VAC.
• If there is a problem with the PE (protective earth) connection or if the mains
power supply has no galvanic connection to earth, the CT Analyzer displays
the error message 901. If you ignore this error message, it is possible to use
the device but safety is no longer given!
12
Safety Instructions for CT Analyzer
Changing Fuses
• Turn off the CT Analyzer and unplug the power cord.
• Ground the test object, and disconnect it from the CT Analyzer. By
disconnecting it you prevent a possibly faulty test object feeding power back
into the CT Analyzer.
• Locate the blown fuse on the side panel of the CT Analyzer and replace it:
2 x T6.3H 250V (6.3 Amps slow-acting high breaking capacity wire fuse
5x20mm).
13
CT Analyzer
14
Designated Use and Features
1.2 Features
• Very small and lightweight (< 8kg/17lb), particularly beneficial for on-site
testing.
• Reduced commissioning time due to fully automatic testing within seconds.
• The test results are displayed in a graphical user interface.
• Allows testing according to IEC 60044-1, IEC 60044-6 or IEEE C57.13 (ANSI
standard).
• Maximum output voltage = 120V (safety).
• Automatic assessment of the test results.
• Automatic analysis of CTs with unknown data ("guesser" function). This
means that for CTs defined according to IEC60044-1, testing can be
performed without any previous specification of CT data by the user. The
CT Analyzer is able to determine the CT data (I-prim, I-sec, class, ratio, etc.)
and to assess whether the connected CT fulfills the specifications or not. The
guesser function is intended as an aid for the user to find out the data of
unknown CTs.
• For measuring transformers: A typical accuracy of 0.02% / 1’ enables field
calibration and verification of class 0.1 CTs for metering.
15
CT Analyzer
16
Designated Use and Features
17
CT Analyzer
18
Hardware
2 Hardware
2.1 Functional Components of the CT Analyzer
2.1.1 Overview
Figure 2-1 provides an overview of the operating and display elements and the
connectors of the CT Analyzer.
Figure 2-1:
CT Analyzer overview Output Display Remote control interface Equipotential
Generator output with context-dependent Serial interface for full connector
keys ("soft keys") remote control (see 13.8)
19
CT Analyzer
ON/OFF switch
Eject button
Press to eject the Compact
Flash card.
20
Hardware
Sec
Measurement input for secondary side of CT,
300VAC max., 500 kΩ input impedance
Prim
Measurement input for primary side of CT,
30VAC max., 150 kΩ input impedance
21
CT Analyzer
2.1.6 Display
Figure 2-5:
Display Card area with tab Soft key description Soft keys, labelled by the soft
fields key description fields
Status
line
Edit fields Display field If 3 points are visible in the lowest soft key
(underlined by (not underlined) description field (see dashed arrow), this
dotted lines) key can be used to open an additional set
of soft keys
22
Hardware
RI
120VDC
RI
Mains Potential
RI: 400V to all other
potential groups
Bandwith
limiter EMC filter
250V 10A T
Output
0...120V max.
Current Shunt
Temp detect
Current measurement
and relay control
Duty cycle control
Overtemp,
overload
Bridge control
Relay control and temperature read
Control for current measurement
RI 300V
MIF
Gain adjust
Range switching
control signals AD-
converter
EMC filter
Sec
Analog Gain control 0...300VAC
input data
RI 300V
EMC filter
Prim
Gain control 0..30VAC
RI 300V
LCD (240x128)
ON/OFF
Contrast-
adjust
LAMP
ON/OFF
Inverter
Keyboard
Digital Signal
Processing
Display Unit and
Keyboard (DUK)
SELV Potential
RI = Reinforced insulation
Remote Control Interface
BI = Basic insulation
23
CT Analyzer
24
Connecting the CT Analyzer
25
CT Analyzer
26
Connecting the CT Analyzer
• Make sure that the primary side of the CT is connected to PE on one side and
open on the other side.
It is absolutely important to avoid coupling of interferences into the primary
circuit during measurement. Therefore, connect the side of the primary circuit
that is able to receive more interferences to PE (e.g. the side with the longer
line length). The ungrounded side should be the side that receives fewer
interferences (refer to figure 3-2).
• Disconnect the hot side of all secondary windings of the CT in order to
remove any load from the CT. Every kind of load remaining on the secondary
side of the CT during measurement leads to incorrect measurement results
or error messages.
• Connect the black socket of the CT Analyzer input "Prim" to the grounded
side of the CT’s primary circuit and the red socket of this input to the open
(ungrounded) side.
• Connect the black sockets of the CT Analyzer "Output" and input "Sec" to
that terminal on the secondary side of the CT that is connected to PE.
• Connect the red sockets of the CT Analyzer "Output" and input "Sec" to the
other (ungrounded) terminal on the secondary side of the CT.
27
CT Analyzer
OK X
Do not use!
4 wire connection: 2 wire connection:
The measurement leads from the Separate measurement leads are
"Output" sockets and input "Sec" of used for the "Output" sockets and
the CT Analyzer are connected to the input "Sec" of the CT Analyzer, but
test object via separate clamps. the measurement leads are
connected to the test object via a
One clamp is used for each
common clamp.
measurement lead! Correct!
Two measurement leads use one
single clamp! Do not use!
28
Connecting the CT Analyzer
Physical Background
When using the 2 wire connection technique as shown in figure 3-3,
considerable measurement errors for the ratio error and other parameters can
occur, if the connection of the clamp to the test object is not ideal and has a
contact resistance. Since this contact resistance will change with each
connection/disconnection of the clamp at the test object, it is furthermore not
possible to reproduce such measurements.
The total contact resistance caused by 2 clamps can be up to 0.5 ohms and thus
can cause an additional burden of up to 12.5 VA. Due to this, the influence of the
contact resistance is the higher the smaller the winding resistance of the CT is:
When measuring a 5 A CT, such a contact resistance can lead to completely
incorrect measurement results while it is practically negligible when measuring
a CT with a relatively high winding resistance of e.g. 10 ohms.
Therefore, measurements on 5 A CTs should exclusively be performed using
the 4 wire connection technique!
The following connection diagrams make this clear.
Figure 3-4:
Connection diagram for Voltage drop caused by contact
4 wire connection resistance of the clamp
technique Ip RCT OUTPUT
S1
Rcl.
P1
UC IL
IC ICT ~
Np Ns
Terminal voltage
Lmain IE
RH Reddy SEC
P2 S2
Rcl.
4 wire connection:
Input "Sec" of the CT Analyzer measures the terminal voltage of the test
object. The measurement is not influenced by the voltage drops caused by
the contact resistance of the clamps. The measurement results delivered
by the CT Analyzer are correct!
29
CT Analyzer
Figure 3-5:
Connection diagram for Voltage drop caused by contact
2 wire connection resistance of the clamp
technique Ip RCT OUTPUT
S1
Rcl.
P1
UC IL
IC ICT ~
Np Ns
Terminal voltage
Voltage behind
Lmain IE
the clamp
RH Reddy
SEC
P2 S2
Rcl.
30
Running an Automatic Parameter Search and CT Test (Guesser Function)
31
CT Analyzer
Start the test by pressing the I / 0 key. The red LED flashes to indicate that the
test is running.
32
Running an Automatic Parameter Search and CT Test (Guesser Function)
33
CT Analyzer
Now you can enter the "Location" and "Object" details and save the test.
The CT Analyzer also performed an automatic assessment of the measured
results using the guessed parameters. This assessment can be viewed on the
Assessment card (figure 4-6). For a detailed description of this card, please
refer to section 5.9 on page 96.
Figure 4-6:
Assessment card after
the test is finished
34
User Interface and Operation
35
CT Analyzer
36
User Interface and Operation
Editing a card
To open the edit mode for a displayed card, press the C U R S O R D O W N key on
the keyboard. The card’s tab is then no longer highlighted (see figure 5-3).
Use the cursor keys to move the cursor and to select the desired edit field. Some
edit fields have soft keys assigned. The soft keys available for an edit field are
displayed if the field is selected.
Figure 5-3:
Edit field selected in CT-
Object card
Display field
(cannot be
selected)
Edit field
Edit field,
selected
37
CT Analyzer
Figure 5-5:
Help system showing
explanatory help text for
a parameter
When the help system is displayed, you can use the soft keys N E X T P A G E and
P R E V . P A G E to scroll through the available help pages.
Pressing the A B O U T soft key opens a page with information about the
CT Analyzer hardware, the installed firmware, the serial number, etc.
38
User Interface and Operation
Figure 5-6:
Help system showing
the wiring diagram for
the CT test
39
CT Analyzer
The following table shows the default settings after initiating a new CT test.
Note: For more detailed information about the parameters and fields in
the CT-Object card, please refer to section 5.4.3 on page 58.
40
User Interface and Operation
Table 5-2:
Default settings for a Parameter Description Default value
new CT test
Location Information about the location of the CT not changed
and (e.g. company, country, station, etc.) and
the CT (e.g. manufacturer, type, serial no.,
Object etc.).
The content of these fields is defined in the
Location settings page and the Object
settings page, respectively (opened by
pressing the D E T A I L S soft key displayed
when the cursor is positioned in these
fields).
For more detailed information, please refer
to section 5.4.2 on page 57.
I-pn Rated primary current. "?"1
I-sn Rated secondary current. "?"1
Standard Standard according to which the test has to As defined in
be performed. the default test
settings
P/M Protection or measurement CT. "?"1
Class Class definition for the CT. "?"1
FS Instrument security factor. "?"1
Only applicable for:
IEC 60044-1, measurement CTs.
ext Rated continuous thermal current Ith. 120%
Only applicable for:
IEC 60044-1, measurement CTs.
RF Thermal current rating factor. 2
Only applicable for:
ANSI 30/45 measurement CTs.
Vb Secondary terminal voltage. "?"1
Only applicable for:
ANSI 30/45 protection CTs.
ALF Accuracy limiting factor. "?"1
Only applicable for:
IEC 60044-1 classes P and PR.
41
CT Analyzer
42
User Interface and Operation
43
CT Analyzer
Available functions
Load Report Loading an existing report or settings file from the
Compact Flash card.
Note: You can also load a file by pressing the L O A D soft
key in the CT-Object card.
Save As Saving the current file to the Compact Flash card.
Saving a file is described in detail on page 47.
Rename File Renaming a selected file on the Compact Flash card.
Only files can be renamed. Renaming of folders is not
possible with the CT Analyzer.
44
User Interface and Operation
Delete File Deleting a selected file or folder from the Compact Flash
card.
Only empty folders can be deleted. If you try to delete a
folder still containing files or subfolders, an error
message is displayed.
To select several neighboring files at a time, keep the
S H I F T key on the keyboard pressed while selecting the
files or folders using the C U R S O R U P / D O W N keys.
Copy / Cut / Paste Copying or cutting a file on the Compact Flash card and
File(s) pasting it at another location on the Compact Flash card.
Copying/cutting and pasting a file is described in detail on
page 47.
Format CF card Formatting the Compact Flash card. All data on the
Compact Flash card will be lost!
Note: In the CT Analyzer file system, the root directory can contain a
maximum of 240 files and the maximum file name length is 240
characters.
45
CT Analyzer
46
User Interface and Operation
Saving a file
1. To save a file, open the main menu and select the list item "File Handling". In
the appearing File Menu, select the item "Save As".
Note: You can also save a file by pressing the S A V E or the S A V E A S
soft key in the CT-Object card.
2. Navigate to the desired folder in the file system where you want to save the
file:
• If you like to save the file with the same name press the S A V E soft key.
A warning window will pop up asking whether the existing file should be
overwritten or not.
• If you want to use a similar file name as an existing one in the selected
folder, select this file using the cursor keys and press S A V E A S or
E N T E R . A new *.xml file entry with the selected file name is added. Edit
this file name and then press again S A V E , S A V E A S or E N T E R .
4. The file system card is closed and the File Menu is displayed. Select the list
item "Paste File(s)".
5. The file system card appears again. Navigate to the desired location in the
file system where you want to paste the file.
6. Press the P A S T E soft key to paste the file.
In case of a cut/paste action, the file is not deleted from its old location until
it has been pasted at its new location.
47
CT Analyzer
5.3.3 Settings
To change the device settings or the test settings for a new CT test, select the
list item "Settings" from the main menu.
Figure 5-12:
Main menu, "Settings"
Use the C U R S O R U P / D O W N
keys to scroll through the list.
48
User Interface and Operation
49
CT Analyzer
Assessment card
Resistance card
CT-Object card1
Excitation card
Comment card
Burden card
Ratio card
Burden card x
Resistance card x
Excitation card x x
Ratio card x x x
Assessment card x
Comment card x
1. The CT-Object card is always required to perform a CT test.
To save your selection and return to the Setting Menu press the B A C K soft key.
To return to the Setting Menu without saving the selection, press the E S C key.
50
User Interface and Operation
Miscellaneous Settings
Figure 5-14:
Miscellaneous Settings
51
CT Analyzer
E.g. a class multiplier of 0.5 means that the maximum accepted tolerance for
the ratio error is only half the standard tolerance (all ratio tolerance values are
multiplied with the entered factor).
Possible values: 0.25 to 1.00
Default value: 1.00
The R E S E T V A L U E S soft key resets all values to their default values.
5.3.4 Tools
Using the item "Tools" in the main menu you can access the update functions of
the CT Analyzer.
Figure 5-15:
Main menu, "Tools"
Note: Instead of using the CT Analyzer functions "Update Text" and "Update
Firmware" described below you can alternatively use the Firmware
Update PC tool contained in the CT Analyzer PC Toolset. For detailed
information, please refer to section 6.6 on page 131.
Available functions
Update Text Using this item you can install a new user interface
language file.
The file system page opened after selecting the "Update
Text" function only displays files named CTUser*.bin.
Please refer to the following subsection.
Update Firmware Using this item you can update the firmware of your
CT Analyzer.
The file system page opened after selecting the "Update
Firmware" function only displays files named
CTAnalyzer*.bin.
Please refer to the following subsection.
52
User Interface and Operation
Update Text
Figure 5-16:
Update text
After selecting "Update Text" you can install a new user interface language file.
This language is then available for selection in the "Language" page.
To install a new language, select the corresponding file CTUser_xxx.bin in
the directory A:\OMICRON\ on the Compact Flash card using the C U R S O R
U P / D O W N keys and press the S E L E C T soft key.
53
CT Analyzer
English is contained in the firmware and does not require a user language file.
Attention: Only install language files that are included in the package of the
installed firmware version.
If you install a language file that does not belong to the same
package, the user interface may become unreadable.
Note: Refer to the OMICRON website for available languages or ask your
distributor for a special language file.
If the text update process is interrupted or fails, the device displays an error
message and automatically resets the user interface to English when it is
switched on the next time.
Update Firmware
Figure 5-17:
Update firmware
54
User Interface and Operation
Attention: It is possible to install older firmware. In this case, the user interface
text is deleted and the device automatically changes to the English
user interface. After you have downgraded the firmware, you also
have to install the user interface language file of this (older)
firmware package.
After an installation of new firmware, the user interface language
automatically changes to English, if the installed user interface text
is no longer compatible. If you are using a user interface language
other than English, install the corresponding new user interface
language file delivered with the new firmware.
Press A B O R T to return to the Tools Menu without updating the firmware.
55
CT Analyzer
56
User Interface and Operation
Opens the Select Cards page where you can select the test cards for
the current test. The Select Cards page is described on page 50.
Attention: When loading a new CT test from the main menu (using
the item "New CT-Test"), the test card selection defined in the default
test settings is used for this new test. The card selection performed in
the CT-Object card will then be discarded.
The text fields "Location" and "Object" are only used for reporting and
documentation purposes. They can be filled by the user after the test is finished.
These fields provide information about the location of the CT and the CT itself.
The content of these fields is defined in the Location settings page and the
Object settings page, respectively.
To open these settings cards, position the cursor on the "Location" or the
"Object" field and then press the D E T A I L S soft key or the E N T E R key.
Table 5-4:
Information in the CT- Card Description
Object card to be filled
by the user Location The Location settings page provides the following text fields.
settings Each field can contain a maximum of 40 numbers or letters.
card
• Company, Country, Station, Feeder:
Company, country, station and feeder where the CT is
installed.
• Phase:
Phase to which the CT is connected.
• IEC-ID:
IEC-ID number of the CT or freely definable information.
57
CT Analyzer
Card Description
Object The Object settings page provides the following text fields.
settings Each field can contain a maximum of 40 numbers or letters.
card
• Manufact.: Manufacturer of the CT under test.
• Type: Type number or description of the CT under test.
• Serial No.: Serial number of the CT under test.
• Core: Number of the tested core.
• Tap: Description of the tap (e.g. 1S1-1S3).
• Optional 1: Optional field for free use.
58
User Interface and Operation
Table 5-5:
Common fields for all Parameter Description
standards, used and/or
determined by the test I-pn Rated primary current.
Possible values: 1 to 99000A (keyboard) or ? soft key.
Resolution: 0.1A
Default value: "?".
If the question mark is entered and a new test is started, the
CT Analyzer automatically tries to guess the nominal primary
current. See the flow chart in section 10.18 on page 198.
I-sn Rated secondary current.
Possible values: 0.0001 to 10A (keyboard) or soft keys 1 . 0 ,
2 . 0 , 5 . 0 , 1 / S Q R T ( 3 ) , 2 / S Q R T ( 3 ) , 5 / S Q R T ( 3 ) or ? .
Resolution: 0.0001A
Default value: "?".
If the question mark is entered and a new test is started, the
CT Analyzer automatically tries to guess the nominal secondary
current. See the flow chart in section 10.18 on page 198.
Standard Standard according to which the test has to be performed.
Possible values: Soft keys 6 0 0 4 4 - 1 , 6 0 0 4 4 - 6 , A N S I 3 0 or
ANSI45.
Default value: Standard defined in the default test settings.
The standard IEEE C57.13 includes two variants which differ in
the definition of the knee point. Therefore either ANSI 30 or ANSI
45 can be selected. For current transformers with non-gapped
cores, the knee is defined as the point where the tangent is at 45°
(ANSI 45) to the abscissa. For current transformers with gapped
cores, the knee is defined as the point where the tangent is at 30°
(ANSI 30) to the abscissa.
Note: Using the ANSI 45 standard for transformers with gapped
cores may possibly deliver incorrect results.
Note: The parameter definition syntax for the different standards
is described in chapter 11 on page 207.
P/M Definition of CT type: Protection or measurement CT.
Possible values: Soft keys P R O T . C T, M E A S . C T or ? .
Default value: "?".
If the question mark is entered and a new test is started, the
CT Analyzer automatically tries to guess the CT type to be used.
See the flow chart in section 10.18 on page 198.
59
CT Analyzer
Parameter Description
Class Accuracy class of the CT.
Possible values: Depending on the selected standard or ? .
The class can be selected by soft keys or is determined during
the test by the guesser function. The class guesser function only
works for the standards IEC 60044-1 and IEEE C57.13
(ANSI 30, ANSI 45). If the "?" is entered, the device selects the
class that first meets the class criteria for the selected standard
and type of CT (protection/measurement CT).
For a detailed description of the parameter definition syntax,
please refer to chapter 11 on page 207.
Definition of the "Class" parameter by the user is only possible, if
the parameters "Standard" and "P/M" have been defined before.
Depending on these parameters, the parameters for the CT class
can be selected using the soft keys.
If the question mark has been entered for the parameter "P/M",
the "Class" cannot be defined by the user but is automatically
determined (guessed) by the CT Analyzer during the test
instead.
VA Nominal burden of the CT, used to calculate the behavior of the
Cosϕ CT at the nominal burden.
Possible values: 0 to 300VA (keyboard) or soft keys 1 . 0 A to
3 0 A or ? to determine the burden using the guesser function
(refer to section 10.19 ”Guesser Algorithm for Burden” on
page 204).
Depending on the guessed burden and the selected test
standard, the corresponding cos ϕ is used (cos ϕ not editable by
the user):
60044-1
60044-1
60044-6
C57.13
C57.13
(ANSI)
(ANSI)
[VA]
PX
M
M
P
continued...
60
User Interface and Operation
Parameter Description
(cont.) If an ANSI standard is selected (ANSI 30, ANSI 45), the
CT Analyzer offers some soft keys (B - 1 , B - 2 , B - 4 , B - 8 ) with
standard ANSI loads for the power. If these soft keys are used,
the power is calculated according to table 9 in IEEE C57.13. If
the rated current is not 5A, the resistance and inductance of the
table is multiplied by the following factor:
5 2
α = ⎛⎝ -------------⎞⎠
I rated
Burden "Burden" and "Cos ϕ" are used to define the operating burden
Cos ϕ connected to the CT. These parameters are used to calculate the
behavior of the CT at the operating burden (connected burden)
and the corresponding cos ϕ.
Possible values for "Burden":
0 to 300VA (keyboard) or soft keys 1 . 0 A to 3 0 A or ? .
Possible values for "Cos ϕ": 0 to 1 (keyboard).
If the "?" was entered or no value was entered, the device tries to
guess the operating burden (refer to section 10.19 ”Guesser
Algorithm for Burden” on page 204).
If the test contains a Burden card, a question mark is
automatically entered to the fields "Burden" and "cos ϕ" and
entering the burden is not possible until the test is finished. In this
case, the value determined during the burden test is automatically
entered after burden measurement. If "I-sn" is defined, the burden
is updated immediately after the burden test is finished. If "I-sn" is
not defined (entry "?"), the burden is updated after the resistance
test.
These parameters can also be changed after the test or in a
loaded test report in order to check the CT behavior at different
burden values.
f Rated frequency of the CT.
Possible values: Integer value between 16 and 400 Hz
(keyboard) or soft keys 1 6 . 7 H Z , 5 0 H Z , 6 0 H Z or 4 0 0 H Z .
Default value: Frequency defined in the default test settings.
Rct Specified winding resistance.
Possible values: 0 to 300 ohms (keyboard) or soft key ? .
Default value: "?".
61
CT Analyzer
Parameter Description
Class mult. Class multiplying factor.
factor
This factor increases the assessment level for the ratio test. E.g.
a class multiplier of 0.5 means that the maximum accepted
tolerance for the ratio error is only half the standard tolerance (all
ratio tolerance values are multiplied with the entered factor).
Possible values: 0.25 to 1.00 (keyboard) or soft key 1 . 0 .
Default value: 1.0.
Delta com- Correction factor for the ratio measurement. This factor enables
pensation the ratio measurement for CTs that are installed inside a delta
winding transformer.
Possible values:
Soft keys R A T I O 1 , R A T I O 2 / 3 or R A T I O 1 / 3 .
Default value: "Ratio 1".
Choose "Ratio 1", if no correction is required.
Choose "Ratio 2/3", if input PRIM is connected to the two
terminals of that transformer winding, the CT is in series.
Choose "Ratio 1/3" if input PRIM is connected to the terminals of
a transformer winding, the CT is not in series with.
62
User Interface and Operation
63
CT Analyzer
Parameter Description
E1 User-defined e.m.f. to verify the excitation current at this specific
e.m.f.
Possible values: 0.1 to 20000V (keyboard) or soft key ? .
Default value: "?".
If the question mark is entered, half the voltage entered or
measured for Ek is used.
Ie1 Maximum allowed excitation current at E1.
Possible values: 0.03mA to 30000mA (keyboard) or soft key ? .
Default value: "?".
If the question mark is entered, the CT Analyzer uses the
excitation current measured at the voltage value defined at E1.
In this case, the assessment for this parameter is OK.
64
User Interface and Operation
Tp = K – 1-
Tp depends on K as follows: ------------
ω
K Dimensioning factor.
Possible values: 1 to 1572 (keyboard).
K depends on Tp as described above for parameter Tp.
V-al Rated equivalent excitation limiting secondary voltage.
Possible values: 0 to 9999V (keyboard) or soft key ? .
Default value: "?".
I-al Accuracy limiting secondary excitation current.
Possible values: 0.03mA to 30A (keyboard) or soft key ? .
Default value: "?".
E1 User-defined e.m.f. to verify the excitation current at this specific
e.m.f.
Possible values: 0.1 to 20000V (keyboard) or soft key ? .
If the question mark is entered, half the voltage entered or
measured for Ek is used.
65
CT Analyzer
Parameter Description
Ie1 Maximum allowed excitation current at E1 (user-defined e.m.f.).
Possible values: 0.03mA to 30A (keyboard) or soft key ? .
If the question mark is entered, the CT Analyzer uses the
excitation current measured at the voltage value defined at E1.
In this case, the assessment for this parameter is OK.
66
User Interface and Operation
Parameter Description
t1 Duration of first current flow. The specified accuracy limit must
not be reached within time "t-al1".
Possible values: 0.000 to 5.000s (keyboard) or soft key
100MS.
Default value: 0.1s.
t2 Duration of second current flow. The specified accuracy limit
must not be reached within time "t-al2".
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.000 to 5.000s (keyboard) or soft key
100MS.
Default value: 0.1s.
t-al1 Permissible time to accuracy limit for first energizing period of
the duty cycle.
Possible values: 0.000 to 5.000s and max. t1 (keyboard) or soft
key 4 0 M S .
Default value: 0.04s.
t-al2 Permissible time to accuracy limit for second energizing period
of the duty cycle.
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.000 to 5.000s and max. t2 (keyboard) or soft
key 4 0 M S .
Default value: 0.04s.
tfr Dead time between first opening and reclosure.
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.00 to 5.00s (keyboard) or soft key 3 0 0 M S .
Default value: 0.3s.
67
CT Analyzer
68
User Interface and Operation
Parameter Description
t2 Duration of second current flow. The specified accuracy limit
must not be reached within time "t-al2".
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.000 to 5.000s (keyboard) or soft key
100MS.
Default value: 0.1s.
t-al1 Permissible time to accuracy limit for first energizing period of
the duty cycle.
Possible values: 0.000 to 5.000s and max. t1 (keyboard) or soft
key 4 0 M S .
Default value: 0.04s.
t-al2 Permissible time to accuracy limit for second energizing period
of the duty cycle.
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.000 to 5.000s and max. t2 (keyboard) or soft
key 4 0 M S .
Default value: 0.04s.
tfr Dead time between first opening and reclosure.
This parameter is only displayed if "Duty" is C-O-C-O.
Possible values: 0.00 to 5.00s (keyboard) or soft key 3 0 0 M S .
Default value: 0.3s.
69
CT Analyzer
70
User Interface and Operation
71
CT Analyzer
If the CT Analyzer cannot reach the desired test current I-test, an overload
message is displayed in the right-hand corner of the status line (see figure 5-21).
72
User Interface and Operation
If the rated secondary current is not known, the result field will
only show "n/a" as long as I-sn is not defined.
Z Calculated as follows:
U rms
Z = -----------
-
I rms
2. In the Burden card, use the default test current or enter the desired test
current to parameter "I-test" and start the burden test by pressing the I / 0
key.
3. A message appears prior to the test notifying you that you should check the
wiring before running the test (figure 5-23).
73
CT Analyzer
Figure 5-23:
Message displayed
prior to burden test
Check the wiring (refer to figure 5-22) and then press the S T A R T T E S T soft
key to start the burden test. Or press C A N C E L T E S T to abort the burden
test.
If you do not want to execute the burden test, press S K I P T E S T . This will
skip the burden test and immediately start the CT test without a previous
burden test. This allows you to repeat the CT test without repeating the
burden test.
74
User Interface and Operation
75
CT Analyzer
Parameter Description
T-ref Reference temperature, i.e temperature the CT is specified for.
Possible values: -40 to 150°C or -40 to 302°F (keyboard).
Default value: 75°C.
This parameter can also be changed after the test in order to
recalculate the test results accordingly.
The winding resistance at reference temperature is calculated
from the winding resistance measured at ambient temperature
(T-meas) and the specified reference temperature.
V DC 234.5 + T ref
R ref = ---------- × ----------------------------------
I DC 234.5 + T meas
76
User Interface and Operation
If the CT Analyzer displays an overload in the Excitation card, either the desired
knee point could not be reached or not all necessary measurement points could
be taken (e.g. in the knee point area, not enough points for a proper calculation
of the knee point could be measured).
77
CT Analyzer
Opens the K-value error graph page (refer to section 5.7.5 on page
88).
This graph is part of Chinese standards and shows the maximum
possible primary current (K * Ipn) that can flow over a specific burden
without exceeding the accuracy limit.
5.7.2 Settings
The upper part of the Excitation card shows the test settings adjusted in the CT-
Object card.
78
User Interface and Operation
Table 5-20:
Excitation card, test
results displayed for
Note: Depending on the selected burden, the Excitation card shows the
IEC 60044-1 results either calculated with the nominal burden (R E S U L T S W I T H
N O M . B U R D E N ) or calculated with the operating burden (R E S U L T S
W I T H O P . B U R D E N ).
The displayed pages are identical to a large extend, except the field
label for the burden parameter ("VA" in case of nominal burden and
"Burden" in case of operating burden) and the result values.
Para- Description Results displayed for
meter IEC 60044-1
meas. CT prot. CT
V-kn Knee point voltage (acc. to standard)
of the uppermost knee point found
V-kn1 x x
(definition of knee point see section
10.15 on page 192).
I-kn Knee point current (acc. to standard)
of the uppermost knee point found. x x
I-kn1
V-kn2 Knee point voltage (acc. to standard)
of the lowest knee point found
x x
(definition of knee point see section
10.15 on page 192).
I-kn2 Knee point current (acc. to standard)
x x
of the lowest knee point found.
ALF Accuracy limiting factor according to
IEC 60044-1 direct measurement
method, calculated for nominal and
operating burden (calculation see
section 10.6.1 on page 184). class P &
class PR
If the CT Analyzer is not able to only
measure up to the actual accuracy
limiting factor, the prefix ">" is
displayed to indicate that the ALF is
larger than the displayed value.
79
CT Analyzer
80
User Interface and Operation
Table 5-21:
Excitation card, test
results displayed for Note: Depending on the selected burden, the Excitation card shows the
IEC 60044-6 results either calculated with the nominal burden (R E S U L T S W I T H
N O M . B U R D E N ) or calculated with the operating burden (R E S U L T S
W I T H O P . B U R D E N ).
The displayed pages are identical to a large extend, except the field
label for the burden parameter ("VA" in case of nominal burden and
"Burden" in case of operating burden) and the result values.
Para- Description Results displayed for
meter IEC 60044-6, class
TPS TPX / TPZ
TPY
V-kn Knee point voltage according to standard
(definition of knee point see section 10.15 x x x
on page 192).
I-kn Knee point current according to standard. x x x
81
CT Analyzer
82
User Interface and Operation
x
This parameter allows the determination
of the working point on the excitation
curve that would be reached with the
entered settings.
ε^ Peak instantaneous error at voltage Emax
(calculation see section 10.7 on page x
187).
Table 5-22:
Excitation card, test
results displayed for
Note: Depending on the selected burden, the Excitation card shows the
ANSI (IEEE C57.13) results either calculated with the nominal burden (R E S U L T S W I T H
N O M . B U R D E N ) or calculated with the operating burden (R E S U L T S
W I T H O P . B U R D E N ).
The displayed pages are identical to a large extend, except the field
label for the burden parameter ("VA" in case of nominal burden and
"Burden" in case of operating burden) and the result values.
Para- Description Results displayed for
meter ANSI (IEEE C57.13)
meas. CT prot. CT
V-kn Knee point voltage according to standard
(definition of knee point see section 10.15 x x
on page 192).
I-kn Knee point current according to standard. x x
83
CT Analyzer
84
User Interface and Operation
Table 5-23:
Definition of the axes in Definition of axes for different standards
the excitation graph for
different standards Standard Vertical axis Horizontal axis
IEC 60044-1 rms terminal voltage rms excitation current
IEC 60044-6 rms emf voltage1 peak excitation current
IEEE C57.13 (ANSI) rms emf voltage1 rms excitation current
1. Calculation of the rms emf voltage see section 10.4 on page 178.
Switches off the display of values on the bottom right of the diagram.
If you have switched off the values, this soft key changes to T E X T O N
to switch the values display on again.
85
CT Analyzer
Closes the excitation graph and brings you back to the Excitation
card.
Moves the cursor to the knee point on the graph according to the
selected standard.
If two or more knee points could be found on the graph, this soft key
is alternately labelled K N E E P O I N T 1 or K N E E P O I N T 2 ,
depending on which knee point the cursor is actually positioned to.
After opening the excitation graph, the cursor shows the uppermost
knee point (knee point 1) and the soft key is labelled K N E E P O I N T 2 :
– Pressing K N E E P O I N T 2 moves the cursor to the lowest knee
point on the graph. The labelling changes to K N E E P O I N T 1 .
– Pressing K N E E P O I N T 1 moves the cursor to the uppermost
knee point on the graph. The labelling changes to
KNEE POINT 2.
Moves the cursor to the knee point on the reference graph. This soft
key is only available if a reference graph is loaded.
If the reference graph loaded has two or more knee points, this soft
key is alternately labelled R E F . K N E E P O I N T 1 or
R E F . K N E E P O I N T 2 , depending on which knee point the cursor is
actually positioned to. After loading the reference graph, the cursor
shows the uppermost knee point (knee point 1) and the soft key is
labelled R E F . K N E E P O I N T 2 :
– Pressing R E F . K N E E P O I N T 2 moves the cursor to the lowest
knee point on the reference graph. The labelling changes to
REF. K NEE P OINT 1 .
– Pressing R E F . K N E E P O I N T 1 moves the cursor to the
uppermost knee point on the reference graph. The labelling
changes to R E F . K N E E P O I N T 2 .
Opens the file system card to select a previous test in order to load the
excitation curve of this test as a reference curve and compare it with
the current one.
The reference curve is displayed as a dotted line in addition to the
excitation curve of the actual test. If a reference curve is loaded, the
values "V-ref", "I-ref" and "L-ref" are displayed in addition to the
measured values.
86
User Interface and Operation
Switches off the reference curve from the display. If you have switched
off the reference curve, the soft key changes to R E F . O N to switch
the reference curve on again.
This soft key is only available if a reference curve has been loaded.
By pressing one of these soft keys you can display the measured
excitation graph and knee point as defined in the respective standard.
Note: The test report only contains the graph for the standard selected
in the CT-Object card.
87
CT Analyzer
Note: The function "K-Value Error Graph" can be switched on or off in the
device settings (M A I N M E N U , entry "Settings" -> S E T T I N G M E N U ,
entry "K-Value Error Graph"). If switched off, the K-value error graph is
not included in the test report.
I ext
-----------------------
- ⋅ 100 = 10 or 5
I sn ⋅ ALFi
Switches off the display of values on the top right of the diagram. If you
have switched off the values, this soft key changes to T E X T O N to
switch the values display on again.
88
User Interface and Operation
Closes the K-value error graph and brings you back to the Excitation
card.
Moves the cursor to that point on the error graph that corresponds to
the nominal burden value defined in the "VA" field of the CT-Object
card (value entered by the user or determined by the guesser function
of the CT Analyzer).
Moves the cursor to that point on the error graph that corresponds to
the operating burden value defined in the "Burden" field of the CT-
Object card (value entered by the user or determined by the guesser
function of the CT Analyzer).
89
CT Analyzer
90
User Interface and Operation
5.8.2 Settings
The following settings can be done in the Ratio test card.
Table 5-24:
Settings in the Ratio Parameter Description
card
I-p Primary current for calculation of the ratio error and phase
displacement with the burden (operating burden) defined in the
CT-Object card.
After the test is finished, it is possible to change the value for the
primary current. The ratio error and/or the phase error are then
recalculated and displayed again. When storing the test results,
the currently displayed measurement results are stored.
Changing this value only influences the results displayed in the
Ratio card (operating burden-related values). It does not affect
the values displayed in the separate pages for the ratio and
phase tables (values related to nominal burden).
Default value: I-pn
91
CT Analyzer
Parameter Description
εt Turns ratio error acc. to IEC 60044-6 class TPS or IEC 60044-1
class PX (calculation see section 10.2 on page 177).
Figure 5-31:
Ratio card, phase table
page
92
User Interface and Operation
Only four values are used for the table on the Ratio card: the value defined on
the CT-Object card and the lowest three values of the standard table "around"
the value on the CT-Object card.
In this example (value defined is 25VA), the values 45.0VA and 2.5VA are
ignored. The following values are used:
• 25.0 VA (defined on CT-Object card)
• 22.5 VA
• 12.5 VA
• 5.0 VA
93
CT Analyzer
If the VA parameter value entered on the CT-Object card is 15, the following
values are used (values 45.0VA and 22.5VA are ignored):
• 15.0 VA
• 12.5 VA
• 5.0 VA
• 2.5 VA
If the VA parameter value entered on the CT-Object card is 60, the following
values are used (values 5.0VA and 2.5VA are ignored):
• 60.0 VA
• 45.0 VA
• 22.5 VA
• 12.5 VA
If the VA parameter value entered on the CT-Object card is B 0.5 (corresponds
to 12.5 VA), the following values are used (value 45.0VA is ignored):
• 22.5 VA
• 12.5 VA
• 5.0 VA
• 2.5 VA
94
User Interface and Operation
Only four values are used for the table on the Ratio card: the value defined on
the CT-Object card and the lowest three values of the standard table "around"
the value on the CT-Object card. If the value defined on the CT-Object card is
a standard value, the four values of the standard table are used.
In this example (value defined is 150VA), the value 200.0VA is ignored. The
following values are used:
• 150.0 VA
• 100.0 VA
• 50.0 VA
• 25.0 VA
If the VA parameter value entered on the CT-Object card is 300.0VA, the
following values are used (value 25.0VA is ignored):
• 300.0 VA
• 200.0 VA
• 100.0 VA
• 50.0 VA
If the VA parameter value entered on the CT-Object card is 200.0VA, the
following values are used (no value ignored since 200.0VA is a standard value):
• 200.0 VA
• 100.0 VA
• 50.0 VA
• 25.0 VA
If the VA parameter value entered on the CT-Object card is 10.0VA, the
following values are used (value 200.0VA is ignored):
• 100.0 VA
• 50.0 VA
• 25.0 VA
• 10.0 VA
95
CT Analyzer
96
User Interface and Operation
meas. CT
class PR
class PX
prot. CT,
prot. CT,
prot. CT,
class P
Class Accuracy class according to standard. x x x x
ε Current ratio error. x x x
Δϕ Phase deviation. x x x
εt Turns ratio error (included in class). x
εc Composite error. x x
FS Instrument security factor (direct
measurement method, calculation see x
section 10.5.1 on page 179).
FSi Instrument security factor (indirect
measurement method, calculation see x
section 10.5.2 on page 181).
ALF Accuracy limiting factor (direct
measurement method, calculation see x x
section 10.6.1 on page 184).
ALFi Accuracy limiting factor (indirect
measurement method, calculation see x x
section 10.6.2 on page 186).
Ek Rated knee point emf. x
Ie Accuracy limiting secondary excitation
x
current.
Ie1 Max. allowed secondary excitation current
x
at E1.
Kx Dimensioning factor (according to
x
IEC 60044-1 class PX).
Rct Secondary winding resistance. x x x x
Ts Secondary time constant (calculation
x x x x
method see section 10.10 on page 188).
Kr Remanence factor (calculation see
x x x x
section 10.12 on page 190).
Burden Manual assessment of the user for the
burden test (no automatic assessment x x x x
performed).
97
CT Analyzer
TPS
TPX
TPY
TPZ
Class Accuracy class according to standard. x x x x
εt Turns ratio error (included in class). x
ε Current ratio error. x x x
Δϕ Phase deviation. x x x
V-al Rated equivalent excitation limiting
x
secondary voltage.
I-al Accuracy limiting secondary excitation
x
current.
Ie1 Max. allowed secondary excitation current
x
at E1.
K*Kssc Dimensioning factor (K) multiplied by the
rated symmetrical short-circuit current x
factor (Kssc).
ε^ Peak instantaneous error at voltage Emax
(calculation see section 10.7 on page x x
187).
Ktd*Kssc Transient dimensioning factor (Ktd)
multiplied by the rated symmetrical short- x x
circuit current factor (Kssc).
Rct DC winding resistance. x x x x
Ts Secondary time constant (calculation
x x x x
method see section 10.10 on page 188).
Kr Remanence factor (calculation see
x x x
section 10.12 on page 190).
Burden Manual assessment of the user for the
burden test (no automatic assessment x x x x
performed).
98
User Interface and Operation
99
CT Analyzer
100
User Interface and Operation
0.6 Class:
at % of rated current
10% 100% 100%*RF
Ratio error (ε) [%] ≤ 1.2 0.6
Phase error (Δϕ) [min] 2600(RCF-1.012) 2600(RCF-1.006)
< Δϕ < < Δϕ <
2600(RCF-0.988) 2600(RCF-0.994)
RCF of the measurement can be calculated from the ratio table:
RCF = 1/((Ratio_Error /100) + 1)
1.2 Class:
at % of rated current
10% 100% 100%*RF
Ratio error (ε) [%] ≤ 2.4 1.2
Phase error (Δϕ) [min] 2600(RCF-1.024) 2600(RCF-1.012)
< Δϕ < < Δϕ <
2600(RCF-0.976) 2600(RCF-0.988)
RCF of the measurement can be calculated from the ratio table:
RCF = 1/((Ratio_Error /100) + 1)
101
CT Analyzer
102
User Interface and Operation
103
CT Analyzer
104
User Interface and Operation
105
CT Analyzer
106
User Interface and Operation
107
CT Analyzer
108
CT Analyzer PC Tools
6 CT Analyzer PC Tools
The CT Analyzer PC Toolset contains a set of PC software that is necessary and
helpful for your work with the CT Analyzer on a PC.
The following tools are installed with the CT Analyzer PC Toolset:
• CTA Start Page
• CTA QuickTest
• CTA Remote Excel File Loader
• Update for CPC Explorer (if installed)
• CTA Remote Control
• Sample software for CTA Remote Control
• CTA remote test sample for Visual Basic (VBA)
• CTA remote test sample for C++
• CTA to Netsim Export
109
CT Analyzer
110
CT Analyzer PC Tools
Figure 6-1:
CTA Start Page
111
CT Analyzer
CTA QuickTest
Click this entry to start the software tool CTA QuickTest. Using the CTA
QuickTest software, the CT Analyzer can be used as a versatile multimeter with
included power source.
Refer to section 6.4 on page 116 for a detailed description of this tool.
CPC Explorer
This entry is only active, if the CPC Explorer software is installed on the PC. If
active, click this entry to start the software tool.
The CPC Explorer software is not included in the CT Analyzer PC Toolset. It has
to be installed separately from the CD ROM "CPC Explorer Software"
accompanying the CT Analyzer. Refer to section 6.9 on page 138 for a detailed
description of this tool.
112
CT Analyzer PC Tools
6.3.3 Setup
Firmware Update
Click this entry to start the CT Analyzer Firmware Update tool. This tool
establishes communication with a connected CT Analyzer and enables you to
read e.g. the serial number and the current version of the firmware and/or to
perform a firmware update or install a new user interface language file.
Refer to section 6.6 on page 131 for a detailed description of this tool.
Settings
Clicking this entry displays the Settings dialog box where you can define the
following:
"General" tab
In the General tab you can select the user interface language for the software
components of the CT Analyzer PC Toolset and the CTA Start Page.
"Remote EFL" tab
Using the "Use custom template" checkbox you can select whether the CTA
Remote Excel File Loader tool should use a custom template file or an
OMICRON template file by default. Depending on your selection, specify the
path and name of the default template and/or the custom template in the
corresponding text field.
Note: The OMICRON template file CT Remote EFL.xlt is by default
stored in a subfolder in the installation path of the CT Analyzer PC
Toolset. If not changed during installation, the path is:
C:\Program Files\OMICRON\
CT Analyzer PC Toolset\RemoteEFL\CT Remote EFL.xlt
113
CT Analyzer
6.3.4 Support
OMICRON News
Clicking this entry opens the OMICRON News window. OMICRON News is an
information service that provides you with relevant messages such as updates
for your installed OMICRON products as well as OMICRON product information
and news.
This service bases on the RSS Dynamic Bookmark Technology. To use this
service, your computer needs access to the Internet. If your computer has no
Internet access, the OMICRON News window will not start.
The upper part of the OMCRON News window lists descriptive headlines of the
available updates and news. The lower part is a browser-like window that
displays the contents of the topic you select in the upper part.
"Show New Here you can select to only display update messages
Messages" and news that were published after your last visit at the
OMICRON news server. If selected, messages before
that date are hidden.
114
CT Analyzer PC Tools
115
CT Analyzer
116
CT Analyzer PC Tools
Output adjustment
Figure 6-3:
CTA QuickTest, Output
group box
Table 6-1:
User interface elements
in the Output group box UI element Description
Mode The generator can be configured as AC or DC source.
Select the corresponding radio button.
V/I The generator can work either in voltage or current
mode. Select the corresponding radio button.
Amplitude Mode "AC" and "V" (voltage source): 0 to 35V
Mode "DC" and "V" (voltage source): 0 to 120V
Mode "AC" and "I" (current source): 0 to 5A
Mode "DC" and "I" (current source): 0 to 10A
Attention: DC current mode is very dangerous!
It is not possible to interrupt the circuit using a standard
relay or a standard circuit breaker. Due to the arc,
contact clearances of up to 10mm are required to switch
off the current.
Frequency Output frequency of the generator.
Possible values: 10 to 400Hz.
Compensate SEC Measurement input SEC of the CT Analyzer has an input
impedance impedance of approx. 470kΩ and can thus influence the
current measurement.
Therefore, it is possible to compensate this error current
by clicking the checkbox "Compensate SEC
impedance". This checkbox is marked by default.
Note: The input impedance of input PRIM is even lower
but cannot be compensated.
ON / OFF button Click this button to start or stop the measurement.
117
CT Analyzer
Measurement Mode
Figure 6-4:
CTA QuickTest,
Measurement Mode
group box
Table 6-2:
User interface elements
in the Measurement UI element Description
Mode group box
RMS Performs a standard true rms measurement of the AC
signal component only. The DC part is removed from
the measured rms result.
Frequency selective Only the fundamental wave of the applied signals is
measured.
Frequency selective Allows high precision measurement of the
(with filter) fundamental wave of an applied signal. All frequencies
outside the output frequency plus/minus the cutoff
frequency are supressed by about 120dB.
Cutoff Frequency
Depending on the selected cutoff frequency, the
measurement time can differ considerably. The typical
time needed for a frequency selective measurement
depends on the cutoff frequency as follows:
Cutoff frequency Measurement time
0.2Hz approx. 1 minute
1.5Hz approx. 15 seconds
10Hz approx. 3 seconds
Integration Time The integration time can be selected in the range
between 0.1 and 1sec.
If the measurement mode "Frequency selective (with
filter)" is used, the software will automatically choose
an approprate integration time depending on the
selected cutoff frequency.
Regulation Only used for AC mode. Inactive in DC mode.
Parameters
If the "Inductive load" checkbox is checked, the output
generator always tries to keep the DC output current
at zero. For this purpose, a current regulator is
permanently running while an AC signal is generated.
continued ...
118
CT Analyzer PC Tools
UI element Description
(cont.) To guarantee proper operation of the current regulator,
the connected load (L and R) has to be defined under
"Regulation Parameters". If you don't know the actual
inductance and resistance of the load, use the default
values (R = 1 Ω, L = 50H). In this case, the regulator
will also regulate the DC current to zero, however it will
need longer to reach this.
Measurement area
Figure 6-5:
CTA QuickTest,
Measurement area
Table 6-3:
User interface elements
in the Measurement UI element Description
area
Column "SEC" Voltage input SEC.
Input voltage ranges (automatic range selection):
0 to 0.3VAC, 0 to 3VAC, 0 to 30VAC, 0 to 300VAC
Input impedance:
470kΩ to 1MΩ depending on the applied signal
Column "PRIM" Voltage input PRIM.
Input voltage ranges (automatic range selection):
0 to 0.03VAC, 0 to 0.3VAC, 0 to 30VAC
Input impedance:
130kΩ to 330kΩ depending on the applied signal
Column "Output (A)" Current measurement.
Input voltage ranges (automatic range selection):
0 to 150mA, 0 to 1.5A, 0 to 3A, 0 to 15A peak
The input impedance of input SEC can be
compensated. Measurement is done using a series
shunt in the generator output.
Row "Range" Displays the actually used voltage/current range.
119
CT Analyzer
UI element Description
Row "DC" DC amplitude of the applied signal.
Row "RMS" RMS amplitude of the applied signal after the selected
filter.
Row "Peak +" Measured positive peak value of the applied signal.
Row "Peak -" Measured negative peak value of the applied signal.
Row "Phase" Phase relative to input SEC (the phase of input SEC is
always 0).
Row "Frequency" Measured frequency.
Results
Figure 6-6:
CTA QuickTest,
Results group box
Table 6-4:
User interface elements
in the Results group UI element Description
box
N (Vsec/Vprim) N = VSEC/VPRIM
Z (Vsec/I) Z = VSEC/IOUTPUT
cos(phi) cos(IPHASE)
L or C X
Phase I = 0 - 360° --> L = -----L-
ω
1 -
Phase I = 0 - 180° --> C = --------------
ω ⋅ XC
XL or XC 2
Phase I = 0 - 360° --> XL = 1 – ( cos α I PHASE ) ⋅ Z
2
Phase I = 0 - 180° --> XC = 1 – ( cos α I PHASE ) ⋅ Z
R
R = cos α I PHASE ⋅ Z
120
CT Analyzer PC Tools
Save Results
The results can be saved to a *.csv file for later use for documentation purposes
or further calculations.
Figure 6-7:
CTA QuickTest, Save
Results group box
Table 6-5:
User interface elements
in the Save Results UI element Description
group box
Add Comment A comment can be entered for each test.
Decimal symbol Character used as decimal separator when storing
data to the *.csv file.
S A V E button Click to save the actual values to the currently open
file. If no file is currently open, the "Save As" dialog will
appear.
S A V E A S button Click to save the actual values to a new file. The "Save
As" dialog appears where you can specify the file.
Connection Status
Figure 6-8:
CTA QuickTest,
connection status
information
121
CT Analyzer
Menu options
Figure 6-9:
CTA QuickTest, menu
bar
Table 6-6:
Menu options available
in CTA QuickTest Menu Description
F I L E menu Using the options C O N N E C T and D I S C O N N E C T you
can connect/disconnect to the CT Analyzer (see also
button C O N N E C T / D I S C O N N E C T ).
Using the options S A V E R E S U L T S and S A V E
R E S U L T S A S you can save the actual values to the
currently open file or to a new file. If necessary, the
"Save As" dialog appears where you can specify the
file (see also buttons S A V E and S A V E A S ).
Using E X I T you can exit CTA QuickTest and return to
the CTA Start Page (see also button E X I T ).
? menu Click the menu option A B O U T C T A Q U I C K T E S T to
open a dialog showing detailed version information
about CTA QuickTest.
122
CT Analyzer PC Tools
Note: The data sheet of the CTA Remote Excel File Loader holds all data the
CT Analyzer can deliver. However, it is not intended to provide a
printable test report. In order to obtain a printable test report, you can
either add an additional worksheet and create your own printable test
report or use one of the pre-made templates delivered by OMICRON.
If you did not change the default installation path of the software, these
templates are located under:
C:\Program Files\OMICRON\CT Analyzer PC Toolset\
RemoteEFL\Templates
You can also access the Templates folder by clicking the Templates
entry in the Support section of the CTA Start Page.
123
CT Analyzer
Note: Upon starting, the CTA Remote Excel File Loader performs a
Microsoft Excel© version check. If the Excel© version installed on your
system is too old, a corresponding message is displayed, notifying you
that that you have to upgrade your Excel version in order to run the
CTA Remote Excel File Loader. Also refer to the system requirements
given in section 6.1 on page 109.
The Welcome page of the CTA Remote Excel File Loader provides the following
functionality:
• Click the N E W T E S T button to start your measurements with the
CT Analyzer using the defined test settings.
• Click the L O A D R E P O R T button to load previously saved CT Analyzer
report files (*.xml).
• Click the H E L P button to launch the CTA Remote Excel File Loader help.
Figure 6-10:
CTA Remote Excel File
Loader Welcome page
124
CT Analyzer PC Tools
125
CT Analyzer
The Test Settings dialog box holds all parameters contained in the
CT Analyzer’s CT-Object card. For a detailed description of the parameters,
please refer to section 5.4.3 on page 58.
To edit the content of the fields "Location" and "Object", click the corresponding
browse button to open the Location dialog or the Object dialog, respectively.
For a more detailed description of the "Location" and "Object" settings, please
refer to section 5.4.2 on page 57.
126
CT Analyzer PC Tools
Changing the test settings in this dialog is only possible after clicking the
C H A N G E S E T T I N G S button.
Table 6-7:
Buttons in the Test Button Description
Settings dialog box
EDIT REPORT Click to edit the location, object and comment fields of a
CT Analyzer report (= version 3.00). Changes to test
settings or test results are not allowed.
SAVE Click to save the report under a name of your choice.
REPORT AS
CHANGE Click to change the test settings. Note that this does not
SETTINGS affect the report area of the worksheet or the embedded
report data.
SAVE Click to save the settings to a settings file. This settings
SETTINGS AS file can later be used as test template by the
CT Analyzer or the CTA Remote Excel File Loader.
Each report file can also be used as a settings file.
Settings files are much smaller than report files.
OPTIONS Opens the Options dialog box. Please refer to ”Options
dialog” on page 128.
LOAD FILE Click to load any CT Analyzer test report or settings file
of your choice into the actual worksheet.
APPLY Click to apply the actual changes to the worksheet. If you
edited a report, the changes will be applied to both the
settings area and the report area of the worksheet.
Futhermore, the embedded report is changed.
Once you applied the actual changes, the C A N C E L
button changes into E X I T .
CANCEL / Click C A N C E L to discard all actual changes and close
EXIT the Test Settings dialog box.
Click E X I T to close the Test Settings dialog box.
START TEST / Click S T A R T T E S T to apply the changes and start
STOP TEST the execution of the test on the CT Analyzer. Once you
have started the test, this button changes to S T O P
TEST.
127
CT Analyzer
Options dialog
Figure 6-13:
CTA Remote Excel File
Loader, Options dialog
box
The Options dialog is displayed, if you click the O P T I O N S button in the Test
Settings dialog box.
Click O K to apply your changes and close the dialog or C A N C E L to close the
dialog without changing the options.
Table 6-8:
Buttons in the Options Button Description
dialog box
Autosave to Select this option to automatically save the result files on
Serialno_Tap your computer.
The resulting file names are a combination of the file
name and the tap description (e.g. 1S1-1S2), separated
by an underscore character. Characters that are not
allowed in filenames are automatically replaced by a
minus character.
If you already saved a file, the file path of the last saved
file is used. When selecting this option for the first time or
when opening the template for the first time, a "Save as"
dialog is displayed.
Example:
File name: 08/58572926
Tap: 1S1-1S2
The results are saved to the file 08-58572926_1S1-1S2
in the folder of the last saved file.
128
CT Analyzer PC Tools
Button Description
Disable Save Starting a new test usually launches a query whether the
Results Dialog existing worksheet data should be overwritten. Select
this option to suppress the query and to overwrite the
existing worksheet data without further notification.
USB port Determine the interface for the communication with the
COM port CT Analyzer: USB port or COM port.
If "USB port" is selected and no device can be found on
the USB port, the software automatically attempts to
locate a device on the COM port, and vice versa.
Language The CTA Remote Excel File Loader supports several
user interface languages. From the "Language" box,
select a language of your choice. The language can be
changed at runtime. Apart from the user interface, this
selection also changes the language of the data sheets.
129
CT Analyzer
Within these two functions, VBA source code can be embedded to handle these
events and to perform required user-defined tasks (see comment "enter your
code to be executed…"). The code is protected by error handling in order to
handle occurring exceptions and to show an appropriate message.
DoUserCodeAfterDataSheetCreation = lRet
Exit Function
ErrHnd:
MsgBox "Error occured: " & Err.Number & ", " & Err.Description
End Function
' enter your code to be executed after data is filled into sheet
DoUserCodeAfterDataSheetIsFilled = lRet
Exit Function
ErrHnd:
MsgBox "Error occured: " & Err.Number & ", " & Err.Description
End Function
130
CT Analyzer PC Tools
131
CT Analyzer
If the text update process is interrupted or fails, the device displays an error
message and automatically resets the user interface to English when it is
switched on the next time.
Attention: Only install language files that are included in the package of the
installed firmware version.
If you install a language file that does not belong to the same
package, the user interface may become unreadable.
Note: Refer to the OMICRON website for available languages or ask your
distributor for a special language file.
Figure 6-14:
Dialog for manual
selection of a serial port
132
CT Analyzer PC Tools
Connected device
The content of these fields is read from the connected CT Analyzer. The
information displayed in these fields cannot be edited.
Figure 6-16:
CT Analyzer Firmware
Update, Connected
device group box
Table 6-9:
User interface elements
in the Connected UI element Description
device group box
Serial number Serial number of the connected CT Analyzer.
Connected over PC port the CT Analyzer is connected to: USB port or
one of the serial COM ports.
Installed firmware Firmware version read from the CT Analyzer
version
133
CT Analyzer
Selected files
If you don’t have the update files available in a .zip file (see "Choose .zip file
containing firmware files" above), you have to specify the firmware file and the
user interface language file to be used for the update manually.
Figure 6-18:
CT Analyzer Firmware
Update, Selected files
group box
Table 6-10:
User interface elements
in the Selected files UI element Description
group box
Select files If checked, you can specify the firmware file and the user
manually interface language file used to update the CT Analyzer
individually.
Firmware file To update the firmware, click the corresponding
B R O W S E button to navigate to the desired firmware file
CTAnalyzer*.bin and select it. The path and the file
name are displayed in the field.
Please note the general information given in section
6.6.1 on page 131.
134
CT Analyzer PC Tools
UI element Description
User interface file To install a new language, click the corresponding
B R O W S E button to navigate to the desired firmware file
CTUser_xxx.bin and select it. The path and the file
name are displayed in the field.
Please note the general information given in section
6.6.1 on page 131.
Update method
Using these options you can select whether you want to update the firmware and
the user interface language file or the user interface language file only.
Figure 6-19:
CT Analyzer Firmware
Update, Update
method group box
135
CT Analyzer
) Refer to the "CTA Remote Control Manual" for a detailed description of the
remote software and how to use it.
136
CT Analyzer PC Tools
If you select this entry, a message dialog appears, notifying you that the data
were exported for use in NetSim.
Within NetSim you can use the CT data by selecting the menu item
PARAMETER | C URRENT T RANSFORMER.
Figure 6-23:
Message after exporting
the results for use in
NetSim
137
CT Analyzer
) Since the CPC Explorer was initially developed for the CPC 100, it contains
functions that are not needed for the CT Analyzer. For the CT Analyzer, some
of these functions are disabled in the CPC Explorer user interface.
138
CT Analyzer PC Tools
File browser
area
139
CT Analyzer
Viewing Files
The representation of the file contents in the right-hand side pane of the
CPC Explorer is generated on the basis of a stylesheet (also refer to ”Default
Stylesheet” on page 141).
To display a file on the PC, either right-click on the file name and select V I E W
from the context-menu, or double-click the file name.
) If you select a folder rather than a single test, right-click the folder name and
select V I E W from the context-menu. The contents of all tests of the entire folder
structure will be displayed in the right-hand side pane.
The toolbar buttons
Clicking on the right-hand side pane activates the file-specific buttons in the
toolbar. Otherwise, these buttons are grayed-out.
The R E F R E S H button renews the display of the left-hand side pane.
The F O N T button changes the size of the font with which the file in the right-
hand side pane is displayed:
• Clicking the F O N T button changes the size of the font to the largest
available.
• Clicking on the "down" symbol on the right-hand side of the button
displays a list of available font sizes.
Clicking on the "down" symbol on the right-hand side of the E D I T button
displays the available applications to edit the displayed HTML file.
1. Choose E D I T W I T H M I C R O S O F T W O R D F O R W I N D O W S to start the
application MS Word for Windows and load the HTML file of the displayed
report.
Note: MS Word for Windows 97 and previous versions do not support
editing HTML files.
2. Choose E D I T W I T H M S H T M L E D I T O R to edit the displayed report in the
WYSIWYG1 HTML editor.
) Contrary to editing the report with MS Word for Windows, the MSHTML mode
does not start an application. Instead, the cursor moves directly to the
beginning of the first line of the report displayed in the right-hand side pane,
and enables direct editing of the test report.
140
CT Analyzer PC Tools
Options
To open the Options dialog box, select T O O L S | O P T I O N S from the pull-down
menu. In this dialog box you can customize the CPC Explorer user interface.
Language
You can change the language of the CPC Explorer user interface. The
CPC Explorer re-starts automatically showing the user interface in the selected
language.
Default Stylesheet
Stylesheets describe how test reports are represented on screens or in print. A
stylesheet contains formatting information, and offers the possibility to influence
the presentation of test reports.
Upon request, OMICRON can provide you with customized XSL stylesheets1.
To apply a stylesheet, enter the path to this file into the entry field "Stylesheet",
or click B R O W S E to search for it.
) You can specify your own stylesheet in order to customize the representation of
the test reports in the right-hand side pane.
This documentation, however, will not describe how to create new stylesheets.
For more detailed information about XML, XSL and HTML please refer to
specialized literature and/or special websites, or contact the OMICRON
customer support.
141
CT Analyzer
142
Application Examples
7 Application Examples
The grounded terminal on the secondary side of the CT always has to be
connected to the black sockets of input "Sec" and the "Output" of the
CT Analyzer. Connecting the red sockets of the CT Analyzer to PE can result in
incorrect measurement and/or cause an automatic abortion of the measurement
with an error message.
143
CT Analyzer
Figure 7-1:
Measurement at a freely
Utility line Utility line
accessible CT
144
Application Examples
Attention: Input PRIM of the CT Analyzer has a low input impedance (330kΩ)
and can thus influence the measurement results.
145
CT Analyzer
L2 L3
Figure 7-3:
Ratio measurement at a L1
delta winding
transformer without
delta compensation
L2 L3
146
Application Examples
147
CT Analyzer
148
Application Examples
Figure 7-5:
Measurement at a H3 H2 H1
bushing-type CT (C) (B) (A)
(W) (V) (U)
CT
H0
(Mp)
(N)
149
CT Analyzer
150
Application Examples
Figure 7-6:
Measurement at a GIS Gas insulated busbar Earthing switch
(SF6) switch gear Utility line
disconnected
Circuit breaker
Burden
max. 3m
max. 100m
151
CT Analyzer
152
Application Examples
Figure 7-7:
Measurement at a bar-
type current transformer
P2 P1
Wrong!
Do not use open loops consisting of
individual single-wire measurement
cables!
Always use original OMICRON coax
measurement cables or twisted-pair
cables!
153
CT Analyzer
154
Application Examples
155
CT Analyzer
Plug-and-socket
connector
156
Application Examples
Primary conductor
Primary insulation
Toroidal core
Secondary winding
157
CT Analyzer
Caution: Do not connect both primary terminals to PE! This would cause
incorrect measurement results. Connecting both primary terminals to PE has the
same effect as a short in the CT.
158
Application Examples
Figure 7-12:
Noise reduction for CT
measurement
Utility line Utility line
Connection required,
if secondary side
terminal cannot be
disconnected
159
CT Analyzer
160
Calibrating the CT Analyzer
161
CT Analyzer
162
Error and Warning Messages
163
CT Analyzer
164
Error and Warning Messages
165
CT Analyzer
166
Error and Warning Messages
167
CT Analyzer
811.xxx Warning [811.xxx] Are you sure you want to delete <File Name>?
Reason: This warning always appears prior to the deletion of a file on the
Compact Flash card.
Solution: Press Y E S to delete the file or N O to return to the file system card
without deleting the file.
812.xxx Warning [812.xxx] Are you sure you want to delete all selected files?
Reason: This warning always appears prior to the deletion of files on the
Compact Flash card.
Solution: Press Y E S to delete the files or N O to return to the file system card
without deleting the files.
813.xxx Warning [813.xxx] One or more parameters could not be found. Some
calculations may not work.
Reason: The report file read from the Compact Flash card is of a newer format
than the device can generate.
Solution: The loaded file is incompatible or the software version on your
CT Analyzer is older than the software version that has generated the
report. To enable your CT Analyzer to read this report, it is necessary
to update the CT Analyzer software.
814.xxx Warning [814.xxx] You tried to load an old report. One or more parameters
could not be found. Some calculations may not work.
Reason: A required parameter cannot be found in the XML file.
Solution: You tried to load an old report. One or more parameters could not be
found. Some calculations may not work.
815.xxx Error [815.xxx] Folder must be empty. Can't remove folder
<Folder Name>.
Reason: The folder selected for deletion is not empty. Only empty folders can
be deleted.
Solution: Navigate to the folder you want to delete. Open the folder and delete
all contained files and subfolders. Then you can delete the empty
folder.
816.xxx Error [816.xxx] You tried to load a report built with a <Device>.
Reason: The xml file seems to be not a CT Analyzer report.
Solution: The report cannot be loaded by the CT Analyzer.
817.xxx Error [817.xxx] Xml file is not an OMICRON report.
Reason: The xml file has no valid OMICRON report style.
Solution: The report cannot be loaded by the CT Analyzer.
818.xxx Error [818.xxx] Moving of folders not supported.
Reason: Moving of folders is not supported by the CT Analyzer.
Solution: Select only files to move. Using the shift key and the cursor up/down
keys it is possible to select any number of files within a folder.
168
Error and Warning Messages
169
CT Analyzer
900.xxx Error [900.xxx] Power supply error, switch off the device and wait 1 minute
to restart.
Reason: The output power was larger than 350VA or the power supply is
defective.
Solution: Switch the CT Analyzer off, wait 1 minute and then switch it on again.
If this error occurs more frequently, you should contact your next
OMICRON service center.
901.xxx Warning [901.xxx] Protective earth (PE) connection in power supply cord
missing. Connect PE to equipotential terminal. Lethal voltage may occur
on housing and all terminals.
Reason: Earth wire break, no earth wire connected or mains supply has no
galvanic connection to protective earth (PE) potential.
Solution: The mains supply has not reference to protective earth or protective
earth is not connected. The mains supply must have galvanic
connection to PE. If you are using an isolation transformer, connect
one supply line of the CT Analyzer to PE.
Danger: Lethal voltages may occur at the housing if the equipotential
bonding terminal of the device is not connected to protective earth
potential!
Attention: If the mains supply is galvanically isolated from earth
potential, the device may become damaged.
Comment: If a galvanically isolated mains supply is used, voltage stress may
occur for the insulation system, the device is not constructed for.
Safety is no longer guranteed! Therefore, always use a mains
power cord with a protective earthing conductor connected to PE of
the mains supply. Connect the equipotential bonding connection of
the device to protective earth in order to prevent electric shock
caused by lethal voltages possibly present at the housing.
If this error is ignored, the device will probably work but safety is no
longer guaranteed.
903.xxx Warning [903.xxx] Excessive reverse power, don't disconnect any wires,
don't switch off mains, wait until power is dissipated.
ATTENTION: Lethal voltages on output terminals!
Reason: The device receives excessive reverse power so that the power
output had to be shortened to prevent overload of the internal output
stage.
Solution: This error is displayed if an excess amount of energy is fed back into
the CT Analyzer. Lethal voltages of up to many kV may occur if wires
are unplugged. The CT Analyzer discharges the connected inductor
with approximately 20Ws, but depending on the connected
inductance the discharge process may take some time.
170
Error and Warning Messages
904.xxx Error [904.xxx] Power supply error. Contact next Omicron service center.
Reason: Temperature limit of secondary power supply exceeded.
Solution: It is not possible to acknowledge this error until the temperature is
back in safe limits. It this error occurs, please contact your next
OMICRON service center.
905.xxx Error [905.xxx] Over-temperature of Power supply, wait until device has
cooled down.
Reason: Overtemperature in power supply. Wait until device has cooled down.
Solution: When this error is acknowledged as long it is active, the popup
window is closed and only the message in the status line remains
active until the overtemperature disappears. If the error is not
acknowledged, the popup window remains active.
Comment: The status line displays the flashing message "Excess temperature".
906.xxx Error [906.xxx] Excess temperature of power supply, wait until device has
cooled down.
Reason: The primary side power supply has reached its warning temperature
limit and entered save mode to cool down.
Solution: When this error is acknowledged as long it is active, the popup
window is closed and only the message in the status line remains
active until the overtemperature disappears. If the error is not
acknowledged, the popup window remains active.
Comment: The status line displays the flashing message "Excess temperature".
908.xxx Error [908.xxx] Device shut down.
Reason: Power failure at measurement interface module.
Solution: Internal power failure. This message is active as long as the internal
power failure exists. If the error message does not disappear, contact
your next OMICRON service center.
910.xxx Error [910.xxx] Internal hardware error.
Reason: Internal hardware error.
Solution: Overload on internal power stage. Try to start the measurement
again.
911.xxx Error [911.xxx] Power supply error, contact next Omicron service center.
Desired voltage xxxx, measured voltage yyyy.
Reason: One supply voltage on the measurement interface boards is out of
tolerance.
Solution: Contact your next OMICRON service center.
171
CT Analyzer
912.xxx Error [912.xxx] Excess internal temperature, wait until device has cooled
down.
Reason: Device-internal temperature at measurement interface > 75°C.
Solution: Excessive heating of the device. Prevent the device from direct
sunlight and wait until it has cooled down. If the error occurs without
previous exposure to direct sunlight, the error is probably caused by
a hardware error. Please contact your next OMICRON service
center.
Safety action:
All measurement processes are stopped and the error is displayed in
the status line and in a popup window. The error is active until the
temperature falls below 60°C.
929.xxx Error [929.xxx] Hardware failure, contact next Omicron service center.
Reason: The circuit for reverse power dissipation does not work properly.
Solution: Contact your next OMICRON service center.
930.xxx Error [930.xxx] Measurement Input Vsec defect, contact next Omicron
service center.
Reason: Power supply for analog input "Sec" is defective.
Solution: Contact your next OMICRON service center.
931.xxx Error [931.xxx] Measurement Input Vprim defect, contact next Omicron
service center.
Reason: Power supply for analog input "Prim" is defective.
Solution: Contact your next OMICRON service center.
932.xxx Error [932.xxx] Data error.
Reason: The sequence of reading of the analog input channels is not correct.
Solution: If this error occurs more frequently, contact your next OMICRON
service center.
933.xxx Error [933.xxx] Hardware failure, contact next Omicron service center.
Reason: Temperature detection circuit on secondary side is defective.
Solution: Send the device to your next OMICRON service center.
934.xxx Error [934.xxx] Reverse power, don't disconnect any wires, don't switch
off mains, wait until power is dissipated. I = xxxxA.
Reason: The device receives reverse power of > 10mA.
Solution: Wait until the power is dissipated within the device and the error
message disappears.
935.xxx Error [935.xxx] No valid CT Analyzer software. Insert a CF-Card with valid
Software and press "Update Firmw.".
Reason: Cannot find valid CT Analyzer software.
Solution: Insert a Compact Flash card with valid software (CTAnalyzer.bin)
in the directory A:\Omicron.
172
Error and Warning Messages
936.xxx Error [936.xxx] Corrupt Calibration Data for voltage inputs. Change to
tools menu and try to reset factory calibration. Until update of calibration
data the device will be not calibrated.
Reason: The calibration data checksum for the analog inputs is not correct.
Solution: Try to update the factory calibration using the tools menu. If this does
not solve the problem, contact your next OMICRON service center.
937.xxx Error [937.xxx] Corrupt Calibration Data for power output. Change to tools
menu and try to update factory calibration. Until update of calibration data
the device will be not calibrated.
Reason: The calibration data checksum for the power output is not correct.
Solution: Try to update the factory calibration using the tools menu. If this does
not solve the problem, contact your next OMICRON service center.
938.xxx Warning [938.xxx] MIF data block 1 corrupt.
Reason: The checksum of the factory settings data is faulty.
Solution: Licensing or device settings data are possibly corrupt. Contact your
next OMICRON service center.
939.xxx Warning [939.xxx] MIF data block 2 corrupt.
Reason: The checksum of the factory settings data is faulty.
Solution: Licensing or device settings data are possibly corrupt. Contact your
next OMICRON service center.
940.xxx Warning [940.xxx] CMOS data block corrupt. Check all device settings.
Press <Clear Values> to work with default values. Press <OK> to work with
corrupt device settings.
Reason: Device settings data block corrupt.
Solution: Check all device settings.
941.xxx Warning [941.xxx] Corrupt factory calibration data for voltage inputs.
Contact next Omicron service center.
Reason: Factory calibration data block for voltage inputs corrupt.
Solution: Contact your next OMICRON service center.
942.xxx Warning [942.xxx] Corrupt factory calibration data for power unit. Contact
next Omicron service center.
Reason: Factory calibration data corrupt.
Solution: Contact your next OMICRON service center.
943.xxx Error [943.xxx] Error overwriting calibration data.
Reason: Factory calibration data cannot be restored.
Solution: Contact your next OMICRON service center.
173
CT Analyzer
944.xxx Warning [944.xxx] Corrupt user text! Change to tools menu and try to
update text. Until update of user text, the device will use default text.
Reason: Checksum for user language support data invalid.
Solution: Copy a user text file (CTUser_xxx.bin) to the directory
A:\Omicron\ on the Compact Flash card and try to update the user
language support data using the "Update Text" function in the tools
menu.
174
Formulas and Definitions
Uc L main IE
RH UCT
Zb
Reddy (VB )
P2 S2
Np Ns
Figure 10-2:
Phasor diagram of a
current transformer
VC VB
R ϕ ICT
ϑ γ ICT CT Iex
β δ
Iex IST
175
CT Analyzer
The calculation of the ratio error is based on the excitation table. With a given
excitation voltage VC, the excitation table allows to find the corresponding
excitation current Iex and the phase between Iex and VC. Since the ratio error is
defined by the primary current Ip and the secondary load current ICT, the
excitation table can be used to determine the value of ICT that corresponds to
the given VC for a particular core. Provided that the rated turns ratio Nrated and
the turns ratio Nturns are known, the ratio error calculationrfor the given primary
current Ip and the burden impedance ZB,ϕ is as follows ( I S is always a
reference pointer):
1. Calculate the real and imaginary parts of the burden impedance:
R B = Z B ⋅ cos ϕ X B = Z B ⋅ sin ϕ
5. Look up the excitation current corresponding to this voltage and the phase ϑ
between them in the excitation table:
I ex ⇔ V C ; ϑ = ∠I E, V S
6. Calculate the phase between the secondary load current and the excitation
current:
β = ϑ–γ
176
Formulas and Definitions
⎛ I – -------------- Ip ⎞
-⎠
⎝ CT N
rated
Err ratio = ----------------------------------
-
Ip
---------------
N rated
r r
10.Calculate the phase displacement as the phase between I ST and I CT :
I ex ⋅ sin β
δ = arg ( I ST ) = arctan ⎛ --------------------------------------⎞
⎝ I CT + I ex ⋅ cos β⎠
N measured – ( I pn ⁄ I sn )
Turns ratio error (%) = ----------------------------------------------------- ⋅ 100
( I pn ⁄ I sn )
177
CT Analyzer
178
Formulas and Definitions
Is (b) Ic (a)
α
Ip / N (c)
179
CT Analyzer
Ip 2 Ip
I s + ⎛ ----⎞ – ⎛ 2 ⋅ I s ⋅ ---- ⋅ cos α⎞
2
Ic =
⎝ N⎠ ⎝ N ⎠
I psc I
- at that point where the composite current is 10 % of
FS = -------- ---p-
I pn N
Ic
ε c = ------------------- ⋅ 100
I pn
------- ⋅ FS
N
The excitation current now delivers an according value on the excitation curve
that can be used to calculate the burden current.
Note 2: The output current is not measured, but calculated instead. Therefore,
the accuracy of FS can differ by up to 10% from the reality.
180
Formulas and Definitions
2 2
Z = ( R burden + R CT_75° ) + X b
2
X b = Z burden ⋅ 1 – cos ϕ
2. Stepping up the current through the load (Z) (CT output current) until the
excitation current reaches 10% of the current flowing through the load (Z)
I ext
( -------- = 0.1 ).
I out
For each step, the calculation V emf = I out ⋅ Z has to be done to read the
corresponding Iext from the excitation graph.
Attention: The e.m.f. voltage has to be used to find out Iext for the calculation
of FSi as defined in the standard IEC 60044-1. Do not use the terminal
voltage to find the corresponding Iext, since this would lead to incorrect
results. Because the standard excitation graph according to IEC 60044-1
only shows the terminal voltage, you should switch to the ANSI excitation
graph on the CT Analyzer. With the tools CTA Remote Excel File Loader or
CTA Remote Control it is possible to get the e.m.f. values from the excitation
graph.
3. After the point where the excitation current is 10% of the output current is
found, the FSi can be calculated as follows:
I out
FSi = --------
I sn
The result is displayed as FSi in the user interface and in all reports.
181
CT Analyzer
2 2 2
Z = ( 10 ⋅ cos ϕ + 3.907 ) + ( 10 ⋅ 1 – cos ϕ ) => 13.33 Ω
V emf = FSi ⋅ I sn ⋅ Z
V emf = 82.1 V
182
Formulas and Definitions
3. Read Iext from the ANSI excitation graph at Vemf (e.g. at 82 V: Iext = 0.588 A).
I ext 0.588
Error [%] = -------------------- ⋅ 100 = ------------------ ⋅ 100 => Error ∼ 9, 5 %
FSi ⋅ I sn 6.16 ⋅ 1
5. FSi must be lower than the FS on the CT-Object card (FSi ≤ FSrated):
5.88 < 10 = true => ok
Attention: Often it is not possible to read the point where the CT reaches an
error of 10% on the excitation graph because the excitation graph is not
measured far enough. In this case, use the highest available point to verify FSi.
183
CT Analyzer
Is (b) Ic (a)
α
Ip / N (c)
184
Formulas and Definitions
Ip 2 Ip
I s + ⎛ ----⎞ – ⎛ 2 ⋅ I s ⋅ ---- ⋅ cos α⎞
2
Ic =
⎝ N⎠ ⎝ N ⎠
I psc I
- at that point where the composite current is 5 % or 10 % of
ALF = -------- ---p-
I pn N
Ic
ε c = ----------------------- ⋅ 100
I pn
------- ⋅ ALF
N
The excitation current now delivers an according value on the excitation curve
that can be used to calculate the burden current.
Note 2: The output current is not measured, but calculated instead. Therefore,
the accuracy of ALF can differ by up to 10% from the reality.
185
CT Analyzer
2 2
Z = ( R burden + R CT_75° ) + X b
2
X b = Z burden ⋅ 1 – cos ϕ
ext I V
emf
2. Determination of the excitation current where -------- ⋅ 10 = 1 with I out = ----------
-
I out Z
3. Reading of the e.m.f. voltage (Val) from the excitation cuve (e.m.f. voltage
that corresponds to the excitation current determined in step 2.).
Attention: The e.m.f. voltage has to be used to find out Iext for the calculation
of ALFi as defined in the standard IEC 60044-1. Do not use the terminal
voltage to find the corresponding Iext, since this would lead to incorrect
results. Because the standard excitation graph according to IEC 60044-1
only shows the terminal voltage, you should switch to the ANSI excitation
graph on the CT Analyzer. With the tools CTA Remote Excel File Loader or
CTA Remote Control it is possible to get the e.m.f. values from the excitation
graph.
V al
4. Calculation of ALF: ALF = -------------
-
Z ⋅ I sn
The result is displayed as ALFi in the user interface and in all reports.
V al
ALF ⋅ I sn ≥ -------
Z
ext I
Val is the voltage where -------- ⋅ 10 = 1 (10% error current).
I out
186
Formulas and Definitions
100 ⋅ Î ext
ε^ ε̂ = --------------------------------
2 ⋅ I sn ⋅ K ssc
Iext Peak excitation current read from the excitation graph at Emax.
2 2
Emax E max = K dt_meas ⋅ K ssc ⋅ ( I sn ⋅ ( R CT + R b ) + X b )
Ktd meas Ktd calculated (parameter from Excitation card), refer to section
10.17 ”Determination of Transient Dimensioning Factor (Ktd)” on
page 196.
Kssc Kssc from the settings (CT-Object card).
RCT Measured winding resistance (RCT) at 75°C.
Rb Burden entered in CT-Object card for parameter "VA" or "Burden",
depending on the selected page on the Excitation card (using the soft
key R E S U L T W I T H N O M . V A or R E S U L T W I T H B U R D E N ).
187
CT Analyzer
Ek
K x = ----------------------------------
-
( R ct + R b ) ⋅ I sn
188
Formulas and Definitions
In other words, Iprim is the primary current where the CT has an error of 10%.
Then, the rated secondary terminal voltage can be calculated as follows:
V B = I CT × Z b
2 2
V c = I CT × X b + ( R b + R ct )
To understand the formulas please refer to the equivalent circuit for current
transformer (refer to section 10.1 on page 175).
Isn rated secondary current
Ipn rated primary current
Iprim primary AC current
VC, UC secondary excitation voltage at main inductance (Lmain)
Rct resistance of secondary winding
Zb external impedance at rated frequency
Rb resistive part of burden
Xb reactance of burden
UCT, VB voltage at secondary terminal
ICT current at the secondary terminals
189
CT Analyzer
Ψr
Remanence Ψr K r = 100 *
Ψs
Coercivity
Magnetizing
force Ψr = remanent flux
H Ψs = saturation flux
Kr = remanence factor in %
190
Formulas and Definitions
Flux [Vs]
0,95
Ψ − Ψ2
L12 = 1 0,9
I1 − I 2 0,85
0,8
Ψ = measured flux 0 2 4 6 8
L12
0.5 < < 1.5 (2)
L23
L13 > 30mH (3)
191
CT Analyzer
Ψn, In = peak flux and peak current at one point of the excitation curve
L 90%
∑L Ln
Lm = ---------------------
20%
192
Formulas and Definitions
For current transformers with non-gapped cores, the knee is defined as the point
(Vkn, Ikn) where the tangent is at 45 degrees to the abscissa. For current
transformers with gapped cores, the knee is defined as the point where the
tangent is at 30 degrees to the abscissa.
S nd ε = 10
TPS 0.25 ε = 10
TPX 0.5 30 0.9 ε = 10
TPY 1.0 60 1.8 ε = 10
TPZ 1.0 180 ± 18 5.3 ± 0.6 ε = 10
193
CT Analyzer
194
Formulas and Definitions
Limits of current error for measuring current transformers (class 3 and 5):
Accuracy class Permitted current (ratio) error (± in %)
at percentage of rated current Ipn
50% of Ipn 120% of Ipn
3 3 3
5 5 5
The load must have a cos ϕ of 0.8 if the load is above 5VA and a cos ϕ of 1 if it
is between 1 and 5VA. A load < 1VA is not allowed (see IEC 60044-1, section
11.2).
195
CT Analyzer
TP
⎛ T ⎞ Ts −T p )
K td = K tf = ωTP ⎜⎜ p ⎟⎟ + 1 = Ktf at tmax
⎝ Ts ⎠
TP
⎛ T ⎞ Ts −T p )
K td = K tf = ωTP ⎜⎜ p ⎟⎟ + 1 = Ktf at tmax
⎝ Ts ⎠
T p * Ts ⎛ Tp ⎞
t max = * ln ⎜⎜ ⎟⎟
(T p − T s ) ⎝ Ts ⎠
Step 2
Calculation of Ktd at accuracy limiting time of cycle 1:
196
Formulas and Definitions
Step 3
Calculation of Ktd at accuracy limiting time of cycle 2 (if defined).
Ktd'' for COCO cycle with consideration of remanence:
Consideration of Remanence
− ( t fr + t ''al )
T
In order to consider the remanence in the term e s
, we limit the value
of the term to the remanence value. Therefore, the value of this term cannot be
smaller than 0.8, if the remanence is 0.8.
In M A I N M E N U -> S E T T I N G S -> K T D C A L C U L A T I O N , you can select whether
you want to have Ktd calculated according to IEC 60044-6 (remanence not
considered) or according to Omicron (remanence considered).
Result: The highest of the values Ktd, Ktd’ or Ktd’’ is the value used for Ktd.
197
CT Analyzer
Overview
Start
Resistance Ratio
measurement measurement
no no
no
no
198
Formulas and Definitions
Start Start
End End
P/M guessing
(protection or measurement CT)
Start
no no
P/M =
P/M =
measurement
protection core
core
End
199
CT Analyzer
Start
no no
no no
Standard values ALF:
5
Second no
Guess FS: Guess ALF: 10
knee point?
FS = Ip/I-pn ALF = Ip/I-pn 15
(see page 179) (see page 184) 20
yes
30
Choose the next Choose the next
higher standard lower standard Guess E1/Ie1:
value from table. value from table.
E1 = knee point Standard values FS:
Ie1 = Iexc at E1 1
1.5
2
5
10
Guess Kx: 15
Kx = Ip/I-pn 20
(see page 188) 30
End
Start
no
P/M = P
?
yes
End
200
Formulas and Definitions
Start
Guess Val/Ial:
yes Val = knee point Second no
Class = TPS (ΔIexc/ΔEk = knee point
? 10% : 100%) ?
Ial = Iexc at Ek
no yes
End
201
CT Analyzer
Start
no P
P/M = M
?
M yes
yes
Class ok?
no
yes
Class ok?
no
yes
Class ok?
no
yes
Class ok?
no
yes
Class ok?
no
Assess class 1
yes
Class ok?
no
Assess class 3
yes
Class ok?
no
End
202
Formulas and Definitions
Start
no P
P/M = M
?
M yes
yes
Class ok?
no
End
203
CT Analyzer
No No
Set Uref equal to the
voltage of the highest Get CT Type Factor = 10 . 0
No ANSI-30 KP found?
point of the excitation
curve
Yes Meas CT? Standard = IEC-1?
KP
U ref = U ANSI 30
Yes Yes
No Standard = IEC-1?
Yes
Kssc specifed?
Burden Table
Yes
Factor = ALF No
Yes = MCT_ANSI
Yes
Yes
No Meas CT? No
Yes
Yes
End
204
Formulas and Definitions
205
CT Analyzer
206
Parameter Definition Syntax for the Different Standards
207
CT Analyzer
Results:
Knee point: 10/50, UKnee, IKnee,
? ALF/FS, TS, LS, Lm, Kr,
Excitation curve, Ratio, εc, RCT, Class
Kx
(?) Results:
Knee point: 10/50, VKnee, IKnee,
RCT ? Kx, Ek, Ie, E1, Ie1, TS, LS, Lm, Kr, Excit.
? curve, Ratio, εc, RCT, Class
Ie
? Assessments:
Ie1 Kx, EK, Ie, E1, Ie1, RCT
? Class:
Ek εc,<= 0.25% (Turns ratio error)
?
E1
Results:
? Knee point: 10/50
FS, TS, LS, Lm, Kr,
Excitation curve [ARMS],
? ? UKnee, IKnee ,
Ratio error (ε, Phase error, εc, Class,
RCT
M
Ext FS (?) Assessments:
(120%) 1-30 According to guessed parameters
? Softkeys
FS 1,
RCT
1,5, 2, 5,
? 10, 15, Results:
0,1 20, 30 Knee point: 10/50
0,2 FS, TS, LS, Lm, Kr,
? Excitation curve [ARMS],
0,2S
UKnee, IKnee , Ratio, εc, Class, RCT
0,5
0,5S
1 Assessments:
3 FS, Fsi, Rct
Class:
Values in brackets () are 5
Ratio and phase error is within limits
default values. according to IEC 60044-1
208
Overcurrent number and
Standard Class Duty cycle definition Results/Assessments
extended definition
RCT
?
Val
?
Ial
Classes TPS and TPZ
?
E1
?
Ie 1
?
TP
(0 ms)
Results:
Knee point: 10/50
UKnee, IKnee, KSSC, Ktd, Ts, LS,
Lm, Kr,
TPZ
Excit. curve (Vemf, IPEAK),
KSSC Ratio, Phase, Polarity, RCT
?
Ktd
?
TP Assessments:
(20 ms) Ktd*KSSC, RCT,
Ts = +- 10% TsSetting,
TS Class:
Values in brackets () are Ratio-Err <= 1%
?
default values. Phase-Err <= 180+-18 min
RCT
?
11.3 Parameter Definition Syntax for IEC 60044-6
Parameter Definition Syntax for the Different Standards
209
210
CT Analyzer
IEC 60044-6
KSSC t1 tal1
TPX C-t1-O
(?) [1-5.000s] (40ms)
Results:
Ktd
t1 Knee point: 10/50,
(?) C-t1-O-tfr- tal1
[1-5.000s] (40ms
VKnee, IKnee, KSSC, Ktd, Ts,
TP C-t2-O LS, Lm, Kr,
(0.100s) )
(20ms)
Excit. curve (Vemf, APEAK),
t2 Val, Ial, e^, Emax, RCT,
[1-5.000s] tal2
RCT (40ms) Ratio, Phase, Polarity, N
(0.100s)
(?)
tfr Assessments:
[1-5.000s] Kdt*Kssc, RCT,
(0.300s) Class:
e^@Kssc*Ktd <= 10%,
Ratio-Err <= 0.5%;
Phase-Err <= 30 min
KSSC
Classes TPX and TPY
TPY t1
tal1
(?) C-t1-O [1-5.000s]
(40ms) Results:
TP (0.100s)
Ktd (20ms)
Knee point: 10/50,
(?) t1 VKnee, IKnee, KSSC, Ktd, Ts, LS,
TS C-t1-O-tfr- tal1 Lm, Kr,
[1-5.000s] (40ms
C-t2-O ) Excit. curve (Vemf, APEAK),
(?) (0.100s)
Val, Ial, EPS^, Emax, RCT,
t2 Ratio, Phase, Polarity, N
RCT [1-5.000s] tal2
(40ms)
(0.100s)
(?) Assessments:
tfr Kdt*Kssc, RCT,
[1-5.000s] Ts = +- 30% TsSetting
(0.300s) Kr <= 10%
Class:
e^@Kssc*Ktd <= 10%,
Values in brackets () are Ratio-Err <= 1.0%;
Phase-Err <= 60 min
default values.
11.4 Parameter Definition Syntax for IEC 60044-6
Parameter Definition Syntax for the Different Standards
Results:
IEEE C57.13 ? Knee point: ANSI 30/45, UKnee, IKnee
Vb, TS, LS, Lm, Kr,
Excitation curve [ARMS], Ratio, RCT
Assessments:
-
P ?
Burden
definition Results:
Knee point: ANSI 30/45, UKnee, IKnee
default Vb, TS, LS, Lm, Kr,
B-2 Excitation curve [ARMS], Ratio, RCT
B-1
B-2 Assessments:
B-4 -
B-8
Vb
100-1200
default 100
Softkeys
Results:
10, 20, 50,
Knee point: ANSI 30/45, UKnee, IKnee,
100, 200,
Vb, TS, LS, Lm, Kr,
400, 800
Excitation curve [ARMS], Ratio, RCT
C
K Assessments:
Vb,
T
Vknee (only K class)
RCT
?
Thermal current
rating factor
Burden definition
default
B-0.2
B-0.1
B-0.2
B-0.5
B-0.9
B-1.8
211
CT Analyzer
212
Care and Cleaning
213
CT Analyzer
214
Technical Data
13 Technical Data
Guaranteed data are specified for an ambient temperature of 23°C ± 5°, a
power supply of 115/230VAC, and after a warm-up time longer than 15 minutes.
Guaranteed data are valid for the period of one year after factory adjustment.
215
CT Analyzer
Table 13-4:
Technical data for
measurement input Measurement input "Prim"
"Prim"
Voltage ranges 0 - 0.03 / 0.3 / 3 / 30VAC (auto ranging)
Accuracy 0.1% guar.
Input impedance 0 - 15V: 330kΩ
15 - 30V: 120kΩ to 330kΩ, depending on the
voltage.
Insulation Reinforced insulation (R) to all other circuits
216
Technical Data
Table 13-7:
Accuracy of ratio values
displayed in the ratio Ratio accuracy for 5 A CTs at rated current
table (5 A CTs )
CT ratio I-sn Rated power1 Typical Guaranteed
accuracy accuracy
0.2 - 1 5 1.0 - 75VA 0.05% 0.1%
1 - 2000 5 0 - 75VA 0.02% 0.05%
2000 - 5000 5 0 - 75VA 0.03% 0.1%
5000 - 10000 5 0 - 75VA 0.05% 0.2%
1. Nominal burden of the CT.
217
CT Analyzer
Table 13-8:
Phase accuracy Phase accuracy at rated current
Resolution 0.01min
Accuracy (cos ϕ 0.8 - 1)
– typical 1 min
– guaranteed 3 min
Table 13-9:
Turns ratio accuracy Turns ratio accuracy
Resolution 0.01 turns
Accuracy
– typical 0.05 %
– guaranteed 0.1 %
218
Technical Data
219
CT Analyzer
1 nc
6 nc
2 RxD (data in / receive)
7 RTS (out)
3 TxD (data out / transmit)
8 CTS (in)
4 nc
9 +5V
5 GND (signal ground)
Do not connect!
For internal use only!
Figure 13-2:
Connection cable for 9-pole (DB9) null modem or
remote control interface crossover cable, 2 x female
Connections required:
1 2 3 4 5 6 7 8 shield
1 2 3 4 5 6 7 8 shield
220
Technical Data
221
CT Analyzer
222
Scope of Delivery and Accessories
223
CT Analyzer
224
Scope of Delivery and Accessories
225
CT Analyzer
Figure 14-1:
Accessories (1)
Battery clamp set with 4mm banana Crocodile clamps for secondary
sockets for primary side connection side connection with 4mm banana
(VEHZ0652) socket (VEHZ0651)
226
Scope of Delivery and Accessories
Figure 14-2:
Accessories (2)
227
CT Analyzer
Figure 14-3:
Accessories (3)
228
Contact Information / Technical Support
Asia, Pacific
OMICRON electronics Asia Ltd, Hong Kong
Phone: +852 2634 0377
E-Mail: support@asia.omicron.at
Web: www.omicron.at
45
Contact Information / Technical Support
46
Index
A
About page (help system) . . . . . . . . . . . . . . . 38
B
accessories . . . . . . . . . . . . . . . . . . . . . . . . . 224 bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
accuracy limiting current for IEC 60044-1 class bar-type CT
PX (Ie) . . . . . . . . . . . . . . . . . . . . . 63, 81–82, 97 application example . . . . . . . . . . . . . . . 152
accuracy limiting factor (ALF, ALFi) basic wiring for burden test . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . 62, 79–80, 97 basic wiring for CT test . . . . . . . . . . . . . . . . . 26
calculation . . . . . . . . . . . . . . . . . . . . . . . 184 block diagram . . . . . . . . . . . . . . . . . . . . . . . . 23
accuracy limiting secondary excitation current Burden (parameter) . . . . . . . . . . . . . . . . . . . 61
(Ial) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65, 98 Burden card . . . . . . . . . . . . . . . . . . . . . . 35, 72
accuracy limiting voltage for IEC 60044-1 class connecting the burden . . . . . . . . . . . . . . . 73
PX (Ek) . . . . . . . . . . . . . . . . . . . . . . . 63, 81–82 overload indication . . . . . . . . . . . . . . . . . 72
address running the test . . . . . . . . . . . . . . . . . . . . 73
OMICRON address . . . . . . . . . . . . . . . . 229 test results . . . . . . . . . . . . . . . . . . . . . . . . 73
ALF, refer to accuracy limiting factor test settings . . . . . . . . . . . . . . . . . . . . . . . 72
ALFi, refer to accuracy limiting factor burden entry logic . . . . . . . . . . . . . . . . . . . . 205
ambient temperature (default setting) . . . . . . 48 Burden parameter . . . . . . . . . . . . . . . . . . . . 73
ANSI burden test
assessed parameters . . . . . . . . . . . . . . . 97 application example . . . . . . . . . . . . . . . 157
class definition . . . . . . . . . . . . . . . . . . . . 193 bushing-type CT
conditions for OK assessment . . . . . . . . 100 application example . . . . . . . . . . . . . . . 148
excitation graph . . . . . . . . . . . . . . . . . . . . 85
excitation test results . . . . . . . . . . . . . . . . 83
parameter definition syntax . . . . . . . . . . 211
application examples C
bar-type CT . . . . . . . . . . . . . . . . . . . . . . 152
calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 161
burden test . . . . . . . . . . . . . . . . . . . . . . . 157
care and cleaning . . . . . . . . . . . . . . . . . . . . 213
bushing-type CT . . . . . . . . . . . . . . . . . . 148
cascading style sheet in CPC Explorer 140–141
delta winding transformer . . . . . . . . . . . 145
changing fuses . . . . . . . . . . . . . . . . . . . . . . . 13
freely accessible CT . . . . . . . . . . . . . . . 143
class definition
gapped core . . . . . . . . . . . . . . . . . . . . . . 155
according to IEC 60044-1 . . . . . . . . . . . 194
GIS (SF6) switch gear . . . . . . . . . . . . . . 150
according to IEC 60044-6 . . . . . . . . . . . 193
multicore CTs . . . . . . . . . . . . . . . . . . . . . 157
according to IEEE C57.13 . . . . . . . . . . . 193
noise reduction techniques . . . . . . . . . . 158
Class multiplying factor . . . . . . . . . . . . . . . . 62
tapped CTs . . . . . . . . . . . . . . . . . . . . . . 154
Class parameter . . . . . . . . . . . . . . . . 60, 97–99
unwound core, excitation curve . . . . . . . 156
cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
wye winding transformer . . . . . . . . . . . . 147
climate conditions . . . . . . . . . . . . . . . . . . . . 221
231
CT Analyzer
232
Index
233
CT Analyzer
IEC 60044-6
H assessed parameters . . . . . . . . . . . . . . . 97
class definition . . . . . . . . . . . . . . . . . . . . 193
hardware class TPS parameters . . . . . . . . . . . . . . . 65
block diagram . . . . . . . . . . . . . . . . . . . . . 23 class TPS results . . . . . . . . . . . . . . . . . . 81
Compact Flash card slot . . . . . . . . . . . . . 20 class TPX parameters . . . . . . . . . . . . . . . 66
display . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 class TPX results . . . . . . . . . . . . . . . . . . 81
fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 class TPY parameters . . . . . . . . . . . . . . . 68
grounding . . . . . . . . . . . . . . . . . . . . . . . . . 20 class TPY results . . . . . . . . . . . . . . . . . . 81
I/0 key . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 class TPZ parameters . . . . . . . . . . . . 70–71
inputs and outputs . . . . . . . . . . . . . . . . . . 21 class TPZ results . . . . . . . . . . . . . . . . . . . 81
mains connection unit . . . . . . . . . . . . . . . 20 conditions for OK assessment . . . . . . . 100
on/off switch . . . . . . . . . . . . . . . . . . . . . . . 20 excitation graph . . . . . . . . . . . . . . . . . . . . 85
overview . . . . . . . . . . . . . . . . . . . . . . . . . . 19 excitation test results . . . . . . . . . . . . . . . . 81
soft keys . . . . . . . . . . . . . . . . . . . . . . . . . . 22 parameter definition syntax . . . . . . 209–210
status LEDs . . . . . . . . . . . . . . . . . . . . . . . 21 IEC-ID field . . . . . . . . . . . . . . . . . . . . . . . . . . 57
version . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 IEEE C57.13
help system . . . . . . . . . . . . . . . . . . . . . . . . . . 38 assessed parameters . . . . . . . . . . . . . . . 97
hotline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229 class definition . . . . . . . . . . . . . . . . . . . . 193
conditions for OK assessment . . . . . . . 100
excitation graph . . . . . . . . . . . . . . . . . . . . 85
excitation test results . . . . . . . . . . . . . . . . 83
I parameter definition syntax . . . . . . . . . . 211
I/0 key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Ikn, refer to knee point current
Ial, refer to accuracy limiting secondary I-meas parameter . . . . . . . . . . . . . . . . . . . . . 73
excitation current inputs
I-DC parameter . . . . . . . . . . . . . . . . . . . . . . . 76 overview . . . . . . . . . . . . . . . . . . . . . . . . . 21
Ie, refer to accuracy limiting current for technical data . . . . . . . . . . . . . . . . . . . . 216
IEC 60044-1 class PX installation
Ie1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 CPC Explorer . . . . . . . . . . . . . . . . . . . . 138
IEC 60044-1 CT Analyzer PC Toolset . . . . . . . . . . . . 110
assessed parameters . . . . . . . . . . . . . . . 97 instrument security factor
class definition . . . . . . . . . . . . . . . . . . . . 194 (FS, FSi) . . . . . . . . . . . . . . . . . . . 64, 80, 84, 97
class P parameters . . . . . . . . . . . . . . . . . 62 calculation . . . . . . . . . . . . . . . . . . . . . . . 179
class PR parameters . . . . . . . . . . . . . . . . 63 insulation coordination . . . . . . . . . . . . . . . . 219
class PX parameters . . . . . . . . . . . . . . . . 63 I-p parameter . . . . . . . . . . . . . . . . . . . . . . . . 91
conditions for OK assessment . . . . . . . . 100 Ipn, refer to rated primary current
excitation graph . . . . . . . . . . . . . . . . . . . . 85 Isn, refer to rated secondary current
excitation test results . . . . . . . . . . . . . . . . 79 Itest, refer to test current for ext. burden
parameter definition syntax . . . . . . . . . . 208
parameters for meas. CTs . . . . . . . . . . . . 64
234
Index
Manufact. field . . . . . . . . . . . . . . . . . . . . . . . 58
K max. allowed secondary excitation
current at E1 (Ie1) . . . . . . . . . . . . . . . . . 97–98
K*Kssc . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 maximum allowed excitation current . . . 81–82
K, refer to secondary loop time constant maximum allowed excitation current at E1 . . 66
knee point current (Ikn) . . . . . . . . . . . 79, 81, 83 measurement cable set . . . . . . . . . . . . . . . 225
knee point voltage (Vkn) . . . . . . . . . . 79, 81, 83 measurement inputs
knee point, calculation . . . . . . . . . . . . . . . . 192 overview . . . . . . . . . . . . . . . . . . . . . . . . 216
Kr, refer to remanence factor . . . . . . . . . . . 190 technical data . . . . . . . . . . . . . . . . . . . . 216
Kssc, refer to rated symmetrical short-circuit minimum computer requirements
current factor for CPC Explorer . . . . . . . . . . . . . . . . . . 138
Ktd calculation (setting) . . . . . . . . . . . . . . . . . 48 for CT Analyzer PC Toolset . . . . . . . . . 109
Ktd*Kssc . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 misc. settings . . . . . . . . . . . . . . . . . . . . . . . . 48
Ktd, refer to transient dimensioning factor multicore CTs
K-value graph graph . . . . . . . . . . . . . . . . . . . 88 application example . . . . . . . . . . . . . . . 157
Kx, refer to dimensioning factor for IEC 60044-1
class PX
N
L navigating in file system . . . . . . . . . . . . . . . . 46
NetSim export . . . . . . . . . . . . . . . . . . . . . . 137
language selection New CT test (main menu) . . . . . . . . . . . . . . 40
CPC Explorer . . . . . . . . . . . . . . . . . . . . . 141 noise reduction for measurement . . . . . . . . 158
CT Analyzer . . . . . . . . . . . . . . . . . . . . . . . 48 nominal burden . . . . . . . . . . . . . . . . . . . . . . 60
CTA Start Page . . . . . . . . . . . . . . . . . . . 115 non-saturated inductance (Lm) . . . . . 80, 82, 84
software tools . . . . . . . . . . . . . . . . . . . . . 115 calculation . . . . . . . . . . . . . . . . . . . . . . . 192
language update . . . . . . . . . . . . . . . . . . 52, 131
licenses update . . . . . . . . . . . . . . . . . . . . . . . 53
Lm, refer to non-saturated inductance
loading a file . . . . . . . . . . . . . . . . . . . . . . . . . 44
Location field . . . . . . . . . . . . . . . . . . . . . . . . . 57
O
Ls, refer to saturated inductance Object field . . . . . . . . . . . . . . . . . . . . . . . . . . 57
OMICRON News service . . . . . . . . . . . . . . 114
on/off switch . . . . . . . . . . . . . . . . . . . . . . . . . 20
operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
M Optional field . . . . . . . . . . . . . . . . . . . . . . . . 58
output
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . 40 overview . . . . . . . . . . . . . . . . . . . . . . . . . 21
File Handling . . . . . . . . . . . . . . . . . . . . . . 44 technical data . . . . . . . . . . . . . . . . . . . . 215
New CT test . . . . . . . . . . . . . . . . . . . . . . . 40 overload indication . . . . . . . . . . . . . . . . . 72, 77
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 48 overview, hardware overview . . . . . . . . . . . . 19
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
mains fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
mains power supply, technical data . . . . . . 215
mains socket . . . . . . . . . . . . . . . . . . . . . . . . . 20
235
CT Analyzer
t-al . . . . . . . . . . . . . . . . . . . . . . . . . . . 67, 69
P tfr . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67, 69
T-meas . . . . . . . . . . . . . . . . . . . . . . . . . . 75
P/M parameter . . . . . . . . . . . . . . . . . . . . . . . 59 Tp . . . . . . . . . . . . . . . . . . . . . 65–66, 68, 70
parameter T-ref . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
ALF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Ts . . . . . . . . . . . . . . . . . . . . . . . . 63, 68, 70
Burden . . . . . . . . . . . . . . . . . . . . . . . . 61, 73 VA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Class . . . . . . . . . . . . . . . . . . . . . . 60, 97–99 V-al . . . . . . . . . . . . . . . . . . . . . . . . . . 65, 98
Class mult. factor . . . . . . . . . . . . . . . . . . . 62 Vb . . . . . . . . . . . . . . . . . . . . . . . . . . . 70, 84
Cos ϕ . . . . . . . . . . . . . . . . . . . . . . . . . 61, 73 V-DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Delta compensation . . . . . . . . . . . . . . . . . 62 V-kn . . . . . . . . . . . . . . . . . . . . . . . 79, 81, 83
Δϕ . . . . . . . . . . . . . . . . . . . . . . . . . . . 97–99 V-meas . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Duty . . . . . . . . . . . . . . . . . . . . . . . . . . 66, 68 Z . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
ε . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97–99 parameter definition syntax . . . . . . . . . . . . 207
E1 . . . . . . . . . . . . . . . . . . . . . 64–65, 81–82 pasting a file . . . . . . . . . . . . . . . . . . . . . . . . . 47
εc . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 PC Toolset . . . . . . . . . . . . . . . . . . . . . . . . . 109
Ek . . . . . . . . . . . . . . . . . . . . . . 63, 81–82, 97 PDF manual . . . . . . . . . . . . . . . . . . . . . . . . . 10
ext. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 peak instantaneous error (ε-peak) . . . . . 83, 98
Frequency . . . . . . . . . . . . . . . . . . . . . . . . 61 calculation . . . . . . . . . . . . . . . . . . . . . . . 187
FS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 permissible time to accuracy (tal) . . . . . . 67, 69
I-al . . . . . . . . . . . . . . . . . . . . . . . . . . . 65, 98 phase accuracy . . . . . . . . . . . . . . . . . . . . . 217
I-DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 phase deviation ε . . . . . . . . . . . . . . . . . . 97–99
Ie . . . . . . . . . . . . . . . . . . . . . . 63, 81–82, 97 phase error . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Ie1 . . . . . . . . . . . . . . . 64, 66, 81–82, 97–98 Phase field . . . . . . . . . . . . . . . . . . . . . . . . . . 57
I-kn . . . . . . . . . . . . . . . . . . . . . . . . 79, 81, 83 Phase parameter . . . . . . . . . . . . . . . . . . . . . 91
I-meas . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 phase table . . . . . . . . . . . . . . . . . . . . . . . . . . 92
I-p . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 accuracy of values . . . . . . . . . . . . . . . . 217
I-pn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 polarity
I-sn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 ok / not ok . . . . . . . . . . . . . . . . . . . . . . . . 91
I-test . . . . . . . . . . . . . . . . . . . . . . . . . . 72, 75 power supply . . . . . . . . . . . . . . . . . . . . 12, 215
K*Kssc . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 primary time constant (Tp) . . . . . 65–66, 68, 70
Kssc . . . . . . . . . . . . . . . . . 65–66, 68, 70, 82
Ktd . . . . . . . . . . . . . . . . . . . . . 66, 68, 70, 82
Ktd*Kssc . . . . . . . . . . . . . . . . . . . . . . . . . 98
Kx . . . . . . . . . . . . . . . . . . . . . . . . . 63, 80, 97
P/M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Q
Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 qualification of operating personnel . . . . . . . 11
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 QuickTest software . . . . . . . . . . . . . . . . . . 116
Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Rct . . . . . . . . . . . . . . . . . . . . . . . . . . . 97–99
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
R-meas . . . . . . . . . . . . . . . . . . . . . . . . . . 76
R-ref . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Standard. . . . . . . . . . . . . . . . . . . . . . . . . . 59
t1, t2 . . . . . . . . . . . . . . . . . . . . . . . . . . 67–68
236
Index
237
CT Analyzer
238
Index
V
VA parameter . . . . . . . . . . . . . . . . . . . . . . . . 60
Val, refer to rated equivalent excitation limiting
secondary voltage
VB, refer to rated secondary terminal voltage
acc. to IEEE C57.13
V-DC parameter . . . . . . . . . . . . . . . . . . . . . . 76
version
CPC Explorer software . . . . . . . . . . . . . . . 2
CT Analyzer software . . . . . . . . . . . . . 2, 38
hardware . . . . . . . . . . . . . . . . . . . . . . . . . 38
language . . . . . . . . . . . . . . . . . . . . . . . . . 38
Vkn, refer to knee point voltage
V-meas parameter . . . . . . . . . . . . . . . . . . . . 73
W
warning messages . . . . . . . . . . . . . . . . . . . 163
weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
winding resistance . . . . . . . . . . . . . . 61, 97–99
winding resistance accuracy . . . . . . . . . . . . 216
wiring diagrams . . . . . . . . . . . . . . . . . . . . . . . 38
wiring for burden test . . . . . . . . . . . . . . . . . . 25
wiring for CT test . . . . . . . . . . . . . . . . . . . . . . 26
4 wire connection technique . . . . . . . . . . 28
wye winding transformer
application example . . . . . . . . . . . . . . . . 147
Z
Z parameter . . . . . . . . . . . . . . . . . . . . . . . . . 73
239
CT Analyzer
240