Sunteți pe pagina 1din 2

1E66-04

en
Bilirubin Calibrator
Bilirubin Calibrator 1E66
ARCHITECT 307009 / R07
S1E6T0
Read Highlighted Changes: Revised July 2016.

This package insert contains information for the use of Bilirubin STORAGE AND STABILITY
Calibrator on the ARCHITECT cSystems.
• Store unopened Bilirubin Calibrator at 2 to 8°C.
Package insert instructions must be carefully followed. Reliability of
• Unopened calibrator is stable until the expiration date when
assay results cannot be guaranteed if there are any deviations from
stored at 2 to 8°C.
the instructions in this package insert.
• Opened calibrator is stable for 7 days when stored at 2 to 8°C
INTENDED USE and kept tightly capped.
For use in the calibration of the Bilirubin assays. PREPARATION OF CALIBRATOR
CONTENTS / MATERIALS PROVIDED Bilirubin Calibrator requires no preparation prior to use.
1E66-04 Bilirubin Calibrator, 3 x 5 mL INSTRUCTIONS FOR USE
Bilirubin Calibrator is prepared in a bovine serum-based solution.
Calibration is performed by running a water blank and the Bilirubin
Analyte levels are adjusted with bilirubin extracts and synthetic
Calibrator set. Water for the blank is provided by the instrument.
derivatives. Preservatives are also present.
1. Verify that the correct calibrator values have been entered into
Actual analyte concentrations for this lot of calibrator are listed in the
the calibration file.
Bilirubin Calibrator Value Sheet, packaged with the calibrator. Verify
2. Allow calibrator to come to room temperature.
that the lot number listed on each calibrator bottle agrees with the lot
number printed on the value sheet. 3. Mix bottle several times by gentle inversion.
When converting the units for an assay, the following items must be 4. Open bottle, place an appropriate amount of each calibrator in a
edited, if applicable, using the appropriate conversion factor. separate sample cup, and place in the assigned positions.
• Calibrator Concentration 5. Cap bottle tightly and return to refrigerated storage immediately
after use.
• Qualitative/Interpretation Range
6. Perform calibration as indicated in the ARCHITECT System
• Linear Range
Operations Manual.
• Retest rules ranges
• Constituent assay ranges for calculated assays INDICATIONS OF INSTABILITY OR DETERIORATION
• Reference Range Instability or deterioration should be suspected if there are visible
• Panic/Extreme Range signs of leakage, extreme turbidity, microbial growth, or if calibration
• Expected QC Values does not meet the appropriate package insert and/or ARCHITECT
System Operations Manual criteria.
STANDARDIZATION
LIMITATIONS OF THE PROCEDURE
Total Bilirubin
6L45 BiliT and 7P32 TBili have been standardized against the Accurate and reproducible results are dependent upon properly
Doumas method/NIST SRM 916.1 functioning instruments and reagents, storage of product as directed,
and good laboratory technique.
Direct Bilirubin
8G63 BilD has been standardized against the Doumas method2 and BIBLIOGRAPHY
confirmed by comparison of human samples. 1. Doumas BT, et al. Candidate reference method for determination of
total bilirubin in serum: development and validation. Clin Chem 1985;
PRECAUTIONS 31(11):1779–1789.
• 2. Doumas BT, et al. Pitfalls in the American Monitor kit methods for
• For In Vitro Diagnostic Use determination of total and “direct” bilirubin. Clin Chem 1982; 28(11):
• Do not use components beyond the expiration date. 2305–2308.
• Do not mix components from different kit lot numbers. 3. US Department of Labor, Occupational Safety and Health
Administration, 29 CFR Part 1910.1030, Bloodborne pathogens.
• Avoid direct exposure to light.
4. US Department of Health and Human Services. Biosafety in
• CAUTION: This product requires the handling of human Microbiological and Biomedical Laboratories. 5th ed. Washington, DC:
specimens. It is recommended that all human-sourced materials US Government Printing Office; December 2009.
be considered potentially infectious and handled in accordance 5. World Health Organization. Laboratory Biosafety Manual. 3rd ed.
with the OSHA Standard on Bloodborne Pathogens. Biosafety Geneva: World Health Organization; 2004.
Level 2 or other appropriate biosafety practices should be 6. Clinical and Laboratory Standards Institute (CLSI). Protection
used for materials that contain or are suspected of containing of Laboratory Workers From Occupationally Acquired Infections;
infectious agents.3-6 Approved Guideline—Fourth Edition. CLSI Document M29-A4. Wayne,
Safety Data Sheets are available at www.abbottdiagnostics.com or PA: CLSI; 2014.
contact your local representative.
For a detailed discussion of safety precautions during system
operation, refer to the ARCHITECT System Operations Manual,
Section 8.

1
Key to Symbols
Consult instructions for use

Manufacturer

Temperature limitation

Use by/Expiration date

Calibrators 1 and 2
Calibrator 1
Calibrator 2
Conventional Units
Authorized Representative in the
European Community
Identifies products to be used
together
Information needed for United
States of America only
In Vitro Diagnostic Medical
Device
Lot Number
Manufactured for
Product of USA
Reagent
List Number
SI Units

The ARCHITECT cSystem family of instruments consists of c4000,


c8000, and c16000 instruments.
ARCHITECT, c4000, c8000, c16000, and cSystem are trademarks of
Abbott Laboratories in various jurisdictions. All other trademarks are
property of their respective owners.

Abbott Laboratories Abbott GmbH & Co. KG


Diagnostics Division Max-Planck-Ring 2
Abbott Park, IL 60064 65205 Wiesbaden
USA Germany
+49-6122-580

Abbott Laboratories
Revised July 2016.
©2005, 2016 Abbott Laboratories

S-ar putea să vă placă și