Sunteți pe pagina 1din 77

URIT-880/881/882

Semi-auto Chemistry Analyzer


Service Manual
(For Engineers)

V1.0

URIT MEDICAL ELECTRONIC CO., LTD.


April 2015 Revision
URIT-880/881/882 Service Manual

Content
Preface ............................................................................................................................................................................... 1
1. Introduction ........................................................................................................................................................................... 2
1.1 Instrument Synopsis .................................................................................................................................................... 2
1.2 Scope of Application .................................................................................................................................................... 2
1.3 Instrument Structure And Operating Principle............................................................................................................ 2
1.4 Main Parameter................................................................................................................................................................... 3
1.5 Appearance And Internal Structure ............................................................................................................................. 5
2. Basic Operational Flow .......................................................................................................................................................... 9
2.1 Operation Process ....................................................................................................................................................... 9
2.2 Outline of The Main Menu Interface ........................................................................................................................... 9
2.3 Concise Operation Process ............................................................................................................................................... 11
3. Acceptance Check And Installation ..................................................................................................................................... 14
3.1 Acceptance Check ...................................................................................................................................................... 14
3.2 Safety Matters ................................................................................................................................................................... 14
3.3 Installation ................................................................................................................................................................. 16
3.4 Precautions For Instrument Use ............................................................................................................................................. 25
3.5 Deposit.................................................................................................................................................................... 26
4. Debugging of The Instrument .............................................................................................................................................. 26
4.1 Debugging Tool and Reagents .......................................................................................................................................... 26
4.2 Appearance and Safety Inspection ............................................................................................................................ 26
4.3 Power on Self-test ............................................................................................................................................................. 27
4.4 Parameter Settings.................................................................................................................................................. 27
4.5 Performance Test ............................................................................................................................................................................... 34
5. Set Up..................................................................................................................................................................................................... 35
5.1 Measurement related Set Up .................................................................................................................................................. 35
5.2 Printer Setup.............................................................................................................................................................. 38
5.3 Language Setup ......................................................................................................................................................... 38
5.4 Time Setup ................................................................................................................................................................. 38
5.5 Screen Setup .............................................................................................................................................................. 38
5.6 Account Setup............................................................................................................................................................ 38
5.7 Authorization Management ...................................................................................................................................... 39
5.8 Driver ......................................................................................................................................................................... 39
5.9 Subject setup ............................................................................................................................................................. 39
5.10 Export and Import.................................................................................................................................................... 39
5.11 PC host Setup........................................................................................................................................................... 41
5.12 LIS/HIS System Setup ............................................................................................................................................... 41
5.13 Factory Reset ................................................................................................................................................................... 41
5.14 Others ...................................................................................................................................................................... 41
6. Systems of The Instrument .................................................................................................................................................. 41
6.1 Sampling System ............................................................................................................................................................... 41
6.2 Optical Path System .......................................................................................................................................................... 44
6.3 Liquid Road System ........................................................................................................................................................... 46
6.4 Hardware Circuit ........................................................................................................................................................ 48
6.5 SMPS .......................................................................................................................................................................... 52
I
URIT-880/881/882 Service Manual
6.6 Incubation Groove ..................................................................................................................................................... 53
7. Instrument Maintenance ..................................................................................................................................................... 56
7.1 Instrument Maintenance ........................................................................................................................................... 56
7.2 Maintenance.............................................................................................................................................................. 58
7.3 Clean & Maintenance of The Instrument .................................................................................................................. 64
7.4 Transportation and Storage .............................................................................................................................................. 66
8. Maintenance Skill ................................................................................................................................................................ 67
8.1 Introduction ............................................................................................................................................................... 67
8.2 General Failures and Solutions .................................................................................................................................. 67
8.3 Notes ......................................................................................................................................................................... 70
Appendix 1:Replaced Spare Parts List ................................................................................................................................... 72
Appendix2: Order and Abbreviations of Test Items............................................................................................................. 73

II
URIT-880/881/882 Service Manual

Preface
This is URIT-880 Semi-auto Chemistry Analyzer Service Manual. It focuses on helping engineers understand the
structure, principle, operation, routine maintenance, fault handling of the instrument. Operator should read through the
service manual and be familiar with it before operating, proper use of the instrument is the key to ensure the testing result is
right and the instrument has no fault.

Symbol explanation
The implication of the graphic identities on semi-auto chemistry analyzer:

Watch out , refer to the attached file Danger! Shock hazards

Look,burn care Biological hazards

Protect ground line Power on

Power off IVD

EFUP Prevent hot and radiation source

Serial number Manufacturing enterprise

Can lead to harm


Waste recycling
to body

Refer to instructions

1
URIT-880/881/882 Service Manual

1. Introduction

1.1 Instrument Synopsis

URIT-880、URIT-881、URIT-882 Chemistry Analyzer ( instrument hereinafter ) are microcomputer, the touch screen
type, filter type semi-auto chemical analyzer which designed basing on photoelectron colorimetric principle. It is available
for qualified test in vitro for body liquid (including serum、urine、cerebrospinal fluid、plasma and so on),and it can test many
kinds of biochemistry project. Please read through operation manual, and take full acknowledge of the instrument’s
function.
The instrument has friendly user interface, humanized design, and it’s easy to operate. Users can operate according to
system,and put sample to liquid inlet pipe.press “start”key to do operating test, it can print the result by built-in thermal
printer.

1.2 Scope of Application

The analyzer applies to medical department,medical scientific research department to do clinical biochemical tests and
immunologic test about quantitative analysis samples of blood serum, blood plasma,urine,cerebrospinal fluid

Notice
Maybe some samples can’t analyze according to testing parameters and the reagent be used in. Please consult
with relevant manufactures and agent.

1.3 Instrument Structure And Operating Principle

1.3.1 Instrument Structure


The instrument is made of sample absorbing system, light path system, liquid line system and hardware circuit.
The difference of each model of the semi-auto chemistry analyzer is as followed:

PS/2
Model Incubation Groove
Interface
built-in incubation
URIT-880 support
groove
URIT-881 Not possess support
URIT-882 Not possess Not support

1.3.2 Operating Principle


The testing principle of the instrument mainly uses Lambert-beer Law,mix the chromogenic reagent with patient’s
sample in a certain proportion,the mixed liquid is absorbed by cuvette,using the mixture’s characteristic of absorbing light
with specific wavelength,and confirm the concentration according to the absorption value.
2
URIT-880/881/882 Service Manual
Lambert-beer’s Law

 A beam of parallel monochromatic light can pass through an object with light-absorbing substance(gas、liquid or solid),
the thickness of the substance is L, concentration is C. When light passing through the object, some photons are absorbed,

light intensity falls to ,The following formula can be deduced:

Proportional constant:ε

Input intensity:
Output
Having measured the I、I0、L, the formula can calculate the concentration C

Picture 1:Process of data conversion

1.4 Main Parameter

1.4.1 Performance Index


(1) wavelength accuracy and repetition
(2) accuracy deviation:not exceeding:±2nm;
(3) half width:not exceeding 10nm
(4) stray light: absorbance no less than 4.0
(5) absorbance linearity
a absorbance :0~≤0.5,not extend±5%;
b absorbance>0.5~≤1.0,not extend±4%;
c absorbance>1.0~≤1.8,not extend±2%;
d absorbance>1.8~≤3,not extend ±5%。
(6) Repetition:CV≤1.0%
(7) Stability:absorbance change≤0.005Abs/20mins
(8) the accuracy of temperature of absorption cell and amplitude:the accuracy of temperature of absorption cell

3
URIT-880/881/882 Service Manual
should not deviate over ±0.2℃;temperature amplitude should not over±0.1℃。
(9) Cross infection rate:≤1.0%(fixed absorption cell (cuvette) is not suitable)。 cross contamination rate:≤1.0%
(10) testing range:-0.3Abs~4.0Abs~4.0Abs

1.4.2 Parameters And Features


(1) Test principle: take the photo-electricity colorimetry principle.
(2) Test method: end-point method,dynamic method(rate method),two-dot method,double wavelength method,
ABS method , polygonometry.
(3) Spectroscope mode: High-resolution optical filter and halogen lamp,fully closed,static state,array,beam-splitting
optical system after chopping wave.Wavelength range 300nm~800nm,8 kinds wavelengths are optional (wavelength are
340nm,405nm,492nm,510nm,546nm,578nm,630nm,700nm).
(4) Test project:≥300。
(5) Light source:6V/10W Long-life halogen lamp,automatic sleep function.
(6) Display screen:7-inch color touch screen,resolution ratio:800×480.
(7) Measuring time:0~999s,program is available.
(8) Delay time:0~999s,program is available.
(9) Colorimetry system:the colorimeter cell (or cuvette) of the instrument is 32μL the volumetric flow type quartz
colorimeter cell (or cuvette), or choose stainless steel quartz colorimeter cell(or cuvette).
(10) Optical path:10mm
(11) Absorb fluid sample size:100μL~9999μL(the accuracy of sample size is can be auto-compensated),500μL is
recommended。
(12) Aspirating accuracy:±30μL
(13) Temperature control:choose room temperature、37℃,cuvette is absorption cell。
(14) Quality control function:it has quality control and statistic program,and it can display quality control curve.
(15) The reaction course monitor:the instrument can display the reaction curve, access some abnormal condition like
substrate exhaustion in real time. The instrument can achieve advanced-data management function through connecting the
professional connection software with computer workstation.
(16) Result storage:parameter setting of 300 ordinary items, parameter setting of 20 curve items, 30000 measurement
results.
(17) Printout:the instrument can connect with printing device,and output combined report with Chinese and English.
(18) Peripheral interface:one SD interface,three USB interfaces,one PS/2 interface,one parallel port,one RJ-45
interface。Notice:URIT-882 semi-auto chemistry analyzer doesn’t support PS/2 interface。
(19) Half width:≤10nm。Half width
(20) Network:instrument supports LIS/HIS system,Inspection department information network can be implemented 。
(21) Extended function:the instrument has information sharing connection port。Extended function
(22) Supply voltage:880/881:AC220V,50Hz, 882 :AC110V,60Hz 。

4
URIT-880/881/882 Service Manual
(23) Power:130VA

5
URIT-880/881/882 Service Manual
(24) Fuse cutout:T1AL 250V.
(25) Ambient temperature:(10~32)℃
(26) Relative humidity:40%~85%
(27) Atmosphere pressure:86kPa~106kPa
(28) External circuitry insulation limited under single failure condition:≤0.1Ω.
(29) EFUP:10 years.

1.5 Appearance And Internal Structure

1.5.1 Appearance

Picture 2: Instrument overview

(1) Switch:the power switch is located in the right side of the instrument panel. “I”means turn on the power;“O”means turn
off the power。
(2) Touch screen:resolution ratio:800×480,display brightness is adjustable. Touch screen:resolution rate 800×480 lattice,
the display brightness can be adjusted.

(3) Liquid inlet pipe:absorb instilled water、calibration solution、 quality control serum、samples and so on。
(4) “start”button:when the instrument is testing,press”start”,and the instrument will start to absorb distilled water or liquid
need to be tested.

6
URIT-880/881/882 Service Manual

Picture 3 Instrument rear view(URIT-880/URIT-881)

Picture 4 Instrument rear view(URIT-882)

(1) Power interface:connect with main power。Power interface:connect main power


(2) Ground pole:protect grounding。Earthing rod:protect earthing
(3) Waste tube’s mouth:after testing,waste liquid discharge to waste system through waste tube’s mouth.please watch
the height of the waste liquid at all times,and empty the waste liquid in time to ensure the test proceeds normally.
(4) Fan’s air outlet: cooling fan applies to emit the instrument’s heat, to ensure the inner temperature in instrument
won’t be overheating.
(5) Internet access:mainly use to exchange data and up-date program with third-party device.
(6) Parallel interface:Standard 25-core parallel socket, using to connect with external printer.
(7) SD card slot:after inserting SD card, users can import or export data,update program.

7
URIT-880/881/882 Service Manual
(8) USB interface No.1:support USB device(mass storage media、USB keyboard)and export data.
(9) USB interface No.2:support USB keyboard.
(10) PS/2 interface:connect PS/2 standard keyboard、bar code scanner.

1.5.2 Internal Structure of The Instrument

Peristaltic electronic filter


URIT-880 main board machine

incubation groove
switching
power supply

light source lamp signal amplified board

图5 internal structure(Ⅰ)

(1) Switching power supply:supply power to main board (model:RT-125A v1:+5V v2:+12V v3:-5V)。
(2) Filter: 880/881:input AC220V; 882:output AC110V。
(3) Inchworm motor:when the peristaltic electrical machine runs,the instrument can absorb liquid.
(4) Signal amplified board:transform the optical signal to electrical signal,and send the amplified signal to the main
board.
(5) Main board of URIT-880:the instrument controls core and peristaltic electrical machine,receives and send data
and signals.

8
URIT-880/881/882 Service Manual

printing interface board

display screen fixed sheet


neutrosphere

Picture 6 instrument’s internal structure(Ⅱ)

(1) Interface board of printer:Transfer the instructions from the main board to built-in thermal printer
(2) Interface board of display:Apply for transiting the data from main board to display screen,and supply power.
(3) A fixed sheet steel on display screen: fix the display screen

9
URIT-880/881/882 Service Manual

2. Basic Operational Flow

2.1 Operation Process

Semi-auto Chemistry Analyzer mainly includes six menus: [test], [test result], [absorbance measurement], [quality
control], [setup], [help].

2.2 Outline of The Main Menu Interface

After logging in system, users can enter the main menu interface,as is shown in the picture 7. Main menu interface is made
of includes six menus : “Test”, “Test result”, “Absorbance measurement”, “Quality control”, “Setup”, “Help”, operator can
enter proper layout with touch of a button.
There are some common icons in the top right corner of the interface: current temperature value(up is cuvette’s

temperature, down is incubation groove’s temperature)、“↓”or“↑”、 ”、system time and system menu key. Among

these,“↓”means the system stops heating,“↑”means the system is heating,when lights, it means the halogen is on.
System menu button includes lock users function and restart the instrument function. When the instrument detects the USB

10
URIT-880/881/882 Service Manual

keyboard,“ ” will appear. When the instrument detects SD card, “ ”will appear, click it, and the content of SD card

appears, when the instrument detects the USB storage ”will appear,click it, and the content of SD card appears.

Picture 7 Main menu interface

Main menu function description

【 Measurement 】 : in measurement interface , allow users to modify chemistry project and parameter,do project
measurement,curve measurement,calibration and quality control test. Among these,curve measurement only applies for
multi-point calibration project.

【Measuring result】:allow users to check the measurement results,credit patient’s information and print report and so on.

【Absorbance measurement】:The absorbance of measured sample.

【QC】:Mainly do mathematical statistics(query quality control data, quality control chart ) and calculation.If users have
input quality control target value and SD value,please use the data users input when drawing quality control graphs,or use
the average value the instrument automatically calculates and SD value as parameter drawing quality control chart.

【Setup】:In setup interface,users can set time,language,screen brightness,printer and authorization management

【Help】:enter the help interface,users can look up some relevant information about the instrument,and ask for assistance

Soft keyboard function description


Input number, English case and Chinese characters in the instrument are mainly by soft keyboard.

 Press“abc”or“中”can switch Chinese-English input methods。When soft keyboard displays“abc”,Chinese


character can be input, and displays ”中”English letters can be input。

 Press“↑”Can switch to English case letters input methods.

11
URIT-880/881/882 Service Manual
 Press “123”,soft keyboard is only allowed to enter numbers. In number keyboard,press“中”,switch to Chinese
character input methods;press“abc”can switch to letter input methods.

 “←”is equivalent to the“Delete”on computer keyboard,move the cursor forward can delete the content on cusor’s
left.

 In Chinese character input mode, when a pinyin possesses many Chinese character,use“↑”to “↓”turn pages, look
through the needed character,get the character you want by touching gently.

 After input, click “Do” to save the current content and exit the soft keyboard.

 Function of“backspace”key is to quit the soft keyboard and back to the top level interface, and not save the
current content.

Notice
Do not use something sharp to touch the surface of the screen,or the screen will damage.

2.3 Concise Operation Process

2.3.1 Turn on
Turn off the general power switch in accordance with normal operation.after self-check, it will enter the main menu
interface automatically.
Click “wash” in absorbance measurement interface or sample measurement interface, absorb alcohol for 5 minutes to
clean dirt and bubbles,then use distilled water to wash pipeline for about 3~5 times. Preheat the instrument 30 minutes to
ensure the light source is stable and reach to the preset temperature, then use the distilled water again to wash the instrument
5 times to ensure the cuvette and pipeline is clean.

2.3.2 A/D Calibration


Please follow the steps to operate calibration(when the instrument is used for the first time or used every week, it needs
to operate A/D calibration):
Main menu—>“setup”interface—>“measurement relevant”interface—>“A/D calibration”interface —> click“credit”,
put distilled water into alumina injector—>execute“absorption” order , absorb distilled water—> click“measurement”—
>A/D value will be automatically corrected , when the A/D value is stable—>click“application”to save —
>click“backspace”(Notice:the A/D value range of the 300nm~800nm wavelength is 2000~4000)
2.3.3 Project Parameter Setting
Click “measurement” in the main interface,choose the project needing to measure, then the system enters to project
parameter setup interface.(Don’t modify the project name by default,if there is new project,please newly create in empty
project )
Take TP as example:choose TP project,and enter to set the TP project parameter.
Basic 1
1.measurement code:TP That is the project’s name,which can be input English name or Chinese name
measurement methods(end point、rate、two-point、dual-wavelength)

12
URIT-880/881/882 Service Manual
2.testing mode:end-point Apply for detect the absorption intensity of the reactant(please set in accordance
3.dominant wavelength:546 with reagent operation manual)
Only apply for dual-wavelength project, subwavelength can’t be the same with
4.subwavelength:405 dominant wavelengths(TP project can’t be set)
Cup(water)、reagent、and sample blank can be chose
5.blank type:cup Absorbance of blank liquid applies for calculate calibration factor ( only end-
6.blank absorbance:-- point method project can be set)
The temperature of the cuvette can be set as room temperature(--)and 37℃,
7.temperature:37 system default is 37℃

Basic 2
1.unit:g/L unit(according to the reagent manual)
2.sample inhaled quantity:500 Adding the quantity of reaction liquid is necessary when doing test,unit is μL,
500μL is recommended
3.delay time:10 The time from finishing absorption to test before(set in accordance with reagent
operation manual)
4.measuring time:3 Read the time which joined the absorbance calculation(set in accordance with
reagent operation manual)
5.print or not:□ printing set of printer(①not print--□ ②print--■)
6.decimal place:1 Decimal place of the measured result can choose 0~4(set in accordance with
reagent operation manual,the project TP is 1)

Threshold value

Threshold value:65-85 Refer to the lower bound and upper bound


Substrate exhaustion Substrate exhaustion:only applies for ratio-method project(TP project can’t be set

□ enable signal Use the function or not(①not use--□ ②use--■)
The reaction of direction: Positive reaction or negative reaction (set in accordance with reagent operation
manual)
Threshold value: Refer to the lower bound and upper bound

Calibration
1.concentration:70.01 Standard value , if the reagent has standard liquid,the input
corresponding value(standard liquid concentration of reagent collocation)
Input factor value according to reagent instruction or after finishing calibration
2.calibration factor:0.0000 test,system will self correct the standard factor.

2.3.4 Sample Test


(1) Press the “run”, and it will enter the test interface.
(2) Immerse alumina injector into cup filled with distilled water,press“rinsing”,wash it once,then press“start”,distilled
water should be blank
(3) When hinting absorb blank solution,use reagent absorbed in instrument as blank.
(4) When hinting absorb blank solution,first,absorb standard liquid as standard to test, then save it. No matter installing for
13
URIT-880/881/882 Service Manual
the first or replacing reagent batch number, it should calibrate again. (If choose calibration,the factors will become invalid,
give the factors according to calibration. If no calibration, calculate the result according to the factor being input)
(5) When it prompts absorb samples, click quality control,and absorbs quality control liquid to do quality control test.
(operate quality control according to the actual laboratory situation)
(6) Then absorb samples, test the first sample;then absorb the second sample,and test the second sample.
(7) In testing process, click retest can test the current number again.
(8) After testing all samples, immerse alumina injector to cup filled with distilled water,press “wash” to clean the cuvette.
(9) After click the “backspace” key, another project can be chosen to start.

2.3.5 Print Sample Report


In main interface,click “measuring result”—>”list of laboratory test report”,then enter to interface of list of laboratory
test report.users not only can choose to print all data;but also can use hands to touch the specified sample result to enter
laboratory sheet interface,press “print”,and start to print laboratory sheet.

2.3.6 Instrument Cleaning


After test, use cleaner fluid wash the cuvette for 5 times, and then use distilled water to clean the cuvette 10~20 times.
After that, absorb the distilled water to make the cuette and pipe line be full of distilled water in order to avoid dust
disturbing the testing result.

2.3.7 Turn off the Instrument


After testing, start the instrument’s washing function,turn off the power at last.After turning off, the associated
maintenance also needs to do,for convenience to normally use next time.
(1) Put reagent into freezer;
(2) Deal with serum tube in accordance with relevant regulation;
(3) Before turning off after everyday use,use cleaner fluid clean the instrument 5 times,;
(4) Use distilled water to clean the cuvette 10~20 times;
(5) The cuvette and pipeline should be filled with distilled water before turning off the power;
(6) Deal with waste liquid in waste system;
(7) Check whether the table-board is stained, if there is,please use soft cloth to wipe.

Biohazard
(1) Some substance in reagent,controlling liquid,standard liquid,cleaner fluid,waste liquid are regulated by
pollution regulation and emission standard anomaly,and consult with reagent producer or distributor.
(2) When dealing with waste liquid, please wear gloves and work clothes to ward off infection,or wear
protective glasses if necessary.

14
URIT-880/881/882 Service Manual

3. Acceptance Check And Installation

3.1 Acceptance Check

(1) After receiving, the receiving units needs to unpack and accept in accordance with the following process:
(2) Open the packing cases carefully,and take out the instrument and assembly.
(3) Check and accept the instrument,it’s amount and it’s condition according to packing list.
(4) If the quantity is not in conformity or the instrument has some damage,please contact with distribution departments
and URIT.
Packing list of URIT-880 Semi-Auto Chemistry Analyzer

No. Items Amount


1 URIT-880 Semi-Auto Chemistry Analyzer 1
2 URIT-880 Operation Manual 1
3 URIT-880 SOP documents 1
4 URIT-880 concise operation process 1
5 Certificate of quality(URIT) 1
6 Protective tube(T2A-250V) 2
7 power line/(3GTJ1-3GTJA) 1
8 Silicone tube(3m*5m) 1
9 Waste liquid bottle(cap drilling)/800 1
10 Thermal printer paper(width:57mm) 1
11 Peristaltic pump tube:1.65mm*3.35mm/PhgrMed BFT NSF-51/( import from Saint-Gobain) 1
12 PTFE tube(ID1.02 OD1.68)( import from Saint-Gobain) 0.5
13 The schematic diagram of sampling tube's installation and replacement 1
14 Packing list of URIT-880 Semi-Auto Chemistry Analyzer 1
15 Service Manual 1
16 list of Poisonous and harmful material 1
17 The "Three Licenses" of URIT-880 1

Warning
(1) Users need to provide appropriate environment and space.
(2) Only technician of URIT can install the instrument.

3.2 Safety Matters

In order to use the instrument safely and effectively, please comply with these following items:
(1) Avoid destroying and burning
Please observe these following precaution to avoid destroying.

15
URIT-880/881/882 Service Manual
Warning
(1) The instrument must install in the environment in accordance with operating manual.
(2) please connect with URIT when instrument needs to change environment.

(2) Protection of bodily injure


In order to avoid injures,please observe these following precaution when using the instrument.

Notice
(1) when the instrument is on work, don’t touch any parts which is moving.
(2) replace the light bulb,turn off the power and let it be cooling to operate again,or it may cause burn.

(3) Protection of biohazard


To prevent biohazard effectively, please comply with these following matters needing attention.
Biohazard
Use the sample incorrectly may cause infection.Please wear gloves and work clothes to ward off
infection,or wear protective glasses if necessary when operating.If contacted with skin,please deal with it in
accordance with user’s work standards and consult doctor.

Notice
Some reagents have extreme acidity or alkalinity, please use with caution, and prevent direct contact reagent
with the hands and clothes directly. If hands or clothes accidentally contact, immediately wash with soap
suds. If eyes contact it accidentally, immediately flush with water,and consult eye doctor.

(4) Treat waste liquid


To prevent pollution and bodily injure caused by waste liquid, please comply with these following matters needing
attention.

Biohazard
Some substance in reagent,controlling liquid,standard liquid,cleaner fluid,waste liquid are regulated by
pollution regulation and emission standard.Please treat it in accordance with relevant provision in medical
waste management regulations.

(5) Operating environment

Caution
(1) Please install the instrument in accordance with the specified environment of the operation manual 。
Installation or operation beyond specified requirement may cause wrong results and system damage.
(2) If users want to change system status, please connect with URIT.

16
URIT-880/881/882 Service Manual

(6) Prevent the disturbance of microwave and noise

Notice
Don’t place any device with abnormal noise near instrument,for the electromagnetic waves may disturb the
instrument’s work, please turn off the mobile phone and transceiver in the installation room,.

(7) System usage

Notice
(1) Please do system usage in accordance with the operation manual.Operate incorrectly may cause wrong
results, even system damage and personal injury.
(2) Use the system for the first time,to ensure the system work properly,please calibrate before operating
quality control.
(3)When using system,operators must do quality control program,or it can not guarantee the result’s
accuracy
(4) Communication interface of the analysis department is set to connect with operation department’s.
Please use the special cable provided by URIT or its agent to connect with analysis department and
operation department.
(5) Don’t touch the screen,mouse and keyboard with hand with water or chemicals.

(8) Other common matters needing to pay attention

Notice
(1) In order to prevent the damage of the keyboard,display,mouse,please do not contact it with wet hands or
chemicals.
(2) check sample’s contamination(dust 、 fibrinogen),check every sample’s quality,whether it has air
bubbles.
(3) Do regular scheduled maintenance,check and replace components,only correct maintenance can receive
correct analysis result.
(4) Replace main components,such as light source lamp,alumina injector,peristaltic pump, cuvette and so
on. Do calibration and refer to various measures,and consult manufacturer.
(5) System needs to set sample size,reagent size,wavelength and calibration value and so on,Please setup in
accordance with the manual of reagent.Check the water quality,standard results,analysis results, and air
bubbles of pipeline.

3.3 Installation

Only technician of URIT Guilin or technician authorized by URIT can install the URIT-880, users need to provide
proper environment and space. When the instrument need to relocate,please connect with URIT.

warning
Only technician of URIT Guilin or technician authorized by URIT can install the URIT-880.

3.3.1 Size and Weight of The Instrument


Size(height includes machine leg):360mm×320mm×145mm(length×width×height)。

17
URIT-880/881/882 Service Manual
URIT-880:net weight:5.9kg,rough weight:11.2Kg。URIT-
881/882:net weight:5.28kg,rough weight:10.58Kg。

3.3.2 Instrument’s Installation Environment


 Only for the use of indoor installation
 The work table should be smooth. (inclination<1/200)
 Excellent ventilation
 The environment should be as clean as possible.
 Avoid direct sun
 Avoid putting next to heat source and air source.
 No corrosive gas and combustible gas

 No shock in work table

 No big noise source and power disturbing.


 Keep stay away from brush type engine and electronic contact device which is often on-off.
 Keep stay away from the devices which may give out microwave,such as:mobile phone、transponders etc.
 Power voltage:880/881:AC220V,50Hz;882 :AC110V,60Hz,three-cores power line,well ground
 Power consumption:130VA
 Environmental temperature:(10~32)℃
 Relatively humidity:40%~85%
 Atmospheric pressure:86kPa~106kPa

Caution
(1) Please install the instrument in accordance with the specified environment of the operation manual。
Installation or operation beyond specified requirement may cause wrong results and system damage.Don’t
put in the place where the instrument can’t disconnecting device easily.

(2) If the system state needs to change,please contact with after-sale department of URIT.
Warning
(2)The power outlet should be grounded correctly.Wrong ground may cause electrical shock and system
damage.
(2)Please be sure of the output voltage of power outlet meets system requirement, and install proper fuse.

18
URIT-880/881/882 Service Manual
3.3.3 Space Requirement

unit:mm
Picture 8 Installment space requirement
(1) Installment space should be larger than the instrument’s dimension,please refer to picture 8.
(2) The installation table should be smooth,and the inclination should be less than 1/200,the table should be stable.

3.3.4 Installation Steps


1) Waste Liquid Pipeline Connection
(1) Remove the instrument from the packing box,lay it out in the working table horizontally and stably.In order to
dissipate heat of the ventilator and disconnect the power switching,the distance between the back of the instrument and wall
or obstacle should be no less than 20cm;the button of the instrument should be kept the air flowing.(Don’t press or damage
the alumina injector when transport)
(2) After putting away instrument, remove the waste liquid tube from packaging bag,check whether the waste liquid tube
is pressed or damaged,and insert one end of the waste teflon tube to the waste liquid outlet,to ensure the waste teflon tube is
flowing,insert the other end of the waste teflon tube to waste system.

19
URIT-880/881/882 Service Manual

Picture 9 Remove waste liquid pipeline and waste liquid bottle

(3) Connect waste liquid pipeline in accordance with identification of the pipeline and waste liquid pipeline. As shown in
the following picture:

20
URIT-880/881/882 Service Manual

Picture 10 Road connection of waste liquid tube

Picture 11 Take the waste liquid tube into the waste systems

Notice
(1) waste systems can not be sealed,doing that could result in peristaltic can not absorb liquid.
(2) empty the waste often, and clean the waste systems. Avoid taking waste tube into the waste liquid
surface.

21
URIT-880/881/882 Service Manual
2). Installation of The Sample Pipeline
(1) Open the optical window cover on instrument, then tear the paper tape on alumina injector gently.

Picture 12 open the optical window cover,and take down the paper tape

(2) Take the alumina injector into the guiding steel tube.

Picture 13 install the alumina injector road


(3) Adjust the height of the alumina injector,and cover the optical window cover.

picture14 cover the optical window cover

21
URIT-880/881/882 Service Manual
3). Powerline Connection
(1) Before connecting with power line, be sure of the instrument is turned off,and connect the ground pole with hospital’s
ground wire.
(2) The connect one end of the power line with the instrument,the other end plugs in electric outlet with sound grounding.

(3)880/881:AC220V input;882:AC110V input。

Power Scoket

Waste Solution Interface

Grounded Pole

picture15 rear view of the instrument

Notice
(1) Please don't share the power outlet with other instruments.
(2) Before connecting the power wire,be sure of the instrument is turned off

4) . Printer Installation
(1) Please comply with these following matters when install the printing paper in built-in thermal printer.
 When installing the built-in thermal printer,please ensure the power is off before.
 Cover the printer’s cover after install printing paper.

Caution
Don’t damage any things with thermoelectric modules,like abrasive paper.Don’t press hard
on thermal printer’s head.
warning
The specification of thew thermal printing paper is 57.5±0.5mm

(2) Printer maintenance


 When the printer work abnormally, please turn off the power.
 When replace scroll,please pay attention to the instrument’s head,whether it attached with scraps of paper,if there
is, puff the it away off gently.
 Keep circuit of the printer clean and no dust.

22
URIT-880/881/882 Service Manual
Notice
To ensure the printer work properly,do not disassemble the printer head.

(3) Install printing paper

Please install the instrument’s head in accordance with the installation example on the printing tape,printing paper
installation as shown in picture 16。

Picture 16 install printing paper

 Open the cover of printer.


 Cut the paper as shown in picture 17.
 Put paper to paper mouth, and set apart proper length of the paper head, then press the printer head.

Picture 17 printing paper

5) .Turn on/off

(1) When operate the instrument for the first time,operators should check whether the waste liquid pipe is
bending,whether the liquid circulates smoothly,whether the waste liquid can be discharged into waste systems.
(2) Turn on the power switch,the instrument can initialize and self-check after instrument starts properly.

23
URIT-880/881/882 Service Manual

Picture 18 power-on self-test

(3) Log in software:choose user name,type user password,then click the “log”,main menu interface can be entered.

Picture 19 the interface of the URIT-880 log-in

24
URIT-880/881/882 Service Manual

Picture 20 URIT-880 main menu interface

(4) Preheating 30 mins after electrifying,make it reach thermal equilibrium and steady laser source.
(5) When the system stabilized,refer to operation manual,apply for 1~2 regular project to test, and evaluate test results.
(6) After confirming the instrument is in normal,turn off the instrument as the normal flow.

3.4 Precautions For Instrument Use

(1) Prevent dust and humidity, the room with air condition is much better,make the surrounding be in 15℃~25℃.
(2) Avoid electric leakage and static electricity. The instrument should be well grounded. If there is no ground wire in
power outlet,it should be grounded in the shell of the instrument.
(3) Ensure the instrument working in steady state, and security.
(4) The instrument using a power outlet alone is recommended.
(5) Use reagent with registration card to test instrument.
(6) After removing from refrigerator,reagent should recover to room temperature to do operating.
(7) In order to avoid the register numbers(ID number) and test results being chaos, different samples should have
different register number(ID number).
(8) After turning on, operate “wash” firstly, then do normal check in half an hour.
(9) After finishing work,,use cleaner fluid clean instrument 5 times,then use distilled water clean the instrument 10~20
times.
(10) The cuvette and pipeline must be filled with the distilled water before turning off.

25
URIT-880/881/882 Service Manual

3.5 Deposit

(1) The instrument can be deposit in the room with 0℃ ~ 45℃,the atmospheric pressure 75kPa ~ 106kPa ,
relative humidity≤95 % , good ventilation after packing,it can’t be deposit with toxic or harmful or erosive
substances
(2) When the instrument not in use,make the cuvette and pipe line be full of distilled water to avoid dust disturb the
testing result。
(3) When reuse the instrument after depositing more than one year,expected accuracy may not be achieved.In order to
ensure test result’s accuracy,recheck the instrument is recommended.

4. Debugging of The Instrument

4.1 Debugging Tool and Reagents

Debugging tool
1 4 1/2 multimeter 1
2 Testo 110 temperature instrument 1
3 calculator 1
4 A head-shaped/cross-followed screwdriver One each
5 No. 4 chave em cruz 1
6 Hexagon wrench 1
7 Tweezers 1
8 Nipper ply 1
9 Wire cutters 1
10 Parallel debugging fixture 1
11 Special tools for Adjusting optical route 1
12 Waste cap 1
13 Waste liquid tube 1
14 Guard tool 1
Debugging reagent
1 Filter the ionised water several
2 Potassium Bichromate Titrant several

4.2 Appearance and Safety Inspection

a. Surface inspection:check whether the graphic symbols on instrument is accurate,clear,even and no obvious nick.
26
URIT-880/881/882 Service Manual
b. Safety test:test by digital multimeter
c. Test the resistance between ground rod of the whole machine and power reserve,requirement≤0.1Ω.
d. Test the resistance between ground rod of the whole machine and metal components sharing on
shell,requirement≤0.1Ω.
e. No anomalies like wire shorted out or wrong connection.

4.3 Power on Self-test

(1) After turning on, the display screen displays normally,the indicator of the circuit board lights normally,the key is
sensitive.
(2) Power output range:V1:5.0±0.2V ; V2:12±0.4V ; V3:-5.0±0.2V。
(3) When cooling fan,heating tube and peristaltic pump work normally,it can lock water stably.
(4) Black level control is normal.Halogen gives out light normally,sleeping mode in half-voltage is normal,

4.4 Parameter Settings

4.4.1 Time Setting


Enter to“setup”interface,click“time and date settings”to check and adjust the current time.
4.4.2 Temperature Regulation
Cuvette、incubation groove’s temperature regulation
(1) Turn on the instrument half an hour,when the temperature is stable,which means it can be calibrated the temperature.
Click “setup” in the main menu,enter the setup interface;as shown in picture 21.

picture21 setup interface

27
URIT-880/881/882 Service Manual

(2) Click“measurement relevant”,then enter to measurement relevant interface. As shown in picture 22

Picture 22 measurement relevant interface

(3) Inject 5ml distilled water into temperature measurement fixture,and insert the fixture into the cuvette. Then insert the
thermodetector’s probe into the fixture. Wait 10 mins to confirm the temperature on the thermodetector is stable.

temperature measurement fixture temperature probe

Picture 23 temperature debugging tool

(4) click“thermostat setup”,enter to thermostat setup interface. As shown in picture 24.

28
URIT-880/881/882 Service Manual

Picture 24 thermostat setup interface

(5) Adjust the calibration factor of the electric (cuvette) based on the actual situation,the temperature the software
displayed should be the same with that the Testo 110 displayed,both are 37°C. When the temperature is stable for some
time,observe whether the temperature reached 37°C,and fine-tune calibrate the calibration factor of electric (cuvette)
continuously.

Notice:calibration factor calculation method:the temperature measured value-temperature value the display screen
displayed+instrument calibration factor display value.

29
URIT-880/881/882 Service Manual

Incubation tank

Picture 25 cuvette thermoregulation

(6) After regulating the cuvette’s temperature,the incubation groove’s temperature also needs to regulate. The method is
the same with cuvette’s thermoregulation,regulate the calibration factor of the electric(incubation device) based on the
actual condition.

Incubation tank

Picture 26 incubation groove thermoregulation


4.4.3 Pipeline Regulation
When the different pipelines are installed properly,turn the pump head by hands to let the alumina injector absorb the
distilled water,observe whether there is weeping or flowing in pipeline;sway the liquid outlet by hands,liquid waving in

30
URIT-880/881/882 Service Manual
alumina injector is not allowed.
Requirement:
(1) Phenomenon of weeping,liquid back is not allowed when the peristaltic is moving ;
(2) Stuck or lose step is not allowed;
(3) Imbibition inaccuracy is not allowed.

fluid discharge tube


cuvette
alumina injector

Picture 27 instrument’s internal structure

Notice
(1)The joints of the alumina injector、peristaltic tube and cuvette should be connected properly to avoid air
leaking or liquid weeping.
(2)After adjusting, empty the liquid in pipeline,make the instrument absorb liquid automatically,let the
liquid goes through the cuvette,and there is 5~8cm liquid column in liquid outlet of cuvette.

4.4.4 Liquid Road Parameters Inspection


Set the liquid road’s parameter, shown in picture 28. Step number/500μl:1100;solution default volume :500;
excitation frequency:3000.

31
URIT-880/881/882 Service Manual

Picture 28 liquid road parameter setting

Notice
After adjusting liquid road ,use pipettor to measure 500ul distilled water,check whether the absorb fluid
amount is 500ul

4.4.5 Optical Path Adjusting


4.4.5.1 halogen lamp Adjusting
(1) Click “setup” of the main menu,choose “test relation”,enter “A/D calibration edition” interface,click imbibition,
inhalating the distilled water.

32
URIT-880/881/882 Service Manual
Picture 29 A/D value calibration credit interface

(2) Adjust the set screw of lamp holder 、set screw of halogen lamp,make A/D signal be the best,and make the A/D value

of wavelength ranged 300nm~800nm ranges from 2000 to 4000.Then tighten the screw, fix the lamp holder and halogen
lamp tightly.

Screw holder

Lamp fixing screw

Picture 30 halogen lamp adjustment


4.4.5.2 A/D Calibration
Enter to A/D calibration interface,click“credit”,it can calibrate A/D value。After absorbing distilled water,
click“measurement”,A/D value can self correct。When A/D value correction sis stable,click “application” to save, then click
“back”.(notice:the A/D value of length 300nm~800nm ranges from 2000 to 4000)

Picture 31 A/D calibration interface


4.4.6 Optical path parameters Inspection
33
URIT-880/881/882 Service Manual
(1) Click “black level setup” in “measurement relation” interface, enter to “black level setup” interface. Check the
“operational amplifier forbidden” value when the programmed gain of the length 340 is “1”.
(2)Use the page-turning buttons on the right of the interface, check “operational amplifier forbidden” value respectively
when the programmed gain of other lengths are “1”, then note them.
(3)The requirement of related value’s range: when the programmed gain of the lengths is “1”, the range of “operational
amplifier forbidden” should be between 5 to 10.

Picture 32 black level setup interface

Notice:
(1) When calibrate A/D value,if the A/D value of the wavelength doesn’t meet the requirement,please wash cuvette

repeatedly,and check whether the cuvette has air bubble or dirt.


(2) If the operational amplifier forbidden value doesn’t meet the requirement, calibrate the A/D value repeatedly is
permitted, then check.

4.4.7 Peripherals Inspection


Connect related devices,check the peripheral interfaces of the instrument respectively:1 SD interface,1 USB interface
1,2 USB interface 2,1 PS/2 interface,1 parallel interface,1 RJ-45interface etc. Whether they can perform normally.

4.5 Performance Test

4.5.1 Imbibition Precision Test


(1) Prepare a pipettor(volume 1000μL),a test tube(volume10mL),some purified water.
(2) Adjust the absorption of the pipettor to 500μL, the fill 500μL purified water to test tube.
(3) According to the second point operation manual of “liquid road examination” to do imbibition operation, press

34
URIT-880/881/882 Service Manual
“imbibition” key,observe the purified water’s inhalation status in the test tube.
(4) If there is surplus of purified water in test tube,that means the absorption amount of the instrument is less than 500μl,
and it needs liquid-moving machine to measure the volume of the surplus of purified water. Method:regard 10μl of
the liquid-moving machine as the grade,and absorb the surplus of the liquid,finally calculate the volume of the liquid
being absorbed.
(5) If the purified water in test tube is stuck out,and there is some blank in alumina injector,that means the absorption
amount of the instrument is more than 500μl.At the time,increase by 10μl by degree,do the absorb liquid test
again,finally, when there is no blank in the alumina injector,record its absorption amount.
(6) The accuracy of the absorption amount :500±30μL.

4.5.2 Absorbance Repeatability Test


(1) cuvette absorbs distilled water,adjust system A/D value.
(2) Enter to absorbance measurement interface,absorb the distilled water again and read cup blank.
(3) Put the potassium dichromate standard solution into alumina injector,press the sample introduction key,and the
potassium dichromate standard solution is absorbed to test. The test result can be recorded automatically,repeat that 20
times,and record the 20-time test data.
(4) Calculate according to relevant formula,the repeatability should be no more than 0.5%.

4.5.3 Absorbance Stability Test


(1) Enter to absorbance test interface,absorb distilled water to test.
(2) Observe the instrument’s absorbance value every 5 mins within 20 mins, and record it.
(3) The absolute value’s maximum should require:the variation of the distilled water’s absorbance should be no more
than 0.005.
(4) If it can’t meet the technical requirement,please check the following order:
a.Zero adjustment again and recheck.
b.Check the light source,whether the A/D value of the signal board and main board is stable.
c.Check whether the temperature in cuvette is stable.
d.Check whether the liquid is qualified.
e.Check whether the cuvette has air bubble.

5. Set Up

5.1 Measurement related Set Up

Measurement relevant includes:A/D calibration、black level setup、light path setup、liquid road system setup、
thermostat setup and other else.
5.1.1 A/D Calibration

35
URIT-880/881/882 Service Manual
Every instrument has passed A/D calibration before factory examination. Different wavelength filter’s water blank A/D
values are different.In general, water blank A/D initial value ranges from 2000 to 4000. Recalibration is needed when the
following situations happen:
(1) instrument installation。
(2) Use after long-term transport.
(3) After replacing new halogen lamp。
(4) After replacing new cuvette。
(5) Long-term usage may cause the filter aging.
(6) Abnormal fluctuations appear when project is testing.
Click"A/D adjustment"—>“credit”,then click "imbibition" to make the cuvette full of distilled water,click
"measurement",system will adjust water blank A/D value of different wavelength.when the adjustment is finished,click
"application",system will save the adjustment result,and the result will be regarded as reference value in future test.
5.1.2 Black Level Setup
When the system boots, it will measure different wavelength’s programmed gain, operational amplifier start using and
the black level value in shut case. Users usually do not need to setup this interface.
When the programmed gains of different wavelengths are 1 and the black level in operational amplifier forbidden
should range from 5~10,if out of range,please refer to the fifth point about “A/D adjustment can’t pass”in List 8-1 common
fault diagnosis and processing.
5.1.3 Optical Path Setup
Optical Path setup interface includes lamp setup and analog to digital converter constraint setup.
(1) Stable time:light source is stable in required time,unit is second. Uses can set according to the actual situation.
(2) Sleep delayed time:when the idle state is beyond a certain time,the light source lamp will be in sleepy state. Uses can
activate the light source lamp through gentle-touch operation interface.
(3) ADC constraint :when read the A/D value,system will guarantee the signal value between “low crisis”and “high
crisis”,the system will pass the programmed gain and switch operation amplifier automatically.If the signal value is
lower than “low crisis”,or high than the “high crisis”,the system will returns to error.Modifying default is usually not
needed.
5.1.4 Liquid Road System Setup
(1) The step number/500μl:to calibrate the absorption amount accuracy.
(2) The default absorption amount:click the absorption of “wash”in measurement interface.
(3) Magnetizing frequency:users usually do not need to modify.
(4) “default”key:restore the factory default settings

36
URIT-880/881/882 Service Manual

Picture 33 liquid road system setting


5.1.5 The Thermostat setup
(1) Thermostat setting includes cuvette setting and incubation device setting;interface includes Proportion Integration
Differentiation setting、electrical characters setting and miscellaneous setting(other settings).
(2) Proportion Integration Differentiation
proportionality coefficient:in order to shorten stabilization duration,when the surrounding temperature is low,properly
higher the parameter,vice versa.
Integration coefficient、differential coefficient、dead zone:there is no need to modify by users.
(3) Electric parameter ( usually no need of modification)
(4) Miscellaneous setting
Initial temperature:temperature-control target temperature which thermostat default after system starts.
stabilization periods:when the temperature deviates the set value,then enter to the waiting duration.This parameter
used to decrease the distortion of sway of temperature reduction.
precision:thermostat’s accuracy. tolerability:
thermostat’s deviation for acceptance “default”key:
restore factory settings.
5.1.6 Other Setup
When test new patient,there are three options includes”display action selection dialog ” 、 ”input patient’s
information”、”no action”.
(1) display action selection dialog:when choose this option,the dialog will pop up,user can choose “yes” to input patient’s
information;click “no” will enter to test directly;choose the “remember this option” at the same time with choosing
“yes” or “no”, the system will remember the user’s option,and it won’t hint next time.
(2) Input patient’s information:choose the option,input patient’s information every time before test.
(3) no action:choose the option,there is no need to input the patient’s information,enter to the test directly.
37
URIT-880/881/882 Service Manual

5.2 Printer Setup

Activate the printer to realize the function of printing laboratory sheet,the printer option
includes: thermal printer,parallel port printer
Touch“system setting”gently—>“printer setting”enter printer setup interface,choose the printer you want to
activate,click “application”.

5.3 Language Setup

The system provides two languages to be chose,Chinese and English.


Touch“system setting”gently—>“language setting”enter language setup interface,choose the language that is needed in
pull-down menu.

5.4 Time Setup

Set system’s current date and time is allowed. Correct system date and time won’t influence the current result
interface’s sample,and also won’t influence the quality control data.
Gently touch “system setting”—>“date and time setting” enter to time setting interface,the interface shows the current
time. Input year、month、day、o’clock、minute、second,click “application”,the system will run the new time.

5.5 Screen Setup

Touch“system setting”gently—>“display screen setting”enter display screen setting interface.


Through clicking or long-pressing “+”or “-” to adjust the backlight brightness of the screen,“+”means enhance the
backlight brightness,“-”means reduce the backlight brightness.
Click “start” key to setup touch sound.

5.6 Account Setup

Add and delete user,setup user type and modify users’ passwords are allowed by system.
Please click the user name or user type in certain area to modify user name or user type. Click “back”after
modifying,the system will save the present setting,after correction,click“back”,the system will save the current setting
automatically. If the power is suddenly off when correction,the current setting is invalid.
The following button will appear after choosing a certain account,user can choose to perform the needing operation:
【add】:add new users.
【delete】:delete the selected serial number of the user's password
【change passwords】:Modify the selected serial number of the user's password
【back】:back to previous menu.
38
URIT-880/881/882 Service Manual

Look!
User No.1 are not allowed to delete, only allowed to change passwords.

5.7 Authorization Management

The system possesses right management function,setting certain function authorization is allowed.Only users who
possess the function can use the function.
● Modify the privilege the function needed:manage the authorization management interface is needed.
● Add user account:add new users

● Modify user account:modifying user’s name、password and user type and so on are allowed
● Modify measurement configuration:modifying measurement’s parameter are allowed
● Operate measurement duty:testing by user is allowed
function key specification:
【Default】means restore the factory default administrative authorization settings.

5.8 Driver

The page shows all the equipment’s loading status.

5.9 Subject setup

The system provides a simple theme.Theme change will come into effect after restart.

5.10 Export and Import

User can do project measurement setting、quality control data、curve measurement setting、test result record and data
immigration of laboratory sheet、user account、administrative authorization setting through SD card、USB and other mobile
storage device.

The methods of importing and exporting operation are as follows:


(1) Insert SD card、U disk.
(2) Click “setup” in main menu, and choose the “export and import” in “setup” page,and enter the “export and import”
page.

39
URIT-880/881/882 Service Manual

Picture 34 setup interface

(3) Choose the data needing to export or import in “exporting and importing” interface.

store path
selection

Picture 35 importing and exporting interface

(4) Choose the program(export or import) you want to execute and storage path.
(5) Choose the item you want through shortcut key(select all or select none) or touching gently.
(6) Click “start”,the system will operate relevant item.

40
URIT-880/881/882 Service Manual

5.11 PC host Setup

Set the PC host.

5.12 LIS/HIS System Setup

Setting related to LIS/HIS communication.

5.13 Factory Reset

When all the parameter being credit is not satisfactory,exclude recredition,the system will provide a special function,
press”set”—>“ restore the factory default settings ”—>“reset”,the chemistry analyzer’s all data will be eliminated and
restore the factory default setting.

Warning
When restore the factory default settings,the data being stored will be deleted,please remeasure the calibration
coefficient and sample absorbance;or record all data before restoring the factory default,you should input the
data by hands after restoring the factory default.

5.14 Others

Set the loudspeaker play the key sound or not.

6. Systems of The Instrument

6.1 Sampling System

6.1.1 Function Introduction


URIT-880 semi-auto chemistry analyzer,the main function of the sample-absorb system is to realize the function of
adding sample and cleaning.1.absorb the reagent sample mixture in cuvette to react and check.2.absorb the distilled water、
alcohol、cleaner to clean the cuvette.
Sample-absorb system is composed by:alumina、liquid inlet guiding tube、”start”key, peristaltic pump component.
6.1.2 Sampling Tube
Tube’s outside diameter is about 1.5mm,length is about 180~200mm,transparent material,one end connects with the
silicone tube whose outside diameter is about 4mm,length is about 30mm.

41
URIT-880/881/882 Service Manual

Picture 36 Sampling tube


6.1.3 Suction Fluid Guide Tube
The sample-absorb tube thread to the sample guiding steel tube, which protect and fix the sample-absorb tube.

Sampling guide tube


Guide tube stator

Picture 37 Suction Fluid Guide Tube


6.1.4 “Start”key
“Start”key is composed of big key and switch component. It mainly be responsible for give out the add sample order.

42
URIT-880/881/882 Service Manual

Sample switch

key

Picture 38 “Start”key
6.1.5 Peristaltic Pump Components
Peristaltic pump component is composed of peristaltic electrical motor、peristaltic pump slide switch. When peristaltic
pump runs,the instrument can add sample or do cleaning.

peristaltic pump assembly

Picture 39 peristaltic pump assembly(Ⅰ)

43
URIT-880/881/882 Service Manual

motor

Picture 40 peristaltic pump assembly(Ⅱ)

6.2 Optical Path System

6.2.1 Function Introduction


This is one of the key system of the instrument, whose function influence the accuracy and precision of the instrument.
The main function of the optical path system is to provide light, split light, receive light signal and convent it to the electrical
signal. Optical path system mainly consists of: source light, light holder components, cuvette, light box components.

6.2.2 Source Light


Light source:halogen lamp,6V/10W

lamp

Lamp tube

Picture 41: source lamp


6.2.3 Light Holder Component

44
URIT-880/881/882 Service Manual

Picture 42 Light Holder Component


6.2.4 Cuvette

Inlet metal tube


Outlet metal tube

Picture 43 cuvette

6.2.5Light Box Component


Wavelength : 340nm,405nm,450nm,492nm,510nm,546nm,578nm,630nm,700nm,800nm
Wavelength accuracy:+2nm.
Linearity range:0-3.0Abs.
Resolution:0.0001Abs.

45
URIT-880/881/882 Service Manual

Picture 44 light box component

Light splier Light flier

Picture 45 light splier and light flier

6.3 Liquid Road System

6.3.1 Liquid Road Map


Sample-absorb tube→cuvette→fluid-discharge tube→peristaltic pump(peristaltic pump tube、peristaltic pump tube
waste system)→waste system→waste flask

46
URIT-880/881/882 Service Manual

sample-absorb tube
fluid-discharge tube

waste system

Picture 46 liquid road map


6.3.2 Fluid Road Line

fluid-discharge tube
peristaltic pump tube
waste system

Peristaltic pump
tube

Picture 47 pipeline
Description:
1. peristaltic pump pipe:locates at rotating axle

2. fluid-discharge tube:locates at cuvette’s outlet to peristaltic pump’s inlet

3. peristaltic pump tube waste system:locates sat peristaltic pump’s outlet to instrument’s waste outlet

Liquid road after-sale codes follow:


no. code name
1 XQB0800009 Sample Teflon tube/800/810/870/after sale
2 XQB0800015 cuvette/800/810/870/after sale
3 5551200353 fluid-discharge tube/880/881/882/after sale
4 5551200342 Peristaltic pump components/880/881/882/after sale
5 5551200352 Peristaltic pump tube/880/881/882/after sale
6 BCP0810011 810 waste system(with plug)/800/810/after sale

47
URIT-880/881/882 Service Manual

6.4 Hardware Circuit

6.4.1 URIT-880 Mainboard


Main board function:the instrument control the core,and control peristaltic electrical machine’s movement,receive and send
data and signal.

6.4.1.1 URIT-880 Mainboard Wiring Diagram

connecting line from temperature-sensi KEY big key fan power heating tube
main board to signal ng connecting line switching supply connecting
amplifier
shielded line

switch power
input line

motor display board 、


printer interface
board connecting
board
lamp

FAN1 fan
power line

Picture 48 URIT-880 main board connection diagram

remark:
FAN1 fan voltage:12.45±0.3V
FAN2 fan voltage:12.45±0.3V
6.4.1.2 URIT-880 Mainboard Test Point
(1) Lamp’s power voltage test point
48
URIT-880/881/882 Service Manual

halogen lamp
power voltage

6V power voltage
test point

Picture 49 main board test point(Ⅰ)


6V voltage testing point:5.98±0.02V
halogen lamp voltage :5.98±0.02V

(2) Signal board power supply and reference voltage test point

switch power
supply input -5V TP

+5V TP

C76Left

Picture 50 main board test point(Ⅱ)


-5V power voltage test point:-5±0.02V
+5Vpower voltage test point:+5±0.02V
C76 left hole test point:4.08V
Switching power supply input:blue:+12±0.4V,yellow:GND ;red:+5.0±0.2V,black:-5.0±0.2V
(3) Temperature signal and test point of the reference voltage

49
URIT-880/881/882 Service Manual

TP15 C77lower

C85lower

R77right C89lower

Picture 51 test point of the main board(Ⅲ)


TP15:2.98±0.02V
C77 down hole:3.45±0.02V
C85 down hole:3.45±0.02V
C89 left down hole:2.35±0.04V
R77 right hole:2.35±0.04V
TMP1 temperature-sensing voltage:0.114±0.005V
TMP2 temperature-sensing voltage:0.114±0.005V

6.4.2 Signal Amplified Board


Function of the signal amplified board:transform optical signal to electrical signal, then transfer the amplified signal to
main board.

50
URIT-880/881/882 Service Manual
Picture 52 signal amplified board
6.4.3 printer interface board
Function of printer interface board:transfer orders to the main board of built-in thermal printer

6.4.4 Display Interface Board

Picture 53 printer interface board


Function of display interface board:transfer data to the main board of the display interface board.

Picture 54 display interface board


6.4.5 Lamp Transfer Board
Function of lamp transfer board:transit main board ,and provide voltage for halogen lamp

51
URIT-880/881/882 Service Manual

Picture 55 lamp transfer board and lamp power line

6.5 SMPS

Mead Well RT-125A SMPS


100-120VAC 3.0A
INPUT
200-240VAC 2.0A
50/60Hz
V1 +5V 12A
OUTPUT V2 +12V 5.5A
V3 -5V 1.0A

52
URIT-880/881/882 Service Manual

Picture 56 switch power supply

6.6 Incubation Groove

6.6.1 The usage of The Incubation Groove


In order to facilitate customers temporarily store、incubate、transfer reaction liquid、preheat reaction liquid,URIT-880
has double row and 8 holes built-in thermostatic incubation groove(picture as shown in 3.7),and applies to test tube that
outside diameter is less than 13mm.
The temperature of Incubation groove is 37°C,when needing to use,insert the test tube containing reaction liquid and
reagent into incubation groove directly,when do test, take out the test tube from the incubation groove directly.

Picture 57 instrument aerial view

53
URIT-880/881/882 Service Manual

Notice
please don’ pour any clutter or liquid into the hole of incubation groove to avoid damage, which would
cause test result.

6.6.2 The Structure of The Incubation Groove

Picture 58 incubation groove aerial view


6.6.3 Heating System

Temperature-sensing

Heating tube

Picture 59 incubation groove broadside

54
URIT-880/881/882 Service Manual

Temperature switch

Picture 60 incubation groove bottom

heating tube

Picture 61 heating tube

55
URIT-880/881/882 Service Manual

Picture 62 circuit of heating tube and temperature switch

Explanation:
1.A heating pipe connects with a temperature switch.
2.The instrument has 2 heating pipe, one is for heating the cuvette, the other is for heating incubation groove.

7. Instrument Maintenance

7.1 Instrument Maintenance

7.1.1 Cuvette cleaning


The cuvette is infinitesimal stainless-steel-quartz window glass that is precise design.The cleanliness of the cuvette’s
cavity and the smooth of the flow system,which directly relates to the accuracy of the measurement result.
notice of usage of instrument:
(1) The new instrument’s cuvette should be washed by medical alcohol for a time before project test,then wash it with
distilled water for a time.
(2) In working process,if finish a certain project sample measurement,please wash the cuvette immediately.when the work
is finished,please wash it many time,then turn off the instrument.
(3) Wash the cuvette with neutral cleaning fluid,then wash it with distilled water at intervals.
(4) If the instrument leaves unused for many days,when reuse it,please wash it with distilled water for many times,before
project test,please wash the cuvette with medical alcohol for a time,the wash it with distilled water for a time.
(5) Avoid serum being absorbed by cuvette,which may cause cuvette or flow system blocked.

56
URIT-880/881/882 Service Manual
(6) Avoid alumina injector break-off.
(7) Pour the waste liquid often,avoid spilling to the worktable, which may corrode the instrument.
(8) Avoid the waste liquid pipe bending,keep the waste liquid pipe flow.
7.1.2 Printer Maintenance
(1) Maintenance of the printer’s head
A. If the printer works abnormally,please turn off the instrument’s power supply.

B.When replace the scroll,please notice the machine head whether it sticks with scraps of paper,if there is,blow it
gently.
C. Keep the circuit of the printer clean and no dust.

Notice
To ensure the printer work properly, do not disassemble the printer head

(2) Install print paper


The instrument has installed printing paper from the factory. Open the cover of the printer,install the printing paper
into machine head according to printing paper’s installing example. The paper installment process as following picture
61.

Picture 63 printing paper installment


Notes:
(1) The specification of thermal printing paper 57.5±0.5mm。
(2) Be ensure of the the power is off before installing the printing paper.
(3) After install printing paper,close the printing cover.
(4) Don’t use something may damage thermal component,like abrasive paper. Don’t press hard the thermal printing head.

7.1.3 Screen Protection


The screen of the instrument is touch screen,touch corresponding image button,the display screen will show
corresponding hint,the hands can get away.

57
URIT-880/881/882 Service Manual

Notice
Don’t use sharp scratch the display screen.

7.1.4 Big Button Switch Protection


Big button switch is plastic key,in order to avoid key damage,and influence service life,don’t be too hard when use it.

big button switch

Picture 64 big button switch

7.2 Maintenance

7.2.1 Alumina Injector Replacement


In the process of using, if the alumina injector is block,and can’t dredge,operator can replace the alumina which is
random distributed. Pipe’s external diameter is about 1.5mm,length is 200mm,transparent material,one end connects with
silicone pipe whose external diameter is about 4mm,length is about 30mm,replacement method is as follow:
(1) Drain the liquid road;

(2) Turn off the instrument and disconnect instrument’s power line;
(3) Open the left optical window cover,gently remove the pipe which plugged in cuvette’s liquid inlet mental pipe
(thinner one,near the front of the instrument,as shown in picture 63;
(4) Remove the pipeline from the liquid inlet guiding steel pipe,reinsert one end of the new liquid inlet guiding steel
pipe with silicone pipe to the cuvette’s liquid inlet mental pipe,then thread the liquid inlet pipe to liquid inlet guiding pipe
freely, then replacement is finished.

Notice
(1) Don’t bend the alumina injector when replace.
(2) To ensure the connection reliability,please insert the pipe to the button of the cuvette’s liquid inlet
mental pipe when do pipe connection. After finishing the pipe installment,crooked place should be transit
smoothly,defection such as sharp angle,wrinkle which may cause poor circulation can’t be allowed.

58
URIT-880/881/882 Service Manual

Picture 65 alumina injector installation

7.2.2 Peristaltic Pump-pipe Replacement


Biohazard
In working process,please wear gloves and working cloth,and wear protective glasses when necessary. If the
waste liquid gets the skin accidentally,please treat it according to user working standard,and consult doctor.
If there are phenomenon,such as no liquid absorbed,or the amount decrease obviously during use of the
instrument,observe whether the back of the peristaltic pump is leakage.(as shown in 64). If it’s leakage,is caused by the
worn-out peristaltic pump tube,and replace after ensure. The instrument will randomly distribute a peristaltic pump
tube(beige),outside diameter is about 3.2mm,length is about 115~125mm. The replace methods as follow:
(1) Drain the liquid in pipeline;

(2) Close the instrument and off the power line;

(3) Turn on the slide switch of peristaltic pump at the back of the instrument;

(4) Take out the peristaltic pump tube and right angle connector from card slot of the peristaltic pump;

(5) Remove the right angle connector from peristaltic pump tube;
(6) Insert one end of the peristaltic pump tube into a right angle connector,thread the other end to the peristaltic
pump pipeline install hole,the peristaltic pump tube should go round the peristaltic pump head and thread out from another
install hole,the connect with another right angle connector;
(7) After connecting the right angle,gently jerk the peristaltic pump tube to completely fit with peristaltic pump,and

there are no faults like distortion、wrinkle;keep the length of the peristaltic pump’s right peristaltic pump tube is similar to
the left peristaltic pump tube. Gently jerk two ends of the peristaltic pump tube to keep it smooth.
(8) Stuck slide switch of the peristaltic pump again.

59
URIT-880/881/882 Service Manual

Picture 66 peristaltic pump’s structure


Notice
The instrument equips spare peristaltic pump pipe,our company does
not accept responsibility for any consequences resulting from using the pipeline that is not our company
provided, which test inaccuracy or instrument failure may cause

Biohazard
When the peristaltic pump pipe contacts with test samples,reagent,standard solution and other dangerous
goods directly, please wear glove.If accidentally touched body, wash it with water immediately and
disinfect,consult with doctor when necessary.
7.2.3 Fuse Protector Replacement
When the fuse protector burnout, take out the fuse protector’s cover,and replace time-delay fuse protector(T1AL
250V)

Picture 67 the placement of the filtering power fuse protector box

60
URIT-880/881/882 Service Manual

Picture 68 filtering power supply structure

Notice
(1) Please turn off the power switch and pull the power line.
(2) Please do not damage the cover of the fuse protector and power interface when replacing the fuse
protector.

7.2.4 Cuvette Replacement


After long time using the reacting cup,it may appear scratches or crack,and it may be so polluted that it’s impossible to
clean,all of these may influence the transmittance and uniformity,which may influence the accuracy and stability of the
results. It’s time to replace cuvette.
The cuvette is made of stainless steel quartz window,sophisticate design,its inside can keep thermal equilibrium,the
temperature is controlled by heating tube component.The cuvette installment has orientation requirement,the inlet is thin,
the outlet is thick;open the “light road cover” when replace the instrument,and empty the liquid in cuvette and liquid
road,press the cuvette with one hand,then remove the alumina injector and peristaltic pump tube,then loose the cuvette’s
fixed screw,take out the cuvette from the cuvette hole. Put the cuvette into the cuvette hole,the fix the fixed screw,finally
insert corresponding pipeline into the outlet when do installment.

Notice
(1) Be sure the cuvette well fixed when install the cuvette,or it may influence the measurement result.
(2) To ensure he connection reliability,please insert the pipe to the button of the cuvette’s liquid inlet mental
pipe when do pipe connection. After finishing the pipe installment,crooked place should be transit
smoothly,defection such as sharp angle,wrinkle which may cause poor circulation can’t be allowed.

61
URIT-880/881/882 Service Manual

图 69 cuvette

Picture 70 light path structure

Biohazard
When the cuvette contacts with test samples,reagent,standard solution and other dangerous goods
directly,please wear glove.If accidentally touched body, wash it with water immediately and disinfect,
consult with doctor when necessary.

Warning
(1) To ensure the cleaning of cuvette, do not touch quartz window of the cuvette.
(2) When test finished,user must clean cuvette: wash the cuvette with distilled water, no sooner the cuvette
filled with distilled water than close system
(3) Clean the cuvette with cleaner fluid at least one time every week, absorb 2ml cleaner fluid, and save liquid
in cuvette for 5 minutes,then absorb 20ml distilled water to wash.

62
URIT-880/881/882 Service Manual
7.2.5 Halogen Lamp Replacement
Because of the filament position of the light source lamp has a great impact on the light path of the optical
performance,if the light source lamp damages or expires the rated using life(2000hours), it’s necessary to replace a new one.
The detailed operations as follows:

(1) Open the optical window cover,use the phillips screwdriver unscrew the screw on window cover,remove the window
cover of the replace-lamp.

screw these 3
screws,then unpack
lamp-replacement
window.

Picture 71 open the lamp window cover

(2) Use the phillips screwdriver unscrew lamp line plugging seat,and remove the lamp line,then unscrew the fixed

screw on lamp holder, remove it and put it aside a proper place,then hold the guide line of the lamp seat, and unscrew the
light source lamp assembly gently.

63
URIT-880/881/882 Service Manual

Picture 72 replace halogen lamp

(3) Reconnect source lamp component’s wire connector with instrument’s internal wire connector,install source lamp
on the installing hole of the source lamp holder carefully.
(4) Start up the instrument,enter the A/D value calibration interface,after absorbing the distilled water,adjusting the

height of the halogen lamp to make the A/D signal reach to the maximum,and the A/D value of 300nm~800nm ranges
2000~4000,tightening the halogen lamp to fix the screw(refer to light path adjustment chapter for detailed steps). After
installment,cover the light window cover and optical window cover;
(5) When installment finished,start the instrument with power.Do gain adjustment,when the instrument is stable, test
the background and test whether the light source lamp is replaced successfully.

Notice
(1) When replace the halogen lamp,please turn off the power of the instrument.
(2) The surface of the lamp should be clean when replacing the lamp.if the surface is dirty,please use
lens cleaning paper to clean.
(3) Do A/D calibration after replacing light source lamp.
(4) Be careful when remove the fixed screw to avoid component dropping into internal instrument.
(5) Before replacing the light source lamp, please turn off the power,and wait at least 15 mins till the light
source cooling.Before light source cooling,in order to avoid scald,don’t touch it.

7.3 Clean & Maintenance of The Instrument

7.3.1 Daily Maintenance


(1) Using the neutral cleaner and wet cloth to wipe the liquid residue in the instrument.

64
URIT-880/881/882 Service Manual
(2) Do 5~10 times pipeline washing program after turning on,do daily cleansing with distilled water
(3) Execute pipeline-washing program 10~20 times,do daily cleaning to pipeline with distilled water.
(4) When not in use,ensure the pipeline system and cuvette should be full of cleaning distilled water.
(5) empty the waste systems.

7.3.2 Weekly Maintenance


Biohazard
please wear gloves and work clothes to ward off infection,or wear protective glasses if necessary when
operating.if contacted with skin,please deal with it in accordance with user’s work standards and consult
doctor.
(1) Use the distilled water to clean the inner wall of the waste systems,or use disinfectant to soak the waste system
when necessary.
(2) Use the cleaner with activating enzymes to soak the cuvette and the alumina injector’s head,after soaking for
some time(if the cleaner with high corrosion, don’t soak too long),and use the distilled water to wash many times.The soak
of the cuvette can be done through “wash” the cleaner being absorbed.
(3) Use percutaneous ethanol to clean 10~15times.
(4) Be ensure the instrument’s power is off, and wipe the instrument’s panels with clean cloth,or wipe it with cloth
dipping clean water or disinfectant when necessary.

Notice
(1) Be careful to operate, and avoid scratch the chassis.
(2) When the cuvette contacts with test samples、reagent、standard solution directly,please wear glove.If
accidentally touched body,wash it with water immediately and disinfect,consult with doctor when necessary.
(3) Don’t free throw the cloth having wiped the instrument’s panel,please properly handle it according to
relevant regulations.

7.3.3 Monthly Maintenance


(1) Maintenance procedure is the same as weekly maintenance.
(2) Check whether the printing paper is enough
(3) Clean the dust in inner instrument with soft brush.

(4) Check the whether the cuvette has bubble、foreign bodies.

Notice
If long time no use,please use protecting bag cover instrument to avoid dust

7.3.4 Unscheduled Maintenance


(1) Cuvette Cleaning

 Background is lower
When the cuvette is polluted or there is air bubble in cuvette, it may cause the background low.,clean the cuvette with
alcoholic solution

65
URIT-880/881/882 Service Manual
(1) Click “wash” key,do pipeline empty wash;

(2) Prepare 5mL alcoholic solution with tube.(alcohol:water=4:1)


(3) Insert the alumina injector into alcoholic solution,press”wash”key,the instrument will start to absorb alcohol.then
wait for 10s,use distilled water wash pipeline 10s~20s,which ends cuvette washing.

 Before doing transform test project


When replacing the project testing or after testing the high average samples,we suggest using distilled water to wash
the cuvette,and the wash time should be no less than 5s every time,or wash it with distilled water after washing with special
cleaner of chemistry analyzer.

 External cleaning of cuvette


When the optical surface of the cuveette is contaminated,users can use cotton fabric dipping percutaneous ethanol to
wipe the optical surface
(1)The instrument can be deposit in the room with -20℃~55℃, relative humidity≤95%, good ventilation after packing,it
can’t be deposit with toxic or harmful or erosive substances
(2) When the instrument is free,cuvette and pipeline should be full of distilled water to avoid dust coming in the
cuvette,which may influence the test result’s accuracy.
(3) The instrument’s storage life is 15 years,reuse it after storing over a year may be not live up to expected measurement
accuracy,to ensure the test results are accuracy,we suggest rechecking the instrument before using.

7.4 Transportation and Storage

7.4.1 Transportation
(1) The transportation of the instrument should accord with the contract and do not mix with poisonous, harmful or
corrosive substances
(2) Transportation environment:-40℃~55℃;RH≤95%
7.4.2 Storage
(1) Well packed instrument allows a drafty place where the temperateness is around -20℃~55℃ and RH≤95%, but

not mixed with poisonous, harmful or corrosive substances


(2) if no performance,cuvette and pipelines should be full of distilled water to avoid dust influence the accuracy of the
instrument.
(3) the storage life of the instrument is 15 years, if there is a long time no use, it should get a test before put it into use.

66
URIT-880/881/882 Service Manual

8. Maintenance Skill

8.1 Introduction

This chapter is about methods of failure search and elimination,if following the methods still can’t eliminate the fault
or more information is needed,please contact with after-sale department of URIT.
This chapter provides some faults phenomenon and handling methods,when the instrument goes wrong,operator can
find out faults' reasons depending on alarm message,and maintain it according to realities of situation.These faults handling
methods the chapter provided follows the principle of part replacement,please check the attachment 1 for main parts
needing to replace.

8.2 General Failures and Solutions

Warning
When doing fault handling,please be careful as much as possible to ensure whether the power is off when
operating again.

Biohazard
When operating liquid road,please wear gloves and work clothes to ward off infection,or wear protective
glasses if necessary.

8.2.1 Troubleshooting
Troubleshooting designed for specially assisting user to check、exclude the fault during instrument operation,the guide
also provides user the methods how to get technical support and help from after-sale service department of URIT. Faulty-
check and fault-exclude skill is based on the in-depth knowledge about the instrument and the cumulative experience in
using process,the more user know,the troubleshooting is easier.
If operator wants to do trouble-find 、 trouble-shoot exactly and immediately, first,he should read through instruction
book and know normal operation and preventative maintenance about instrument.
In general,troubleshooting should follow these three steps:
(1) Fault verification
(2) Fault classification
(3) Fault dropping
Step 1:fault verification
Operators can not only be sure of what the fault is, but also know what the right is in normal condition. Only know the
fault correctly can removal the fault.
Step 2:fault classification
Fault classification of the instrument are mainly these three types:
(1) Fault relates to hardware.

67
URIT-880/881/882 Service Manual
(2) Fault relates to software.
(3) Test fault relates to sample analysis
Fault relates to hardware、software only can be maintained by URIT or corresponding qualification engineer authorized
by URIT, fault relates to sample analysis test can be figured out on operator’s own under the guidance of the URIT’s
engineer.
Step 3:fault dropping
Maintenance engineer adopt proper methods、step troubleshoot.If the operator can figure out the faults own their own
or figure out the fault under the guidance of URIT’s engineer, the deferred time will be decreased greatly.
8.2.2 Technical Support
If the instrument breaks down and needs URIT’s technical support,please contact with the after-sales department of
URIT. Users for technical support must provide the detailed matter description and related information,the details as
follows:
(1) instrument’s model number
(2) instrument’s serial number
(3) describe the faults phenomenon and operation environment in detail and clearly (eg:operate in what window or in
what state)
(4) the data and reports relates to fault
There are some common faults and treatment methods in this chapter,when the instrument has faults,operator can find
out faults’ reasons according to the warning from the instrument,and follow the troubleshooting methods and steps in the list
to operate.
8.2.3 Processing Method of Common Failures
There are some common faults and treatment methods in list 1,if users can’t troubleshooting after following the
methods or need URIT’s technical support,please contact with URIT’s after-sale department.

Table 1 Common fault diagnosis and processing

No. Phenomenon Possible cause Processing method


(1) connect the power circuit wrongly (1) connect power circuit rightly
Power switch is not on
1 (2) no electric current in power socket (2) replace fuse protector(T1AL
work
(3) fuse protector melt 250V)
Display is disorder when After shutting off for 1~2 mins,plug
2 memory may be abnormal
the power is on in power again
(1) unplug and plug the main
The fan not run,the bulb (1) the voltage of main board bad contact
board’s power line
3 not light,when the power (2) the main board is damaged
(2) replace the damaged main board
is on (3) the fuse protector is damaged
(3) replace fuse protector

Fan runs but the bulb is


4 the bulb burnt out Replace halogen lamp
not light,when the power

68
URIT-880/881/882 Service Manual
is on
(1) the pipeline leak (1) check the leak area,then
No water running when
5 (2) the peristaltic pump is aging or leak reconnect the pipeline.
washing
caused by breakage (2) replace the peristaltic pump tube

(1)o-line control cable kerosene stuck


(1) stuck the on-line cable tightly
Computer failing on-line tightly
6 (2) choose the right communication
with instrument (2)the communication port for the
port for the instrument
instrument is wrong
(1) big key failure
(2) peristaltic pump doesn’t install properly (1) replace big key switch
(3) peristaltic pump tube is aging and (2) reinstall peristaltic pump
Press big key and no
7 inelastic or peristaltic pump tube rupture and (3) replace peristaltic pump
liquid was absorbed leaking (4) reset the motion parameter of
(4) peristaltic pump’s motion parameter the peristaltic pump
setting is improper

Operate the
(1) on-line control cable is too long
instrument,but not (1) replace the specific cable
(2) on-line control cable connection is bad
8 execute other (2) fix the cable
(3) instrument’s accessory is interfered by
functions(computer link (3) move away high power device
strong electromagnetic interference
status)
Turn off power supply,restart the
Crash caused by the external network’s
9 Key failures instrument after waiting for several
disturb
minutes
(1) printer is damaged
(1) check printer’s connection
10 printer not print (2) printer and instrument are loose
(2) replace printer
connection
(1) small bubble in cuvette (1) reset the peristaltic pump’s
(2) peristaltic pump’s silicone aging or motion parameter
break make sample-absorb inaccuracy (2) replace peristaltic pump’s

(3) fluid-absorb is abnormal(amount is silicone tube


small or doesn’t absorb) (3) wash with medical alcohol for a
Result inaccuracy or
11 (4) alumina injector、waste system breakage time,then wash the cuvette with
repeatability is bad distilled water
(5) alumina injector or flow cuvette block
(6) light source lamp is aging (4) replace cuvette

(7) biochemistry test parameter setting is (5) replace light source lamp

improper (6) reset the biochemistry checking

(8) power with no ground wires parameter according to reagent

69
URIT-880/881/882 Service Manual
specification
(7)connect the instrument’s ground
pole with earth.
(1) Replace light source lamp
(1) lamp is burnt out
(2) Rewash the cuvette,and fill
(2) cuvette with no enough water or the
with distilled water
12 A/D value can’t be passed cuvette with little bulbs
(3) Replace cuvette
(3) cuvette is dirty
(4) Correct the light path setting
(4) the setup parameter light pathis wrong
parameter

(1) display screen is damaged (1) Change display screen

display screen fails to (2) display screen makes poor contact with (2) Extract or change the power
13
work display board or damaged. strip
(3) display board is damaged (3) Change display board

8.3 Notes

To ensure operation safety and the long-term stabilization of the device, please read through the manual and take full
acknowledge of the function and operation and maintenance knowledge before operating.
Mal-operation or users do not comply with the manual manufacturer or agent provided, which may cause the
instrument damage or personal injures.
The following conventions run through the whole operation manual to indicate the information is particular emphasis.

Symbol Implication
Remind the operator operate in accordance with the instructions, doing so may
result in loss of life or serious injury. serious injury includes blindness、trauma、
Warning thermal injury(temperature is too high)、cold injury(temperature is too low)、
electric shock、fracture、intoxication and other complication or need to spend
night in hospital for observation or long-term extramural hospital treatment.
Remind the operator operate in accordance with the instructions,doing so may
Caution
result in system damage or distort the result.
Remind the operator operate in accordance with the instructions,in cas e of
personal injure、physical trauma、may distort the result. It also hint the source of
Notice infection.Personal injury includes minor burn which doesn’t need to
hospitalization caused by electric shock or drug allergy.physical includes
damage caused by building and other contents、domesticated animals and pets.

70
URIT-880/881/882 Service Manual
Remind the operator operate in accordance with the instructions,doing so may
Biohazard result in biohazard.biohazard is a kind of hazard,which is caused by biotic
factors damaging to environment and organism’s health.

71
URIT-880/881/882 Service Manual

Appendix 1:Replaced Spare Parts List

No. Name Remark


1 Fuse protector T1AL 250V
2 Switch power
3 Rocker switch
4 Power filter
5 Stepping motor(peristaltic pump) With plug
6 Power line
7 Main board
8 Peristaltic pump line
9 Peristaltic pump
10 cuvette
11 Ventilating fan
12 Halogen lamp 6V 10W
13 Temperature sensor With plug
14 Light filter 340nm~700nm
15 Signal board
16 Print interface board
17 Display interface board
18 on-line control cable
19 Display screen

Warning
(1) Peristaltic pump line, fuse protector, cuvette can be replaced by users, please replace in accordance
with instruction book. If unsuitable operation makes instrument damage, URIT won’t be responsible
for the instrument’s security, reliability, movement condition any more.
(2) What the assembly needs to replace should be provided by URIT, please contact with URT to get
technical support.

72
URIT-880/881/882 Service Manual

Appendix2: Order and Abbreviations of Test Items

NO. Item NO. Item

1 ALT 19 K

2 r-GT 20 Cl

3 ALP 21 Ca

4 CHE 22 Fe

5 D-Bil 23 P

6 T-Bil 24 Mg

7 ALB 25 AST

8 TP 26 LDH-L

9 UREA 27 CK

10 BUN 28 CK-MB

11 Cr 29 α-HBDH

12 UA 30 LD-1

13 GLU 31 PK

14 CHO 32 Ams (blood)

15 LDL-C 33 Ams (urine)

16 HDL-C 34 Hb

17 TG 35 CO2

18 Na 36 ACP

73