Documente Academic
Documente Profesional
Documente Cultură
HART Transmitter
Calibration
Application Note
Introduction
In today’s process plants, most
new field instruments are smart
digital instruments. Smart im-
plies a microprocessor-based
instrument with extra functional-
ity and digital compensation,
supporting multiple sensor types
or multiple variables. These in-
struments generally offer better
accuracy, long-term stability,
and reliability than conventional
analog instruments.
The most common class of
smart instruments incorporates
the HART protocol, with more
than five million HART instru-
ments in use in 100,000 plants
worldwide. HART, an acronym
for Highway Addressable Remote
Transducer, is an industry stan-
dard that defines the communi-
cations protocol between smart
field devices and a control sys-
tem that employs traditional
4-20 mA wiring.
Two capabilities are required
to properly service HART instru-
ments: precision analog source
and measure capability and digi-
tal communication capability.
Until recently, this required two
separate tools, a calibrator and a
communicator. Today, the capa-
bilities of those two tools are
available in a single HART Docu-
menting Process Calibrator that
can help you quickly and effec-
tively service HART instruments.
HART calibration Regular calibration is also pru- How are HART
is required! dent since performance checks instruments properly
will often uncover problems not calibrated?
A common misconception is directly caused by the instru-
that the accuracy and stability mentation, such as solidified or To calibrate a HART instrument
of HART instruments eliminate congealed pressure lines, in- consistent with its application, it
the need for calibration. stallation of an incorrect ther- is very helpful to understand the
Another misconception is that mocouple type, or other errors functional structure of a typical
calibration can be accomplished and faults. HART transmitter. The article in
by re-ranging field instruments A calibration procedure con- Appendix A, by Kenneth L.
using only a HART communica- sists of a verification (As Found) Holladay of Southwest Research
tor. Still another misconception test, adjustment to within ac- Institute, describes a typical
is that the control system can ceptable tolerance if necessary, HART instrument and defines
remotely calibrate smart instru- and a final verification (As Left) both proper and improper cali-
ments. These are not true. All test if an adjustment has been bration practices. Originally
instruments drift. Re-ranging made. Data from the calibration published in Intech, May 1996, it
with just a communicator is not are collected and used to com- is reprinted with permission of
calibration. A precision calibra- plete a report of calibration, the author.
tor or standard is required. documenting instrument per-
Regular performance verifica- Note: If you are unfamiliar with HART
formance over time. calibration or need a review, this is an
tion with a calibrator traceable All instruments, even HART excellent point to stop and read the article
to national standards is neces- instruments, must be calibrated in Appendix A. It covers the basics of HART
sary due to: on a regular, preventive main- instrumentation and addresses issues critical
to instrument maintenance.
1. Shifts in performance of tenance schedule. The calibra-
electronic instruments over tion interval should be set short HART instruments consist of
time, due to exposure of the enough to insure that an in- three distinct sections (see
electronics and the primary strument never drifts out of Figure 1). Proper HART calibra-
sensing element to tempera- tolerance, yet long enough to tion may involve either or both
ture, humidity, pollutants, avoid unnecessary calibrations. sensor trim and output trim.
vibration, and other field Alternatively, the interval may Adjusting range values (LRV and
environmental factors. be determined by critical pro- URV) without a calibrator is not
2. Regulations governing occu- cess requirements, e.g., calibra- calibration. Performing an output
pational safety, consumer tion before each batch. trim while ignoring the input
safety, and environmental section is not proper calibration.
protection. Adjusting range values with a
3. Quality programs such as calibrator may be a practical
ISO 9000 standards for all calibration alternative for instru-
instruments that impact ments operated in 4-20 mA
product quality. analog mode, provided that the
PV and PVAO are not used for
4. Commercial requirements process control.
such as weights, measures,
and custody transfer.
Figure 1
PVAO
(Digital representation PV LRV
of the Primary Variable (Primary Variable
Analog Output) Lower Range Value)
PV URV
(Primary Variable
Upper Range Value)
Figure 2
New tool speeds gether to complete field calibra- a single key enters the HART
calibration tions. However, the desire to mode and displays the essential
carry less equipment and to HART information in the Active
Today, instrument maintenance perform maintenance in the Device Screen, shown in Figure
is moving out of the shop and field has created a need for a 2. Additional HART functionality
into the field. This reduces pro- new class of calibration tool. is accessed with only a few
cess interruptions and avoids The new 744 Documenting more keystrokes, per the menu
the time and expense of return- Process Calibrator from Fluke is tree in Figure 3.
ing instruments to the shop. the first powerful yet easy-to-
Portable communicators and use tool for field calibration of
calibrators are often used to- HART instrumentation. Pressing
• Software Version
• Final assembly number
Device Sensor 2
Setup • LRV, URV
• Change Sensor Type
Identification
• Damping
• Change Sensor
• Transfer function Connections
(Dual sensor
temperature devices)
Figure 3
No communicator is • Interrogate HART devices to • Read and write HART tag and
required! determine type, manufacturer, message fields to re-label
model, tag-ID, PV, and PVAO smart transmitters
The 744 requires no external Perform automated HART
box or communicator for every- • • Clone additional transmitters
sensor trim and output trim with basic HART configuration
day HART calibration and main- for selected devices
tenance. It supports many data
popular models of HART trans- • Adjust ranging, damping, and
mitters, with more device- other basic process–configu-
specific command support than ration settings
any other HART field calibrator.
communication connection is
not important. A separate
250 ohm resistor is not neces-
sary because the 744 incorpo-
rates a resistor in series with
the 24V loop supply through its
mA jacks. The 3051 in this ex-
Figure 5
ample is configured for psi units.
Figure 6
HART Connection
Hand Pump
744 DOCUMENTING PROCESS CALIBRATOR
S I G N A L
+ –
T E S T
MEAS
SOURCE
mA SETUP
V TC
V
Hz RTD
Pressure
7 8 9 CLEAR
( ZERO) Module
4 5 6
1 2 3
0 . ENTER
Pressure Input
V mA mA V
RTD
RTD
CAT
30V SOURCE MEAS 300V
MAX MAX
TC
30V 30V
MAX MAX
Red
Figure 4
HART Connection
TC
+ TC
744 DOCUMENTING PROCESS CALIBRATOR –
2 3
1 4
– 5
+
T
3144
Transmitter
MEAS
SOURCE
mA SETUP
V TC
V RTD
Hz
CLEAR
7 8 9 (ZERO)
4 5 6
1 2 3
0 . ENTER
V mA mA V
RTD
RTD +
30V
MAX
SOURCE
30V
CAT
MEAS
30V
300V
MAX
TC
Red
MAX MAX
Black
Figure 10
Figure 12 Figure 15
CLEAR
TEST DC PWR
7 8 9
– ++ –
instruments not supported by the
( ZERO)
4 5 6
0
3
. ENTER
Red
744 (to perform the As Found V mA mA V
CAT
30V SOURCE MEAS 300V TC
Figure 17
2. Press the key again and ensures that errors are calcu- variable (PVAO) is in the
you are prompted to select lated correctly. upper right corner of the
the 744 configuration (Figure display (Figure 23). This is
20). Move the cursor to normally a 4 mA signal. The
MEAS mA, SOURCE T/C real-time mA value as mea-
typ K, and press . (If you
ENTER
sured by the Fluke 744, is in
were verifying the digital PV the center of the display.
instead of the mA output, i.e., Press the Fetch softkey to
the transmitter has a non- load the measured mA value.
zero HART poll address, you Press the Send softkey to
would select MEAS PV, send the value to the trans-
SOURCE T/C typ K instead.) mitter to trim the output
section for the 4 mA value.
Press Continue for the 20
mA adjustment and repeat
Figure 21 this step.
4. Press Done and then press
Auto Test. Once the test is
complete, an error summary
table is displayed (Figure 22).
Test errors exceeding the
tolerance are highlighted. If
the test passed, i.e., if no
errors are highlighted, ad-
Figure 20 justment is not required.
Figure 23
100% test points. (If the error to display the basic setup
summary table is no longer parameters shown in Figure
displayed, you can use the 24. To ENTER the new LRV,
Review Memory softkey to move the cursor to Lower
recall the As Left data.) Range Value and press . ENTER
Return the 744 to the normal Type the new LRV and press
Measure/Source screen dis- . Also type in the new
ENTER
playing the As Left softkey URV and press . Press the ENTER
Figure A2
HART Transmitter Block Diagram
DocuMint ™
OnTimeSupport ™
Fluke DPC/TRACK ™
ProCALV Meridium
Intools
Ordering Information
FLUKE-744 Documenting Process Calibrator-HART
Includes: TL224 Industrial Test Leads (two sets), AC220
Test Clips (2 sets), TP220 Test Probes (1 set), BP7235
NiMH Battery Pack, BC7217 Battery Charger, serial port
cable, HART communications cable, DPC/TRACK
Sample with free PC communication utility software,
Instruction Manual, HART User’s Manual, NIST-
traceable calibration certificate and data, three-year
warranty.
Optional Accessories
Fluke-700 Pxx Pressure Modules
Included with each Fluke Pressure
Module: BSP-ISO to NPT Adapter(s)
(except with P29 - P31), Instruction
Sheet, NIST traceable calibration
Fluke. Keeping your world
report and data, one-year warranty. up and running.
Fluke-700BCW Bar Code Wand Fluke Corporation
Fluke-700PTP Pneumatic Test Pump, 360 psi/25 bar PO Box 9090, Everett, WA USA 98206
ANAG E M E NT
T
QU
EM
90
TI
R
FI O
ED IS
TO M E E T