Sunteți pe pagina 1din 16

®

HART Transmitter
Calibration
Application Note

Introduction
In today’s process plants, most
new field instruments are smart
digital instruments. Smart im-
plies a microprocessor-based
instrument with extra functional-
ity and digital compensation,
supporting multiple sensor types
or multiple variables. These in-
struments generally offer better
accuracy, long-term stability,
and reliability than conventional
analog instruments.
The most common class of
smart instruments incorporates
the HART protocol, with more
than five million HART instru-
ments in use in 100,000 plants
worldwide. HART, an acronym
for Highway Addressable Remote
Transducer, is an industry stan-
dard that defines the communi-
cations protocol between smart
field devices and a control sys-
tem that employs traditional
4-20 mA wiring.
Two capabilities are required
to properly service HART instru-
ments: precision analog source
and measure capability and digi-
tal communication capability.
Until recently, this required two
separate tools, a calibrator and a
communicator. Today, the capa-
bilities of those two tools are
available in a single HART Docu-
menting Process Calibrator that
can help you quickly and effec-
tively service HART instruments.
HART calibration Regular calibration is also pru- How are HART
is required! dent since performance checks instruments properly
will often uncover problems not calibrated?
A common misconception is directly caused by the instru-
that the accuracy and stability mentation, such as solidified or To calibrate a HART instrument
of HART instruments eliminate congealed pressure lines, in- consistent with its application, it
the need for calibration. stallation of an incorrect ther- is very helpful to understand the
Another misconception is that mocouple type, or other errors functional structure of a typical
calibration can be accomplished and faults. HART transmitter. The article in
by re-ranging field instruments A calibration procedure con- Appendix A, by Kenneth L.
using only a HART communica- sists of a verification (As Found) Holladay of Southwest Research
tor. Still another misconception test, adjustment to within ac- Institute, describes a typical
is that the control system can ceptable tolerance if necessary, HART instrument and defines
remotely calibrate smart instru- and a final verification (As Left) both proper and improper cali-
ments. These are not true. All test if an adjustment has been bration practices. Originally
instruments drift. Re-ranging made. Data from the calibration published in Intech, May 1996, it
with just a communicator is not are collected and used to com- is reprinted with permission of
calibration. A precision calibra- plete a report of calibration, the author.
tor or standard is required. documenting instrument per-
Regular performance verifica- Note: If you are unfamiliar with HART
formance over time. calibration or need a review, this is an
tion with a calibrator traceable All instruments, even HART excellent point to stop and read the article
to national standards is neces- instruments, must be calibrated in Appendix A. It covers the basics of HART
sary due to: on a regular, preventive main- instrumentation and addresses issues critical
to instrument maintenance.
1. Shifts in performance of tenance schedule. The calibra-
electronic instruments over tion interval should be set short HART instruments consist of
time, due to exposure of the enough to insure that an in- three distinct sections (see
electronics and the primary strument never drifts out of Figure 1). Proper HART calibra-
sensing element to tempera- tolerance, yet long enough to tion may involve either or both
ture, humidity, pollutants, avoid unnecessary calibrations. sensor trim and output trim.
vibration, and other field Alternatively, the interval may Adjusting range values (LRV and
environmental factors. be determined by critical pro- URV) without a calibrator is not
2. Regulations governing occu- cess requirements, e.g., calibra- calibration. Performing an output
pational safety, consumer tion before each batch. trim while ignoring the input
safety, and environmental section is not proper calibration.
protection. Adjusting range values with a
3. Quality programs such as calibrator may be a practical
ISO 9000 standards for all calibration alternative for instru-
instruments that impact ments operated in 4-20 mA
product quality. analog mode, provided that the
PV and PVAO are not used for
4. Commercial requirements process control.
such as weights, measures,
and custody transfer.

Sensor Input PV Conversion PVAO Output 4-20 mA Analog


Analog
Input Section Section Section mA Output
(digital (digital
input) 4-20 mA)

Sensor Trim LRV/URV Adjust Output Trim

Figure 1

2 Fluke Corporation Hart Transmitter Calibration


Analog Measure Value
Model Number
Analog Source Value
PV Tag ID
(Primary Variable)

PVAO
(Digital representation PV LRV
of the Primary Variable (Primary Variable
Analog Output) Lower Range Value)

PV URV
(Primary Variable
Upper Range Value)

Figure 2

New tool speeds gether to complete field calibra- a single key enters the HART
calibration tions. However, the desire to mode and displays the essential
carry less equipment and to HART information in the Active
Today, instrument maintenance perform maintenance in the Device Screen, shown in Figure
is moving out of the shop and field has created a need for a 2. Additional HART functionality
into the field. This reduces pro- new class of calibration tool. is accessed with only a few
cess interruptions and avoids The new 744 Documenting more keystrokes, per the menu
the time and expense of return- Process Calibrator from Fluke is tree in Figure 3.
ing instruments to the shop. the first powerful yet easy-to-
Portable communicators and use tool for field calibration of
calibrators are often used to- HART instrumentation. Pressing

3 Fluke Corporation Hart Transmitter Calibration


• Tag • Keypad input
• View process
variables • PV units • Adjust URV, LRV to
Process Basic • LRV, URV applied values
• View variable
map • Damping
• Re-map process • Transfer function
variables Sensor 1
(Dual sensor • Sensor serial number
temperature • Sensor lower and
Detailed • Config Sensor
devices) upper limits
(coriolis) • Config Output
• Sensor minimum span
Temperature
Sensor devices only:
• Change Sensor Type
• Change Sensor
Connections

• Software Version
• Final assembly number
Device Sensor 2
Setup • LRV, URV
• Change Sensor Type
Identification
• Damping
• Change Sensor
• Transfer function Connections
(Dual sensor
temperature devices)

HART • Dual Sensor Config


Output (Dual sensor
temperature devices)

Loop test • Manufacturer


Service • Model
Pressure • Write protect
zero trim HART • Device HART ID
• Alarm state
Information • Software revision
• HART poll address
Output trim • Hardware revision
• HART burst mode
• Number of preambles
Sensor trim • HART burst command
Abort

Figure 3

No communicator is • Interrogate HART devices to • Read and write HART tag and
required! determine type, manufacturer, message fields to re-label
model, tag-ID, PV, and PVAO smart transmitters
The 744 requires no external Perform automated HART
box or communicator for every- • • Clone additional transmitters
sensor trim and output trim with basic HART configuration
day HART calibration and main- for selected devices
tenance. It supports many data
popular models of HART trans- • Adjust ranging, damping, and
mitters, with more device- other basic process–configu-
specific command support than ration settings
any other HART field calibrator.

4 Fluke Corporation Hart Transmitter Calibration


Versatile HART protocol HART operating modes perform the vast majority of
support supported day-to-day operations you nor-
mally perform with a separate
The 744 supports the commands • For Point to Point operation, communicator. The HART capa-
contained in HART protocol ver- the most commonly used bility of the 744 is comparable
sion 5.7. With 2 MB of memory, mode, connects the 744 to a to that of the model 275 HART
the 744 supports a substantial single HART device in a 4-20 communicator, with the excep-
set of HART instructions: mA loop. tion of the DD interpreter. While
• Universal commands — • In Multi-Drop mode, several the DD interpreter enables the
provide functions that are HART instruments can be 275 communicator to read com-
implemented in all field bussed together. The 744 mand set libraries from any
devices, for example, read searches for each, identifies HART supplier, it offers capabili-
manufacturer and device addresses in use, and allows ties far beyond those generally
type, read primary variable you to select the instrument required for daily HART instru-
(PV), or read current output for calibration and related ment maintenance.
and percent of span operations.
• Common practice com- • In Burst Mode, the HART HART calibration
mands — provide functions instrument transmits bursts of applications
that are common to many but data without waiting to be The following examples demon-
not all field devices, for interrogated by a master unit. strate how the 744 makes HART
example read multiple vari- The 744 can take transmitters calibration an efficient opera-
ables, set damping time, or out of burst mode during test tion. The 744 enables easy
perform loop test or calibration, then later hookup using its HART cable,
• Device-specific commands — restore them to burst mode. fast access to the most impor-
provide functions that are tant HART data, automatic
unique to a particular field Is there still a role for the branching to appropriate adjust-
device, for example sensor communicator? ment choices, automatic
trim. The 744 Version 2.5 Commissioning a HART instru- completion of test templates,
supports these devices: ment or modifying HART vari- and automatic fetching and
ables not supported by the 744 sending of analog readings
requires the use of a communi- during trim.
cator. The 744 is designed to
Manufacturer Pressure Instruments Temperature Instruments Coriolis Instruments
ABB/Kent-Taylor 600T 658T1
1
ABB/ Contrans P,
Hartmann & Braun AS 800 Series
Endress & Hauser CERABAR S, TMT 1221, TMT 1821,
CERABAR M, TMT 1621
DELTABAR S
Foxboro Eckardt TI/RTT201
Foxboro/Invensys I/A Pressure
Fuji FCX FRC
FCXAZ
Honeywell ST3000 STT25T1, STT25H1
Micro Motion 2000
2000 IS
9701
9712
9739
Moore Products 3441
Rosemount 1151 3044C
2088 644
3001C 3144
3051, 3051S 3244, 3144P
Siemens SITRANS P DS
SITRANS P ES
SMAR LD301 TT3011
Viatran I/A Pressure
Wika UNITRANS T32H1
Yokogawa EJA YTA 110, 310 and 320
1
Table 1 Sensor Trim not supported

5 Fluke Corporation Hart Transmitter Calibration


Procedure 2. Press the key again and
Example 1 you are prompted to select
1. Power on the Fluke 744 the 744 configuration (Figure
Calibration of a Calibrator. Press the red 6). Selecting MEAS mA,
Rosemount 3051 HART key followed by the Loop SOURCE psi will configure
Pressure Transmitter Power softkey and the 744 the calibrator to measure the
will display the basic HART analog mA output and the
Basic connections information for the 3051 pressure being applied si-
This example assumes that the (Figure 5). multaneously to the trans-
transmitter is isolated from the mitter input and the pressure
process and is not electrically module. (Selecting MEAS PV,
connected to a loop power sup- SOURCE psi will configure
ply. Make basic connections to the 744 to evaluate the digi-
the 3051 per the diagram in tal PV output from the trans-
Figure 4. Polarity of the HART mitter.) Press to select.
ENTER

communication connection is
not important. A separate
250 ohm resistor is not neces-
sary because the 744 incorpo-
rates a resistor in series with
the 24V loop supply through its
mA jacks. The 3051 in this ex-
Figure 5
ample is configured for psi units.

Figure 6

HART Connection

Hand Pump
744 DOCUMENTING PROCESS CALIBRATOR

S I G N A L

+ –
T E S T

MEAS
SOURCE
mA SETUP

V TC
V
Hz RTD
Pressure
7 8 9 CLEAR
( ZERO) Module
4 5 6

1 2 3

0 . ENTER
Pressure Input
V mA mA V
RTD
RTD

CAT
30V SOURCE MEAS 300V
MAX MAX
TC
30V 30V
MAX MAX
Red

mA Measure, 24V Loop Black

Figure 4

6 Fluke Corporation Hart Transmitter Calibration


3. Vent the pressure line and 5. If the As Found test failed 7. Select Output Trim and
press to zero the pressure
CLEAR
(ZERO) (i.e., there were highlighted pressENTER . The value of the
module. Press the As Found errors in the error summary primary variable (PVAO) is in
softkey, and then press table), adjustment is neces-
ENTER the upper right corner of the
to select Instrument for a sary. Press the Adjust display. This is normally a
linear transmitter calibration. softkey. Select Sensor Trim 4 mA signal. The mA value,
(If the 3051 is configured for and press . (Do not select
ENTER as constantly measured by
square root output, select Pressure Zero Trim. It is the the Fluke 744, is in the cen-
Instrument.) Notice that same as trimming the lower ter of the display. Press the
the calibration template is sensor point at zero, which is Fetch softkey to load the
automatically completed with useful for pressure transmit- measured mA value. Press
the exception of Tolerance. ters that do not offer Sensor Send to send the value to
Fill in the appropriate test Trim.) The 744 screen should the 3051 to trim the output
tolerance and press Done. look like Figure 8. section for the 4 mA value.
4. Press the Manual Test Press Continue for the 20
softkey to begin calibration. mA trim and repeat this step.
Apply the input pressures as 8. After completing Output Trim,
instructed in the SOURCE press the Done softkey and
screen. Press the Accept proceed with the As Left
Point softkey when the cor- verification test. Press the As
rect pressure is applied for Left softkey. Press Done and
each point. When the test is then press Manual Test.
complete, the error summary Apply the requested pres-
table is displayed (Figure 7). sures and press Accept Point
Test errors exceeding the when the readings are
tolerance are highlighted. stable. On completion an
When done viewing the Figure 8
error summary table is dis-
table, press the Done softkey. played. If none of the errors
Press Done again to accept, are highlighted (Figure 9),
6. Select Perform user trim – the 3051 passes the calibra-
or ENTER to change the tag, both and press . Zero the
ENTER
serial number or ID fields. tion test. If errors are high-
pressure module (vented to lighted, the test has failed
atmosphere) by pressing .
CLEAR
(ZERO)

and further adjustment is


Press the Continue softkey required. Return to step 5 for
and you are prompted for the adjustment of the 3051.
Lower Trim value. For best
results, apply the LRV pres-
sure and press Fetch to load
the value being measured by
the pressure module. Press
Trim. Then press Continue
to move to the Upper Trim.
As before, apply the URV
pressure, press Fetch, and
press Trim. If the 3051 is
Figure 7 used with the digital PV
output, skip to step 8 and
perform the As Left test. If
the 4-20 mA analog output
is used in the process, con- Figure 9
tinue on to step 7.

7 Fluke Corporation Hart Transmitter Calibration


Procedure 2. Press the key again and
Example 2 you are prompted to select
1. Power on the Fluke 744 the 744 configuration (Fig-
Calibration of a Calibrator. Press the red ure 12). Selecting MEAS mA,
Rosemount 3144 HART key followed by the Loop SOURCE T/C typ K config-
Power softkey. Press to ENTER
Temperature Transmitter bypass the warning screens
ures the calibrator to mea-
sure the analog mA output of
Basic connections and the 744 will display the the transmitter and source
This example assumes that the basic HART information for the correct temperature
transmitter is isolated from the the 3144 (Figure 11). stimulus at the 3144 input.
process and is not electrically (Selecting MEAS PV,
connected to a loop power SOURCE T/C typ K will
supply. Make basic connections configure the 744 to evalu-
to the 3144 per the diagram in ate the digital PV output
Figure 10. Polarity of the HART from the transmitter.) Press
communication connection is to select.
ENTER

not important. A separate


250 ohm resistor is not neces-
sary because the 744 incorpo-
rates a resistor in series with
the 24V loop supply through its
mA jacks. The 3144 in this ex-
ample is configured for a type K
Figure 11
thermocouple sensor with a
span of 0-300 °C.

HART Connection

TC
+ TC
744 DOCUMENTING PROCESS CALIBRATOR –

2 3
1 4
– 5
+
T

3144
Transmitter
MEAS
SOURCE
mA SETUP

V TC
V RTD
Hz

CLEAR
7 8 9 (ZERO)

4 5 6

1 2 3

0 . ENTER

V mA mA V
RTD
RTD +
30V
MAX
SOURCE
30V
CAT
MEAS
30V
300V
MAX
TC
Red
MAX MAX

Black

Figure 10

8 Fluke Corporation Hart Transmitter Calibration


5. If the As Found test failed
(i.e., there were highlighted
errors in the error summary
table), adjustment is neces-
sary. Press the Adjust softkey.
Select Sensor Trim and
press . Select Perform
ENTER

user trim – both and press


. The 744 screen should
ENTER

look like Figure 14.

Figure 12 Figure 15

3. Press the As Found softkey, 8. After completing Output Trim,


and then press to select
ENTER
press the Done softkey and
Instrument for a linear proceed with the As Left
transmitter calibration. Notice verification test. Press the As
that the calibration template Left softkey. Press Done and
is automatically completed then press Auto Test. On
with the exception of the completion, an error summary
Tolerance. Fill in the appro- table is displayed. If errors
priate test tolerance and are highlighted, the test has
press the Done softkey. Figure 14
failed and further adjustment
4. Press the Auto Test softkey is required. Return to step 5
to begin calibration. Once 6. For best results, press LRV to for adjustment of the 3144.
the test is complete, an error apply the LRV for the Lower
summary table is displayed Trim value. Press Trim and
(Figure 13). Test errors ex- then Continue to move to
ceeding the tolerance are the Upper Trim. Press URV,
highlighted. When done press Trim, and then press
viewing the table, press the Done. If the 3144 is used
Done softkey. Press Done with the digital PV output,
again to accept, or to ENTER skip to step 8 and perform
change the tag, serial num- the As Left test. If the analog
ber or ID fields. 4-20 mA output is used in
the process, continue on to
step 7.
7. Select Output Trim and
press . The value of the
ENTER
Figure 16
primary variable (PVAO) is in
the upper right corner of the
display. (Figure 18). This is
normally a 4 mA signal. The
mA value, as constantly
measured by the Fluke 744,
is in the center of the dis-
play. Press the Fetch softkey
to load the measured mA
Figure 13 value. Press Send to send
the value to the 3144 to trim
the output section for the 4
mA value. Press Continue for
the 20 mA trim and repeat
this step.

9 Fluke Corporation Hart Transmitter Calibration


adjustment is required, this is not found at 0 the 744 will
Example 3 example first performs an Output begin polling addresses 1 to 15.
Trim to bring the instrument The 744 also displays a non-
Calibration of HART within tolerance. Failing that, zero address with the basic
instruments using the example performs an adjust- HART information.
universal commands ment to the Lower and Upper
Range Values (LRV and URV) to Basic connections
The 744 supports a majority of compensate for input section
the installed workload of HART This example assumes that the
error. transmitter is isolated from the
transmitters – see Table 1 – by
supporting sensor trim, which Note: Appendix A explains that these adjust- process and is not electrically
employs device-specific com- ments do not constitute a proper HART cali- connected to a loop power sup-
mands that are unique to a par- bration. While this is true, these adjustments ply. Make basic connections to
are a practical calibration alternative for the transmitter per the diagram
ticular instrument. So how can instruments operated in 4-20 mA analog
you calibrate instruments that mode if error corrections are not large in Figure 18. Polarity of the
are not supported by the 744? HART communication connec-
The short answer is that the How to determine digital tion is not important. A sepa-
744 supports a substantial set of or analog? rate 250 ohm resistor is not
the universal HART commands necessary because the 744
and the common practice HART The transmitter is in digital incorporates a resistor in series
commands. The 744 can com- mode if its HART Poll Address is with the 24V loop supply
municate with virtually any set between 1 to 15. An address through its mA jacks. This ex-
HART instrument and, in most of 0 (zero) sets it to 4-20 mA ample assumes a type K ther-
cases, can complete a calibration analog output mode. The 744 mocouple transmitter with an
procedure (except for sensor trim will automatically connect to a input range of 0-100 °C, 4-20
for unsupported instruments). device at address 0; if a device mA output, and a 0.25% test
This example applies to tolerance.
instruments used in analog
mode (4-20 mA). If the instru- 744 DOCUMENTING PROCESS CALIBRATOR

ment is operated in digital


mode, i.e., its PV is the output
variable that is used for control,
a calibration of the Input Section
is all that is needed. Adjustment
will require a Sensor Trim, (see MEAS
SOURCE
mA SETUP

Figure 17) which means that for V


V
Hz
TC
RTD

CLEAR
TEST DC PWR
7 8 9
– ++ –
instruments not supported by the
( ZERO)

4 5 6

744 you will need to use both a 1 2

0
3

. ENTER
Red
744 (to perform the As Found V mA mA V

and As Left tests and record the


RTD
RTD

CAT
30V SOURCE MEAS 300V TC

results) and a communicator (to


MAX 30V 30V MAX
MAX MAX

perform sensor trim).


For instruments used in ana-
log mode, i.e., where the 4-20 Black
mA analog output is used for
control, the 744 can be used for
calibration. After performing an
Figure 18
As Found and determining that

Sensor Input PV Conversion PVAO Output 4-20 mA Analog


Analog
Input Section Section Section mA Output
(digital (digital
input) 4-20 mA)

Sensor Trim LRV/URV Adjust Output Trim

Figure 17

10 Fluke Corporation Hart Transmitter Calibration


Procedure 3. Press the As Found softkey
and press to select
ENTER
1. Power on the Fluke 744 Instrument calibration. Move
Calibrator. Press the key the cursor to Tolerance and
and the Loop Power softkey ENTER the appropriate test
(if loop power is not already tolerance (0.25% in this ex-
supplied). Press until any
ENTER
ample). Verify that the 0%
device warnings are cleared Value and 100% Value are
and the basic HART informa- the proper, nominal operat-
tion is displayed (Figure 19). ing values for the transmitter
(0.0 °C and 100.0 °C in this
example, Figure 21). If the
Lower (0%) and Upper Figure 22
(100%) Range Values (LRV
and URV) have been previ- If errors are highlighted,
ously modified for calibration adjustment is necessary by
purposes, you will need to performing an Output Trim.
ENTER the nominal values. Press Done to leave the re-
For example, if a previous sults screen, edit the tag,
calibration modified the URV serial number or ID fields as
to 100.2 °C, you need to necessary, and press Done
manually ENTER the nomi- again.
nal value of 100.0 °C for the 5. Press the Adjust softkey,
Figure 19
100% Value. Entering nomi- select Output Trim and press
nal zero and span values . The value of the primary
ENTER

2. Press the key again and ensures that errors are calcu- variable (PVAO) is in the
you are prompted to select lated correctly. upper right corner of the
the 744 configuration (Figure display (Figure 23). This is
20). Move the cursor to normally a 4 mA signal. The
MEAS mA, SOURCE T/C real-time mA value as mea-
typ K, and press . (If you
ENTER
sured by the Fluke 744, is in
were verifying the digital PV the center of the display.
instead of the mA output, i.e., Press the Fetch softkey to
the transmitter has a non- load the measured mA value.
zero HART poll address, you Press the Send softkey to
would select MEAS PV, send the value to the trans-
SOURCE T/C typ K instead.) mitter to trim the output
section for the 4 mA value.
Press Continue for the 20
mA adjustment and repeat
Figure 21 this step.
4. Press Done and then press
Auto Test. Once the test is
complete, an error summary
table is displayed (Figure 22).
Test errors exceeding the
tolerance are highlighted. If
the test passed, i.e., if no
errors are highlighted, ad-
Figure 20 justment is not required.

Figure 23

11 Fluke Corporation Hart Transmitter Calibration


6. Now perform an As Left test. 9. Calculate the new LRV or
Press As Left, press Done, URV by multiplying the span
and then press Auto Test. On by the error in percent and
completion the error sum- adding that to the old value.
mary table is displayed. If If our example has the fol-
errors are highlighted, the lowing nominal Source val-
test has failed and further ues, errors, and old LRV and
adjustment is required. URV values:
Note: If the failure error is large, sensor trim Source Error% Old LRV/
adjustment with a communicator may be nec- URV
essary. Often, however, adjustment can be ac-
complished with a 744 by modifying the LRV 0% Value 0.0 ºC 0.84% 0.4 ºC
(Lower Range Value) and URV (Upper Range 100%Value 100.0 ºC -2.41% 102.0 ºC
Value) to compensate for Input Section error. Figure 24
Table 2
7. In the case of a pressure 11. Now press Done and then
transmitter that has on- Calculate the new LRV and press Abort 3 times. Per-
board Zero and Span adjust- URV as follows: form a new As Found test
ment buttons, calibration is by pressing As Found. (Re-
LRVnew = LRVold + (Span × Error0%)
easy. Simply apply a cali- member to make sure that
brated source at the LRV and LRVnew = 0.4 ºC + (100.0 ºC × 0.84%) the original, nominal zero
URV values and press the and span values are shown
respective Zero and Span LRVnew = 0.4 ºC + (100.0 ºC × 0.0084)
as the 0% Value and 100%
buttons on the transmitter. LRVnew = 0.4 ºC + 0.8 ºC Value.) Press Done and
Then verify the condition of LRVnew = 1.2 ºC then press Auto Test. On
the transmitter by complet- completion, the error sum-
ing an As Left test as in step URVnew = URVold + (Span × Error100%) mary table is displayed.
6. Many HART transmitters If errors are highlighted,
do not have physical adjust- URVnew = 102.0 ºC + (100.0 ºC × the test has failed – repeat
ments and need either a –2.41%) the adjustment or trim
communicator or a Fluke 744 URVnew = 102.0 ºC + (100.0 ºC × sensor section with a
to adjust the LRV and URV –0.0241) communicator.
values. For those cases, pro-
URVnew = 102.0 ºC – 2.4 ºC Note: If you encounter any difficulty with
ceed to step 8. any of these examples, you may call 1-800-
8. The error summary table URV = 99.6 ºC 44-FLUKE for assistance (1-800-443-5853).
new
(displayed from step 6) pro-
vides the data necessary to
make LRV and URV changes. 10. Press and then press the
Write down the ERROR % Setup softkey. Select Basic
values for the failed 0% and from the menu and press ENTER

100% test points. (If the error to display the basic setup
summary table is no longer parameters shown in Figure
displayed, you can use the 24. To ENTER the new LRV,
Review Memory softkey to move the cursor to Lower
recall the As Left data.) Range Value and press . ENTER

Return the 744 to the normal Type the new LRV and press
Measure/Source screen dis- . Also type in the new
ENTER

playing the As Left softkey URV and press . Press the ENTER

by pressing the Done softkey Send softkey.


3 times.

12 Fluke Corporation Hart Transmitter Calibration


Appendix A a physical quantity into an elec- variable using a communicator,
trical signal. However, the simi- this is the value that you see.
larity ends there. Instead of a The second box is strictly a
Calibrating HART purely mechanical or electrical mathematical conversion from
Transmitters path between the input and the the process variable to the
resulting 4-20 mA output signal, equivalent milliamp representa-
By Kenneth L. Holladay, P.E.
a HART transmitter has a micro- tion. The range values of the
Calibrating a conventional processor that manipulates the instrument (related to the zero
input data. As shown in Figure and span values) are used in
instrument A2, there are typically three conjunction with the transfer
For a conventional 4-20 mA calculation sections involved, function to calculate this value.
instrument, a multiple point test and each of these sections may Although a linear transfer func-
that stimulates the input and be individually tested and tion is the most common, pres-
measures the output is sufficient adjusted. sure transmitters often have a
to characterize the overall accu- Just prior to the first box, the square root option. Other special
racy of the transmitter. The instrument’s microprocessor instruments may implement
normal calibration adjustment measures some electrical prop- common mathematical transfor-
involves setting only the zero erty that is affected by the pro- mations or user defined break
value and the span value, since cess variable of interest. The point tables. The output of the
there is effectively only one measured value may be milli- second block is a digital repre-
adjustable operation between volts, capacitance, reluctance, sentation of the desired instru-
the input and output as illus- inductance, frequency, or some ment output. When you read the
trated below. other property. However, before loop current using a communi-
it can be used by the micropro- cator, this is the value that you
Zero and Span cessor, it must be transformed see. Many HART instruments
Adjustments to a digital count by an analog support a command which puts
to digital (A/D) converter. the instrument into a fixed out-
Sensing 4-20 mA In the first box, the micropro- put test mode. This overrides the
Element Analog Output cessor must rely upon some form normal output of the second
Electronics of equation or table to relate the block and substitutes a specified
raw count value of the electrical output value.
measurement to the actual prop- The third box is the output
Figure A1 erty (PV) of interest such as tem- section where the calculated
Conventional Transmitter Block Diagram
perature, pressure, or flow. The output value is converted to a
This procedure is often re- principle form of this table is count value that can be loaded
ferred to as a Zero and Span usually established by the into a digital to analog con-
Calibration. If the relationship manufacturer, but most HART verter. This produces the actual
between the input and output instruments include commands analog electrical signal. Once
range of the instrument is not to perform field adjustments. again the microprocessor must
linear, then you must know the This is often referred to as a rely on some internal calibration
transfer function before you can sensor trim. The output of the factors to get the output correct.
calculate expected outputs for first box is a digital representa- Adjusting these factors is often
each input value. Without tion of the process variable. referred to as a current loop trim
knowing the expected output When you read the process or 4-20 mA trim.
values, you cannot calculate the
performance errors. High and Low Range and High and Low
Sensor Trim Transfer Function Output Trim
Calibrating a HART
instrument
A/D D/A
Counts

For a HART instrument, a mul- Counts PV mA Counts


mA
PV

tiple point test between input


and output does not provide an
Counts PV mA
accurate representation of the
transmitter’s operation. Just like Input Conversion Output
a conventional transmitter, the Section Section Section
measurement process begins
with a technology that converts PV may be mA may be set
read digitally and read digitally

Figure A2
HART Transmitter Block Diagram

13 Fluke Corporation Hart Transmitter Calibration


HART calibration always a linear relationship Testing overall performance
requirements between the input and output, After calibrating both the Input
Based on this analysis, you can and both are recorded in the and Output sections, a HART
see why a proper calibration same engineering units. In gen- transmitter should operate
procedure for a HART instru- eral, the desired accuracy for this correctly. The middle block in
ment is significantly different test will be the manufacturer’s Figure A2 only involves compu-
than for a conventional instru- accuracy specification. tations. That is why you can
ment. The specific calibration If the test does not pass, then change the range, units, and
requirements depend upon the follow the manufacturer’s rec- transfer function without neces-
application. ommended procedure for trim- sarily affecting the calibration.
If the application uses the ming the input section. This Notice also that even if the in-
digital representation of the may be called a sensor trim and strument has an unusual trans-
process variable for monitoring typically involves one or two fer function, it only operates in
or control, then the sensor input trim points. Pressure transmit- the conversion of the input
section must be explicitly tested ters also often have a zero trim, value to a milliamp output value,
and adjusted. Note that this where the input calculation is and therefore is not involved in
reading is completely indepen- adjusted to read exactly zero the testing or calibration of ei-
dent of the milliamp output, and (not low range). Do not confuse ther the input or output sections.
has nothing to do with the zero a trim with any form of re-rang- If there is a desire to validate
or span settings. The PV as read ing or any procedure that in- the overall performance of a
via HART communication con- volves using zero and span HART transmitter, run a Zero
tinues to be accurate even buttons. and Span test just like a con-
when it is outside the assigned ventional instrument. As you
output range. For example, a Calibrating the output will see in a moment, however,
range 2 Rosemount 3051c has section passing this test does not nec-
sensor limits of -250 to +250 Again, the same basic multiple essarily indicate that the trans-
inches of water. If you set the point test and adjust technique mitter is operating correctly.
range to 0 - 100 inches of is employed, but with a new
water, and then apply a pres- definition for input. To run a test, Effect of damping on test
sure of 150 inches of water, the use a communicator to put the performance
analog output will saturate at transmitter into a fixed current Many HART instruments support
just above 20 milliamps. How- output mode. The input value for a parameter called damping. If
ever, a communicator can still the test is the mA value that you this is not set to zero, it can
read the correct pressure. instruct the transmitter to pro- have an adverse effect on tests
If the current loop output is duce. The output value is ob- and adjustments. Damping in-
not used (that is the transmitter tained using a calibrator to duces a delay between a
is used as a digital only device),measure the resulting current. change in the instrument input
then the input section calibra- This test also implies a linear and the detection of that
tion is all that is required. If the
relationship between the input change in the digital value for
application uses the milliamp and output, and both are re- the instrument input reading
output, then the output section corded in the same engineering and the corresponding instru-
must be explicitly tested and units (milliamps). The desired ment output value. This damp-
calibrated. Note that this cali- accuracy for this test should also ing induced delay may exceed
bration is independent of the reflect the manufacturer’s accu- the settling time used in the test
input section, and again, has racy specification. or calibration. The settling time
nothing to do with the zero and If the test does not pass, then is the amount of time the test or
span settings. follow the manufacturer’s rec- calibration waits between set-
ommended procedure for trim- ting the input and reading the
Calibrating the input section ming the output section. This resulting output. It is advisable
The same basic multiple point may be called a 4-20 mA trim, a to adjust the instrument’s
test and adjust technique is current loop trim, or a D/A trim. damping value to zero prior to
employed, but with a new defi- The trim procedure should re- performing tests or adjustments.
nition for output. To run a test, quire two trim points close to or After calibration, be sure to
use a calibrator to measure the just outside of 4 and 20 mA. Do return the damping constant to
applied input, but read the not confuse this with any form its required value.
associated output (PV) with a of re-ranging or any procedure
communicator. Error calculations that involves using zero and
are simpler since there is span buttons.

14 Fluke Corporation Hart Transmitter Calibration


Operations that are NOT zero shift. Thus with both ports button. Suppose there is also a
proper calibrations vented (zero), its output is 4.16 digital indicator in the loop that
mA instead of 4.00 mA, and displays 0.0 at 4 mA, and 100.0
Digital range change when applying 100 inches of at 20 mA. During testing, it read
There is a common misconcep- water, the output is 20.16 mA 1.0 with both ports vented, and
tion that changing the range of instead of 20.00 mA. To fix this it read 101.0 with 100 inches of
a HART instrument by using a he vents both ports and presses water applied. Using the com-
communicator somehow cali- the zero button on the trans- municator, the technician per-
brates the instrument. Remem- mitter. The output goes to 4.00 forms a current loop trim so that
ber that a true calibration mA, so it appears that the ad- the display reads correctly at 0
requires a reference standard, justment was successful. and 100, essentially correcting
usually in the form of one or However, if he now checks the output to be 4 and 20 mA
more pieces of calibration the transmitter with a commu- respectively.
equipment to provide an input nicator, he will find that the While this also appears to be
and measure the resulting out- range is 1 to 101 inches of wa- successful, there is a fundamen-
put. Therefore, since a range ter, and the PV is 1 inch of tal problem with this procedure.
change does not reference any water instead of 0. The zero To begin with, the communica-
external calibration standards, it and span buttons changed the tor will show that the PV still
is really a configuration change, range (the second block). This reads 1 and 101 inches of water
not a calibration. Notice that in is the only action that the in- at the test points, and the digi-
the HART transmitter block dia- strument can take under these tal reading of the mA output still
gram (Figure 2), changing the conditions since it does not reads 4.16 and 20.16 mA, even
range only affects the second know the actual value of the though the actual output is 4
block. It has no effect on the reference input. Only by using and 20 mA. The calibration
digital process variable as read a digital command which con- problem in the input section has
by a communicator. veys the reference value can been hidden by introducing a
the instrument make the appro- compensating error in the out-
Zero and span adjustment priate internal adjustments. put section, so that neither of
The proper way to correct a the digital readings agrees with
Using only the zero and span zero shift condition is to use a the calibration standards.
adjustments to calibrate a HART zero trim. This adjusts the in-
transmitter (the standard prac- strument input block so that the As published in Intech, May 1996 and also in
tice associated with conven- digital PV agrees with the cali- HART Book 8, July 1998. Reprinted with the
tional transmitters) often bration standard. If you intend
permission of the author.
corrupts the internal digital to use the digital process values
readings. You may not have for trending, statistical calcula-
noticed this if you never use a tions, or maintenance tracking,
communicator to read the range then you should disable the
or digital process data. As external zero and span buttons
shown in Figure 2, there is and avoid using them entirely.
more than one output to con-
sider. The digital PV and Loop current adjustment
milliamp values read by a com-
municator are also outputs, just Another observed practice
like the analog current loop. among instrument technicians
Consider what happens is to use a hand-held commu-
when using the external zero nicator to adjust the current
and span buttons to adjust a loop so that an accurate input
HART instrument. Suppose that to the instrument agrees with
an instrument technician in- some display device on the
stalls and tests a differential loop. If you are using a
pressure transmitter that was Rosemount model 268 commu-
set at the factory for a range of nicator, this is a “current loop
0 to 100 inches of water. Test- trim using other scale.” Refer
ing the transmitter reveals that again to the zero drift example
it now has a 1 inch of water just before pressing the zero

15 Fluke Corporation Hart Transmitter Calibration


Documentation of results
The scheduling of calibrations, creation of proce-
dures and documentation of your calibration
results are facilitated by a number of instrumen-
tation management software packages:

DocuMint ™

PRM (Plant Resource


Manager) from Yokogawa
Electric Corporation.

OnTimeSupport ™

Fluke DPC/TRACK ™

ProCALV Meridium

Intools

Ordering Information
FLUKE-744 Documenting Process Calibrator-HART
Includes: TL224 Industrial Test Leads (two sets), AC220
Test Clips (2 sets), TP220 Test Probes (1 set), BP7235
NiMH Battery Pack, BC7217 Battery Charger, serial port
cable, HART communications cable, DPC/TRACK
Sample with free PC communication utility software,
Instruction Manual, HART User’s Manual, NIST-
traceable calibration certificate and data, three-year
warranty.

Optional Accessories
Fluke-700 Pxx Pressure Modules
Included with each Fluke Pressure
Module: BSP-ISO to NPT Adapter(s)
(except with P29 - P31), Instruction
Sheet, NIST traceable calibration
Fluke. Keeping your world
report and data, one-year warranty. up and running.
Fluke-700BCW Bar Code Wand Fluke Corporation
Fluke-700PTP Pneumatic Test Pump, 360 psi/25 bar PO Box 9090, Everett, WA USA 98206
ANAG E M E NT

Fluke-700TC1 TC Mini-Plug Kit, 9 types Fluke Europe B.V. IT


YM S
YS
AL

T
QU

EM

Fluke-700TC2 TC Mini-Plug Kit, JKTERS PO Box 1186, 5602 BD ISO 9001


BE9005 Battery Eliminator Eindhoven, The Netherlands
01
CE

90

TI
R

FI O
ED IS
TO M E E T

BC7217 Battery Charger For more information call:


BP7217 NiCd Battery Pack U.S.A. (800) 443-5853 or Fax (425) 446-5116
Europe/M-East/Africa +1 (31 40) 2 675 200 or
BP7235 NiMH Battery Pack Fax +1 (31 40) 2 675 222
C700 Hard Carrying Case Canada 1-800-36-FLUKE or Fax (905) 890-6866
C781 Soft Carrying Case Other countries +1 (425) 446-5500 or
C789 Soft Carrying Case Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2004 Fluke Corporation. All rights reserved.
Printed in U.S.A. 12/2004 1262439 A-US-N Rev F
16 Fluke Corporation Hart Transmitter Calibration

S-ar putea să vă placă și