Xylem Analytics Germany GmbH

Sensortechnik Meinsberg
Kurt-Schwabe-Str. 6, 04736 Waldheim, Germany
Tel +49 34327 6230, Fax +49 34327 623-79
info@meinsberg.de www.meinsberg.de 12/2016

CS4 – free chlorine sensor


indicator Free chlorine with reduced pH-dependence
e. g. swimming pool water, drinking water, sea water
Application Surfactants (tensides) are partially tolerated.
Membrane covered, amperometric potentiostatic 3-electrode
Measuring system
system with electronic inside
- current output analog
- not galvanically isolated electronics
Electronic - output signal: analog (analog-out/analog)
- only valid for sensors with 2-pole-terminal and
CE-mark: poti for (restricted) adjustment of measuring range
Slope of a sensor can vary production-related or application-
related between 65% and 150% of the nominal slope

-> Recommendation to determine the suitable measuring range


Information about the
or the suitable sensor:
measuring range of
Concentration to be measured x factor 1.5 = measuring range
sensors with 4-20 mA
of the sensor

Example: Concentration to be measured 1.6 ppm x 1.5 = 2.4


-> recommended sensor with a measuring range of 5 ppm
Accuracy - Measuring range 2 mg/l: at 0.4 mg/l <1%
after calibration at repeatability at 1.6 mg/l <1%
conditions (25°C, pH 7.2 in drinking - Measuring range 20 mg/l: at 4 mg/l <1%
water) of the upper full scale
at 16 mg/l <3%
Working temperature 0 – <45 °C (no ice crystals in the measuring water)
Temperature
Automatically, by an integrated temperature sensor
compensation
Max. allowed working
3 bar, no pressure impulses and/or vibrations
pressure
Flow rate approx. 15-30L/h, small flow rate dependence is given
pH-range pH 4 – pH 9, reduced dependence on pH-value
Run-in time First start-up approx. 2 h
Response time T90: approx. 2 min
Zero point adjustment Not necessary
Slope calibration At the device, by analytical determination, DPD-1-Method
Connection 2-pole terminal (2 x 1 mm²)
Microporous hydrophilic Membrane, PVC-U, stainless steel
material
1.4571
diameter: approx. 25 mm
Size
Length: approx. 220 mm (2-pol-terminal)
Probe: Frost-protected, dry and without electrolyte
no limit
at >5 - <40 °C
storage Membrane cap: Used membrane caps can not be stored!
Electrolyte: in original bottle protected from sunlight
min. 1 year
at >10 - <35 °C
Regularly control of the measuring signal, min. once a week
The following specifications depend on the water quality:
maintenance
Change of the membrane cap: once a year
Change of the electrolyte: once a year
Measuring range:

Measuring
resolution Nominal
range Output Voltage supply
in ppm slope
in ppm
CS4MA2 0.01…2.00 0,01 8.0 mA/ppm
CS4MA5 0.01…5.00 0,01 3.2 mA/ppm 12…30 VDC
CS4MA10 0.01…10.00 0,01 4…20 mA 1.6 mA/ppm RL 50Ω…900Ω
uncalibrated
CS4MA20 0.01…20.00 0,01 0.8 mA/ppm
CS4MA-200 0.03… 20.00 0,01 0.08 mA/ppm

www.wtw.com www.si-analytics.com www.ebro.com www.meinsberg.de


Xylem Analytics Germany GmbH Amtsgericht München HRB 145685 Bank: Deutsche Bank AG, BLZ 700 700 10, Kto.-Nr. 201585700
Dr.-Karl-Slevogt-Str. 1 USt-IdNr./ VAT No.: DE275429984 IBAN: DE16700700100201585700
82362 Weilheim, Germany Managing Directors: BIC: DEUTDEMMXXX
Dr. Robert Reining · Ulrich Schwab · Matthias Rohwedder Steuernummer: 119/116/90391

S-ar putea să vă placă și