A59439 AA

July 2008
Combination Calomel pHree Electrodes pH/Reference/ATC (3-in-1) Electrodes
A51705 12mm x 150mm epoxy sealed 511052 12mm x 150mm 3-in-1 electrode for pHi
A51706 12mm x 150mm glass sealed series 2xx, 3xx, 6xx meters
511053 12mm x 150mm 3-in-1 electrode for pHi
A51707 12mm x 150mm epoxy refillable
series 2xx, 3xx, 6xx (with waterproof
Beckman Coulter Sealed and Refillable Electrodes
A51708 12mm x 150mm glass refillable
connector) pH/Ref/ATC (3-in-1) Electrodes
A51709 9.5mm x 300mm epoxy sealed
A57198 12mm x 150mm 3-in-1 electrode for pHi Combination Ag/AgCI Electrodes
A51710 9.5mm x 300mm glass sealed series 4xx and 5xx Combination Calomel pHree Electrodes
A51711 6mm x 150mm epoxy sealed Combination Low Ionic Strength Electrodes
A51712 6mm x 150mm glass sealed
pH Buffer Solutions Combination pHRESH Research-Grade Electrodes
582517 pH 4.00 Buffer, Red, 6 x 475 ml bottle pH and Reference Electrodes
Combination Ag/AgCI Electrodes 582521 pH 7.00 Buffer, Green, 6 x 475 ml bottle
Beckman Coulter has a comprehensive offering of 3-in-1, combination, pH, reference, and specialty
A57177 12mm x 150mm epoxy sealed 582525 pH 10.01 Buffer, Blue, 6 x 475 ml bottle electrodes in epoxy or glass bodies. Refillable or sealed models offer the same superior performance.
A57186 12mm x 150mm epoxy refillable 566575 pH assortment:
A57178 12mm x 150mm double-junction epoxy 3 x 475 ml pH 7.00 Helpful Operating Tips Figure 1. Removing the soaker bottle
sealed 2 x 475 ml pH 4.00 1. Depending on the model, your new Beckman
1 x 475 ml pH 10.01 Coulter electrode is shipped in either a soaker
A57184 12mm x 150mm double-junction flat bulb
epoxy sealed 3005 pH 4.00 Buffer, powdered, 12 pack storage bottle or a protective cap. Remove these
3006 pH 6.86 Buffer, powdered, 12 pack from the electrode, and retain for storage when
A57185 9.5mm x 150mm double-junction
the electrode is not in use (See Figure 1).
spear-tipped epoxy sealed 3007 pH 7.00 Buffer, powdered, 12 pack
2. If using refillable models, ensure fill solutions are
A57181 12mm x 150mm glass refillable 3008 pH 7.41 Buffer, powdered, 12 pack
within 2.5 in. (6.4 cm) of the fill hole. Ensure fill
A57179 9.5mm x 300mm epoxy sealed 3009 pH 9.18 Buffer, powdered, 12 pack hole is open during use to prevent vacuums.
A57182 9.5mm x 300mm glass refillable 3019 pH 10.01 Buffer, powdered, 12 pack Sealed models require no special attention.
A57180 6mm x 150mm epoxy sealed 3010 pH 12.45 Buffer, powdered, 12 pack 3. Combination electrodes combine pH and
A57187 6mm x 150mm epoxy refillable reference cells in one electrode. Use of pH
Refill Solutions
A57183 6mm x 150mm glass refillable electrodes require use of reference electrodes.
A51713 Calomel pHree refill solution, 4 x 118 ml Attach appropriate connector to mating
bottle connection on meter. If using a pH/Ref/ATC
pH and Reference Electrodes
A60217 Ag/AgCI refill solution, 4 x 118 ml bottle (3-in-1) electrode, an additional connection is
A57188 pH measuring half cell, epoxy
A58729 Low ionic strength refill solution, provided for the temperature connection. If using
A57189 pH measuring half cell, glass Refillable Models
4 x 118 ml bottle a separate temperature sensor, attach it to the
A57190 Calomel pHree reference only, epoxy appropriate mating connection. The fill holes should be open when measurements
511469 pHRESH refill solution, 1 x 120 ml bottle
sealed are being made, when the junction is being renewed
4. Vigorously stir the electrode in the sample buffer
A57192 Calomel pHree reference only, glass or the electrode is being refilled. It should be closed
NOTE or rinse solution. This action will bring the
refillable at other times. The reference solution level should
Please visit www.beckmancoulter.com for a solution to the surface of the electrode more
not be allowed to drop more than 2.5 in. (6.4 cm)
A57193 Ag/AgCI reference only, epoxy sealed complete list of electrochemistry meters, quickly and improve speed of response.
below the refill line in the electrode.
A57194 Ag/AgCI reference only, glass refillable electrodes, and accessories. 5. When calibrating, use a buffer close in value to
that expected from the sample. This will minimize Standard models use a sleeve to cover the fill hole.
span errors. Before use, slide the sleeve to open the fill hole and
Combination Low Ionic Strength Electrodes NOTE
allow air flow for measurement.
A57195 12mm x 150mm epoxy refillable For MSDS information, please visit 6. Keep buffers and samples at the same
temperature. This will eliminate the need to The pHRESH series electrodes feature Rapid-Renew
A57196 12mm x 150mm glass refillable www.beckmancoulter.com and click on
correct values for temperature effects. Reference Junctions. To open the sleeve-type
Customer Support, then MSDS.
reference junction, press down on the spring-loaded
Combination pHRESH Research-Grade 7. pH readings stabilize faster in some solutions than
Top Cap. While the junction is open, the solution
Electrodes NOTE in other solutions. Allow time for reading to
inside the reference flows out of the electrode. This
511110 12mm x 150mm epoxy refillable If you experience problems with your Beckman stabilize. In general, buffers provide stable
flow flushes out contaminants. When the Top Cap is
Coulter electrode that these procedures cannot readings within several seconds.
511111 12mm x 150mm glass refillable released, a new reference junction is formed. These
resolve, call your local Beckman dealer. In the electrodes also feature a Twist-Seal design
U.S., please call the toll-free Product Support reference solution fill hole. The hole is opened or
line at (800) 742-2345. closed by rotating the knurled knob.

©2008 Beckman Coulter, Inc. 1


4 4300 North Harbor Boulevard, Fullerton, CA 92834-3100
Printed in USA.
Calibration Procedure Figure 2. Calibrating the electrode Electrode Reconditioning
As a rule, follow the procedures recommended by the When reconditioning is required due to electrode aging, the following chemical treatments may be tried.
pH meter manufacturer, keeping in mind the Helpful They are presented in the order of the severity of attack on the pH glass and may not improve, and in some
Operating Techniques in this guide. The cases may actually deteriorate, electrode performance.
frequency of calibration is a function of the
Figure 4. Reconditioning the electrode
electrode, the pH meter, and the solutions to which WARNING
the electrode is exposed. The electrode and meter
should always be calibrated together, with the Use proper precautions when handling these
calibration frequency determined by experience. Use hazardous chemicals. Ammonium bifluoride
the following procedure for both calibration in and hydrofluoric acid (HF) are extremely
buffers and for sample measurements (See Figure 2): hazardous and should only be used by qualified
1. Remove the electrode from its soaker bottle or personnel.
cap. Save the bottle or cap for reuse.
警告
2. Vigorously stir the electrode in a rinse solution.
3. Shake the electrode with a snap action to remove 请在处理这些有害化学物质时采取适当
residual drops of solution. 的预防措施。氟化氢铵和氢氟酸 (HF)
4. Vigorously stir the electrode in the buffer or 极为有害,仅限合格的人员使用。
sample and allow the electrode to rest against the
wall of the beaker. 1. As shown in Step 1, immerse the electrode tip in
0.1 M HCI for fifteen seconds, rinse in deionized
5. Allow reading to stabilize and take the reading.
water, then immerse tip in 0.1 M NaOH for fifteen
6. Repeat these steps for each sample or buffer seconds. Rinse the electrode in deionized water
determination. again. Repeat this sequence three times, then
Electrode Storage recheck the performance of the electrode. If
performance has not been restored, proceed to
When pH readings are made infrequently (several
Step 2.
days or weeks apart, for example), the electrode can
be stored by replacing it in its soaker bottle or cap. 2. Immerse the tip of the electrode in a 20% solution
First, slide the cap onto the electrode. Next, slide the of ammonium bifluoride (NH4F•HF) for two to
o-ring onto the electrode, insert the electrode into three minutes, rinse in deionized water, and
the bottle, and tighten the cap. If storage solution is recheck performance. If performance has not
not available, fill the bottle with pH 4 buffer. been restored, proceed to Step 3.
3. Immerse electrode in 5% HF for ten to fifteen
Electrode Cleaning Figure 3. Cleaning the electrode seconds, then rinse thoroughly in deionized water
Coating on the measuring surface can lead to (Step 3a). Rinse in 5N HCI, rinse thoroughly in
erroneous readings, including shortened span deionized water, and recheck performance
(slope). Never use a brush or abrasive on any pH (Step 3b). If performance has not been restored, it
electrode (see Figure 3). Instead, use the following is time to replace your electrode.
techniques to clean the electrode measurement bulb:
1. Soft coating can be removed by vigorous stirring WARNING
or by using a squirt bottle filled with water.
Always follow laboratory safety procedures,
2. Organic, chemical, or hard coatings should be
including wearing protective gloves and eye
removed by soaking the electrode in a 5-10%
hydrochloric acid (HCI) solution for a few wear. Avoid consuming, inhaling, or coming
into direct contact with any chemicals or
minutes.
solutions.
3. Grease, oil, or other surfactants can be removed
by rinsing with a solution of 50% acetone and
警告
50% isopropyl alcohol. Then rinse with deionized
water and soak in pH 4 buffer for one hour. 始终按照实验室安全程序操作,包括佩
4. Protein can be removed by soaking the electrode 戴防护手套和眼罩。避免吞食、吸入或
in a 10% bleach solution (9 parts deionized water 直接接触化学物质或溶液
to 1 part bleach) for fifteen minutes. Rinse in
deionized water and soak in pH 4 buffer for fifteen
minutes. Rinse again in deionized water.