Documente Academic
Documente Profesional
Documente Cultură
Introduction
1.1 Introduction
Some amount suspended particle can be measured when it is stable enough for analysis. When a
light of beam is passed through a turbid sample, some light were scattered, reflected, refracted, or
transmitted by those suspended particle. These four mechanisms is a concept analysis on turbidimetry
and nephelometry. The light intensity that was passed through or scattered (opalescence) is a function
of suspended particle concentration. For diluted concentration, both methods can give a good turbidity
result. Turbidity is commonly known in NTU (nepheloturbidity unit) or FTU (formazinturbidity unit).
The influence of suspended particle optical nature caused by concentration and type of compound in
dispersion phase, so that in this condition will be semi-empirical. Therefore, it is necessary to make
calibration curve in quantitative analysis (nephelometry and turbidimetry). In order to have good
reproducibility, high quality of sample storage and preparation are required. One of important condition
to consider in suspension preparation is that the particle should be really fine, thus it is not easily
vaporized. Light intensity of irradiation is highly affected by the size and stability of suspension, thus
if it is well prepared, then it is possible to obtain quantitative anlaytical results. This method is able to
determine anion sulphate quantitatively by turned it into discrete suspension using Ba2+ cation. The
formation of BaSO particles that are finely and stable may show good result in turbidimetry quantitative
analysis. The concept analysis between turbidimeter and spectrophotometer is very different. In a
spectrophotometer, light intensity was determined by absorbance value. Thus, this apparatus is suitable
to use on sample solution (nor suspension or colloid). UV-vis spectrophotometer is commonly used on
a clear and transparent solution. On some circumstance, turbidity can be measured using uv-vis
spectrophotometer. However, repetitive procedures have to be performed to ensure its accuracy and
stability
1.2 Objective
1) Students understand the analysis technique of discrete particles as turbidity, and able to
perform step by step of turbidimetry analysis
2) Students are understand the concept of a beam of light passed through a sample and its
equivalencies to the turbidity
3) Students are able to correctly use and choose laboratory apparatus for turbidimetry analysis
and can distinguish the function of turbidimeter and spectrophometer
Transmissometry and nephelometry are two of the most common optical metrics used in
research and monitoring of the Earth’s oceans, lakes, and streams. Both of these measurements
relate to what we perceive as the clarity of the water, and both provide vital information in
numerous studies of natural processes and human activities’ impact upon water bodies.
Applications involving these measurements range from monitoring drinking water suitability to
understanding how carbon is transferred into and transported within ocean waters. Nephelometry
refers to measurements made by optical scattering sensors, often referred to as turbidity sensors or
nephelometers. These sensors project a beam of light into the water and measure the radiant flux
of light scattered into the direction of a receiver. Since the receiver signal increases with greater
numbers of particles, the device infers the concentration of suspended particles in the water.
(C. Moore, 2001)
Spectrophotometer
turbidimeter
Turbidimeter turned on
Distilled water
Distilled water is added into 50mL volumetric flask
0,01M K2SO4
2N HCl
2N HCl
glycerol-alcohol
solution (1:2)
10mL of glycerol-alcohol solution is added into each flask
glycerol-alcohol
solution (1:2)
0,1% BaCl2
1. 2mL of BaCl2 is added into each flask
2. The contents of the flask are swirled to assure adequate mixing
3. The volume is made up to the bottom of the solution meniscus lines up
exactly with the volume mark by adding distilled water
Turbidity
Characteristics of aquadest:
Obtain 5 units of 50 mL Colorless
1. volumetric flask and add 2 mL
Clear
of aquadest into the first
Odorless
volumetric flask
Normal temperature
Characteristics of K2SO4:
Liquid
Colourless Take 0.01 M K2SO4 solution
Take and fill other flasks with
2. Odorless
0.5, 1.0, 1.5, and 2.0 mL of 0.01
M K2SO4 solution Normal temperature
-Take it using pro pipette and
measuring pipette.
-Add it respectively.
Add it carefully.
5. Add some amount of aquadest
Add it to maximum line of
into each flask
dilution
Mix by swirling or
6. inverting the contents until Mix to homogeneity by
substrate are completely gentle orbital shaking
dissolved
Characteristics of MgSO4:
Liquid.
Transfer it into a beaker
15. Colorless.
glass
Odorless.
Normal temperature.
Characteristics of MgSO4:
Measure the turbidity of 1 Liquid.
16. mL of MgSO4 (sample) Colorless.
using turbidimeter. Odorless.
Normal temperature.
The turbidity result of
MgSO4 (13,2 NTU).
Characteristics of MgSO4:
Measure the absorbance of 1 Liquid.
17. mL of MgSO4 (sample) Colorless.
using spectrophotometer. Odorless.
Normal temp.
The absorbance result of
MgSO4 (0,021).
3.2. Discussion
The 7th practicum conducted on Thursday, 16th of November 2017 in Water Treatment
Laboratory, Environmental Engineering Department, ITS at 08.30 AM until 12.00 PM. In this
exercise, we learned the basic principals of spectrophotometry and serial dilution and their
practical applications. A spectrophotometer is a very powerful tool used in both the biological
and chemical sciences yet operates by simply shining a beam of light, filtered to a specific
wavelength (or very narrow range of wavelengths), through a sample and onto a light meter.
Some basic properties of the sample can be determined by the wavelengths and amount of light
absorbed by the sample. The instruments we use in lab are: measuring pipette 5 mL and 10
mL, propipette, pipette dropper, wash bottle, turbidimeter, uv-vis spectrophotometer,
volumetric flask 50 mL and 100 mL, and beaker glass. While the materials requirement are
Aquadest, 0.01 M of K2SO4, 10% of BaCl2, acidic acid, MgSO4.
Due to recent advances in science and technology, the detection of specimen color or
wavelength using a spectrophotometer, colorimeter, or turbidimeter is frequently used in most
laboratories and hospitals. The color measuring tool is the basic measuring equipment essential
to most research activity fields, such as physics, biotechnology and food engineering (Kim,
2015).
Turbidity is a commonly used parameter to determine water quality and is used to quantify
water clarity. Turbidity can be easily measured using turbidimeter and the turbidity reading is
simple and fast. The nephelometric turbidity method of determining TSS concentration is
based on the theory that light scattering increases with the concentration of particles. Thus,
turbidity level has the potential to estimate the concentration of TSS. However, turbidity is
also dependent on other factors such as the size, shape, colour, and reflectivity of the particles
(Daphne, 2011).
First of all, we have to prepare the solution. Here, we will have 6 solutions, there are :
aquadest, 0.5 mL K2SO4, 1.0 mL K2SO4, 1.5 mL K2SO4, 2.0 mL K2SO4, and MgSO4. We take
wash bottle and pour aquadest into first volumetric flask. Measure aquadest as much as 2 mL
using measuring pipette and propipette. Take other volumetric flasks filled with 0.01 M K2SO4
in the amount of 0.5 mL, 1 mL, 1.5 mL, and 2 mL respectively. K2SO4 are taken and moved
using propipette and measuring pipette. Aquadest is colorless, has liquid texture, and no smell.
K2SO4 is colorless, has liquid texture, and no smell.
All of the solutions in each flask are added with salt acid in the amount of 2 mL. Salt acid
is used to replace HCl and glycerol alcohol which are catalisator to increase the rate of solution
and used to stabilize the suspension and ensure the suspense spread evenly, not forming
precipitate, so that the turbidity can be measured. Salt acid is taken and moved using propipette
and measuring pipette. The texture of salt acid is liquid, it has no color and no smell. After
added by salt acid, the physical properties of solutions in each flask stay the same. Add 2 mL
of 10% BaCl2 into each flask. Mix all the solution until it is homogenous so that we will obtain
an accurate result. After the addition of BaCl2 the solution becomes slightly turbid.
The reaction between K2SO4 and BaCl2 is as follows:
𝐾2 𝑆𝑂4(𝑎𝑞) + 𝐵𝑎𝐶𝑙2(𝑎𝑞) ⇄ 𝐵𝑎𝑆𝑂4(𝑠) + 2𝐾𝐶𝑙(𝑎𝑞)
From that equation we can see the amount of BaSO4 precipitate is equal to the amount of
K2SO4 that is used in this experiment.
To find the mole of K2SO4 we should use the following formula:
a. 0,5 mL K2SO4 0,01M :
Mol = 0,01 x 5.10-4
= 5 x 10-6
b. 1 mL K2SO4 0,01M :
Mol = 0,01 x 10-3
= 10-5
c. 1,5 mL K2SO4 0,01 M :
Mol = 0,01 x 1,5.10-3
= 1,5 x 10-5
d. 2 mL K2SO4 0,01 M :
Mol = 0,01 x 2.10-3
= 2.10-5
Mol BaCl2 10% =
10
Mass = 100 x 1000 = 100gr
𝑚𝑎𝑠𝑠 100
Mol = x Vol = 207 x 2.10-3 = 9,6 x 10-4
𝑀𝑟
Aquadest is added into each flask until maximum line of dilution or miniscus line
and mixed until homogeneous. Aquadest is added to dilute all of the solutions in each flask.
After diluted with aquadest, all of the solutions become more turbid. After that, go to the
spectrophotometer and wipe the cuvette using tissue. Make sure the cuvette is in a clean
condition because if not it could affect the result. Should always be reminded that don’t
touch the rough sides. Then, don’t forget to calibrate the spectrophotometer first before
putting the other solutions. Insert the cuvette into the sample holder and put it inside the
spectrophotometer, then begin measuring the solution by pressing the read button. So after
calibration with aquadest, proceed with the 0.5 mL solution, then calibrate again then
proceed with 1 mL solution and so on. Wait until the number is stable, then note the result.
Record all data that we already have in tabel:
Vol K2SO4 (mL) Kadar BaSO4 (ppm) Turbidity (NTU) Absorbance (A)
0,5 23,3 4,48 0,13
1 46,6 7,69 0,19
1,5 69,9 9,8 0,21
2 93,2 18,3 0,28
A calibration curve displaying Turbidity vs. Concentration was created using Excel
by using the increasing concentrations of the five standard solutions for the x values, and
their corresponding turbidity for the y values.
Turbidity
20 18.3
18
y = 0.1858x - 0.69
16 R² = 0.9058
14
12
Turbidity (NTU)
9.8
10 7.96 Y-Values
8
Linear (Y-Values)
6 4.48
4
2
0
0 20 40 60 80 100
Amount of BaSO4(ppm)
Absorbance
0.3
y = 0.002x + 0.085
0.25 R² = 0.9625 0.28
0.2
Absorbansi (A)
0.21
0.19
0.15
Series 1
0.1 0.13 Linear (Series 1)
0.05
0
0 20 40 60 80 100
Amount of BaSO4(ppm)
The plot above is a calibration plot from the data recorded. It shows a linear
relationship between absorbance and concentration. The higher the BaSO4 concentration,
the greater the absorbance value. This is because as the concentration increases, there are
more molecules in the solution, and more light is blocked. We have a straight line equation
that relates absorbance to concentration, y = 0.002x + 0.085, where y represents absorbance
value, and x is concentration value. This is the line of best fit through the data.
Both graph are plotting concentration in ppm, but the first graph plots is vs.
turbidity and the second graph plots is vs. absorbance. They are directly proportional to its
concentration. We can see from the data above, as the concentration increases, the
absorbance also increases as well as turbidity value. And if we are looking for “Best-Fit
Line", we will see that the best line that represents the data is linear/straightforward.
Lastly, substitute the absorbance and turbidity of the solution of unknown
concentration (MgSO4) into the equation determined as y and solve for x, where x
represents concentration.
The calculation for the MgSO4 Solution is:
Turbidimeter Spectrophotometer
𝑦 = 0.1858𝑥 − 0.69 𝑦 = 0.002x + 0.085
13,2 = 0.1858𝑥 − 0.69 0,021 = 0.002x + 0.085
13,89 = 0,1885𝑥 −0,064 = 0,002𝑥
𝑥 = 73,6 𝑝𝑝𝑚 x = −32 𝑝𝑝𝑚
The results were not quite as expected, since the data was askew due to a great
amount of experimental error in Part 1 of the lab. This error may occurred from not adding
the correct amount of solutions to each beaker, throwing off the absorption rate and then
the calibration curve. The absorbances of each of the five solutions being wrong also
affected the linear equation obtained as well as the linear equation being not a perfect
straight line. The impact of this experimental error in Part 1 affected the rest of the lab, not
allowing for perfect results. However there are common errors in calibration plots such as
spectrophotomer is not calibrated, abs readings are incorrect, diluted samples are prepared
incorrectly or contaminated, inappropriate wavelength chosen for calibration graph, or the
calibration line is not a "best fit" line.
4. Conclusion
From the experiment, we can take the conclusion as follows:
1) Turbidimeters work based on the optical phenomena that occur when incident light through
water body is scattered by the existence of foreign particles which are suspended with it.
Students are able to perform the step by step procedure of turbidimetry. First, make the
solution that want to be measured then proceed to the turbimeter. Before measuring the
solution, we have to calibrate the turbidimeter first by aquadest solution. Press the ‘read’
button, and the result will be shown in NTU.
2) The particles suspended in the water will scatter a light beam focused on them. The
scattered light is then measured at various angles from the incident light path. This is now
accepted as a more precise measure of turbidity. If more light is able to reach the detector
it means there are many small particles scattering the source beam, less light reaching the
detector means fewer particles. The amount of light scattered is influenced by many aspects
of the particles like color, shape, and reflectivity.
3) The material that needed in this experiment are K2SO4 solution containing different volume
of K2SO4, there are; 0.5 mL, 1.0 mL, 1.5 mL, and 2.0 mL. Beside that, there are aquadest
and acid salt which used to dissolve the solution. Turbidimetry is method for determining
the amount of cloudiness, or turbidity, in a solution based upon measurement of the effect
of this turbidity upon the transmission and scattering of light. Turbidity in a liquid is caused
by the presence of finely divided suspended particles. A spectrophotometer is employed
to measure the amount of light that a sample absorbs. The instrument operates by passing
a beam of light through a sample and measuring the intensity of light reaching a detector
5. In Depth Knowledge
1) Describe why alcohol was added in this experiment?
2) Is there any difference on the liniearity of both graphic standard? Explain. Describe the
similarity of concept analysis between turbidimeter and uvspectrophotometer!
3) Based on the literature, describe the schematic of turbidimeter and uvspectrophotometer!
Answer:
1) Alcohol was added in this experiment to stabilize the solution so that the suspended particle
in solution will not precipitate when reacted. If precipitation happens, it will interfere
measurement of absorbance value in spectrophotometer and turbidity in turbidimeter, thus
the accuracy of result will decrease.
2) There is difference on the linearity of both graphic standard because the concept analysis
of both apparatus is different. In spectrophotometer, light intensity is determined by
absorbance value. While the concept analysis of turbidimeter is based on four mechanism
when a light of beam is passed through a turbid sample, some light were scattered,
reflected, refracted, or transmitted by those suspended particle. The similarity of both
concept analysis is both are based on interaction of beam of light with the suspended
particle in solution.
3) The schematic of turbidimeter and uv spectrophotometer:
Spectrophotometer:
Turbidimeter:
Beam of light will go through fraction grating which work like a prism that separate the
lights into its component wavelengths, the grating is rotated so that only specific
wavelength of light reaches the exit slit. Then the light interacts with the sample. The
detector measure absorbance and transmittence (amount of light that passed completely
through the sample straight to detector). The detector senses the light being transmitted
through the sample and convert this information into digital display.
6. References
Grobbelaar, J.U. 2016. Turbidity. Encyclopedia of Inland Waters 1(1): 699–704
Orwin, John F. and C. Chris Smart. 2005. An inexpensive turbidimeter for monitoring
suspended sediment. Geomorphology 68(2): 3-15