Sunteți pe pagina 1din 4

Design Specifications

DS-M/T 3750A
August, 1998
Model 3750A (Supersedes issue dated April, 1997)

Ar-MiteTM Low Flow Armored Flowmeter


• Wetted metal parts 316L stainless steel
• Type NEMA 4X aluminum housing, epoxy
painted
• Front panel mounting
• Optional hermetically sealed reed switch
alarm
• Interchangeable scale - without removing
the meter from the pipeline

DESCRIPTION
The Brooks® Ar-MiteTM is a reliable, low flow metal tube
flowmeter with 316L stainless steel wetted parts. The
magnetically coupled indicator provides a highly reliable
method of indication. The reed switch alarm uses a
unique two sensor design, eliminating the need for
"biased" reed switches which can affect meter
performance. This model is a practical and economical
approach to low flow rate indication for high pressure
and difficult to handle fluids.
Model M/T 3750A
SPECIFICATIONS

WARNING
Do not operate this instrument in excess of the
Materials of Construction:
specifications below. Failure to heed this warning
can result in serious personal injury and/or damage
to the equipment. Meter Tube Assembly
316L stainless steel
Scales
Length - 55mm, nominal Clean-out Plug, Float Stop
Types - Direct reading scales in engineering units for air or 316L stainless steel
water at standard conditions; millimeter reference with
air or water calibration data sheet. Housing
Epoxy painted aluminum
Connections
1/4" female NPT, horizontal inlet and outlet Float
316L stainless steel (integral magnet)
Accuracy
±5% full scale
Valve Plug or Optional Valve
Repeatability 316L stainless steel (316 SS for NRS Valves)
1% full scale
O-rings
Ratings Standard: Viton®; Optional: Buna-N, Ethlylene Proplylene,
Maximum fluid pressure: 1500 psig (103 bar) Kalrez®, Teflon® (without valve)
Optional maximum operating pressure: 4000 psig (276
bar) no valve Float Stop Spring
Maximum fluid temperature: 400°F Inconel® 625

1
Figure 1 Dimensions

Figure 2 Mounting Dimensions

2
Meter Dimensions
Refer to Figures 1 and 2

Ordering Information and Model Code


Refer to Table 1

Capacities
Flow Range
Meter Water Air*
Size gph lph scfh slpm
1 0.035-0.35 0.13-1.3 0.2-2.2 0.1-1.0
2 0.095-0.95 0.36-3.6 0.5-5 0.2-2.3
3 0.28-2.8 1-10.6 1.3-13 0.6-6.0
4 0.55-5.5 2.1-21 2.4-24 1.1-11
5 1.1-11 4.2-42 5.5-55 2.6-26
6 2.6-26 10-100 12-120 5.6-56
*Air flows are at 70°F (21.1°C) and 14.7 psia (101.3kPa)

Optional Equipment FM
Cartridge Valve or Precision NRS Valve on inlet
Integrally piped flow controller (See DS-8800, 8900) APPROVED
Special scales
Reed Switch Alarm Figure 3 Model MT 3750A with Alarm

Specifications - Reed Switch Alarm Nonincendive, FM and CSA/NRTL/C Approved, CL l, ll, lll,
High, low or dual setpoints, latched output, over full range. Div. 2, Gps A-G
NEMA 4X Enclosure Max. Voltage 30 Vdc*
Maximum Fluid Temperature: 250°F Max. Current 100 mA*
Maximum Ambient Temperature: 125°F Max. Contact rating 3 Watts* (product of current and
voltage)
Intrinsic Safety, FM and CSA/NRTL*/C Approved, CL l, ll, *maximum switch contact specifications
lll, Div. 1, Gps A-G Specifications Subject to Change Without Notice
Entity Parameters, Vmax=30V, lmax=100mA, Li=0, Ci=0
FM and CSA/NRTL/C Approved IS Barrier required or FM TRADEMARKS
and CSA/NRTL/C Approved Power Supply/Relay Ar-Mite ................ Brooks Instrument Division, Emerson Electric Co.
Isolator available from Brooks. Brooks ................. Brooks Instrument Division, Emerson Electric Co.
*Canadian Standards Association/Nationally Recognized Testing Inconel ................................................. Inco Alloys International, Inc.
Laboratory accredited by the US Occupational Safety and Health Kalrez ........................................................... DuPont Dow Elastomers
Administration (OSHA) and Canada. CSA, FM, UL, ETL are all Teflon ................................................. E.I. DuPont de Nemours & Co.
currently approved NRTL test laboratories. Viton ............................................................. DuPont Dow Elastomers

Figure 4 Typical Alarm Circuit

3
Table 1 Ordering Information and Model Code
MODEL ARMORED PURGE METER
3750A 55 MM
| CODE METER SIZE
| 1 SIZE 1
| 2 SIZE 2
| 3 SIZE 3
| 4 SIZE 4
| 5 SIZE 5
| 6 SIZE 6
| | CODE MATERIALS OF CONSTRUCTION
| | A 316L STN. STL. - VITON O-RINGS
| | B 316L STN. STL. - TEFLON O-RINGS (NO VALVE UNITS ONLY)
| | C 316L STN. STL. - BUNA O-RINGS
| | D 316L STN. STL. - KALREZ O-RINGS
| | E 316L STN. STL. - EPM O-RINGS
| | F 316L STN. STL. - ALL WELDED, MUST BE NO VALVE UNIT (VALVE CODE A) AND HIGH PRESSURE (OPTION CODE M)
| | | CODE VALVE/CONTROLLER CONFIGURATION
| | | A NO VALVE OR CONTROLLER
| | | B LOW CARTRIDGE VALVE ON INLET TYPICAL FOR SIZES 1 & 2 (SEE NOTE 1)
| | | C MEDIUM CARTRIDGE VALVE ON INLET TYPICAL FOR SIZES 3 & 4 (SEE NOTE 1)
| | | D HIGH CARTRIDGE VALVE ON INLET TYPICAL FOR SIZES 5 & 6 (SEE NOTE 1)
| | | E MOUNTED TO 8802/8902 STN. STL. FLOW CONTROLLER W/ TEFLON DIAPHRAGM
| | | F MOUNTED TO 8805 STN. STL. FLOW CONTROLLER W/ TEFLON DIAPHRAGM
| | | G MOUNTED TO 8802/8902 BRASS FLOW CONTROLLER W/ TEFLON DIAPHRAGM
| | | H SIZE 1 NRS VALVE ON INLET
| | | J SIZE 2 NRS VALVE ON INLET
| | | K SIZE 3 NRS VALVE ON INLET
| | | L SIZE 4 NRS VALVE ON INLET TYPICAL FOR SIZES 1 & 2 (SEE NOTE 1)
| | | M SIZE 5 NRS VALVE ON INLET TYPICAL FOR SIZE 3 (SEE NOTE 1)
| | | N SIZE 6 NRS VALVE ON INLET TYPICAL FOR SIZES 4, 5 & 6 (SEE NOTE 1)
| | | P VALVE CAVITY ON INLET (NO VALVE ASSEMBLY OR PLUG)
| | | Q MOUNTED TO 8802/8902 STN. STL. FLOW CONTROLLER W/ BUNA-N DIAPHRAGM
| | | R MOUNTED TO 8805 STN. STL. FLOW CONTROLLER W/ BUNA-N DIAPHRAGM
| | | S MOUNTED TO 8802/8902 BRASS FLOW CONTROLLER W/ BUNA-N DIAPHRAGM
| | | | CODE SCALE INSCRIPTION
| | | | B MILLIMETER
| | | | C % OF MAXIMUM FLOW - AIR @ STP
| | | | D % OF MAXIMUM FLOW - WATER
| | | | E 0 - 100 , LINEAR
| | | | F SCFH AIR @ STP
| | | | G GPH WATER
| | | | H SLPM AIR @ STP
| | | | J LPH WATER
| | | | K SPECIAL INSCRIPTION , UNCALIBRATED
| | | | L SPECIAL GAS INSCRIPTION , CALIBRATED
| | | | M SPECIAL WATER INSCRIPTION , CALIBRATED
| | | | N SPECIAL OIL INSCRIPTION , CALIBRATED
| | | | | CODE CONNECTION ORIENTATION
| | | | | 1 INLET & OUTLET - BACK
| | | | | | CODE ELECTRICAL OUTPUT
| | | | | | A NONE (INDICATOR ONLY)
| | | | | | B REED SWITCH ALARM, METER MOUNTED ENCLOSURE
| | | | | | | CODE CONNECTION TYPE
| | | | | | | 1 1/4" FEMALE NPT, INTEGRAL
| | | | | | | 2 1/4" MALE COMPRESSION
| | | | | | | | CODE ACCESSORIES
| | | | | | | | A NONE
| | | | | | | | B FLUSH PANEL MOUNTING BEZEL (NOT AVAIL. W/ALARM)
| | | | | | | | C NO BROOKS IDENTIFICATION
| | | | | | | | D DEGREASE FOR OXYGEN SERVICE PER J834Z169AAA
| | | | | | | | E SPECIAL CLEANING PER J834Z172AAA
| | | | | | | | F PER NACE SPEC. MR-01-75
| | | | | | | | G CALIBRATE FOR NIST TRACEABILITY
| | | | | | | | H 115V RELAY, 1 CHANNEL
| | | | | | | | J 220V RELAY, 1 CHANNEL
| | | | | | | | K 115V RELAY, 2 CHANNEL
| | | | | | | | L 220V RELAY, 2 CHANNEL
| | | | | | | | M HIGH PRESSURE (4000 PSIG), MUST BE NO VALVE UNIT (CODE A)
| | | | | | | | |
3750A 1 A B F 1 A 1 A TYPICAL MODEL CODE

NOTE 1: VALVE SELECTION IS BASED ON AIR AT 10 PSIG INLET. SEE DS-BLIV OR DS-8503 FOR ADDITIONAL INFORMATION.

ISO 9001
QUALITY SYSTEM
Certificate of Conformity
The Quality System at Brooks Instrument
conforms to the quality standards set
forth in ISO 9001.

S-ar putea să vă placă și