The combustible gas Toxgard II • Real-time clock for time and date
LED display instructions for applying zero and
span gas.The Toxgard II Monitor automatically
event stamping.
Monitor requires just a simple makes any necessary adjustments. There’s
• Logging of minimum, maximum, no guesswork with the Toxgard II Gas Monitor,
connection to its remote sensor. and average gas concentrations over as actual gas readout is displayed
pre-determined time intervals. during calibration.
• Selectable output signal lockout
When calibration is complete, the
during calibration to inhibit alarms
Toxgard II Monitor automatically returns
and control actions.
to normal operation.
2
Strobes
Air-line Regulator
Federal Horn
Side-Mount
Battery Pack Sensor
LP4 Streamline®
Low Profile
Mini Sounder
Bottom-Mount Internal
Sensor Pump
Gas – simply select the gas to be monitored. Strobes – up to 2 top-mounted strobes Flow Cap – used whenever non-pressurized
Sensor location – while bottom mount is can be selected to activate from Warning gas sample is to be delivered to the sensor.
the typical selection, side mount option is and Alarm levels. Handle – for easy carrying of unit when
available particularly if Toxgard II Monitor Battery Backup – provides power if unit used as transportable monitor.
will be used as a transportable gas monitor. input power is interrupted. Also allows Key lock (not shown) - if it is necessary
Monitor can then be allowed to sit on any for unit transportable operation. to lock enclosure during operation for
flat surface during operation. Internal Pump – for applications where it is security purposes.
Audible Alarms – all units are standard with necessary to draw samples to the unit.
70 db sona-alert, 93 db Federal horn, or Air-line Regulator – allows the unit to
95dB sounder with selectable output tones monitor breathing air lines, typically for
can be selected as alternative choices. low levels of carbon monoxide.
3
An internal 2-button keypad allows access to all setup parameters Specifications
and stored values including calibration initiation. In addition,
2 types of remote communication devices are available for the Gas types Combustibles; oxygen; toxics
Toxgard II Monitor. The Toxgard II Calibrator and Controller both
Temperature Range:
provide a digitally encoded signal for non-intrusive calibration and -20˚ to +50˚C
Toxics and oxygen
access to advanced features. This patented technique eliminates -4˚ to +122˚F
(range on some models may differ)
tools as well as the need to open the enclosure during setup or
Temperature Range: -40˚ to +90˚C
calibration.
Combustibles -40˚ to + 194˚F
Calibrator Zero Drift < 5%/Yr., typically
The Toxgard II Calibrator offers the industry’s simplest method of Span Drift < 10%/Yr., typically
calibration.This easy-to-use, 3-button device allows for calibration Noise Less than 1% FS
of the Toxgard II Monitor. The Toxgard II Calibrator is available Accuracy: Repeatability ±1% FS or 2 ppm
for those who do not need to access the Toxgard II Monitor’s ± 2% FS
advanced features. (Combustible; O2; CO)
Accuracy: Linearity
± 10% FS or 2 ppm
Controller (others)
The Toxgard II Controller provides complete access to all features
through its full-function keypad. Step Change Response
T20 O2 and toxics < 12 sec. (Typ. 6 sec.)
The controller is used to: T20 O2 and toxics < 30 sec. (Typ 12 sec.)
• Set the real-time clock.
T50 combustibles
T90 combustibles
< 8 sec.
< 20 sec.
• Set alarm levels.
Humidity 0-95% RH, non-condensing
• Change span gas values.
Sensor Life
• Display date of last calibration. Combustibles 3 years typically
• Display minimum, maximum, Toxics and O2 2 years typically
and average gas values. Full replacement warranty 1 year
• Change address.
Hazardous Area Rating
General Purpose; XP Remote Sensor.
• Set future calibration time/date.
Assembly can be installed in Class I, Div I areas.
Power Input 110-220 VAC power input
In addition to providing basic calibration functions,
the controller is completely password-protected. Signal Output 4-20 mA: 2-wire current source
Internal Relays Fault, Caution, Warning, Alarm
Accessories Internal Relay Contact Rating 5 amp @ 125 VAC; 5 amp @ 30 VDC
Remote Sensor Assembly
Battery Back Up (12V lead acid)
Allows sensor to be separated from transmitter for difficult-to-reach Battery operating time with no
installations. options; minimal alarm activity
Toxics and O2 up to 24 hr.
Ordering Information Combustibles up to 10 hr.
All Toxgard II Monitors are manufactured using MSA’s unique
Assemble-To-Order (ATO) process, allowing you to select any listed Battery Back Up (12V lead acid)
options to ensure that your Toxgard II Monitor satisfies Battery operating time with
your individual requirements. full options; depending on
alarm activity
E-chem 1-4 hr..
Combustibles 0.5-2 hr.
Physical dimensions
Enclosure 10” W x 12” H x 5” D
Toxics and O2 13 lbs.
Note: This bulletin contains only a general description of the products shown. Corporate Center MSA Mexico Offices and representatives worldwide
While uses and performance capabilities are described, under no circumstances 1000 Cranberry Woods Drive, Phone 01 800 672 7222 For further information:
shall the products be used by untrained or unqualified individuals and not until Cranberry Township, PA 16066 USA Fax 52-44 2227 3943
the product instructions including any warnings or cautions provided have Phone 724-776-8600
MSA International
been thoroughly read and understood. Only they contain the www.MSAnet.com
Phone 724-776-8626
complete and detailed information concerning proper use and U.S. Customer Service Center Toll Free: 1-800-672-7777
care of these products. Phone 1-800-MSA-INST FAX 724-741-1559
Fax 1-800-967-0398
ID 07-0949-MC / October 2011 MSA Canada
© MSA 2011 Printed in U.S.A. Phone 1-800-672-2222
Fax 1-800-967-0398