Sunteți pe pagina 1din 1

Technical Data Sheet

Leica EG1160- Paraffin Embedding Center

Type Leica EG1160 Leica EG1160 Leica EG1160


Mark of conformity CE CE CE
Nominal voltage: 230V AC±10% 120V AC±10% 100V AC±10%
Nominal frequency: 50 Hz 60 Hz 50/60 Hz
Power input: 800 VA 800 VA 800 VA

Operating Temperature range: 18-35 °C/air conditioned room recommended


Relative humidity: < 60 %

Site requirements:
Distance of back of the unit to wall: 15 cm

Dimensions
Length 910 mm/35.83 inches
Depth: 575 mm/22.64 inches
Height: 320 mm/12.60 inches
Weight ca. 50 kg/110.13 lbs

Cooling System:
Cut-out pressure 25 bar
Safety factor 3
Refrigerant 160 g ± 5 g refrigerant R 134A
Compressor oil 150 ccm EMKARATE RL-15S

Capacities
Paraffin reservoir 3l
Cassette bath: approx.100 cassettes
Cassette tray 8 cassettes
Cold plate: 70 cassettes

Temperature/Dimension/Capacity

Functional unit Temperature Heated Separate Warm up tim Dimensions Storage


range separately tempera.Control prior to oper. Capacity
Paraffin reservoir 45-70 ° C + + 4h
± 1°C increments
Mold warmer 45-70 ° C + + 1h
± 1°C increments
Cassette bath 45-70 ° C + + 1.5 app. 100 cass.
± 1°C increments
Work Area 45-70 ° C + + 1h 270 x 230 mm
± 1°C increments minus cold spot
Cold Plate -5 °C - - 1h 450 x 285 mm 70 cassettes
17,72x 11.22 inches
Cold Spot -5 °C 70 x 60 mm

Foreceps holder right (3 holder) 70°C + -


Foreceps holder left (not basic instrument) (6) 70°C + -
Paraffin dispenser + pump 45-70 ° C + -

Stand: May 2006