User Manual ddRCompact™ Series

Chapter 4: Technical Data

Technical Data

Caution: US Federal Law restricts this device to use by or on the


order of a physician.

Mise en garde: Les lois fédérales des États-Unis limitent l'utilisation


de cet appareil à un médecin ou à un tiers agissant
sur son ordre.

Revision Date By Reason for Change

01 2008-10-14 SRCH / WOL First Official Release

This document is the English original version, edited and supplied by Swissray Medical AG.
All rights reserved by Swissray Medical AG.

2008-10-14 / WOL SHT_37_100_031_01_E 1/5


User Manual ddRCompact™ Series
Chapter 4: Technical Data

TABLE OF CONTENTS

Section Page

4 Technical Data .......................................................................................... 3


4.1 Dimensions................................................................................................. 5

2008-10-14 / WOL SHT_37_100_031_01_E 2/5


User Manual ddRCompact™ Series
Chapter 4: Technical Data

4 Technical Data
ddR System Stand
Minimum room height 2.60 m / 102.4”
Optimum room height 2.80 m / 110.2”
C-Arm rotation -35° … +135° motorized
Detector swivel range -45° … +45° motorized, Shift function
Central beam – floor - distance 46 cm - 190 cm / 18.1” – 74.8” motorized
Floor – detector – distance min. 63 cm / 24.8” motorized
Source image distance (SID) 103 cm - 180 cm / 40.6” – 70.9”
Tube rotation ddRCompact™ +12° … -12° motorized
ddRCompact™ Plus: +12° … -180° motorized

HD-3000™ Detector
Technology High Definition Silicon Solid State
Cesium iodide scintillator (CsI:TI)
A/D conversion 14 bit / 16’384 grey levels
Active detector area 43 cm x 43 cm / 17” x 17”
Spatial resolution up to 3.3 lp/mm
Active Pixel matrix 4096 x 4096 pixels
Pixel size 103 microns
Sensitivity up to 400 speed film/screen combination

Anti Scatter Grid


Type Fixed anti scatter grid, removable
Ratio 8:1, fiber interspaces 85 L/cm
Transmission Primary 74.5 %

X-ray HV Generator
Power 80 kW at 100 kHz
Automatic exposure control 3-field ionization chamber

X-ray Tube
Optitop 150/40/80 HC-100L Double focus, high speed rotating anode tube
Focal spots 0.6 / 1.0 mm
Heat capacity 763 000 HU

Collimator
Multileaf collimator AL 02 Operation motorized (remote controlled) or
manual
Rotation +/- 45° manual
Integrated internal filters 0.1 mm Cu / 0.2 mm Cu / 0.3 mm Cu
Total filtration (X-ray tube + collimator) 2.5 mm Al equivalent
DAP meter (dose area product) optional

2008-10-14 / WOL SHT_37_100_031_01_E 3/5


User Manual ddRCompact™ Series
Chapter 4: Technical Data

Full System Control Console


eXpert™3000
LCD system information on stand

Workstation
SwissVision® XG for ddRCompact™ with DICOM Worklist

Environmental Conditions
According to IEC / EN 60601-1, UL 60601-1

Operating
Ambient temperature +10°C to +40°C
Recommended room temperature +22°C to +25°C
Relative air humidity 30% to 75%, non condensing
Air pressure 700 hPa to 1060 hPa

Storage / Transport
Ambient temperature 0°C to +50°C
Relative air humidity 10% to 80%, non condensing
Air pressure 700 hPa to 1060 hPa

Main Supply
Power consumption ddR system 1 x 208 / 230 VAC, 10 Amps, 50 Hz - 60 Hz
Power consumption X-ray generator 3 x 400 VAC, 5 Amps (160 Amps peak) Europe or
3 x 480 VAC, 5 Amps (140 Amps peak) USA

Weight
Overall 720 kg / 1587 lb

2008-10-14 / WOL SHT_37_100_031_01_E 4/5


User Manual ddRCompact™ Series
Chapter 4: Technical Data

4.1 Dimensions

Certifications FDA (DHHS), UL, CSA and CE


Subject to technical modification. The units are manufactured according to
FDA, UL, CSA and CE requirements.

Our units and equipment’s must only be installed, serviced and put into operation by qualified
staff.

2008-10-14 / WOL SHT_37_100_031_01_E 5/5