Documente Academic
Documente Profesional
Documente Cultură
© 2010 - 2016 Natus Medical Incorporated or one of its subsidiaries. All rights reserved.
Natus is a registered trademark of Natus Medical Incorporated. All product names appearing on this document
are trademarks or registered trademarks owned, licensed to, promoted or distributed by Natus Medical
Incorporated, its subsidiaries or affiliates. All other trademarks are the property of their respective owners.
Natus Neurology Incorporated
Blank page.
Contact information
Technical Support
Natus Neurology Incorporated
3150 Pleasant View Road Domestic
Middleton, WI USA 53562-3530 Natus Neurology Incorporated
608-829-8500 3150 Pleasant View
1 800-356-0007 Middleton, WI USA 53562
Fax: 608-829-8589 1-800-356-0007
www.Natus.com madison.helpdesk@natus.com
www.Natus.com
Natus EU Authorized Representative
Natus Manufacturing Limited International
IDA Business Park Natus Neurology Incorporated
Gort, Co.Galway, Ireland Phone: 0049 (0) 180 501 5544
Fax: 0049 (0) 89 83942777
Email: service.europe@natus.com
www.Natus.com
Blank page.
Contact information.............................................................................................................................................iii
Blank page.
Blank Page
Safety precautions
The following items are suitable for the Intra-Operative patient environment: EEG main
unit, Nicolet VikingQuest Combo EEG, OR amplifiers - including Nicolet EEGwireless
32/64 channel amplifier, Nicolet Cortical Stimulator, Natus Neurology Incorporated cart,
Natus Neurology Incorporated isolation transformer, and all approved Natus Neurology
Incorporated cables for use with the system. Only Natus Neurology Incorporated
approved peripheral devices should be used with this system. Devices not specifically
approved by Natus Neurology Incorporated are not suitable for the Intra-Operative patient
environment.
Do not use extension cords to power the system. Extension cords can cause system to be
over the ground integrity and leakage current limits.
When used simultaneously with high frequency surgical equipment, burns may occur at
the site of any recording or stimulation electrode.
Electrodes of inadequate size or unsuitable application could provoke skin reactions or
burns. Natus Neurology Incorporated recommends the use of disposable, pre-gelled,
silver/silver chloride electrodes with a contact area of 20 mm diameter or larger. These
electrodes contain an adhesive that maintains the electrode skin position and contact
during electrical stimulation and contact during recording and electrical stimulation.
During direct nerve stimulation with the Nicolet VikingQuest EEG combo unit (i.e.,
stimulation of an exposed nerve or nerve root), it is recommended that a "low level
stimulator output" be used if available. If not, limit the duration to 100 us or less and the
intensity to 10 mA or less. Avoid using constant voltage.
Electrode burns to the patient are variable due to numerous conditions that must be
understood and monitored during the use of the equipment. Some variables to be aware of,
but not limited to, are:
• Time of exposure (All connections should be checked periodically when used in excess
of 30 minutes in one place.
• Concentration of currents (burns are caused by current density). Always use the
largest possible of contact surface.
• Patient condition.
Summary
This document is intended to inform you about some of the most commonly applicable
safety precautions as they relate to the use of Nicolet brand EEG monitoring systems for
Intra-Operative Monitoring applications. As more electrical and electronic devices are
introduced into this hostile environment, there will be the potential for harm to you and/or
the patient.
At Natus Neurology Incorporated, commitment to safety and quality is paramount. For
this reason, we are providing the following information about the use of Nicolet brand
EEG systems for neurophysiological monitoring during various surgical procedures.
Please refer to IOM Safety Guide and Card for additional details.
Introduction
There are a number of unique requirements related to the surgical monitoring environment
that must be satisfied before a monitoring system can be used safely for Intra-operative
Monitoring. These requirements concern the safety of both the patient and the person
operating the monitoring system. Remember to follow all the cautions and warnings for
your system. The reason we suggest using the OR head-box for intra-operative monitoring
is to provide the patient with protection from potential burns at the electrode skin tissue
sites caused by radio frequency currents from the electro-surgical units. Remember to
follow all the cautions and warnings for your system.
We believe that the risk of an accident or an injury to either person can be minimized if
you are knowledgeable about the safety issues and guidelines discussed on the following
pages.
Although some ESUs contain circuitry that will detect if the return electrode is not making
proper contact with the patient and will activate an alarm in such cases, we strongly
recommend that you independently verify the return electrode status. In the event of loss
of the return electrode connection, the RF signal from the ESU will seek another pathway
back to the ESU.
As a result, any and all electrodes (both stimulating and recording), which connect the
patient to the system, or any other device connected to the patient, could potentially
provide such a pathway. If the ESU signal flows through these EEG electrodes, then a
high density RF current may flow through the tissue to the EEG electrode interface and an
“electrode burn” can result.
Electrode burns to the patient are variable due to numerous conditions that must be
understood and monitored during the use of the equipment. Some variables to be aware of,
but not limited to, are:
• Time of exposure (All connections should be checked periodically when used in
excess of 30 minutes in one place.
• Concentration of currents (burns are caused by current density). Always use the
largest possible of contact surface.
• Patient condition.
Also, it should never be assumed that there is no possibility of an electrode burn
occurring when the system power is off. Even when the system is turned off, a passive
pathway for RF currents to flow may exist and allow the ESU signal to produce electrode
burns.
All system components that the operator can contact are either grounded or isolated. In
addition, the amplifier and electrical stimulator subsystems meet safety isolation
requirements to protect the patient. Natus Neurology Incorporated has designed the
system to minimize the risk of such pathways being established and causing burns.
Natus Neurology Incorporated has designed the system to minimize the risk of such
pathways being established and causing burns.
a. First, in every amplifier subsystem, the patient is safety isolated from earth
ground.
b. Second, special RF choke or resistive devices, which act to inhibit the flow of RF
currents through the recording electrodes, are included as part of Natus
Neurology Incorporated’s “O.R. approved Amplifiers and Head boxes.” Models
of these head boxes specifically designed for Natus Neurology Incorporated
products are:
• Nicolet EEG System:
1. C64 OR Amplifier
2. C64 OR/SSU Amplifier
3. Nic36 OR Headbox and Amplifier
4. V44 OR Headbox
5. Nicolet Cortical Stimulator
6. Nicolet EEG Wireless Amplifier 32/64 Channel
• Nicolet Cortical Stimulator:
As a stand alone unit
As stated previously, misuse of the ESU by members of the surgical team can also cause
burns at the electrode locations. For example, unnecessary and prolonged activation of the
ESU should be avoided to reduce the possibility of burns caused by RF leakage currents.
Also, activation of the ESU when the hand piece either is not making contact with the
patient (i.e., "open circuit" usage) or is applied very near (for example, less than 20 cm)
to any of the recording and stimulating electrodes should never be done.
It is also highly recommended to minimize electrode length before the protective
headbox. A maximum electrode length of 12 inches (30 cm) is recommended to
minimize RF leakage currents.
Most ESU manufacturers’ user documentation advises that the patient should not come
into direct contact with any grounded metal objects (for example, the surgical table frame,
instrument table, etc.). In some surgical procedures (for example, those involving the use
of non-insulated head frames), it may not be possible to prevent contact between the
patient and a grounded metal object. Use of extreme caution to maximize patient safety is
advised in these situations.
Never intentionally “ground” a patient by connecting them to a grounded metal object
or earth ground in order to overcome interference problems in your monitoring.
On occasion, two or more ESUs may be connected to a patient and in use simultaneously.
For dual monopolar ESU applications, the location of each ESU return electrode is
particularly important. ESU manufacturers advise that each ESU return electrode should
be positioned as close as possible to the respective site of surgery (that is, the point of
application of the ESU signal to the patient) and care should be taken to avoid the
possibility of the two ESU return electrodes touching.
Given that the responsibility of the set up and use of the ESUs lies with the surgical team,
assurance that these and other cautions against ESU misuse (as described by the ESU
manufacturers) are heeded and followed is not the responsibility of the person performing
neurophysiological monitoring. Being aware of these cautions, however, may help you to
minimize the risks of injury to the patient.
Current Limit
A current limit circuit controls the maximum amount of stimulating current that may flow
out of the Stimulus Control Unit. The current may be limited to ±15 mA. It may also be
decided in some intraoperative monitoring applications to switch between different
electrical stimulating sites.
Use of a custom-made or commercially-available stimulus switching apparatus for
these applications can introduce a safety risk to the patient in several ways.
Such devices typically are simply connected to the output jacks of the electrical stimulator
subsystem on a Nicolet brand system. Since these outputs are isolated for reasons of
patient safety, the switching device therefore must not provide any connection to earth
ground to avoid defeating the safety isolation. Natus Neurology Incorporated has
approved electrode and stimulus switching devices available.
The wiring and cabling associated with the switching device may also introduce additional
capacitance to earth ground and in turn increase current leakage, thereby making the
stimulator subsystem no longer conform to the safety standards. Even when packing and
cabling for the switching device is carefully managed, the device can still represent a
patient safety hazard because it may provide a low impedance path to earth ground for RF
currents. The use of a stimulus switching device can affect the system functionality
because additional capacitance in the switching device can act to distort the shape of the
electrical stimulus pulse being delivered to the patient, thereby altering its effectiveness as
a stimulus.
Use only Natus Neurology Incorporated-supplied and/or approved electrode and stimulus
switching devices. Please see the third paragraph on this page.
The applicable stimulation probe is the Natus Neurology Incorporated Bipolar
Stimulation Probe with 2.2mm diameter tips. The applicable stimulation electrodes are
any commercially available 2.3mm diameter Strip and Grid Electrodes. In the United
States only, the stimulation electrodes and stimulation probes must have been cleared to
market by the Food and Drug Administration (FDA).
Blank page.
Blank Page
Safety precautions
The following items are suitable for the patient environment: Main unit, Natus Neurology
Incorporated cart, Natus Neurology Incorporated isolation transformer, and all approved
Natus Neurology Incorporated transducers for use with the system. Only Natus Neurology
Incorporated approved peripheral devices should be used with this system. Devices not
specifically approved by Natus Neurology Incorporated are not suitable for the patient
environment.
Do not use extension cords to power the system. Extension cords can cause ground
integrity and impedance problems.
When used simultaneously with high frequency surgical equipment, burns may occur at
the site of any recording or stimulation electrode.
Electrodes of inadequate size or unsuitable application could provoke skin reactions or
burns. Natus Neurology Incorporated recommends the use of disposable, pre-gelled,
silver/silver chloride electrodes with a contact area of 20 mm diameter or larger. These
electrodes contain an adhesive that maintains the electrode skin position and contact
during electrical stimulation.
During direct nerve stimulation (i.e., stimulation of an exposed nerve or nerve root), it is
recommended that a "low level stimulator output" be used if available. If not, limit the
duration to 100 us or less and the intensity to 10 mA or less. Avoid using constant voltage.
Transcranial electrical motor evoked potential stimulation is not recommended for use on
patients that meet the following exclusion criteria:
a. History of head injury, stroke, seizures, epilepsy or other neurologic or
psychiatric disorders.
b. History of cerebral aneurysm (i.e., with an intracerebral vascular clip).
c. Subjects with any type of implanted biomedical device (i.e., pacemaker, infusion
pump).
d. Subjects with metal fragments or plates in their head.
Patient movement may occur as a result of transcranial electrical motor evoked potential
stimulation.
Electrode burns are variable that occur on patients due to numerous conditions that must
be understood and monitored during the use of the equipment. Some variables to be aware
of, but not limited to, are:
• Time of exposure (All connections should be checked periodically when used in
excess of 30 minutes in on place.
• Concentration of currents (Burns are caused by current density). Always use as large
of contact surface as possible.
• Patient condition.
Summary
This document is intended to inform you about some of the most commonly applicable
safety precautions as they relate to the use of monitoring systems for Intra-Operative
Monitoring applications. As more electrical and electronic devices are introduced into this
environment, there will continue to be the potential for harm to you and/or the patient.
In order to minimize these risks, you should be as informed as possible about the system
and its monitoring capabilities and to communicate effectively with members of the
surgical team and biomedical electronics personnel in your facility about the safety
requirements necessary to perform monitoring effectively.
Introduction
At Natus Neurology Incorporated, commitment to safety and quality is paramount. For
this reason, we are providing the following information about the use of Nicolet EP/EMG/
IOM systems for neurophysiological monitoring during various surgical procedures.
There are a number of unique requirements related to the surgical monitoring environment
that must be satisfied before a monitoring system can be used safely for Intra-Operative
Monitoring. These requirements concern the safety of both the patient and the person
operating the monitoring system. Remember to follow all the cautions and warnings for
your system.
We believe that the risk of an accident or an injury to either person can be minimized if
you are knowledgeable about the safety issues and guidelines discussed in the following
pages.
Although some ESUs contain circuitry that will detect if the return electrode is not making
proper contact with the patient and will activate an alarm in such cases, we strongly
recommend that you independently verify the return electrode status. In the event of loss
of the return electrode connection, the RF signal from the ESU will seek another pathway
back to the ESU.
As a result, any and all electrodes (both stimulating and recording), which connect the
patient to the system, or any other device connected to the patient, could potentially
provide such a pathway. If the ESU signal is able to flow through these electrodes, then a
high density RF current may flow through the tissue-to-electrode interface and an
“electrode burn” can result.
Also, it should never be assumed that there is no possibility of an electrode burn
occurring when the system power is off. Even when the system is turned off, a passive
pathway for RF currents to flow may exist and allow the ESU signal to produce electrode
burns.
Natus Neurology Incorporated has designed the system to minimize the risk of such
pathways being established and causing burns.
a. First, in every amplifier subsystem, the patient common is safety isolated from
earth ground.
b. Second, special RF choke or resistive devices, which act to inhibit the flow of
RF currents through the recording electrodes, are included as part of Natus
Neurology Incorporated’s “O.R. Electrode Headboxes.” Models of these head
boxes specifically designed for and Natus Neurology Incorporated products are:
• For the Nicolet EDX: HB-7 Headbox
• For the Nicolet VikingSelect: Nicolet brand ET 16A, ET 16B, and ET 208
Headboxes and the IES D9 Stimulus Switching Unit
• For the Nicolet Viking II, II3, III and IV: Nicolet brand ET 206, ET 207,
ET 208, and ET 216 Headboxes and the IES 9 Stimulus Switching Unit
• For Nicolet Endeavor O.R.: Adaptors and the Headboxes 1 & 2
• For Nicolet Endeavor CR: HB-1 and 2 Headboxes, SP1 and 2 Stimulus
Switching Unit
Use of these head boxes or in-line RF filter is mandatory in all Intra-Operative
Monitoring uses of the systems, and all recording electrodes must be plugged
into these head boxes. Never connect even a single electrode directly onto the
amplifier subsystem during any monitoring application.
c. Third, RF choke devices have also been included in the electrical stimulator
subsystem. Special stimulator switching output boxes, the IES-D9, IES- 16, SP-1
or SP-2 contain the RF chokes and are available for use on Nicolet Viking,
Nicolet EDX and Nicolet Endeavor CR systems. Use of this of approved
stimulator is mandatory any time these Nicolet brand systems are used for
intra-operative monitoring.
As stated previously, misuse of the ESU by members of the surgical team can also cause
burns at the electrode locations. For example, unnecessary and prolonged activation of the
ESU should be avoided to reduce the possibility of burns caused by RF leakage currents.
Also, activation of the ESU when the hand piece either is not making contact with the
patient (i.e., "open circuit" usage) or is applied very near (for example, less than 20 cm)
to any of the recording and stimulating electrodes should never be done.
It is highly recommended to minimize electrode length between the patient and the
protected headbox.
A maximum electrode length of 12” (30 cm) is recommended to minimize RF leakage
currents.
Most ESU manufacturers’ user documentation advises that the patient should not come
into direct contact with any grounded metal objects (for example, the surgical table frame,
instrument table, etc.). In some surgical procedures (for example, those involving the use
of non-insulated head frames), it may not be possible to prevent contact between the
patient and a grounded metal object. Use of extreme caution to maximize patient safety is
advised in these situations.
Never intentionally “ground” a patient by connecting them to a grounded metal object
or earth ground in order to overcome interference problems in your monitoring.
On occasion, two or more ESUs may be connected to a patient and in use simultaneously.
For dual monopolar ESU applications, the location of each ESU return electrode is
particularly important. ESU manufacturers advise that each ESU return electrode should
be positioned as close as possible to the respective site of surgery (that is, the point of
application of the ESU signal to the patient) and care should be taken to avoid the
possibility of the two ESU return electrodes touching.
Given that the responsibility of the set up and use of the ESUs lies with the surgical team,
assurance that these and other cautions against ESU misuse (as described by the ESU
manufacturers) are heeded and followed is not the responsibility of the person performing
neurophysiological monitoring. Being aware of these cautions, however, may help you to
minimize the risks of injury to the patient.
2. Adding an electrode switching box can increase current leakage to ground, thereby
making the system no longer conform to the safety standards (as described initially in
this document). Specifically, the wiring and cabling associated with the switching
device may introduce additional capacitance to earth ground which in turn increases
leakage.
The wiring and cabling associated with the switching device may also introduce additional
capacitance to earth ground and in turn increase current leakage, thereby making the
stimulator subsystem no longer conform to the safety standards. Even when packing and
cabling for the switching device is carefully managed, the device can still represent a
patient safety hazard because it may provide a low impedance path to earth ground for RF
currents. The use of a stimulus switching device can affect the system functionality
because additional capacitance in the switching device can act to distort the shape of the
electrical stimulus pulse being delivered to the patient, thereby altering its effectiveness as
a stimulus.
Use only Natus Neurology Incorporated-supplied and/or approved electrode and stimulus
switching devices. Please see the last paragraph on page 2-11.
Please keep these safety guidelines on or near your Nicolet EEG Monitoring system.
• Use only approved devices for recording or stimulating electrode switching on Nicolet brand
systems.
• Never use an extension cord to power the system.
• Always use the appropriate Natus Neurology Incorporated O.R. Electrode Headboxes and
Amplifiers for connecting all recording electrodes to the Nicolet brand system.
• Always use the largest electrode possible for the Nicolet brand system’s patient "ground",
especially when used with needle or hookwire electrodes. Never use the ESU return electrode
as the Nicolet brand system patient ground, or place the Nicolet brand system patient
"ground" electrode beneath or further away than the ESU return electrode. Never directly
connect the patient or the Nicolet system patient "ground" electrode to grounded metal objects
or to earth ground.
• Minimize electrode length between patient and protected headbox.
• Verify that the ESU return electrode is optimally on and properly applied to the patient.
• Advise the surgeon to avoid prolonged ESU activation. Never activate the ESU when the ESU
probe is not in contact with the patient.
• Avoid ESU use close to the monitoring electrodes.
• Turning the Nicolet brand system power off does not eliminate the possibility of current flow
through the monitoring electrodes.
• Periodically check the system ground integrity, the system leakage current and the leakage
current of the amplifier and electrical stimulator subsystems. Verify that line power
receptacles are properly grounded.
Please keep these safety guidelines on or near your Nicolet brand system.
• Verify that the ESU return electrode is optimally located on and properly applied to the
patient.
• Advise the surgeon to avoid prolonged ESU activation. Never activate ESU when the ESU
pencil is not in contact with the patient.
• Avoid ESU use close to the monitoring electrodes.
• Always use the appropriate Natus Neurology Incorporated O.R. Electrode Headboxes for
connecting all recording electrodes to the Nicolet brand system. Always use the appropriate
Natus Neurology Incorporated O.R. Stimulator boxes for connecting all stimulating
electrodes.
• Always use the largest electrode possible for the Nicolet brand system’s patient common,
especially when used with needle or hookwire electrodes. Never use the ESU return electrode
as the Nicolet brand system patient ground, or place the Nicolet brand system patient common
electrode beneath the ESU return electrode. Never directly connect the patient or the Nicolet
system patient common electrode to grounded metal objects or to earth ground.
• Minimize electrode length between patient and protected headbox.
• Turning the system power off does not eliminate the possibility of current flow through the
monitoring electrodes.
• Use only approved devices for recording or stimulating electrode switching on Nicolet brand
systems.
• Periodically check the system ground integrity, the system leakage current and the leakage
current of the amplifier and electrical stimulator subsystems. Verify that line power
receptacles are properly grounded.