Document No.

: S503-E004
Revision : Original

SPECIFICATIONS
for
Mobile X-ray System: MUX-10 Series

Medical Systems Division

1
Purpose

The mobile X-ray system has been designed to deliver exceptional mobile X-ray
performance and image quality. It will also ensure optimum operator and patient
safety.

1. Features

Compact design
The unit, with its exceptional forward view, can be moved in limited spaces
confidently and with complete control.

Easy positioning
The unit may be moved into position easily and speedily with X-ray tube part
routable around the support stand.

High-frequency inverter
Use of Shimadzu's high-frequency inverter technology, with a maximum
frequency of 60 kHz, generates the high voltage with the lowest ripple factor
available, ensuring efficient X-ray generation.

Anatomical programs
Anatomical programs are installed as standard, facilitating the setting of
optimum exposure conditions.

2
2. Use Conditions

Environmental Conditions
Use environment
Ambient temperature : 10℃ - 40℃

Relative humidity : 30% - 85% (no dew condensation)

Atmospheric pressure : 700 hPa - 1060 hPa

Storage environment
(This condition applies to a package of equipment that is not for
transportation or storing, but packed by user’s convenience.)

Ambient temperature : -10℃ - 60℃

Relative humidity : 30% - 85% (no dew condensation)

Atmospheric pressure : 700 hPa - 1060 hPa

Power supply
AC power

System : Single phase AC

Frequency : 50/60 Hz

Standard voltages : 100, 110, 120, 200, 220, 230, 240 V

(set one of the voltages when installing)

Voltage variation range : ±10% of standard voltages

Rated power supply : More than 1.5kVA

Earth
Earth terminal : Earth resistance of no more than 100Ω

Additional earth terminal : Earth resistance of no more than 100Ω

3
EMC ( Electro-Magnetic Compatibility )
This system belongs to Group 1 and Class A equipment in accordance with
EN60601-1-2:1993, EN55011:1991.

The system uses radio-frequency energy for its internal function only and this is
not intended to deliver energy to the patient. However, radio-frequency energy
can affect other highly sensitive equipment.

The system main power line in the clinical site should be connected to the
domestic power sources which are separated from the public main network.

3. Conformity to IEC Standards

IEC standards to which the system conforms are below:

IEC 60601-1:1988 +A1 : 1991 + A2 : 1995

IEC 60601-1-2:1993

IEC 60601-1-3:1994

IEC 60601-1-4:1996

IEC 60601-2-7:1998

IEC 60601-2-28:1993

IEC 60601-2-32:1994

4
4. Classification of equipment

The method of protection against electric shock


Class I equipment

The degree of protection against electric shock


Type B equipment for x-ray tube assembly

The degree of protection against harmful ingress of water


Ordinary equipment

The degree of safety of application in the presence of a flammable


Anesthetic mixtures with air or with oxygen or nitrous oxide

Equipment is not suitable for use in the presence of a flammable anesthetic


mixture with air or with oxygen or nitrous oxide

The mode of operation


Continuous operation with intermittent loading

5
5. Specifications

MUX-10
Use environment
Atmospheric pressure 700 ~ 1060hPa
Ambient temperature 10 ~ 40℃
Relative humidity 30 ~ 85%

Common specifications
Total width 640mm
Total length 1130mm
Height of support stand 1930mm
Weight 240kg
Cassette storage 10 lightweight cassettes: 14×17 (half size)
Focus height 650 ~ 2010mm
Tube arm length 660 ~ 1000mm
Support stand rotation range ±270deg min.
Tube rotation angle ±180deg
Tube rotation around tube axis Forward: 90 deg min.; Backward: 20 deg max.

Power supply
Rated power supply 1.5kVA
Supply voltage Single phase AC 100,110,120,200,220,230,240V
Supply frequency 50/60Hz

Options
Remote controller Type Infrared type
Function Ready operation, X-ray operation, Collimator lamp ON/OFF
Model name SSZ-R190
Manufacturer SEITECH
Printer Type Thermal Printer
Function The following information can be printed out after exposure:
Year, month, hour, minute,Measured voltage, Measured mAs
Model name
Manufacturer SANEI
Note The RS232C connector is specified to be connected with a printer
above.
Photo-timer (AEC) Type Ionization chamber type with two detection fields.
Model name MOBILE-AID
Manufacturer AID
200mAs option The current time product can be set to 200mAs.
(The standard is 100mAs)
Protective screen Acrylic containing lead resin.
option

6
MUX-10
High-voltage generator
12.5kW (100kV、125mA、0.1s)
Nominal electric power (The product of the current allowed to flow for 0.1 seconds at a
tube voltage of 100 kV and tube voltage)
The nominal maximum tube voltage
and the maximum tube current
allowed to flow at the nominal 125kV、100mA
maximum tube voltage
The maximum tube current and the
maximum tube
voltage that allows the maximum tube 160mA、 80kV
current to flow
The combination of the tube voltage 125kV、100mA
Ratings and tube current that output the 100kV、125mA
maximum electrical power 80kV、160mA
The minimum value of the current
time product 0.32mAs
The maximum value of the current 100mAs
time product (200mAs with option)
The minimum value of nominal
irradiation time 10 ms
(in photo-timer exposure)
Standard With 200mAs option
40 ~ 42kV 0.32 ~ 100mAs 0.32 ~ 200mAs
The current time
43 ~ 52kV 0.32 ~ 80mAs 0.32 ~160mAs
product that can be
53 ~ 67kV 0.32 ~ 63mAs 0.32 ~ 125mAs
set with the X-ray tube
68 ~ 85kV 0.32 ~ 50mAs 0.32 ~ 100mAs
voltage
86 ~ 106kV 0.32 ~ 40mAs 0.32 ~ 80mAs
107 ~ 125kV 0.32 ~ 32mAs 0.32 ~ 63mAs
Setting and display of Range of the setting: 40 ~ 125kV、in 1-kV increments
tube voltage for
exposure Display: Digital display
0.32, 0.36, 0.40, 0.45, 0.50, 0.56, 0.63, 0.71, 0.80,
0.90, 1.0,1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8,
3.2, 3.6, 4.0, 4.5,5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10,
Setting and display of Range of the setting: 11, 12, 14, 16, 18, 20, 22,25,28, 32, 36, 40,
radiographic current
time product. 45, 50, 56, 63, 71, 80, 90, 100,
110*,125*,140*,160*,180*,200* mAs
*Optional
Display: Digital display
Up to 72 kinds of anatomical programs can be stored.
(72 kinds = 8 body parts × 3 orientations × 3 body thickness)
The contents of anatomical program stored in memory are:
Anatomical program (a) Exposure condition (kV, mAs)
(b) Method of exposure
(c) Selection of photo-timing pickup field of photo-timer
(d) Setting of film density with photo-timer
Photo-timer filming is available in
combination with optional photo-timer
receptor
Direct photo-timer filming (optional) (a) It is possible to choose manual filming or photo-timer filming.
(b) Film density can be adjusted in seven steps with the density
adjuster.(-3, -2, -1, 0, +1, +2, +3)
(c) It is possible to choose either of two combinations of the
photo-timing pickup fields.

7
MUX-10
Collimator R-20CA
Shape Rectangular
Field Max. field 430 × 430 mm SID 1m
Min. field 0 × 0mm (leaves overlapped)
Average illumination Over 160 lx
Contrast ratio of edge Over 3
Accuracy 2% of SID in use
Light field
Display of center Black cross lines
Type of lamp Halogen lamp (12V, 100W), #55939
Illuminating period 30 seconds, with automatic turn-off time switch
SID indicated 1, 1.5, 2 m
Field size
indication Dimensions indicated 20(8), 23(9), 25(10), 28(11), 30(12), 36(14), 43(17) cm(in.)
Accuracy of SID indication 5%
Drive of leaves Manual
Mounting dimension (Focus to fit surface distance) 56 mm
External dimension (W × D × H) 202 × 211 × 197 mm

X-ray tube Assembly*


Model Name 0.7U161CS-36
Spot size 0.7mm
Target angle 16deg
Nominal X-ray Radiography 125kV
tube Voltage
X-ray tube Max. heat content 750kJ(1060kHU)
Assembly
Max. continuous heat content 120W(170HU/s)
X-ray Tube Max. anode heat content 100kJ(140kHU)
Max. anode cool content 360W(440HU/s)
Max. continuous heat content 180W(170HU/s)
Weight 12.5kg

Minimum specific filtration


Total 2.5 mm Al 70 kV
X-ray tube assembly Equivalent to 1.5 mm Al 70 kV
Multi-leaf collimator Equivalent to 1.0 mm Al 70 kV
Classification of the system
Method of protection against electric shock ClassⅠ
Degree of protection against electric shock Type-B for x-ray tube assembly
Classification according to operation mode Continuous operation with intermittent load
*For detail about X-ray-tube assembly, please refer to X-ray tube operation manual (M535-E218, M535-
E219).

8
6. Dimensional Drawing of the System

weight : 240kg±10kg
( 265kg±10kg with 200mAs option)

Dimensional drawing of the system