Sunteți pe pagina 1din 113

The purpose of the service manual is to make it easier to carry out installation, troubleshooting and

repair of the DAC UNIVERSAL autoclave.

Content

1. Installation of the DAC UNIVERSAL..................................................................................3


1.1 Requirement concerning placing ....................................................................................... 3
1.2 How to install the DAC UNIVERSAL? ................................................................................ 3
1.3 Before starting the autoclave - IMPORTANT!..................................................................... 4
1.4 Yearly service DAC Universal ........................................................................................... 6
2. Error code list ................................................................................................................ 12
2.1 Error 64 – Safety system has been released .................................................................... 13
2.2 Error 65 – Safety system is defective ............................................................................... 14
2.3 Error 66 – High temperature during sterilisation................................................................ 15
2.4 Error 67 – Sterilisation time is to short ............................................................................. 16
2.5 Error 68 – Too high pressure during pressure test............................................................ 17
2.6 Error 70 – The pressure does not drop following the pressure test .................................... 18
2.7 Error 71 – Bad water ...................................................................................................... 19
2.8 Error 72 – Lid does not open........................................................................................... 20
2.9 Error 73 – The sterilisation temperature has not been reached during back-flush ............... 22
2.10 Error 74 - The right temperature for building up pressure for sterilisation has not been
reached fast enough ................................................................................................................ 24
2.11 Error 75 – The sterilisation temperature has not been reached fast enough ....................... 26
2.12 Error 77– Air valve for lubrication is defective................................................................... 27
2.13 Error 78 – Valve 7 is not opening .................................................................................... 28
2.14 Error 79 – Valve 6 does not open.................................................................................... 29
2.15 Error 80 - The sterilisation temperature has not been reached during water reduction ........ 30
2.16 Error 81 – Liquid error .................................................................................................... 31
2.17 Error 82 – The temperature sensor has reached a point where it cannot measure .............. 32
2.18 Error 83 – Temperature sensor chamber in saturation ...................................................... 33
2.19 Error 84 – The temperature during sterilisation is below the temperature chosen................ 34
2.20 Error 85 – The pressure sensor has reached a point where it cannot measure ................... 36
2.21 Error 86 – Activation of the safety ring during closing of the lid .......................................... 37
2.22 Error 87 – The safety ring is dismounted.......................................................................... 38
2.23 Error 88 – The steam pressure is too low during sterilisation ............................................. 39
2.24 Error 89 – The steam pressure is too high during sterilisation............................................ 40
2.25 Error 90 – Pressure too low during pressure test.............................................................. 41
2.26 Error 91 – The autoclave is leaking due to leakage in the chamber during pressure test ..... 42
2.27 Error 92 – The autoclave has been disconnected during sterilisation ................................. 44
2.28 Error 93 – The steam generator is too hot ........................................................................ 45
2.29 Error 95 – water consumption during standby .................................................................. 46
2.30 Error 96 – No pressure from the expel valves................................................................... 47
2.31 Error without code .......................................................................................................... 48
3. Instruction for repair ....................................................................................................... 50
3.1 Instructions for dismounting the casings .......................................................................... 51
3.2 Complete chamber......................................................................................................... 53
3.3 Opening/closing system ................................................................................................. 59
3.4 Water tank/Conductivity sensor....................................................................................... 61
3.5 Plastic top ..................................................................................................................... 65
3.6 Valve battery ................................................................................................................. 66
3.7 Lid ................................................................................................................................ 72
3.8 Pressure sensor............................................................................................................. 77
3.9 Temperature sensor ....................................................................................................... 78
3.10 Fans and back plate....................................................................................................... 79
3.11 Steam generator ............................................................................................................ 82
3.12 Mainboard ..................................................................................................................... 85
3.13 Automatic water supply .................................................................................................. 87
3.14 Pumps .......................................................................................................................... 88
3.15 Pressure regulator ......................................................................................................... 90
3.16 Ground Connection at the main board. ............................................................................ 92
3.17 Oil dispenser ................................................................................................................. 93
3.18 Safety parts ................................................................................................................... 95
4. Software........................................................................................................................ 96
4.1 ServiceSoftware............................................................................................................. 97
4.2 Download new software in the autoclave ....................................................................... 103
1
5. Spore test.................................................................................................................... 107
6. Pressure and temperature ............................................................................................ 108
7. Appendix – Diagram..................................................................................................... 112
8. Revision history ........................................................................................................... 113

2
1. Installation of the DAC UNIVERSAL

The purpose of this chapter is to provide the user with information about how the installation of the
DAC UNIVERSAL should take place. It is recommended that the user manual be examined thoroughly
before the installation is carried out.

Please note that the autoclave must be left to warm to the ambient temperature before it is connected.
Note : Do not keep the DAC Universal below 0°.

1.1 Requirement concerning placing


The DAC UNIVERSAL should be placed in an open space on a flat surface.

Recommended minimum space to the wall: 10 cm.

Furthermore, there must be enough space to allow the autoclave to open (upright). Total height of the
DAC UNIVERSAL when open: 60 cm.

1.2 How to install the DAC UNIVERSAL?


Before the DAC UNIVERSAL is ready for use, the following connections are required:

90-240 volt and 50-60 Hz


The DAC UNIVERSAL is supplied with a main lead with plug, which must be connected to the
autoclave (see figure 1.2.1). When the DAC UNIVERSAL is connected to power the lid will then open
and error 92 can possibly occur; in this case press “C” on the display to reset the autoclave.

Compressed air adjusted between 5-8 bars: 60 litres pr. min.


A tube size 6/4mm is required for the connection between the DAC UNIVERSAL and the external
compressor. The compressed air tubing is inserted in the top screw coupling (see figure 1.2.1). The
tube must be held in while simultaneously the nut is fastened with a 12mm spanner. To dismantle the
tube the nut is unfastened and the tube can easily be drawn away.

The DAC UNIVERSAL has a built-in pressure regulator therefore the autoclave will function at 5-8
bars. The compressed air is used during the cleaning and lubrication cycle. The compressed air is
specifically used for flushing the internal channels, during internal cleaning and for the pulse wash
during the external cleaning. Furthermore the compressed air is used for distributing the oil during the
lubrication cycle.

The compressed air to the DAC UNIVERSAL must be dry and must not contain any oil remains.
Please note that an air filter (20 micron) must be installed on the inlet tube before the DAC
UNIVERSAL. Please contact the supplier of your compressed air unit if you have any questions
regarding this issue.

Drain for used water and oil


Residual water emerges through the drain outlet. Therefore the drain outlet (the middle screw coupling
- see figure 1.2.1) must be connected with a drain tank (order no. 700040 – includes 3 meters of PTFE
tube 6/4 mm and mounting instructions) in order to collect the residual water. For this connection you
must use a 6/4mm PTFE tube. The PTFE tube must not be longer than 3 meters.

The fitting and the removal of the tube are accomplished in the same way as the connection of
compressed air. The tube must be connected without pressure to the drain. When connecting the
drain tank to the drain of the DAC UNIVERSAL it is important to fix the connecting piece thoroughly.

Be aware not to handle the tube for the drain when it is hot or when the machine is in operation.

Warning
Oil and hot steam is discharged through the drain from the autoclave.
If using another type of tube than the PTFE tube, the tube must be steam resistant at approx. 137ºC.

One of the following connections must be made:

Constant connection to demineralised/destilled water


Dismantle the stopper on the water inlet.

3
The constant connection of demineralised/destilled water is through a 6/4mm tube connected to the
bottom screw coupling (see figure 1.2.1). The fitting and the dismantling of the tube are done in the
same way as the connection of compressed air. Maximum water pressure: 4 bars.
It is important to have a water filter (20 micron) in front of the DAC UNIVERSAL.
In order for the autoclave to work with an automatic connection to demineralised/destilled water the
autoclave has to be put in an “automatic water inlet mode”. This is done through the menu system of
the DAC Universal under “settings” and automatic water is set to “on" (see the user manual).
A maximum water conductivity of 3.0 µS (microsiemens) is required.

Manual supply of demineralised/destilled water


Leave the stopper on the water inlet.
All functions (cleaning and sterilisation) of the DAC UNIVERSAL require demineralised or destilled
water. Therefore, it is necessary to frequently supply the internal water tank with water if the constant
connection is not chosen. 1 filled water tank = approx. 6 complete cycles.
Volume in water tank: 2.0 l.
Look in section 3.4 where the water tank is and this is where to fill in the water (See the user manual).
A maximum water conductivity of 3.0 µS (microsiemens) is required.

1.2.1 Connection of the DAC UNIVERSAL to a Printer


The DAC UNIVERSAL is prepared for connection to a printer in order to print autoclave reports.
It is also possible to connect to a computer in order to use the service software for the DAC
UNIVERSAL (see section 4).

Figure 1.2.1: Installation

Connect to computer or
printer (COM)

Drain filter lid


Connect to 90-240 volt & The water runs through the
50-60 Hz filter when it is led from the
chamber. The filter must be
cleaned on a regular basis
look in the instruction
Connect to compressed manual.
air (compressor) .

Connect to drain

Connect to
demineralised/destilled
water - optional

1.3 Before starting the autoclave - IMPORTANT!


Before handling the autoclave please be aware that the weight of the machine is app. 25 kilograms
(=50 pounds). Therefore when lifting the machine, please use a lifting technique adjusted to the weight
of the machine. And please be aware not to hold under the machine when it is hot.

In order to be able to use the machine please go through steps 1-3

1. Connect the machine to a power supply.

The DAC UNIVERSAL is connected between 90 – 240 Volt and 50 –60 Hz with the cable that comes
with the DAC UNIVERSAL.

2.Mount the oil bottle

4
The oil bottle has a capacity of 0.2 litres of oil.
This quantity of oil is sufficient for approximately 250 cycles. It is recommended to use the type of oil,
which the manufacturer of the instruments recommends. Mount the oil bottle by opening the blue top
lid of the tank system
1. Mount the in oil bottle beside the water tank. See how on Figure 3.17.1

2.The oil bottle is dismounted in the opposite sequence. See section 3.17.1 about mounting and
dismounting the oil bottle.

Please check the instructions from the manufacturer of the oil, which security precautions you have to
follow when handling the oil.

Check the oil in the chamber bottom.


Choose the Lubricate button at the display and press for more than 1 sec.
Through visual control, check that oil appears in the bottom adapter in the sterilisation chamber.
It is very important to see the oil enter the bottom adapter in the chamber, if there is no oil the
instruments are not lubricated.
Maybe it is necessary to hold the oil bottom down for several air shots (about 5 air shots), before the
oil is coming out on a new DAC UNIVERSAL.
If oil is not coming from the bottom adapter. See section 2.30.6

The oil bottle must be change when the message ”Low oil” is visible in the display, or if the error 81
has come. The message automatically disappears when the oil bottle is change.

3.Fill the water tank

The water tank has a capacity of approximately 2.0 litres of water.

This quantity of water is sufficient for approximately 6 normal cycles. It is necessary to use distilled or
demineralised water. A maximum water conductivity of 3.0 µS (microsiemens) is required.

Please note that the water should be poured carefully into the tank. Otherwise, the water may run in
between the tank and the autoclave itself, resulting in water on the table.
Note the following is necessary to prevent air bobbles in the tubes.
Water is led through the tubes in the following way:

1. The lid must be up


2. Enter the menu system by pressing the “Menu/Enter” button
3. Press the “Up” or the “Down” button until you reach “Manual”.
4. Press the “Menu/Enter” button to select “Manual” mode.
5. Press the “Up” or “Down” button in order to reach “Water” in the menu.
6. When “Water” is marked you can press “Menu/Enter” in order to start pumping water into the
tubes.
7. When water becomes visible in the sterilisation chamber, there is water in the tubes and the
“Menu/Enter” button can be touched again, in order to stop the water from being pumped into the
chamber.
8. Press “Clear” until the normal display appears.

It is also possible to choose the “rinse/flush” button at the display and press for more than 1 sec.
Hold the button down and through visual control check that water appears in the bottom adapter in the
sterilisation chamber.

There is now water in the tubes.

The water tank must be filled when the message ”Low water” is visible in the display or if the error 81
has occurred. The message “Low Water” automatically goes out when the water tank is filled.

When you have ensured that there is both oil and water in the tubes, the instruments are placed on
the adapters on the lid and the autoclaving cycle can be started (see the instruction manual). If the
adapters are not mounted on the lid when the machine is delivered, please refer to section 3.7.

NOTE! At each adapter position there is a marking, which is the number of the adapter position.
At position one it is not possible to place an instrument, which is longer than 14,5 cm. (See section
3.7, figure 3.7.1.) at position 2 – 6 it is not possible to place an instrument which is longer than 16 cm.

5
1.4 Yearly service DAC Universal

Every year or after 2500 runs it is recommended to perform a yearly service.


The DAC Universal will flash with the alarm led without sound, when it is time for yearly service.

Yearly service kit


DAC Universal
English
Article no. 728095

Code Description Qty

360000Screw, M2x10 for old adapter type 12


360220Screw, M3x16 for valve 14
360221Screw, M3x20 for valve 4 and 8 4
360222Screw, M3x10 for valve 18
720058Angular screw fitting for valve 8 1
320901Filter 1
364105O-ring for filter housing 1
728083Gasket for anchor solenoid valve 1-3, 5-7 + 9 7
708087O-ring for solenoid valve 9
708088O-ring for solenoid valve 9
708089O-ring for solenoid valve 9
708234O-ring for bottom adapter 9,25x1,78 2
708265Gasket for adapter incl. Screws M2x6 6
708260Gasket for lid 1
708249O-ring for guiding tube on light lid 4
728096O-rings and gasket for valve 12 (Expel valve) 1

How to perform a yearly service:


Dismount the front casing and the plastic top see section 3.1

Change of gasket and o-rings in valve 1-3 , 6-7 and 9.

Dismount the coil by


means of an 11-mm
spanner.

6
Dismount the valve
from the block by
means of a torx 10
screwdriver.

Dismount the valve


by means of a torx
10 screwdriver.

Change the o-ring


no. 708089 and the
gasket no.708083

Change the o-rings


against the valve
block no. 708087
and 708088

The mounting sequence is the opposite of the above dismounting.


Remember to change the screws 360022 and 360020.

Dismounting valve 5.

Open the back plate with a Torx 20 screwdriver see section 3.1
Dismount the coil by means of an 11-mm spanner.
Dismount the 4 screws with a torx 10 screwdriver and open the valve.
change o-ring 708089 and gasket 708083.

Change the o-ring


no. 708089 and the
gasket no.708083
inside valve 5.

The mounting sequence of valve 5 is the opposite of the above dismounting.

7
Dismounting valve 8:

Dismount the PTFE


tube at the top of
valve 8 by means of
a 12 mm spanner.

Dismount nut at
valve 8 with a 9-mm
spanner.

Then valve 8 has to be dismounted as the above valves (1-3, 6-7 and 9)
Change the o-rings 708087, 708088 and 708089 the gasket in this valve is in the core
and do not have to be changed.
The mounting sequence of valve 8 is the opposite of the above dismounting and be careful not to
mount the nut to tight, because it will break.
Remember to change the screws 360022 and 360021.

Dismounting valve 4:

Dismount the PTFE


tube at the top of
valve 4 by means of
a 9 mm spanner
and dismount the
banjo bolt with the
same spanner.

Then valve 4 has to be dismounted as the above valves (1-3, 6-7 and 9)
Change the o-rings 708087, 708088 and 708089 the gasket, in this valve is in the core
and do not have to be changed (see figure 3.6.1 where valve 4 is placed).

The mounting sequence of valve 4 is the opposite of the above dismounting.


Remember to change the screws 360022 and 360021.

Dismounting of valve 12

Dismount the tube in front valve 4 for more space .


Dismount the 4 screws at the top of valve 12.
Turn the filter house and dismount the 6/4 PTFE tube with a 12-mm spanner.

Dismount the 4 screws.


Note there is a spring
inside.

8
Dismount the 6/4 mm
PTFE tube with a 12-
mm-spanner.

Dismount the tube to


valve 4 here for more
space.

press the core out of the valve and change the o-rings and gasket. See figure 3.6.9.

Change the 3 o-rings


and remember to use
silicone grease on the
two o-rings inside the
valve.

Change the gasket on


the core.

Mounting of valve

The mounting sequence of valve 12 is the opposite of the above dismounting.

Valve 10 and 11 does not need maintenance.

Change the o-ring in the chamber:

By means of a Sond
it is possible to
change the two o-
rings at bottom
adapter no. 708234

The mounting sequence of the o-rings is the opposite of the above dismounting, be careful not to
place the o-rings wrong so it blocks the drive channel.

9
Change the filter and filter lid o -ring:

Change the filter


320901 by means of
filter key and
change the o-ring
364105 at the filter
lid.

Change of gaskets and o -rings at the lid:

Dismount the
screws at each
adapter by means of
an allen key and
remove the
adapters.

Change the adapter


gasket 708265 at
each adapter.

The mounting sequence of the adapters is the opposite of the above dismounting.

10
Change the o-rings at the guiding tube:

Turn the big metal


tube for dismounting.

Dismount the bottom


adapter by removing
the two screws with a
screwdriver and take
off the adapter.

Change the two o-


rings at each guiding
tube 708249.

The mounting sequence of the bottom adapter is the opposite of the above dismounting.

Change of the black gasket at the lid:

Dismount the gasket


708260 with a
screwdriver like this.

The mounting sequence of the black gasket is the opposite of the above dismounting.
Remember to press the gasket down in the outer and inner ring at the lid.
It is also recommended to change the o-rings at the instrument adapters.
Now the yearly service has been carried out.

11
2. Error code list

Chapter 2 should be used for trouble shooting when the autoclave gives an error.

Chapter 2 is elaborated in such a way that it can be used as a book/work of reference.

Under each error the different error possibilities are listed. Moreover various advise about how to find
the cause of the error is also listed.

12
2.1 Error 64 – Safety system has been released

This error can occur if an error has released the safety system, which cuts the power to the motor,
pumps, valves and the heating element.

The internal safety system has blocked the power to the motor, the steam generator, the pumps and
the valves
Restart the autoclave by disconnecting and reconnecting the power. This resets the safety system.

13
2.2 Error 65 – Safety system is defective

This error can occur if the internal safety system cannot cut the 24V. power to the motor.

Check whether there is a counter-pressure in the drain system.

Valve 1, 2 or 3 is leaking air


Fill water into the bottom of the chamber, until it covers the bottom adapter. If one of the air valves is
leaking air, bubbles will appear. Note that air must be connected to the autoclave.
Valve 1 is leaking air, if bubbles appear from one or more of the 4 holes at the bottom of the chamber.
Valve 2 is leaking air, if bubbles appear from top hole in the bottom adapter.
Valve 3 is leaking air, if bubbles appear from the side hole on the bottom adapter. See section 3.6
about valve battery, for changing the valve and/or gasket and o-rings.
It is also possible to test the valves via a PC with service software. See section 4.1 to activate the
service software (remember to empty the chamber for water before testing the air valves).

Replace the main board


Pressing the “C” button cannot reset error 65. It has to be reset via the computer and service
software. See section 4.1.
Replace the main board. See section 3.12

14
2.3 Error 66 – High temperature during sterilisation

This error may occur if during sterilisation the temperature exceeds 3°C over the chosen program
(more that 137°C).

This error can occur if:

The temperature sensor is defective


Change the temperature sensor. See section 3.9 for change of the temperature sensor.

The pressure sensor is defective


Change the pressure sensor. See section 3.8 for change of the pressure sensor.

15
2.4 Error 67 – Sterilisation time is to short

This error may occur if the sterilisation time is to short.

This error can occur if:

The main board is defective


Change the main board. See section 3.12 for change of the main board.

16
2.5 Error 68 – Too high pressure during pressure test

This error may occur if during pressure test the pressure exceeds 4.8 bars absolut pressure.

This error can occur if:

The pressure regulator is defective or incorrectly adjusted


Adjust the pressure regulator. See section 3.15 about the pressure regulator.
Change the pressure regulator. See section 3.15 about the pressure regulator.

Valve 1, 2 or 3 is leaking air


Fill water into the bottom of the chamber, until it covers the bottom adapter. If one of the air valves is
leaking air, bubbles will appear. Note that air must be connected to the autoclave.
Valve 1 is leaking air, if bubbles appear from one or more of the 4 holes at the bottom of the chamber.
Valve 2 is leaking air, if bubbles appear from the top hole in the bottom adapter.
Valve 3 is leaking air, if bubbles appear from the side hole on the bottom adapter. See section 3.6
about valve battery, for changing the valve and/or gasket and o-rings.
It is also possible to test the valves via a PC with service software. See section 4.1 to activate the
service software (remember to empty the chamber for water before testing the air valves).

The pressure sensor is defective


Change the pressure sensor. See section 3.8 about the pressure sensor.

17
2.6 Error 70 – The pressure does not drop following the pressure test

This error occurs if the pressure does not drop within 20 seconds after the initial pressure test.

During the pressure test, pressure is built up by opening valve 3 for 3 sec. and close then valve 1. The
pressure must rise to 4.0 bar absolut pressure . Then valve 1 closes and the pressure must be stable
during 10 seconds. If the pressure test is approved, the pressure is led out through the filter and valve
7, 6 and 9.

This error can occur if:

The filter is blocked


Change the filter in the filter housing on the back plate.
The filter house could be blocked in the inlet or outlet of the filter house.

Filter house is blocked


The filter house could be blocked in the inlet or outlet of the filter house.
Dismounts the two tubing’s at the filter house and check that there is no blockings in the fittings. See
section 3.10.5 about dismounting the filter house.

The drain tube is blocked


By dismounting the drain tube on the autoclave, it is possible to check whether air is coming out of the
drain outlet.
If air is coming out of the drain outlet, an additional check of the drain system must be carried out.
There must be no counter-pressure in the drain outlet.
If air is not coming out through the drain outlet, check the drain system of the autoclave through valve
9. See section 3.6 about valve battery.
The drain tube must be of heat resistant material: Nitram Dental recommends the use of a Teflon tube
(PFTE 6/4 mm).

Too much residual water in the chamber from the last test run
If there is too much water left in the chamber from the last process, there is no time for it to be pressed
out before the 20 seconds have passed, which is the time that the machine has to empty itself. Restart
the autoclave. The pressure test will press the water out.

Possible drain tank is full


Empty the tank. There must be no counter-pressure in the drain system. This can be remedied by
means of a small air-escape hole in the drain tank. A counter-pressure can occur if you use a drain
tank, which is too small, that is too small or if the air-escape hole is too small. The autoclave is tested
with an original Nitram drain tank, with an air-escape hole of 6 mm and with a min. content of 5 litres.
Nitram recommends the use of an original Nitram drain tank (order no. 700040).

Valve 9 is defective
Valve 9 can be activated via a PC with the service software. See section 4.1 to activate the service
software.
If it is clicking when activated, the valve is okay.
If it is not clicking, the valve is defective. See section 3.6 about valve battery for changing the valve
and/or gasket and o-rings.

18
2.7 Error 71 – Bad water

This error occurs if:

The conductivity of the water in the water tank is inferior to the limit value for the machine (maximum
value: 3 µs)

Empty the water tank for water and fill in new water with conductivity, which is below the limit value (3
µs). See section 3.4 to empty the water tank.

The conductivity sensor is defective and has to be changed. See section 3.4.

19
2.8 Error 72 – Lid does not open

The closing system has jammed during opening.

When the autoclave have completed the cycle and is half open press “C” button to start opening the
chamber, a timer starts. After 50 seconds this timer must receive a signal from the limit switch on the
chamber (see Figure 2.1.1).

Figure 2.1.1

Limit switch

If the switch has not been activated after 50 seconds, the following errors are possible:

Check: That there is no pressure in the chamber, open the drain filter lid see figure 1.2.1

The contact is not activated


The lever arm on the limit switch is incorrectly adjusted. The arm must hit the trace in the supporting
tube and be activated, when the spindle goes up (see Figure 2.1.1). Adjust the arm on the limit switch
by bending it up or down carefully, if necessary.

The surface of the lid holder for lid is not parallel with the surface of the chamber
This means that the 3 tension rods tighten towards each other. Normally it is possible to see if the lid
holder is not parallel with the chamber. If you suspect that this is the case, loosen the 3 screws on the
lid holder for lid and adjust the lid holder. See section 3.3.1 for adjustment of the lid holder.

The timing belt under the motor plate is loose


Press the “Menu/Enter” button. Press the arrows buttons up or down, till you reach “manual mode” on
the display. Then press the “Menu/Enter” button again and go to “close” by means of the arrow
buttons. Press the “Menu/Enter” button again to start the running of the motor. Press the “Menu/Enter”
button to stop the running of the motor. When the motor is running and the belt is turning, look for the
2 pointed screws on each timing belt, to check whether they are loose. They have to be changed and
re-tightened. See section 3.2 about complete chamber. The lid holder must then be adjusted. See
section 3.3.1.

Belt too tight


Check the belt by pressing the belt together in between the motor and the supporting tube with the
limit switch. The belt should give bend approx. 10-15 mm. See section 3.2.4 about complete chamber
for loosening the belt.

Belt too loose


You can clearly hear if the belt is tooting. Tighten the belt (see section 3.2.4) and adjust the lid holder
subsequently (see section 3.3.1)

20
Motor is defective
There is no movement when the motor is started. Check if there is power coming from the mainboard.
If there is power (24V DC) from the mainboard, the motor must be changed. See section 3.2.3 about
complete chamber for changing the motor.

No power supply from the mainboard


Check the tension at the wire connection for the motor (24 V DC) and check the wires for gaps. If the
wires for the motor have been short-circuited, the power supply on the main board will be damaged
and it must be changed. See section 3.12.2 for changing the main board.

21
2.9 Error 73 – The sterilisation temperature has not been reached during back-flush

This error occurs if the temperature does not rise to 134 degrees during the first back-flush cycle.

This error can occur if:

There is lack of water in the water tank:


No water in the tank.
Pour in water if the display indicates low water
Verify that water is present in the tank. The sensor may be defective. See section 3.4 about changing
the tank sensor.

Tubes are jammed or blocked under the tank.


Choose the “rinse/flush” button at the display and press for more than 5 sec. Check whether the little
water pump is lead water into the chamber. If this is not the case, dismount the water tank and check
blocked or jammed tubes, see section 3.4 for dismounting the tank.

The steam generator /the non-return valve is blocked/defective


Dismount the bottom tube after the non-return valve. See section 3.11.5. There must be water
emerging through the non-return valve when the steam pump is activated. See figure 2.9.1.
Use the service software to activate the steam pump. See section 4.1.
Try to blow air through the bottom tube. By means of compressed air, it is possible to blow air from the
bottom tube at the non-return valve and out through the steam generator, to the bottom of the
chamber to see if there is free passage through the steam generator (The steam generator can be
defective because of a small passage). See figure 2.9.1.

Figure 2.9.1

Water must come out of


the non-return valve, when
the steam pump is
activated.

NOTE! Avoid water on


electrical parts.

Try to blow air through the


bottom tube into the
chamber. There must be
nearly no resistance.

22
The steam pump is defective
If you can hear the steam pump is running, dismount the non-return valve at the steam pump (see
figure 2.9.1) and check whether the steam pump gives any water when activated. There must be a
strong jet of water and the water must not touch the electrical parts. It is possible to activate the steam
pump via a PC with service software. See section 4.1.
If you cannot hear the steam pump, check whether the pump is defective by mounting a pump which
is working (either a new one from the service box or the little water pump, which is then dismounted).
See section 3.14.1, for changing the steam pump. If the new pump is not functioning, check the cords
or change the mainboard. See section 3.12.2, for changing the mainboard.

Valve 6,7,9 is leaking (see valve diagram)


Check if steam pressure disappears through the drain outlet and the temperature cannot rise to
134°C. See section 3.6 about valve battery, for changing the valve or gasket and o-rings.

Valve 8 is not closed (see valve diagram)


If valve 8 is not closing at 107°C, the steam pressure disappears through the drain outlet and the
temperature cannot rise to 134°C. See section 3.6.5 about valve battery for dismounting the valves.

23
2.10 Error 74 - The right temperature for building up pressure for sterilisation has not been
reached fast enough

This error occurs if the temperature cannot reach 107°C before 2.5 minutes has passed after start of
the sterilisation (when “Hot rinse” is shown in the display).

This error occurs if:

There is lack of water in the water tank:


No water in the tank.
Pour in water if the display indicates low water.
Verify that water is present in the tank. The sensor may be defective. See section 3.4 about changing
the tank sensor.

Tubes are jammed or blocked under the tank.


Choose the “rinse/flush” button at the display and press for more than 5 sec. Check whether the little
water pump is leading water into the chamber. If this is not the case, dismount the water tank and
check blocked or jammed tubes, see section 3.4 for dismounting the tank.

The steam pump is defective


If you can hear that the steam pump is running, dismount the non-return valve at the steam pump (see
Figure 2.10.1) and check whether the steam pump gives any water when activated. There must be a
strong jet of water and the water must not touch the electrical parts. It is possible to activate the steam
pump via a PC with service software. See section 4.1.
If you cannot hear the steam pump, check whether the pump is defective by, mounting a pump that is
working (either a new one from the service box or the little water pump, which is then dismounted).
See section 3.14.1, for changing the steam pump. If the new pump is not functioning check the cords
or change the mainboard. See section 3.12.2, for changing the mainboard.
Figure 2.10.1

Check whether
water is coming
out of the pump

The steam generator /the non-return valve is blocked/defective


Dismount the bottom tube after the non-return valve. There must be water emerging through the non-
return valve when the steam pump is running (make sure that the water does not touch the electrical
parts). Use the service software to activate the steam pump. See section 4.1
Try to blow air through the bottom tube. By means of compressed air, it is possible to blow air from the
bottom tube at the non-return valve and out through the steam generator, at the bottom of the
chamber if there is free passage through the steam generator (see figure 2.10.2). (The steam
generator can be defective if there is a small passage).

24
Figure 2.10.2

Dismount here and blow the


steam generator through to
check whether there is
passage to the chamber. The
resistance must be very low.

Too many instruments in the basket.


o It is not possible to have more than 5 solid instruments in the basket. Look into the user
manual how to use the basket.

25
2.11 Error 75 – The sterilisation temperature has not been reached fast enough

This error occurs if the temperature cannot rise from 107° C to 134° C in 2.5 minutes.

The error can occur if:

There is lack of water in the water tank:


No water in the tank.
Pour in water if the display indicates low water
Verify that water is present in the tank. The sensor may be defective. See section 3.4 about changing
the tank sensor.

Tubes are jammed or blocked under the tank.


Choose the “rinse/flush” button at the display and press for more than 5 sec. Check whether the little
water pump is leading water into the chamber. If this is not the case, dismount the water tank and
check blocked or jammed tubes, see section 3.4 for dismounting the tank.

The steam generator /the non-return valve is blocked/defective


Dismount the bottom tube after the non-return valve. See section 3.11.5. There must be water
emerging through the non-return valve when the steam pump is activated. See figure 2.9.1. Use the
service software to activate the steam pump. See section 4.1
Try to blow air through the bottom tube. By means of compressed air, it is possible to blow air from the
bottom tube at the non-return valve and out through the steam generator, at the bottom of the
chamber if there is free passage through the steam generator (The steam generator can be defective
if there is a small passage). See figure 2.9.1.

The steam pump is defective


If you can hear that the steam pump is running, dismount the non-return valve at the steam pump (see
figure 2.9.1) and check whether the steam pump gives any water when activated. There must be a
strong jet of water and the water must not touch the electrical parts. It is possible to activate the steam
pump via a PC with service software. See section 4.1.
If you cannot hear the steam pump, check whether the pump is defective by mounting a pump which
is working (either a new one from the service box or the water pump, which is then dismounted). See
section 3.14.1, for changing the steam pump. If the new pump is not functioning check the cords or
change the mainboard. See section 3.12.2, for changing the mainboard.

Valve 6,7,9 is leaking (see valve diagram)


Check if steam pressure disappears through the drain outlet and the temperature cannot rise to
134°C. See section 3.6 about valve battery, for changing the valve or gasket and o-rings.

Valve 8 is not closed (see valve diagram)


If valve 8 is not closing at 107°C, the steam pressure disappears through the drain outlet and the
temperature cannot rise to 134°C. See section 3.6.5 about valve battery, for dismounting the valves.

26
2.12 Error 77– Air valve for lubrication is defective

This error occurs if the pressure does not rise in the chamber when air valve 3 is opening during
lubrication.

The error can occur if:

The lid is defective.


There is not free passage through the lid
If the gaskets under the adapters are defective the passage through the instruments can be blocked.
See section 3.7.1 about lid for changing the adapter gaskets.
Check the gasket in the lid and the channels are without blockings. See section 3.7
Try with a new lid to see if it is working

Valve 3 is not being activated


Check with a PC in the service software whether the valve 3 is clicking. See section 4.1 to activate
valve 3.
If valve 3 is not clicking, the following are the problem:
The cords are loose at the coil. See section 3.6.2 about valve battery for correct mounting of the
cords.
The cords can have a bad connection at the main board.
The coil is defective. See section 3.6.3 for replacement of the coil.
Valve 3 is defective. See section 3.6.3 for replacement of the valve.

The air flow to the autoclave is too low


In general it is not recommended to have several installations on a long tube with an external diameter
of only 6-mm. There must be airflow of minimum 60 l/min.

Defective main board


See section 3.12.2 for replacement of the main board.

27
2.13 Error 78 – Valve 7 is not opening

This error occurs in connection with the pressure test or the first back-flush cycle.
Error 78 occurs if the pressure does not drop fast enough when valve 7 opens after pressure test or in
the first back-flush when the temperature reaches 134°C, valve 7 is opening to lead steam through the
drive cannels of the instruments and out in the drain. Thereby the temperature and the pressure will
drop a bit in the chamber. If there is no pressure drop error 78 will occur.

The error can occur if:

The lid is defective.


There is not free passage through the lid
If the gaskets under the adapters are defective the passage through the instruments can be blocked.
See section 3.7.1 about lid for changing the adapter gaskets.
Check the gasket in the lid and the channels are without blockings. See section 3.7
Try with a new lid to see if it is working

Valve 7 is not being activated


Check with a PC in the service software whether the valve 7 is clicking. See section 4.1 to activate
valve 7.
If valve 7 is not clicking, the following could cause the problem.
The gasket and o-rings inside the valve are defective. See section 3.6 for replacement of the gasket
and o-rings.
The cords are loose at the coil. See section 3.6.2 about valve battery for correct mounting of the
cords.
The cords can have a bad connection at the main board.
The coil is defective. See section 3.6.3 for replacement of the coil.
Valve 7 is defective. See section 3.6.3 for replacement of the valve.

The autoclave is placed in a space which is too hot, whereby the valve is failing
Recommended distance to the wall is 10 cm. This is specifically applying to the fan at the behind of
the machine.
The autoclave must not be installed in a small closed room as e.g. a closed cupboard.

The fan at the back of the machine is defective or blocked by foreign bodies
The fan is blowing hot air out at the back of the autoclave. See section 3.10.1 about back plate for
replacement of the fan.

The drain is blocked


The tube can be blocked or bent. Check that there is free passage through the tube. The tube must
not be more than 3 m long and minimum of the 6-mm PTFE tube type.

The drain tank is full


Empty the tank. There must be no counter-pressure in the drain system. This can be remedied by
means of an air-escape hole of min. 6 mm in the drain tank. A counter-pressure can occur if you use a
drain tank, which is too small or if the air-escape hole is too small. We advise that you use an original
Nitram drain tank (order no. 700040) with an air-escape hole of 6 mm.

Defective main board


See section 3.12.2 for replacement of the main board.

28
2.14 Error 79 – Valve 6 does not open

This error occurs in connection with pressure test or the first back-flush cycle.
Error 79 occurs if the pressure does not drop fast enough when valve 6 opens after pressure test or in
the first back-flush when the temperature reaches 134°C, valve 6 is opening, to lead steam through
the spray channels of the instruments and out into the drain. Thereby the temperature and the
pressure will drop a bit in the chamber. If there is no pressure drop error 79 will occur.

The error can occur if:

The lid is defective. There is not free passage through the lid
If the gaskets under the adapters are defective, the passage through the instruments can be blocked.
See section 3.7.1 about lid for changing the adapter gaskets.
Check that the gasket in the lid and the channels are without blockings. See section 3.7
Try with a new lid to see if it is working

Valve 6 is not being activated


Check with a PC in the service software, whether the valve 6 is clicking. See section 4.1 to activate
valve 6.
If valve 6 is not clicking, the following could cause the problem.
The gasket and o-rings inside the valve are defective. See section 3.6 for replacement of the gasket
and o-rings.
The cords are loose at the coil. See section 3.6.2 about valve battery, for correct mounting of the
cords.
The cords can have a bad connection at the main board.
The coil is defective. See section 3.6.3 for replacement of the coil.
Valve 6 is defective. See section 3.6.3 for replacement of the valve.

The autoclave is placed in a space which is too hot, whereby the valve is failing
Recommended distance to the wall is 10 cm. This is specifically applying to the fan at the behind of
the machine.
The autoclave must not be installed in a small closed room as e.g. a closed cupboard.

The fan at the back of the machine is defective or blocked by foreign bodies
The fan is blowing hot air out at the back of the autoclave. See section 3.10.1 about back plate for
replacement of the fan.

The drain is blocked


The tube can be blocked. Check that there is free passage through the tube. The tube must not be
more than 3 m long and minimum of the 6-mm PTFE tube type.

The drain tank is full


Empty the tank. There must be no counter-pressure in the drain system. This can be remedied by
means of an air-escape hole of 6 mm in the drain tank. A counter-pressure can occur if you use a
drain tank, which is too small or if the air-escape hole is too small. We advise that you use an original
Nitram drain tank (order no. 700040) with an air-escape hole of 6 mm.

Defective main board


See section 3.12.2 for replacement of the main board.

29
2.15 Error 80 - The sterilisation temperature has not been reached during water reduction

This error can occur if:

There is lack of water in the water tank:


No water in the tank.
Pour in water if the display indicates low water
Verify that water is present in the tank. The sensor may be defective. See section 3.4 about changing
the tank sensor.

Tubes are jammed or blocked under the tank.


Choose the “rinse/flush” button at the display and press for more than 5 sec. Check whether the little
water pump is leading water into the chamber. If this is not the case, dismount the water tank and
check blocked or jammed tubes, see section 3.4 for dismounting the tank.

The steam generator /the non-return valve is blocked/defective


Dismount the bottom tube after the non-return valve. See section 3.11.5. There must be water
emerging through the non-return valve, when the steam pump is activated. See figure 2.9.1. Use the
service software, to activate the steam pump. See section 4.1
Try to blow air through the bottom tube. By means of compressed air, it is possible to blow air from the
bottom tube at the non-return valve and out through the steam generator, at the bottom of the
chamber, if there is free passage through the steam generator (The steam generator can be defective
if there is a small passage). See figure 2.9.1.

The steam pump is defective


If you can hear that the steam pump is running, dismount the non-return valve at the steam pump (see
section 3.11.5) and check whether the steam pump gives any water when activated. There must be a
strong jet of water and the water must not touch the electrical parts. It is possible to activate the steam
pump via a PC with service software. See section 4.1.
If you cannot hear the steam pump, check whether the pump is defective, by mounting a pump that is
working (either a new one from the service box or the little water pump, which is then dismounted).
See section 3.14.1, for changing the steam pump. If the new pump is not functioning, check the cords
or change the mainboard. See section 3.12.2, for changing the mainboard.

Valve 6,7,9 is leaking (see valve diagram)


Check if steam pressure disappears through the drain outlet and the temperature cannot rise to
134°C. See section 3.6 about valve battery, for changing the valve or gasket and o-rings.

Valve 8 is not closed (see valve diagram)


If valve 8 is not closing at 107°C, the steam pressure disappears through the drain outlet and the
temperature cannot rise to 134°C. See section 3.6.5 about valve battery for dismounting the valve.

Defective main board


See section 3.12.2 for replacement of the main board.

30
2.16 Error 81 – Liquid error

This error occurs if the start button is activated while a signal from the liquid sensor in the water tank
indicate that there is no water in the tank or if the oil bottle is empty, press “C” button to cancel.

The error can occur if:

There is no water in the water tank.

Fill in water (see section 3.4 where the water tank is placed).

There is lack of water or oil.

The liquid sensor is defective


The sensor is blocked. This can be disclosed by opening the water tank, touching the sensor with a
stick. If thereby it is possible to switch off the “low water” in the display and the error re-occurs, the
sensor must be changed. See section 3.4 about replacement of water sensor.

There is no oil in the oil bottle.

Change the oil bottle (see section 3.17.1 about oil dispenser)

Pressure switch at the oil dispenser is defective.

Change the pressure switch (see section 3.17.3 about oil dispenser).

Main board are not able to receive the information from liquid sensor or pressure switch.

Change the main board (see section 3.12)

31
2.17 Error 82 – The temperature sensor has reached a point where it cannot measure

This error occurs if the temperature sensor in the steam generator reaches a point, where it measures
its maximum value.
The steam generator temperature sensor is pre-calibrated in the mainboard, to regulate the steam
generator to 300°C.

This error can occur if:

The steam generator is out of calibration on mainboard


Change mainboard. See section 3.12. All main boards have been calibrated according to the steam
generator.

Insulation of temperature sensor in the steam generator


The temperature sensor in the steam generator must the insulated with silicone armoured glass flex.
The insulation must go all the way down through the insulation around the steam generator. A strip
must bind the three cords constituting the sensor.

Figure 2.17.1

The insulation must


go all the way down
into the steam
generator

Steam generator is defective


The steam generator thermo sensor is defective change the steam generator. See section 3.11.3 and
3.11.4.

32
2.18 Error 83 – Temperature sensor chamber in saturation

This error occurs if the temperature sensor is defect.

Replace the temperature sensor, see section 3.9.

33
2.19 Error 84 – The temperature during sterilisation is below the temperature chosen

This error occurs if the sterilisation temperature drops below the chosen program during sterilisation.
The autoclave is measuring 100 times per second. This means that you cannot necessarily see on the
display of the autoclave that the temperature has dropped.
Note: The DAC UNIVERSAL is in sterilisation steering at the pressure.

The error can occur if:

The autoclave runs out of water during the sterilisation

The steam generator does not contain enough energy for the entire sterilisation
There is an abnormal drop of energy, due to a leakage in the autoclave.
Check whether liquid is coming out anywhere during the sterilisation
Check whether steam is coming out through the drain. If this is the case, check the drain valve 6, 7, 8,
or 9. See section 3.6 about valve battery.
The temperature switch on the steam generator has been disconnected. See section 3.11 about
steam generator.

Defective temperature sensor


Change the temperature sensor. See section 3.9 about changing the temperature sensor.

The steam pump is defective


If you can hear the steam pump is running, dismount the non-return valve at the steam pump (see
figure 2.9.1) and check whether the steam pump gives any water when activated. There must be a
strong jet of water and the water must not touch the electrical parts. It is possible to activate the steam
pump via a PC with service software. See section 4.1.
If you cannot hear the steam pump, check whether the pump is defective by mounting a pump which
is working (either a new one from the service box or the little water pump, which is then dismounted).
See section 3.14.1, for changing the steam pump. If the new pump is not functioning, check the cords
or change the mainboard. See section 3.12.2, for changing the mainboard.

The non-return valve between the steam pump and the steam generator is defective
Dismount the bottom tube after the non-return valve See figure 2.9.1. There must be a strong jet of
water running through the non-return valve when the steam pump is activated and the wat er must not
touch the electrical parts. Use the service software to activate the steam pump. See section 4.1.

Valve 1, 2 or 3 is leaking air


Fill water into the bottom of the chamber, until it covers the bottom adapter. If one of the air valves is
leaking air, bubbles will appear. Note that air must be connected to the autoclave.
Valve 1 is leaking air, if bubbles appear from one or more of the 4 holes at the bottom of the chamber.
Valve 2 is leaking air, if bubbles appear from the top hole in the bottom adapter.
Valve 3 is leaking air, if bubbles appear from the side hole on the bottom adapter. See section 3.6
about valve battery, for changing the valve and/or gasket and o-rings.
It is also possible to test the valves via a PC with service software. See section 4.1 to activate the
service software. (empty the chamber for water before activating the air valves)

The filter in the filter house between valve 8 and the chamber is blocked.
Open the filter house and check the filter if the filter is blocked change the complete filter house.
spare part no. 728090.

34
Valve 8 does not open to ventilate the chamber when the DAC Universal heats up the chamber.
Dismount the valve 8. see section 3.6.5
Check that there is no dirt in the valve and look at the gasket that it is not too big.

Bad gasket. Good gasket levels


with the surface.

The pressure sensor is defective


Change the pressure sensor. See section 3.8 about changing the pressure sensor.

35
2.20 Error 85 – The pressure sensor has reached a point where it cannot measure

This error can occur if the pressure sensor is defective.

Replace the pressure sensor. See section 3.8 about the pressure sensor.

Valve 1, 2 or 3 is leaking air


Fill water into the bottom of the chamber, until it covers the bottom adapter. If one of the air valves is
leaking air, bubbles will appear. Note that air must be connected to the autoclave.
Valve 1 is leaking air, if bubbles appear from one or more of the 4 holes at the bottom of the chamber.
Valve 2 is leaking air, if bubbles appear from the top hole in the bottom adapter.
Valve 3 is leaking air, if bubbles appear from the side hole on the bottom adapter. See section 3.6
about valve battery, for changing the valve and/or gasket and o-rings.
It is also possible to test the valves via a PC with service software. See section 4.1 to activate the
service software (remember to empty the chamber for water before testing the air valves).

36
2.21 Error 86 – Activation of the safety ring during closing of the lid

The safety system consists of an aluminium ring (safety ring) and a contact folio stuck on top of the
chamber. The contact folio comprises 9 contacts. On the lower of the safety ring are mounted plastic
stags, which can activate one of the contacts in the contact folio, when the safety ring is touched. The
safety ring is elevated from the contact folio by means of adhesive stickers. See Figure 2.21.1.

Figure 2.21.1

The error can occur if:

The user has touched the safety ring


Press C and restart the autoclave.

The lid is not placed right in the holder, so that it touches the safety ring
Press C, place the lid correctly and restart the autoclave.

The three tension rods that goes through the safety ring, touches the safety ring sufficiently enough to
press the ring a bit downwards
Adjust the safety ring manually, so that it does not touch the tension rods.

The small adhesive stickers separating the safety ring and the contact folio are permanently
compressed
Check whether there is air between the plastic stags and the contact folio. (See Figure 2.21.1).

Ventilation under the contact folio is missing


The chamber flange, on which the contact folio is mounted, is equipped with three ventilation holes.
Check whether the contact folio is correctly placed over the ventilation holes. If this is not the case
vacuum may occur in the contact folio, resulting in the contacts being pressed down. Change the
contact folio. See section 3.2.5.

37
2.22 Error 87 – The safety ring is dismounted

The mainboard is constantly checking, whether the contact folio is mounted correctly as regards to
power. If the contact folio is not mounted on the mainboard error 87 will occur.

This error can occur if:

The plug on the mainboard drops off


Check whether the plug is correctly mounted on the mainboard. See section 3.12.1 for an overview of
the cords.

The contact folio is defective


The two outer and the 2 inner cords in the plug must be in contact. The resistance between the 2 outer
pins is checked by means of an ohmmeter. It must be approx. 60 ohms. The resistance between the 2
inner pins must also be approx. 60 ohms. See section 3.2.5 about changing the safety ring.

38
2.23 Error 88 – The steam pressure is too low during sterilisation

This error occurs if the pressure drops below 1.95 bar at 134°C and 0.95 bar at 121°C during the
sterilisation.

The error can occur if:

The pressure sensor is defective


See section 3.8 about the pressure sensor.

The autoclave runs out of water during the sterilisation

The autoclave can be leaking


See whether there is a visible leakage during the sterilisation (liquid leakage).
Check whether steam is coming out through the drain. If this is the case, check the drain valves 6, 7, 8
and 9. See section 3.6 about valve battery, to check and possible change the valves or o-
rings/gaskets in the valves.

39
2.24 Error 89 – The steam pressure is too high during sterilisation

This error occurs if the pressure in the chamber is too high during sterilisation > 2.4 bar.

The error can occur if:

Valve 8 is defective
The tube between the chamber and valve 8 is blocked.
The banjo bolt on valve 8 is blocked due to foreign bodies. Separate and clean it. See section 3.6.5
and 3.6.6.

Valve 1, 2 or 3 is leaking air


Fill water into the bottom of the chamber, until it covers the bottom adapter. If one of the air valves is
leaking air, bubbles will appear. Note that air must be connected to the autoclave.
Valve 1 is leaking air, if bubbles appear from one or more of the 4 holes at the bottom of the chamber.
Valve 2 is leaking air, if bubbles appear from the top hole in the bottom adapter.
Valve 3 is leaking air, if bubbles appear from the side hole on the bottom adapter. See section 3.6
about valve battery, for changing the valve and/or gasket and o-rings.
It is also possible to test the valves via a PC with service software. See section 4.1 to activate the
service software (remember to empty the chamber for water before testing the air valves).

The pressure sensor is defective


See section 3.8 to replace the pressure sensor.

40
2.25 Error 90 – Pressure too low during pressure test

When the lid has closed completely valve 1 is opening till the pressure reaches 4.0 bars absolute
pressure in the chamber. If the 4.0 bars has not been reached within 20 seconds error 90 will occur.

The error can occur if:

The lid of the filter house is not mounted or is not tightened enough.

The pressure to the autoclave is too low


Check whether the pressure to the autoclave is between 5-8 bars.

The autoclave is leaking


Normally this leakage is distinctly heard unless the leakage is going through the drain tube.
The pressure disappears through the drain valve. Dismount the drain tube. If pressure is coming out of
the drain, valve 6, 7, 8 or 9 are leaking. See section 3.6 about valve battery.
The lid is not completely closed.
If the opening/closing system is descending too slowly, there is a risk that the lid is not completely
closed before opening of valve 1. Here you can try to place the lid in the chamber manually and
subsequently start the machine, so that the lid holder presses on the blue plastic. If this helps the
opening/closing system has to be adjusted. See section 3.3 about adjustment of the opening/closing
system.

The pressure sensor is measuring incorrectly


Change the pressure sensor. See section 3.8 about the pressure sensor.

The large black gasket in the lid is not fitting closely. See section 3.7 about lid for replacement of the
gasket.

The motor is defective. See section 3.2.3 and section 3.2.4 to replace the motor.

A tube can be leaking inside the machine. You can probably clearly see or hear the leakage, as it has
to be relatively large to give error 90.

41
2.26 Error 91 – The autoclave is leaking due to leakage in the chamber during pressure test

When the lid has closed completely, air valve 1 is opening till the pressure reaches 4.0 bars absolut
pressure in the chamber.
The pressure value measured is stored and 10 seconds later the pressure is re-measured. The
difference between the measured values 1 and 2 must not exceed 0.15 bar. If the difference exceeds
0.15 bar error 91 will occur.

The error can occur if:

There is a leak somewhere in the autoclave


This leak can occur in several places in the autoclave and can be checked in the following two ways:
Spray soapy water on the places you believe to be leaking and see whether bubbles appear during
the process. NOTE! Make sure not to get soap water on the electrical parts.
Start the autoclave. Wait 10 seconds after you hear valve 1 clicking for the first time. Bend the tube for
the pressure sensor, so that the pressure at the sensor can be maintained. Let go of the break when
you can hear that the autoclave is letting its pressure out. Hereafter the autoclave will run one
sterilisation, and normally you can see where the leak is because water and steam will escape when
there is pressure in the autoclave.

Check for leakage at the following places:

Black gasket between lid and chamber


Place a finger in the V-trace in the gasket or change the gasket. See section 3.7.3 about lid for
changing the gasket.

Under the blue plastic on the lid


Tighten the 4 screws under the blue plastic top. Note that the gasket under the centre column can also
be leaking. See section 3.7.3 about the lid.

At the lid for the filter house


Tighten the lid or change the O-ring on the lid.

All tubes inside the autoclave with connection to the chamber

All bolts on the valve battery. See figure 2.26.1

Figure 2.26.1

The bolts can


be leaking

Look for leaks


around this
bolt too.

Check all assemblings between valves and manifold on the valve battery

42
Figure 2.26.2

Leakage can occur


between the valve and
the valve block, if the
o-rings in between
them are defective

Leaking drain valves


Dismount the drain tube with 12-mm. spanner at drain inlet. If pressure is coming out of drain inlet in
the 10 sec. the DAC Universal should be able to hold the pressure, valve 6, 7, 8 or 9 can be leaking.
See section 3.6 about valve battery.

Bubbles in the water tank


The non-return valve in the small water pump is defective. Change the pump. See section 3.14.2.

Leakage at the bottom adapter at the bottom of the chamber


See section 3.2.1 for replacement of complete chamber.

43
2.27 Error 92 – The autoclave has been disconnected during sterilisation

The error can occur if:

The power to the autoclave is disconnected when the autoclave is running a cycle
Press C and restart.

The cycle has been cancelled Press C and restart.

44
2.28 Error 93 – The steam generator is too hot

If the temperature sensor in the steam generator exceeds 350°C, error 93 occurs. This also releases
the internal safety system, that cuts all power to the components, except for the fan. This is done to
avoid a too hot steam generator and that the thermo switch at the steam generator is released.

This error can occur if:

The temperature in the autoclave becomes to high whereby the regulation on the mainboard does not
function
The fan is defective. See section 3.10.1 about back plate for changing the large fan B.
The autoclave is placed in a too hot spot.
The autoclave is placed too close to a wall with the back plate (minimum 10-cm from wall).

Bad connection in temperature sensor plug at the main board


Dismount the plug at the main board; see figure 3.12.1 for placing of the thermo sensor plug.
Dismount the cords from the plug. Adjust the contacts and replace the cords in the plug. See figure
3.4.2 for dismounting of the cords.
Re-mount the plug at the main board.

Steam generator is defective


The steam generator thermo sensor is defective change the steam generator. See section 3.11.3 and
3.11.4.

The control on the main board is defective


Replace the main board. See section 3.12.

To reset error 93 it is not sufficient to press ”C”. The power to the autoclave must be disconnected for
10 sec. to deactivate the internal safety system.

45
2.29 Error 95 – water consumption during standby

If the level gauge in the water tank is giving a low-level signal while the autoclave is on standby, error
95 will occur. This does also release the internal safety system that cuts all power to the components,
except for the fan.

This error can occur if:

Leaks in the tubing system to the water tank


Look for leaks in the water tube system under tank and to the water pump and steam pump. See
Section 3.4 and 3.14.

A short-circuit has resulted in constant activation of the water pump


Changing the main board will repair the short-circuit (see section 3.12).

To reset error 95 it is not enough to press ”C”. The power to the autoclave must be disconnected to
deactivate the internal safety system.

46
2.30 Error 96 – No pressure from the expel valves

When the steam pressure is out of the chamber after the sterilization and expel is chosen in the menu
of the DAC Universal valve number 11 will open and the pressure from valve 11 opens valve 12 (the
expel valve at the bottom adapter). If the pressure in the chamber, is more than 1.5 bars for 10 sec.
after opening of valve 12 error 96 will occur.

This error can occur if:

Valve 11 does not open:


The gasket and o-rings inside valve 11 are defective. See section 3.6.7 for replacement of the gasket
and o-ring.
The cords are loose at the coil. See section 3.6.2 about valve battery, for correct mounting of the
cords.
The cords can have a bad connection at the main board.
The coil is defective. See section 3.6.7 for replacement of the coil.
Valve 11 is defective. See section 3.6.7 for replacement of the valve.

Valve 12 does not open:


The gasket and o-rings inside valve 12 are defective. See section 3.6. for replacement of the gasket
and o-ring (see section 3.6.11 how to change the o-rings and gasket).

Main board does not open valve 11:


If there is no power for valve 11 when the DAC Universal is running the expel cycle change the main
board (see section 3.12.2).

47
2.31 Error without code

2.31.1 The machine will not open


Press the “Menu/Enter” button. Press the arrow buttons till you reach the manual mode. Press the
“Menu/Enter” button again and move forward to close via the arrow buttons. Press the “Menu/enter”
button to close. If the lid is moving a bit downwards, there is connection to the motor and the following
3 possibilities exist:

Pressure in chamber
Press ”C”. Bypassing the filter is opening a drain valve.
Dismount the drain tube at the drain tank. Note – be very careful it can be hot (135 degrees).
Check the filter at the back plate and drain tube for free passage before pressing ”Start” again.

Limit switch is activated.


The spacer pipe with the limit switch is rotated and the contact is activated. See section 3.2 about
complete chamber
The motor can be defective. See section 3.2 about complete chamber for changing the motor.
If a cord for the motor is jammed between the casing and the bottom plate, the motor control on the
mainboard can burn off. See section 3.12 for changing the mainboard.

The lid holder is jammed on the safety ring


Note! The autoclave must not be started without the lid being mounted.
Remember always to place the lid in the holder before pressing start.
In case the autoclave is being started without the lid in the holder and the holder has pressed the
safety ring down and is unable to go up - please follow below procedure:

1. Place the adapter key in this


opening to reset and open the
machine.

2. Press carefully the adapter key in the


opening until you hear a little click. The
machine will go up for about 5 seconds
and stop “Error 86” will occur.
Afterwards press C to open the
machine completely. The autoclave can
now be used again.

2.31.2 The machine will not close

The lid holder for lid must be adjusted


See section 3.3.1 about opening/closing system.
Motor for the lid defect. See section 3.2.3 and 3.2.4 about the motor.
The motor control on the mainboard is burned off check the 24VDC at the motor plugs when the motor
is started. See section 3.12 for changing the mainboard.

48
2.31.3 The machine will not take in water

Check whether water is being led to the autoclave from the external water supply

Check the cords and the function of valve 5


Via a PC with service software it is possible to activate valve 5. See section 4.1.
For repair of valve 5, see section 3.6.4 about valve 5.
Remember to set automatic water inlet mode to on in the menu of the DAC Universal.

2.31.4 Water in the chamber following the process

Clean or change the filter. Check that the drain tube is not blocked or broken. Empty the drain tank if
this is full.
Valve 9 does not open. See section 3.6 about repair of valve 9.

2.31.5 Metallic sound when the pump is activated

Check whether there are any loose screws under the bottom plate. Check that the rubber suspensions
on all pumps are not defective. See section 3.14.

2.31.6 No oil from the oil dispenser.

Check the oil bottle if there is pressure in the bottle and oil is coming out when pressing the valve of
the oil bottle. See section 3.17.1 how to change the oil bottle.
Check that valve 3 and 10 are working (see section 4.1 about service software how to check the
valves) and there is air pressure at the DAC Universal (more than 5 bars and max 8 bars).
Check the oil dispenser to see if oil is coming out. See section 3.17.2

2.31.7 Error caused by the lid

The 4 bolts with copper washers under blue plastic can leak
Put a bit of soapy water on the bolts during process. If bubbles appear they are leaking. See section
3.7.2 for changing the gaskets under the centre column and copper washers under the four bolts.

Black gasket between chamber and lid can be leaking


If steam is coming out in connection with the sterilisation, the gasket has to be changed. See section
3.7.3 for changing the gasket.

2.31.8 No communication with the PC (ServiceSoftware, Downloader etc.)

Check that a COM cable is used (DB09 – Straight through)

2.31.9 The DAC Universal will not heat up.

Under the safety ring the contact folio is defective


The two outer and the 2 inner cords in the plug must be in contact. See figure 3.12.1 where the safety
folio is connected. The resistance between the 2 outer pins is checked by means of an ohmmeter. It
must be approx. 60 ohms. The resistance between the 2 inner pins must also be approx. 60 ohms.
See section 3.2.5 about changing the safety ring.

Steam generator is defective


The thermo sensor is defective change the steam generator. See section 3.11
The steam generator is defective, the resistance between the red and the black cord has to be approx
56 ohms, if not change the steam generator. See section 3.11

Main board is defective


Change the main board. See section 3.12

49
3. Instruction for repair

The purpose of this chapter is to make it easier to repair the DAC UNIVERSAL.
The chapter is built up around each part in the machine.

Content:
Page

913.1 The casings of the autoclave ............................................................................................... 51

3.2 Complete chamber ................................................................................................................. 53

3.3 Opening/closing system.......................................................................................................... 59

3.4 Water tank/Conductivity sensor ............................................................................................... 61

3.5 Plastic top.............................................................................................................................. 65

3.6 Valve battery .......................................................................................................................... 66

3.7 Lid ......................................................................................................................................... 72

3.8 Pressure sensor ..................................................................................................................... 77

3.9 Temperature sensor ............................................................................................................... 78

3.10 Fans and back plate ............................................................................................................. 79

3.11 Steam generator................................................................................................................... 82

3.12 Mainboard............................................................................................................................ 85

3.13 Automatic water supply ......................................................................................................... 87

3.14 Pumps ................................................................................................................................. 88

3.15 Pressure regulator ................................................................................................................ 90

3.16 Ground Connection at the main board. ................................................................................... 92

3.17 Oil dispenser........................................................................................................................ 93

3.18 Safety parts…………………………………………………………………………………………….….94

50
3.1 Instructions for dismounting the casings

The plastic top is dismounted by loosening the lock fittings placed as shown on Figure 3.1.1. Lifting
from the back plate and forwards then draws off the plastic top. The flat cable to the display on the
plastic top is dismounted. The casing is fixed with 5 thread rods, which are mounted from the top
through the casing.
Figure 3.1.1

Location of the
lock fittings

Figure 3.1.2 shows an overview of the placing of the thread rods. The arrows indicate the placing of
the thread rods. The bottom plate on the picture is shown from the top.
An 8-mm ring/star spanner/socket spanner loosens the casing rods.

Figure 3.1.2

Thread rods for tank


at back plate

Thread rods
5 for casing

51
Figure 3.1.3

Plastic top

Large side
Trilateral casing
casing

Small side
casing

Back plate Tank casing

Fixation screw

Figure 3.1.3 shows an overview of the different casings.

Loosening fixation screw M4x10 Torx screws on the back plate to open the back plate.

The casing of the DAC UNIVERSAL consists of the following:


Trilateral casing
Small side casing
Large side casing
Back plate
Plastic top
Tank casing

52
3.2 Complete chamber

Figure 3.2.1

All white casings have to be dismounted. See section 3.1.

3.2.1 Instructions for dismounting the complete chamber

Dismount the 3 connections (temperature sensor, top switch and safety ring) on the main board. See
figure 3.2.2.
Remove the cords from the connectors at the temperature sensor and top switch. See figure 3.4.2.
Figure 3.2.2

Dismount the
connections here
(temperature sensor,
top switch, safety ring)

53
Dismount the cords to the motor. For later mounting purpose note that the grey cords is placed at the
top and the brown one is placed at the bottom. See figure 3.2.3.

Figure 3.2.3
Dismount the tube at
the filter here.

Loosen the silicone


tubes here

Dismount the tube at the


cross fitting with a 12-mm
spanner. (Connection to
the chamber)

Dismount the PTFE tube


here. The PTFE tube
witch is connected to the
bottom adapter spray
channel.

Dismount the cords


for the motor here
Dismount the PFTE tube over the valves (at the filter). See figure 3.2.3.
Loosen the silicone tube from the water pump . This is done by loosening the clamps and drawing the
tubes off the connecting pieces. See figure 3.2.3
Loosen with an 12-mm spanner the PTFE tube to the bottom adapter spray channel in the chamber.
Dismount the tube mounted at the cross fitting and with connection to the chamber block with a 12-
mm spanner.
The water tank and tank casing are dismounted. See section 3.4.1 and figure 3.4.3 Thread rods for
loosening the tank casing. Loosen the metal tube from the steam generator. See figure 3.2.4.
Hold anti-clock wise with an 12-mm spanner and loosen the metal tube from the steam generator with
a 10-mm spanner.

Figure 3.2.4

Loosen the metal


tube from the steam
generator here

54
Loosen the connection to the drive channels at the oil dispenser and the connection to valve 10. See
figure 3.2.5.

Figure 3.2.5

Loosen the
connection here
with a 12-mm
spanner.

Now it is possible to lift the small water pump out of the holders. See figure 3.14.2
Then slightly tilt the machine to loosen the 4 screws under the bottom plate. See figure 3.2.7

Figure 3.2.7

Screws holding the


complete chamber

Now it is possible to remove the chamber from the machine.

55
3.2.2 Mounting of complete chamber
The mounting sequence is the opposite of the above dismounting.
Please make sure to re-mount the tubes for water pump correctly.
Make sure that the cords (colours) are mounted correctly on the pump and the motor.

3.2.3 Instructions for dismounting the motor


Lift the small water pump out of the holder and dismount the holder (two screws torx 20) for the pump
and valve 10 and 11 then it is possible to get an 8 mm fork spanner under the holder to reach the nut
for belt tightener at the motor plate. See figure 3.2.3.1.
Disengage the belt tightener by loosening the nut with an 8-mm fork spanner. See figure 3.2.3.1.
Dismount the belt wheel on the motor with a 2 mm Allen key (two screws).
The motor is fixed with 3 screws, which are disengaged with a 10-mm fork spanner.
Now the motor can be removed.
Dismount the cords to the motor. For later mounting purpose note that the grey cords is placed at the
top and the brown one is placed at the bottom. See figure 3.2.3.

Figure 3.2.3.1

Nut for belt


tightener.

3.2.4 Instructions for mounting the motor:


The mounting sequence is the opposite of the above dismounting.
Please note the following during the mounting:
To get full benefit from the function of the belt tightener, the motor must be rotated a half turn clock-
wise. If power supply is not available, the motor can be rotated by means of tongs. See figure 3.2.4.1.
Then the belt tightener is fixed by means of an 8-mm fork spanner. See figure 3.2.3.1.
The thin aluminium washer must be mounted in between the motor and the motor plate.
The thick aluminium washer with ball bearing must be mounted under the motor plate.
One of the pointed screws in the belt wheel, must be mounted in such a way, that it hits the flat side of
the motor axle.
The lid holder for lid is adjusted. See section 3.3

56
Figure 3.2.4.1

Belt wheel
is rotated
clockwise

3.2.5 Instructions for dismounting the safety ring


Dismounting:
Remove the lid.
Dismount the lid holder for lid by loosening the 3 screws. See figure 3.2.5.1.

Figure 3.2.5.1

Screws for
dismounting the
lid holder for lid

Remove the attached safety ring, by carefully loosening the adhesive stickers with a knife. Remove
possible remains of the adhesive stickers on the safety ring. See figure 3.2.5.2.
The plug to the contact folio is dismounted from the mainboard.
By means of a knife the contact folio is carefully cut off the chamber flange.

57
Figure 3.2.5.2

Stickers to be removed
from the safety ring with
a knife

3.2.6 Instructions for mounting the safety ring


The mounting sequence is the opposite of the above dismounting.
Please note the following during the mounting:
New contact folio is attached to the chamber and the plug is connected to the mainboard.
Make sure that the folio does not touch the tension rods.
The safety ring with adhesive stickers is mounted. The adhesive stickers are attached by means of
silicone glue.
The lid holder for lid is adjusted. See section 3.3

3.2.7 Instructions for dismounting the limit switch


Dismounting:
Dismount the plug from mainboard. See figure 3.2.5.3.
Dismount 2 screws M2 x 12 on print for limit switch. Now the limit board, with the limit switch, can be
removed.
For placing of the limit switch see figure 2.1.1. at section 2.8
Figure 3.2.5.3

Dismount the limit


switch here

3.2.8 Instructions for mounting the limit switch


The mounting sequence is the opposite of the above dismounting.
Make sure that the limit switch can move freely.

58
3.3 Opening/closing system

3.3.1 Adjustment of the semi-circle for lid


The following instruction has to be followed, if an adjustment block for the lid holder for lid is available
(Figure 3.3.1).
Dismount the 3 screws on the top of the lid holder for lid when it is open. See figure 3.2.5.1
Screw the 3 tension rods down so they are approx. 4 cm above the chamber.
Place the adjustment block in the chamber, in such a way that the three arms fit the tension rods.
Hold the adjustment block down by one hand.
Screw the 3 tension rods up by the other hand, in such a way that they touch the adjustment block.
Remove the adjustment block carefully.
Carefully place the lid holder on the tension rods, in such a way that the holes in the lid holder fit the
holes in the tension rods.
Mount the 3 screws (M6 x 12). The screws are mounted with LocTite 243. Make sure that the tension
rods do not turn during the mounting.
Figure 3.3.1

The following instruction has to be followed, if no adjustment block is available for the lid holder.
Dismount the 3 screws on the top of the lid holder when it is open. See figure 3.2.5.1.
Place the lid in the chamber.
The 3 tension rods are adjusted by means of a slide gauge, in such a way that their heights measured
from the lid are identical (Figure 3.3.2).
Carefully place the lid holder on the tension rods, in such a way that the holes in the lid holder fit the
holes in the tension rods.
Mount the 3 screws (M6 x 12). The screws are mounted with LocTite 243. Make sure that the tension
rods do not turn during the mounting.

59
Figure 3.3.2

60
3.4 Water tank/Conductivity sensor
Reasons for changing the tank:
Defective liquid sensor.

Defective tank.
Leaking conductivity sensor

Defective conductivity sensor

Before doing anything empty the tank for water.

Water tank:
1. Press the “Menu/Enter” button
2. Use the “up” or “down” button until you reach “Manual”.
3. Press the “Menu/Enter” button to select “Manual” mode.
4. Press the “Up” or “Down” button in order to reach “Draintank” in the menu.
5. When “Draintank” is marked, you can press “Menu/Enter” in order to open valve 5. At
the same time, ensure that the automatic water inlet is disconnected, as the water runs
out of water inlet. When the water is out, press the “Enter” button to close valve 5.
6. Pressing the “C” button until the normal display appears.

3.4.1 Dismounting the water tank:

The plastic top of the tank and the trilateral casing are dismounted (see section 3.1).Remove the filter
at the top of the water tank (lift up in the filter).
Lift up in the tank. See figure 3.4.1. Note: Do not lift up in the tank before emptying the tank.

Figure 3.4.1

Remove the filter


and lift up in the
water tank.

1.

3.4.2 Mount the water tank.

The mounting sequence is the opposite of the above dismounting.


Press the water tank down over the water tank adapter and be very careful, not to damage the water
sensor or the conductivity pins.
Note: The water tank has to be sealed by the o-ring at the water tank adapter. See figure 3.4.4.
Mount the water filter again.

3.4.3 Dismount the water tank adapter with sensors.


Dismount the water tank. See section 3.4.1
Dismount the cords from the water sensor at the main board. The water sensor plug is placed at the
top of the main board. (Figure 3.4.3) The cords are removed from the plug by pincers (Figure 3.4. 2).
Disengage the cords by cutting the strips and dismount the cords (make sure not to damage the
remaining cords).

61
Figure 3.4.2

This model shows


how to dismount the
cords from the plug.
NOTE: this is not the
water sensor plug.

Figure 3.4.3

Plug for sensor in


tank

Thread rods for


loosening the tank
casing.

Quick coupling for


the water tubes
(blue).

Loosening and removing the three thread rods disengages the tank casing. The thread rods are
loosened with an 8-mm fork spanner. See Figure 3.4.3.
Now the tank casing can be lifted up and removed.

Dismount the two screws Torx 20 at the bottom plate, to loosen the water tank adapter. See figure
3.4.4
Press at the outer ring at the quick coupling to loosen the tubes. See figure 3.4.3 to see where they
are placed, now it is possible to remove the water tank adapter.

62
Figure 3.4.4

This o-ring seals


between the water
tank and the water
tank adapter.

Water tank adapter

Loosen the two


screws

Dismount the water sensor from the water tank adapter. See figure 3.4.5

Figure 3.4.5

Unscrew the water


sensor and take the
cord out. When
mounting, remember
the o-ring.

Water tank adapter

63
To change the conductivity sensor pins. See figure 3.4.6

Figure 3.4.6

The conductivity
sensor pins are very
easy to take out, just
lift up in the pins.
Remember the little
o-ring when mounting
the pins.

Water tank adapter

3.4.4 Instructions for mounting the water tank


The mounting sequence is the opposite of the above dismounting.

3.4.5 Tubes under the tank


Examine the tubes under the tank and check that the tubes are not bent.
In case of bent tubes try to straighten them – otherwise the tubes are shortened. A bent tube closes
for water to the autoclave. See figure 3.4.3

64
3.5 Plastic top
Reasons for changing the display board:
The display is defective.
The display folio is defective.

3.5.1 Instructions for dismounting the display


The plastic top is dismounted (see section 3.1).
The flat cable between display and keyboard is drawn off the display.
Loosening 4 self-cutting screws disengages the display (see figure 3.5.1).
Please note that the display can be attached to the display folio.

3.5.2 Instructions for mounting the display


The mounting sequence is the opposite of the above dismounting

3.5.3 Instruction for dismounting the display folio


The display folio is glued to the plastic top and is disengaged by carefully loosening it from the edge.
The folio can now be drawn off (see figure 3.5.2).

3.5.4 Instructions for mounting the display folio


The mounting sequence is the opposite of the above dismounting.
Make sure that the flat cable is led all the way through the plastic top, before the display folio is
attached to the plastic top.
Figure 3.5.1

4 self-cutting screws
disengages the display.

Figure 3.5.2

The display folio is


disengaged by carefully
loosening it from the
edge.

65
3.6 Valve battery
Check of valves.
Check of coils.
Dismounting of all valve type.

In order to be able to carry out repairs on the valve battery the casing and the plastic top should be
dismounted. (See section 3.1).

3.6.1 Check of valves


Connect the autoclave to a PC and use the service software program. See section 4.1 under the tab
“Manual” for further instructions.
Test the valves by means of the button for the individual valves. You should be able to hear that they
are clicking.
Valve 11 and 12 are not able to be tested with the service software.
Before opening any of the valves turn off the air pressure and water pressure.

Figure 3.6.1

Valve 4 Valve 8

Valve 3 Valve 7

Valve 2 Valve 6

Valve 1 Valve 9

3.6.2 Check of coils


The valve coil on the valves is equipped with a Bürkert text. The colour of the cords on each valve can
be found on figure 3.6.1. It is possible to dismount the connectors of each valve and mount again (just
draw off and press on).
Figure 3.6.2

Valve 8,4

Valve
1,2,3,6,7 & 9

66
3.6.3 Dismounting of all valve types except valve 5, 8, 4 and expel valve 11 and 12.
Dismount the valve from the valve battery.
Dismount the coil by means of an 11-mm fork spanner. See figure 3.6.3.
Remove the coil.
Disengage the 2 screws that hold the valve house to the block, by means of Torx 10 screwdriver.
Remove the valve house from the block. Open the valve by a Torx 9 screwdriver. Be aware of the 3 o-
rings, the gasket inside the valve and the spring.
Change all O-rings and gaskets on the valve. See table for type of O-rings and gasket before
assembling.
Assembling takes place in the opposite sequence.
See table for spare parts.

Valve Valve Gasket O-ring 1 O-ring 2 O-ring 3 Complete


No Coil type Core on core 3,0*1,5 12,0*1,5 13,0*1,2 valve
1 134077 2,0 NC 335019 708083 708087 708088 708089 728038
2 134077 2,0 NC 335019 708083 708087 708088 708089 728038
3 134077 2,4 NC 335019 708083 708087 708088 708089 728039
4 134076 1,6 NC 335022 708087 708088 708089 708036
5 134076 1,6 NC 335021 708083 708087 708088 708089 728034
6 134076 2,0 NC 335021 708083 708087 708088 708089 708035
7 134076 2,0 NC 335021 708083 708087 708088 708089 708035
8 134076 1,6 NO 335022 708087 708088 708089 728036
9 137226 2,0 NC 335021 708083 708087 708088 708089 728035
10 134076 1,6 NO 335022 708087 708088 708089 728036
11 134076 1,6 NO 335022 708087 708088 708089 728036

Figure 3.6.3

Dismount the coil by


means of an 11-mm
spanner.
Lift up to remove the
coil.

3.6.4 Dismounting valve 5 (valve for external water supply).


Make sure that the external water supply to the machine is disconnected and remove the tube from
the DAC Universal with a 12-mm spanner.
Dismount the plastic top and loosen the fixation screw, at the back plate. See section 3.1.
Open the back plate and with a 12-mm spanner, it is possible to loosen the connector for the water
tube (maybe it is necessary to hold at the other side). See figure 3.6.4.a.
Dismount the tube between the tank and valve 5. See figure 3.6.4.a.
Dismount the coil by means of an 11-mm spanner. See figure 3.6.3. how to loosen the coil and then lift
up, to dismount the coil from the valve.

67
Disengage the 4 screws which hold the valve house by means of Torx 9 screwdriver. Remove the
valve house from the block. Be aware of the o-ring, the gasket inside the valve and the spring. See
figure 3.6.4.b.

Hold against here


Figure 3.6.4a when dismounting Loosen the
valve 5. connector for
the water tube Dismount the
with a 12-mm blue tube with a
spanner 12 mm spanner

Figure 3.6.4b

Mounting of valve 5
The mounting sequence is the opposite of the above dismounting
Be aware of the right gasket inside the valve.
See table for spare parts.

3.6.5 Dismounting valve 8


Loosen the banjo bolt, Teflon tube on the top of valve 8 by a 12-mm fork spanner. See figure 3.6.5
Loosen the banjo bolt at the top of valve 8 with a 9-mm fork spanner.
Dismount the coil by means of an 11-mm fork spanner.
Remove the coil.
Disengage the 2 screws which hold the valve house to the block by means of Torx 10 screwdriver.
Remove the valve house from the block. Open the valve by a Torx 9 screwdriver.
Be aware of the 3 o-rings and the spring inside.

68
Figure 3.6.5

Banjo on valve 8

3.6.6 Mounting of valve 8


The mounting sequence is the opposite of the above dismounting
Be aware of the right gasket inside the valve.
See table for spare parts.

3.6.7 Dismounting of valve 10 and 11

Dismount the Trilateral casing, small side casing and large side casing (see section 3.1)
Loosen and remove the filter with an 8-mm fork spanner. See Figure 3.4.36.
Loosen the coil with an 11-mm spanner and remove the coil. See figure 3.6.3.
Remove the water pump from the holder and let it stay there.
Dismount the tubes at booth sides by means of a 12-mm spanner and remove the tubes. See figure
3.6.6
Dismount the 4 screws which hold the valve at the plate below, by means of a Torx 10 screwdriver.
First remove valve 10 and then valve 11.
Disengage the 4 screws which hold the valve house by means of Torx 9 screwdriver. Remove the
valve house from the block. Be aware of the o-ring, the gasket inside the valve and the spring.
See table for spare parts.

69
Figure 3.6.6

Filter is placed
here at the top.
Dismount with a
8 mm spanner.

Valve 11 is placed
here.

Dismount the
tubes by means of
a 12-mm spanner
at booth sides of
the valves.

Valve 10 is placed
here.

The 4 screws
which hold the
valve.
3.6.8 Mounting of valve 10 and 11

The mounting sequence is the opposite of the above dismounting.

3.6.9 Dismounting of valve 4

Dismount the PTFE


tube at the top of
valve 4 by means of
a 9 mm spanner
and dismount the
banjo bolt with the
same spanner.

Dismount the PTFE tube at the top of valve 4 by means of a 9 mm spanner and dismount the banjo
bolt with the same spanner.
Then valve 4 has to be dismounted as the above valves (1-3, 6-7 and 9). See section 3.6.3
Change the o-rings 708087, 708088 and 708089, the gasket in this valve is in the core
and do not have to be changed.

3.6.10 Mounting of valve 4

The mounting sequence of valve 4 is the opposite of the above dismounting.

3.6.11 Dismounting of valve 12

Dismount the tube to valve 4. See figure 3.6.8


Dismount the 4 screws at the top of valve 12. See figure 3.6.7.
Turn the filter house and dismount the 6/4 PTFE tube with a 12-mm spanner. See figure 3.6.8.

70
Figure 3.6.7

Dismount the 4 screws.


Note there is a spring
inside.

Figure 3.6.8

Dismount the 6/4 mm


PTFE tube with a 12-
mm-spanner.

Dismount the tube to


valve 4 here for more
space.

press the core out of the valve and change the o-rings and gasket. See figure 3.6.9.
Figure 3.6.9

Change the 3 o-rings


and remember to use
silicone grease on the
two o-rings inside the
valve.

Change the gasket on


the core.

3.6.12 Mounting of valve

The mounting sequence of valve 12 is the opposite of the above dismounting.

71
3.7 Lid
Mounting the adapters.
Dismounting the lid.
Mounting the lid.

3.7.1 Mounting the adapters


Figure 3.7.1

Spray channel (outer hole)


Drive channel (inner hole)

Dots are the marking for


adapter position. One dot
is position one and two
dots are position two and
so on.

Flat side of adapter base


Adapter gasket

NOTE! At each adapter position there is a marking with a the number of the adapter position.
At position one, it is not possible to place an instrument which is longer than 14,5 cm. At position two-
six the instrument cannot be longer than 16 cm.

First place the gasket for the adapter on the distributing plate over the spray and drive channels.
The adapter is placed over the gasket.
The flat side of the base on the adapter is turned towards the centre column.
Clean the screws and threaded screw holes on the lid, with alcohol.
Apply just a small drop of Loctite 270 to the tip of the screw. See figure 3.7.2.
Place the screws and fasten them with an Allen key.
NOTE! All the new screws are pre-applied with blue Loctite on the tip.
Figure 3.7.2

72
3.7.2 Dismounting of adapters

The dismounting sequence is the opposite of the mounting sequence.

3.7.3 Dismounting the lid


Remove the blue folio on the top of the lid. It can be removed without being torn.
Loosen the big screw by means of a large straight slot screwdriver. See figure 3.7.3.
Figure 3.7.3

Loosen the screw and


remove the blue plastic
top

Loosen the 4 screws

Loosen the 4 M5x12 mm by means of an Allen key. Note that copper washers are mounted. Now the
complete centre column can be removed.
The black gasket can now be removed from the centre column.
Clean the area on the distribution plate and the centre column carefully, as the gasket is not flexible.
NOTE that it may be necessary to use a knife, to clean the surfaces sufficiently. See figure 3.7.4 and
3.7.5.

73
Figure 3.7.4

Remove old gasket from


the centre column and
clean it carefully

Use a minus screwdriver to


remove the large black gasket

Loosen the 6 screws which hold the distribution plate. See figure 3.7.5.
Figure 3.7.5

6 screws to be
loosened

Remove the old gasket


from the distribution plate
and clean it carefully

Remove the distribution plate and the gasket below. See figure 3.7.6.

74
Figure 3.7.6

Distribution plate

Protection tube for Gasket below


centre column distribution plate

Dismount the protection tube for the centre column. See figure 3.7.6.
Dismount the bottom adapter from the centre column by loosening the 2 side screws. See figure 3.7.7.
Figure 3.7.7

Side screw to be loosened


at the bottom adapter

Change the o-rings on the fluid guiding tubes (two pieces on each tube). NOTE that the spring is
loose. See figure 3.7.8.

75
Figure 3.7.8

Spring

O-rings to be changed

Bottom adapter

Dismount the large black gasket for lid. Use a small minus screwdriver and place it under the black
gasket and pull the gasket off. See figure 3.7.3.

3.7.4 Mounting the lid


The mounting sequence is the opposite of the dismounting.
Please note the following:
The large black gasket must be pressed into the lid with the small minus screwdriver.
The 4 screws which tighten the centre column, should be tightened crosswise, remember to change
the copper washers.

NOTE! When mounting the lid it is important that all o-rings and gaskets are new.

76
3.8 Pressure sensor
Changing the pressure sensor.
Changing the tube on the pressure sensor.

3.8.1 Instructions for dismounting the absolute pressure sensor .


Dismount the plastic top and trilateral, large side, small side casing of the DAC Universal
(see section 3.1.1).
Dismount the PTFE tube from the absolute pressure sensor at the filter housing (see figure 3.8.2).

Figure 3.8.2

Dismount the
PTFE tube from
the absolute
pressure sensor
here.

Dismount the plug at the main board (see figure 3.8.3)


Dismount the thread rod that holds the absolute pressure sensor (see figure 3.8.3).
Figure 3.8.3

Dismount the
thread rod here.

Absolute pressure
sensor.

Dismount the plug


for the absolute
pressure sensor.

3.8.2 Mounting of absolute pressure sensor:


The mounting sequence is the opposite of the above dismounting
Mount the casing and plastic top again (see section 3.1.1 in the service manual).

77
3.9 Temperature sensor
Changing the temperature sensor.
Dismount the plastic top (see section 3.1):
Dismount the plug on the print (see figure 3.9.2).
The cords are disengaged from the plug by means of pincers (see figure 3.4.2)
Dismount the sensor by means of a high 13-mm top with extension, so it can reach above the
chamber edge (Figure 3.9.1).

Figure 3.9.1

The temperature sensor


is placed below the tool.
It is dismounted in this
way.

Mounting the temperature sensor:


Mount a new sensor. Loctite 620 is used for sealing between the temperature sensor and the
chamber.
Connectors are mounted in the plug.
Mount the plug on the mainboard.
Mount the casing and the plastic top (se section 3.1).
Figure 3.9.2

Dismount the plug for


the temperature
sensor from the
mainboard

78
3.10 Fans and back plate
Changing the fans.
Changing the filter house.

3.10.1 Instructions for dismounting the large fan (fan B at the mainboard)
Dismount the plastic top, the casing and the tank (see section 3.1 and section 3.4)
Dismount the cords from the fan at the mainboard (see 3.12.1). The cords are disengaged from the
plug by means of pincers (Figure 3.4.2).
Loosen the fixation screw on the back plate and open it. See figure 3.1.3
Dismount the fan and the grid for fan by disengaging 4 pcs. M4x12 screws.

3.10.2 Instructions for mounting of the large fan


The mounting sequence is the opposite of the above dismounting.
Figure 3.10.1

Large fan (fan B)

3.10.3 Instructions for dismounting the small fan (fan A at the mainboard)
Dismount the plastic top, the casing (see section 3.1 )
Dismount the cords from the fan at the mainboard (see 3.12.1). The cords are disengaged from the
plug by means of pincers (Figure 3.4.2).
Dismount the fan by disengaging 4 pcs. M4x25 screws.
Cut all the strips holding the cords. Make sure not to damage the cords. See figure 3.10.2.

3.10.4 Instructions for mounting of the small fan


The mounting sequence is the opposite of the above dismounting.

79
Figure 3.10.2

Cut the strips here

Small fan (fan A)

3.10.5 Instructions for dismounting the filter house


Dismount the plastic top (see section 3.1)
Loosen the fixation screw on the back plate and open it. See figure 3.1.3.
Screw off the lid for filter house.
Loosening the 2 screws when the back plate is open disengages the filter. M3x6 screws (figure 3.10.3)
Disengaging the unions (figure 3.10.4) loosens drain tubes on the filter.
Figure 3.10.3

Screws holding
the filter house

80
Figure 3.10.4

Loosen the tubes


from the filter house
with a 12-mm and 9-
mm spanner

3.10.6 Instructions for mounting of the filter house


The mounting sequence is the opposite of the above dismounting
Please note that the filter house can be hot.

81
3.11 Steam generator
Reset of the thermo-switch.
Test of steam generator (flow).
Changing the steam generator.
Changing the non-return valve

3.11.1 Reset of thermo switch


Remove the plastic top and the trilateral casing. (See section 3.1).
Press the button on the thermo switch down either by a finger or a screwdriver. Note that the button
must be pressed totally down. See figure 3.11.1. Note: The steam generator can be hot.
Re-mount the trilateral casing and the plastic top.
Figure 3.11.1

Thermo switch

Tube after non-


return valve

Non-return valve

82
Test of steam generator (Flow)
Dismount the plastic top and the trilateral casing (see section 3.1).
Dismount the bottom tube after the non-return valve mounted on the steam pump (10-mm key) see
figure 3.11.1.
Blow air through the bottom tube and the steam generator. It can be expedient to place a cloth or
similar thing over the chamber, to avoid water sprays.
If air is coming up through the holes at the bottom of the chamber, there is passage through the steam
generator. However, there must be a distinct flow.
If there is no air coming up, or only a bit of air is coming through to the chamber, the steam generator
is blocked and must be changed (see section 3.11.2)
If there is free passage through the steam generator, the non-return valve is checked in the same way.
The non-return valve is dismounted at the steam pump and is blown through. See section 3.11.5.
Figure 3.11.2

Blow air through the


steam generator to the
tank to check whether
there is passage for air

3.11.2 Instructions for dismounting the steam generator


Dismount the plastic top, the casing and the tank casing (see section 3.1 and 3.4).
Loosen the fixation screw on the back plate and open it. See figure 3.1.3
Dismount the plugs from the steam generator sensor at the mainboard (see figure 3.12.1). The
sensor cords are disengaged by means of pincers (Figure 3.4.2). Dismount the cords (take care not to
damage the remaining cords).
Dismount the main power supply to the steam generator at the bottom of the mainboard (see figure
3.12.1)
Dismount the ground connection (see figure 3.11.3)
Dismount the bottom tube connections to the steam generator. I.e. between the steam generator and
the chamber (see figure 3.2.4) and between the steam pump and the steam generator (after the non-
return valve). Use a 10-mm key and hold contra with a 16-mm key. See figure 3.11.1.
The steam generator is mounted on the bottom plate with 4 pcs. M4x10 screws, which are loosened,
whereby the steam generator is disengaged from the bottom plate.
Now the steam generator can be removed.
83
Figure 3.11.3

Ground
connection

3.11.3 Instructions for mounting the steam generator.


Mount a new steam generator in the opposite sequence.
Be aware of the following:
Glass flex must be mounted on both bottom tubes on the steam generator.
The bottom tube and the non-return valve must be mounted in order for it not to scrape against the
other parts of the machine.
The bottom tubes must not break; they must have soft bows.
Note : After mounting the steam generator, it is very important to adjust the thermo sensor version in
the service software. See section 4.1 under “Static data”.

3.11.4 Instructions for dismounting the non-return valve


Dismount the plastic top and the trilateral casing (see figure 3.1)
Dismount the bottom tube after the non-return valve (see figure 3.11.1) Dismount with a 10-mm
spanner. Hold contra with a 16-mm spanner.
Lift up the steam pump from the holder and loosen (turn) the non-return valve between the pump and
the 90-degree bending.

3.11.5 Instructions for mounting the non-return valve


The mounting sequence of the non-return valve is the opposite of the dismounting. Be aware of the
flow direction of the non-return valve.
The non-return valve is mounted with Teflon tape at the pump (be careful not to mount too much
Teflon tape because the steam pump will break).
The bottom tube is mounted with a non-return valve with a 10-mm spanner. Hold contra with a 16-mm
spanner.

84
3.12 Mainboard
Overview of cords/tension.
Changing the mainboard
Clamping of cords.
Specifications on fuses

3.12.1 Overview of cords


The below pictures shows the main board with a description of the cords and the tension.

Figure 3.12.1

Display Temperature sensor


Safety folio Limit
(safety ring) switch Tank sensor

Fan A
Fan B

Steam generator,
thermo sensor

Valve battery 11 valves

Pressure sensor

Motor, braun/grey

Red/blue, not in use

Water pump, Yellow/black

Steam pump, orange/White

Conductivity sensor, orange/white


or black/red.

Steam generator

85
3.12.2 Instructions for dismounting the mainboard
Dismount the plastic top (see section 3.1)
The back plate is opened by loosening the fixation screw (see figure 3.1.3) and all cords are
dismounted. See figure 2.12.1. Note that the ground cords are mounted in the bottom plate.
Loosening 7 pcs M3x6 screws from the back plate, 2 pcs. self-cutting screws around the main supply
plug and 2 pcs. stags around the Com plug disengages the mainboard. See figure 3.12.2
Figure 3.12.2

Bolts for COM plug

7 screws holding the


mainboard

Screws for
power plug

3.12.3 Instructions for mounting the mainboard:


The mounting sequence is the opposite of the above dismounting.
Please note correct ESD handling of the mainboard.
Mount the cords cf. section 3.12.1.

3.12.4 Specification on fuses

Three fuses are placed on the mainboard.


Characteristics: Miniature fuses
Size: 5 x 20 mm
Type: T2A 2 ampere fuse
T6,3A 6,3 ampere fuse
T12,5A 12,5 ampere fuse

86
3.13 Automatic water supply

Water can be connected from an ion exchange system, in order for the autoclave to take water in
automatically.
Water from the ion exchange system, is lead into the autoclave in a 6-mm tube and is connected to
the water inlet on the back plate Note: there has to be a water filter (20 micron) mounted at the water
inlet.
Remember that the level of microsiemens must be 3µS or lower.
If automatic water supply is used, this must be stored as “on” in the memory of the autoclave, by using
the menu point ”Automatic water” under the menu ”Settings”. See the instruction manual for detailed
description.
See figure 1.2.1 where to connect the water supply.

Nitram recommends the use of a NitraDem unit, for automatic water supply, part no. 750100.

87
3.14 Pumps
Changing the steam pump.
Changing the water pump.

3.14.1 Instructions for changing the steam pump


Empty the water tank. See section 3.4.
Dismount the plastic top and the casing (see section 3.1).
Dismount the connectors on the steam pump. Note the colour sequence (top white, bottom orange).
Dismount the bottom tube after the non-return valve. Dismount with a 10-mm spanner. Hold contra
with a 16-mm spanner.
Draw the silicone tube off the steam pump.
Now it is possible to remove the pump from the holder.
Dismount the non-return valve. (See section 3.11.5)
Re-mount in the opposite sequence.
When remounting the non-return valve, Teflon tape must be used (be careful not to mount too much
teflon tape because the steam pump will break).
Figure 3.14.1

Dismount the tube here

Dismount the tube here

Dismount the cords for the


pump here

3.14.2 Instructions for changing the water pump.


Empty the water tank. See section 3.4.
Dismount the plastic top and the casing. (See section 3.1).
The pump is mounted on the side of the chamber. (See figure 3.14.2)
Dismount the connectors on the pump. Note the colour sequence (Black/Yellow cords).
Draw the tube connection from the tank off the pump.
Dismount the thick tube connections against the valve battery, by loosening the clamp.
Now it is possible to remove the pump from the holder.
Re-mount in the opposite sequence.

88
Figure 3.14.2

Water pump

89
3.15 Pressure regulator

3.15.1 Change of the pressure regulator

Dismount all the casings including the tank and oil bottle (remember to empty the water tank). See
section 3.1 and section 3.4 about casings and tank.
Dismount the washer for the pressure regulator. See figure 3.15.1.

Figure 3.15.1

Dismount the washer here

Loosen both teflon tubes on the regulator with a 12-mm spanner. See figure 3.15.2.

Figure 3.15.2

Now the pressure regulator is detached.

The pressure regulator is mounted in the opposite sequence.

90
3.15.2 Adjustment of the pressure regulator

The machine must be connected to power, drain and compressed air ready for start. (the inlet
pressure must be between 5 and 8 bar).
Loosen the fixation screw holding the back plate (see figure 3.1.3 ) and open it. Take care not to
damage the mainboard. Connect a computer with the ServiceSoftware installed and open the software
under “manual”. See section 4.1.
Mount the spec. tool with manometer on the button adapter in the chamber (see figure 3.15.4) Note:
Hold it down and never press on valve 2. Click on valve 3 in the ServiceSoftware, to let air into the
button adapter. Check the manometer to see, whether the pressure is 5 bar as required. If the
manometer does not show 5 bars, the pressure regulator must be adjusted to 5 bars, by means of the
filter key. See figure 3.15.3.
When the 5 bars appears at the manometer, repeat the above 2 or 3 times, to be sure that you have
the right adjustment.
Note: The filter key must not touch the main board.
Figure 3.15.3

Figure 3.15.4 Spec. tool with manometer.

Place the spec. tool at the


button adapter in the
chamber here and hold it
down with your hand.
Note: The chamber might
be hot - be careful.

When valve 3 is opened


with the service software
there would be 5 bar air
pressure on the
manometer, if the pressure
regulator is adjusted right.

Finally the back plate is closed, the fixation screws are tightened and the adjustment has finished.

91
3.16 Ground Connection at the main board.

See figure 3.16.1 where ground is connected to the button plate.


Figure 3.16.1

Ground from main board


to the button plate.

92
3.17 Oil dispenser
Replacement of the oil bottle
Check the oil in chamber.
Change the oil dispenser

3.17.1 Replacement of the oil bottle

o Open the blue top lid of the tank system.


o Unscrew the oil bottle. See how on Figure 3.17.1
o Lift up in the oil bottle and remove the bottle.
o The oil bottle is mounted in the opposite sequence.

Figure 3.17.1

Unscrew the oil bottle in


this direction.

3.17.2 Check the oil in the chamber

Oil in the chamber bottom.


o Choose the Lubricate button at the display and press for more than 1 sec.
o Through visual control, check that oil appears in the bottom adapter, in the sterilisation
chamber. It is necessary to press the lubricate button 4-5 times before oil appears.
o If there is no oil, change the oil bottle. See section 3.17.1
o If there is no oil, check valve 3 and 10. See section 4.1 about the Service software how to
test the valves.

3.17.3 Change the oil dispenser

o The plastic top of the tank and the trilateral casing are dismounted (see section 3.1).
o Loosening and removing the three thread rods disengages the tank casing. The thread rods
are loosened with an 8-mm fork spanner. See figure 3.4.3
o Remove the blue top from the tank.
o Remove the water tank. See section 3.4
o Remove the oil bottle. See section 3.17
o Remove the tank casing carefully. Thread rods for loosening the tank see Figure 3.4.3
o Dismount the connectors for the pressure switch. See figure 3.17.2
o Dismount the tubes from the dispenser. See Figure 3.17.2

93
Figure 3.17.2

The pressure switch and


the connectors needs to
be dismounted.

Dismount the tubes with


a 12-mm spanner and
take the tube out.

o Dismount the 3 screws holding the oil dispenser. See figure 3.17.3.
Figure 3.17.3

Unscrew the 3 screws with


a Torx 20 screwdriver.
Remove the dispenser.

o Dismount the pressure switch, outlet for the oil and pressure connection from valve 10.
o Remember when mounting it is important to change the gasket between the pressure
switch and the dispenser.
Figure 3.17.4

Dismount the pressure


switch from the oil
dispenser here.

Dismount the oil outlet


here.

Dismount the connection to


valve 10 here.

3.17.4 Mounting the oil dispenser.

o The oil dispenser is mounted in the opposite sequence.

94
3.18 Safety parts

Safety valve.

Safety valve will open if


the pressure is more
than 3,5 bars relative in
the chamber.

Safety ring.

Press the safety ring


and the motor for the lid
stops.

There are 9 contacts in the safety ring, which are checked, before start of a cycle and the circuit is
always checked, when the power is on.

Watchdog.

Watchdog relays no. one


and two, switches off the
power to the pumps,
motor, valves and heater.

If the microprocessor does not send a signal to the relays they will switch off the power to the pumps,
valves, motor and heater.

95
4. Software
Programs for carrying out service are available for the DAC Universal.
The purpose of the programs is to make it easier, to carry out the service and download the most
recent software in the machine.

Content:
Page

4.1 ServiceSoftware ..................................................................................................................... 97

4.2 Download new software in the DAC Universal ........................................................................ 102

96
4.1 ServiceSoftware
The service software program is used for error finding in the DAC UNIVERSAL. By means of the
program, it is possible to monitor the temperature and the pressure in the chamber, the temperature of
the steam generator and of the water quality (conductivity). It is also possible to activate the valves or
to open/close the lid.
Connect the DAC UNIVERSAL to power and connect the machine from the COM connector (See
Figure 1.2.1) with a 9 pin straight through RS232 cable, to a computer with servicesoftware installed.

You are able to get the servicesoftware from our web side or from a service CD. (See section 4.2)

Open the servicesoftware:

Select the COM port: If your computer has a serial RS 232 port use COM 1 and click “OK” and then
the servicesoftware opens.
If your computer only has a USB port then you need to use the COM port that works with your
computers USB connection.

Press here to select


the COM port.

Press “OK” to open


the servicesoftware.

Now the servicesoftware opens.

If your computer does not have a serial RS 232 port but only a USB port you will need an adapter like
this. (spare part no. 700086)

97
The picture below shows the program under the tab “AD”:

Press “AD” to show


the tab AD.

These fields are used


to set the various
curves

Updating the speed of


the graphs

Resets the window for


a new curve

To update settings
press “Update
parameters”

When the coloured fields are marked, a graph showing the parameter chosen, will appear. The scale
of the y-axis is indicated in divisions from the a/d converter, which will be between 0-1024 irrespective
of the graph required.

In the field at the right just below the graph, it is possible to indicate the up-dating speed of the graphs.
The default value is 1000.

Press Update parameters for reading the data from the DAC UNIVERSAL.
Press “Clear graph” for resetting the curves.
The field Runs 134°C, Runs 121°C and 134 °C 18 min. Shows how many runs the DAC UNIVERSAL
have made in total in the different programs.
The field “Software version” indicates the version of software in the DAC UNIVERSAL.
“Packets requested” are indicating the communication between the autoclave and the computer. This
value should be between 1-5. The larger the queue, the longer is the time, from activation of one
button till the action is happening on the autoclave.
The field ”state” indicates which state (chapter) the software is in. State 3 is standby.
The field “Error” indicates a possible error code.
The field ”Chamber pressure” indicates the pressure in the chamber (red curve).
The field “Chamber temperature” indicates the temperature in the chamber (green curve).
The field “ Heatelement temperatur” indicates the temperature in the heater (blue curve).
The field “ Conductivity sensor” indicates the water quality (pink curve).

98
The picture below shows the possibilities under the tab ”Manual”:

Press “Manual” to
show the tab Manual.

Activate and deactivate


the valves on these 11
buttons

Activate the steam


pump

Activate the water


pump

Steam generator is red


when it is turned on.

Motor up, down and


stop

The 11 buttons (V1 – V11) are used to activate or deactivate the valves. It is also possible to press F1
–F9 on the keyboard at the computer to activate or deactivate from valve 1 - 9.

Next to the buttons, the current status of the valves, are shown. If they are red, they are activated and
open, if they are black, they are closed.
However, valve 8 is an exception. As it is normally open, whereas the others are normally closed, the
valve is red when closed, and black when it is open.

By means of the “Up”, “Down”, and “stop motor” it is possible to activate the motor, in order to open or
close the machine.

It is also possible to see, whether the steam generator is heating and whether some of the pumps are
running.

With the buttons “Steam” and “Water” it is possible to activate or deactivate the steam and water
pumps.

NOTE! When pressing the steam button, hot steam is emerging from the chamber. Keep away from
the chamber and be ready to stop the steam pump, by pressing the steam button once again.

99
The picture below shows the possibilities under the tab ”Static data”:
Click on ”Options” and
”Enable programming”
to turn on the set
buttons

Adjustment of
water quality

Heating element
setpoint

Main board is adjusted


to version 8.

Pressure sensor is
adjusted to Keller
absolut sensor (above
serial no 4490).

Important: Thermo
sensor version is
adjusted to the colour
of the cords from the
steam generator
temperature sensor.
Look at the top of main
board to see the colour

Oil system is adjusted


to the oil system the
DAC Universal use.
Press “Static data” to Adjustment: “Expel unit implemented” or Normally “oil
show the tab Static data. “Expel unit not implemented” (drying cycle). dispenser” but old
version “oil pump”.
Pressing “Enable programming” under the menu point “Options”, activates the
buttons.

The “Heating element set point”, indicates which A/D value the autoclave consider as 300°C / 572°F.
There is a tolerance on the electronic components, for which reason it is necessary to calibrate each
board. However, the boards are pre-calibrated from the factory.

By adjusting the conductivity, it is possible to change DAC UNIVERSAL’s accepted water quality limit.
Enter the field “water quality limit” and write the indicated value by pressing the arrows. Default
setting is 130. Press “set water quality”. (Do not adjust to more than 130 because this will block the
steam generator and the steam generator has to be exchanged).

Hardware version:

“Mainboard version” is adjusted to version 8.


“Pressure sensor” is adjusted to the pressure sensor witch is inside the DAC Universal.
“Thermo sensor version” is adjusted to the colour of the cords from the temperature sensor in the
steam generator see figure 3.12.1 to find the cords from the steam generator, thermo sensor.
“Oil system” is adjusted to the oil system the DAC Universal use (normally oil dispenser if there is an
Oil bottle mounted in the DAC Universal).
“Expel unit” is adjusted to “Expel unit implemented” (3.3.6 software or more) if the machine has a
serial no. above 101000 Sirona DAC and 10000 Nitram DAC. Below this serial no. adjust to “Expel
unit not implemented”.

Press “Set HW version” after adjusting the Hardware version.

100
The picture below shows the possibilities under the tab ”Batch”:

Press “Batch” to
show the tab
Batch.

Set serial number

After adjustment of
batch press ”Set
Batch”.

Pressing “Enable programming” under the menu point “Options”, activates the buttons.

From here it is possible to read or set the serial number. If the serial number should be changed, (new
Mainboard) the new serial number is written in the field and the button “Set Batch” is pushed. Note:
The serial number must be 6 digits. Id. no.112 is written as 000112.

The important parts in the DAC Universal, has an identification number. If one of the parts has to be
changed, it is possible to change the ID. number in the column, which is related to the replaced part.
The new part number will be stored in the memory, when Set Batch is pressed.
So it is not necessary to open the DAC Universal, to see the ID number of the main parts.

101
The picture below shows the possibilities under the tab ”Service Tools”: Press here to reset error
65.

Press “Service Tools” to


show the tab Service
Tools.

Clear runs until service


( 2500 runs ).

Set the V sign in auto


test to start test.

Press the arrow up or


down to set the number
of test runs.

Total testruns performed

Date for the last yearly


service.

When was the DAC


Universal installed for
the first time.
Reset of error 65:
Press the button “ Reset error 65” and press yes to the warning.

Runs until service:


After one year or 2500 runs, a service must be carried out. See section 1.4 about yearly service.
The alarm led will flash without a sound and then the customer knows that it is time for yearly service.
Reset of the service runs: Pressing “Enable programming” under the menu point “Options”, activates
the button “Clear Service Runs” then press “Clear Service Runs” and press yes to yearly service has
been carried out.

Date for last service: This will inform the service technician when the last yearly service was made.
This date will be changed into the day for the next yearly service when the technician press the button
“Clear Service Runs”. (Please remember that it is not ok to press “Clear Service Runs” before a yearly
service is carried out. See section 1.4).

Date for start of use: This is the date where the DAC Universal was started for the first time (the start
button was pressed for the first time after leaving Nitram Dental).

Automatic cycling:
When a service has been carried out, it is necessary to test the DAC Universal before it is shipped to
the customer.
Press “ Clear “ button on the DAC Universal before starting Automatic cycling.
Press the arrows in “Automatic cycling” to set the numbers of test runs.
Set the V sign in “Auto test” to start the Automatic cycling program. (To stop test remove the V sign).
Now the DAC Universal will test itself automatically for chosen number of runs, without pressing start
every time.
About Download software to autoclave. See section 4.2

The tab ”Nitram” contains some possibilities for changing the basic settings for the autoclave. These
parameters are protected by a password. Thereby vital functions are not changed by error.
102
4.2 Download new software in the autoclave

From a service CD-Rom, you are able to download the programs into the DAC UNIVERSAL, or it is
possible to get and download the new revision of software/Servicesoftware from our web side
www.nitramdental.com. You just need the username and password of your company.
Usernames and passwords are something each company have received and is informed about.

Open our home page


(www.nitramdental.com).
Press “contact”, ”Dealer
Customer Service”, “login”
and then write username,
password and press
“service” then “software”.

When you want to download new software to the DAC UNIVERSAL, connect the DAC UNIVERSAL to
power and connect the machine from the COM connector (See Figure 1.2.1) with a 9 pin straight
through RS232 cable to a computer.

4.2.1 Download software automatic.


Open the Service Software and press “Service Tools”

1. Press Download automatic


2. Press the version of software
3. Press open
4. Now the computer downloads software into the DAC Universal

First you will have to download the software file ( “ 3_3_6.hex “ or more).
Then you will have to press “Download Language”. Below the button “Download Language” the
service software writes the language file you need to download. On the figure 4.3 the service software
writes: Memory version needed for this software : I. then press the language file “Language_I” and
press “open”. Now you are able to see a countdown at “Packets requested” when the counter is down
to 1-3 the software is installed.

103
Figure 4.1

Press version of
software

Press open to start


downloading the
software.

Press “Download
sw automatic” to
start downloading.

Figure 4.2

When the download


is finished, the
software is installed.

104
Figure 4.3

2. Press
“Language_I” and
press “open”

3. After pressing
“open” you will see a
countdown.
When the counter is
down to 1-3 the
language file is
installed.

1. Press “Download
Language”

4.2.2 Download manually.

1. Press “Download manually”


2. Press “Load SW” ( Load software)
3. Press version of software
4. Press “open”
5. Press “Boot Mode”
6. Press “Program”

Figure 4.4

Press Load SW

Press Download
manually

105
Figure 4.5

Press version of
software and press
“open”.

Figure 4.6

When the download


is finished the
software is installed.

Press “Boot Mode”


and then “Program”
when the bottom is
on.

Remember to download the language file as in automatic download. (See figure 4.3)
This was the two different ways of downloading software to the DAC Universal.

106
5. Spore test
Description of procedure:
Nitram Dental has used spore tests from the Danish Serum Institute, which is a certified institute.
The test is made in the form of a letter test and the letter with the spore, is fixed to the centre column
of the lid by means of a holder for indicator spare part no. 700018. Subsequently the autoclave
must run a cycle and the sample must be sent for testing, in order to check for growth in the sample.
It is also possible to place a 3M biological test, in the same way as described as above.

107
6. Pressure and temperature

A 134 degrees standard cycle:

Sec. Temp. Pres.


2.484 19.406 0.976
7.484 19.406 0.981
12.484 19.406 0.976
17.484 19.406 0.976
22.484 19.406 0.976
27.484 19.406 0.976
32.484 19.406 0.976
37.484 19.406 0.976
42.484 19.406 0.976
47.484 19.406 0.976
52.484 19.406 0.976
57.484 19.406 0.976
62.484 19.406 0.976
67.484 21.007 2.509
72.484 24.208 4.391
77.484 23.408 4.313
82.484 22.607 4.021
87.484 20.207 2.256
92.484 18.606 1.323
97.484 17.806 1.649
102.484 19.406 2.162
107.484 19.406 2.515
112.484 20.207 2.989
117.484 20.207 3.05
122.484 20.207 3.105
127.484 18.606 1.599
132.484 17.806 1.406
137.484 17.806 1.009
142.484 17.806 0.976
147.484 17.806 0.976
108
152.484 18.606 0.976
157.484 18.606 0.976
162.484 19.406 1.047
167.484 21.007 1.24
172.484 22.607 1.246
177.484 22.607 1.24
182.484 23.408 1.29
187.484 24.208 2.851
192.484 25.808 4.137
197.484 24.208 3.491
202.484 22.607 2.338
207.484 21.007 1.356
212.484 20.207 1.02
217.484 20.207 0.976
222.484 20.207 0.976
227.484 21.007 0.976
232.484 21.007 0.976
237.484 21.807 0.981
242.484 33.249 1.163
247.484 38.951 1.202
252.484 40.472 1.323
257.484 40.852 1.301
262.484 40.472 1.251
267.484 40.852 3.392
272.484 40.472 4.236
277.484 38.571 3.32
282.484 36.67 2.901
287.484 36.29 2.824
292.484 35.53 2.774
297.484 34.769 1.505
302.484 33.629 1.003
307.484 49.512 1.108
312.484 70.806 0.992
317.484 78.206 0.987
322.484 83.622 0.998
327.484 90.963 0.998
332.484 95.591 1.003
337.484 98.19 1.047
342.484 101.238 1.152
347.484 104.735 1.284
352.484 108.321 1.489
357.484 112.803 1.698
362.484 117.286 1.924
367.484 121.212 2.173
372.484 125.035 2.443
377.484 129.012 2.752
382.484 133.002 3.089
387.484 132.694 2.901
392.484 130.465 2.719
397.484 131.841 3.039
402.484 134.622 3.155
407.484 135.007 3.144
412.484 134.005 3.188
417.484 134.313 3.204
422.484 135.47 3.26
427.484 135.624 3.155
432.484 135.316 3.072
437.484 135.007 3.039

109
442.484 134.699 3.078
447.484 135.084 3.138
452.484 135.084 3.155
457.484 135.007 3.182
462.484 135.161 3.171
467.484 135.084 3.155
472.484 135.007 3.166
477.484 135.007 3.177
482.484 135.007 3.149
487.484 135.084 3.155
492.484 135.007 3.16
497.484 135.007 3.16
502.484 135.007 3.155
507.484 135.007 3.177
512.484 135.007 3.16
517.484 135.007 3.144
522.484 135.084 3.16
527.484 135.007 3.149
532.484 135.007 3.149
537.484 135.007 3.149
542.484 135.007 3.155
547.484 135.007 3.16
552.484 135.007 3.149
557.484 135.007 3.149
562.484 135.007 3.149
567.484 135.007 3.177
572.484 135.007 3.177
577.484 135.007 3.171
582.484 135.007 3.171
587.484 135.007 3.16
592.484 135.007 3.171
597.484 135.007 3.171
602.484 135.007 3.16
607.484 135.084 3.149
612.484 135.007 3.155
617.484 135.007 3.16
622.484 135.007 3.171
627.484 135.007 3.166
632.484 135.007 3.155
637.484 135.007 3.166
642.484 135.007 3.171
647.484 135.084 3.155
652.484 135.007 3.171
657.484 135.084 3.16
662.484 135.007 3.177
667.484 135.084 3.138
672.484 134.93 3.039
677.484 134.467 3.078
682.484 134.39 3.133
687.484 134.545 3.204
692.484 134.699 3.249
697.484 134.776 3.249
702.484 134.853 3.227
707.484 132.848 2.719
712.484 122.971 1.787
717.484 114.865 1.45
722.484 109.934 1.29
727.484 107.155 1.136

110
732.484 103.48 1.08
737.484 100.252 0.981
742.484 99.894 0.981
747.484 99.625 0.976
752.484 99.356 0.976
757.484 99.087 0.976
762.484 98.908 0.976
767.484 98.639 0.976
772.484 98.011 0.976
777.484 97.652 0.976
782.484 97.473 0.976
787.484 97.384 0.976
792.484 97.384 0.976
797.484 97.384 0.976
802.484 97.294 0.976
807.484 97.115 0.976

111
7. Appendix – Diagram

Overview of diagram

Dr Sp
iv ra

Valve 12

112
8. Revision history

Revision Date (Y-M-D) Responsible Description of change


No
US 1.1 2004-02-04 MH Initial release.
US 1.2 2004-05-12 MH
US 1.3 2004-12-23 MH
US 1.4 2005-02-04 MH
US 1.5 2005-06-09 MH
EU 1.6 2005-10-10 MH
EU 1.8 2007-06-07 MH

113

S-ar putea să vă placă și