Documente Academic
Documente Profesional
Documente Cultură
I. INTRODUCTION
Method validation is the process used to confirm that the proposed analytical
procedures is suitable for its intended use. Results from method validation it is used to
judge the quality, reliability and consistency of analytical results; it is an integral part of
a good analytical practice. The further documentation it is the process of defining an
analytical requirement and confirms that the method under consideration has
performance capabilities consistent with what the application requires.
The analytical method validation has been performed as per ICH (International
Commission of Harmonization) guideline Q2A and Q2B
II. PURPOSE
The objective of the present work was to develop validated analytical method
with the help of which we can quantify CLORSULON from the injectable solution
IVERMECTIN-CLORSULON INJ.
Therefore, there is always a need to develop validated analytical method which is
precise, accurate, selective, and sensitive and can be used for routine analysis and
stability studies of the drug product IVERMECTIN-CLORSULON INJ.
1. Label claim
table no.01 IVERMECTIN-CLORSULON INJ composition per one milliliter of
solution:
3. Method description
The ingredient Clorsulon is identified and quantitatively assayed
with a high-performance liquid chromatographic method. This permits the identification
and quantification of the ingredient in one run, also separation and identification of the
impurities. Clorsulon is identified by comparing the retention time with the reference
standard. Quantifications followed by measurement of the peak area of Clorsulon.
3.1. Equipment
a) HPLC system
-flow: 1.0 ml/min
-UV detector 254 ±10 nm; D2 -lamp
-Runtime 70 minutes
-column temperature 30°C± 2°C
- Column Symmeetry C8, 250x4 mm,5µ
-sample loop 20µl
3.2. Chemicals
b) Mobile phase
• Phase A: weight 700 g water HPLC grade in a beaker. Add 236 gram of
acetonitril HPLC grade. Homogenise. Add 1.05 g acetic acid, homogenise.
• Phase B: Acetonitril HPLC grade
𝐕𝐝𝟐 𝐕𝐝𝟏
Dilution factor, F1 = 𝐱 , where
𝐕𝐩 𝐕𝐬
Vs – the volume of assay sample (ex,1,7 ml IVERMECTIN-CLORSULON INJ)
Vd1 – the volume the first volumetric flask (ex,50ml)
Vp – the volume transferred by pipetting (ex, 1 ml of sample)
V d2 – the volume of second volumetric flask (ex, 20 ml)
F1 = 588,2353 (20/1x50/1,7), if the dilution of the sample is identical as described in
this document
Transformation factor, F2 =1000 (transformation from µg to mg),
𝑭𝟏
so = 𝟎. 𝟓𝟖𝟖𝟐𝟒
𝑭𝟐
𝒍𝒂𝒃𝒆𝒍 𝒄𝒍𝒂𝒊𝒎 𝒐𝒇 𝒄𝒍𝒐𝒓𝒔𝒖𝒍𝒐𝒏 (𝒎𝒈/𝒎𝒍) 𝒐𝒃𝒕𝒂𝒊𝒏𝒆𝒅
% Assay of Clorsulon = 𝒙𝟏𝟎
𝒍𝒂𝒃𝒆𝒍 𝒄𝒍𝒂𝒊𝒎 𝒐𝒇 𝒄𝒍𝒐𝒓𝒔𝒖𝒍𝒐𝒏 (𝒎𝒈/𝒎𝒍) 𝒕𝒂𝒓𝒈𝒆𝒕𝒆𝒅
1. System suitability
System suitability is used to verify that the system is adequate for the analysis to
be performed.
6. Linearity
The linearity of an analytical procedure is its ability (within a given range) to
obtain test results that are directly proportional to the concentration of analyte in
the sample.
Conclusion:
The acceptance criteria were fulfilled
2. Specificity
Conclusion:
Clorsulon assay method for IVERMECTIN-CLORSULON INJ, solution for
injection, is said to specific as per assessment of the specificity study.
3. Accuracy
Table No 06: Preparation of accuracy levels as per following table
level no %level prep. First dilution Second dilution
no. Qty of Wt. Of Diluted Pipetted Pipetted
placebo API to
(mL) (mg) (mL)
LEVEL 1 1,7 85 50 1 20
1 50 2 1,7 85 50 1 20
3 1,7 85 50 1 20
LEVEL 1 1,7 170 50 1 20
2 100 2 1,7 170 50 1 20
3 1,7 170 50 1 20
LEVEL 1 1,7 255 50 1 20
3 100 2 1,7 255 50 1 20
3 1,7 255 50 1 20
Tab
le No. 07: % Recovery at difference levels for CLORSULON
Level % Level wrt. Concentration Concentration %
No. working added in μg per recovered in μg Recovery
conc. mL per mL
85 85.20 100.23
1 50 85 85.23 100.27
85 85.17 100.20
170 166.56 97.98
2 100 170 166.79 98.11
170 166.50 97.94
255 247.76 97.16
3 150 255 247.95 97.23
255 248.45 97.43
Mean % Recovery 98,51
Minimum % Recovery 97,23
Maximum % Recovery 100,27
StDev 1,34
RSD 1,3575
Concentrations recovered (μg per mL) were calculated using seven points standard
calibration curve (peaks areas versus concentrations at this levels).
The results and the equation of this calibration curve it is attached to this document.
Using the calibration curve equation, the formula by which were calculated
concentration recovered is:
Conclusion:
Assay method for CLORSULON is found accurate as recovery results and
individual recoveries are well within the acceptance criteria.
4. Precision -Repeatability
Table No. 09: Repeatability test results obtained for sample solution of IVERMECTIN-
CLORSULON INJ, solution for injection.
IVERMECTIN- CLORSULON
CLORSULON Area Retention
INJ time
sample no.
1 353506251 11.987
2 354090278 11.983
3 354005999 11.99
4 353417974 11.98
5 354683895 11.978
6 353657977 11.977
AVERAGE 353893729 11.9825
SD 470671.7319 0.0052
RSD 0.1330 0.0431
Conclusion:
Assay method for CLORSULON in IVERMECTIN-CLORSULON INJ , solution
for injection has precision due to RSD < 2% obtained in sample solution
repeatability study.
5. Intermediate precision
The results for this intermediate precision were calculated with standard solution
of clorsulon 170 µg per mL prepared for each analyst/each day of assay.
Table No. 10: Results of Intermediate precision study for clorsulon for analyst 1
ANALYST 1 Area Label Claim of % ASSAY OF
Clorsulon CLORSULON CLORSULON
sample 1 350186351 99.2946 99.29
sample 2 353009621 100.0952 100.10
sample 3 353426966 100.2135 100.21
sample 4 353569546 100.2539 100.25
sample 5 352667158 99.9981 100.00
sample 6 353034335 100.1022 100.10
Table No. 11 Results of Intermediate precision study for clorsulon for analyst 2
ANALYST 2 Area Label Claim of % ASSAY OF
Clorsulon CLORSULON CLORSULON
sample 1 340197770 100.1162 100.12
sample 2 338196597 99.5273 99.53
sample 3 338779171 99.6988 99.70
sample 4 339480415 99.9051 99.91
sample 5 338554322 99.6326 99.63
sample 6 338630230 99.6549 99.65
Table No. 13: Results of Intermediate precision study for clorsulon / day 1
DAY 1 Area Label Claim of % ASSAY OF
Clorsulon CLORSULON CLORSULON
sample 1 349978714 99.4334 99.43
sample 2 353003611 100.2928 100.29
sample 3 352917255 100.2683 100.27
sample 4 353829334 100.5274 100.53
sample 5 352525112 100.1569 100.16
sample 6 351923466 99.9859 99.99
Table No. 14: Results of Intermediate precision study for clorsulon /day 2
DAY 1 Area Label Claim of % ASSAY OF
Clorsulon CLORSULON CLORSULON
sample 1 349521806 100.1650 100.16
sample 2 348697530 99.9288 99.93
sample 3 348058796 99.7457 99.75
sample 4 348224125 99.7931 99.79
sample 5 348144315 99.7702 99.77
sample 6 347844110 99.6842 99.68
6. Linearity
Linearity was carried out with 5 concentration levels within a range of 50 to
150% of the target concentration of analyte (170 µg/ml) by using test sample.
Linearity was prepared from the stock solution of working standard; concentration of
the solutions are 86,128,170,212 and 254µg/ml.
Evaluation:
Table No. 16: Linearity study data for clorsulon
No of Linearity % Level w.r.t. Concentration Area
levels Working conc. µg/ml Clorsulon
1 50% 86 182898506
2 75% 128 266426383
3 100% 170 350065789
4 125% 212 436620568
5 150% 254 528039316
510000000
460000000
410000000
AREA
360000000
310000000
260000000
210000000
160000000
70 90 110 130 150 170 190 210 230 250 270
CONCENTRATION µg/ml
The correlation coefficient of the regression line is 0.9996, greater than 0.999.
The y-intercept is 1,3 % of target concentration response, less than 2%.
The acceptance criteria were fulfilled.
Conclusion:
Clorsulon method assay for IVERMECTIN-CLORSULON INJ is found linear within
the specified range.
7. Range
The data obtained during the linearity and accuracy studies will be used to assess the
range of the method.
434000000
381000000
328000000
275000000
222000000
169000000
116000000
63000000
10000000
0 44 88 132 176 220 264 308 352
CONCENTRATION LEVELS
8. Robustness
a. Evaluation of results without variation of test parameters
For this test the results from precision- repeatability data are used. The results for
this set of data was calculated with standard solution of CLORSULON -170 µg/ml;
Peak area of clorsulon= 352673937, and they are presented in table no.20.
Table No. 20: Results for Precision -repeatability data for robustness
IVERMECTIN- AREA Label Claim of % ASSAY OF
CLORSULON CLORSULON CLORSULON CLORSULON
INJ sample no.
1 353506251 100.236 100.24
2 354090278 100.402 100.40
3 354005999 100.378 100.38
4 353417974 100.211 100.21
5 354683895 100.570 100.57
6 353657977 100.279 100.28
MEAN 353893729 100.346 100.35
Robustness 1
Table No. 22 Results of Robustness at the flow rate of 1.1 ml /min for sample solution
IVERMECTIN- AREA Label Claim of % ASSAY OF
CLORSULON CLORSULON CLORSULON CLORSULON
INJ sample no.
1 331023704 99.236 99.24
2 334963121 100.417 100.42
3 332982315 99.823 99.82
MEAN % assay value for robustness 99.83
MEAN % Assay value for Precision Study 100.35
% Absolute difference 0.525
Robustness 2
Table No. 23 Results of Robustness at the column oven temperature 35°C sample
solution.
IVERMECTIN- AREA Label Claim of % ASSAY OF
CLORSULON CLORSULON CLORSULON CLORSULON
INJ sample no.
1 335560427 99.316 99.32
2 337238268 99.813 99.81
3 338630230 100.225 100.23
MEAN % assay value for robustness 99.78
MEAN % Assay value for Precision Study (repeatability) 100.35
% Absolute difference 0.565
Conclusion:
Robustness studies signified that the results of the method remained unaffected by
small, deliberate changes in the flow rate and column temperature.
9. Solution stability
Solution stability study for standard solution has been conducted for 72 hours by
keeping the solution on bench top at normal illuminated laboratory condition stored in
tightly closed low actinic glassware (as prescribed in USP -Clorsulon monograph 29),
Sample solution stability at bench top condition (at normal illuminated laboratory
condition). This study has done for 72 hours for standard and sample solutions
Standard and test solution stability summary:
Solution stability study for standard and test solution has been established by
calculating %relative difference of the standard area (clorsulon) and test solution area at
each time interval with respect to initial solutions area.
Evaluation:
Table No. 24: Solution stability study of standard solution at bench top condition for
CLORSULON (i.e. at normal illuminated laboratory condition) as given below:
Solution stability for Standard solution
Status for Area Recovery Relative %
standard Clorsulon difference
solution
initial 353567028 100.00 -
6 hours 354138567 100.16 0.16
12 hours 355890151 100.66 0.66
24 hours 358631808 101.43 1.43
36 hours 357871579 101.22 1.22
48 hours 357351823 101.07 1.07
72 hours 358070283 101.27 1.27
Table No. 25: Solution stability study of sample solution at bench top condition for
CLORSULON (i.e. at normal illuminated laboratory condition) as given below:
Conclusion:
-Standard solution is stable up-to 72 hours at normal illuminated laboratory condition
on bench top stored in tightly closed low actinic glassware.
-Test solution is stable up-to 24 hours at normal illuminated laboratory condition on
bench top stored in tightly closed low actinic glassware.
VII. CONCLUSION
The approach described in this document was used to develop and validate a
liquid chromatographic analytical method that can be used for determination of
Clorsulon in a pharmaceutical dosage form (injection).
The proposed method was validated by testing its linearity, accuracy, precision,
robustness and stability of solutions in accordance with ICH Q2A and ICH Q2B
guidelines.
The results of the analysis of solution for injection by the proposed method are
highly reproducible, reliable, and are in good agreement with the label claims of the
drug.
The excipients that are present in the pharmaceutical formulation of the assayed
samples did not interfere with Clorsulon.
It may be said that the proposed method is precise, sensitive, and accurate, so that
these can be used as standard method for determination of Clorsulon in IVERMECTIN-
CLORSULON INJ solution for injection, using the HPLC systems with PDA detector.
VIII. DEFINITION AND ABREVIATIONS
1. USP - United States Pharmacopoeia
2. NLS -not less then
3. NMT -not more then
4. Std sol -standard solution
5. CLO -clorsulon
6. R.T. -retention time
7. A -peak area
8. wt -weight
9. API -active pharmaceutical ingredient
10. Qty -quantity
11. w.r.t. -with respect to
12. Placebo -dosage form that contains all other ingredients except the
active ones
13. SD -standard deviation, is given by
1
SD= σ =√ ∑𝑁
i=1(𝑥𝑖 − 𝑥)
2
𝑁
𝑆𝐷
14. %RSD -Relative Standard Deviation= × 100 ,is the standard
𝑥
deviation as a fraction of the mean multiplied by 100.
15. Calibration curve - relationship between the signal and the concentration of
analyte in multiple point standardization, when a
. calibration curve is a straight-line, we represent it using
. the following mathematical equation: y= a + bx, where y
. is the signal (Astd) and x is the analyte’s concentration
(Cstd) .The constants a and b are, respectively, the
. calibration curve’s expected y-intercept and its expected
. slope.
2
16.Correlation coefficient (R ) - is a statistical measure that calculates the strength of
the relationship between variables (signal-concentration)
The formulas return a value between -1 and 1, where:
• 1 indicates a strong positive relationship.
• -1 indicates a strong negative relationship.
• A result of zero indicates no relationship at all.
𝑐ℎ𝑎𝑛𝑔𝑒 𝑖𝑛 𝑦 𝑦2−𝑦1
17. Slope = =
𝑐𝑎𝑛𝑔𝑒 𝑖𝑛 𝑥 𝑥2−𝑥1
18. Intercept -is the point where the function crosses the y-axis.
19. Spiking - The addition of a small known amounts of a known
compound to a standard, sample, or placebo, typically for
the purpose of confirming the performance of an analytical
method.
20. Excipient - All other intentionally added components of a medicinal
veterinary product except the active(s) constituent(s)
21. Acceptance criteria - Numerical limits, ranges, or other suitable measures for
acceptance of the results of analytical procedures.
22. Validation - the procedures involved in checking data or programs for
correctness, compliance with standards and conformance with
the requirement specifications.