2. Mount YL6500GC on the support, but you need to consider the transfer line must
be away from YL6500GC left wall.
1
YL6900 GC/MS
4. Rotate four rubber legs under the support to adjust the height between Transfer
line and the hole, and then move carefully MSD to left side of GC as close as
possible so that Transfer line is inserted in the ceramic chamber.
2) PC Communication setting
1. Connect an enclosed LAN to rear side of MS and LAN Hub.
2. Connect RS 15 pin cable to GC remote In/Out port and MS.
3. Connect two Power cables to;
A. Main power(220V)
B. Rotary pump
4. Set IP address for PC as below procedure.
2
YL6900 GC/MS
3
YL6900 GC/MS
5. Oil level sight glass: Fill oil around 70~80% level as indicated.
5) Column installation
1. Inlet side: adjust 2cm from bottom of nut and end of column as below then
tighten the nut to capillary injector.
4
YL6900 GC/MS
ii. After inserting Column in Column measuring tool, adjust column length
around 3~4cm from the outlet of tool.
iii. Cut and align the end of column to measuring point as below picture
and remove the tool.
iv. To prevent from contamination, perform clean procedure (A) once again
and then connect column to Mass transfer line.
5
YL6900 GC/MS
1) Power On
Turn GC power ON (IP is same as general setting.)
2) GC analysis parameter
1. Oven temperature: 40 ℃
5. Remote time setting (200 min -> 3000 min) as below access.
System -> Special -> Press Next button -> Remote Access
*Under GC/MSD mode, column flow rate is displayed around 0.9mL/min at oven
temperature 40℃ with 15m, 0.25mm column, He carrier gas. It is for pressure
*Splitless mode
Set Auto Readyrun to off from System -> Special -> Run start. When GC is Ready
status, press Ready Run button and then wait until Ready LED is flickering. After
end of run, split flow rate is not reduced to 0mL/min and flows according to its split
ratio setting.
3) MSD Setting
1. Turn on Yellow switch at rear side of MSD. (Red switch is Service mode purpose.
6
YL6900 GC/MS
7
YL6900 GC/MS
7. Check Turbo pump rate is higher than 99.9%, Fore pressure value is between
50~70mTorr and Ion source, Transfer line temperature value.
8
YL6900 GC/MS
9
YL6900 GC/MS
6) Specified range spectrum: Check background noise level to see MSD status.
10
YL6900 GC/MS
7) Air/Water Spectrum
11
YL6900 GC/MS
* Ensure that mass spectrum of calibration gas contains all its typical peaks: 69, 131, 219,
264, 414 and 502 m/z. The highest peak may be 69 or 219 depending on selected settings.
You need to check above m/z accuracy and W(resolution) value. W is 0.5~0.7 and m/z is
within ±0.1.
12
YL6900 GC/MS
9) Automatic tune
Once 6 hours elapsed after system activation and operating, you can run automatic tune.
Automatic tune is necessary when you install MSD initially or do not use for few days or
Calibration gas spectrum is out of range. (m/z: ±0.1, W: 0.5~0.7)
13
YL6900 GC/MS
14
YL6900 GC/MS
4. When Ion source and Transfer line temperature reaches to lower than 125℃and Turbo
pump rate is 0 %, close YL Tune software and then turn off Yellow switch at rear side of
MSD.
5. Turn off carrier gas and remove MSD front cover. After that, open Vent valve to counter
clock wise. At this time, you need to be aware that near to Vent valve must be clean
because atmospheric air is introduced into MSD.
15
YL6900 GC/MS
YL Clarity Software
1) Instrument setting
1. Install YL Clarity installation file (version 6.1.0.130 or higher), and MSD patch file.
For the patch file, run ‘!regasm.bat’ file in MSD patch folder.
2. Insert a dongle key and enter the user code during installation.
16
YL6900 GC/MS
17
YL6900 GC/MS
7. Delete not used GC detectors and set Method change option to’ Do Not Send
18
YL6900 GC/MS
2) Instrument Control
1. Click on Login button and make and enter name of project.
a: Device Monitor
19
YL6900 GC/MS
20
YL6900 GC/MS
3> Start time: Set ‘Start time’ after elution of solvent peak.
6. When all parameter reaches to its setting point, press Start button of GC after
injection.
3) Data Processing
1. Integration
1> Open a data file by this icon , and choose the desired file.
2> The data file created by MSD displays two signals, one is Scan and another
one is SIM. You must select either Scan or SIM which you used the method.
21
YL6900 GC/MS
3> Set ‘Result table signal’ on ‘MS method’ window to Scan or SIM used by the
method.
22
YL6900 GC/MS
2. Qulification
-There are three types of NIST library search options, Single compound search,
Automatic compound search and Target compound search for Qulification procedure
udner Scan mode. On Chromato gram window, you are able to see three icons
on the top.
Cllick on icon and place a vertical line to apex of desired peak. NIST library
window shows the result of compound, probability, specturm and match factor.
23
YL6900 GC/MS
(3) Exclude the compounds lower than Min Match factor value.
(4) Set NIST Library source file. Normally, ‘mainlib’ file is necessary for single MS.
(5) After search procedure is done, the highest probability compound is displayed
based on Retention time of peak. Select the compound and press ‘Add all
Selected to Method’ button to copy the compound information to MS method
window.
24
YL6900 GC/MS
(2) Enter desired compound name in Search box and then press Ok button.
(3) Min. Match Factor is used for searching compound by match factor.
25
YL6900 GC/MS
(4) Choose desired RT with Match factor and copy the information to MS
method window by clicking on ‘Add Selected to Method’ button.
3. Calibration
(1) Open a SIM data file and select ‘Add compound w/o Library spectrum’ MS
menu on Chromatogram window.
(2) Place a vertical line to apex of peak and enter name of compound.
(3) Check compound name and retention time are copied to MS method and
change Spectrum to ‘SIM spectrum’, Quantify on option to ‘SIM’.
26
YL6900 GC/MS
(4) Add the other compounds as above procedure and then click on ‘Create MS
calibration’ button.
(6) Open the Standard file which you add compound information to MS method
for calibration curve from the menu. At this time, Response value is 0.
27
YL6900 GC/MS
(7) Click on Responses from Quant. Signals icon and response values are
copied to the table, and enter Amount value of each compound.
28
YL6900 GC/MS
-In order to create a calibration curve for Scan file, you need to enter name of
STD compounds and retention time by NIST library search or ‘Add compound
w/o Library spectrum’ function into MS method as same as SIM calibration curve.
When compound is entered by NIST Library search, ID and Library source file
information is shown in MS method table as below picture.
After enter name of compound and retention time, make a calibration curve as
same as SIM file procedure.
4. Quantitation
29
YL6900 GC/MS
2> Qualify name of compound and retention time into MS method window by NIST
library search option (Scan) or Add compound w/o Library spectrum (Scan or
SIM).
3> Move to result table and click on set button to apply a calibration curve.
30
YL6900 GC/MS
4> After applying a calibration curve, result table will show amount of each
compound.
31