GPT100

Gas Detector Operation Manual


Precautions: 6 Troubleshooting and solution
To avoid personal safety injury, instrument damage and potential dangerous accident,
please read this manual before you use the device. Phenomenon Possible Reasons Solutions

Low voltage Charge in time


1. Brief introduction
GPT100 portable gas detector (hereinafter called Unable to turn on Contact the distributor or
Detector down
detector) is a new type of detector for detecting gas the detector manufacturer for repair.
leakage from the gas pipelines, gas valve, and other Contact the distributor or
Circuit malfunction
possible places. With imported high quality manufacturer for repair.
semi-conductor sensor and embedded Contact the distributor or
Inaccurate
microcontroller, it has very high sensitivity, excellent Disabled sensor manufacturer to replace
indication
man-machine operation interface and various the sensor
adaptive ability.
7. Notice:

The housing of the detector is made of high strength 7.1 Prevent the detector from falling down from high places or strenuous vibration.

engineering plastics and compound anti-slip rubber. 7.2 When there is interferential high-concentration

It is waterproof, dustproof, and with good handle. gas, the detector may not work normally.
7.3 Please operate and handle in strict accordance
2. Features and Specifications
with the instruction, otherwise the result may be
2.1 Features incorrect or you may destroy the detector.
MCU control, low consumption 7.4 The detector should not be stored or used under the circumstance with caustic gas
High resolution STN LCD (such as CL2), or be used or stored under other rigorous circumstances (including
Adjustable sensitivity excessive high and low temperature, high humidity, electromagnetic field and strong
Low battery alert , Intrinsically safe design sunlight).
2.2 Specifications 7.5 After long-term use, if there is dust on the surface of the detector, please clean it
Detecting Range: 0-10,000ppm (based on Methane) lightly with clean soft cloth, instead of caustic impregnant or hard things. Otherwise, the
Target gas: combustible gas body of the detector may be destroyed.
Response time: <5s 7.5 To ensure the detecting accuracy, the detector should be calibrated termly, and the
Alarm: Digital indication, audible and visual alarm calibration period should not be more than one year.
Working condition: Tem:-20~60℃ Hum: <95%RH 7.7 Please send the abandoned Li batteries from the detectors to the appointed places
Power: Rechargeable Li battery DC3.6V 2100mAh or our company. Don’t throw them into the dustbin at random.
Charging time: 4h~6h 7.8 Any malfunction not included in this manual, please contact us for solutions.
4.4 Battery charging Working time: 12h
Life of gas sensor: 2 years
Charge the battery when there is ico displayed in the LCD. When the detector is
Protection level: IP54
deactivated, please connect the plug of the charger to the charging socket of the
Weight: about 300g(including the battery)
detector, then connect the AC connector plug of the charger with the 110V~240V AC
Dimensions: L*W*H 170mm×77mm×33mm
power. The detector wil be activated automatically and indicates that it is being charged.

3. Configuration and functions


When there is ico, the charging is completed. Please remove the charger and the
3.1 Appearance
detector can work normally.
Attention: In order to avoid fire or explosion, please do not charge the detector in
the working area because when plugging or unplugging the charger may create
spark; avoid charging the detector when it is in working condition, otherwise the
charge speed would be affected. Flexible
1 5 Buzzer hole
gooseneck
Slot for
2 LCD screen 6 detecting rod
5. Calibration Charging
3 7 Gas sensor
socket
Press both and into 180 seconds warm up time when the detector is closed,
when warm up finished, the detector will enter into calibration interface, you could see a 4 Operation buttons

display interface as below:

0 PPM 196
1000 PPM
10000 PPM
3.2 Display Elements

EXIT SAVE OK External power or charging


Very low voltage, please charge it
finished
Firstly, please make zero calibration in clean air, you could see dynamic AD value of the Battery is used up, will
Full voltage
sensor on screen, when the value is stabled, press “OK”, then this calibration is finished; automatically turn off in 30s

press to choose calibration point, input standard gas into the detector according
Battery partially consumed Sensor fault alert
to concentration showed on the screen, when the sensor is stabled, press “OK”. Please
press “SAVE” to make the save when all calibration are finished, otherwise the Low voltage and it can last
calibration is invalid. (Before leaving factory, all detectors need to be calibrated very Sound
about 5 hours
well, and the manufacturer make calibration before detector leaves factory)
Low voltage Muting
3.3 Operation buttons MAX:The maximum danger level of this detection.
TIMES:Full scale range(ppm). At 10 times, the full range is 1000ppm, and at 100 times,
Button Function
the full range is 10000ppm.

z To active the detector,press and hold the key for 3s. Black column:Indicates the proportion of full scale. For example, if TIMES is 10 and
S.V is 15, it means the gas concentration is 15 * 10 = 150ppm. And if TIMES is 100, S.V
z When the device is deactivated, press the key and then press
is 15, it means the gas concentration is 15 * 100 = 1500ppm.
the key for more than 2s to enter calibration status.
z Function key,please refer to the indication on the top of the screen. S.V MAX TIMES
z To switch the detecting range between 1,000ppm (10times) and 56 3 10
10,000ppm (100times), please press key
z To switch the detecting range between 1,000ppm (10times) and

10,000ppm (100times), please press key

z Dangerous level returns to zero


Pressing and can adjust the sensitivity(detecting range. The detector can

z Zeroing achieve the best working condition by adjusting this key, which makes the detection
faster and more accurate.
z Function key,please refer to the indication on the top of the screen.
Specific operating method:
z Switch of indicating sound
1. Ensure the detector stay in clean air for at least 20s befor detecting. Now the

4. Operating instruction detector has the highest sensitivity and ability of detecting the minimum gas leakage.
2. Put the sensor of the detector near the position with possible gas leakage and
4.1 Activating the detector
moving it slowly, detecting the gas leakage by checking the indication column. The
To activate the detector, press for 3s. longer is the black column, the higher is the gas leakage. Check the S.V. value to

4.2 Deactivating the detector confirm the changing value. Similarly, the higher is the value, the higher is the gas
leakage. If the gas leakage is very big and the black column is full, then you may use
Press ,and then press before the count down completed to deactivate
the detector. , to adjust the detecting range.

4.3 Detecting 3. MAX indicates the danger level of the gas leakage at the moment. Normally 1-2
means small leakage, the concentration of the gas leakage around the sensor is about
Gas leakage detecting: As shown in the picture below: the first line on the top of the
1500ppm; 3-5 means higher gas leakage concentration, about 1500-5000ppm; 6-9
screen indicates the detection and the pre-set value.
means serious gas leakage, related solutions need to be adopted immediately.
S.V: Show value