Operation / Service Manual

Ear Washing & Automatic Drainage System


Model : EWS-300 & ADS-300 _2016.07.Rev.3

Manufactured by
MEDSTAR CO., LTD.
35-12, Goraedeul-gil, Uiwang-si, Gyeonggi-do,Korea, ZIP Code : 16071
Tel : 82-31-457-1111 Fax : 82-31-457-2670 Since 1997
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Home Page : http://www.medstar.co.kr
E - mail : president@medstar.co.kr, trade@medstar.co.kr

Ear Washing & Automatic Drainage System_2016.07.Rev.3 1


► Contents
1. Brief.................................................................3
2. Specification....................................................4
3. Structural Lay-Out & Components...................4~8
3-1. Structural Lay-Out 3-3. Ear Washing Main Body
3-2. Components 3-4. Automatic Drainage Body
4. Installation.......................................................9~11
4-1. Preparation for Installation
4-2. Installation
5. Operation.......................................................12
5-1. Checking before Operation
5-2. Operation Sequence
6. Legend Indicator….………..……………………12~14
7. Troubleshooting………………………………….14~28
7-1. Troubleshooting on which Parts didn’t work for injecting warm water
from the hand-piece
7-2. Troubleshooting for the Overflow of the Automatic Drainage System
7-3. Troubleshooting on No Water Flow at the Ear Washing System
7-4. Guideline on Controlling the Water Pressure at the EWS-300
7-5. Trouble-Shooting on the Sensor in the Hand-Piece of Ear Irrigation
7-6.Troubleshooting on the Hand-Piece Shaft Replacement
7-7. Electrical Block Diagram on the EWS+ADS
7-8. Heater Module Diagram

8. Product Information Sticker….…………………28


9. Warranty Certificate……………………..….......29

Ear Washing & Automatic Drainage System_2016.07.Rev.3 2


1. Brief
Thanks for your purchasing this product.
We guarantee you that this product is passed on the internal inspection and test
requirements, exactly without any defect in the workmanship for operation.
The period of warranty and guarantee time is one year from the date of operation.
However, even within this period of time, if it doesn’t work properly caused by force
majeure, unilateral modification without manufacturer’s approval, misuse, in particular,
careless usage, unlikely the guideline on this manual, it may be beyond warranty and
guarantee.
Out of the warranty / guarantee period of time, end-users shall pay for all the costs of the
repair, spare parts, and transportation.

1-1. Precaution Before Use


Read the following instruction before use.
Warning ! : This product is only used for the medical purpose by the doctor or authorized
expert in the medical field.
The failure to read the instruction carefully and misuse might cause to hurt the
human body and damage the product.
This product operates by the electronic and electrical control, under which the
authorized technicians are only admitted to repair and shoot the troubles, if it
happens.
This product might be interfered with the surrounding medical / electrical
devices, which normally generate their own electromagnetic fields.
Peruse how to install, operate and shoot the trouble on the product.

1-2. Summary of Ear Washing & Automatic Drainage System


The ear washing system is functioning to inject the water at 37°Celsius, body
temperature, into the ear canal and soften, take out ear wax, sometimes, it could be used
for nystagmus stimulation.
In the meantime, its collective water into the funnel type collector reversely flows to the
container at 4 liter, at the allowable max. high level, making the sensor recognize and
transmit its sensing to main PCB, which makes motor start and open the container, thus
automatically drained out to connection of the waste water pipeline.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 3


2. Specification
Discipline Specification

1. Ear Washing System


Power Input AC 220 V, 50~60 Hz
Water Temperature 37°C
Water Flow Rate Max. 450ml / Min.
Water Supply Hose G 3/4’’
Water Drainage G 1/4’’
Dimension 250x130x432mm
Weight 5Kg
2. Automatic Drainage System
Power Input AC 220 V, 50~60 Hz
Water Tank Capacity 4 Liter
Dimension 240x400x400mm
Weight 13Kg

3. Structural Lay-Out & Components


3-1. Structural Lay-Out

Ear Washing & Automatic Drainage System_2016.07.Rev.3 4


3-2. Components

<Ear Washing Body w/ Hand Piece Hanger> <Automatic Drainage Main Body>

<Ear Washing Hand Piece> <Water Basin w/ Hose>

<Ear Washing Drainage Hose> <Filter>

< Power Cable>


<Water Pipe Connection Hose to Hand Piece>

<Electrical Cable between Ear Washing & Auto. Drainage>

Ear Washing & Automatic Drainage System_2016.07.Rev.3 5


3-3. Ear Washing Main Body

<Front >

No Description Function

① Main Power Turn on / off the power.

② Hand Piece Sensor Put the hand piece extremely into touching the sensor inside.

Press the start button, before lifting up the hand piece.


③ Start / Pause Button
Press the pause button, to stop using the ear washing system.
Legend ( Temperature,
④ Display the current status on ear washing system.
- -,etc )Indicator

⑤ Hand Piece Hanger Hook up the hand piece around the hanger.

Collect the water from the ear canal and flow into the waste
⑥ Water Basin
container to be drained automatically.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 6


7

< Rear >

No Description Function

Some warm water drain out by reaching 37°C, while


⑦ Water-Out Adaptor
most water is circulating to warm up to 37°C.

Supply the purified water through filter from the raw


⑧ Water- In Adaptor
water pipe line.

⑨ Electrical Power Socket Supply the electrical power to the socket.

Connect the electrical cable to the outlet of the


⑩ Outlet automatic drainage system, to control the start-up /
stoppage of the motor for automatic drainage.
Displaying “HH” on the legend indicator, meaning low
Button forcing Solenoid Valve open to pressure of water, press this button to open the

flow water solenoid valve forcefully so as to flow water into the
inside line, regardless of water pressure.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 7


3-4. Automatic Drainage Body

2 1

MODEL : ADS-300
PRODUCT NAME : Automatic Drainage System
Voltage & Frequency : AC 220 Vac, 50/60Hz
Power Consumption : Max 39 W Weight : 11 kg
Country of Origin : South Korea

: S/N :

MEDSTAR.CO.,LTD
35-12, Goraedeul-gil, Uiwang-si, Gyeonggi-do, Korea
Tel : +82-31-457-1111 Fax : +82-31-457-2670

6
5

No Description Function

Collective water from the basin are flowing


① Rinse- In to the waste water tank for automatic
drainage.
② Making sensor start-up motor / open the
Drain Out
valve, waste water Drains out its pipe line.
Connect the electrical cable to the outlet of

③ the ear washing system(⑩) in rear, to


Outlet
control the start-up / stoppage of the motor
for automatic drainage.
④ Make waste water easily flow into its tank
Air Ventilation Hose
and filled.
Some natural drain out water line, while in
⑤ not using the hand- piece, connects to this
Drainage In Hose
drain in hose and it flows to the waste
water container.
⑥ Taking out the plug, waste water in the
Emergent Waste Water Plug Out
container forcefully drains out.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 8


4. Installation
4-1. Preparation for Installation
1) Be sure to make an enough space to install this equipment.
2) Ear washing system and automatic drainage system are connected to the
socket through its electrical cable, more than 1.5 meter.
The socket meet the requirements of IEC 309, at 220 volt, 15A and it has the
The electrical circuit breaker and on / off switch.
3) Its system is operated and stored in the following environmental condition.
4) The post pole and its bracket with the arm are required to install the built-out ear
washing system and automatic drainage system.

Description Conditions
Operating Temperature 10~40℃
Operating Humidity 10~90%
Storage Temperature -20~70℃
Storage Humidity 10~90%

4-2. Installation
1) Check all of components are put into the box.
2) Prepare the hexagonal wrench to unfasten the arm.
3) Using the hexagonal wrench, fasten the arm as shown in the photo, below.

4) Put the plastic bushing on the hole of the arm, as shown in the photo, below.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 9


5) The male round bar on the bottom of the ear washing body is inserted into the female
hole of the arm, carefully, as shown below.

6) Fasten the nut, putting the bushing, for fixing the ear washing body, as shown below.

7) Put the hand piece into the holder as shown below.

8) Connect the electrical power on the rear of the ear washing body to the outlet of the
automatic drainage system, in addition, insert the collective water pipe line from the
basin into the socket of “rinse-in”.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 10


9) Insert the automatic drain water pipe into the hole of “drain out”, below, and taking out it
from the hole, first push the outer adaptor evenly by one hand’s two fingers and you
can take out its pipe line by another hand, as shown below.

10) Insert the nozzle into the hole until its click sounds in an arrow direction, as shown
below.

11) The water-in, water-out pipe line are inserted into the holes on the rear of the ear
washing system, as shown below.

12) You can take out the fuse holder below the electrical power socket, and replace it by
the new one.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 11


5. Operation
5-1. Checking before Operation
1) Install the product, guiding on the manual.
2) Check whether it’s connected, exactly and turn on main power switch.

5-2. Operation Sequence


1) Lifting up the hand piece, you’re ready for injecting the water, already warming up to at
37°Celsius.
2) Making patient holding up the water basin below the ear of patient and inject the warm
water into the ear canal.
3) After injecting water, it’ll naturally flow out into the water basin and collect into the
container of the automatic drainage system.
4) Going up the Level of the water, the micro control PCB, getting the signal from the
sensor at the allowable max. level automatically orders solenoid valve to open and
drain water out to the waste water pipe.
5) After using the ear washing system, hook up the hand piece and warm up the water up
to 37°C and then it’s changed to the idle mode to save the electricity.

6. Legend Indicator
Legend Indicator Description Troubleshooting

Water pressure is too weak, in this 1. Check the water pressure,


case, water pressure sensor never min. more than 0.1 MPa to
makes check vale open, resulting make the check valve open.
in blockage of water flow. 2. Check the filter, being
clogged, replace it by the new
one.
Buzer sounds for three seconds, 1. Check whether warm water
when the temperature is over 40° is draining out well or not.
degree. 2. Check whether heater is
working properly, otherwise,
it’s be replaced.

Ready for warming up by 37°


degree.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 12


Hand piece doesn’t put into the Put the hand piece into the
bottom position of the hanger, in bottom position deeply.
which sensor is installed.

The signal from the temperature 1. Check temperature sensor.


sensor to main board(PCB) might 2. Check main board.
not communicated well.

Heater is not connected to the Connect the electrical power to


electrical power, at AC 220V. the heater.

The sensor from the “Water-In” is Check the sensor in the


not working properly. “Water-In”, otherwise, replace
it by the new one.

The sensor from the “Water-Out” Check the sensor in the


is not working properly. “Water-Out”,otherwise, replace
it by the new one.

The sensor is short from the 1. Check the integral electrical


“Water-In”. line.
2. Replace the sensor in the
“Water-In” and other
electrical line, depending on
its test result.
The sensor is short from the 1. Check the integral electrical
“Water-Out”. line.
2. Replace the sensor in the
“Water-Out” and other
electrical line, depending on
its test result.

The water flow value doesn’t 1. Check whether water flows


change, which means water or not.
doesn’t flow at all. 2. Check whether water flow
sensor works or not,
otherwise, it’s replaced by the

Ear Washing & Automatic Drainage System_2016.07.Rev.3 13


new one.

The set temperature is different 1. Check whether water is


from the actual temperature. flowing well.
2. Check the relevant sensors,
otherwise, they’re replaced by
the new ones.

7. Troubleshooting
7-1. Troubleshooting on which Parts didn’t work for injecting warm water from the hand-
piece

Ear washing system (hereinafter called EWS) appeared “A1” and changed to “36~39”, sometimes,
“E9:” on the indicator.

However, warm water is not injected into the hand-piece, while warm water drains out.
You’ve connected the hand-piece to the warm water drain adaptor as a temporary expedient.
Therefore, our tentative conclusion is that the cause not to inject warm water into the hand-piece falls
into five cases.

First, possibility to show the voltage below DC 12V, at the ground point, which is able to operate the
relay for opening three way cock solenoid valve.
In this case, the transformer, supposedly functioning to rectify the voltage from AC 220V to 12V, should
be replaced, while you’re kindly requested to notify us of the rated / used voltage in Qatar.

Secondly, possibility not to operate the relay properly for opening three way cock solenoid valve.
In this case, the relay should be replaced by the new one.

Thirdly, possibility not to turn on LED of the “ 3 WAY ON” on the main PCB.
In this case, the main PCB should be replaced by the new one.

Fourthly, possibility not to operate the three way cock solenoid valve, itself.
In this case, three way cock solenoid valve should be replaced by the new one.

Fifthly, possibility to combine any case above with another.


In this case, you’re to precisely examine which parts it didn’t work properly.

So, we’d like you to undertake the following troubleshooting procedure on which parts it didn’t work,
exactly.

1.Mandatorily Prior Preparation for Examination

1-1. Place the current hand-piece, which you connected to the warm water drain adaptor, into the original
position.
1-2. Check whether the temperature to show from 36℃ to 39℃, returns to the normal one at 37℃ on
the indicator, otherwise, you should increase to open water flow rate control knob as much as showing
Max.37℃, inside EWS, as I sent you its location before. The reason why it shows the temperature up to
39℃ is the shortage of water volume / flow rate.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 14


2. Examination Process on which parts didn’t work properly

2-1. Case I : Shortage of Voltage for the Operation of Relay


Check whether it shows the voltage at DC 12V, or above, at the ground point (U8, 7812) or not by voltage
meter, as shown in the attached diagram.
Should it designate 12 voltage, or above, it’s no problem, otherwise, showing it below 12V, it’s the
problem of the transformer, which should be replaced by the appropriate one.
Should the stable 220V input into the transformer, it’ll result in showing the 12V at the U8(7812).
In this regard, please be sure to notify us of the used / rated voltage in your country.

2-2. Case II : Defective Relay


There are two ways to examine that the relay works or not, one is by eyes, the other by voltage meter.
When the indicator changes to show 37℃, you can also take a look at whether the relay works or not by
your eyes.
Also, check whether it shows the voltage w/ 220V at the “CON55”” connector by voltage meter.

Should it show the voltage at 220V, the relay works, properly, otherwise, the relay is defective, which
should be replaced by the new one.

2-3. Case III : Not Turning On the LED of “3WAY ON” on the Main PCB
When the indicator shows “ 37”, the LED of “3WAY ON” on the main PCB should turn on.
Otherwise, the main PCB is not working, properly, which should be replaced by the new one.

2-4. Case IV : Three Way Cock Solenoid Valve Breakage


When showing the “37” on the indicator and the voltage of the relay at the “CON55” shows 220V, and after
checking the housing of three way cock solenoid valve, precisely, with no problem detected, three way
cock solenoid valve might break down.
In this case, it should be replaced by the new one.

2-5. Case V : Combined


When the voltage at the “CON55” doesn’t show 220V, the relay doesn’t work properly.
In this case, you should examine whether the three way cock solenoid valve works properly, or not.
The way to examine it is that after connecting the “GND PATTERN” with “3 WAY VALVE CONTROL PIN”
by the electrical cable, or wire, so-called ‘make a shortage’, the three way cock solenoid valve is supposed
to forcefully open.
Should the three way cock solenoid valve open, it’s working, otherwise, it’s out of order.

7-2. Troubleshooting for the Overflow of the Automatic Drainage System


I. Easy, Quick Test Method 
Dismantling  the  waste  water  container  with  sensors,  highest  /  lowest  seen,  undertake  the 
following procedure : 
‐ Filled with water at the highest level sensor, check whether or not level sensor is activated to 
make the drain pump start up its operation. 
If so, the drain pump is working, the level sensor is working, so check whether the drainage 
pipe and the container outlet were plugged or not. 
  If not, the drain pump is broken, or the level sensor is out of order. 
II. Standard Test Method 
1. In case the Drain Pump is out of order 
At the waiting mode, showing “ ‐ ‐ “ on the LED display of the ear washing system(EWS),   
pressing the button of “start / pause” for 3 seconds, it immediately checks the water level on 
the container of the automatic drainage system(ADS) : 

Ear Washing & Automatic Drainage System_2016.07.Rev.3 15


 
Without placing the lowest level on it, the drain pump is supposed to start up operation, while 
it’s placed at the lowest level, the drain pump is supposed to stop. 
In case the drain pump is operating, check whether the drain pipe / container outlet might be 
plugged, or not. 
* Checking the waste water level by sensor, the drain pump is supposed to start‐up operation at 
the highest level, while it’s supposed to stop at the lowest level. 
In case the drain pump doesn’t work, you need to check the followings : 
          ‐  Check  whether  “drain  on  LED”  is  turning  on  at  the  “CHECK  POINT”  of  the  MCU  board, 
referring to the attached excel sheet. 
            Turning on, MCU board is no problem, vice versa, MCU board is not working. 
          ‐  Check  whether  the  “drain  relay,  RY25”  is  measured  /  well  connected,  by  more  than  12 
voltage,  as  indicated  in  the  “GND  +12V”  of  excel  sheet,  otherwise,  below  11v,  its 
transformer  should  be  changed  from  12  volt  to  15  volt  under  AC  220voltage,  this  is  the 
defect of the MCU board.   
          ‐ Check whether it measures 220 voltage by tester at the “drain pump connector, CON54” : 
            When  measuring  220volt,  and  if  the  drain  pump  doesn’t  work,  the  drain  pump  must  be 
broken. 
  2. In case Level Sensor is out of order 
      Taking a look at the three colored electrical cables (white, black, green, CON52) on the right 
bottom of the MCU board, its voltage is functioned to work how to drain ; 
      Under 5volt at the white electrical cable, using the tester, connect an electrode to the black / 
green colored electricity, while the other electrode is connected to “GND” on the MCU board 
of excel sheet and compare its(black, green) voltage to the following table. 
      In this case, you’re to adjust water level at the highest / lowest by your judgement. 
      When automatic drain is made : 
Black Cable  Green Cable  Drainage  Water Level 
5Volt  0Volt  Off  Lowest 
0Volt  0Volt  Off  Middle 
0Volt  5Volt  On  Highest 
 
      When  forceful  drain  is  made,  for  instance,  pressing  the  button  of  “start/pause”  for  3‐5 
seconds : 
Black Cable  Green Cable  Drainage  Water Level   
5Volt  0Volt  Off  Lowest 
0Volt  0Volt  On  Middle 
0Volt  5Volt  On  Highest 
 
      *  Regardless  of  the  drain  mode(automatic,  forceful),  the  drain  pump  is  always  supposed  to 
stop at the lowest level. 
        Without varying from the voltage along with the water level, level sensor must be broken. 
        Be sure to check whether the “LED On” of the “drain on” on the MCU board is activated or 
not. 
  If so, MCU board is working, otherwise, it’s broken, as mentioned above.     
  
7-3. Troubleshooting on No Water Flow at the Ear Washing System

1. Check whether the faucet / valve of the raw water pipeline is fully open or not, otherwise, the
weak pressurized water below about 2.3 MPa can’t make the pressure sensor inside the ear

Ear Washing & Automatic Drainage System_2016.07.Rev.3 16


washing system open the check valve, which results in never supplying water to the next
water pipe line, to be able to start to warm up, referring to the following photos, we tested in
our lab.

- Check whether raw water faucet valve is fully open. – Water Supply Closed by Valve

- Water is filled into the pipe, showing its


pressure, at 0.3 Mpa. Press. Sensor Purified Water Supply Check Valve
inside Ear Washing System

- Check whether the filter is flogging or not.

2. Check whether every cables are connected into the right sockets / holes, exactly, as follows :

Ear Washing & Automatic Drainage System_2016.07.Rev.3 17


- Power, Water-In, Water-Out, Electrical Cable into - Rinse-In from the Water Basin, Drain-Out, Electrical
Outlet Between E.W.S & A.D.S Cable to the Outlet, in which sensor transmits its signal
to the main PCB, which thus orders motor to start up
, resulting in waste water drainage.

- Connection Line for Hand-Piece - Warm Water-Out Line, Drain Line well Connected

3. Turn on the power, A1 displays on the indicator, which means it’s ready for starting to warm
up water, and taking 2~3 minutes, you can see the temperature at 36~37 ° Celcius,
simultaneously, warm water is flowing out from the warm water pipeline, until the indicator
shows idle for energy saving, as “-, -“.

- It’s ready for starting to warm up water. – Displaying temperature, warm water continues to
flow.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 18


-

- Finish to warm up at 37° Celcius. – Change to Idle Mode, Energy Saving,


Warm water-out stops.

- Warm water continues to flow while in displaying - Press the Start button, lift up the hand-piece and press
the temperature, at 36~37°Celcius. the button of the hand-piece, water injects, warm
water-out continues to flow.

- Water injects, warm water in a scanty amount – Warm water stops flowing, while in showing idle
Continues to flow. signal, ”-, -“.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 19


4. Check whether water flows before hand-piece or not as follows :

- Take out the tip of the hand-piece. – Unfasten the bolt to be connected to the hand
-piece.

- Check whether water is flowing from the tip of


the warm water pipe, connected to the hand-Piece.

7-4. Guideline on Controlling the Water Pressure at the EWS-300

1. As a micro-control of the pressure, you can get it done at the irrigation hand-piece.

a
<Fig. 1>

Ear Washing & Automatic Drainage System_2016.07.Rev.3 20


<Fig. 2>

2. Turning around the nut clockwise as shown in the red-circle of the <Fig. 1>, and it will change to
the one, as shown in the red-circle of the <Fig. 2> and the ear irrigation pressure decreases,
proportionally.
Be sure to have a caution that turning round it counter-clockwise entirely as part to increase its
pressure, the nut can be loose apart.

3. Should the micro-control of its pressure can’t meet you, as a macro-control of it,
please refer to the followings:

< Fig. 3>

Ear Washing & Automatic Drainage System_2016.07.Rev.3 21


4. Taking out the case of the ear irrigation system, you can see the water-inlet pressure control
valve, Inside as shown in the red-circle of the <Fig.3> by which too much water inlet pressure
can Decrease not to be able to impact this system.
5. Turning around the pressure control valve in a directional way of +/ -, it increases / decreases,
in taking a look at the pressure gauge.

6. After controlling the water pressure by its valve as shown in the red-circle in a macro-way,
Turning around the knob counter-clockwise, as shown in the green-circle of the <Fig. 3>,
its pressure can be increased in a micro-way.

7-5. Trouble-Shooting on the Sensor in the Hand-Piece of Ear Irrigation

1. Check whether the line of sensor is cut off or not, allegedly caused by the
damage or the poor soldering, as shown in the pink circle.
2. Unless it's no trouble, replace the sensor by the new one, we’ll send you.
3. When replacing the sensor, be sure to install it in a vertical direction, while it's better to fix it by
the bond of 401, included, with the new sensor, instead of the both sided tape, which might be
also acceptable.
4. Temporarily fixing the sensor at first, check whether the sensor works without any trouble, after
the hand-piece places onto its guided hole, as shown above.
If it works very well, be sure to fix it permanently.
5. A scanty amount of bond applies to the flat part of the sensor.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 22


7-6. Troubleshooting on the Error of Sensor in the Ear Irrigation System
1. As indicated in the <Fig. 1>, below, cutting the electrical cable, connected to the sensor of the
ear irrigation system, and unfasten the four bolts, by a screw driver, as shown in the <Fig. 2>.
2. Take out the sheath of the hand-Piece.

<Fig. 1> <Fig. 2>

3. Unfasten the bolts by a screw driver, which are attached to the main body and side of the ear
irrigation system, as shown in the <Fig. 3>, while keep the bolts safe without losing them.

<Fig.3>

4. After taking out the case, and take out the two pins of the cable by hands, indicated by the red
arrow, below, from the third socket of the right side, in the <Fig. 4-1, 2>.

<Fig. 4-1> <Fig. 4-2>

Ear Washing & Automatic Drainage System_2016.07.Rev.3 23


5. Fasten the new hand-piece sheath, you’ve got from us, by the four bolts, and insert the cable
into the hole, as indicated in the <Fig. 5>.
-

<Fig. 5>

6. After going through the cable, insert the pin into the socket, as indicated in the <Fig.6>.
Not having the polarity, its color is useless.( No need to distinguish the right from the left by its
color.)

Pay Attention to the inserted


location of the pin, like this.

<Fig. 6>

7. After inserting the pin into the socket, above, and insert it into the socket from the third right
side as shown in the <Fig. 7>.

<Fig. 7>

8. Put the cover of the ear irrigation system and fasten the bolts by a screw driver.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 24


7-6.Troubleshooting on the Hand-Piece Shaft Replacement
1. Unfasten the four cross recessed bolts by the screwdriver, as shown in the <Fig.1>.

<Fig. 1>

2. Take out the main part as shown in the <Fig.2>.

<Fig. 2>

3. Unfasten the rear part and take out its spring and cover, as shown in the <Fig.3>.

<Fig. 3>

Ear Washing & Automatic Drainage System_2016.07.Rev.3 25


4. Unfasten the front flow volume control lever, and take out its spring and lever, as shown in the
<Fig.4>
<Fig. 4>

5. Pushing down the protruded shaft, as shown in the <Fig. 5-1> and take out it, as shown in the
<Fig. 5-2>.

<Fig. 5-1> <Fig. 5-2>

6. Replace the disassembled parts by the new ones, enclosed.


Assemble the first, shaft, secondly, rear spring and cover, thirdly, front spring and flow volume
control lever into the hand-piece.

7. Keep in mind that in fastening the four cross recessed bolts of the hand-piece too forcefully, by
screw driver, it might be broken down.

Ear Washing & Automatic Drainage System_2016.07.Rev.3 26


7-7. Electrical Block Diagram on the EWS+ADS

ADS-30 0 EW S-10 0
Aut o Dra in Syst em Ea r Irrig a t on Syst em

1 : 흑색 0.75 ( Water Motor ) 1 : 흑색 ( Water Motor )


2 : 적색 0.75 ( Water Motor ) 2 : 백색 ( Water Motor )
1
7 1 3 : 녹색 AWG22 (Water Sensor On ) 2 7 3 : 회색 (Water Sensor On )
8 4 : 황색 (Water Sensor Gnd )
6
8 2 4 : 백색 AWG22 (Water Sensor Gnd ) 3
6

3
5 : 흑색 AWG22 (Water Sensor Off ) 4 5 5 : 적색 (Water Sensor Off )
5
4 6 : 적색 AWG22 ( Led Dc09V + ) 6 : 청색 ( Led Dc09V + )
7 : 황색 AWG22 ( Led Dc09V - ) 7 : 갈색 ( Led Dc09V - )
8 : 녹황색 ( Gnd ) 8 : 녹색 ( Gnd )

Main Connector
220 0

Trans

Main Swit ch
12 0
M ED-14 4 0

FMD Board




Eg-CrW

Ac2 20 v Ac12 v
Eg-CrH
Eg-Zn

Eg-Cr

Tra nsForme r
BuF

AC22 0 In
청황흑백

Heat .Mod.

Heat

f use

Hall.Sen
Co-1
Co-2
Co-3
Co-4
Co-5

면 가 공

Water Heater
Press.Sen

M icro Sw it ch Heat
Sol.
Water.Sen




Start.Pause

Sol.
Drain
TO-14 36

3wa

Drain

3way
M ED-14 4 3
청황 흑 백
Ea r Irrig a t ion
규정하지않음

100S
25S
6.3S
0.8S

1 2
Pressure Sensor

4 3

Solenoid Valve
Connector

6 . Led Dc09V +
8. Gnd
7. Led Dc09V -

5. Water Sensor Off


4 . Water Sensor Gnd
3. Water Sensor On
2. Water Motor
1. Water Motor

Dc 9 v LED Boa rd
Dc5v
Dc12v

3WAY Solenoid Valve


Dc9v
Dc9v

Connector
-

Stop
WILO PM-015NM

정지
Water Pump

Water Sensor
Wat er Sensor

Start
Blue LED Bar

Blue LED Bar

시작
Water Sensor
Tel: 031-4 57-1111

Stop : 청
Start : 적
Gnd : 백
Fa x: 031-4 57-26 70

Ear Washing & Automatic Drainage System_2016.07.Rev.3 27


7-8. Heater Module Diagram

8. Product Information Sticker


No Classification Description
1 Product Name Ear Washing & Automatic Drainage System
2 Brand Name EWS-300 & ADS-300
MEDSTAR Co., Ltd.
3 Manufacturer 130-1, Ojeon-dong, Uiwang-si, Gyeonggi-do, Korea, 437-819
Tel. : +82-31-457-1111, Fax. : +82-31-457-2670
Inject warm water into the canal of ear and take out
4 Used Field cerumen, and sometimes check the nystagmus
stimulation.
5 Manufacturing No. Indicated after the Completion of Manufacturing
6 Manufacturing Date Indicated after the Completion of Manufacturing
7 Weight EWS-300 : 10kg, ADS-300 : 20Kg
8 Packaging 2 Sets / Carton Box
9 Operation, Caution, Storage Refer to the operation manual.
10 Electricity AC 220V, 50/60 ㎐, 20W
11 Electrical Class & Type Class I, Type B

Ear Washing & Automatic Drainage System_2016.07.Rev.3 28


Warranty Certificate
 MEDSTAR Co., Ltd is to certify that this product is guaranteed and warranted for a
period of one(1) year.
 In case of not working properly, please contact the local agent / representative with its
detailed phenomena presented.
 The function, shape, etc, of the product for improvement might be subject to change
without prior notice.

Ear Washing, Automatic


Product Name Warranty Period of Time
Drainage System
Brand Name EWS-300, ADS-300
One(1) Year
Manufacturing No. Serial No.:
Purchase Date MM / DD / YY Until. MM / DD / YY
Purchase Place
Telephone No. +82-31-457-1111

 Complimentary After-Service Scope


The product, which didn’t work properly at the normal operation, can be repaired and
trouble-shot, free of charge, within one year from the date of purchase.

 Charged After-Service Scope

1. In case of proper work


• When asking for the after-service, at the normal operation, you’ll be charged on the
after-service.
• So, be sure to read the operation manual carefully and get the reliable information from
the seller.

2. In case of Non- Proper Work by Misuse


• Misuse by End-User, or out of Order from the voluntary modification, or Disassembly
• Not working properly by Non-Authorized Personnel.
• Not working properly, using the Unsuitable Electricity.
• Out of Order, caused by droppage in the Movement.
• When it didn’t work, using the non-authentic spare parts made by other company.

3. Others
• It didn’t work by force majeure.
• Consumables Life ended.