SGS MINERALS SERVICES – T3 SGS 1317 03-2016

LABORATORY QUALITY CONTROL (SLIM)

SCOPE QUALITY CONTROL Note: The frequencies specified are the


minimum expected and may not apply
This summary describes the Quality
CRITERIA to all sample types due to sample size,
Control protocols established at SGS sample matrix, and analytical method
Minerals Services geochemistry QUALITY CONTROL INSERTION AND applied (e.g. trade analysis). In the
laboratories to ensure test results are FREQUENCY case of samples received as pulps the
consistently monitored and accepted as fit preparation blank is substituted with an
Our QC material insertion frequency
for purpose. additional method blank for pulps and
is 14% for exploration and ore grade
for solutions with a spike replicate and/
samples which includes sample reduction
or a spike blank in order to retain the QC
LABORATORY blanks and duplicates, method blanks,
frequency.
INFORMATION weighed pulp replicates and reference
materials.
MANAGEMENT (SLIM) CRITERIA FOR ACCEPTANCE AND
Process control and metal accounting
Every Minerals Services laboratory in REJECTION OF BLANKS, REFERENCE
solid samples are typically run in duplicate
the SGS network operates a Laboratory with added insertion of blanks and
MATERIALS, REPLICATES AND
Information Management System called reference materials for a QC frequency DUPLICATES
SLIM. The SLIM system has been of 115%. For process control samples Blanks
designed to be the “backbone” of SGS not performed in duplicate, the frequency
geochem labs and includes our method Sample reduction blanks, method blanks,
insertion rate is ~20% which includes
set up protocols (analytes, ranges, reagent blanks are used to assess (and
sample reduction blanks and duplicates,
internal QC materials that include blanks, correct for when appropriate) responses
weighed pulp replicates and reference
duplicates, replicates, and reference other than those inherent to the blank
materials.
materials) as well as their frequency of material. If failure occurs, that cannot
For process control and metal accounting be accounted for based on set rules for
insertion and tolerance requirements.
solution samples the QC insertion is 14- exemption; a minimum of 25% of the
SLIM is also used to monitor our internal
20% which includes blanks, spike blanks, samples including the failed blank is
processes for instruments, equipment,
spike replicates and solution standards. repeated. Repeated sample failure results
sample tracking, storage and reporting
formats. SLIM uses a secure and SLIM automatically flags whenever a in investigation and repeating the entire
complete audit trail to ensure traceability QC material fails to meet established batch.
and confidentiality. The system will also statistical rules preset in the system.
allow the lab to transfer all data to a web The SLIM QC module is based on the
portal called QLab™ for easy electronic Thompson and Howarth precision curve
access of your data. SLIM is typically which sets tolerance requirements that
compatible with the most frequently used are associated with the detection limit and
databases such as Maxwell / Acquire etc. expected precision of the analyte within
the method. These rules are based on
SLIM is the link to all of our internal
rigorous method validation requirements
quality control criteria outlined below.
established for our methodology.
SGS MINERALS SERVICES – T3 SGS 1317 2

Reference Materials, Replicates and Reports can be provided which allow for QC REPORTS
Duplicates method data and/or customer specific
QC reports can be customer specific
When inputting the acceptable values data review. SLIM has key tools that
reports and as such are tailored to each
and associated tolerances for duplicates, are used to look at these processes as
client’s requirements. They produce
replicates and reference materials described below.
information on bias, precision, sample
consideration is given to the fitness volumes and turn- around times and can
for purpose of the method (what are CONTROL CHARTS AND STATISTICAL be used at the request of a customer or
the customer’s needs) as well as the PROCESS CONTROL (SPC) as part of an investigative or remedial
certification process of reference material process.
These are monitored on a regular and
when compared to the method for
on-going basis; daily, weekly, monthly etc.
which the material will be used. Repeats
although the frequency will depend upon
are performed based on a percentage
the method utilization Shewhart Control
of reference material, duplicate and
chart rules apply that are statistically CONTACT INFORMATION
replicate failures to the total number of
based to flag and monitor warnings,
these materials inserted in a batch and Email us at minerals@sgs.com
actions, trends and bias for individual
range from 25-100%. These are clearly www.sgs.com/mining
reference materials or set up to monitor
specified in our globally controlled internal
all method based reference materials in a
quality procedures.
single frame.

QUALITY NOTES KEY PERFORMANCE INDICATOR REPORTS


(KPIS)
In those cases where results are
outside of the method range or where The weekly KPI reports contain pivot
© SGS Group Management SA – 2016 – All rights reserved - SGS is a registered trademark of SGS Group Management SA

suspected sample homogeneity and/or tables listing z-scores for every internal
matrix problems are encountered, this standard used in the laboratory. These
information is noted in SLIM and reported tables can be sorted by scheme and/
to the customer. or analyte. The warning limit for bias is
a z-score of ± 0.5 and the warning limit
for precision is 1.0 z-score. Sections
QUALITY TREND ANALYSIS falling outside these limits require further
In addition to being accepted or rejected inspection.
on a batch basis the performance of A weekly KPI Report is reviewed at the
quality control samples is assessed on weekly production meeting by all senior
a short term and long term basis. This is laboratory staff. It is particularly useful
critical to the detection and remediation as it gives the user an overview of the
of systematic biases and trends. The laboratory and its sections combined with
performance of reference materials and the ability to drill down to an individual
blanks must be monitored in a short standard level. Control charts can then be
term context (e.g. an instrument batch run on targeted standards to determine
or daily basis) and a long term context. underlying trends.