What  we  are  going  to  study?  
Course  details:  
 1.  Electron  paramagne@c  resonance  
 2.  Super  conduc@ng  Quantum  Interference  Device  (SQUID)  
 3.  Introduc@on  to  Op@cal  methods  
   -­‐  UV-­‐Visible  
   -­‐  Infra  red  
   -­‐  Fluorimetry  
   -­‐  Atomic  absorp@on  
                                   4.  ScaQering  of  Radia@on   -­‐  Principle  
   -­‐  Nephelometry   -­‐  Instrumenta&on    
   -­‐  Turbidimetry   -­‐  Methods  
   -­‐  Raman  Spectroscopy   -­‐  Applica&ons  
 5.  Electro-­‐analy@cal  methods  
   -­‐    Cyclic  voltammetry  
   -­‐  Polarography  
   -­‐  Amperometry  
   -­‐  Conductometry  
 6.  Thermogravimetry  (TGA)  and  Differen@al  thermal  analysis  (DTA)  
 7.  Gas  Chromatography  and  high  performance  liquid  chromatography  (HPLC)    
Text  Books  and  References  

Text  and  Reference  books:  


 
 
1.  Electron  paramagne@c  Resonance:  A  prac@@oner’s  tool  kit  
-­‐  Marina  Brustolon  and  Elio  Giamell  
2.  Electron  Paramagne@c  Resonance  by  John  A  Weil  and  James  R.  Bolton  
 
3.  NMR,  NQR,  EPR  and  Mossbauer  spectroscopy  in  Inorganic  Chemistry  by  
R.  V.  Parish  
Electron  Paramagen&c  Resonance  (EPR)  

Electron  Paramagne@c  Resonance  


Detects  the  unpaired  electron  in  the  
Electron  Spin  Resonance  or    
sample  under  inves@ga@on  and  its  
Electro  Magne@c  Resonance  
electronic  structure  around  it  

What  is  an  EPR  spectrum?      


It  is  usually  a  plot  of  first  deriva@ve  of  absorp@on  of  microwave  frequency  
radia@on  against  the  magne@c  field  intensity.    

Magne@c  field  (G)  


Electron  Paramagen&c  Resonance  (EPR)  

EPR  

Con@nuous  wave  method   Pulsed  method  

Low  intensity  con@nuous  irradia@on   Short  pulses  of  high  power  microwave  
(microwave)  of  sample     radia@ons  are  sent  to  the  sample  and  
the  sample  response  is  recorded  in  the  
absence  of  radia@on.  
What  are  the  main  features  of  an  EPR  spectrum?  
Main  parameters  and  features  of  EPR  spectrum:  

1.  Posi@on  of  absorp@on:  Magne@c  field  value  at  which  the  absorp@on  takes  place  
2.  Number,  Separa@on,  rela@ve  intensity  of  the  lines  
3.  Width  and  Shape  of  the  line  

Structure  of  the  species  under  inves&ga&on        


Their  interac&on  with  environment  
Dynamic  process  associated  with  it.  
Magne&c  moment  of  an  electron  

Defini@ve  spin  value  of  the  vector,  Sz  =  ±  1/2  

These   spin   states   are   differ   in   orienta@on  


of   the   angular   momentum   in   space   but  
Representa@on  of  electron  spin   both   spin   state   has   same   magnitude   of  
angular  momentum  vector     angular  momentum.  
Origin  of  EPR  spectra_electron  spin  in  a  magne&c  field(Zeeman  effect)  

The  existence  of  magne@c  moment  associated  with  the  electron  spin  responsible  
for  the  energy  separa@on  between  the  spin  up  and  spin  down  

In  the  presence  of  magne&c  field  (B):  

When  B  is  parallel  to  z  axis,  then  S  will  take  the  value  of  Sz  
Electron  spin  in  a  magne&c  field  (Zeeman  effect)  

These  two  energy  levels  are  degenerate  in  the  absence  of  magne@c  field    

E
n
Spliing   of   electronic   energy  
e
level   in   the   presence   of  
r
magne@c   field   called   Zeeman  
g
spliing  or  Zeeman  Effect  
y

B  =  0   B  >  0  

In  presence  of  Magne@c  field,  the  degeneracy  of  these  states  will  be    lihed,  based  on  
the  alignment  of  the  moments  with  the  external  magne@c  field,  one  state  will  in  
lower  in  energy  and  the  other  state  will  be  higher  in  energy      

Hence  the  spliing  of  energy  levels  depends  on  strength  of  the  magne@c  field.  
Electron  spin  in  a  magne&c  field  (Zeeman  effect)  

Ms  =  +½  
Ms  
 

±½  
ΔBpp  
Energy  

ΔE=hν=gβB  

Ms  =  -­‐½  
B  =  0   B  >  0  
Magne@c  Field  (B)  
h Planck’s constant 6.626196 x 10-27 erg.sec
ν frequency (GHz or MHz)
g
β
g-factor (approximately 2.0023)
Bohr magneton (9.2741 x 10-21 erg.Gauss-1)
hν  =  gβB  
B magnetic field (Gauss or mT) ν =  (gβ/h)B  =  2.8024  x  B  MHz  
 
EPR  is  the  resonant  absorp@on  of  microwave  radia@on   for  B  =  3480  G      ν    =  9.75  GHz    (X-­‐band)
by  paramagne@c  systems  in  the  presence  of  an  applied   for  B  =  420  G      ν    =  1.2  GHz  (L-­‐band)  
magne@c  field   for  B  =  110  G    ν    =  300  MHz  (Radiofre
Macroscopic  collec&on  of  electron  spins  
Ø  Previously  what  we  have  described  the  situa@on  for  an  isolated  electron,  however  in  
a  real  system  the  electrons  will  be  in  the  order  of  1010  or  even  more.  

Ø  There  will  be  interac@on  between  the  electron  or  with  environment  

Ø  Not  only  that  the  electron  will  be  confined  in  an  orbital  of  an  atom  or  a  molecule  

Even  for  the  macroscopic  system,  in  the  absence  of  Magne&c  field  these  electrons  will  be  
equally  distributed  between  the  spin  up  state  and  spin  down  state  (due  to  degeneracy  of  
these  two  states  (spin  up  (α)  and  spin  down(β)))    

In   the   presence   of   magne@c   field,   distribu@on   of   electron   between   the   spin   down   to  
the   spin   up   states   are   decided   by   the   thermal   equilibrium   with   the   laice   or  
environment.  

The  ra@o  between  the  number  (N)  of  α  spin  to  the  β  spin  given  by  Boltzmann’s  
distribu@on  equa@on  

kB  =  Boltzmann  constant    =  1.3806  ×  10-­‐23  J  K  


T  =  Absolute  temperature  of  the  laice.  
Spin  laTce  relaxa&on  

The  equa@on  becomes  

Popula@on  usually  at  lower  energy  state  will  be  higher,  hence  the  transi@on  from  β  
spin  state  to  α  spin  state  occurs.  
Due  to  con@nuous  irradia@on  of  electromagne@c  radia@on,  it  is  possible  to  have  
equal  distribu@on  of  electron  in  spin  down  (β)  state  as  well  as  spin  up  state  (α).  In  
this  situa@on,  epr  signal  will  saturate  or  disappears.    
 
However,   in   reality   even   if   you   run   the   experiment   con@nuously   for   long   @me,  
satura@on  doesn’t  occur  due  to  spin-­‐laice  interac@on.  Ie.  The  absorbed  energy  
will  be  lost  due  to  interac@on  with  environment  or  laice.    
 
This  spin  laice  relaxa@on  process  occurs  very  rapidly  which  is  usually  reported  as  
inverse  of  the  characteris@c  @me  T1  than  the  s@mulated  emission.  
Schema&c  representa&on  to  explain  spin  laTce  relaxa&on  
Spin-­‐laTce  relaxa&on  (longitudinal)  
Spin-­‐Laice  relaxa@on  @me  (T1):  This  relates  the  characteris@c  life&me  of  the  electronic  state  
and  is  determined  by  the  exchange  of  energy  via  thermal  vibra&ons  of  the  laTce.    

Simplest  individual  resonance  


line  referred  to  Spin  packet.  
Spin-­‐laTce  relaxa&on  rate  (1/T1)  
Spin-­‐Laice   relaxa@on   @me   (T1),   which   is   very   temperature   dependent,   measures   the  
characteris@c   @me   for   recovery   of   the   magne@za@on   of   the   paramagne@c   system   along  
the  applied  field  direc@on  aher  equilibrium  has  been  disturbed.  

Direct  process   Raman  Process   Orbach  Process  


Direct  process  
This  process  is  limited  to  electron-­‐phonon  interac@on  where  the  phonon  frequencies  are  
exactly  equal  the  applied  microwave  frquency.  
Mechanism:    Modula@on  of  ligand  field  interacts  with  the  vibra@onal  modes  of  the  laice    
Raman  process  
Here,  the  phonon  of  frequency  ν1  is  scaQered  by  the  paramagne@c  center  and  emerge  
with  a  different  frequency  ν2,  the  frequency  difference  being  equal  to  the  microwave  
frequency.    
Orbach  process  
Thermal  ac@va@on  from  one  electronic  states  of  EPR  transi@on  and  subsequent  emission  
or  absorp@on  of  a  phonon.  
Spin-­‐Spin  relaxa&on  (T2)  
Spin-­‐Spin  relaxa@on  is  concerned  with  mutual  spin  flips  caused  by  dipolar  and  exchange  
interac@on  between  the  assembly  of  spins  making  up  the  sample.    

Physical   significance:   Spin-­‐Spin   interac@on   is   concentra&on   dependent,   temperature  


independent   and   measures   the   @me   taken   for   oscilla@ng   component   of   the  
magne@za@on   in   the   plane   normal   to   the   applied   magne@c   field   to   get   out-­‐of-­‐phase  
with  one  another.  

Due   to   spin-­‐spin   interac@ons,   ohen   the   EPR   lines   obtained   are   broad,   within   that   fine  
structural  details  are  buried.  Example:  Hyperfine  interac@ons  (which  will  give  us  the  atoms  
surrounded  by  the  metal  ions)  
Observa&on  of  magne&c  resonance  
Epr  resonance  condi@on  can  be  achieved  by  two  ways  
 1.  Under  constant  magne@c  field,  changing  the  frequency  of  microwave  irradia@on  intensity.  
 2.  Under  constant  irradia@on  of  microwave  of  par@cular  intensity,  sweeping  the  magne@c  
field.    

Selec&on   rule   for  


o b s e r v i n g   t h e  
EPR  resonance  
 ΔMs  =  ±  1  
Advantage  of  recording  EPR  in  constant  microwave  frequency  
Example  1:  

Absorp&on     First  deriva&ve  of  absorp&on    


Advantage  of  recording  EPR  in  constant  microwave  frequency  
Example  2:  
Appears   to   be   an   isotropic  
Spectrum.  However  due  to  spin-­‐
orbit   coupling   (very   small),   the  
resonance  transi@on  occurs  very  
c l o s e l y ,   r e s u l t s   i n   l i n e  
broadening   or   line   width  
increases.  

Due   to   increased   quanta   of  


energy   of   microwave,     the  
distance   between   the   peak  
(resonance   condi@on)   also  
increases  linearly,  results  in  high  
EPR  spectra  of  Dibenzothiophene  radical  recorded  in       resolu@on  anisotropic  spectra.    
a)  X-­‐band  (9.5  GHz)  b)  W-­‐Band  (95  GHz)  
Instrumenta&on  of  CW-­‐EPR  

There  are  four  important  components  in  CW-­‐EPR:    


 1.  Microwave  resonator  
 2.  Microwave  bridge  
 3.  Magnet  System  
     4.  Control  electronics  
Microwave  Resonator  
Microwave  resonator  is  the  sample  cell.  ie  This  is  where  the  sample  under  inves@ga@on  
needs  to  be  place  (sample  holder)  
 
In  addi@on,  other  two  main  purpose  of  microwave  resonator  is  
Ø  To  create  sufficient  microwave  magne@c  field  at  the  sample  posi@on  to  drive  the  transi@on  
Ø  To  efficiently  convert  the  sample  response  into  detectable  microwave  signal  
   
Both  of  these  factor  depends  on  the  value  of  quality  factor  Q  
L-­‐Band  =  1  GHz  
S-­‐Band  =  ~3  GHz  
X-­‐Band  =  9.5  GHz  
Q-­‐Band  =~35  GHZ  
νres  =  Resonance  frequency  of  the  resonator   W-­‐Band  =  95  GHZ  
Δν  =  Width  of  the  resonance  curve  
Microwave  Resonator  

Microwave   energy   is   stored   in   resonator   leads   to   an   enhancement   in   microwave  


magne@c  field  (B1)  and  electric  field  (E1)  compare  to  its  free  space  value  

B1  generated  by  microwave  power  P  is  given  by  

B1  =  c  ×  (Q×P)1/2  

Where   c   =   Conversion   factor   of   a   resonator   the   value   ohen   depends  


resonator  size  (Smaller  the  resonator,  larger  the  conversion  factor  

Also   more   importantly   the   peak   to   peak   amplitude   (ΔSpp)   is  


also  depends  on  the  Quality  factor  (Q)  
 
ΔSpp  α  η  ×  Q  
Note:  

ΔSpp  α  Q  
ΔSpp  α  νres  
Microwave  Resonator  
According   to   the   above   equa@on   listed   in   the   previous   slide,   the   signal   amplitude   will  
increases   with   increasing   sample   volume,   along   with   Q.   However,   ohen   Q   factor   is  
lowered   upon   increasing   the   sample   volume,   hence   the   op@mum   sample   needs   to   be  
used  for  corresponding  instrument  depends  on  the  frequency  

v  The  benefits  of  resonator  Q  factor  with  respect  to  B1  (microwave  magne@c  
field)  and  the  sensi@vity  of  the  instrument  (ie,  resolu@on  of  peak  to  peak  
separa@on)  
v   Microwave  frequency  doesn’t  have  large  window  to  sweep  across,  unlike  
the  external  magne@c  field    
v  Hence,  EPR  is  recorded  with  fixed  microwave  frequency  with  constant  
intensity  by  sweeping  the  external  magne&c  field  across.      
Microwave  Resonator  

Microwave                              Electromagne@c  radia@on                        Magne@c  field  component  (B1)  and  


Electric  field  component  (E1)  (⊥  to  each  other)      

Microwave  Electric  and  magne@c  field  components  

Magne@c  field  component  (B1)                                      Magne@c  resonance  transi@on    


Electric  field  component  (E1)                                              Non-­‐resonant  microwave  absorp@on  

This  causes  decrease   Through  electric  dipole  


in  Q  factor  and   moment  of  the  sample  
undesired  sample  
hea@ng.  
Hence  EPR  resonator  has  been  designed  based  on  electrodynamics  principle  
where  the  sample  felt  the  B1  and  minimizes  the  E1  at  the  sample  posi@on    
II.  Microwave  Bridge  

Microwave  frequencies  are  tradi@onally  classified  in  term  of    a  bands.  

L-­‐  Band          1  GHz              Mainly  used  for  biological  sample  and  for  live  animals  study  (To  
reduce  the  absorp@on  of  electric  field  component)  

S-­‐Band  (3  GHz),  X-­‐Band  (9.5  GHz),  Q-­‐Band  (34  GHz),  K-­‐Band  (24  GHz),  W-­‐Band  (95  GHz)  

Usually  the  higher  frequency  EPR  instrument  is  used  to  get  detailed  insights  about  the  
structure.  Lower  frequency  (L  and  S  band)  instruments  are  used    for  biological  applica@on.  
“Cri&cal  coupling”  
The  complete  absorp@on  of  CW  microwave  power  by  the  resonator  is  called  Cri@cal  coupling  
or  matching.  
Reference  arm  

T o   K e e p   t h e  
detector   in   Linear  
response  

Large   and   small  


signals   will   be   given  
in  correct  amplitude  
ra@o  

When  the  magne@c  field  is  swept  through,  resonance  microwave  energy  is  absorbed  
by  the  sample  that  causes  a  change  of  the  resonator  Q  factor.  It  is  this  change  that  
results  in  the  reflec@on  of  microwaves  from  the  resonator  that  in  turn  is  registered  as  
the  EPR  signal.  
Automa&c  Frequency  Control  (AFC)  

The  change  in  the  resonator  Q  factor,  is  EPR  dispersion  signal  
 
The  change  in  the  matching  condi@on  of  resonator  (that  is  resonator  reflects  the  
microwave  signal)  is  EPR  absorp@on  
 
EPR  absorp@on  and  EPR  dispersion  process  appear  in  parallel.  
 
AFC  separates,  more  precisely  eliminates  the  dispersion  signal,  only  absorp@on  signal  is  
detected.  
 
Not  only  that  AFC  is  linked  to  oscillator  frequency  which  is  locked  to  the  resonator  
frequency.  
 
     
III.  Magnet  system  

Magnet  
Power  supply  unit   Magnet  system  
Field  sensor  
Field  Regulator  

Lower  than  W-­‐band  (95  GHz),  we  can  use  electromagnets     Magnet  system  
  supplies  the  magne@c  
W-­‐band  and  above,  we  have  to  use  superconduc@ng  magnet   field  

X-­‐Band  (9.5  GHz)  =  0.3  T  or  3000  G  


Q-­‐Band  (35  GHz)  =  ~1.3  T  or  13  kG  

Homogeneity  of  magne@c  field  at  the  sample  posi@on  is  an  another  important  
factor  to  be  keep  in  mind  
Magnet  system  

Hall  sensor  is  installed  in  between  the  


Measuring  and  control  the   gap  of  electromagnet:  
system  to  get  the  correct   Ø  To  give  us  accurate  value  of  Magne@c  
magne@c  field  value     field  posi@on,  line  width,  separa@on  
Ø  Mounted  off  center,  to  avoid  the  
field  modula@on  applied  to  resonator  
Ø  Needs  calibra@on  (sample  posi@on  to  
Hall  sensor  posi@on,  since  it  is  
Alterna&vely  sample  calibra&on  can  be  done  by  using   mounted  offcenter)  
standard  sample:   Ø   NMR  Gauss  meter  used  to  calibrate  
2,2-­‐Diphenyl-­‐1-­‐picryl-­‐hydrazyl  (DPPH)  =  2.0036   the  Hall  sensor  with  very  high  
Cr  =  1.979  (Cr  in  MgO  matrix)  
3+ 3+
precision  (10-­‐6  to  10-­‐7)  
Mn2+  =  2.0011    
Fundamentals  

1.  g  values  
-­‐  Spin-­‐orbit  coupling  
-­‐  Spin  orbit  coupling  arising  from  the  ground  state  (in-­‐Phase)  
-­‐  Spin-­‐orbit  coupling  arising  from  mixing  of  ground  state  with  excited  
states  (Out-­‐of-­‐phase)  
2.  Fine  structure  
-­‐  Zero  field  spliing  (Due  to  spin-­‐orbit  coupling  and  spin-­‐spin  interac@on)  

3.  Hyperfine  interac@on  

4.  Isotropic  and  anisotropic  EPR  Spectra  


Spin-­‐orbit  coupling  basic  concepts  
To  Explain  the  concept  we  will  consider  the  transi@on  metal  ions  

In  gaseous  ion  all  five  d-­‐orbitals  are  degenerate  and  splits  in  octahedral  field  as  below  

eg  orbital  

Δo  

T2g  orbital  
Spin-­‐orbit  coupling  basic  concepts  
Consider  an  electron  in  an  orbital,  where  electron  spins  by  itself  (creates  spin  magne@c  
moment),  which  revolves  around  the  orbital  (hence  orbital  angular  momentum).  

If  total  spin  angular  momentum  (s)  and  total  orbital  


µtotal   angular   momentum   (l)   are   independent   or   not  
coupled   with   each   other   the   resultant   magne@c  
moment  of  the  system  is  given  by  
µorbital  
µspin   [1]  
e-­‐  
However,  in  reality  both  spin  and  orbital  angular  
momentum  are  coupled  together,  ie.  S  and  L  are  
no   longer   good   quantum   numbers,   this   requires  
different   quantum   number   “J”   called   total  
The   total   angular   momentum   quantum   angular   momentum.   The   resultant   magne@c  
number  takes  the  absolute  values   moment  is  given  by  
J  =  |L+S|..|L+S-­‐1|..  |L-­‐S|  
g   is   Lande’s   g-­‐factor   which   has  
defini@ve   value   for   type   of   compounds  
which   depends   on   the   spin   and   orbital  
angular  momentum  
Rules  for  predic&ng  the  ground  state  Term  Symbol  (free  gaseous  ion)  

General  nota@on  for  term  Symbol:  


Where  S  is  spin  mul@plicity  (S  =  2s+1)  [s  =  n/2]  
L  is  the  TERM  corresponds  to  the  highest  L  value  
J  is  total  angular  momentum  (J  =  |l+s|..|l+s-­‐1|..|l-­‐s|)  
1)  For  any  electronic  configura@on,  states  with  highest  S  value  (  S  =  n/2  where  n  =  
number  of  unpaired  electrons)  is  lowest  in  energy  
2)   States  with  highest  L  will  be  in  lowest  energy  

3)  J  with  |l  -­‐  s|  value  will  be  in  lowest  energy  if  the  orbital  is  less  than  half  filled  
and  J  with  |l  +  s|  value  will  be  in  lowest  energy  if  the  orbital  is  greater  than  half  
filled.    
Example:     L  value  TERM  
d1  configura&on:            0                              S  
         1                              P  
↑            2              D  
+2   +1   0   -­‐1   -­‐2            3                              F  
S  =  2(1/2)+1  =  2            4                              G  
l  =  1×  (+2)  =  2            5                              H  
J  =    l-­‐s  =  2-­‐1/2  =  3/2            6                                I  

Work  out  for  d2  to  d10  configura&on?  


Ground  state  Term  Symbol  for  gaseous  ion  
m
ML=Σml
l

Configuration Example 2 1 0 -1 -2 L S Ground Term


1 3+ 2
d Ti 2 1/2 D
2 3+ 3
d V 3 1 F
3 3+ 4
d Cr 3 3/2 F
4 3+ 5
d Mn 2 2 D
5 2+ 6
d Mn 0 5/2 S
6 2+ 5
d Fe 2 2 D
7 2+ 4
d Co 3 3/2 F
8 2+ 3
d Ni 3 1 F
9 2+ 2
d Cu 2 1/2 D
General  comparison  of  Term  Symbol  for  gaseous  ion  d1-­‐d9  

ml ml
2 1 0 -2 ML S Example
-1 Configuration Ground Term
2 1 0 -1 -2 ML S
1 3+ 2
d2 1/2 Ti D 2 1/2
2 3+ 3 2
d3 1 V F 3 1
3 3+ 4 3
d3 3/2 Cr F
4
3 3/2
4 2+ 5
d2 2 Cr D 5 2 2
5 2+ 6
d0 5/2 Mn S 6 0 5/2
6
d2 2
2+ 5
Fe D 5 2 2
7 2+ 4 4
d3 3/2 Co F
3
3 3/2
8 2+ 3
d3 1 Ni F 2 3 1
9 2+ 2
d2 1/2 Cu D 2 1/2
Energies  of  J  levels  in  V(III)  ion  
3λ  
J  =4   Energies  of  J  states  can  be  calculated  by  

3F  

-­‐λ  
J  =3  
+2  
+1  
-­‐4λ   J  =2   0  
-­‐1  
In  magne@c  field   -­‐2  
l  =  3  
s  =  1   Energy  difference  between  the  J  states  is  calculated  by    
J  =  |l+s|  ..|l+s-­‐1|..|l-­‐s|
J  =  4,3,2   ΔEJ,  J+1=  λ(J+1)
ΔE3,2          =  3  λ
 ΔE4,3          =  4  λ  
Microstates  
“Microstates:  In  Terms  of  Molecular  Energy  Levels”,  

In  contrast,  a  microstate  for  a  system  is  all  about  @me  and  the  energy  of  the  molecules  in  
that  system.  "In  a  system  its  energy  is  constantly  being  redistributed  among  its  par@cles.  In  
liquids  and  gases,  the  par@cles  themselves  are  constantly  redistribu@ng  in  loca@on  as  well  
as   changing   in   the   quanta   (the   individual   amount   of   energy   that   each   molecule   has)   due   to  
their   incessantly   colliding,   bouncing   off   each   other   with   (usually)   a   different   amount   of  
energy  for  each  molecule  aher  the  collision..  Each  specific  way,  each  arrangement  of  the  
energy  of  each  molecule  in  the  whole  system  at  one  instant  is  called  a  microstate."  

hQp://entropysite.oxy.edu/microstate/index.html  
Microstates  in  V(III)  ion  
For  V(III)  ion,  d2  configura@on,  hence  n=  10  and  r  =  2  
Total  number  of  microstates  available  for  d2  ion  is  =  10!/2!8!  =  45  

3λ   To  calculate  number  of  microstate  available  for  a  par@cular    


J  =4  
9   ground  state  term  symbol  is  given  by    
             (spin  mul@plicity  (2s  +  1)  ×  Orbital  mul@plicity(2l  +1))    

3F   For  3F  =  3×7  =  21  states  


21   (because  s  =  1  and  l  =  3)  

-­‐λ  
J  =3  
7   +2  
+1  
-­‐4λ   J  =2   0  
5  
-­‐1  
-­‐2  
In  magne@c  field  
J  split  maintain  the  center  of  gravity  ie  the  energy  of  the  higher  energy  state  (J=4)  is  
compensated  by  the  sum  of  energies  of  lower  energy  state  (J  =  2  and  3)  
9*3-­‐7*1-­‐5*4  =  0  
Microstates  for  d2  ion  

For  a  d2  ion  maximum  l  value  that  we  can  obtain  is  =  4  and  other  possible  l  values  are  3,2,1,0    

l  =  4  
↑↓  
+2   +1   0   -­‐1   -­‐2  
l  =  4;  s  =  0
1G

Microstates:  Spin  multiplicity  (2s  +1)  ×  Orbital  multiplicity  (2l  +1)


1×  9  =  9

l  =  3  
↑   ↑  
+2   +1   0   -­‐1   -­‐2  
l  =  3;  s  =  1
3F

Microstates:  Spin  multiplicity  (2s  +1)  ×  Orbital  multiplicity  (2l  +1)


3×  7  =  21
Microstates  for  d2  ion  
l  =  2   l  =  2  
↑   ↑   ↑↓  
+2   +1   0   -­‐1   -­‐2   +2   +1   0   -­‐1   -­‐2  
l  =  2;  s  =  1 l  =  2;  s  =  0
3D 1D

Microstates:  Spin  multiplicity   Microstates:  Spin  multiplicity  


(2s  +1)  ×  Orbital  multiplicity   (2s  +1)  ×  Orbital  multiplicity  
(2l  +1) (2l  +1)
3×  5  =  15 1×  5  =  5

l  =  1   l  =  0  
↑   ↑   ↑↓  
+2   +1   0   -­‐1   -­‐2   +2   +1   0   -­‐1   -­‐2  
l  =  1;  s  =  1 l  =  0;  s  =  0
3P 1S

Microstates:  Spin  multiplicity   Microstates:  Spin  multiplicity  


(2s  +1)  ×  Orbital  multiplicity   (2s  +1)  ×  Orbital  multiplicity  
(2l  +1) (2l  +1)
3×  5  =  15 1×  5  =  5
Rela&onship  between  Term  symbols  and  Tanabe-­‐Sugano  diagram  

1G

3F

3D 1D

3P

1S
Orbital  contribu&on  to  the  magne&c  moment  

When  Orbital  contribu@on  arises  ?  


 In  order  that  an  electron  can  have  orbital  angular  momentum,  it  must  be  possible  
to   transform   the   orbital   that   it   occupies   into     an   exactly   equivalent   and   degenerate  
orbital  by  rota<on.  
 
The  electron  is  then  effec@vely  rota@ng  about  the  axis  used  for  the  orbital  rota@on.  

(the  circula@on  of  an  electron  around  the  nucleus  is  like  the  circula@on  of  a  current  
through  the  turns  of  a  solenoid)  
Spin-­‐Orbit  coupling  in  transi&on  metal  octahedral  complexes  
The  degenerate  t2g  orbitals  (dxy,  dxz,  dyz)  can  be  interconverted  by  90°  rota@ons  

eg  orbital  

Δo  

T2g  orbital  

C4  

Thus,  an  electron  in  a  t2g  orbital  can  


C4   contribute  to  orbital  angular  
momentum  

e.g.  the  dxz  orbital  is  transformed  into  the  dyz  orbital  by  a  rota@on  of  90°  about  the  z-­‐axis  –  
during  this  rota@on  the  electron  is  orbi@ng  the  nucleus  
Is  spin-­‐orbit  coupling  will  be  there  in  eg  orbital  in  Oh?  

The  eg  orbitals  cannot  be  interconverted  by  rota@on  due  to  their  nature  of  shape  
of  orbital  

The  unpaired  electron  in  eg  orbital  cannot  contribute  to  spin-­‐orbit  coupling  
Is  t2g  to  eg  inter-­‐conversion  possible  by  rota&on?  

Yes.  Inter  conversion  is  possible.  

However,   in   octahedral   field   or   any  


ligand   field,   t 2g   and   e g   orbital  
separated   by   Δ0   (in   Oh   field),   hence  
the  mixing  of  dxy  and  dx2-­‐y2  will  depend  
on  the  energy  gap  between  these  two  
level.  

dxy  /  dx2-­‐y2  orbital  mo@on  about  z  axis  


dxz  /  dyz    orbital  mo@on  about  z  axis  
dxz  /  dxy    orbital  mo@on  about  x  axis  
dyz  /  dxy    orbital  mo@on  about  y  axis  
Orbital  contribu&on  to  the  magne&c  moment:  High  spin  octahedral  dn  
ions  
Think  of  possible  t2g  electron  arrangements  

d1  
dxz   dyz   dxy   dxz   dyz   dxy   dxz   dyz   dxy  

Possible  t2g  arrangements  =  3   d1  e.g.  Ti(III)  


Orbital  contribu@on  =   YES  

d2  
dxz   dyz   dxy   dxz   dyz   dxy   dxz   dyz   dxy  

Possible  t2g  arrangements  =  3  


Orbital  contribu@on  =  yes   d2  e.g.  V(III)  
Orbital  contribu&on  to  the  magne&c  moment:  High  spin  octahedral  dn  
ions  

But,  electrons  in  t2g  orbitals  will  not  always  contribute  to  orbital  
angular  momentum  
e.g.  consider  octahedral  Cr(III)  d3,  t2g3  
an  electron  in  the  dxz  orbital  cannot  by  rota@on  be  placed  in  e.g.  
the  dyz  orbital  because  this  orbital  already  contains  an  electron  
with  the  same  spin  as  the  incoming  electron  

dxz   dyz   dxy   dxz   dyz   dxy  

Therefore,  there  is  only  one  possible  t2g  arrangement,    


and      NO  
             orbital  angular  momentum  
Orbital  contribu&on  to  the  magne&c  moment:  High  spin  octahedral  dn  
ions  
Think  of  possible  t2g  electron  arrangements    
 

d4  high  spin   t2g  half-­‐filled  


dxz   dyz   dxy  

Possible  t2g  arrangements  =  1   d4  e.g.  Cr(II)  


Orbital  contribu@on  =     NO  

d5  high  spin   t2g  half-­‐filled  


dxz   dyz   dxy  

Possible  t2g  arrangements  =  1   d5  e.g.  Mn(II),    


Orbital  contribu@on  =   NO                            Fe(III)  
Orbital  contribu&on  to  the  magne&c  moment:  High  spin  octahedral  dn  
ions  
Think  of  possible  t2g  electron  arrangements    
 

d6  high  spin  
dxz   dyz   dxy   dxz   dyz   dxy   dxz   dyz   dxy  

Possible  t2g  arrangements  =  3   d6  e.g.  Fe(II)  


Orbital  contribu@on  =   YES  

d7  high  spin  

Possible  t2g  arrangements  =  3   d7  e.g.  Co(II)  


Orbital  contribu@on  =   YES  
Orbital  contribu&on  to  the  magne&c  moment:  High  spin  octahedral  dn  
ions  
Think  of  possible  t2g  electron  arrangements    
 

Ni(II)  d8   t2g  filled  

Possible  t2g  arrangements  =  1   d8  e.g.  Ni(II)  


Orbital  contribu@on  =   NO  

Cu(II)  d9   t2g  filled  

Possible  t2g  arrangements  =  1   d9  e.g.  Cu(II)  


Orbital  contribu@on  =   NO  
Orbital  contribu&on  to  the  magne&c  moment:  Contribu&on  due  to  
excited  state  
think  of  possible  t2g  electron  arrangements    
 

t2g  not  comp.  


Ni(II)  d8   t2g  filled  
filled  

Excited  state  
Possible  t2g  arrangements  =  1  
Orbital  contribu@on  =   NO  (due  to  ground  state)  

d8  high  spin  ES  

Possible  t2g  arrangements  =  3  


Orbital  contribu@on  =   YES  

μexp  >  μs  for  Oct.  Ni2+  


How  to  dis&nguish  Ni  Td  and  Oh  complexes?  
Therefore  for  Oh  Ni(II)  the  magne&c  moments  are  larger  if  the  GS-­‐ES  gap  is  small.  
BUT  FOR  Tetrahedral  Ni(II)  situa&on  is  en&rely  different  ….  

d8  HS  Tetrahedral  

Possible  t2g  arrangements  =  3  


Orbital  contribu@on  =   YES  

Μexp  Oh  Ni(II)  is  smaller  than  Μexp  Td  Ni(II)    

Exp.  NiCl42-­‐,  Ni(HMPA)42+  (HMPA=hexamethyl  phosphoramide)  


have  magne@c  moment  larger  than  4  BM.    
(larger  the  distor@on  smaller  the  magne@c  moment  )  
 
USEFUL  IN  DISCRIMINATING  Oh  vs  Td  structures.  
How  does  this  spin-­‐orbit  coupling  affect  the  g-­‐value?  
Ø  For  free  or  isolated  electron,  the  g-­‐value  is  determined  as  2.0023  

First  
  deriva@ve   Absorp@on   Measure  B  (“resonance  field”)  for  
given  applied  υ.  
 

Magne@c  field  (G)  


Ø  Presence  of  orbital  angular  momentum  can  interact  with  spin  angular  momentum  
construc@vely  or  it  may  oppose  each  other.  In  such  cases,  the  observed  magne@c  
moment  either  higher  than  or  lower  than  (respec@vely)  compare  to  free  electron.  
Hence  there  will  be  change  in  g-­‐value  of  complex.  

Ø  Magne@c   field   in   linear   rela@onship   with   magne@c   moment,   ie   larger   the   moment,  
spliing  of  moment  in  magne@c  field  will  be  higher.    

Ø  From  g-­‐value  we  can  deduce  the  electronic  environment  of  an  atom  or  ion.    
What  controls  g-­‐value?  

Where    
a  denotes  covalency,  usually  a2  will  be  <  1    (if  an  unpaired  electron  
delocalized,  the  spin-­‐orbit  coupling  will  be  reduced.  
λ  =  Spin  orbit  coupling  constant  
ΔE  =  Energy  gap  between  the  orbitals    
(Bigger  ΔE  →  less  mixing  →  smaller  shih  of  g  from  ge.  Note  ΔE  
corresponds  to  an  electronic  dd  transi&on!)  
n  =  depends  on  the  iden@ty  of  orbitals  which  are  being  mixed  

λ  =  Posi@ve  for  less  than  half  filled  orbitals  


λ  =  Nega@ve  for  more  than  half  filled  orbital  
Hence,   the   g-­‐value   is   higher   than   the   free   electron   g-­‐value   (ge)   for   more   than   half   filled  
electronic  configura@on  and  the  same  will  be  less  the  free  electron  g-­‐value  (ge)  for  less  than  
half  filled  d-­‐orbital.    
Predic&on  of  n  value  from  Magic  Pentagon  