For
Digital Conductivity
Meter
1
SPECIFICATIONS
1) Conductivity
Range : 0 to 1000 milli mho in 6 ranges
Range Selection : Automatic
0 to 20 ms/cm
20 to 200 ms/cm
200 to 2 ms/cm
2 to 20 ms/cm
20 to 200 ms/cm
200 to 1000 ms/cm
Resolution : 0.1 ms/cm
Cell Constant : Adjustable
Temperature co-eff. : Adjustable between 0 to 4%
Temperature Comp. : Automatic for entire range
Accuracy : +0.5% of range + 1 digit
2) TDS
Range : 0 to 1000 ppt in 6 ranges
Range Selection : Automatic
Resolution : 0.1 ppm
TDS Factor : 0.5 and adjustable manually
3) Salinity
Range : 0 to 50 ppt
Resolution : 0.1 ppt
Accuracy : +0.5% of range + 1 digit
Temperature Comp. : Automatic
Salinity Factor : .75 and Adjustable Manually
4) Temperature
Resolution : 0.1 ºC
Accuracy : + 0.2 %
2
INTRODUCTION
KEYBOARD
The instrument consists of six keys. The function of each key is as follows:
START
- MICROCONTROLLER -
-COND-TDS-SAL-
SYSTEM CHECK
SYSTEM CHECK
DATE TIME
PROGRAM MODE:
To enter into program mode, press PRG Key, The display will show
Calibrate Memory
(PROGRAM MODE)
Date/Time Print
NOTE: To enter into any mode in program mode press ENT key
CALIBRATION MODE
To calibrate the instrument take the cursor below Calibrate & Press ENT key. Display Shows
CAL
4
TEMPERATURE CALIBRATION
Cal TEMP
MANL LO HI UNCAL
AUTO MODE
Connect temp.calibrator to the 3-pin temp socket & put knob of calibrator at low temp.
Take the cursor below Lo Press ENT key, display will show:
CAL TEMP LO °C
If the temperature is not as per your standard value, press INC / DEC key till display shows the
correct value. Press ENT key, the display is adjusted as per your standard value. And display will
shows
Cal TEMP
MANL LO HI UNCAL
Now put the knob mark of calibrator at high temp. & take the cursor below HI & press ENT key,
CAL TEMP HI °C
Press INC/DEC key till display shows the correct value and press ENT key, the display will show:
5
Cal TEMP
MANL LO HI UNCAL
MANUAL MODE
In manual mode you can set the temp. that will display when temp. Probe is not
connected. Take the cursor below MANUAL & Press ENT key
XXX.X
Press INC/DEC key to set your temp. in manual mode press ENT key, the display
Will shows
Cal TEMP
MANL LO HI UNCAL
UNCAL MODE
With the help of UNCAL you can uncalibrate the Temp. Calibration
NOTE:
1) Do not enter the same value for LO & HI TEMPERATURE. There should be more than 20 degree
cel. Difference between LO & HI temperature
2) If same values entered for low & high temp .then it shows TEMP ERROR-RECAL
6
CONDUCTIVITY CALIBRATION
NOTE:
For conductivity measurement you should know the cell constant of the cell you are using and the temperature co-
efficient of the solution. There are two ways of entering the cell constant. One is directly enter the cell constant of
the cell. Another is by checking the conductivity. The second method is useful when you don’t know the cell
constant. You know the conductivity of the solution what you have to do is simply adjust the conductivity the
correct value. Cell constant will automatically be displayed when you are checking cell constant. Take the cursor
below COND & Press ENTER key. Display shows:
Cal CND
CELLCNST TREFCOF
Now take the cursor below CELLCNST. & Then Press ENT Key display shows
Cal CELLCNST
BYVAL BYMEASURE
Now take the cursor below BY VAL if you know the cell constant value & Press ENT key display shows cell
constant of the solution 1.000 constant of your cell. If you know the cell constant, then you can straight way enter
the cell constant with INC/DEC key. Press ENT & then PRG key. Now take the cursor below TREFCOF & Press
ENT key display shows
SET TREF-COF
REFTEMP TEMPCOF
Take the cursor below REFTEMP & Press ENT key display shows
SET REFTEMP
25.0 ºC 20.0 ºC
Now you can select here one of these given reference temp. by Pressing ENT key. Also as shown above you can
select temp. Coefficient. So selects the TEMPCOF & Press ENT key, Display shows
Here you can adjust the temperature co-efficient of the solution. If you don't know the temperature co-efficient of
solution Press ENT key otherwise adjust with the help of INC/DEC key & Press ENT key. Press PRG key & again
PRG key display shows the Calibration mode.
TDS CALIBRATION
Take the cursor below the TDS and Press ENT Key display shows:
Here you can adjust the TDS FACTOR, CELLCONSTANT & TEMP COEFICIENT in the same way as you have
done for conductivity. TDS 0.50 (by default)
If you don’t want to change this factor press PRG key. Display shows the calibration mode
7
SALINITY CALIBRATION
Take the cursor below the SAL in Calibration mode and Press ENT Key display shows:
Here you can adjust the SAL FACTOR, CELLCONSTANT & TEMP COEFICIENT in the same way as you have
done for conductivity. SAL 0.75 (by default) If you don't want to change this factor press PRG key. Display shows
the Calibration mode.
1. It is always recommended to check the cell constant of conductivity cell before every measurement. It is done
as below:
a. Dip the cell in the solution whose conductivity is known.
b. Keep the instrument in cal mode on conductivity parameter.
Adjust the displayed value with the help of INC / DEC key to the correct value. Press the ENT key.
2. Then press the CAL key. It will be show the correct cell constant of the conductivity cell. Conductivity cell
may be cleaned by using a warm, 50% detergent solution and scrubbing with very fine brush. Care must be taken
that” PLATINISATION” is not damaged by excessive scrubbing. For more deposits a 0.2% solution of
hydrochloric acid may be used.
3) Don't finger the cell electrodes.
4) Select cell of proper cell constant for accurate readings.
ROUTINE MAINTAINENCE
The design of the instrument is such that no routine maintenance is required to be carried out by
the customer.
As with any high impedance instrument, it should not be exposed unnecessarily to dirty
atmosphere, and it should be covered by a plastic dust cover when not in use.
WARNING