Technical Publication

Ver 2.0

INSTALLATION
EVA-HF325/525
TABLE OF CONTENS

1. INTRODUCTION ------------------------------------------------- 3

2. COMPOSITION -------------------------------------------------- 4

3. ELECTR IC ENVIRONMEN T------------------------------------ 5

4. PACKING ------------------------------------------------------------ 7

5. DIMENSION---------------------------------------------------------- 8

6. INSTALLATION----------------------------------------------------- 9

7. installation trouble shooting------------------------------------ 15

2
1. INTRODUCTION

Installation person, operator and patient always be careful their


head at the tube or assembling collimator

Installation person, operator and patient always be careful their


leg and foot under the table when moved or used machine.

Installation person, operator and patient always


be careful their hand at the sharp part or need the
electronic safety

3
2. COMPOSITION

TUBE
COLLIMETER
STAND
TABLE
X-RAY CONTROLLER
TABLE MOVING S/W
HV TRANSFORMER
RAIL
RAIL SUPPORT
CABLE DUCT
STAND GUIDE
HANDLE BAR

3. ELECTRIC ENVIRONMENT

4
Table. Line Power requirement

GENERATOR MODEL EVA-HF325 EVA-HF525

Max.Power kW 37.5kW 40kW

Maximum mA 300mA 500mA

Maximum kVp 125kVp 125kVp

Power Line 210 -230VAC, Single-Phase, 50/60Hz

Table. Line impedance requirement

LINE VOLTAGE GENERATOR POWER( 37.5 kW)

208 VAC 0.053Ω

230 VAC 0.065Ω

240 VAC 0.070Ω

The installation should comply with all the electrical


requirements indicated in this document. These requirements
should be upgraded if Local Standards are more stringent.

RMS Line current during a X-ray exposure, generator stand-by consumption(W),


the differential sensitivity(mA) and the thermomagnetic breaker should be:
Maximum power line impedance :

Table. Magnetic breaker requirement

5
LINE VOLTAGE SINGLE-PHASE( 37.5kW )

208VAC 273 A

230VAC 245 A

240VAC 232 A

Minimum kVA required Maximum kW ×1.2

Stand-by Consumption 500 W

Differential Sensitivity 30 mA

50% of the RMS line current


Thermomagnetic Breaker
(RMS = momentary line current based on 100ms X-ray exposures)

Below 210VAC requires an auxiliary boost transformer to adequate the line voltage to the
generator input

Table. Wire size requirement


(SINGLE-PHASE 1Ø)

Wire size at ;
Line voltage
15m 30m

208 VAC 50 105

230 VAC 50 105

240 VAC 50 95

6
4. PACKING

tube assembly
stand guide

The generator is shipped in Table to facilitate transport and


installation.
Upon receipt of the X-ray unit and associated equipment,
inspect all shipping containers for signs of damage. If
damage is found, notify the carrier or his agent
immediately.

5. DIMENSION

7
1. tube forward and backward movement range : 185mm
2. X-axis center rotation angle range : over ± 90°
3. Z-axis center rotation angel range : over ± 90°
4. tube max upward movement : above ground 1720 mm
5. tube max downward movement : above ground 650 mm
6. tube up and down movement range : 1170 mm
7. tube left and right movement range : 1940 mm
8. table forward and backward movement range : 200mm
9. table left and right range : 600mm

207

74

20
200

110

120

245

Total weight : 430kg

6. Installation

8
braker stopper

Insert the plate


Insert the stopper.
Connect the braker at both guide and fixed after release the bolt adjust the space (about 2mm
spacing)

Connect the rail.


Connect the stand-guide.
Insert the rail-supporter.

9
Centering weight

Fixing bolt

Insert stand to stand-guide and tightened the bolt.


(It must fixed with adjust make the table level with the stand..)

A angle

B angle

A angle = B angle Side-view

Fix the center weight and remove two fixing bolt.


(when pull out the bolt , the center weight will slowly down)

10
connect the tube to the stand.
connect the doct.
input the wire inner rail space and pull out.
connect the wire to the connecter.
(warning : when assembling, be careful not cutting the wire of the connecter)

11
Assemble front and back case.

insert rail-cover into rail and sticking.


(Release the dish-head-bolt and tighten with the rail-cover by the bolt.)

Cable Connection
Two types(high voltage cable and 10p cable) are used. When you are connecting
high voltage cable, be careful of its polarity.
*CAUTION: Cable must be connected deep into the hole.

Cathod( )

12
Anode( )
And connect 10p cable to the terminal board according to the following labels.

Lock(+) : stand-driving magnetic conductor power (+)


Lock(-) : stand-driving magnetic conductor power (-)
CM(AC24V) : collimator power
CM(AC24V) : collimator power
BUCKY 1 : table bulky power cable
BUCKY 1 : table bulky power cable
BUCKY 2 : stand bulky power cable
BUCKY 2 : stand bulky power cable
BUCKY 3 : bulky power cable(spare)
BUCKY 3 : bulky power cable(spare)
TUBE TEMP : TUBE over-heat protection cable
TUBE TEMP RTN : TUBE over-heat protection cable
W(COM) : TUBE ROTOR connection cable

13
G(MAIN) : TUBE ROTOR connection cable
B(PHASE) : TUBE ROTOR connection cable
DC24V : handle bar power cable(LAMPpower)
DC0V : handle bar power cable(LAMP power)
110VAC : TABLE power cable
0VAC : TABLE power cable
GND : ground

7. Installation trouble shooting


7.1 bucky stand trouble shooting

14
1) locker interval(top cover off)

two bolts release and calibrate..


(doing move bucky device)

2)If x-ray film image moved, you should fix bucky to wall

first, remove bolts(1) and top cover off


second fix “S”plate to top cover(2)cover on..

3)Bucky stand input signal

1. GND
2. DC+(locker power)
3. DC-(locker power)
4. START(BUCKY2—moving motor power110VAC)
5. START(BUCKY2— moving motor power 110VAC)

SEE PICTURE (5P CABLE)

15
pic1. Bucky stand wire

BUCKY 2/BUCKY2
DC+/DC-
Pic2. connection location

16
3)if grid moving speed too fast or too slow, you can change motor speed

move slow motor speed

17
side bolts view front bolts view

1) side bolts off and side cover off


2) front bolts off and front wood cover off
3) transformer output wire change..(refer schematic)

18
7.2 Table bucky trouble shooting
1. remove rubber stopper.

2. remove locker.

2. move top table a little.

3. check bucky assembly.

grid
spring
4. check hitter

19
hitter
ready s/w push --Æ hitter hitting spring
(check 24VDC in hitter when ready s/w is pushed)
5. check bucky signal

to hitter(24vdc when ready s/w pushed)


from bucky sigmal(AC110V when ready s/w pushed)
bucky relay(k6)

6. check bucky signal

bucky 1 signal(110vac when ready s/w


pushed)

20
7. bucky squency

ready s/w

ready

exposure s/w

x-ray

hitter signal

hitter moving signal : when ready s/w push--Æ move hitter


when exposure s/w push Æ release hitter

21
7.3 change brocken bearing in table

Move

first, remove rubber stopper on top-table and remove breaker and move top-table to
right-side little.(consider room size).
rubber stopper on top-table breaker

Second, take out bearing assembly(refer picture)

6800 bearing

M5 bolts

picture: bearing assembly


first, remove M5 bolts and take out bearing assembly

22
M6 bolts

second, change broken bearing with same bearing


and put together reverse method

7.4 Locker interval change

fix with volts and wind tightly.

interval : 2-3mm
when pushed foot s/w, check interval of locker
push foot s/w Æ release locker
not push foot s/w Æ attached locker

Locker fixing method

23