APPLIES TO:
Vivid i/q BT10.2.0 systems
PROBLEM:
ON BT10 systems, the CardioLab data, workflow context box does not appear (it has been inserted in
Figure 1-1 below).
"CardioLab
data,
CardioLab data, workflow workflow"
- On BT10
systems this
does not appear
Figure 1-1 "CardioLab data, workflow" context box shown above as an inset
SOLUTION:
See "work-around" described below.
1.) On the system console press Patient button and in the screen that displays, select Patient List.
The Search/Create Patient screen display - see Figure 1-2 .
2.) From the Dataflow drop-down list, select the required CardioLab data flow see Figure 1-2.
3.) Click the yellow folder on screen - see Figure 1-2 - and the Select Input context box opens - see
Figure 1-3.
4.) Select Worklist (shown in Figure 1-3) to open the Search/Create Patient list - see Figure 1-4.
Work list
6.) Click the Select Patient soft-key at the bottom of the monitor screen - Figure 1-4.
The system starts to process the CardioLab examination.
Technical Support
GEHC - Americas GEHC - Europe GEHC - Asia
REVISION HISTORY: