ACCURATE CALIBRATION
Client Name Date of Calibration 08/02/2011
CENTER AL KHOBAR
2. Environmental Condition :
Temperature: 25 ± 2 ºC
Humidity: 35 to 65 % RH
3. DETAILS OF STANDARD INSTRUMENT:
5. Calibration Results:
Simulation Method:
6.Remarks/Conclusion:
UNCERTAINTY BUDGET:
Sensitivit Uncertaint
Probability Standard y y
Source of Estimate Limits Distribution – Uncertaint coefficien contributio Degree of
Uncertainty s (µm) (µm) Type A or B y (µm) t n (µm) freedom
Xi xi ± Dx i - Factor u(x i) ci u i (y) ni
Normal
Repeatability
(Ua) - Type A 0.004 1 0.004 5
- Ö5
Standard
Instrument Normal
Uncertainty 6 3 3.000 1 3.000 ¥
(Ub1) Type B - K=2
Rectangular
Standard Inst.
Accu.(Ub2) 20.0 - - Type B 11.547 1 11.547 ¥
- Ö3
Standard Inst. Rectangular
Acc tn Slip 0.60 - - Type B 0.346 1 0.346 ¥
(Ub3) - Ö3
Standard Inst. Rectangular
Acc ts Slip 0.20 - - Type B 0.115 1 0.115 ¥
(Ub4) - Ö3
Thermal Rectangular
expansion 57.50 - - Type B 33.199 1 33.199 ¥
(Ub5) - Ö3
Combined
Uncertainity 35.28 ¥
Uc (Y)
Expanded
uncertainty k=2 70.56 ¥
(U)