Performa, Alpha RT, Alpha IQ, Alpha ST, Alpha III

Training Manual
This document is for training
use only. It is not intended to
replace any applicable service
documentation and should not
be used while conducting
service on any equipment.

Service Manual
October 2002 31687-4AN
PLACE FOR OPERATION
INSTRUCTIONS

PLACE FOR INSTALLATION


INSTRUCTIONS

FUNCTIONAL DESCRIPTION

PREVENTIVE MAINTENANCE

TROUBLESHOOTING AND
ERROR MESSAGES

REPAIR INSTRUCTIONS

SCHEMATICS
SERVICE MANUAL FOR ALPHA IQ/RT/ST/III AND PERFORMA

Notes/Cautions/Warnings
Update Information
Service Manual Contents

COPYRIGHT  2002 by Instrumentarium Imaging

Documentation, trademark and the software are copyrighted with all


rights reserved. Under the copyright laws the documentation may not be
copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form in whole or part, without the prior
written permission of Instrumentarium Imaging.

The original language of this manual is English.

Instrumentarium Imaging reserves the right to revise this publication from


time to time and to make changes in the content of it without obligation to
notify any person of such revision or changes.

Microsoft and Windows are trademarks of Microsoft Corporation in the


United States of America and other countries.

MANUFACTURED BY: Instrumentarium Imaging


P.O.Box 20
FIN-04301 Tuusula
FINLAND
Tel. +358-10-394 6500
Fax. +358-10-394 6501
e-mail: mammo@fi.instrumentarium.com
Internet: http://www.InstrumentariumImaging.com
Performa, Alpha IQ, RT, ST, III Service Manual

NOTES/CAUTIONS/WARNINGS

NOTE! ONLY ENGINEERS TRAINED BY INSTRUMENTARIUM ARE


ALLOWED TO SERVICE THE UNIT. PARTS INSTALLED BY
NONTRAINED PERSONNEL CARRY NO WARRANTY!

CAUTION! ALWAYS CHECK FOR PROPER SEATING ON ALL THE


CONNECTORS LOCATED IN THE AREAS YOU HAVE WORKED.
EQUIPMENT DAMAGE CAN RESULT IF POWER IS APPLIED OR
EXPOSURES MADE WITH IMPROPERLY SEATED CONNECTORS!

WARNING! ALWAYS INSTALL THE PROPER FUSES INTO THE UNIT.


FAILURE TO DO SO JEOPARDIZES PATIENT, OPERATOR AND
EQUIPMENT SAFETY. (SEE TABLE IN REPAIR SECTION)!

WARNING! THIS UNIT HAS FLOATING GROUNDS. ALWAYS MAKE ALL


SCOPE MEASUREMENTS DIFFERENTIALLY AND ALWAYS BE
CAREFUL WHEN SERVICING THE UNIT!

CAUTION! WHEN HANDLING CIRCUIT BOARDS, ALWAYS UTILIZE


GROUND WRIST STRAPS OR GROUNDING KITS TO PREVENT
DAMAGE TO THE "MOS" TYPE CHIPS. ALWAYS KEEP CIRCUIT
BOARDS IN A STATIC PROOF BAG.

WARNING! ALWAYS DISCONNECT THE UNIT FROM LINE SUPPLY BY


REMOVING THE PLUG FROM WALL OUTLET BEFORE OPENING
THE COVERS.

WARNING! THE X-RAY TUBE UTILIZES A BERYLLIUM WINDOW. DO NOT


PERFORM ANY OPERATIONS ON BERYLLIUM WHICH PRODUCE
DUST OR FUMES, SUCH AS GRINDING, GRIT BLASTING OR ACID
CLEANING. BERYLLIUM DUST OR FUMES ARE HIGHLY TOXIC
AND BREATHING THEM CAN RESULT IN SERIOUS PERSONAL
INJURY OR DEATH.

WARNING! THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND


OPERATOR UNLESS SAFE EXPOSURE FACTORS AND
INSTRUCTIONS ARE OBSERVED!

31687 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

UPDATE INFORMATION

From time to time, update sheets will be issued to bring your service manual
up-to-date. For ease of updating, the new information will be page for page, or
section for section, replaceable. You will also get a new update table so you
can check the contents of your manual to verify you have the latest
information. You can also check what has been changed from the previous
revision.

Version: Manual date: Details revisited:


31687-4AN October 2002 Software block diagrams
7.31/7.71
Revision levels of schematics
removed
Address information updated
31687-4AM May 2000 33892 ECS driver board ver.
7.30
ECS Troubleshooting
Measurement of the feedback
circuitry
31687-4AL January 2000 Section 7 / Schematics
Software block diagram 7.30
31687-4AK May 1998 Section 7 / Schematics
32305 AMD board ver. 3.0

32220 Inverter board ver. 1.3

32245 CPU board ver. 1.3

33270 SMD AEC board ver 1.1

33035 C&Z driver board ver 1.4

Software block diagram 7.29

October 2002 31687


Performa, Alpha IQ, RT, ST, III Service Manual

31687-4AJ November 1996 Section 7 / Schematics


Main wiring diagram

Casopto board OPTIPOINT


33280
AEC Display board
OPTIPOINT 33275
SMD AEC board OPTIPOINT
33270
5V Power Supply board (Only
ST) 32962
Sensor board OPTIPOINT
32735
31687-4AI October 1995 Details
31687-4AH February 1995 C&Z Drive Board (33035)
31687-4AG December 1994 SW 7.xx, New service program
menu
31687-4AF May - June 1994 New type generator
31687-4AE December 1993 SW 5.09, INF-mode, Print mode
31687-4AD February 1993 Details
31687-4AC February 1993 Details
31687-4AB March 1992 Section 7 (Bucky control)
31687-4AA March 1992 Section 7 (Pulse, Base, Power
boards)
31687-4A March 1992 New manual

31687 October 2002


Performa, Alpha IQ, RT, ST, III Service Manual

TABLE OF CONTENTS
3 FUNCTIONAL DESCRIPTION .............................................................................3-1
3.1 SOFTWARE BLOCK DIAGRAM..............................................................3-1
3.2 KEYBOARD FUNCTIONS .........................................................................3-1
3.2.1 Main modes (nor-, SER- and Pr -mode)..............................................3-1
3.2.2 Program (Pr) mode ..............................................................................3-3
3.2.3 Service program menu.........................................................................3-4
3.2.3.1 SETUP-mode (SUP and Par programs; see 3.1) .....................3-7
3.2.3.2 Calibrations (see section 7 software block diagram)...............3-8
3.2.3.3 Diagnostics (see 3.1) .............................................................3-10
3.2.3.4 Integrated Quality Control History........................................3-10
3.2.3.5 AEC adjustment parameters ..................................................3-11

3.3 USING THE SERIAL PORT IN FOR COMMUNICATION ................3-15


3.3.1 The Cable...........................................................................................3-15
3.3.2 Communication protocol ...................................................................3-16
3.3.3 Data formats ......................................................................................3-16

3.4 THEORY OF OPERATIONS....................................................................3-19


3.4.1 Process controller ..............................................................................3-19
3.4.1.1 The CPU BOARD .................................................................3-19
3.4.2 User interface.....................................................................................3-21
3.4.2.1 Switch and microswitch reading............................................3-21
3.4.2.2 Collimator lamp operation.....................................................3-21
3.4.2.3 Cassette sensing CASOPTO BOARD...................................3-22
3.4.2.4 Cassette holder sensing CSENSE BOARD...........................3-22
3.4.3 X-RAY GENERATION AND CONTROL ......................................3-23
3.4.3.1 INVERTER BOARD ............................................................3-24
3.4.3.2 The high voltage unit.............................................................3-27
3.4.3.3 mA generation timing............................................................3-28
3.4.3.4 FILAMENT CONTROL BOARD ........................................3-28
3.4.3.5 Anode motor drive AMD BOARD .......................................3-32
3.4.3.6 Mains transformer .................................................................3-32
3.4.3.7 Energy storage capacitor .......................................................3-32
3.4.3.8 Motor phase shift capacitor ...................................................3-32

31687 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.3.9 Tube housing assembly..........................................................3-32


3.4.3.10 X-ray tube bias circuit (M101G) .........................................3-32
3.4.4 Motor control .....................................................................................3-34
3.4.4.1 Motor control - general..........................................................3-34
3.4.4.2 C&Z DRIVER BOARD ........................................................3-34
3.4.4.3 Back up release ......................................................................3-36
3.4.4.4 Anode motor ..........................................................................3-37
3.4.4.5 Bucky motor ..........................................................................3-37
3.4.5 Automatic exposure control (AEC) ...................................................3-38
3.4.5.1 AEC - general ........................................................................3-38
3.4.5.2 The Detector ..........................................................................3-38
3.4.5.3 AEC BOARD ........................................................................3-38
3.4.5.4 The CPU ................................................................................3-39

3.5 DESCRIPTION OF CONTROL ELECTRONICS..................................3-40


3.5.1 Mains transformer and SUPPLY board.............................................3-40
3.5.2 CPU board .........................................................................................3-40
3.5.3 DISPLAY BOARD ...........................................................................3-42
3.5.4 C&Z-DRIVER BOARD....................................................................3-42
3.5.5 CD-BOARD ......................................................................................3-43
3.5.6 C-ARM CONTROL BOARD ...........................................................3-43
3.5.7 C-ARM DISPLAY BOARD .............................................................3-44
3.5.8 ECS DRIVER BOARD .....................................................................3-44

4 PREVENTIVE MAINTENANCE ...........................................................................4-1


4.1 X-RAY TUBE PERFORMANCE CHECK ................................................4-1
4.1.1 Half-value layer (HVL) measurement .................................................4-1
4.1.2 Tube efficiency check on the cassette holder ......................................4-2
4.1.3 Focal spot measurement procedure .....................................................4-3
4.1.4 Image quality evaluation procedure.....................................................4-5
4.1.5 Measurement of the feedback circuitry ...............................................4-7

4.2 AUTOMATIC EXPOSURE CONTROL EVALUATION........................4-8


4.2.1 Quick AEC test ....................................................................................4-8
4.2.3 Object thickness compensation ...........................................................4-8
4.2.2 kV compensation ...............................................................................4-10

October 2002 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4.3 MECHANICAL CHECKS .........................................................................4-11


4.3.1 Compression system ..........................................................................4-11
4.3.2 C-arm rotation brake..........................................................................4-11
4.3.3 Column and carriage..........................................................................4-11
4.3.4 Compression force checking .............................................................4-11
4.3.5 Testing the x-ray field illumination...................................................4-11

5 TROUBLESHOOTING & ERROR MESSAGES .................................................5-1


5.1 Error message chart ......................................................................................5-1
5.2 Useful information to have when calling for assistance .............................5-2
5.3 E.01 Error message explanation ..................................................................5-3
5.4 E.02 Error message explanation ..................................................................5-4
5.4.1 Testing of the new Performa, Alpha generator....................................5-5
5.4.2 Testing of the old Alpha generator ......................................................5-6
5.5 E.03 Error message explanation ..................................................................5-7
5.6 E.04 Error message explanation ..................................................................5-8
5.7 E.05 Error message explanation (software 5.08 and higher).....................5-9
5.8 E.06 ERROR MESSAGE EXPLANATION...............................................5-9
5.9 E.07 ERROR MESSAGE EXPLANATION...............................................5-9
5.10 E.08 ERROR MESSAGE EXPLANATION.............................................5-10
5.11 E.09 Error message explanation (PERFORMA, RT/IQ) ........................5-10
5.12 E.10 Error message explanation ................................................................5-10
5.13 E.11 Error message explanation ................................................................5-11
5.14 E.12 Error message explanation ................................................................5-11
5.15 E.r1 Error message explanation.................................................................5-11
5.16 E.r2 ERROR MESSAGE EXPLANATION .............................................5-11
5.17 E.r3 Error message explanation.................................................................5-11
5.18 E.r4 Error message explanation.................................................................5-12
5.19 E.r5 Error message explanation.................................................................5-12
5.20 S.Lo Error message explanation ................................................................5-12
5.21 C.Br Error message explanation (Software 7.27 and up)........................5-12
5.22 C.bu Error message explanation................................................................5-12
5.23 C.CA Error message explanation ..............................................................5-13
5.24 C.Ch Error message explanation ...............................................................5-13
5.25 C.Cn Error message explanation ...............................................................5-14
5.26 C.Co Error message explanation ...............................................................5-14
5.27 C.ES Error message explanation ...............................................................5-14
5.28 C.PS Error message explanation................................................................5-14
5.29 C.dE Error message explanation ...............................................................5-15
5.30 O.FL Error message explanation...............................................................5-15
5.31 R.EL Error message explanation...............................................................5-15
5.32 R.ES Error message explanation ...............................................................5-16
5.33 (U.FL Error message explanation) not used .............................................5-16
5.34 S.YS (Software 7.27 and up).......................................................................5-16
5.35 t.OF (Software 7.27 and up) .......................................................................5-16
5.36 E.AF (Software 7.27 and up) ......................................................................5-17
5.37 No power indications...................................................................................5-17

31687 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.38 AEC Inoperative ..........................................................................................5-17


5.39 Software diagnostic mode ...........................................................................5-18
5.39.1 Entering Diagnostic Mode.................................................................5-18
5.39.2 Error readout......................................................................................5-18
5.39.3 Switch and signal status display ........................................................5-19

5.40 SUPPLY VOLTAGES ................................................................................5-20


5.41 Grid lines ......................................................................................................5-21
5.42 ECS Trouble shooting .................................................................................5-22

6 REPAIR INSTRUCTIONS ......................................................................................6-1


6.1 FUSE RATING / LOCATION CHART......................................................6-1
6.2 COLUMN REPAIR.......................................................................................6-2
6.2.1 Tilt down the column...........................................................................6-2
6.2.2 Replacing the floor plate......................................................................6-2
6.2.3 Replacing the Z-motor.........................................................................6-3
6.2.4 Replacing the column cap....................................................................6-4
6.2.5 Replacing the mechanical stopper .......................................................6-6
6.2.6 Replacing the wires..............................................................................6-6
6.2.7 Replacing the counterweight ...............................................................6-7

6.3 POWER UNIT REPAIR...............................................................................6-8


6.3.1 Removal and assembly of the rear cover.............................................6-8
6.3.2 Replacing the POWER / INVERTER board .......................................6-8
6.3.3 Replacing other parts ...........................................................................6-8

6.4 CARRIAGE REPAIR ...................................................................................6-9


6.4.1 Removal and assembly of the side covers ...........................................6-9
6.4.2 Replacing the mains transformers .......................................................6-9
6.4.3 Replacing the boards on the electronic plate .....................................6-11
6.4.4 Replacing the SUPPLY board ...........................................................6-11
6.4.5 Replacing the magnetic brake and its assembly ................................6-12
6.4.6 Replacing the MGF DISPLAY board................................................6-15
6.5 C-ARM REPAIR .........................................................................................6-16
6.5.1 Removal and assembly of the C-arm covers .....................................6-16
6.5.2 Replacing the handgrips ....................................................................6-16
6.5.3 Replacing the brake switch................................................................6-16
6.5.4 Replacing the vertical drive control switch .......................................6-16

October 2002 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.5.5 Replacing the x-ray tube....................................................................6-19


6.5.6 Replacing the high voltage unit .........................................................6-19
6.5.7 Replacing the AMD board.................................................................6-19
6.5.8 Replacing the compression motor .....................................................6-20
6.5.9 Replacing the compression system....................................................6-20
6.5.10 Replacing the compression thickness measurement potentiometer ..6-20
6.5.11 Replacing the compression force measurement system (str. gages) .6-20
6.5.12 Replacing the C-arm angulation measurement system (trim pot) .....6-20
6.5.13 Replacing the cassette holder base cover ..........................................6-23
6.5.14 Replacing the cassette holder base ....................................................6-23
6.5.15 Replacing the collimator....................................................................6-25
6.5.16 Replacing the light field bulb ............................................................6-25
6.5.17 Replacing the beam filter mechanism ...............................................6-26
6.5.18 Replacing the compression display board .........................................6-26
6.5.19 Replacing the detector display board.................................................6-26
6.5.20 ECS repair .........................................................................................6-27

7 SCHEMATICS ..........................................................................................................7-1

31687 October 2002


Performa, Alpha IQ, RT, ST, III Service Manual

PERFORMA, ALPHA SERVICE MANUAL CONTENTS


3 FUNCTIONAL DESCRIPTION

3.1 SOFTWARE BLOCK DIAGRAM


PROGRAM VERSION 7.30 - See Section 7.23

3.2 KEYBOARD FUNCTIONS

3.2.1 Main modes (nor-, SER- and Pr -mode)


Exposure and other functions are possible only in main modes: nor and, SEr
mode. When the power is switched on to the device, the program is in
nor(=user)-mode, which functions as explained in this manual. In
SEr(=service)-mode certain checks like the presence of the cassette holder,
cassette, and diaphragm are bypassed so that service personnel can test the
device without any delays. In service mode the kV and mAs display shows the
AEC integration time in ms. If the AUTOREL button is pressed the timer
display shows the exposure time and the kV display shows the kV (Normally
timer display shows SEr).

NOTE! A cooling time of 5 to 55 sec. between exposure is implemented also in the


service mode. This will protect the unit from not being overloaded during
testing.

31687 3-1 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

The main functions achieved from the panel are listed below. The field
"display" indicates what is seen in the display. '&' means simultaneous
pressing.

KEY FUNCTION DISPLAY


kV+ Increase kV kV
kV- Decrease kV kV
mAs+ Increase mAs, not in AEC, go to IQC mAs
mAs- Decrease mAs, not in AEC, go to IQC mAs
DEN+ Increase exposure time by 10% DEN
DEN- Decrease exposure time by 10% DEN
AUTOREL Set automatic release on/off AUTOREL led
In SEr-mode display exp.time TIMER
AUTOEXP Switch between manual and AUTO EXP led,
leds for a,b,c
AEC-modes

DEN-& kV- Switch between SEr- and nor-mode TIMER 'Ser'


Switch from AEC- to nor-mode
DEN+ & - Enter pr- from nor - mode TIMER 'Pr'
Enter service program menu from SEr- mAs 'Pr' &
DEN+ & -
mode kV '51'
Return to service mode from program TIMER 'Ser'
DEN + & -
menu

October 2002 3-2 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.2.2 Program (Pr) mode


This mode can be entered from nor- and SEr-mode. Usually this is done when
the unit is installed to adjust exposure parameters to the conditions at the site.
The preset kV can be programmed in semi AEC mode from 23 to 35 kV for
bucky and cassette tunnel, and 25-35 for magnification. The film/screen
combinations (FI) are:

mAs-display film cassette/screen


FI1 Min - RE Min - R (Kodak/Kodak)
FI2 Microvision Min - R (Dupont/Kodak)
FI3 Min - RE Min - R/Medium
FI4 AGFA MR3+ AGFA MR Detail
FI5 Fuji M-1 Fuji MI-MA
HI-Mammo M1
FI6 Konica AGFA MR Detail

In manual mode default kVs can be programmed from 20 to 35 kV.

Pressing AUTOREL sends all the data for all exposures after last RESET to
serial port (RS-232). The port can be connected to a printer. Pressing
AUTOEXP enters the AEC stations, which can be programmed with the key
functions as described below. (In software versions 5.06 and below)

KEY FUNCTION DISPLAY


kV+ Increase preset kV kV
kV- Decrease preset kV kV
mAs+ Set film/ screen combination/ mAs
mAs in manual mode
mAs- Set film/ screen combination/ mAs
mAs in manual mode
DEN+ Increase internal density DEN
DEN- Decrease internal density DEN
DEN+ & - Return to main program all
AUTOREL Send last exposure data through RS READY-led
port
AUTOEXP Change from AEC channel a -> b -> c corresponding led

31687 3-3 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.2.3 Service program menu


Service program menu contains special programs for system setups, calibration
and diagnostics. The menu can be entered from service mode by pressing
Den+ & Den- buttons simultaneously for approximately one second.

In the kV display you will see the program number and in the mAs display the
program name. You can scroll the available programs with kV+ and kV-
buttons. To enter the desired program press the Autoexp button.

Return to the menu from the special programs with Autoexp button. To leave
the menu and enter the service mode press density+ and density- buttons
simultaneously. To resume the normal operation (nor-mode) press kV-and
density- buttons simultaneously for 3 seconds.

List of special programs:

Program # Program name Description


51 Pr Density and default value programming
52 SUP Set up parameter programming
53 Par Performa, Alpha parameter programming
54* CAn* C-arm angle display calibration
56* Cfo* Compression force display calibration
56* Cth* Compression thickness display calibration
57 Ano Anode run test
58 CEC Check error counters
59 Cin Check inputs
60 FoC Focus measurement mode
61 bUC Bucky measurement mode
62 Inf Print Performa, Alpha information
63* iqc Integrated Quality Control Parameters
64* AEC AEC Adjustment Parameters
65* tAr AEC Target time Adjustment

* Performa, Alpha RT/IQ only

October 2002 3-4 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Setup parameters:

The unit setup parameters can be altered in two programs:

In the setup section (program #52 SUP):

A Compression auto-release time


L Label type
H Label text
i i/o device
F Default filter selection

In the unit parameter section (program #53 Par)

S Serial number of the unit


C Country code
r Tube Efficiency mo/rh (mR/mAs)
b Anode braking mode
t Mode (Performa/III/ST/RT/IQ)
I Collimator select
d C-arm Control board select

The calibration programs:

The various measurements and displays can be calibrated in calibration


programs:

#54 C.An C-arm angulation display calibration


#55 C.fo Compression force display calibration
#56 C.th Compression thickness display calibration
#57 A.no Anode rotation program

The diagnostic programs:

The following programs are for system diagnostics:

#58 CEC Check error counters


#59 Cin Check CPU inputs
#60 FoC Focus measurement mode
#61 bUC Bucky measurement mode
#62 Inf System setup print out

31687 3-5 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

The Integrated Quality Control program (Software version 7.27 and up):

#63 iqc Integrated Quality Control Program

The AEC Adjustment and parameters programs (Software version 7.27


and up):

#64 AEC AEC parameters


#65 tAr AEC Target Time Adjustment Program

October 2002 3-6 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.2.3.1 SETUP-mode (SUP and Par programs; see 3.1)


Factory installation data (can also be set in the field) has 12 alterable items.
Items to be set up are:

parameter range default description


DENSITY- value
display
Par 1 S 0-9999 Serial number of the unit

Par 2 C 9-16 10 Country code


Par 3 r 0-25.5 - mR/mAs output @25kV
Separate for Mo/Rh
Par 4 b 0-1 0 Anode brake mode (1=on/0=off)
Par 5 t 1-5 4 Mode, (1=AlphaIII, 2=Performa/RT/IQ,
(Full-Man), 3=Performa/RT/IQ(Semi-
Man), 4=Performa/RT/IQ(Full-Semi-Man),
5=Performa/RT/IQ(Semi-Full-Man)).
Par 6 I 0-1 1 Collimator select (1= interlock, 0= non-
interlock)
Par 7 d 0-1 0 0=C-arm control board (32310), 1=CD-
board (32010)
SUP 1 A 0-25 6 Auto release time, (seconds)
SUP 2 L 0-5 0 Label type (0= print off)
SUP 3 H 0-60/255 32 Label programming mode
60 ascii chars
SUP 4 I 0-2 0 Serial port mode, (0=no I/O device,
1=printer(1200bd),2=PC(9600bd))
SUP 5 F 0-3 1 Default beam filter programming,
(Performa/ (0=Rhodium, 1=Molybden, 2=previously
RT/IQ) used, 3= "autofilter": use mAs+/- keys to
select the compression distance where filter
0-1 (ST)
is automatically changed

NOTE! Reset the CPU after changing the parameters.

NOTE! Items are updated to EEPROM when kV+, kV- or AUTOEXP-button is


pressed.

Changing the parameters:

Enter the special program menu as described previously and select program
#52 SUP or #53 Par with kV+ and kV- buttons. Press Autoexp button to enter
the desired program.

31687 3-7 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

FUNCTION DISPLAY AFF.


KEY
kV+ &kV- Scrolls the parameters to be set DEN
DEN+ Increase parameter value kV& mAs
mAs+ ONLY IN "h" Next character kV & mAs
position
DEN- Decrease parameter value kV& mAs
mAs- ONLY IN "h" Previous kV & mAs
character position
Autoexp Switch back to program menu

Press Autoexp to leave the program and return to the special program menu.

To store the new values press Density + and Density - simultaneously until
service mode is entered. Press kV- and Density- until normal mode is entered.
Cut the power from the mains switch for few seconds.

3.2.3.2 Calibrations (see section 7 software block diagram)

C-arm angle display calibration

The c-arm angle display can be calibrated with special program #54 CAn.

Enter the special program menu as described previously and select program
#54 CAn with kV+ and kV- buttons. Press Autoexp button to enter this
program.

The timer display will show CAn and kV and mAs displays will show to 00.
Make sure that the c-arm is in 0 degree angle. Press Autorel; the 0 degree
angle is now programmed. The control unit shows to 180. Turn the c-arm to
180 degree angle. If you want to use another angle for calibration you can
change the display with density +/- buttons. (You can abort the programming
and return to the program menu with Autoexp button).

Press Autorel. If the calibration was successful the mAs display will show
PAS and you are returned to the program menu. If the calibration did not
succeed you will be returned to the beginning and "to 00" will displayed again.

Compression force display calibration

The compression force display can be calibrated with special program #55 Cfo.

NOTE! Make sure the C-arm angle display is calibrated before calibrating the
compression force display (Pr #54 Can). Also the offset voltage on C-arm
control board must be 500mV when there is no stress on compression
paddle.

October 2002 3-8 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Enter the special program menu as described previously and select program
#55 Cfo with kV+ and kV- buttons. Press Autoexp button to enter this
program.

The timer display will show Cfo and kV and mAs displays will show to 50.
Make a compression of 50 N. Press Autorel; the 50 N level is now
programmed. The control unit shows to 200. Make a compression of 200 N. If
you have a different level of compression you can change the display with
density +/- buttons. (You can abort the programming and return to the
program menu with Autoexp button).

Press Autorel. If the calibration was successful the mAs display will show
PAS and you are returned to the program menu. If the calibration did not
succeed you will be returned to the beginning and to 50 will displayed again.

Compression thickness display calibration

The compression thickness display can be calibrated with special program #56
Cth. Enter the special program menu as described previously and select
program #56 Cth with kV+ and kV- buttons. Press Autoexp button to enter
this program.

The timer display will show Cth and kV and mAs displays will show to 00.
Drive the compression paddle completely down against the bucky. Press
Autorel; the 0.0 cm level is now programmed. The control unit shows to 60.
Place 60 mm acrylic on the bucky and drive the compression paddle down. If
you have 50 mm acrylic you can change the display with density +/- buttons.
(You can abort the programming and return to the program menu with
Autoexp button).

Press Autorel. If the calibration was successful the mAs display will show
PAS and you are returned to the program menu. If the calibration did not
succeed you will be returned to the beginning and to 00 will displayed again.

Anode run voltage calibration

The anode run voltage can be adjusted with special program #57 Ano.

Enter the special program menu as described previously and select program
#57 Ano with kV+ and kV- buttons. Press Autoexp button to enter this
program.You can now run the anode motor by pressing the exposure button.
The Autorel button can be used to set the anode brake on/off.

31687 3-9 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.2.3.3 Diagnostics (see 3.1)


The programs #58 -#62 are for system diagnosis.

In Pr #58 CEC the following counters can be checked:


# exposures, E01, E02, E03, E04, E06, E07,E08, E09, E10, E11, E12, S.Lo,
Er1, Er2, Er3, Er4, Er5, EAF, SYS and Err (last error).

FUNCTION DISPLAY AFF.


KEY
AUTOEXP Enter or quit error count
program
kV+/- Scroll counters
mAs+ kV and mAs
Reset the displayed counter

In #59 Cin program all the input-ports can be examined in the input-ports
check routine and the following internal counters for errors can be examined
and RESET in the error- exposure counter routine:

FUNCTION DISPLAY AFF.


KEY
AUTOEXP Enter input-ports check routine Lo -> Hi = line ok
AUTOEXP Return to program menu
DEN+ & - Returns to SEr-mode timer shows"SEr

For information on #60 FoC focus measurement program refer to section 4.1.3
Focal spot measurement procedure.

#62 INF program prints all preprogrammed default values and all
programmable tables. THIS DOCUMENT SHOULD BE FILED AND USED
AS A BASE ADJUSTMENT DOCUMENT.

3.2.3.4 Integrated Quality Control History


In the program #63 iqc, 20 latest IQC correction coefficients can be seen.
Scroll with kV+ or kV- . The kV display shows the reference number and the
mAs display shows the coefficient. IQC-coefficient is calculated from a
formula:

New IQC-coefficient = Old IQC-coefficient * (Wanted optical


density/measured optical density)

Wanted optical density is defined in the program #64 AEC parameter d. In


IQC mode the filter change is not allowed. Filter selection is made in the
program #52 SUP parameter F (If F=2 then iqc used filter is molybdenum).

The allowed change of the coefficient is ±40%, except Germany where ±25%.
If the change is bigger a C.PS error code is seen.

October 2002 3-10 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.2.3.5 AEC adjustment parameters


Functional description of AEC full automatic mode

The initial kV is automatically selected based on the compressed breast


thickness. In the beginning of the exposure the first estimation of the exposure
time at initial kV is made by measuring the radiation coming through the
breast. Based on this, the kV is automatically changed up or down, if
necessary, to meet the target time. If the kV is changed, a second estimation
of the exposure time at this kV is made by measuring the radiation and the kV
is automatically changed again, if necessary, to meet the target time. After
the exposure the final kV is displayed.

The Auto kV performance can be optimized and configured according to the


user preferences. The configurable items are the initial kV (initial kV versus
breast thickness table), amount of allowable kV change (Max kV change in
Pr-mode, enter from normal mode with Den+/Den- 3sec.) and desired exposure
time range (target time). In program #64 AEC, parameter I determines the
min/max values for target time. The target time is setup in program # 65 tAr
by giving two thicknesses and the respective desired exposure times. The
target times for other thicknesses are calculated with linear estimation. The
system selects the kV so that the desired target time can be reached. The
maximum allowable kV-change is more limiting, however, and thus a different
exposure time than the target time may result.

Thickness compensation adjustment

Software versions 7.27 and greater utilize a thickness compensation method


which compensates for variations in optical density (O.D) of the film.
The rotation point can be programmed. The amount of rotation and
compensation curve slopes below and above the rotation point can be
programmed separately.

Rotation point P

Rotation point is the exposure time where the O.D is adjusted correctly.
Thickness compensation is achieved by altering the exposure time around this
point. Default value is 1.00.

Amount of rotation r1 and r2

Parameters r1 and r2 determine the amount of compensation with exposure


times shorter or longer than the rotation point respectively. r1 determines the
amount of compensation at the point where exposure time is half the rotation
point. r2 determines the amount of compensation at exposure time twice the
rotation point.

The values of r1 and r2 can vary from 0.70 to 1.30. If the parameter is 1.30 it
means that exposure times are increased by 30%. If the parameter is 0.80 it
means that the exposure times are decreased by 20% etc. Default values are 1.0
for both.

31687 3-11 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Compensation curve slope parameters S1 and S2

Parameters S1 and S2 determine the curve slope for exposure times shorter or
longer than the rotation point respectively. S1 determines the slope at the point
where exposure time is lower than the rotation point. S2 determines the
amount of compensation at exposure time higher than the rotation point. Refer
to the curves below for effects of the S parameters. The rotation point is 1.0
seconds and the amount of rotation (r1 and r2) is 0.8, 0.9, 1.1 or 1.2 (= 80%,
90%, 110%, 120%). Default values are 200 for both.

Compensation curve slope parameters S1 and S2

Parameters S1 and S2 determine the curve slope for exposure times shorter or
longer than the rotation point respectively. S1 determines the slope at the
point where exposure time is lower than the rotation point. S2 determines the
amount of compensation at exposure time higher than the rotation point. Refer
to the curves below for effects of the S parameters. The rotation point is 1.0
seconds and the amount of rotation (r1 and r2) is 0.8, 0.9, 1.1 or 1.2 (= 80%,
90%, 110%, 120%). Default values are 200 for both.
1.4 1.4
S = 700 S = 700
S = 600
1.3 S = 500 1.3 S=0
S = 350 r2 = 1.2
S = 200
1.2 S = 100 1.2
r2 = 1.1 S = 50
1.1 S=0 1.1
P = 1.0sec P = 1.0sec
1 1

0.9 0.9
r2 = 0.9 S=0
S = 50
0.8 S = 100 0.8
S = 200
S = 350 r2 = 0.8 S=0
0.7 S = 500 0.7
S = 600
S = 700 S = 700
0.6 0.6
1 1.5 2 2.5 3 3.5 4 4.5 5 1 1.5 2 2.5 3 3.5 4 4.5 5
Exposure time / sec. Exposure time / sec.

Figure 3.1 Effect of parameter S2


S=0
1.4 1.4
S=0
S = 700
1.3 1.3

r1 = 1.2
1.2 1.2
S = 700 r1 = 1.1
1.1 1.1

P = 1.0sec P = 1.0sec
1 1

0.9 0.9
S = 700
r1 = 0.9
0.8 0.8
r1 = 0.8
0.7 0.7
S=0 S = 700
S=0
0.6 0.6
0 0.2 0.4 0.6 0.8 1 0 0.2 0.4 0.6 0.8 1
Exposure time / sec. Exposure time / sec.

Figure 3.2 Effect of parameter S1

October 2002 3-12 31687


Performa, Alpha IQ, RT, ST, III Service Manual

In programs #64 AEC and #65 tAr the AEC is adjusted.

KEY FUNCTION
kV+&- Scroll parameters
DEN+ & - To change the value of the parameter

Parameters to be set in program #64 AEC are:

parameter range default value description


d 0.1-2.55 1.3 Wanted optical density (iqc) (only in
Performa/RT/IQ)
h 0 /23-35 27 IQC-kV programming (only in
Performa/RT/IQ) 0=sensitometry based
iqc, 23-35=phantom based iqc.
I Small MIN/MAX (SF): Target time MIN/MAX values in full
focus: 0.50/5.00 automatic exposure (only in
0.30-10.00 MIN/MAX (LF): Performa/RT/IQ).
Large 0.50/3.50 NOTE: Separate target times for
focus: different Cassette holder/filter
0.30-5.00 combinations.
c - - Thickness compensation parameters.
Press autorel to scroll P,r1,r2,S1,S2.
P 0.30-2.55 1.0 Rotation point of compensation curve
(Exposure time in seconds where wanted
optical density was achieved).
r1 0.30-1.30 1.0 Compensation factor for exposure times
shorter than P.
r2 0.30-1.30 1.0 Compensation factor for exposure times
longer than P.
S1 0-700 200 Compensation parameter for very short
exp times (fine tuning).
0-700 200 Compensation parameter for very long
S2 exp times (fine tuning).
g 0.20-2.00 around 1 Software gain for optical density
adjustment. Replaces the hardware gain
trimmer.
E 23-35 See table 2.5 in Automatic kV default tables
installation programming (Only in Performa/RT/IQ)
manual

Parameters in program #65 tAr to be set are:

1. Target exposure time for 1.5 cm thick breast (factory default = 0.50 sec.).
2. Target exposure time for 9.5 cm thick breast (factory default = 2.40 sec.).

The software calculates the other thicknesses from these values.

31687 3-13 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

To adjust target times:

1. Go into program #65 tAr (press autoexp) You will see in density
display 0, in kV display the compressed thickness and in mAs display
the target time for the compressed thickness. You can check the target
times for different thicknesses now by driving the paddle.
2. Drive the paddle to 1.5 cm.
3. Press Autorel
4. Change the value with Den +/- buttons (Default 0.50 sec.).
5. Press Autorel to accept
6. Drive paddle to 9.5 cm
7. Change the value with Den +/- buttons (Default 2.40 sec.).
8. Press Autorel to accept. You will see PAS or ERR depending on the
new values if they were acceptable or not.

When to adjust target times:

AEC adjustment quick fix guide


Problem Probable Reason Action Comments
No contrast with Full AEC kVs with large Increase the Target time with Results in lower
large breasts breasts too high. large breast thickness. (Pr# 65 kV with large
tAr program) breasts
No contrast with full AEC kVs with small Increase the Target time with Results in lower
small breasts breasts too high. small breast thickness. (Pr# 65 kV with small
tAr program) breasts.
mAs too high, long Full AEC kVs low Decrease the Target time values Low kVs yield
exposure times long mAs
O.fl errors or Start kV too low and/or Change the full AEC kV table
maximum allowable kV (Pr# 64 AEC parameter E) or
"white images"
change too small. Increase the kV change value (Pr
mode)
EAf errors Poor connection between Check the cabling.
CPU and AEC boards.
Adjust the AEC board offset
Incorrectly adjusted AEC voltages.
offset voltages
EAf error at start up AEC board supply voltages Check the AEC board supply
problems voltages.
or autozeroing circuit does If you are using the 32315 AEC
not work at start up. board with software 7.27 or later,
you can remove the capacitor
C35 (22µF)
"white images" O.fl, U.fl or C.cn error code Check the bucky connection
(apply compression)

October 2002 3-14 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.3 USING THE SERIAL PORT IN FOR COMMUNICATION

3.3.1 The Cable


For the cable, there are two options because there are two ways to connect a
serial port of a PC. Some older models of personal computers use a 25-pin
D-connector for serial communications while normally 9-pin connector is
used. The cable for both options is illustrated below:

Alpha/Performa

Alpha/Performa

Choose either the 9-pin or pin 25-pin version.

To see that the communication works properly, you might want to add a pair of
LED's between GND and both data lines. This can be done with two 1000 ohm
resistors connected in series with the LEDs. The direction of leds is not
important because both (input and output) channels have voltage levels from
-12 to +12 volt.

31687 3-15 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.3.2 Communication protocol


The serial connection is an ordinary asynchronous receiver/transmitter.

The serial port is initialized in Performa, ALPHA III/ST/RT/IQ as follows:

1200 baud (9600 baud with 5.09 software and if Dataview is connected)
8 data bits
no parity (always 0)
1 stop bit

The effect of the 'no parity bit' and the stop bit is a total of two stop bits, so
receiving end must be configured correspondingly. The data byte looks the
following:

START 8-BIT DATA 2 STOP

The data can be received with a communications program (for example


Procomm) to PC and be printed afterwards on paper. This might be more
convenient than a printer with RS 2 because it is not a standard equipment.

3.3.3 Data formats


Performa, ALPHA III/ST/RT/IQ can send two different groups of data:

1) Diagnostic information about the system, like serial number, exposure


parameters and number of error situations. This information can be sent
to serial channel by pressing AEXP in the diA/inf-mode. This is meant
to be sent to either PC or printer. An example of this data is displayed
on the next page:

October 2002 3-16 31687


Performa, Alpha IQ, RT, ST, III Service Manual

INSTRUMENTARIUM Imaging, MGF 101/110 information, SW v.7.30, 01-Jan-2000


========================================================================
DEVICE#: 08224 ALPHA MODE: RT(4)
COUNTRY CODE: 010 ANODE BRAKING IS OFF
CARM BOARD: MCU
RADIATION EFFECIENCY 10.2(Mo), 08.1(Rh) mR/mAs at 25 kV
AUTOREL TIME(s): 006 SW gain for AEC-board: 1.00
LABEL TYPE: 000 I/O DEVICE: PRINTER
LABEL TEXT IS:

IQC multiplier: 1.00 IQC MODE: NO EXPOSURE


WANTED DENSITY: 1.30
DEFAULT FILTER: Molybdenium
EXPOSURE COUNTER: 0001276
ERROR CNT: E01 E02 E03 E04 E07 E08 E09 E10
00000 00000 00003 00000 00000 00000 00010 00005
E11 E12 SLO ER1 ER2 ER3 ER4 ER5 EAF SYS
00016 00013 00000 00000 00001 00000 00000 00000 00050 00000
PROGRAMMED VALUES: Full
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV f/s den kV f/s den kV f/s den
A= 00004 00001 00007 00004 00001 00007 00000 00004 00007
B= 00004 00001 00007 00004 00001 00007 00004 00001 00007
C= 00004 00001 00007 00004 00001 00007 00004 00001 00007
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV f/s den kV f/s den kV f/s den
A= 00004 00001 00007 00004 00001 00007 00004 00001 00007
B= 00004 00001 00007 00004 00001 00007 00004 00001 00007
C= 00004 00001 00007 00004 00001 00007 00004 00001 00007
PROGRAMMED VALUES: Semi
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV f/s den kV f/s den kV f/s den
A= 00027 00001 00007 00027 00001 00007 00027 00001 00007
B= 00027 00001 00007 00027 00001 00007 00027 00001 00007
C= 00027 00001 00007 00027 00001 00007 00027 00001 00007
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV f/s den kV f/s den kV f/s den
A= 00027 00001 00007 00027 00001 00007 00027 00001 00007
B= 00027 00001 00007 00027 00001 00007 00027 00001 00007
C= 00027 00001 00007 00027 00001 00007 00027 00001 00007
PROGRAMMED VALUES: Man
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV mAs kV mAs kV mAs
00027 00012 00027 00012 00028 00012
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV mAs kV mAs kV mAs
00027 00012 00027 00012 00027 00012
PRINTED VALUES OF DENSITY: 0..14 = -7..+7
FULL-AEC TARGET TIME variables + limits:
BUCKY: Mo: 0.25 +0.10 0.50 3.50
BUCKY: Rh: 0.25 +0.10 0.50 3.50
MAGN: Mo: 0.25 +0.30 0.50 5.00
MAGN: Rh: 0.25 +0.30 0.50 5.00
CONTACT:Mo: 0.20 +0.20 0.50 3.70
CONTACT:Rh: 0.15 +0.20 0.40 3.50
THICKNESS COMPENSATION variables:
BUCKY: Mo: 1.00 1.00 1.00 250 250
BUCKY: Rh: 1.00 1.00 1.00 250 250
MAGN: Mo: 1.00 1.00 1.00 250 250
MAGN: Rh: 1.00 1.00 1.00 250 250
CONTACT:Mo: 1.00 1.00 1.00 250 250
CONTACT:Rh: 1.00 1.00 1.00 250 250
FULL-AEC kV-TABLES:
mm: 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
BUCKY:
Mo: 23 23 23 23 23 23 23 23 24 24 25 26 26 27 28 29 30 31 32 33
Rh: 23 23 23 23 23 23 23 23 24 24 25 26 26 27 28 29 30 31 32 33
MAGN:
Mo: 25 25 25 25 25 25 25 25 26 26 26 27 27 28 28 28 29 29 30 30
Rh: 25 25 25 25 25 25 25 25 26 26 26 27 27 28 28 28 29 29 30 30
CONTACT:
Mo: 23 23 23 23 23 23 23 23 23 23 24 24 24 25 25 25 26 26 27 28
Rh: 23 23 23 23 23 23 23 23 23 23 24 24 24 25 25 25 26 26 27 28
EPROM CHECKSUM: Calc = 52255 Set = 52255

31687 3-17 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

2) After each exposure the exposure data is sent to serial port if the L-
parameter (in SETUP) is > 0. If the L-parameter (in SETUP) is 1 the
data is printed on one line, for example:

02693,01429,335,027,00234,7,7,1,100,080,150,001,Rh,0
0123, OK ,B,FULL

02693 Exposure counter


01429 integration time
335 exposure time
027 kV
00234 true mAs
7 programmed density
7 density
1 film/screen
100 IQC -multiplier
080 compression thickness
150 compression force
001 angle of C-Arm
Rh Filter
00123 dose
OK error code
B technique ( M / N )
FULL exposure mode (SEMI / MAN / IQC)

If the L-parameter (in SETUP) is > 1 the printout is:

B Rh 0.02s 28kV 002mAs +0 HOSPITALxNAMEx


6.5cm 200N 180^/R 0.06mGy CODExxxxxxxxxx
___/___/___ R1 12345 RADIOLOGISTxxx
_________________________ DEPARTMENTxxxx

Explanations:
B Bucky / Cassette tunnel / Magnification
Rh Rhodium / Molybdenum filter
+0 Density correction
28kV kV
002mAs mAs
0.02s Exposure time
6.5cm Breast thickness
200N Compression force
180^/R C-arm tilt angle, ^= degrees (), to the /Right or /Left
0.06mGy Average patient dose
___/___/___ Date is handwritten after printing the label
R1 Site of the detector (L1, L2, C1, C2, C3, C4, R1, R2)
12345 Exposure counter, up till 99999
_____________________ Patient name is handwritten after printing the label

October 2002 3-18 31687


Performa, Alpha IQ, RT, ST, III Service Manual

If the L-parameter is 2 the consecutive labels are printed 5 rows apart


If the L-parameter is 3 the consecutive labels are printed 3 rows apart
L=4 and L=5 include exposure information and are used for testing purposes
only.

3.4 THEORY OF OPERATIONS

3.4.1 Process controller

3.4.1.1 The CPU BOARD


The CPU board is the heart of Performa, ALPHA III/ST/RT/IQ. It controls
every function in the unit, except the ROTATIONAL BRAKE. The 8031
microprocessor on the CPU board controls kV GENERATION ,mA
GENERATION (via the FILAMENT CONTROL board) and the rest of the
functions through latches and buffers directly and uses a keyboard controller
(8279) to handle the control panel..

The I/O-control (latches and buffers) controls the vertical drive motor,
compression motor, bucky, anode motor, foot switches, microswitch reading,
and the AEC. The 8279 reads and decodes the control panel switches, as well
as drives the control panel display.

Exposure times in both manual and AEC mode are calculated and controlled
by software and the microprocessor.

31687 3-19 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

COMPRESSION

October 2002 3-20 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.4.2 User interface

3.4.2.1 Switch and microswitch reading


The control panel switches on the PERFORMA, ALPHA III/ST/RT/IQ are
configured in a row/column matrix (fig. 3.1) that is strobed and read by the
CPU board. D15 on the CPU board is an 8279, which is a keyboard controller.
It will strobe each column looking for a return on one of the rows, which it will
only find when a switch is pressed. For example, if the circled switch is closed,
the resulting timing chart will be as given in fig. 3.2.

S0 S1 S2 S7 S0

S1

R0 S2
S3
R1 S4
S5

R2 S6
S7
R0
R1
R2
R7 R3

This will let the CPU know that the switch S2/R0 is pressed. The software
defines that the switch is the kV-button and decreases the kVref on the PULSE
board.

The hand switch (exposure button) signal is fed through the CPU board to the
pulse board where it becomes an enable for the exposure circuitry before it is
fed back to the CPU board as signal EXP2 on connector X5 pin 9. This allows
the operator to stop the exposure in the event of a CPU board malfunction.

3.4.2.2 Collimator lamp operation


When the COMPRESSION DOWN switch is activated, in addition to the CPU
turning on the compression motor, it also sends a signal called PROJ RELAY,
and starts a software timer. The PROJ RELAY signal is fed to the SUPPLY
board where it activates relay K1 turning on the collimator lamp. The lamp is
shut off when the timer runs out ( about 8 sec.) or the hand switch is pressed
and the exposure sequence is started.

31687 3-21 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.2.3 Cassette sensing CASOPTO BOARD


Cassette sensing is accomplished by using two (one in optipoint) optical
sensors (fig. 3.3). Each sensor has two halves, an infrared source and an
infrared detector.

Figure 3.3. Cassette sensing

The cassette bottom forms the reflective surface needed. There is a green LED
which comes on to indicate that the cassette is being sensed and the CASNOR-
signal is sent to the CPU.

3.4.2.4 Cassette holder sensing CSENSE BOARD


The cassette holder sensor has four reed relays in it, three of them (CHL,CHR
and K4) are used to decode cassette holders for the CPU. The fourth read relay
(CHON) ensures the cassette holder is completely attached. The identification
is done with magnets attached to the holders.

CHL CHR For small cass. holder and small


Cas holder 1 1 bucky the K4 is also activated
Bucky 0 1
Mag box 1 0
Stereotact 0 0

0 = Not activated
1 = Activated

CSense Board:

October 2002 3-22 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.4.3 X-RAY GENERATION AND CONTROL

X-ray generator block diagram

The generator of the unit is a modern, constant power, high frequency,


feedback-controlled DC x-ray generator consisting of the following parts:

1. Inverter board
2. Filament Control board
3. High voltage unit
4. Anode motor drive (AMD) board
5. Mains transformer
6. Energy storage capacitor
7. Tube housing

Technical specifications are:

Input power: 3500 W ( 6500 VA) max


Output power: 2500 W max
Output voltage : 20 - 35 kV
Output current : 10 - 100 mA
Frequency: 20 - 90 kHz
Line voltage: 220/ 240 f 10% 50/60Hz
IRMS(rest): 1A
IRMS(exposure): 30 A
Ipeak: 55 A

31687 3-23 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.3.1 INVERTER BOARD


The INVERTER Board controls the high voltage (kV) generation. The kV
level is digitally set by the CPU board.

The Inverter Board contains a high voltage and a low voltage section. The high
voltage section is isolated from the low voltage section with pulse transformers
TF1, TF2 and relays K1,K2.

WARNING! Voltages in the high voltage part of the INVERTER Board are deadly.
The peak to peak voltage exceeds 700V in normal operation.

High voltage section:

The high voltage section consists of a MOSFET transistor H-bridge which


converts the rectified line voltage (+310 VDC) into high frequency current.
This current is taken to the HIGH VOLTAGE UNIT where it is transformed
into high voltage. The inductors L1, L2 and capacitor C8 act as parts of
resonant circuit which shapes the bridge current. The RMS value of the
current is controlled by the H-bridge's operating frequency. The MOSFET
driving pulses are generated at the low voltage section of the board and are
provided by the pulse transformers TF1 and TF2.

TF3 is a current transformer which monitors the bridge current. If the bridge
current exceeds the pre-adjusted current limit the control circuit in the low
voltage section turns the MOSFETs off.

Timing:

The line voltage is applied through the connector X1. When an exposure is
initiated, relays K1 and K2 provide the rest of the board with line voltage.
Relays K1 and K2 are activated with signals RG1 and RG2 respectively. K1
applies voltage to the bridge rectifier D3 through the current limiting resistor
R1 which allows controlled charging of the energy storage capacitor
(connected to X2). K2 is activated approximately 0.8 seconds later and
provides the circuit with full power. Finally, the activation of the EXPENA
signal (TP17) enables the H-bridge drive circuitry.

Low voltage section:

kV control:
The high voltage is regulated by adjusting the H-bridge's operating frequency.
This is done by comparing the feedback signal (HVFB) from the high voltage
unit with the reference from the CPU BOARD and taking the error signal to
the frequency modulator circuit (ICA1, ICA4, ICD1, ICD2). The output is
taken to ICA5 and ICA6 which drive the pulse transformers TF1 and TF2.

October 2002 3-24 31687


Performa, Alpha IQ, RT, ST, III Service Manual

EXPENA signal:
An exposure is initiated by the CPU BOARD by activation of the EXPENA
signal. This enables pulse transformer drivers (ICA5, ICA6) and allows the
kV reference signal to rise providing "soft start".

TUBEFAIL signal:
TUBEFAIL signal is activated if the kVFB signal does not reach the kV
reference. This typically occurs during a high voltage breakdown. A
SHUTDOWN for the kV control is provided. As the high voltage breakdown
is characteristic to x-ray tubes, and hence may occur during normal operation,
the exposure is continued with a new activation of the EXPENA signal. The
CPU Board allows three (3) tube failures per exposure before an error
message (E02) is displayed.

Test Switch:
Test switch S1 allows an exposure sequence to continue regardless of the
kVFB signal. In "TEST" position the TUBEFAIL signal is forced down .

Note! The Test Switch S1 should always be at "NORMAL" position whenever


the high voltage part of the INVERTER BOARD is energized. Failing to
do this may damage the INVERTER BOARD, HIGH VOLTAGE UNIT
and the X-RAY TUBE.

An exposure without X-rays:


It is possible to check the operation of kV control circuitry without X-rays.

1. Unplug the connector X1


2. Set the test switch S1 to "TEST" position
3 Unplug the connector X4 on the FILAMENT CONTROL Board
4. Set the test switch on FILAMENT CONTROL Board to "TEST"
position
5. Set the jumper X8 ("FBSEL") on the FILAMENT CONTROL
Board to position 1-2 (left hand position)
6. Initiate the exposure

The control circuitry is driving the MOSFETS with the minimum frequency
(app. 20kHz).

Back Up Timer:
The Back Up Timer is located on the FILAMENT CONTROL BOARD and it
disables both the +34REL voltage and the EXPENA signal providing a
SHUTDOWN approximately fifteen (15) seconds after the activation of the
EXPENA signal. This is a safety feature only and does not affect normal
operation.

31687 3-25 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

kV reference:
The level of the used kV is set by the CPU BOARD with signals DATA0 -
DATA7. This 8 bit digital word contains the kV reference . The kV reference
is converted into an analog signal in D/A converter ICA2 (TP24).

The analog kV reference (TP24) and the kV feedback (TP22) signals


correspond to actual kV as follows: kV = DVM read-out * 10 kV

For example, if the digital multimeter (DVM) read-out is 2.5 V the actual high
voltage is 25kV.

kV adjustment:
The calibration of the kV is done by adjusting the D/A converter's (ICA2)
reference voltage with trimmer potentiometer R63.

Note! The kV calibration is performed at the factory. The trimmer


potentiometer R63 is sealed . Manipulation of the potentiometer can be
carried out by authorized personnel only. Improper settings of the
potentiometer can damage the INVERTER BOARD, HIGH VOLTAGE
UNIT and the X-RAY TUBE

External indicators:
Connector X6 provides relay contacts for X-ray system status indication. X6
pins 1 and 2 are short circuited when the unit is on. X6 pins 3 and 4 are short
circuited when X-rays are being generated. The contacts are rated for 5A,
24VDC /240VAC.

Fuses:
High Voltage fuse in the INVERTER BOARD (F1) is to protect the board in
case of a short circuit.
-F1 20 AF /500V High voltage!

List of LED indicators

LED indicator Signal


H1 +310VDC
H2 EXPENA signal
H3 Bridge Current
H4-H5 Pulse A (MOSFET drive pulses)
H6-H7 Pulse B (MOSFET drive pulses)
H8 +15V
H9 +34V
H10 -15V
H11 +25V
H12 +34VREL

October 2002 3-26 31687


Performa, Alpha IQ, RT, ST, III Service Manual

List of Test Points


Test Point Signal
TP1 +310V GND High voltage!
TP2 +310 VDC High voltage!
TP3 High voltage!
TP4 High Voltage!
TP5-TP7 MOSFET gate High Voltage!
TP8-TP10 MOSFET gate High voltage!
TP11-TP13 MOSFET gate High voltage!
TP14-TP16 MOSFET gate High Voltage!
TP17 EXPENA signal
TP18 SHUTDOWN signal
TP19 GND
TP20 Operating frequency
TP21 Pulse A
TP22 kV Feedback
TP23 Pulse B
TP24 kV Reference
TP25 GND
TP26 +34V
TP27 +15V
TP28 GND
TP29 GND
TP30 -15V
TP31 -34V
TP32 +34VREL
TP33 +25V

Schematics, Layouts

Refer to the documents in section 7.

3.4.3.2 The high voltage unit


In the secondary of the high voltage transformer, there is a voltage doubler. As
one half of the inverter is conducting, secondary current flows through one set
of diodes, charging one of the 1.2 nF capacitors in the tank. When the other
half is conducting, the current flows through the other set of diodes and
charges the other 1.2 nF capacitor. Since the capacitors are connected in series,
the voltage across the capacitors is double the voltage of secondary. There is a
450 Mohm resistor connected to the high voltage output forming the upper leg
of the voltage division. On the Inverter board is the lower leg over which the
control circuit measures the voltage. The modulator compares this voltage to
that of a reference and changes the output frequency, if necessary, to maintain
the proper kV.

The output of the hV-unit is directly connected to the anode of the tube with a
high voltage cable. The unit also contains a rubber bottle because of the
expansion of oil volume due to rising temperature.

31687 3-27 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.3.3 mA generation timing


mA generation timing is controlled with four signals: FOCUS, PREHEAT,
PREH and EXT. FOCUS is a latch output which changes state depending on
whether the mag-box (small focal spot) or either the bucky or cassette tunnel
(large focal spot) is attached to the machine. It selects which mA feedback
resistor is used on the FILAMENT CONTROL board and which filament is
energized. PREHEAT becomes active 55 milliseconds after the hand switch is
depressed. It powers relay K2 on the FILAMENT CONTROL board and
supplies power to the filament transformer TR1. During the time between
PREHEAT and EXT, the pulse width modulator (ICA8) uses filament voltage
as its feedback voltage. The feedback voltage is gain is controlled with
trimmer potentiometers R28 and R96 to achieve proper tube current rise at the
beginning of an exposure. (See PREHEAT ADJUSTMENT in the Adjustments
& Alignments Section).

When EXT goes active, actual exposure begins. We are now no longer
interested in filament voltage; now, we must regulate tube current. To
accomplish this, the common of the filaments is connected to ground through
one of the feedback resistors (R77 or R78) on the FILAMENT board. This
signal called mA FEEDBACK (MAFB) is now fed to the pulse width
modulator by the analog switch (ICA10). Its output was switched from
filament voltage to MAFB with the signal PREH. The PREH signal is
activated approximately 7 ms after the EXT signal. The duty cycle of the
pulse width will vary depending on what is required to maintain proper mA.

3.4.3.4 FILAMENT CONTROL BOARD


The CPU board begins the mA generation cycle when it sees the EXP2 signal
from the FILAMENT CONTROL Board. The EXP2 is activated from the
exposure button. The control of the mA and kV are handled through the serial
to parallel converter / latch (ICD3). The latch is configured to have four major
functions:

1. Control/Timing
2. Serial output back to the CPU board
3. kV reference
4. mA reference

kV reference
The kV reference is converted into eight (8) parallel DATA bits which contain
the kV reference information for the INVERTER BOARD. The kV reference
occupies ICD3 pins 23 through 30.

mA reference
The mA reference is converted into eight (8) parallel DATA bits which contain
the mA reference information. The mA reference occupies ICD3 pins 11
through 18.

October 2002 3-28 31687


Performa, Alpha IQ, RT, ST, III Service Manual

mA control:
The mA control adjusts the X-ray tube's filament heating level to provide the
desired tube current (mA). This is done by regulating the filament voltage
with "FLYBACK" type pulse width modulated switch mode power supply (T1,
T2, TR1, D6, C12, ICA8). The control circuitry utilizes current mode control
which provides better stabilization and protects the semiconductor switches
against over current.

The operation consists of two consecutive sequences: the preheat sequence and
normal mA control sequence. The preheat sequence is used for heating of the
filament to the desired level prior to the exposure, in order to achieve optimum
mA rise at the start up. During the preheat sequence the control feedback
comes from the filament voltage, and during an exposure the mAFB signal is
used, providing the actual tube current monitoring.

Preheat adjustment:
The Preheat adjustment is done separately for large and small focus with
trimmer potentiometers R28 and R96 respectively. (See PREHEAT
ADJUSTMENT in the Adjustments & Alignments Section) The FOCUS
signal selects the corresponding trimmer with relay K3. During the preheat
adjustment measure the tube current from TP4.

EXT signal:
The EXT signal enables the kV generation.

FOCUS signal:
The FOCUS signal changes the mA feedback circuit according to the selected
X-ray tube focal spot size. It also changes the filament that is being powered.
When FOCUS signal is high (led H1) a small focus (0.1) is selected.

PREH signal:
The PREH signal changes the filament control mode from preheat to normal
exposure. The jumper X8 is used to disable the PREH signal during testing.

PREHREL signal:
A preheat sequence is initiated before each exposure (before activation of the
EXT signal) with the PREHREL signal. This activates the relay K2, providing
+34V input voltage for the filament power supply and the INVERTER
BOARD. The PREHREL signal enables the mA control circuit also.

mA feedback:
Feedback voltage to the filament power supply controller ICA8 is selected
with the analog switches ICA10. The source depends on the control signal
EXT as follows:

STATUS EXT (TP8) FEEDBACK SOURCE


preheat sequence passive (0V) Filament voltage
normal exposure active (+15V) mA feedback
stand by passive (0V) Filament voltage

31687 3-29 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Jumper X8:
The jumper X8 is used for keeping the filament control in preheat mode during
the whole exposure. This is intended for testing purposes only. The jumper
should normally short circuit pins 2 and 3 (right hand position) Note: Always
remember to leave the jumper in the right hand position after servicing the
unit.

mACHK signal
Comparator ICA6 monitors the mA feedback signal and activates the mAOk
signal when mA is present.

Buzzer
Buzzer (BZ1) is activated with signal RG2. It indicates that X-rays are being
generated. (Note: The buzzer is activated during an exposure without X-rays
also; see section 3.4.3.1:Timing)

C-arm angle measurement


The c-arm angle measurement is done with a potentiometer in the c-arm brake
assembly. The potentiometer gives a voltage proportional to the angle. This
voltage is taken to the 12 bit, serial mode analog to digital converter ICA12.
The control signals and data signal comes from the CPU board to connector
X12.

FB MISS signal
The FB MISS signal is activated if the feedback cable is not connected. This is
to protect the high voltage transformer in case the feedback connector is
accidentally left open. Activation of this signal disables the EXP2 signal to the
CPU board. This feature requires that the unit has cabling that short circuits
pins 1 and 2 in X4. If the unit does not have that kind of cabling you can use
the board if you install a jumper connector X11 that grounds the pin 2.

Back Up Timer:
The Back Up Timer disables both the K2 and the EXPENA signal for the
INVERTER Board providing a SHUTDOWN approximately fifteen (15)
seconds after the activation of the EXT signal. This is a safety feature only and
does not affect normal operation.

Test Switch:
The test switch S1 allows an exposure sequence to continue regardless of the
mAOK' signal which indicates that tube current is present.

X-ray tube bias


The filament is biased by having the tube current going trough a bias resistor (
0, 220 or 300 ohms).The value of the resistor is determined at the factory
separately for each tube and should therefore not be changed or modified.

October 2002 3-30 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Supply voltages:
The +15 V supply voltage is regulated from the +34 V with switching
regulator ICA11, and the -15 V is regulated from the -34V with linear
regulator ICA9.

Fuses: Fuses F1 and F2 protect the board in case of short circuits

- F1 6,3 AT (=slow), 6,3 x 32 mm


- F2 6,3 AT (=slow), 6,3 x 32 mm

List of LED LED


indicators indicator Signal

H1
H2 Filament voltage
H3 TubeFail signal (E02 error)
H4 PrehRel
H5 RG1
H6 RG2
H7 EXT
H8 +15V
H9 +34V
H10 -15V

List of Test Points Test Point Signal

TP1 Filament Voltage


TP2 MosFet gate
TP3 Filament Feedback
TP4 mA Feedback
TP5 PrehRel
TP6 RG1
TP7 RG2
TP8 EXT signal

TP9 +34V
TP10 +15V
TP11 -15V
TP12 GND
TP13 GND
TP14 -34V
TP15 mA Reference

Schematics, Layouts

Refer to the documents in section 7

31687 3-31 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.3.5 Anode motor drive AMD BOARD


The unit utilizes a rotating anode tube. The anode plate is accelerated with 230
VAC before the emission of electrons from cathode. During exposure, stator
needs a certain running voltage (40-100 VAC), and after exposure the anode
plate can be braked.

All this is done with the phase control of a triac (V1). The control element is
Telefunken TEA1007 (A7). Acceleration is achieved directly with mains
voltage. During exposure, the firing angle of the triac is set to a level that
corresponds the required RMS voltage. Braking is done by half wave rectified
mains voltage. Acceleration and braking times are approx. 1 second. Safety
features are also taken into account by monitoring the current of the triac.
Should errors be detected, the microprocessor does not allow exposures. The
exchange of information between the microprocessor and this board is
optoisolated because the anode motor drive board is floating on mains voltage.

3.4.3.6 Mains transformer


The power and control side of the unit uses only one 50/60 Hz transformer. It
is a two slot 120 VA EI transformer. The first slot contains primary 220 V and
secondary 15 V for the protection circuit of the power transistors ( this is done
because the protection circuit is floating on mains voltage ), the second slot
contains the actual secondary 2 x 24 V for AUX POWER. A grounded static
shield is situated between these slots.

3.4.3.7 Energy storage capacitor


The energy storage capacitor is used for smoothing the rectified mains voltage
fed to the POWER board. It is an aluminum electrolytic capacitor 2200 uF /
400 V with a 4K7 /11W bleeder resistor.

3.4.3.8 Motor phase shift capacitor


The motor phase shift capacitor is needed to rotate the anode plate. Its
capacitance is 40 uF / 450 V.

3.4.3.9 Tube housing assembly


Contains expansion systems and thermal switches and tube insert.

3.4.3.10 X-ray tube bias circuit (M101G)


The filament of the M101G X-ray tube is biased by having the tube current
going trough a bias resistor (0, 220 or 300 ohms).The value of the resistor is
determined at the factory separately for each tube and should therefore not be
changed or modified.

October 2002 3-32 31687


Performa, Alpha IQ, RT, ST, III Service Manual

31687 3-33 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.4 Motor control

3.4.4.1 Motor control - general


As you can see from the block diagram to the left, all interfacing between the
microprocessor and the motors is handled by 8255 peripheral controllers (D8
and D16) on the CPU board. The one exception to this is BHOME signal,
which goes directly to the 8031.

The microprocessor writes to and reads from the 8255 when its chip select (pin
6) is active. Information is transferred back and forth on the data bus. Because
of this, the outputs of the 8255 are the effective origin of the signals, and the
inputs to it are effectively the final destination of all the signals.

3.4.4.2 C&Z DRIVER BOARD


This board controls the compression and vertical drive motors (also called C-
and Z movements). It also provides regulated +24V dc-voltage to the bucky.

Both C- and Z- motors are +24V dc-motors which are driven with pulse width
modulated (PWM) power supplies that can provide variable voltage control
thus enabling a rough speed control. Each drive circuit consists of control
circuit and two power transistors. The output voltage consists of +34 Volt
pulses whose width determines the RMS voltage of the motor. This voltage is
proportional to the reference voltage applied to the control circuitry. The
operating frequency for both drives is app. 30kHz.

Carriage motor (Z-motor)


This is also called vertical drive motor. The vertical drive has two speeds.
The movement is started with slow speed that allows accurate positioning of
the carriage. After app. 2 seconds the faster speed is activated.

The CPU controls the vertical movement with two signals:

ZDIR signal:
This signal sets the status of the relay K3. It determines the polarity of Z-
motor voltage that is, the direction of the carriage.

ZRUN signal:
This signal activates the PWM circuit (ICA5) which applies voltage to the
motor. The RC circuit R60 and C21 determine the slow speed time interval.

Lower Limit, Upper Limit


These signals provide a way to limit the movement. When Lower/Upper Limit
signal is low only upwards/downwards direction is enabled. This is for future
purposes.

October 2002 3-34 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Compression motor (C-motor)


The compression motor has three different speeds: fixed high speed up,
trimmer potentiometer controlled high speed down and trimmer potentiometer
controlled low speed down. The purpose of these speeds is to provide "soft"
compression. This is done by reducing the speed of the compression paddle as
soon as the paddle reaches the breast.

The control of the compression is done by monitoring the compression motor's


current which gives us a rough idea of what the actual torque of the motor is.
The current of the motor is measured with resistor R29 and then compared to
two predetermined threshold levels. The first threshold level, speed threshold,
determines the force that the speed of the compression is reduced. Turn the
trimmer potentiometer R43 to set this level. The second threshold level,
current limit, determines the final compression force. Turn the trimmer
potentiometer R33 to set the compression force.

The CPU controls the compression movement with following signals:

CDIR signal:
This signal sets the status of the relay K1. It determines the polarity of the C-
motor voltage that is, the direction of the compression paddle.

CRUN signal:
This signal activates the PWM circuit (ICA4) which applies voltage to the
motor. The reference voltage is selected from three different speed references
with analog switch ICD5. The choice depends on the direction of the
compression paddle and the speed threshold level.

SPEED signal: (for future purposes)


This signal enables the CPU to reduce the compression speed regardless of the
speed threshold level.

CMAN signal:
This signal activates the relay K2 which connects diode D10 across the motor
terminals. The purpose of this diode is to prevent the motor from rotating when
the breast is under compression.

+24Vin Bucky:
Linear regulator ICA6 provides regulated +24V dc-voltage to bucky control
board.

31687 3-35 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Adjustment of the compression speeds:

1. Adjust the high speed compression to app. 2cm /1 seconds.


2. Turn trimmer R43 until only low speed is on.
3. Adjust the low speed compression down to app. 1cm /1 seconds
4. Adjust the trimmer pot R43 so that the paddle switches to low speed
when it reaches the breast. Make sure the paddle travels in high speed
(free air) at all angles. If not increase the speed threshold level.
5. Adjust the final compression force with trimmer R33. The compression
force limit can be adjusted by software setting in service program
(version 7.36 and up). In this case compression force with trimmer R33
is adjusted to be 25-27 kg (53-59lbs) and force stop has been set by
software.

List of LED:s: LED Signal


H1 +34Vdc
H2 +15Vdc
H3 +24Vdc for bucky
H4 C-motor voltage
H5 Z-motor voltage
H6 Upper Limit
H7 Lower Limit

List of testpoints: Test point Signal


TP1 +34Vdc
TP2 +15Vdc
TP3 +24Vdc for bucky
TP4 low speed reference
TP5 high speed reference
TP6 speed reference paddle up
TP7 vertical drive speed reference
TP8, TP9 C-motor voltage
TP10, TP11 Z-motor voltage
TP12 GND
TP13 GND

Schematics and layouts: Refer to section 7.

3.4.4.3 Back up release


The units are equipped with a compression back up feature. If the line voltage
suddenly disappears the compression paddle is driven up. This is implemented
with a back up transformer in the back of the unit. The back up transformer is
energized whenever the unit is connected to the line supply. This back up
transformer provides power for the relay K4 on the Filament Control board.
When this relay is powered the compression motor is driven normally. When
the relay is not activated the motor is driven upwards with the voltage in the
capacitors C20 -C31 on the Filament Control board. The status of the relay K4
is indicated with a LED next to the main switch.

October 2002 3-36 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.4.4.4 Anode motor


After the hand switch has been depressed, the microprocessor will signal the
8255 (D8) on the CPU board to output four signals that are ANSTART,
ANENA, ANRUN and ANSTOP. ANRUN is the first signal to go active, it is
normally high to extend opto coupler life on the ANODE MOTOR DRIVER
(AMD). It will go low in preparation for the ANENA signal so there is no
arching when K1 relay on the AMD is pulled. After ANENA has enabled the
anode motor, ANRUN goes high supplying the motor with 40 V. Seventy
milliseconds later ANSTART goes low turning on the opto coupler A4, which
pulls pin 6 of the motor controller (A7) high. This turns the triac V1
completely on supplying the rotor with 220 VAC. After nearly 1.5 seconds of
boosting, ANSTART goes high again ending the boost phase so the voltage
applied to the rotor drops back down to 40 V. After the exposure is finished,
ANRUN goes low removing all power from the rotor so there will be no
arching when ANSTOP pulls relay K2 to remove the phase shift capacitor and
winding from the circuitry for the braking.

An anode brake can be selected in the SETUP mode.


The braking is done by applying a half wave rectified voltage to the main
stator winding. This produces a stationary magnetic field acting on the rotor.
The phase shift winding does not get any voltage.

After the relay K2 has settled, ANRUN goes high again applying 40V half
wave rectified voltage to the stator. To begin braking shortly after that
ANSTART goes low applying the full 220 V half wave rectified voltage to the
stator.

STARTOK and RUNOK are two signals that the CPU monitors on during the
boost and run phase respectively. They will go low if the voltage during each
phase falls between the window set up by the two comparators for each signal.

3.4.4.5 Bucky motor


At the beginning of the exposure, the CPU checks that the grid is in its proper
position using the bucky home signal coming from the BUCKY CONTROL
board. The proper position is just after the reversal of the grid direction. After
this and about 100 milliseconds before the exposure, the bucky on signal (pin 7
of X1) goes high, which pulls the relay K1 and thus connects the bucky motor
to the BUCKY CONTROL board.

The BUCKY CONTROL board uses a photo interrupter to sense the position
of the grid. The input photodiode of this sensor is connected between pins 5
and 6 of X3, and the output photo-transistor is connected across pins 3 and 4 of
X3 on the board. During the normal speed, the phototransistor conducts and
thus grounds the base of transistor V7 and connects the bucky motor to the
adjustable output of A1. Just before the change of the grid direction the
phototransistor stops conducting, and the bucky motor is connected to a higher
voltage through relay K2 thus making the change of the direction as quick as
possible. After the change the speed returns to normal again. After the
exposure the grid is driven back to the proper position.

31687 3-37 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.4.5 Automatic exposure control (AEC)

3.4.5.1 AEC - general


There AEC consists basically of three parts to the circuit, the DETECTOR, the
AEC board and the CPU board.

3.4.5.2 The Detector


The detector is a package containing 34 diodes connected in parallel. As X-ray
strikes the diodes, they begin to leak current from the cathode to anode. The
amount of current each diode leaks is related to the amount of X-ray it
receives. Since the diodes are connected in parallel, the total current draw of
the detector is the total of all diodes current draw.

3.4.5.3 AEC BOARD


There are four basic connections to the AEC board. The main power is
connected to X1, unregulated +18VDC supply to X1/1, analog ground to X1/2
and unregulated -18VDC supply to X1/3. The detector connects to X2
(cathode), X3 (detector shield) and X4 (anode). The detector can be measured
as a diode.

The connector X6 is for static shield. The purpose of the shield is to protect the
phototimer board from major electrical interferences and X-rays.

The CPU is connected with the phototimer through connectors X1/8


(AUTOEXP, information from the CPU) and X1/7 (AECFRQ information to
the CPU). The CPU also provides the +5VDC supply through X1/5 = +5VDC,
X1/6 = ground.

The signal from the detector is taken to the first amplifier ICA1 which is the
preamplifier. Besides amplifying it also does the current to voltage conversion.
Trimpot R2 is for adjusting the offset voltage of ICA1.
(NOTE: adjusted to 0 - offset in the factory)

October 2002 3-38 31687


Performa, Alpha IQ, RT, ST, III Service Manual

The second amplifier A2 has a smaller gain which is adjustable to take into
account differences between detectors, x-ray tubes and film/screen
combinations. Trimpot R6 is for adjusting the offset voltage of A2.
(NOTE: adjusted to 0 - offset in the factory)

The AEC board replaces the phototimer board in the units manufactured since
December 1994). In the AEC board the signal from the detector is taken to
voltage to frequency converter after amplification. The software for this board
is 7.xx or higher (6.35 and 6.55 also). The CPU board needs to be of version
1.2 or higher.

3.4.5.4 The CPU


The CPU board looks at the pulses from the AEC board. The frequency of
these pulses represents the radiation through the breast. The CPU converts the
frequency into a variable called INTEGRATION time. The microprocessor
plugs this time into a formula and calculates the exposure time. Integration
time has a nonlinear effect on the exposure time, which is important to keep in
mind when making adjustments.

31687 3-39 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.5 DESCRIPTION OF CONTROL ELECTRONICS

3.5.1 Mains transformer and SUPPLY board


T1 T2

220V 24V
220V 24V

18V
240V 240V

24V 15V
260V 260V

Figure 3.5.1 Mains transformer connections

The SUPPLY board supplies all the voltages that are not used to produce
X-rays. The collimator bulb voltage is on newer models regulated with a triac
and why the voltage should be measured with the voltage meter at the
AC-range.

3.5.2 CPU board


A block diagram of CPU board is in figure 3.7.

Intel 80C32 (in version 1.07 and below 80C31) (ICD1) microprocessor
running at a clock frequency of 12 MHz is used as the nucleus of the CPU
board.

Port 0 of the processor is used as a combined data bus and the lower half of the
address bus. Port 2 is used as the upper half of the address bus. Circuit ICD27
(74HC373) latches the address code into the address bus when the external
memory fetch is used. Circuit ICD21 (74HC245) with programmable Gate
Array Logic (GAL) circuit ICD22 (22V10), ICD7 (74HC32) and processor
signals PSEN,ALE, WR and RD control the direction of data transmission in
the data bus/lower half of the address bus.

Because the EA-pin of the processor is tied low, all the instructions are fetched
from the external memory. The printed circuit board includes three IC-sockets
for program chips. One is for the program memory circuit (ICD16) and the
second is for static RAM circuit (ICD14). The third is for a nonvolatile
memory EEPROM (ICD15).

October 2002 3-40 31687


Performa, Alpha IQ, RT, ST, III Service Manual

ICD22 (GAL 22V10) is used as an address decoder for data memory and
I/O-circuitry (ICD8, ICD12, ICD13, ICD18, ICD19, ICD23, ICD26 and half of
the ICD5). The address-map of the system can be presented:

0000 - FFFF program memory (64k)

8000 - 9FFF nonvolatile memory (8k)


A000 - CFFF I/O
D000 - EFFF data memory
F000 - FFFF not used

ICD8 (8279) is a keyboard/display controller which independently takes care


of user interface and displays of the unit.

CPU has six 8-bit I/O ports. Three of them is configured as inputs and the rest
three as outputs

The RESET and WATCHDOG circuitry consists of ICA2 (TL7705A).

31687 3-41 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.5.3 DISPLAY BOARD


The block diagram of DISPLAY board is presented in figure 3.5.3

KEYBOARD/
DISPLAY
SEGMENT
CONTROLLER
DRIVER
DISPLAYS
AND
LEDS
DIGIT
SELECT

KEYBOARD

Figure 3.5.3 Block diagram of DISPLAY board

The DISPLAY board is located in the control box and is connected by a cable
to the controller circuit D15 (8279), which is located in the CPU board.

This board includes circuits D1 (ULN2003A), D2 and D3 (74HC138). D1 is a


segment driver for the LEDs (light emitting diodes) and the seven segment
displays. D2 and D3 are decoders, with which the controller selects the digits.

Signals from the keyboard are conducted via this board to an 8279 controller
matrix.

3.5.4 C&Z-DRIVER BOARD


This board controls the functions of compression (C) motor and vertical
movement (Z) motor. C&Z-drive board also supplies the voltage for the bucky.

October 2002 3-42 31687


Performa, Alpha IQ, RT, ST, III Service Manual

3.5.5 CD-BOARD
This board is installed in Alpha MT only. The CD-board measures the breast
thickness. The signal is sent to the CPU board which controls the display on
the CD board. All information is passed through a serial link. On the CD board
locates also the electronic for the beam filter selection and the driver circuit for
the filter change motor.

3.5.6 C-ARM CONTROL BOARD


This board is installed in PERFORMA, ALPHA RT/IQ only. It measures the
compression thickness and force and controls the displays on the c-arm. It also
controls the motor that drives the beam filter change mechanism.

The C-arm Control board is based on Intel 80C32 micro controller. It has a
software of its own which is independent from the software on the CPU board.
The only requirement is that the software is of version 7.01 or higher.

The C-arm Control board communicates with the CPU board with a serial link.
An error in this communication is indicated as E.12 error on the control panel.

The C-arm Control board is divided into digital part and analog part. The
digital part controls the i/o signals and c-arm display board. The analog part
measures the compression thickness and force and drives the filter motor.

The compression force is measured with strain gages in the compression


mechanism. The signal is brought to the amplifier ICA4. The offset voltage of
the amplifier must be set to 500mV when there is no strain on the compression
paddle. The voltage is measured between testpoints TP5 and TP6.

List of LED indicators


LED indicator signal

H1 RXD
H2 TXD

H3 +5V
H4 RST (reset)

List of testpoints
Testpoints signal

TP1 +25V
TP2 +5V
TP3 GND
TP4 Compression thickness measurement
TP5 Compression force measurement
(amplifier offset voltage- set to 500mV)
TP6 GND

31687 3-43 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

3.5.7 C-ARM DISPLAY BOARD


This board is installed in PERFORMA, ALPHA RT/IQ only. It contains 8
seven segment displays, two LEDs to indicate which beam filter is selected
and a filter change switch. The seven segment displays are controlled by ICD4
on C-arm Control board. The LEDs and the switch are handled by regular i/o-
control on C-arm Control board.

3.5.8 ECS DRIVER BOARD


The ECS Driver Board Controls the C-arm movement of the ECS-200 device.
It is an electronic circuit board that requires software based control from
Performa, Alpha unit. Software version of the unit must be 7.27 or greater.

The C-arm movement can be activated in four different ways:

1. Up and down from footpedal in either C-arm or ECS mode.

2. Up and down from either one the two switches located on both sides of
the C-arm covers.

3. Automatic return of the C-arm to it's lowest position after an exposure


if the autorelease function in the Performa, Alpha control panel has
been selected.

4. Return of the C-arm to it's lowest position from the return switch
located on both sides of the C-arm covers.

Up, Down and Return switches are also available on the circuit board for
service purposes.

Motor control

The C-arm movement is handled by a pulse width modulated (PWM) DC


motor. The speed of the C-arm and ECS are separately adjusted with trimmer
potentiometers R37 (C-arm) and R59 (ECS respectively. The PWM control
circuit ICA3 (TL494) controls the motor voltage according the reference
voltage (TP3). The motor is activated by pulling the *MOTOR ENABLE
signal down.

The direction of the C-arm is determined by relay K2. The default direction of
the C-arm is down.

Movement limits

The C-arm movement limits are sensed with two photointerrupters on the
electronic circuit board (ICD9 and ICD10). When the limit is reached a metal
plate enters the photointerrupter slot turning the photointerrupter transistor off.

October 2002 3-44 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Control signals:

Up and down

The user can move the C-arm up and down with a switch located on the cover
or with a footpedal. The *CARM UP signal activates relay K2 and changes the
direction of the motor drive. The motor drive is enabled if the C-arm is not in
it's upper limit.

*CARM DWN signal moves the C-arm down. The motor drive is enabled if
the C-arm is not in it's lower limit.

Return

The user can return the C-arm to it's normal position (down) by pressing the
return switch. This activates D-flip flop ICD6 if the relay K2 is not activated
(that is if the direction of the C-arm is down). When the lower limit is reached
the flip flop is reset and motor drive is disabled.

Autorelease

The autorelease function returns the C-arm down after an exposure. The
information of an exposure is derived from control signals for ANODE
MOTOR DRIVE board (connectors X5 and X6). ANENA signal is active
during preheat sequence and an exposure.

If the compression paddle is driven up after exposure, the C-arm is returned


down after the compression paddle has reached it's upper limit. The
information of compression paddle movement is derived from compression
motor encoder (connector X8).

Compression current limit

The information of the compression reaching it's current limit is derived from
the compression motor voltage (connector X3 pins 10 and 11). When the
breast is being compressed (paddle moving) the optocoupler ICD11 A and B
are activated.

List of test points:

TP1 +34V
TP2 +15V
TP3 Motor speed reference
TP4 Motor voltage
TP5 Motor voltage
TP6 GND
TP7 GND
TP8 +5V

31687 3-45 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

List of LED indicators:

H1 +34V
H2 +15V
H3 C-arm movement enabled
H4 +5V
H5 C-arm shifted (located in the ECS top cover)
H6 C-arm shifted (located in the ECS top cover)
H7 LOWER LIMIT
H8 UPPER LIMIT
H9 C-arm direction up
H10 Motor enabled

October 2002 3-46 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4 PREVENTIVE MAINTENANCE

4.1 X-RAY TUBE PERFORMANCE CHECK

4.1.1 Half-value layer (HVL) measurement


Test equipment required: Radiation meter
0.1 mm Aluminum Plates (Alloy 1100)

Enter the service mode by simultaneously pressing the kV- and density-
buttons and holding them for 3 seconds. Insert the 18 x 24 diaphragm into its
mount. Collimate the X-ray beam by a lead plate with a hole placed close the
tube and restrict the beam to an area slightly larger than the ionization
chamber. Use 25 kV and measure the radiation using 0 - 0.1 - 0.2 - 0.3 - 0.4
mmAl.

Plot your data on the chart and connect data points with a line. Take the 0.0
mmAl reading and divide it by two. Find where that reading would intersect
your graph and determine the half-value layer from the aluminum axis. Use
mAs that gives you about 400 mR with 0 mmAl.

0.0
0.1
0.2
0.3
0.4

NOTE! HVL @ 25 kV MUST be more than .25 mmAl, and should be less than .35
mmAl.

Table 4.1. HVL measurement chart

31687 4-1 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

4.1.2 Tube efficiency check on the cassette holder


Test equipment required: Radiation meter

Center the radiation meter on the image receptor so that its ion chamber is as
close as possible to the chest wall edge of the image receptor.

Place a piece of lead under the meter to block back scatter.

For small focal spot testing, remove the image receptor. The "MAGN" light in
the control panel will lit up, make the exposures in service mode For large
focus, insert a cassette holder and make the measurements on the cassette
holder top.

Enter the service mode by simultaneously pressing the kV- and density-
buttons and holding them for 3 seconds.

Make exposures as directed by the table below, comparing your results with
the data given on the table 4.2.

mR/
mAs
kV SMALL FOCUS LARGE FOCUS
DESIRED ACTUAL CHECK DESIRED ACTUAL CHECK
25 3-9 3-9
29 8 - 15 7 - 14
35 16 - 23 10 - 18

Table 4.2. Tube efficiency

+ All exposures are 100 mAs without the paddle!

+ If the output of the tube consistently falls below the desired level, check kV,
mA, preheat.

October 2002 4-2 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4.1.3 Focal spot measurement procedure


Test equipment required: Star phantom
Slit focal
Special focal spot measurement
diaphragm
Measurement of Large Focus

Insert a bucky or a cassette holder. Enter service mode by simultaneously


pressing kV- and density- buttons and holding them until the timer display
shows "ser".

Press density+ and density- buttons simultaneously until the timer display
shows "51 Pr". Press kV+ button until the timer display shows "Foc" which is
the focal spot measurement mode. Press autoexp to enter the program. The
exposure parameters are automatically adjusted according to the table below.

Focus Insert kV mA T.exp.


Large focus M101G 28 50 0.1
M107 25 70 0.1
M101, M113, 25 80 0.1
M113SP
Small focus M101/M107 25 10 0.3
and
M101G,M113,
M113SP

The default setting of the Large Focus is 28 kV, 50 mA.Large focus


measurement parameters for other inserts can be scrolled by pressing the
mAs+ button.

Remove the bucky or cassette holder

Install the star pattern phantom or slit focal to the special focal spot
measurement diaphragm to have the phantom at a correct angle. The
diaphragm can be ordered from Instrumentarium Imaging. NOTE: Each tube
model has a diaphragm of its own.

Place a normal cassette in line with the front edge of the cassette holder
support. Put a guide under the tube side of the cassette so that the cassette
raises 12 mm (0.5 inc.) from the surface.

Make an exposure.

Develop the film and fill in the form for large focal spot to calculate the focal
spot size. Measure also the magnified diameter of star.

31687 4-3 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Measurement of Small Focus

Enter the service mode by pressing density+ and density- simultaneously until
"ser" is displayed on the timer display.

Install a magnification table, 1.6 or 1.8

Go back to the focus measurement mode as described in previously.

Remove the magnification table and place the cassette on the cassette holder
support as described in previously.

Make an exposure

Fill out the form for small focal spot to calculate the small focal spot.

Bw = Blur diameter of width


Parallel to anode-cathode axis

BL = Blur diameter of length


Perpendicular to anode-cathode axis

Fw = C * BW/(M - 1) , FL = C * BL / (M - 1)

C = .00872 for .5 degree star


= .017 for 1 degree star
= .0349 for 2 degree star (measurements of the small focal spot are
not possible using 2 degree star)

Magnified diameter of star, DS = _________________mm

Optical magnification, M = DS/45= _________________mm

Large focal spot Small focal spot


Bw = ________________mm Bw = _________________mm
Bl = ________________mm Bl = _________________mm
Fw = ________________mm Fw = _________________mm

Fl = ________________mm Fl = _________________mm

October 2002 4-4 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4.1.4 Image quality evaluation procedure


1. Go to the film processor evaluation procedure described in the
Installation Instructions , 2.14.2

2. Place a loaded cassette into the bucky and select AEC operation using
settings (kV) as you would use for an average 4.5 cm compressed
breast.

3. Place the RMI 156 phantom on the bucky in the same position as a
breast. The nipple indent marker should be positioned away from the
chest wall. Use a standard 18 x 24 or 24 x 30 size compression paddle.

4. Set the AEC detector to an inward position under the wax portion of
the phantom.

5. Make an exposure and process the exposed film. This image will
represent the imaging abilities of your machine using these clinical
factors.

6. If the image is over or under exposed, make a AEC adjustment and


repeat the exposure.

7. Once an acceptable image has been obtained, have one or more viewers
identify and record which objects are seen. Use a magnifying glass to
look for the structures as you would on a patients mammogram. Use
the phantom scoring procedure on the following pages to score system
performance.

8. Make another exposure using magnification table and score that image.

9. Mark both films with the date, kV, mAs and their scores and file them
for future reference.

31687 4-5 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

IMAGE SCORING WITH THE RMI 156 PHANTOM

Figure 4.1. RMI 156 phantom

FIBERS SPECS MASSES


Number Size Points Number Size Points Number Size Points
1. 1.56 1 7. .54 1 12. 2.0 1
2. 1.12 1 8. .40 1 13. 1.0 1
3. .89 3 9. .32 6 14. .75 1
4. .75 5 10. .24 7 15. .50 7
5. .54 9 11. .16 10 16. .25 10
6. .40 10

As a minimum, the system shall image:

1) 4 Fibers - 1, 2, 3, 4
2) 2 Specs - 7, 8, 9
3) 3 Masses - 12, 13, 14

To meet the ACR criteria the minimum acceptable score would the yield a
minimum score of 10 for fibers, 8 for specs and 3 for masses. Total score
would the be 21.

If your mammographic unit is significantly worse, you may wish to check the
accuracy of the kVp, the focal spot size, the half value layer, and the image
processor, You could also consult a medical physicist or engineer who
specializes in diagnostic radiology.

October 2002 4-6 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4.1.5 Measurement of the feedback circuitry

CAUTION! The readings from the following resistance measurements depend on the
probe polarity.

Figure 4.2 Measurement of the feedback circuitry

Remove the low voltage connector X33 from the HV-unit and measure on the
cable connector end the resistance between ground and pin X33/8 (This is the
kV-feedback resistance on the INVERTER Board).

It should be 20-45 k:. (If the power is ON, the value is 45 kohms).

If 0Ω check the cabling between the connector and the FILAMENT


CONTROL Board, and the cabling between FILAMENT CONTROL Board
DQG WKH ,19(57(5 %RDUG ,I WKH UHDGLQJ LV DSS  N  FKDQJH WKH SUREH
polarity. Plug in connector X33.

CAUTION! X36 is hot with power on.

Remove the connector X36 from the tube and measure on the cable connector
end the resistance between ground and pin X36/5 (this is the mA feedback
resistor on the FILAMENT CONTROL Board). It should be about 115 :
when small focus on and 30 : when large focus on. Plug in connector X36.

NOTE! mA feedback will read 30 : (large focus) only with power off.

31687 4-7 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

4.2 AUTOMATIC EXPOSURE CONTROL EVALUATION

4.2.1 Quick AEC test


Testing procedure:

1 Attach the cassette.


2 Position an acrylic plate carefully so that it covers the AEC detector.
3 Select 25 kV and density 0.
4 Make an exposure using the semi-AEC.
5 Develop the film.
6 Measure the film density. If the density is not between 1.3 and 1.5, the
reprogramming of AEC is recommended.

Carry out this test once a week.

4.2.3 Object thickness compensation


Measure the optical density for 20 and 40 millimeters of acrylic. Use also 60
millimeters if available. Use 25 kV without bucky and 27 kV with bucky and
magnification. Measure the optical density for both molybdenum and rhodium
filter. Plot the values in the table below.

ACR mm O.D. Bucky) O.D. (Cas tunnel) O.D. (Magn)


20 / / /
40 / / /
60 / / /

1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
mm 20 40 60

Table 4.5. Object thickness compensation

October 2002 4-8 31687


Performa, Alpha IQ, RT, ST, III Service Manual

If the optical density is higher with 20 mm acrylic than with 40 mm acrylic the
thickness compensation has to be adjusted.

1. Adjust the AEC so that the wanted optical density is reached 40 mm


acrylic. Notice the exposure time
2. Go into program mode (Pr) and select -2 Density
3. Adjust the gain on the AEC board so that the same exposure time is
reached as in point 1 with 40 mm acrylic.
4. Repeat the test 4.2.3
5. If compensation not correct select more - Density

NOTE: In some cases + Density should be used instead of - Density

31687 4-9 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

4.2.2 kV compensation
Measure the optical density for each kV station, using molybdenum and
rhodium filter if available Use 40 mm ACR.

kV O.D. (Bucky) O.D. (Cas tunnel) O.D. (Magn)


23 / / /
24 / / /
25 / / /
26 / / /
27 / / /
28 / / /
29 / / /
30 / / /
31 / / /
32 / / /
33 / / /
34 / / /
35 / / /

1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
kV 23 24 25 26 27 28 29 30 31 32 33 34 35

Table 4.4. kV compensation, x = molybdenum filter, o = rhodium filter

October 2002 4-10 31687


Performa, Alpha IQ, RT, ST, III Service Manual

4.3 MECHANICAL CHECKS

4.3.1 Compression system


The part of the equipment which undergoes the most use is the compression
system, so it needs regular attention and treatment when necessary. Spread
some grease (MBDE or other high-quality grease) on the compression driving
axle and support joints. If the driving wheels are noisy or the hand wheels
difficult to rotate, we strongly recommend that you remove the C-arm and
grease the inside of the compression system. Check also the digital
compression display according to the test described in the Installation
Instruction.

4.3.2 C-arm rotation brake


Check that the brake has 110 N (24 lbs.) of holding force, if necessary tighten
the four nyloc nuts to increase holding force if necessary. Check that only 20 N
(4 lbs.) of force is needed when the brake is disengaged. If required, loosen the
four nyloc nuts or remove one disc spring pair. Recheck the 110N (24 lbs.)
holding force.

4.3.3 Column and carriage


Check all cables, especially the cables coming from the C-arm. Check also the
counter weight wire condition.

4.3.4 Compression force checking


Place a scale under the paddle and compress it like a breast using hand wheels.
The mechanical clutch should stop the compression movement before the
pressure reaches 250 N (55 lbs.). If the compression exceeds this, adjust the
clutch by removing the back cover of the C-arm and turning the sealed and
labeled screw. Do not forget to lock the screw after adjustment.

4.3.5 Testing the x-ray field illumination


Select the desired film size, load the film into the cassette and put the cassette
and diaphragm in their places. Select the manual mode (25 kV, 20 mAs). Place
a piece of white paper (18 x 24 and/or 24 x 30) on the cassette holder so that
the light field can be seen clearly. Place for example some coins in the corners
of the area illuminated on the cassette holder and make an exposure. Develop
the film. The coins on the film should not be more than +- 5 mm out of the
exposed field. If they are, adjustment must be performed for the light field.

31687 4-11 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

THIS PAGE HAS INTENTIONALLY BEEN LEFT BLANK.

October 2002 4-12 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5 TROUBLESHOOTING & ERROR MESSAGES

5.1 Error message chart


Error code Signal line Short explanation of the error
E.01 MAOK No tube current or the preheat level far too low.
E.02 GENOK No kV.
E.03 STARTOK Anode too hot or the anode motor does not start.
E.04 SERIAL Communication error between FILAMENT CONTROL board
and CPU board.
E.05 N/A Exposure counter exceeds the preset limit.
E.07 RUNOK Stator voltage too high during exposure.
E08 Bad connection in breast thickness measurement potentiometer.

E09 Beam filter is not in its position.

E10 Compression force display error.

E11 C-arm angle display error.

E12 Link error between CPU and C-arm Control board.

E.r1 N/A Check sum error in the EPROM.


E.r2 N/A Check sum error in the EEPROM.
E.r3 N/A External RAM error.
E.r4 EPROM error in C-arm display board.

E.r5 RAM error in C-arm display board.

C.br N/A Brake released during exposure.


C.bU BHOME Bucky does not find the HOME position.
C.CA CASNOK/ CASOPTO board does not sense the cassette or it wasn't changed
after last exposure.
CASBUC
C.CH CHON CPU does not sense the cassette holder.
C.Cn BHOME CPU does not sense the bucky connection.
C.Co BLE CPU does not sense the diaphragm or wrong size of diaphragm.
C.dE N/A Exposure time is less than integration time.
C.ES Exposure button line is activated when unit is turned on.

C.PS IQC, System control out of range > ± 40%.

O.FL N/A Exposure time exceeded 5 seconds.


r.EL EXP2 Exposure button released prematurely.
r.ES N/A Microprocessor reset.

31687 5-1 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

U.FL N/A Integration time is less than 20 millisecs.


S.Lo LOW 1 Low line voltage or 2. Fault in the INVERTER board.
S.YS CPU error.
t.oF Motor time out error.
E.AF Communication error CPU - AEC.

Table 5.1. Error message chart

5.2 Useful information to have when calling for assistance


a) General (required information)

1. Serial number of the unit.


2. Name of the operator who experienced the problem.
3. Software revision level. (Shown on mAs-display and C-arm Control
board when power is switched on)

b) For phototiming questions (AEC)

1. Film/screen combination used.


2. Temperature of the developer.
3. What is the gross fog, speed index, contrast index of the film
developed in the processor?
4. Was any work done on processor just prior to problem?
5. Did problem appear suddenly or gradually?
6. Breast compression, kV selected, type of view, mAs of exposure,
detector positioning, filter selected, which imaging mode?

c) Other problems

1. What is the history of the problem?


2. Error codes on the display, error codes in error counter (service
program menu/58, cnt/r
3. What have you done to attempt a fix?
4. Is it hard failure or intermittent?
5. If intermittent, how often per day?
6. Did problem appear as a result of some other work performed on the
unit?
7. Is any other equipment exhibiting intermittent problems?

October 2002 5-2 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.3 E.01 Error message explanation


NO TUBE CURRENT

The FILAMENT CONTROL board compares the mA-feedback voltage


MAFB to a reference value that is about 80% of the actual mA-reference
voltage. When the feedback exceeds this reference value it sends a MAOK
signal to the CPU board. If the CPU does not get this signal in 0.2 seconds
after EXT signal, the processor terminates the exposure and gives error
message E.01.

This can depend of a problem either in kV- or in mA- circuitry, because if the
high voltage (kV) is absent, anode current (mA) can not be produced. .

a. CONNECTORS. Check that all connectors on INVERTER and


FILAMENT CONTROL boards are properly installed

b. SUPPLY VOLTAGES. Check that all supply voltages are present


(See section 5.28).

c. FILAMENT MOSFETs. Turn the unit off. Connect the common of


your DVM to the leg of resistor R38 and plus to the leg of resistor R50.
You should read .4 volts using diode range.

d. mA REFERENCE. Turn the unit on and measure the mA reference


voltage from test point TP15. At 27kV (large focus) the reference
voltage should be appr. 2,78 volts.

e. SOFTSTART. Go to PERFORMA, ALPHA softstart procedure, (2.12


on Installation Instructions Manual).

NOTE: IF THE PREHEAT CAN NOT BE ADJUSTED AS DESCRIBED IN THE


SOFTSTART PROCEDURE, CHANGE THE HIGH VOLTAGE
TRANSFORMER.

OTHER REASONS

- Very slow tube current rise (easily with small focus)

- No kV

- No filament in the X-ray tube

- Connector to filaments open

31687 5-3 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.4 E.02 Error message explanation


NO KV

On units with INVERTER and FILAMENT CONTROL Boards, the E.02 error
indicates that there is no kV present. This can happen if either the INVERTER
board or the HIGH VOLTAGE UNIT is faulty. The problem can also be that
the FILAMENT CONROL board does not provide necessary control signals
for the Inverter board.

a. CONNECTORS. Check that all connectors on INVERTER and


FILAMENT CONTROL boards are properly installed

b. MOSFETs. DISCONNECT ALL CONNECTORS FROM


INVERTER BOARD 1. Connect the common of your DVM to test
point TP3 and plus to test point TP1. You should read .4 volts in diode
range. 2. Connect the common of your DVM to test point TP4 and
plus to test point TP1. You should read .4 volts in diode range. 3.
Connect the common of your DVM to test point TP2 and plus to test
point TP3. You should read .4 volts in diode range. 4. Connect the
common of your DVM to test point TP2 and plus to test point TP4.
You should read .4 volts in diode range.

c. FUSE. Measure that the fuse has not been blown

d. SUPPLY VOLTAGES. Turn the unit on. Check that all supply
voltages are present (See section 5.28).

e. MOSFET DRIVER. Go to service mode. Make an exposure without


X-rays. (See section 3.4.3.8; remember jumper X8 on FILAMENT
CONTROL Board) Make sure that the LEDs H4, H5, H6, H7, H11
and H12 are lit during the exposure.

f. kV REFERENCE. Turn the unit on and measure the kV reference


voltage from test point TP24. At 27kV the reference voltage should be
app. 2,7 volts.

NOTE! IF THE INVERTER BOARD APPEARS TO WORK CORRECTLY


CHANGE THE HIGH VOLTAGE TRANSFORMER.

OTHER REASONS
- Inverter board
- HV-unit

IF HV-UNIT CHANGED:
- Check kV
- Check Preheat
- Check AEC

October 2002 5-4 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.4.1 Testing of the new Alpha generator


(new generator consisting of the inverter and filament board)

Follow the testing procedure if the Inverter Board seems to be working well,
the system gives E02 error messages and it is not clear if there is a problem
with the HV-unit, the HV-cable or with the X-ray tube

1) Open connector X4 on Filament Board.


This disconnect the filaments from their power supply

2) On Filament Board move jumper X8 between pins 1 & 2.


The jumper is normally between 2 & 3. This forces the filament supply
always to take it's feedback from the filament voltage even during an
exposure. Normally during an exposure the feedback is taken from the
tube current.

3) On Filament Board move Switch S1 to "TEST" position.


This disables the CPU from getting the error message (E01) because of
the missing tube current.

4) Disconnect HV-cable from the HV- unit and start taking exposures
with 20kV and gradually increase the kV setting.

5) If you can take exposures up to 35kV, the HV-transformer is good

6) Connect the HV-cable into the transformer and disconnect it from the
tube.

7) WITH A PLASTIC ROPE OR EQUIVALENT PULL THE LOOSE


END OF THE HV-CABLE AWAY FROM THE PERFORMA,
ALPHA. THIS IS TO MINIMIZE THE POSSIBILITY OF ARCING
TO THE CHASSIS.

8) Start making exposures with 20kV and increase the setting gradually.

CAUTION! BE EXTREMELY CAREFUL!!!

9) If you can make exposures up till 35kV without E02-errors, the HV-
transformer and the HV-cable are good and the reason for the error
message can be in the x-ray tube.

31687 5-5 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.4.2 Testing of the old Alpha generator


(old generator consisting of the Pulse Board, Base Drive and Power Board)

Follow the testing procedure if the Power Board seems to be working well, the
system gives E02 error messages and it is not clear if there is a problem with
the HV-unit, the HV-cable or with the X-ray tube.

1) Disconnect filaments by opening connector X2 on Base Drive.


This disconnects the filaments from their power supply

2) Remove ic-chip A7 (opto) on Pulse Board and put a short between pins
4 and 5.

This disables the CPU from getting the error message (E01) because of the
missing tube current.

3) Disconnect HV-cable from the HV- unit and start making exposures
with 20kV and increase the setting gradually.

4) If you can make exposures up till 35kV without E02-errors, the HV-
transformer is good.

5) Connect the HV-cable into the transformer and disconnect it from the
tube.

6) WITH A PLASTIC ROPE OR EQUIVALENT PULL THE LOOSE


END OF THE HV-CABLE AWAY FROM THE ALPHA. THIS IS TO
MINIMIZE THE POSSIBILITY OF ARCING TO THE CHASSIS.

7) Start making exposures with 20kV and increase the setting gradually.

CAUTION! BE EXTREMELY CAREFUL!!!

8) If you can make exposures up till 35kV without E02-errors, the HV-
transformer and the HV-cable are good and the reason for the error
message can be in the x-ray tube.

October 2002 5-6 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.5 E.03 Error message explanation


X-RAY TUBE IS HOT

The condition of the AMD board during pre-heat is controlled by the CPU
through the STARTOK signal. A disturbance generates E03.

a. CONNECTORS. Check that all connectors coming to the AMD


board are properly and right way installed. Check also that connector
X36 on the X-ray tube is properly installed.

b. THERMAL SWITCH Disconnect X36, measure between pins 1 and


3, and 2 and 3 on the tube side, if the circuit is open, the thermal
switch is open. Measure the resistance between pins 1 and 2. The
resistance should be about 66 ohms If the circuit is open and the
resistance is right, wait about 45 minutes and check if the thermal
switch is closed. If the resistance shows OL (over load) , the coils of
the anode motor are broken, change the tube.

c. LINE VOLTAGE. Check that the AMD board does get 220VAC
between TP1 and TP2. If not check cabling and connections on X9.
Measure the running voltage with a TRMS meter, between X4 pin 2
and X5 pin 2. . It should be about 30VAC. If not, adjust with trimmer
R100.

d. OTHER PROBLEM. Change the AMD board

HINT! When the user gets the E03, leave the unit ON because the fan cools the
tube.

NOTE! If measured with a nonTRMS meter, reading is about 20 - 30 VAC.

31687 5-7 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.6 E.04 Error message explanation


Problem in serial communications between CPU and FILAMENT BOARD.

The CPU board sends serial data to the serial-to-parallel latch (D3) located on
the PULSE board (or FILAMENT CONTROL board). The CPU receives the
same signal back from pin 2 of D3 (buffered and isolated through A9 and
A5).If the received signal is different from the transmitted signal an E04 error
message is indicated.

a. CPU - FILAMENT CONTROL BOARD CABLE Check that the


cable between FILAMENT CONTROL and CPU board is properly
installed (connector X5 on CPU board, connector X1 on filament
control).

b. FUSES F1 - F4. Check that the fuses on the FILAMENT BOARD


(F1-F4 BASE DRIVE Board) are not blown.

c. FILAMENT BOARD (BASE DRIVE) VOLTAGE. Using the


negative side of capacitors. C11 - C14 as ground, measure the voltage
across C11 - C14 (+32VDC - +37VDC), from the case of
TO3-regulator A3 (+20VDC), from fuse F3 (-15VDC) and from fuse
F4 (+15VDC). If one of the voltages are missing or out of range,
change the FILAMENT BOARD (BASE DRIVE) board.

d. D1, D2 or A4 BROKEN. If spare circuits available, change the


circuits, if not, change the PULSE board.

e. 8031 PROCESSOR BROKEN. If spare circuits available change the


microprocessor 8031, if not, change the CPU board.

E.04 with software 6.01 and higher (units with INVERTER and FILAMENT
CONTROL boards).

f. SUPPLY VOLTAGES. Turn the unit on. Check that all supply
voltages are present on FILAMENT CONTROL Board. (See section
5.28).

October 2002 5-8 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.7 E.05 Error message explanation (software 5.08 and higher)


Error E05 indicates that the preset exposure counter limit has been exceeded.
The unit does not allow further exposures until the limit has been adjusted.

USAGE
- Entering in the limit setting: Press Den- and mAs- all the time and
switch the power on. When display starts blinking, keep pressing Den-
and mAs- and reset the program by pressing AutoExp.

- Changing the limit: Press Den- or Den+, 0 means no limit (OFF), other
values indicate amount of the accepted exposures. Altogether there are
50-50 000 possible limit variations.

- Exiting the limit setting: Press kV-.

5.8 E.06 ERROR MESSAGE EXPLANATION


Error E06 is caused if there is a communication problem between the Alpha ID
(Dataview) PC-controller and the Performa, Alpha RT/IQ. The Alpha ID
(Dataview) sends a check request to the Performa, Alpha unit, 12 seconds after
the last communication. If Dataview does not get any respond a E06 error code
is displayed.

a. Check the cable connection between the Performa, Alpha unit and the
Alpha ID (Dataview).
b. Check that the Performa, Alpha RT/IQ program is correctly selected in
the setup mode. (If Dataview, "i" is always 0 or 1. With Alpha ID i
=2.).
c. Check that Performa, Alpha unit is turned on.
d. Check that the Alpha ID (Dataview) is turned on.
e. Check Alpha ID (Dataview) setup (see Alpha ID (Dataview) Operation
Instructions).

ATTENTION!
With Alpha ID2:
a) using Delta 16, i=3
b) without Delta 16, i=1

5.9 E.07 ERROR MESSAGE EXPLANATION


Problem flagged by signal RUNOK generated by the ANODE MOTOR
DRIVER. Run voltage is too high.

a. RUN VOLTAGE. Measure the run voltage, it should be around 30 V.


Go to Anode Run Voltage Procedure in the Alignment & Adjustment
section.

If nothing else works, change the AMD board and/or CPU. Check cabling.

31687 5-9 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.10 E.08 ERROR MESSAGE EXPLANATION


The digital compression thickness display does get an analog voltage
information from the slide potentiometer located on the C-arm frame. On the
Performa, Alpha RT/IQ the kV is set automatically based on the compression
thickness. If the analog voltage is missing, the Auto-kV would not work
properly. When the compression is driven to the lowest position the voltage
output is not 0V.If a connection problem occurs, the voltage drops to 0V and a
E08 error message will be activated. Exposures are inhibited during the error
message.

a. Check the slide potentiometer connection and clean the potentiometer


surface.

5.11 E.09 Error message explanation (PERFORMA, RT/IQ)


PERFORMA, RT/IQ ERROR WITH FILTER

The beam filter position is controlled by two microswitches. If the filter is not
in position a E09 error message will be activated.

a. Check the beam filter mechanism, microswitches, C-arm control board


and motor.

5.12 E.10 Error message explanation


Error in the compression force measurement.

The micro controller on the C-arm Control board gets incorrect readings from
the analog to digital converter.

a. Check that the offset voltage of the ICA4 amplifier is 500mV when
there is no strain on the compression paddle. (Measure between TP5
and TP6).
b. Check the cabling.
c. Check that the C-arm angle measurement is correctly calibrated. The c-
arm angle value is used for compensation in the force measurement
routine.

E.10 occurs also if the compression paddle is driven past the upper
microswitch against the mechanical limit.

d. Drive paddle totally up.

October 2002 5-10 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.13 E.11 Error message explanation


Error in the c-arm angle measurement

The CPU board gets incorrect readings from the analog to digital converter on
the Filament Control board.

a. Check the cabling


b. Check the potentiometer reading with an ohm meter.
c. Check the FILAMENT BOARD
d. Check measurement potentiometer (max. position 5 kohms, 2.5 kohms
when 0 degrees.

5.14 E.12 Error message explanation


Error in the serial link between CPU board and C-arm Control board

The communication between CPU board and C-arm Control board is not
working.
a. Check the cabling
b. Check the supply voltages on both boards.
c. Change C-arm control board and /or CPU

5.15 E.r1 Error message explanation


Error in the EPROM checksum (Program chip).

a. Check the seating of the chip


b. Change the EPROM or the whole CPU board.

After changing software: Pr-mode - scroll through (Press Autoexp) - come out.

5.16 E.r2 ERROR MESSAGE EXPLANATION


Error in the EEPROM.

a. Change the EEPROM or the whole CPU board.


After changing software: Pr-mode - scroll through(Press Autoexp) -
come out - calibrate the displays.

5.17 E.r3 Error message explanation


Error in the external RAM.

a. Change the RAM or the whole CPU board.


After changing software: Pr-mode - scroll through(Press Autoexp) -
come out.

31687 5-11 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.18 E.r4 Error message explanation


Error in the C-arm Control board EPROM (Software chip).

a. Change the EPROM or the whole C-arm Control board.

5.19 E.r5 Error message explanation


Error in the C-arm Control board EEPROM

a. Change the EEPROM or the whole C-arm Control board.

5.20 S.Lo Error message explanation


POWER SUPPLY VOLTAGE TOO LOW

The FILAMENT CONTROL board monitors the line voltage by comparing


the unregulated, rectified 24VAC voltage to a fixed voltage on the
FILAMENT CONTROL board. If the rectified 24VAC (+34V) voltage falls
below app. 20V during the preheat sequence the LOW signal is activated
telling the CPU to inhibit the exposure

a. Check the line voltage


b. Check the +34V voltage on FILAMENT CONTROL board
c. Check the mains transformers
d. Check mains voltage during exposure (it should not go under
190VAC). If mains voltage is too low, use a step up transformer.

5.21 C.Br Error message explanation (Software 7.27 and up)


a. Check that the patient has not pressed the brake button during
exposure.
b. Check the brake circuit

5.22 C.bu Error message explanation


Bucky home signal not sensed after exposure. Bhome activated for over
500ms. (the turn around takes too long,/ Bhome is continuously activated).
What is most likely happening: After an exposure => The grid will be
positioned so that the movement is stopped immediately after the turn around =
when leaving biome. If that does not happen (the D connector is not properly
seated?) => C.bu

a. CONNECTOR X4. Check that the bucky control cable connector X4


on the CPU is properly installed.

b. BUCKY CONTROL BOARD. Remove the carbon cover of the


bucky. Check that all signals are received from X4 on the CPU board to
the BUCKY CONTROL board.

October 2002 5-12 31687


Performa, Alpha IQ, RT, ST, III Service Manual

c. VISUAL INSPECTION. Check visually that the grid can move freely.

d. FAULTY CPU BOARD. Change the CPU board.

5.23 C.CA Error message explanation


Cassette sensing board does not sense the cassette

a. Adjust opto gain from trimmer R11 on Cas-opto board (Optipoint).

a. CASSETTE CHANGE. Make sure that the customer does not use an
other cassette to push out the loaded cassette from the cassette holder.
The CAS-OPTO board does not interpret this as a cassette change.

b. CASNOR SIGNAL. Check the CASNOR signal receives to the CPU


board on X2 pin 19. If signal is received, change D27, D8 and D15
if spare circuit available, else change the whole CPU board. If signal is
not received change the CAS-OPTO board. .

5.24 C.Ch Error message explanation


Cassette holder is not being sensed

a. CASSETTE HOLDER. Cassette holder is not properly installed.

b. CODE MAGNETS. Check that the code magnets located under the
cassette holder are present

Seen from the bottom of the cassette holder, with the breast side against chest.
small /large
NORM. CASS. TUNNEL X X X X/O X = magnet
BUCKY O X X X/O O = empty
MAGNIFICATION O X O O
STEREOTACTIC O X O O

c. C-SENSE BOARD

31687 5-13 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.25 C.Cn Error message explanation


Bucky connection is not sensed. Bhome is not activated at all. (5sec time
limit)

a. BUCKY CONNECTION. Check that the bucky is properly attached.


Control also that the connector pins on the bucky are not bent.

b. BUCKY SIGNALS. Perform test as described in 3.4.4.5.

c. GRID IS STUCK.

5.26 C.Co Error message explanation


Diaphragm is not sensed.

a. MICROSWITCH. Check that the microswitch is properly aligned

b. CABLING. Measure on the CPU board that pins 8 and 9 on X2 are


shorted when the diaphragm microswitch is closed. If yes, change D16
and D21 on the CPU board if spare circuits available, else change the
whole CPU board. If pins 8 and 9 are not shorted, change the
microswitch

c. COLLIMATOR INTERLOCK. Wrong film size vs. diaphragm.

5.27 C.ES Error message explanation


If the exposure button is damaged and short circuited a C-ES error message
and a beep is activated. Exposures are inhibited until the error has been
corrected.

a. Check the exposure button line (see dia mode)

5.28 C.PS Error message explanation


If the correction rate in the IQC system control mode is more than +/-40% (+/-
20% in Germany) compared to the exposure time calibrated during the
installation, a C.PS error message will be activated.

NOTE! Exposures are not inhibited.

a. Film processor has drifted too much. Call for film processor service.

October 2002 5-14 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.29 C.dE Error message explanation


(With the old software the microprocessor has calculated that exposure time
would be less than integration time.)

New software: Too much radiation to detector.

a. TOO HIGH kV in Semi-AEC. Decrease the kV with 1 - 2kV.

b. AEC DETECTOR. If a patient with small breasts, check that the


detector has been in the position nearest the chest wall

c. AMPLIFIER GAIN. Check that he gain adjustment on the


AMPLIFIER board is not set too high.

5.30 O.FL Error message explanation


Problem: The time interval between the signal AUTEXP from the
CPU board to the PHOTOTIMER board and CNTRDY
from the AMPLIFIER board back to the CPU board is too
long causing exposure times that would exceed the
maximum

a. TOO LOW kV. Increase the kV with 1 - 2kV.

b. AMPLIFIER GAIN. The gain adjustment on the PHOTOTIMER /


AEC board is set too low

c. WITH SOFTWARE 7.27 AND UP. Change the full AEC kV table
(Pr#62 AEC parameter E) or increase the kV max. change value (Pr
mode)

d. CHECK THE BUCKY CONNECTION (apply some compression on


the bucky)

- Exposure monitoring: O.FL and kV display flashing

5.31 R.EL Error message explanation


The hand switch was released prior to the end of an exposure.

a. HAND SWITCH. Check the connections in the hand switch

b. CABLING. Check the exposure cable.

With software 6.xx a broken AEC board may give R.EL.

31687 5-15 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.32 R.ES Error message explanation


Indicates a reset of the microprocessor on the CPU board. Normally
activated when the machine is turned on or the reset button on CPU
board is pressed. Otherwise it indicates a disruption of the voltage
applied to the CPU board.

a. DISTURBANCE. Control that no spike, generated as a result of


an E02 error, or some static electricity caused the R.ES code.

OTHER REASONS
- CPU supply voltage
- Missing HV-grease
- Bad grounding
- Bad HV-unit
- Bad 5V cabling from CPU

5.33 (U.FL Error message explanation) not used


With software up to 6.xx (integration time based AEC): Integration time is less
than 20 milliseconds. Exposures resulting from such short integration time are
too dark.

a. CHECK THE BUCKY CONNECTION (apply some compression on


the bucky)

With software 7.xx and up: Phototimer frequency out of range.


- Check AEC system
- Check diaphragm
- Check detector

5.34 S.YS (Software 7.27 and up)


Cpu error.

a. Check CPU supply voltage


b. Check cabling
c. Change the RAM circuit on CPU
d. Change the CPU

5.35 t.OF (Software 7.27 and up)


Motor timeout error. Motor runs too long (Vertical movement, filter motor,
compression motor, ECS motor)

a. Check motor cabling


b. Check stop limit microswitches (filter, compression, ECS)

October 2002 5-16 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.36 E.AF (Software 7.27 and up)


Poor connection between CPU and AEC boards. Incorrectly adjusted AEC
offset voltages. Autozeroing circuit does not work at start up.

a. Check the cabling.


b. Adjust the AEC board offset voltages.
c. If you are using the 32315 AEC board with software 7.27 or later, you
can remove the capacitor C35 (22uF) on the 32315 AEC board.

5.37 No power indications


a. Check the mains switch.
b. Check that the power cable and the display cable are connected.
c. Check the circuit breaker (proper main fuses 15 A slow) in the rear
bottom.

Open the side covers and check that three LED:s (H1-H3) on the SUPPLY
board are lit. If they are not, check fuses on the SUPPLY board, supply cables,
and power supply connections. Measure that proper secondary voltages are
available at supply transformers T1 and T2.

If LEDs on the SUPPLY board are still not lit replace the SUPPLY board.

5.38 AEC Inoperative


Check cabling to the PHOTOTIMER / AEC board, supply voltages ± 12V,
+5V on the PHOTOTIMER / AEC , and continuity of AUTOEXP and
CNTRDY signals.

Measure the AUTOEXP signal with an oscilloscope. The CPU should set the
signal up 30ms after the radiation rise. If the amplifier does not receive the
signal, check the CPU board.

Measure the CNTRDY / AECFRQ signal coming from the amplifier. Make
sure you have no absorption in the radiation field. If you have a
PHOTOTIMER board the signal should go down for 150µs. If the signal does
not go down, change the amplifier and the detector. If you have an AEC board
the AECFRQ signal should contain high frequency.

To check whether the CPU sees this frequency press autorel button in service
mode. If the numbers displayed in the kV and mAs window show 00 then
there is something wrong with the cabling or AEC board.

- Check grounding.
- Check Cassettes/films/screens
- Check detector positioning
- Check HV-grease
- Check spot diaphragm

31687 5-17 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.39 Software diagnostic mode


The software includes a diagnostic mode which presently has two parts. The
first is a software counter which records the number of error messages seen
and also the number of exposures. The second gives a readout of switch and
signal status.

5.39.1 Entering Diagnostic Mode


1. Turn the unit on

2. Enter "service" (SER) mode by simultaneously pressing the kV- and


density- buttons.

3. Enter the service mode special program menu by simultaneously


pressing density+ and density- buttons until the "Pr 51" mode is
displayed on the timer display. Press kV+ until "58 CeC" is displayed.
Press Autoexp to enter the program. To return to the menu press
"autoexp" again.

5.39.2 Error readout


By pressing the "kV+" button the occurrence of the different error messages
can be controlled. The "kV" & "mAs" displays are read together to give the
total error occurrence.

For example:

kV display = 00

mAs display = 010

error code xx occurred 10 times

October 2002 5-18 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.39.3 Switch and signal status display


Select special program "59 Cin" and enter it by pressing "autoexp". This
program is the switch and signal status display. The timer display indicates
which device and port on the CPU board is reading the switch. The kV display
indicates the bit connected to and the mAs display tells the state. A change
from LO to HI indicates that the line activated when pressing a press button is
ok.

The normal function of the press button is inhibited in this mode.

SWITCH INTERFACE IC/PIN on CPU BIT NORMAL


DEVICE CARD STATUS
kV- dS2 D29/13 + b0 (R0) LOW
D15/38
kV+ dS2 D29/13 + b1 (R1) LOW
D15/39
mAs- dS2 D29/13 + D15/1 b2 (R2) LOW
mAs+ dS2 D29/13 + D15/2 b3 (R3) LOW
density- dS2 D29/13 + D15/5 b4 (R4) LOW
density+ dS2 D29/13 + D15/6 b5 (R5) LOW
AUTO REL EXIT FROM THIS MODE
AUTO EXP dS3 D29/13 + D15/7 b6 (R6) LOW
Handswitch CP1 D1/2 b1 LOW
Comp up P2A D16/37 b7 LOW
Comp down P2A D16/38 b6 LOW
Z up P2A D16/3 b1 LOW
Z down P2A D16/4 b0 LOW
COLL micro P2b D16/25 b7 LOW
CHL micro P2b D16/24 b5 LOW
CHR micro P2b D16/24 b6 LOW
CHON micro P2b D16/22 b4 LOW
COMP micro P2A D16/2 b2 LOW
CASNOR P1C D8/12 b5 LOW
CASBUC CP1 D1/1 b0 LOW

Table 5.2. Switch status display chart

31687 5-19 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.40 SUPPLY VOLTAGES

SUPPLY BOARD
a) TP7 to TP1 brake 26 Vdc
b) TP7 to TP2 33 Vdc
c) TP7 to TP3 lamp 11.5 Vdc**(R8 set to 160 lx in USA)
d) TP7 to TP4 Vin CPU 23 Vdc
e) TP7 to TP5 +AEC in 20 Vdc
f) TP7 to TP6 -AEC in -20.5 Vdc

CPU BOARD
h) across C36 Vin CPU 5.1 Vdc

C&Z BOARD
i) TP1 to TP Comp up/dn 22 Vdc
j) TP3 to TP4 Vert up/dn 25 Vdc
k) TP8 to TP6 Vin bucky 24 Vdc
l) TP8 to TP7 Vin CPU 5.1 Vdc

PULSE BOARD
m) Anode V3 tp R25 (D/A) 10 Vdc
n) V3 t0 A16 - 18 pin 16 5 Vdc

BASE DRIVE BOARD


o) See soft start procedure

FILAMENT CONTROL BOARD


p) TP12 to TP9 35.0 Vdc
q) TP12 to TP10 +Vcc +15 Vdc
r) TP12 to TP11 -Vcc -15 Vdc

INVERTER BOARD
s) TP29 to TP27 +Vcc +15 Vdc
t) TP29 to TP30 -Vcc -15 Vdc

C-ARM CONTROL BOARD


u) TP2 to TP3 +Vcc +5 Vdc

** = in Performa, Alpha RT/IQ units in USA voltage is about 9 Vdc.

October 2002 5-20 31687


Performa, Alpha IQ, RT, ST, III Service Manual

5.41 Grid lines


The grid in the bucky is stuck, too slow or connector are bad.

a. Check cassette holder recognition


- magnets under the bucky
- Check reed relays on Csense board

b. Check grid speed


- Speed should be about 1.2-1.4 sec for slow movement time
- Adjustment from the Bucky Control board using the Pr #61 bUC
- Check the speed when c-arm tilted 0, -90 and +90 degrees

c. Supply voltages to the bucky


- Check that the pins in connectors are not bent

d. Grid alignment and movement


- is the grid stuck (motor and gearbox maybe broken).
- sides of the grid may not touch the sides of the frames.
- The grid has to be close to the carbon fiber in the front but may not
touch.
- is the grid properly attached to the blocks holding it.
- when the breast is compressed, does the carbon fiber top bend and
touch the mechanism.

e. Grid movement detection


- is the opto isolator recognizing the movement.

31687 5-21 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

5.42 ECS Trouble shooting

Problem Items to check

The ECS does not move. - The signal from the switches reaches the X3 on
ECS board
- The motor is connected to the X6 on the ECS
board
- The optocouplers for movement limits are ok
(see leds H7 and H8)
- The motor works (use external voltage supply)

The autoreturn function does - The autoexp led on Performa, Alpha control
not work. panel is lit
- The connectors X2, X3, X4 and X5 are properly
connected

The ECS does not stop when - The connector X3 is not properly connected
the compression reaches the
current limit.

The return function does not - The signals from the switches do not reach
work. connector X3 on ECS board

The ECS moves too slow. - Adjust the speed reference from trimmer pot
R37 or R59 (down / up)

The ECS moves too fast. - Adjust the speed reference from trimmer pot
R37 or R59 (down / up)

C-arm brake does not work. - Check the microswitch on the side of the ECS
mechanism
- Check the photointerrupters and the metal plate
sliding to the photointerrupter slot

ECS does not move from - Check pulses from compression motor encoder
ECS pedal.

October 2002 5-22 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6 REPAIR INSTRUCTIONS

6.1 FUSE RATING / LOCATION CHART


LOCATION RATING PURPOSE
EXTERNAL: AUTOMATIC 15 A/ 250 V Fuse to the 220 line
FUSE OR F1, F2 15 ASB/ 250 V Fuses to the 220 line
SUPPLY BOARD:
F1 4 ASB/ 32 V 24 VAC prior to the
regulation, for AMD, C, Z
and the break
F2 5 ASB/ 32 V 18 VAC prior to regulation
for the CPU and lamp
F3 500 mAF/32V 15 VAC prior to regulation
for the AMPLIFIER board
FILAMENT BOARD: (PULSE board*)
F1 6 A T/ 32 V 24 VAC to regulators
F2 6 A T/32 V 24 VAC to regulators
F3* 1 ASB/32 V -15 VDC to PULSE board
F4* 1 ASB/32V +15 VDC to PULSE board
INVERTER BOARD: (POWER board*)
F1 20AF/500V Dcinput to INVERTER board
(16AFF/500V*) DC-input to POWER board
F2* 16 AFF/ 500 V DC-input to POWER board
F3* 1 ASB/ 32 V 15 VAC to protection circuit
ANODE MOTOR DRIVER:
F1 10 ASB/ 250 V 220 VAC to rotor

Table 6.1. Fuse rating & location table (* Old type generator)

Fuse rating should always be higher than the voltage in the circuit to be
protected.

WARNING! Always install the proper fuse into the unit. Failure to do so jeopardizes
the safety of the patient, operator and system.

31687 6-1 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.2 COLUMN REPAIR

NOTE! The counterweight is heavy. Be careful when handling it.

NOTE! Remember to insert the counterweight assembly pin in the column before
starting the repair.

6.2.1 Tilt down the column


Remove the power supply plug. Remove the C-arm. Remove the two bolts on
both sides of the floor plate (A). Open the two wall bolts and take slowly down
the column. If the unit is installed on a freestanding base plate, remove the
sidepieces by opening the screws inside. Do not let the column fall free on the
floor. Put some soft material below the column to avoid damage.

6.2.2 Replacing the floor plate


When the column is lying on the floor, remove the two bolts (B) from the floor
plate on both sides of the column.

220 0
40

Figure 6.1 Floor plate Figure 6.1b Performa floor plate

October 2002 6-2 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.2.3 Replacing the Z-motor


Insert the counter weight pin and drive the carriage up until the counterweight
rests against the pin. Lift up the carriage (this requires two persons while
otherwise the c-arm may turn and the carriage fall down causing damage to the
unit) and put two boards (25 x 100 x 1400 mm for the short column, 25 x 100
x 1620 for the long column) under the carriage sides. Secure the boards straps.
Loosen the four lower bolts, and lift of the column cap.

NOTE! If no boards available, take down the column as described in 6.2.1

CAUTION! Be careful not to drop any parts into the column.

Remove the two screws on top of the column. Lift up the cap and remove it
from the top of the column. Be careful not to scratch the column. Unplug the
connector. Loosen the two lock screws (C) in the pulley. Unscrew and remove
the motor assembly bolt (A) of the pulley. Remove the two screws (B) which
attach the motor plate to the cap, see figure 6.2. Replace the Z-motor.

To achieve correct shaft alignment the motor assembly bolt (A) must be
tightened (max. 15Nm) holder screws. Check the alignment by running the
motor with slack wires before installing the cap assembly.

31687 6-3 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.2.4 Replacing the column cap


Remove the cap of the column as described in 6.2.3. Remove motor as
described in 6.2.3.

Remove the cable coming through the top of the cap. Replace the old cap.

Place the wires first on the new pulley, and fit the pulley to the new column
cap. Push the motor assembly axle (D) in place against the motor axle, and
tighten the screws so that they go into the holes of the motor assembly axle
(D). Fasten the motor plate to the cap with screws (B), see figure 6.2. Screw
the motor assembly bolt (A) in place by using torque, max 15 Nm (133 in lbs.).
Check the assembly of the wires and the column cap and lift the column so that
the counterweight pulls the wires, and position the column cap in its place.
Fasten the four bolts, and connect the motor plug.

Figure 6.2 Column cap

October 2002 6-4 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Figure 6.2b New version of Column cap

In the new version of column cap the tightening torque of nut 5477 is 4Nm at
the factory.

The washers 3587 are spring washers and they are installed so that last four
washers are turned around in pairs of two facing each other (pattern would be
>><><>) making a spring package.

31687 6-5 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.2.5 Replacing the mechanical stopper


Tilt down the column as described in 6.2.1 and remove the top cover. Remove
the screws of the limiter, located in the back of the column, and remove the
limiter. Tighten the new limiter screws, and replace the column cap.

6.2.6 Replacing the wires


Take down the column as described in 6.2.1 and remove the column cap. Be
careful not to scratch the column. Release the wires by moving the carriage
upwards. Unplug the connector. Remove the mechanical stopper as described
in 6.2.5. Pull out the counterweight of the column and detach the wires.
Loosen the two lock screws (C) in the pulley. Unscrew and remove the motor
assembly bolt (A). Remove the two screws (B) which attach the motor plate to
the cap and remove the motor, see figure 6.2.
Remove the motor assembly axle (D).

NOTE! The motor assembly axle (D) moves through the motor end only.

NOTE! When replacing the wires, pay attention to the order of the wires in the
fastening plate.

Pull the carriage out of the column to


facilitate changing the wires and checking
the assembly. Remove back cover of the
carriage. Remove Inverter board. Remove
the assembly plate of the power unit.
Open the nut that fastens the wire
fastening plate and replace the wires.

Place fastening plate, assembly plate of


the power unit, Inverter board and
assemble the column cap as described in
6.2.4.Push the carriage and counterweight
back, and screw the mechanical stopper
and the column cap back in place. Figure 6.2c Sliding unit of the
carriage

October 2002 6-6 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.2.7 Replacing the counterweight


Tilt down the column as described in 6.2.1 and remove the column cap and the
limiter. Pull the counterweight out of the column so that you can release the
wires. After changing the counterweight check the assembly again, carefully to
avoid damages.

Figure 6.3. Counterweight

31687 6-7 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.3 POWER UNIT REPAIR

6.3.1 Removal and assembly of the rear cover


Unplug the mains cord. To remove the rear cover, unscrew the four screws on
it, two at the top and two at the bottom.

6.3.2 Replacing the POWER / INVERTER board


Unplug the mains cord. Remove the rear cover and disconnect all cables from
the power/ inverter board. Remove the plastic nuts from the corners of the
power board, and replace the board. Go to chapter 5, ALIGNMENT &
ADJUSTMENT and check the kV, mA, preheat, and protection circuit.

6.3.3 Replacing other parts


The capacitor is installed with a plastic holder. Open the plastic nut and replace
the capacitor. Beware of electrical shocks.

Figure 6.4. Inverter board

October 2002 6-8 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.4 CARRIAGE REPAIR

6.4.1 Removal and assembly of the side covers


Loosen the screw holding the side cover and take the cover out. When
reassembling, first put the upper side of the cover on its holders.

6.4.2 Replacing the mains transformers


Remove the power supply plug. Remove the side covers. Remove all cables
connected to the transformer, and open the four screws at the bottom of the
transformer base. After replacing the transformer, check the cables carefully.

Figure 6.5. Carriage

31687 6-9 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Figure 6.5b. Performa carriage

October 2002 6-10 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.4.3 Replacing the boards on the electronic plate


Open the screws holding the corners of the electronic plate. Pull out the whole
electronic plate to the left side. Remove the cables and nuts in the corners of
the board, which you intend to replace. Check the cables and assembly
carefully. If you replace the PULSE board go to ALIGNMENT &
ADJUSTMENT section and check kV, mAs, and preheat. If you replace the
C&Z drive check compression force. If you replace the CPU board, EE-prom
or the software go to Setup chapter on 2.13 in the Installation Instructions
manual.

6.4.4 Replacing the SUPPLY board


Remove all cables and screws holding the board and replace it. Check the
cables and assembly carefully. After changing, go to ALIGNMENT &
ADJUSTMENT section and check light intensity.

Figure 6.6. Electronic plate

31687 6-11 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.4.5 Replacing the magnetic brake and its assembly


Remove the side covers and pull out the electronic plate with the cables.
Disconnect all cable connectors and pull loose connectors through the center
axis on the C-arm.

Remove the four nuts behind the brake, and remove the other parts in the
following order:
- locking plate with the counter plate for magnetic brake.
- the disc springs
- the magnetic brake connected to the assembly plate.

See figures 6.7a, b and c.

Unscrew the friction plate and replace it with a new one. Replace the other
parts: magnetic brake, 4x7 pairs of disc springs, locking plate, and finally the
locking nuts.

Tighten the four nuts so that the locking plate is completely shut. Then
unscrew them half a turn.

The brake is adjusted using a 20 VDC power supply connected to the brake.
The adjustment is performed either by tightening or loosening the holding nuts.
The brake should be adjusted in such a way that the holding force measured
from the cassette holder should be less than 20 N (4 lbs.) with the brake off
and greater than 110 N (24 lbs.) with the brake on.

C-arm Brake adjustment

Check that the brake has 110N (27lbs) of


holding force measured at the cassette holder. If
not, open the side covers of the carriage and
tighten the nyloc nuts (3).

If the brake does not release properly, open the


nuts (1), and adjust the space (4) with the
screws (2). The space should be about 0.1 - 0.4
mm.

Figure 6.7a

October 2002 6-12 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Figure 6.7b

Figure 6.7c

31687 6-13 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Figure 6.7d Performa

October 2002 6-14 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.4.6 Replacing the MGF DISPLAY board


Open the four screws at the bottom of the display box and remove the rear
cover of the box. Open the holder screws, remove the cables and replace the
board.

Figure 6.8. Display board Figure 6.8b Performa display board

31687 6-15 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.5 C-ARM REPAIR

6.5.1 Removal and assembly of the C-arm covers


Remove the four screws on top of the top cover. Lift up the cover and remove
it. Remove the two screws near the collimator opening to remove the lower
cover. Turn the C-arm in a lateral position to remove the back cover of the
C-arm. Disconnect the cables for pcb boards and

6.5.2 Replacing the handgrips


Remove the C-arm and the back cover. Disconnect the handgrip buttons and
unscrew the handgrips. Attach new handgrips and connect the buttons. Replace
the C-arm and the covers and check the assembly.

6.5.3 Replacing the brake switch


Unscrew the switch, disconnect the wires, and replace the switch.

6.5.4 Replacing the vertical drive control switch


The switches are located in the lower cover of the C-arm. Therefore to remove
them, remove first the upper cover and then the lower cover and push the
switch out. Install a new one and test the movement direction.

October 2002 6-16 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Figure 6.9. C-arm

31687 6-17 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Figure 6.9. Performa C-arm

October 2002 6-18 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.5.5 Replacing the x-ray tube


Remove the C-arm covers and diaphragm, and disconnect the plugs of the tube
and unscrew the four bolts below the tube(1/4 x 20 UNC). The bolts inside the
collimator require a long hexagonal wrench. Go to ALIGNMENT &
ADJUSTMENT section and check kV, mAs and preheat.

6.5.6 Replacing the high voltage unit


Disconnect the ground wires and the plugs of the high voltage unit, and loosen
the four screws, two on each side of the unit. After replacing it, check the
assembly and the ground cables carefully. Go to ALIGNMENT &
ADJUSTMENT section and check kV, mAs and preheat.

6.5.7 Replacing the AMD board


Unplug the mains cord. Remove the upper cover of the C-arm and disconnect
all cables leading to the anode motor drive board. Replace the board, reconnect
all cables, and check the assembly. Go to ALIGNMENT & ADJUSTMENT
section and check the run voltage

NOTE! ALIGNMENT & ADJUSTMENT section is in the Installation Instructions


manual.

Figure 6.10. Anode motor drive board

31687 6-19 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.5.8 Replacing the compression motor


Remove the upper cover of the C-arm and unscrew the motor. Replace it and
connect the motor plug. Go to ALIGNMENT & ADJUSTMENT section and
check the compression force and adjust if needed.

6.5.9 Replacing the compression system


Remove the C-arm and the back cover. Unscrew the four bolts holding the
compression unit and remove it. Before assembling and tightening the bolts of
the new unit, be sure that the beveled gears are perfectly placed, and also that
the trapezoidal shaft has no slack. Go to ALIGNMENT & ADJUSTMENT
section (in Installation Manual) and check the compression force and adjust if
needed. On PERFORMA, RT/IQ make sure that the cable for the strain gages
is properly placed and does not enter the compression mechanism.

6.5.10 Replacing the compression thickness measurement potentiometer


Remove the c-arm. Remove the high voltage transformer. Remove the four
screws on the potentiometer (two up and two down). Change the
potentiometer. Check also the potentiometer glide located on the compression
system mechanical clutch bracket.

6.5.11 Replacing the compression force measurement system (strain gages)


Remove the C-arm and the back cover. Unscrew two bolts in the gear box and
two bolts in the bearing block holding the compression unit and remove it. In
the gear box detach the clutch and the cogwheel and release the shaft. Replace
the part with strain gages and attach the shaft again. In the gearbox attach the
cogwheel and the clutch. Install the compression mechanism in to the C-arm
with the four bolts. Tighten the two bolts in the gearbox. Tighten the two bolts
in the bearing block with your hand only and remove vertical slack from the
shaft by pushing the bearing block towards the gearbox. Then tighten the two
bolts in the bearing block. Install the C-arm. Go to the ALIGNMENT &
ADJUSTMENT section (in Installation Manual) and check the compression
force and adjust if needed. Make sure that the cable for the strain gages is
properly placed and does not enter the compression mechanism.

Calibrate the system with Pr 55 Cfo program in the service mode program
menu.

6.5.12 Replacing the C-arm angulation measurement system (trim pot)


Unscrew the two screws holding the trim pot and change the trimmer. Make
sure with an ohm meter that the trimmer is exactly in the center position when
the c-arm angle is zero. Turn the c-arm carefully and check that the trimmer
pot does not reach it's end before the c-arm hits the mechanical stops.
Calibrate the angle measurement with Pr 54 Can program in the service mode
program menu.

October 2002 6-20 31687


Performa, Alpha IQ, RT, ST, III Service Manual

Figure 6.11. Old Compression system

31687 6-21 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Figure 6.11b. New Compression system

October 2002 6-22 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.5.13 Replacing the cassette holder base cover


Remove the cassette holder or Bucky device and open the screws of the cover.
Disconnect the plugs from the cover, and replace the cover, the detector, or the
CAS-OPTO board in the cover.

6.5.14 Replacing the cassette holder base


Remove the base cover and take the cables out of the base. Unscrew the two
attachment bolts and replace the base.

Figure 6.12 Cassette holder base

31687 6-23 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

Figure 6.12b Cassette holder base with Optipoint (Vectre in USA)

October 2002 6-24 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.5.15 Replacing the collimator


Remove the upper, lower cover of the C-arm and the collimator plate.
Disconnect the lamp cable. Unscrew the four collimator bolts on both sides
and replace the collimator. To remove the lamp base, disconnect the cable,
open the lock screw and unscrew the base. Go to the ALIGNMENT &
ADJUSTMENT section and adjust the x-ray field and light field after
replacement.

6.5.16 Replacing the light field bulb


Turn the unit off and leave it for a while to allow the bulb to cool. Remove the
collimator plate and the old bulb by pulling out. When fitting the new one,
avoid touching the bulb with your fingers. Go to the ALIGNMENT &
ADJUSTMENT section and check the light field alignment.

Figure 6.13 Replacing the collimator

Figure 6.14 Collimator

31687 6-25 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

6.5.17 Replacing the beam filter mechanism


Remove the upper and lower cover from the c-arm. Remove the X-ray tube.
The beam filter on Alpha III is attached directly on the lead frame. When
replacing the filter change the whole lead frame.

On PERFORMA, RT/IQ both Molybdenum and Rhodium filter are molded in


the plastic filter frame. When replacing the filter change the whole plastic
frame.

6.5.18 Replacing the compression display board


Remove the upper and lower cover from the c-arm. The compression display
board is located on the lower cover. Remove the attachment screws and
replace the board.

6.5.19 Replacing the detector display board


Remove the bottom plate of the detector display housing. Remove connector
X1 from the detector display board and open the screw attaching the board to
the housing. Open the white plastic on the cassette holder base. Remove the
cable connectors from the two microswitches on both sides of the AEC
detector. Replace the detector display board.

October 2002 6-26 31687


Performa, Alpha IQ, RT, ST, III Service Manual

6.5.20 ECS repair

31687 6-27 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

October 2002 6-28 31687


Performa, Alpha IQ, RT, ST, III Service Manual

31687 6-29 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

October 2002 6-30 31687


Performa, Alpha IQ, RT, ST, III Service Manual

7 SCHEMATICS
7.1 Main wiring diagram .............................................................................31812-1J_

7.2 Supply board ...................................................................... 32030-3S_, 32030-3K_

7.3 CPU board .......................................................................... 32245-3K_, 32245-3S_

7.4a C & Z Driver board (soft comp.) ................................... 33035-3K_, 33035-3S_

7.5a Display board ....................................................................................... 91420-3B_


7.5b Display board .......................................................................................91419-4C_

7.6a Cas-opto board OPTIPOINT ........................................... 33280-3_, 33280-3K_


7.6b Cas-opto board ....................................................................................92983-4C_
7.6c Cas-opto board ..................................................................................... 92984-4B_

7.7a Bucky control board ............................................................................92670-4C_


7.7b Bucky control board ............................................................................ 92744-3B_

7.8 Anode Motor Driver board ............................................... 32305-3S_, 32305-4K_

7.9 High voltage unit .....................................................................................32370-4S_

7.10a Cassette holder sense .........................................................................31438-4K_


7.10b Cassette holder sense ..........................................................................31439-4S_

7.11 Compression Display board ............................................ 32010-3S_, 32010-4K_

7.12a AEC Display board OPTIPOINT ................................. 33275-4S_ 33275-4K_


7.12b Detector Display board ................................................. 32005-4S_, 32005-4K_

7.13 Inverter board .................................................................. 32220-3S_, 32220-3K_

7.14 Filament Control board .................................................. 32225-3S_, 32225-3K_

7.15 C-arm Control board ...................................................... 32310-3S_, 32310-3K_

7.16 C-arm Display board ....................................................... 32335-3S_, 32335-4K_

7.17a SMD AEC board OPTIPOINT .................................... 33270-3S_, 33270-3K_


7.17b AEC board ..................................................................... 32315-3S_, 32315-4K_
7.17c Phototimer board ........................................................... 32345-3S_, 32345-4K_

7.18 5 V Power Supply board OPTIPOINT (Only ST) ........ 32962-4S_, 32962-4K_

7.19 Sensor board OPTIPOINT ............................................. 32735-4S_, 32735-3K_

31687 7-1 October 2002


Service Manual Performa, Alpha IQ, RT, ST, III

7.20 ECS Driver Board ............................................................ 32720-3S_, 32720-3K_

7.21 Wiring diagram ECS 200......................................................................32862-2J_

7.22 Detector Optipoint................................................................................33285-3K_

7.23 Software Block Diagram ................................................. 37267-2L_, 37268-2L_

October 2002 7-2 31687


Performa, Alpha IQ, RT, ST, III Service Manual

THIS PAGE HAS INTENTIONALLY BEEN LEFT BLANK.

31687 7-3 October 2002


MANUFACTURER:
Instrumentarium Imaging
P.O.Box 20, FIN-04301 Tuusula, Finland
Tel. + 358 10 394 6500 • Fax + 358 10 394 6501
info@fi.instrumentarium.com • www.InstrumentariumImaging.com

INSTRUMENTARIUM IMAGING OFFICES OUTSIDE OF FINLAND:


Instrumentarium Imaging Inc.
300 West Edgerton Avenue, Milwaukee, Wisconsin 53207, USA
Tel. +1 800 558 6120, +1 414 747 1030 • Fax +1 414 481 8665
info@usa.instrumentarium.com • usa.instrumentarium.com

Instrumentarium Imaging France S.A.R.L.


4, Avenue des Roses, 94386 Bonneuil Sur Marne Cedex, France
Tel. +33 1 43 39 51 51 • Fax +33 1 43 39 75 75
instrumentarium.imaging@wanadoo.fr

Instrumentarium Imaging Italia S.R.L.


Via Cassanese, 100
20090 Segrate (MI), Italy
Tel. +39 02 21 30 28 1 • Fax +39 02 21 30 28 60
instrumentarium@it.instrumentarium.com

Instrumentarium Imaging Singapore


152 Beach Road
#12-03A Gateway East
189721 Singapore
Tel. +65 6391 8600 • Fax: +65 6396 3009
singapore@sg.instrumentarium.com

Instrumentarium Imaging Ziehm GmbH


Isarstrasse 40, D-90451 Nuremberg, Germany
Tel. +49 911 642 070 • Fax +49 911 642 0739
info@ziehm-eu.com

Instrumentarium Imaging Ziehm Inc.


4181 Latham Street Riverside, CA 92501, USA
Tel. + 1 909 781 6660 • Fax +1 909 781 6457
mail@ziehm.com
1 2 3 4 5 6 7 8
MERKKI MUUT.NO MUUTOS PIIRT. TARK. HYV.

GND BRAKE SWITCH


9.9.99

BRAKE SWITCH
MICRO SWITCH E 31812 VAIHDETTU R14 100K->121K JaT

MICRO SWITCH

GND BRAKE
25.4.96 13.6.96 14.6.96
A D 31281 LISATTY LIITIN X4 A

GND AMD

GND extra
LTu HK JJS

GND C&Z

Vin AMD
Vin C&Z
BRAKE

GND CPU
GND

Vin CPU
PROJ LAMP1

PROJ LAMP2

PROJ RELAY

GND AMPLIFIER
+ AMPLIFIER

- AMPLIFIER
X2
X4

X3
14

15

11

10

13

12
2k7
1

1
10k R20 C10

TP2

TP5

TP6
TP9
R22

TP7
B 10k 100nF B

1N4148
TP8

V11
R16

2
BYV27 R21

10k
10k

R26

10k
BT 139

R25

R27
100k
TP1

CNY17 IV

ICD1
R24

TP4
10k
V10

1N4148

V12
V7

R15
10k

4
100k
BDX33A

10k

R25
R7
R11

C11
C3
10k

+
C 100nF C

6
V9

13
39 V

V8 4700 F

4
C12

ICA1
10k 10k
100nF
C6

SGen
Csync

Vsync
100nF R5 R6

SStart
2k2/5W

1N4004
-Vcc

1
14

VCC

V5
R10 C4 C5

Range

NC
+ +

Rmin
Cvco
10k

R17
TDA1185A

1k21
11
FB
C2 470 F 470 F
R4 +

C1 C9 4700 F

10
12
9

NC 3
NC 5
8
+ +
H2
2200 F R14 1 F 4k32
D H3 D
H1 R18 R2 4k75
5K62 121k

1N4004
100k R3

V6
R1 R12
R13 R19

V2

1N4004
10k 10k 10k

1N4004
V3

V4
V1

C8
+
C7
+ 10 F

100 F

F3
0.5AT
E E
4AT

F1

F2
5AT
3

6
1

X1

X1

X1

X1
X1

X1

X2

15 VAC
18 VAC
GND
24 VAC

YLEISTOLERANSSI
PIIRT. 23.09.92 JJS LAITE
MGF - 101/110 SFS 4011
F F
TARK. 27.05.93 MTL SUHDE
LIITTYY
HYV. 27.05.93 TI :
NO
INSTRUMENTARIUM NIMI
SUPPLY CARD
imaging V 2.1
MGF 32030-3SE

1 2 3 4 5 6 7 8
+5V 1 X1 RN6
ICD8 GND 23 9X10k 1
*KBDCS 22 1 RN27 6 X5

C111
CS/ B3 2 B3 9

100n
*WR 11 WR/ B2 3 4 B2 8

C110

C112
*RD 10 4 IRQ

100n
RD/ IRQ B1 5 6 B1 7 5 6

100n
RST 9 RST SHIFT 36 N/C B0 7 8 B0 6 +5V 4 X5
3MHZ 3 ICD9
+5V CLK CNTL/STB 37 1 ABB 4X470R A3 5 ICD2
RN2
N/C 11
A0 21 A0 BD 23 EN1 A2 4 EXPO 4 3 1 2 N/C 12
19 EN2 A3 1 RN28 2 EXP 9
ICA1 L4962 L1 A1 3 74HC14 4X10k
A2 3 4 A0 2 ICD2 SERIAL 5
X9 2 Vin CPU 7 2 D7 19 D7 B3 28 2 18 B3 A1 5 6 RN2 LATCH ENA 3
N/C 8 1 N/C D6 18 29 4 16 B2 2 1 3 4
150uH D6 B2 A0 7 8 CLOCK 2
N/C 9 3 N/C D5 17 D5 B1 30 6 14 B1 74HC14 4X10k DATA 1
N/C 16 6 N/C D4 16 31 8 12 B0 4X470R E ICD2
D4 B0 EXP1 GND 14
15 D3 15 D3 A3 24 11 9 A3 SERI 10 11 EXP1 13
D2 14 D2 A2 25 13 7 A2 RN3 9X10k 1 74HC14 C119
14 5 4 11 10 D1 13 26 15 5 A1

475R
13 12 D1 A1

R13
D0 12 D0 A0 27 17 3 A0 100n

475R
MPTE-5
3 4 2 5 6 7 8 9 10

R34
74HCT244

+
470u
C39

C37
ICD10 *PPI1BIN

22u
D3

C46
N/C 8

1000u

100n
D1

+
MPTE-36
R7

C26
1 EN1 R6 7 *PPI1BOUT
R6 7 RN21

BYV27-100
H1 19 RD6 8 R6 21 X1

RB
EN2 R5 6 R5 EXP1 22 X1

D2
+

4k75
RD5 5 6 R5 20

R11
2u2

R9

R8
C34

2N2

15k
4k32
C36
R4 5 R4 EXP1 GND 25
+5V 2 18 N/C R3 2 35 RD4 3 4 R4 19 ICD5
RD6 4 16 R6 R3 SL3 RD3 1 2 R3 18 19 26
R2 1 R2 SL2 34 EN
RD5 6 14 R5 R1 39 33 4X10k
RD4 8 12 R4 R1 SL1 1 X4

33n

330p
C20

C25
R0 38 R0 SL0 32 RD2 1 RN22 2 R2 17 X1
RD3 11 9 R3 LATCH EN 11 9 1 RN20 2 5

C114
RD2 13 7 R2 RD1 3 4 R1 16 CLOCKO 13 7 3 4 23

100n
X9 1 GND RD1 15 5 R1 8279-5 ICD6 A RD0 5 6 R0 15 DATAO 15 5 5 6 24
RD0 17 3 R0 1 EN N/C 7 8 N/C AECFRQI 3 17 7 8 AECFRQ 7
ICD28 74HC244 4X10k 74HC244 4X470 2
*PPI1AIN 1 8
+5V +5V EN1 2 18 7 RN25 8 SL3 13 X1 RN17 9
19 EN2 4 16 5 6 SL2 12
1 9X10k 11
6 14 3 4 SL1 11 CEND 15

C115
WDRFS 2 18 D7 8 12 1 2 SL0 10

100n
(10u)

22U
PROJRELO 4 16 D6 S2 14

C70
74HCT244 4X10k 18 X4

+
AUTEXPO 6 14 D5 ICD4 10 9
+5V +5V ICD11 16 S3 24
BUCKYONO 8 12 D4 15 N/C 1 RN26 2 CASBUCO 8 9 5
RN2
6
19
*ANENAO 11 9 D3 3 4 DMUX 0 3 4 20
2 1 14 N/C 74HC14 4X10k 21
(4u7) ICD37 16 ANRUNO 13 7 D2 2 13 N/C 5 6 N/C ICD4
C51 1 2 ZCOM 22
C69 *ANSTARTO 15 5 D1 1 N/C 7 8 N/C RN2

1N5818

1N5818
1 2 12 BHOMEO 6 5

D8

D9
+5V (4u7) C65 *ANSTOPO 17 3 D0 3 7 8 6
R16

+ 3 6 6 11 N/C 4X1k
4k75

100n
C66 22U & 4 74HC14 4X10k CASBUC 12
4
22U
+ RS232 PORT1 74HC244 IB 4 5 10 N/C BHOMEO 10
"RESET" *RST + 5 5 9 N/C

C116
ICA2 8 (10u) 6

100n
S1 22U R35 ICD29
7 S 6 RST TXD 11 14 TXD1 1 X8 ICD12 7 7 N/C WDRFS
3 4 2 *PPI1BIN 1
RI RO 5 H2 RXD 12 13 100R RXD1 2
19
EN1 *RST 1 R 74HCT138 8
3 C2 Vr 1 10 7 GND 3 EN2 *PPI1AOUT 11
1N4148

C1 ICA4
1 2 N/C 9 8 N/C 1 16 N/C 20 X3
D4

TL7705 4 ZDIRO 2 18 D7 7 RN23

P6KE15 P6KE15
D7 3 1D 2 WDRFS 2 15 8

D11
MAX232 15 ZRUNO 4 16 D6
C55

D6 4 5 PROJRELO 3 14 5 6 PROJREL 7 X2
100n

(SP/AD232) CRUNO 6 14 D5
+

D5 7 6 *ANENAO 4 13 3 4 *ANENA 25
2U2

CDIRO 8 12 D4
C52

R15

D4 8 9 ANRUNO 5 12 1 2 ANRUN 24
1K5

1K5
R17

CMANO 11 9 D3

D10
D3 13 12 *ANSTARTO 6 11 4X470
NOCON0 13 7 D2 D2 14 15 *ANSTOPO 7 10 1 RN24 2 *ANSTART 23 X2
NOCON1 15 5 D1 D1 17 16 8 9 N/C 3 4 *ANSTOP 22
NOCON2 17 3 D0 D0 18 19 5 6 BUCKYON 27
C4 ULN2003
+5V 74HC244 BUCKYONO 7 8 AUTEXP 26
74HC273
1uF
221k

AUTEXPO
R4

ICD13 4X470
*RST 1 R
1 X10 +5V *PPI1BOUT 11 +5V 1 X2 +5V 1 X3
+5V C1
V5 R5 1 RN18 2 3
2 D7 3 2 ZDIRO ZDIR 5 X3 GND 2 X2 GND 2
ICD14 28 ICD15 28 1D 3 4
2K D6 4 5 ZRUNO ZRUN 6 30 4

C80
1N4148 ICD16 28 5 6

100n
3 A0 10 A0 10 A0 A0 10 A0 D5 7 6 CRUNO CRUN 7 9 16
C53 A0 7 8
A1 9 A1 9 A1 A1 9 A1 D4 8 9 CDIRO CDIR 8 15 CCOM 18
1N4148

C81
A2 8 A1 A2 8 A2 8 1 RN19 4X470

100n
V6

4 A2 A2 D3 13 12 CMANO 2 CMAN 9 17
470n A2
A3 7 A3 7 A3 A3 7 A3 D2 14 15 NOCON0 3 4 CSPEED 14 33
A4 6 A3 A4 6 A4 6
A4 A4 A4 D1 17 16 NOCON1 5 6 D16EXP1 13 3 RN17 RN12
A5 5 A5 5 A5 A5 5 A5 D0 18 19 NOCON2 7 8 D16RSENA 15 4 9X10k 9X10k
WDRFS A6 4 A5 A6 4 A6 4
1 1

14 Z1 A6 A6 A6 74HC273 4X470
A7 3 A7 3 A7 A7 3 A7
N/C 1 12Mhz 8 12MHz A8 25 A7 A8 25 A8 25 ICD17 RN4
A8 A8 A8 2 1 8 7 8 9 3 X4
A9 24 A9 24 A9 A9 24 A9 ICD18
1 2
7 A10 21 A9 A10 21 A10 21 4 3 3 4 4
A10 A10 A10 EN1 1 6 5
A11 23 A11 23 A11 A11 23 A11 19 *PPI1C 4X10k RN11
A12 2 A11 A12 2 A12 2 EN2 8 9 1 2 CASNOR 19 X2
A12 A12 A12 10 11 CHSMALL 18
A13 26 4 3
A13 *RAMCS 20 CS1 *ROMCS 20 CS1 D7 18 2 12 13
ICD1 40 A14 27 26 26 D6 16 4 RN5 4X10k
12MHz 19 18 N/C ICD27 20 A14 CS2 CS2 74HC14 1 2 LOW 7 X5
X1 X2 A15 1 A15 *WR 27 WE *WR 27 WE D5 14 6 MA 10
RST 9 29 *PSEN ICD2 D 3 4
R PSEN 1 20 CS *RD 22 OE *RD 22 OE D4 12 8 GENOK 8
31 30 ALE EN 8 9 5 6
EA ALE ALE 11 *PSEN 22 WR D3 9 11 N/C N/C
C1 D2 7 13 74HC14
7 8

11 AD0 11 AD0 R33 LIMN 10 X3


P1 P0 D0 11 AD0 D0 D0 D1 5 15 ICD2 C 4X10k 4 2 3

C82

C83

C84

C85

C86

C87

C88

C89
CASBUCO 1 0 0 39 AD0 AD0 3 1D 2 A0 D1 12 AD1 D1 12 AD1 D0 3 17 6 5 R30

100n

100n

100n

100n

100n

100n

100n

100n
EXPO 2 38 AD1 AD1 4 5 A1 D1 12 AD1 13 AD2 13 AD2 10k 10k
1 1 D2 13 AD2 D2 D2 74HC244 74HC14
BHOMEO 3 2 2 37 AD2 AD2 7 6 A2 D3 15 AD3 D3 15 AD3 1 RN6
LATCH EN 4 36 AD3 AD3 8 9 A3 D3 15 AD3 16 AD4 16 AD4
3 3 D4 16 AD4 D4 D4 ICD20 9X10k
CLOCKO 5 4 4 35 AD4 AD4 13 12 A4 D5 17 AD5 D5 17 AD5 RN7
DATAO 6 34 AD5 AD5 14 15 A5 D5 17 AD5 18 AD6 18 AD6 2 1 1 2 CUP 17 X3
5 5 D6 18 AD6 D6 D6 ICD19 4 3 CDOWN 19
SERI 7 6 6 33 AD6 AD6 17 16 A6 D7 19 AD7 D7 19 AD7 1
3 4

IRQ 8 32 AD7 AD7 18 19 A7 D7 19 AD7 EN1 6 5 5 6 12


7 7 EN2 19 *PPI2A 8 9 11
P3 P2 74HC373 10 27512 14 6264 14 2864A 14 7 8

RXD 10 21 A8 10 11 RN8
0 RXD 0 D7 18 2 4X10k
TXD 11 22 A9 CBUS 12 13 1 2 BON 17 X4
1 TXD 1 D6 16 4 CEND 16
N/C 12 23 A10 74HC14 3 4
2 INT0 2 DBUS D5 14 6
5 6 ZUP 25
N/C 13 3 INT1 3 24 A11 D4 12 8 ICD4 B
N/C 14 25 A12 4 3 7 8 ZDOWN 26
4 T0 4 ABUS ICD21 D3 9 11
AECFRQI 15 5 T1 5 26 A13 D2 7 13 74HC14 4X10k 3 4 5 6
ICD22 *IO 19

7
5
3
1
*WR 16 6 WR 6 27 A14 G3 D1 5 15 ICD4 A

RN14
*RD 1

4X10k
C90

C91

C92

C93

C94

C95

C96

C97
*RD 17 7 RD 7 28 A15 ALE 1 3EN1[BA] D0 3 17 2 1

100n

100n

100n

100n

100n

100n

100n

100n
*PSEN 2 23 *RAMCS 3EN2[AB] 9X10k 1

8
6
4
2
80C32 20 74HC244 74HC14
*WR 3 22 *ROMCS AD0 2 18 D0 RN17 B
*RD 4 21 *KBDCS 1 ICD24
ICD7 A 2 RN10
A15 5 20 *PPI1A *WR 1 1 AD1 3 17 D1 2 1 1 2 BLE 8 X2
A14 6 19 *PPI1B 3 *PPI2CL ICD23 4 3 CHR 14
2 AD2 4 16 D2 3 4

PDBUS
A13 7 18 *PPI1C EN1 1 6 5 CHL 16
*PPI2C AD3 5 15 D3 5 6
A12 8 17 *PPI2A 74HCT32 EN2 19 *PPI2B 8 9 CHON 32
ICD7 B AD4 6 14 D4 7 8
A1 9 16 *PPI2B AD5 7 13 D5 10 11 RN11 *RUNOK 21
A0 10 15 *PPI2C 4 1 6 *PPI2CH D7 18 2 12 13 4X10k 5 6 *STARTOK 20
*RD 5 AD6 8 12 D6 D6 16 4
N/C 11 14 *IO AD7 9 11 D7 74HC14 8 7
N/C 13 74HCT32 D5 14 6 13 X4
74HC245 D4 12 8 ICD4 F RN4 4X10k
ICD7 C 12 13 5 6 14
22V10 *WR 9 1 D3 9 11
8 *PPI1AOUT 7 8
*PPI1A 10 D2 7 13 74HC14 5 6 7 4 3 2 A
D1 5 15 ICD4 E 4X10k RN17 R32

C100

C101

C102

C103

C104

C105
C98

C99
74HCT32 D0 3 17 10 11 10k

100n

100n

100n

100n

100n

100n

100n

100n
ICD7 D 9X10k
12 1 74HC244 74HC14 1
11 *PPI1AIN RN12 9X10k
*RD 13
+5V 74HCT32 ICD5 A ICD3 D RN13
ICD6 B EN 1 *PPI2CH 8 9 1 2 FILTER1 5 X2
*PPI1B 19 74HC14 ICD3 C 3 4 ANG LN 6
EN 6 5 5 6 FILTER2 10
D7 18 2 ICD3 B 74HC14 7 8 CD DATA IN 11
*WR 11 9 *PPI1BOUT D6 16 4 4 3 N/C 12
4X10k
C22

C60

C15

C17

C18

C19

C30

C31

C32

C33

C35

C40

C41

C42

*RD 13 7 *PPI1BIN D5 14 6 74HC14 ICD3 A N/C 13


100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

7
5
3
1
D4 12 8 2 1 N/C 28

RN15
15 5 N/C

4X10k
17 3 N/C 74HC244 74HC14 N/C 29

C106

C107

C108

C109

8
6
4
2
2N2
74HCT244 ICD26

100n

100n
10N
ICD3 E
11 10 N/C *RST 1 R
+5V *PPI2CL 11
74HC14 C1
ICD25 ICA3 1 RN29 2
ICD3 F CTRDIV16 D7 3 2 1 18 3 4 ARELOFF 37 X2
13 12 N/C 1D
1 5CT=0 D6 4 5 2 17 5 6 ARELON 38
74HC14 9 M1 D5 7 6 3 16 7 8 *ANG CS 39
ICD2 F M2 3CT=15 15 N/C D4 8 9 4 15 4X470 ANG CLK 40
C117

C118
C43

C44

C45

C48

C47

C49

C59

C58

C57

C56

C50

C61

13 12 N/C 10 D3 13 12 5 14 CD LATCH ENABLE 36


100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

100N

G3
74HC14 7 G4 D2 14 15 6 13 CD CLOCK 35
12MHz 2 C5/2,3,4+ D1 17 16 7 12 CD CS (ADC) 34
D0 18 19 8 11 CD DATA OUT 31
3 1,5D [1] 14 N/C 74HC273 ULN2803A
4 [2] 13 3MHZ
5 [4] 12 N/C YLEISTOLERANSSI
PIIRT. 7.12.1993 SK LAITE
6 [8] 11 N/C MGF 101/110 SFS 4011
MERKKI MUUT.NO MUUTOS PIIRT. TARK. HYV. TARK. TI
74HC163 SUHDE
LIITTYY
D 31529 C109 = 2.2n. 9.9.97 HK 9.9.97LTu 9.9.97JJS HYV. JJS :
NO
E 31732 R27 , R28 = 4x470R. 1.12.98 NJ INSTRUMENTARIUM NIMI
CPU BOARD
C 31272 eps-200 SIGNAL NAMES ADDED 18.4.96JJS 8.5.96HK 8.5.96JJS
imaging V 1.4
MGF 32245-3SE
TP1 TP2 TP3

2
4x10K

4x10K
1 RN1

3 RN1

4x10K
1 RN2
+15V +34V
ICA1 L4962 L1 D4
X3 1 Vin C&Z +34V 7 IN 2 +24 Vin BUCKY 1 X6
OUT
N/C 8 1 N/C
N/C 9 3 N/C 150uH 1N4004

GREEN
H3
"+24V"
N/C 16 6 N/C ICA6 2 X6
15 SOFT

9K09
FRQ 3 2

R6
GREEN
OSC GND COM FB +15V

"+34V"

H1

H2
1

"+15V"

GREEN
14 4 5 12 13 11 10
LM317T

P6KE18
+

220n

R16

3K9
C7

C8

D3
P6KE47

470u
+
1000u
C1

C2

220n

121R
D1

R9
10K
R1
+34V

100n
C11
15K

BYV27
R4

4K75

2K21
D2

R5
R24

R2

R7
4K75

+
2u2

2n2

4K32
C3

C4

R3
56R2
TP12 +15V

+
12

2K2
R8

C9
330p

100u
33n
C5

C6
ICA2 VCC T1 1

BYV27

221R
R26
12 R17 D5 R23

D6
X3 2 14 2 V1- C1 8 3
13 1
LM324 20K V1+ C2 11 P6KE18 150R/5W
15 V2- BD140 T2 2

BY229
2

CXX

D7
100n
16 V2+ 9

R18

C10
1M

100n
E1 1
3 FB E2 10
14 VREF BUS36 3
13 OC 6

56R2
R25

C13
RT

1n
X2 10 ICD4 ICD5 16 7 8 4 DTC CT 5
+34V 13 VDD VE VS GND
& 11 6
12 11 Inh TL494 7 ICA4

C12

R22
A

10n

3K32
4011 TP6 +15V 10
+15V B

H4
2 K1 9 C MUX

1N4004
ICD6

D11

RED
X2 2 CDIR 1 16 13 X0

BY229
1 14

D8
ICD3 X1

2
"0" UP 2 15 1 18 G2R-2 15 X2

4x10K

R14
12

10K
1 RN3
"1" DOWN X3 X 3 R28

R27

4K75
TLP521-4 ULN2804 1 X4
TP4 TP5 5 10K
+15V X5
R47

100K

2 CBA X

P6KE18
FIXED X6

4K75
100n
C14

R21
4 0 0 0 FIXED

D9
+15V SPEED X7
4051B 0 0 1 FIXED

R15

10K
0 1 0 LOW

8
+15V 0 1 1 HIGH R85

4x10K

4x10K
R13

10K

5 RN1

7 RN1
ICD6 1 0 0 LOW +15V
X2 6 SPEED 5 12 1 0 1 LOW 3K32

C20
ICD1 1 1 0 LOW 3K65 X5

1
1
"0" HIGH SPD 6 11 13 12 1 1 1 LOW
ICD2 R34
"1" LOW SPD TLP521-4 40106 3 4 2 X5

40106 LOW SPEED HIGH SPEED ICD2


R49

10K

L H R38

R11

R12
5 6
TP13 40106 20K TP8 TP9
D10

50K

50K
ICD2 +15V
1 2 R46

+
+15V 7 3

R37

C19
10K
1u
+15V 1N4004
ICD6 40106 200K
X2 4 CRUN 3 14

RN3 6

RN3 8
2K21

R84

4x10K

4x10K
K2
4 13 5
3 K1

R44
10K
TLP521-4 X1

7
7 5 CMOT 1

H8
LOW SPEED
R48

10K

ICA2 G2R-2
7 5
LM324 6
9 LM324
8 10
"0" LOW SPEED ICA2
X2 9 N/C "1" HIGH SPEED

50K
R45

R43

R33
+15V

10n
R41

C15
68K1
100n
C18
FORCE

500R
16K2
ICD7 1K8 THRESHOLD
X7 6 10 7
R77
X7 5 9 8 R31 R30 4 K1
LM393 8

R42

R40
10K

1K
TLP521-4 2 10K 10K 8 6 CMOT 2 X1
+15V 1 3
ICD6 ICA3 4 G2R-2
X2 1 +5V 10 7 R51

1K8 R35

0.5R/5W
30R1
R32

R29
X2 5 CLIM 9 8

C17

C16
4u7

100n
TLP521-4 150K
R50

2K2

+34V

+15V 2 K2
1N4004

+15V
D12

ICD7
X2 3 CMAN 1 16
ICD3 1
2 15 2 17 G2R-2
R79 D20
1 R52 2

TLP521-4 ULN2804 10K

+
1N4004

C24

10u
100K

R65

+15V +15V +15V +15V TP7 100K


ICD7

51K1
R69
X2 8 ZRUN 3 14 +15V
ICD1 14 ICD1 +34V
D13 R59
4

4 13 1 2 3 4
681R
R66
4x10K
100K

RN3
R63

40106 40106 1K
1 R53 4

TLP521-4 1N4004 LM324 12 14 R73


ICD2 ICD3 4 ICA2 VCC VREF 13
10K

R60 9 8 4 15 R61 3 R36 56R2


+ 1 2 OC 4
3

2 V1- TP10 TP11


X7 7 N/C 121K 40106 47K5 11 20K 1 V1+ DTC D15 T3
ULN2804 R70

221R
BYV27
R75
15 8

D17
V2-
+

1n

C1
C21

R64

C22
1M

+15V 16 11
10

V2+ C2 P6KE18 150R/5W T4


BD140

6
X7 8 N/C 3

3 RN2 4
FB
4 7
+

6 9
C33

R62

4x10K

4x10K
10n

5 RN2
121K

RT E1

BY229

C34
1 R54 2

BUS36

D19
R74
R82 5 E2 10

56R2

C26
CT
10K

1n
GND 3 K3
ICD1
475K

3K32

3K32
470n
C35

R78

R67

R68

ICD7 4K75 TL494 7 ICA5 7 5 ZMOT 1 X4


X2 7 ZDIR C 5 12 5 6 +34V
4
40106 X4

H5
"0" UP 6 11 1N4004 R72 8 6 ZMOT 2
"1" DOWN R80

RED
TLP521-4 10K

BY229
2 K3 1K G2R-2

D18
1N4004

D21
D14

ICD2 ICD3 ICD3

P6KE18
X7 1 ICD3 R81

8K25
C23

D16

R71

C25

1u
1 13 12 5 14 6 13
10n

4K75
3 16

R76
G2R-2 200K 40106
ULN2804 ULN2804
X7 3 +15V ULN2804 +15V
R58 +15V
+
C36

10

2K2 ICD3
ICD2 14
N/C 7 12 N/C
+15V 11 10 N/C
GREEN

14
ICD4 +15V
R55

10K

1 ICD4
H6

ICD1 3 R83 9 40106 ULN2804


X7 2 A R19 9 8 2 & 10 F ABC Z-MOTOR N/C
4k75 8 & 010 ON
10K 40106

8
LOWER LIMIT 4011 011 OFF ICD3
4011 100n ICA3

100n

100n

100n

100n

100n
C27

C28

C29

C30

C31

C32
100 OFF N/C 8 11 N/C N/C 4 K2

4x10K
7 RN2
5 7 N/C
+15V 101 ON 6
R57 110 ON LM393 ULN2804 N/C 8 6 N/C

2K2 111 ON N/C


GREEN

ICD4
R56

10K

YLEISTOLERANSSI
5 PIIRT. 06.04.95 LTU LAITE
H7

ICD1 4 MGF 101 / 110 SFS 4011


X7 4 B R20 11 10 6 & MERKKI MUUT.NO MUUTOS PIIRT. TARK. HYV. TARK. 02.06.95 HK SUHDE
LIITTYY
10K 40106 D 31332 Kompression puristusvoima muutettu 16.8.96 AM 27.8.97HK 27.8.98JJ HYV. 02.06.95 JJS :
UPPER LIMIT 4011
NO
C 31149 Uusi ohjauslogiikka kompressiolle 17.10.95RN 26.1.96HK 26.1.96JJ INSTRUMENTARIUM NIMI
C&Z DRIVER BOARD
MG 33035-3SE
E 31686 Kompression puristusv.muut. R85 = 3K32 13.8.98 NJ
imaging V 1.5
92744-8.sch-1 - Thu Jun 15 09:11:20 2000
+24V +5V

IC1 L4962 L1
+24V X1 1 7 IN OUT 2
NC 8 1 NC +5V
NC 9 3 NC 150 H
NC 16 6 NC
15 SOFT FRQ
OSC GND COM FB
14 4 5 12 13 11 10

C10

C12
C7

C8
100n

100n

100n

100n
+
100u
C1

+
C5

100u
BYV27
V1
15K
R2

4k75
R3
2n2
+

4K32
2u2
C2

C3

R1

33n
C4
GND X1 2

+5V
+5V

CPUVCC X2 8 1 X4
+5V 2
OPTIO
4
6
8
2

2
8
6
4
1 RN8

3 RN9
330R

1k
1 40 IC8
3
5
7

1
7
5
20 38 10 X4
37 9
MCT2E ' IC7 36 8
1 2 39 DATA 35 7
40106 34 6
CD_DATA_OUT X2 6 33 5
IC3 32 4
31 3
MCT2E ' IC7 OPTIO
4 RN2 3
3 4 22 STROBE 30 2 1
29 4 H1
40106
CD_LATCH_ENABLE X2 3 28 2 RN1 1 330R 5
IC6 27 4 3 10 1
UCN5818 26 6 5 9 6
25 8 7 8
MCT2E ' IC7 24 7
330R 8 RN2 7 2
5 6 21 CLK 23
6 5 3
40106
CD_CLOCK X2 4 6 RN3 5 330R
IC5 18 8 7
17 9 H2
+5V 16 8 RN4 7 330R 10
15 6 5 5 1
14 4 3 4 6
13 2 1 3
10k

12 2
R9

330R
' IC7 11 4 RN3 3 8
13 12 19 BLANKING 2 1 7
10 9 3 4 5 6 7 8 8 7 6 5 4 3 9 10 10 9 8 7 6 5 4 3
330R

RN11
40106

RN5

RN6
9 10 3 4 5 6 7 8
10u

10k

10k

10k
+
C9

1 1 1
+24V +24V +24V +24V

+5V +24V

1N4004

1N4004

1N4004

1N4004
V6

V7

V8

V9
MCT2E ' IC7 16 8
11 10
3 47R/5W 6 X1
40106 7
CD_CS(ADC) X2 5 H3 R11
IC4 330R 2
1 6 5 X1
R6 3 4 5 6 7 8 9 IC2
+5V H4 10 14 4 X1

RN7
330R
+5V +5V 15

10k
R7 1 L293B 11 3 X1
+5V
1N4148

X3 3
V11

10k

10k
RN7

RN7

VCCOUT IC11

1N4004

1N4004

1N4004

1N4004
1 8

V5

V4

V3

V2
VCC

AIN X3 2 2 REF+ 7 4 5 12 13
IN CL
3 DOUT 6
4 REF- 5
X3 1
1N4148

GNDOUT *CS
C11
V10

GND
100n

TLC549
+5V
1k5

R10

MCT2E 220R

CD_DATA_IN X2 2 R8

CPUGND X2 7
IC12 +5V IC10 14
VCC 3
1 6
2
4k75

4 8
R4

'
IC7 5 11
100k 9 8 9
R5 40106 10
12 13
GND
C6
100n

74HC126 7
S1

YLEISTOLERANSSI
PIIRT. 15.1.93 OPO LAITE
ALPHA RT SFS 4011
MERKKI MUUT.NO MUUTOS PIIRT. TARK. HYV. TARK.
SUHDE
LIITTYY
HYV. :
NO
30600 PUSKUROIDAAN AD-MUUNTIMEN SARJALIIKENNE 16.5.92 TI INSTRUMENTARIUM NIMI
COMPRESSION DISPLAY
MG 32010-3S
imaging 32010A3S.DWG
+15V TP17
X3
+15V

2 RN7 1
"EXP" +15V

2 RN1 1
9X10k
ICD1 H2

4X10k
11 10 R32
ICD1
40106 3K32

10k
X3 11 *EXPENA 13 12

R33
TYPE 40106
R35
X3 18 TUBEFAIL R36 R37
TYPE 47k5
6K81 CURRENT LIMIT 475R
LM339 R38
9 R39 TL074
14 2 39R2
8 1

BYV27

BYV27
+15V ICA4 1k 3
+15V ICA3

D13

D14
R40
TF3 SEC

750R
6k81

R41

33p
C9
NORMAL +15V

R104
39R2
RN7 1
TP18 3 4

1R

R42
1W

332K
R43
1 S1

9X10k

H3
ICA4 3 2 ICD2 14

BYV27

BYV27
10 13 4 8 13 "SHUTDOWN"

8 RN1 7
X3 9 S Q

D15

D16
11 11 +15V BRIDGE CURRENT
10 LM339 6 5 C1

4X10k
14 9 1D

R44

C10

10n
10k
D17 R45 10 12 D18
17 TEST R R46
TYPE +15V
SD101A 1k 4013 7 ICD3 14 500R
2 & 1N4005
R47 R48
8 9

SD101A
+15V 100k 1

D19
10k

C11

22n
4023 7

4K75
R49

500p
4 ICA3

C12
X3 15 HFB R50 5 "OP FRQ" TP19
7 TP20
16 10k 6
TYPE SD101A
11 TL074 +15V
C13

D20
10n
43k2

2 T13
R86

4 RN1 3
TP21
C54

10n

1 R51 2

RN1 6
"PULSE A" +25V +25V

4X10k
kV-FEEDBACK 1

4X10k
10k
+25V
3
ICA1 D21
1k82

TP22 IRF642
R87

2 R52 1
-15V -15V 5

5
7

10k
6 ICD1 ICD3 ICA5 1

+
C14

R53

100k
LM324 R54 1N4148 TF1 PRIM

22
SHUTDOWN 5 6 3 Vs

2
R55 R56

LM336-5.0
5 & 6 2

D22
500R 40106 ENABLE 4 2 1

1
3k65 4 10k 5 PULSE

10k
R57
3
4023

C16

10n

4K75
R58
RED

1k1

+
+15V D23 +15V D24 GND

100k
R60

C15

R61
10k
+15V R64 R105

22
R63
3 SG3635
1k 5k11 R65
kV-ADJUSTMENT 1N4148 ICD1 14 1N4148 H4
6 RN7 1

1 C17 R66 R67 2k


4 3
9X10k

9X10k
RN3

200k 40106 7 12k1 RED


R68

9k09
2 3 4 5 6 7 8 9 10 ICD1 47p "PULSE B" H5

220p

100p
1 RN4 C20 TP23

C18

C19
X3 12 CHIPSEL 2 9 8 +15V R69 +15V ICD2 "PULSE A"
TYPE 6 1 +25V +25V
A 4X10k 40106 ICA2 100k S Q +25V
100n 3
12 /CS OUT2 2 kV-REFERENCE C1 D25
3 RN4 4 R70 R71 5

3 RN7 1

2 R72 1
X3 13 CLK 13 /WR VDD 14 [1V ~ 10kV] kVfb 1D
8 DATA0 5 6 4 R 2 ICD3 ICA6 1

10k

+
10k 1M

C21

R73

100k
9X10k
TP24 1N4148 TF2 PRIM

22
7 DATA1 7 8 ICA4 ICD1 13 Vs
ICA1 4 4013 R74
TYPE 4X10k 11 3 ICA1 13 R75 2 1 2 12 & 10 2 ENABLE 4 2 1
10 D0 GND 15 10 R76 14 5 11 5
8 kVref 12 LM339 40106 10k PULSE
X3 6 DATA2 6 RN5 5 9 D1 VREF 9 LM324 68K1 4023

4K75
D2 LM324 10k R79 RED

100k
R77

R78
+
1k1
5 DATA3 4 3 8 GND

R80

C22

22
D3

220n
C23
4 DATA4 2 1 7 1M 3 SG3635
TYPE 6 D4 16 R81
4X10k
X3 3 DATA5 6 RN6 5 5 D5 RFB 1 +15V +15V H6
2k
2 DATA6 4 3 4 D6 OUT1 TP25
1 DATA7 2 1 D7
TYPE AD7524 SG3635 TRUTH TABLE H7 RED
4X10k

6k81

6k81
PULSE ENABLE OUTPUT "PULSE B"

R82

R83
D26 D27 1 1 TRI STATE
1 0 HIGH
X3 20 RG2 1N4148 1N4148 0 1 LOW
19 RG1

C24

R84

R85

C25
5k62
0 0 HIGH

100n

100n
1k
TYPE
f MIN f MAX
+34V

1N4004 1N4004

D1 D2

X1
JH2

JH1
K1

K2

11
X1 1
K1 JH2 K2 JH1
36MB100 TP2
R1

LINE F1
R110
20AF
VOLTAGE 100R/25W D3

TP5 T1 TP11 T7
C1

100n

R4 R18
TP36
1

TP39

1
47R TP6 T2 TP12 T8
"310V" 47R
R5 R19
H1
1

1
X1 2 47R TP7 T3 TP13 T9
3

3
47R
P6KE18

R16

P6KE18
K1 JH2 R6

1M
6n8
X4
D4

C4
R20

R30

1M
6n8
D8

C6
SE1

SE2

1
47R 47R
X1
TF2

3 N/C

TF1
TP4
P6KE18

R7

R8

R9
1K

1K

1K

P6KE18

R21

R22

R23
1K

1K

1K
D5

TF3 PRIM

D9
L1
4

C8

4
X1 4 N/C 1 2 HP1 1 X4
1 2
HP1 2
2 5
T10
TP37
+

681K

TP8 T4 TP40 TP14


C2

C3

R2
100n

47

R10 R24 T11


TP38 TP35
1

1
X2 47R TP9 T5 47R TP15
R11 R25 T12
1

1
R3

T6 TP16
200k

47R 47R
5

5
P6KE18

P6KE18
TP10
R17

D10

R31
X2 R12 R26
1M

1M
6n8

6n8
D6

C5

C7
2
SE1

SE2
1

1
47R 47R 4 X4
CAPACITOR 3
TF1

TF2

TP1 TP34
P6KE18

R13

R14

R15

P6KE18

R27

R28

R29
1K

1K

1K

1K

1K

1K
D11
D7
6

X2 1

TP3

L2
HPI 5 X4
2 1
HPI 6

CONNECTORS: +15V +15V +15V +15V

X1: LINE VOLTAGE 1 4 LM324


RN5 8

RN6 8

TL074 TL074 3 ICA4


9X10k

X2: BULK CAPACITOR 7


4X10k

4X10k

2 13 9 1
RN7

1 14 8 6
X3: FILAMENT CONTROL BOARD (control signals) 3 12 10
YLEISTOLERANSSI
5 7 8 9 10 11 ICA1 ICA3 ICA3 12 LM339 PIIRT. 05.04.94 JJS LAITE
ALPHA III/ST/RT
7

X4: HIGH VOLTAGE UNIT PIIRT.


SFS 4011
MERKKI MUUT.NO MUUTOS TARK. HYV. TARK. 19.05.94 HK
X5: FILAMENT CONTROL BOARD (supply voltages) LIITTYY
SUHDE
C 31413 LISATTY D38 23.1.97 AM 31.1.97 HK 31.1.97JJS HYV. 19.05.94 JJS :
X6: WARNING LIGHT INTERFACE NO
D 31636 Muutetu R55 =3k65 ja R63 = 10k. 23.3.98 NJ 20.4.98 AM 20.4.98 HK INSTRUMENTARIUM NIMI
INVERTER BOARD PAGE 1(2)
MG 32220-3SE
E 31735 R1 and R110 = 100R. 30.11.98NJ
imaging 32220D1S.DWG (kV CONTROL) V 1.5
+34V TP26 TP27 +15V

X5
1 RN2 2 3 RN2 4
ICA7 L4962 L3
1 +34V A 4X10k 7 IN 2 B 4X10k
OUT
N/C 8 1 N/C
N/C 9 3 N/C 150 H
H8

GREEN
N/C 16 6 N/C

GREEN
15 SOFT

R88
FRQ

9k1
H9
OSC GND COM FB
14 4 5 12 13 11 10
"+34V"

/50V

P6KE47
+
C26

D28

P6KE18
D29
+
C27
BYV27
1000

D30

470
6K81
R89

R90

15K

4K75
R91

R92

3k32
4K32
2n2
+
C28

C29

R93
2u2
TP28 TP29

33n

330p
C30

C31
X5 2
6
TYPE
R94

H10
1k
F/50V

+
C32

P6KE18
90R9

D34
F
+
470

C33
10
R95
LM337 1
CONTACTS FOR WARNING LIGHTS
3k32

X5 5 -34V
R96

2 3
-34V ICA8 -15V
D31

1N4004 RG2 4 X6
D
8 RN2 7

TP30
4X10k

VB24
VB24
X-RAYS!

1N4004
5 RN2

D36
6
TP31

K4
C 4X10k K4
3 X6

+34V 2 X6

VB24
1N4004 VB24
X-RAY UNIT ON

1N4004

D35
D38

K3
TP32 TP33 +25V K3

1 X6
1 RN8 2 3 RN8 4
ICA9 L4962 L4
X5 3 +34VREL A 4X10k 7 IN 2 B 4X10k
OUT
N/C 8 1 N/C
N/C 9 3 N/C 150 H H11
N/C 16 6 N/C
RED

H12
15 SOFT FRQ
R97

18k

OSC GND COM FB RED


14 4 5 12 13 11 10
"+34VREL"
+

P6KE27
C34

D32
470
P6KE47
D37

BYV27
D33
6K81
R98

R99

15K

4K75
R100

R101

6k8
2n2

4K32
+
C35

C36
10u

R102

33n

330p
C37

C38

X5 4

-15V +15V +15V +15V +15V +15V +15V +15V


4X10k D

4X10k C
8 RN8 7

6 RN8 5

100n

100n

100n

100n

100n

100n

100n

100n
C39

C40

C41

C42

C43

C44

C45

C46

YLEISTOLERANSSI
PIIRT. 05.04.94 JJS LAITE
ALPHA III/ST/RT SFS 4011
PIIRT.
MERKKI MUUT.NO MUUTOS TARK. HYV. TARK. 19.05.94 HK SUHDE
LIITTYY
C 31413 LISATTY D38 23.1.97 AM 31.1.97 HK 31.1.97JJS HYV. 19.05.94 JJS :
NO
D 31636 Sivu 1 muuttunut. 23.3.98 NJ 20.4.98 AM 20.4.98 HK INSTRUMENTARIUM NIMI
INVERTER BOARD PAGE 2(2)
MG 32220-3SE
E 31735 Sivu 1 muuttunut. 30.11.98NJ
imaging 32220D2S.DWG (kV CONTROL) V1.5
TP1

TP2
UPPER RIGHT

TP4
+5V +5V
1 +5V *RHOD +5V
+25V *MOLY +5V +5V

1N4148
R24

R25
VCCOUT 3 X3

D1
10k

10k

R26

10k
ICA1
ICA2 L4962 L1 8 VCC 1
10k
X1 1 +25V
N/C
7 IN
8
OUT 2
1 N/C
H1 H2 ADCL
ADCOUT0
7
6
CL
DOUT
REF+
IN
2
3
AIN 2 X3
X3
150uH ICD1 R1
N/C 9 3 N/C *ADSEL 5 REF- 4 GND 1 X3

1N4148
1 *CS

D2

C1

100n
N/C 16 6 N/C EN1 GND

15 SOFT FRQ 19 EN2 TLC549


OSC GND COM FB
14 4 5 12 13 11 10 *LED0 2 18 1 RN1 2 +5V

475R
R2
*LED1 4 16 3 4 +5V R11
RXD 6 14 5 6 1 X4
TXD 8 12 7 8 51k1
MPTE-36

MPTE-5

TP5
+

100n
*ADCS 11 9 R12 20k

470
D3

C2

D4

C3

C4

C5
470

4X330R +5V TYPE

22
H3

BYV27-100
ADCLK 13 7 3 X4
+5V 51k1

4K32

4K75
D5
15K
15 5
2 2
+

R4

R5

R6
R3
2n2
C6

C7

R27
17 3 *ADSEL

10k
"+5V" ICA3 4 X4
74HCT244 ADCL 8 VCC 1 7 1 8 R7
ADCOUT0 ADCL 7 REF+ 2 R8 6 3
ADCOUT1 ADCOUT1 6
CL IN
3 ICA4 2 1k50
5 X4
X4

330p
DOUT 10k

33n
C8

C9
ADDATA0 *ADSEL 5 REF- 4 LT1077 4 R9
*CS 6 X4

TP3

C10

100n
ADDATA1 GND

X1 2 GND GND TLC549 1k50


7 X4

TP6
R10

1M5

C54

C53
8

33p

33p
R13
1M5 9
+5V +5V +5V TYPE
N/C 10 X4

A0 10 ICD2 28 ICD3 28 ICD4 19


+5V A0 A0 10 A0 5 12 AD0 +5V
A1 9 A0 DA0 ID0
A2 8 A1 A1 9 A1 A1 6 DA1 ID1 11 AD1
14 XTAL1 A3 7 A2 A2 8 A2 A2 10 DA2 ID2 13 AD2
A3 A3 7 14 AD3 1 X5
N/C 1 12MHz 8 12MHz A4 6 A3 DISPMODE 9 ID3 2
A5 5 A4 A4 6 A4 MODE ID7 7 AD7
A5 A5 5 *DISPWR 8 TYPE
7 A6 4 A5 WRITE SEG A 3 X5
A7 3 A6 A6 4 A6
A7 A7 3 SEG B 4
+5V +5V A8 25 A7 SEG A 16 SEG C 5
A9 24 A8 A8 25 A8 17
SEG B SEG D 6
A10 21 A9
A10
A9
A10
24
21
A9 SEG C 20 SEG E 7 X5
C11 A11 23 A10 SEG D 18 SEG F 8
A12 2 A11 A11 23 A11 21
4K75

+5V +5V SEG E SEG G 9


R16

A13 26 A12 A12 2 A12 22


100n +5V A13 SEG F D.P 10
"RESET" A14 27 *RAMCS 20 CS1 SEG G 23 TYPE
ICA5 8 *RST ICD5 40 A14 26 15
S1 A15 1 CS2 D.P DIGIT1 11 X5

+5V
7 S 6 RST 12MHz 19 18 N/C ICD6 20 A15 *WR 27
3 4 2 X1 X2 20 WE DIGIT2 12
RI RO 5 H4 RST 9 R PSEN 29 *PSEN *PSEN 22
CS *RD 22 OE DIGIT 1 4
DIGIT3 13
3 C2 Vr 1 1 WR 25
1N4148

31 EA ALE 30 ALE EN DIGIT 2 DIGIT4 14


1 2 ALE 11 3
D6

TL7705 4 C1 DIGIT 3 DIGIT5 15


D0 11 AD0 D0 11 AD0 DIGIT 4 1
P1 P0 D1 12 AD1 12 AD1 26 DIGIT6 16
100n
C12

R_ENA 1 0 0 39 AD0 AD0 3 1D 2 A0 D1 DIGIT 5 DIGIT7 17


R_DATAIN 2 38 AD1 AD1 4 5 A1 D2 13 AD2 D2 13 AD2 DIGIT 6 2
1 1 D3 15 AD3 15 AD3 27 DIGIT8 18
R_DATAOUT 3 37 AD2 AD2 7 6 A2 D3 DIGIT 7
R17

R18
1K5

1K5
2 2
+

2 2 TYPE
C13

R_R/W 4 36 AD3 AD3 8 9 A3 D4 16 AD4 D4 16 AD4 DIGIT 8 24


3 3 D5 17 AD5 17 AD5 *RHOD 19 X5
*ADCS 5 4 4 35 AD4 AD4 13 12 A4 D5
D7 C14 ADCLK 6 34 AD5 AD5 14 15 A5 D6 18 AD6 D6 18 AD6 ICM7218D 28 *MOLY 20
WDFRS 5 5
ADDATA0 7 33 AD6 AD6 17 16 A6 D7 19 AD7 D7 19 AD7 CODEB TYPE
6 6 21 X5
1N4148 470n ADDATA1 8 7 7 32 AD7 AD7 18 19 A7 27C512 14 MODE
6264 14 22
C15

47n

1M5
R19

P3 P2 74ACT373 10 TYPE
1N4148

RXD 10 0 RXD 0 21 A8 +25V +25V +25V +25V


D8

TXD 11 22 A9 C-BUS N/C 24 X5


1 TXD 1 ICD7
R_CLK 12 23 A10 N/C 25
2 INT0 2 AD-BUS *RST 1 R +5V +25V
WDFRS 13 3 INT1 3 24 A11 *LAT00 11 N/C 26
C1

1N4004

1N4004

1N4004

1N4004
*LED0 14 25 A12 A-BUS TYPE

D10

D11

D12
D9
4 T0 4
*LED1 15 5 T1 5 26 A13 D7 3 2 16 8 ICA6
*WR 16 27 A14 1D VSS VS
6 WR 6 +5V D6 4 5 3 47R/5W 6 X1
*RD 17 28 A15
7 RD 7 D5 7 6 FILMOT1 7 INP2 OUT1
D4 8 9 FILMOT2 2 INPUT1 R20
80C32 20 ICD8 20 OUT2 6 5 X1
D3 13 12 FILMOTENA 1 CE1
ICD9 *IO 19 G3 D2 14 15 (XMOTENA) 9 CE2 X1
ALE 1 *RD 1 3EN1[BA] D1 17 16 (XMOT2) 10 INPUT3 OUT4 14 4 X1
*PSEN 2 19 *RAMCS 3EN2[AB] D0 18 19 (XMOT1) 15 INPUT4
+5V *WR 3 18 *IO OUT3 11 3 X1
74HCT273
*RD 4 17 *DISPWR AD7 2 1 18 D7 L293B
R_DATAOUT 5 16 *LAT00 2

1N4004

1N4004

1N4004

1N4004
AD6 3 17 D6

D13

D14

D15

D16
X2 8 VCC1 A15 6 15 *LAT01
A14 7 14 *BUF00 AD5 4 16 D5 4 5 12 13
AD4 5 15 D4
2
8
6
4

2
8
6
4

A13 8 13 *BUF01
4X330R

AD3 6 14 D3
3 RN6

3 RN7

A1 9 12 R_BDATAOUT
R21

1K5

ICD10
4X1K

A0 11 AD2 7 13 D2
AD1 8 12 D1 *RST 1 R
1
7
5

1
7
5

MCT2E N/C 16V8-15 AD0 9 11 D0 *LAT01 11 C1


74ACT245 10 D7 3 2 (xout7) 1 RN8 2 (xout7) 20 X6
1D
X2 4 CD_CLK D6 4 5 (xout6) 3 4 (xout6) 19
ICA7 ' ICD11 D5 7 6 (xout5) 5 6 (xout5) 18
MCT2E N/C CODEB 19 D4 8 9 (xout4) 7 8 (xout4) 17
EN D3 13 12 (xout3) (xout3) 16
4X330R 1 RN9 2
D2 14 15 (xout2) (xout2) 15
X2 3 CD_ENA D1 17 16 (xout1) 3 4 (xout1) 14
MODE 11 9 DISPMODE 5 6
ICA8 ' ICD11 13 7 N/C D0 18 19 (xout0) (xout0) 13
MCT2E N/C 1 7 8
EN 15 5 N/C 74HCT273
17 3 N/C 4X330R +5V
X6
X2 6 CD_DATAIN 1 RN10 2 2 18 R_CLK 74HCT244
1 X6
ICA9 3 4 4 16 R_ENA 2
MCT2E N/C 5 6 6 14 R_DATAIN 3
7 8 8 12 R_R/W +5V +5V
4
4X10K 74HCT244 TYPE
X2 5 CD_R/W 1

9X10K
100n

ICA10
C16

ICD12 +5V

RN11
N/C MCT2E R22 R_BDATAOUT 1 *BUF00
EN1

C17

C18

C19

C20
100n

100n

100n

100n
X2 2 CD_DATAOUT 330R EN2 19 *BUF00 2 3 4 5 6 7 8 9 10 1 X7
+5V 2
1 RN12

N/C
X2 7 GND1 D7 18 2 (xin6) 2 (xin6) 3
ICA11 D6 16 4 (xin5) 3 4 (xin5) 4

1N5818
1N5818
D17

D18
C21 C22 ICD13 16 D5 14 6 (xin4) 5 6 (xin4) 5
X2 1 N/C 1 C1+ VCC 2 C23 D4 12 8 (xin3) 7 8 (xin3) 6

100n 100n
3
4
C1- V+
V-
6 D3
D2
9
7
11
13
(xin2)
(xin1)
4X10K 1 RN13 2
(xin2)
(xin1)
7
8
X7
C2+ 100n 3 4
5 C2- D1 5 15 (xin0) (xin0) 9
TXD 11 14 R23 tXD 1 X8 D0 3 17 5 6 10
T1I T1O 7 8
RXD 12 R1O R1I 13 100R TYPE 74HCT244 11
10 T2I T2O 7 rXD 2 X8 4X10K 12
N/C 9 R2O R2I 8 TYPE 13

C24

C25

C26

C27
100n

100n

100n

100n
GND 3 X8 14
MAX232 15 TYPE TYPE

P6KE15
+5V

D19
N/C 4 X8
TYPE
1

9X10K
ICD14 FILTBUT 23 X5
X5

P6KE15

RN14
1 *BUF01

D20
EN1 TYPE

C28

C29

C30

C31
19 *BUF01

100n

100n

100n

100n
EN2 2 3 4 5 6 7 8 9 10
+5V
1 RN15

N/C
D7 18 2 (xin14) 2 PROJBUT 12 X6
D6 16 4 (xin13) 3 4 (xin13) 11
D5 14 6 (xin12) 5 6 (xin12) 10
D4 12 8 (xin11) 7 8 (xin11) 9
D3 9 11 (xin10) 4X10K 1 RN16 2
OPDGS 8 X6
D2 7 13 (xin9) OPD2 7
C32

C33

C34

C35

C36

C37

C38

C39

C40
100n

100n

100n

100n

100n

100n

100n

100n

100n

D1 5 15 (xin8) 3 4 OPD1 6
D0 3 17 (xin7) 5 6 OPD4 5
7 8
74HCT244
+5V 4X10K

C41

C42

C43

C44
100n

100n

100n

100n
- An asterisk (*) before a signal name means
C45

C46

C47

C48

C49

C50

C51

C52

an active low signal.


100n

100n

100n

100n

100n

100n

100n

100n

YLEISTOLERANSSI
PIIRT. 02.02.94 HSY/JJS LAITE
MGF MT / RT SFS 4011
MERKKI MUUT.NO MUUTOS PIIRT. TARK. HYV.
TARK.
SUHDE
LIITTYY
- 30841 HYV. :
17.01.00 NO
C 31954 Changed R7, R9 INSTRUMENTARIUM NIMI
C_ARM CONTROL BOARD
NJ
MG 32310-3SC
B 31135 Optipoint datalines added 18.04.96
JJS
imaging V 2.1