Documente Academic
Documente Profesional
Documente Cultură
Manual
GC1000 Mark II
Process Gas Chromatograph
Basic Operation and Startup
IM 11B03A03-02E
IM 11B03A03-02E
2nd Edition
<Toc> <Ind> <Rev> <Introduction> i
◆ Notice
● Regarding This Manual
1. This Manual should be passed on to the end user.
2. Read this manual carefully and fully understand how to operate this product before
you start operation.
3. Yokogawa makes no warranty of any kind with regard to this material, but not limited
to, implied warranties of merchantability for particular purpose.
4. All rights reserved. No part of this manual may be reproduced in any form without
Yokogawa’s written permission.
5. Great effort has been expended to ensure that the descriptions in this manual are
correct. Should you, however, come across a questionable area or note an inconsis-
tency, a telephone call or letter to Yokogawa.co.,ltd. noting the questionable area
would be highly appreciated.
6. The contents of this manual are subject to change without prior notice.
WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice, condition
or the like, which, if not correctly performed or adhered to, could result in injury or
death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls attention to a procedure, practice, condi-
tion or the like, which, if not correctly performed or adhered to, could result in damage
to or destruction of part or all of the product.
IMPORTANT:
Indicates that operating the hardware or software in this manner may damage it or
lead to system failure.
NOTE:
Draws attention to information essential for understanding the operation and features.
TIP:
Gives information that complements the present topic.
Media No. IM 11B03A03-02E 2nd Edition : July. 2002 (YK) IM 11B03A03-02E 2nd Edition : July. 1. 2002-00
All Rights Reserved Copyright © 2001, Yokogawa Electric Corporation
<Toc> <Ind> <Rev> <Introduction> ii
See Also:
Gives the reference locations for further information on the topic.
Alternating current
Indicates the power switch state “ON”.
3. If protection / safety circuits are to be used for the product or the system controlled by
it, they should be installed outside of the product.
4. When you replace parts or consumables of the product, use those specified by our
company.
5. Do not modify the product.
◆ Warning/Caution Labels
● To ensure safety operation of this equipment, warning/caution labels are at-
tached on the equipment as follows. Check these labels for your safety opera-
tion.
A B
The inside of the enclosure is high temperature after turning off the power.
Don’t touch the inside of the enclosure or the compponents, the protective gas
should be suoolied for more than one hour after turning off the power.
D E
F0002.EPS
◆ Introduction
Thank you for purchasing the GC1000 Mark II process gas chromatograph.
This manual describes the technical information on basic operation and startup of Model
GC1000D / GC1000S / GC1000T / GC1000E / GC1000W / GC1000C (Hereafter, it is
abbreviated as GC1000 Mark II) Process Gas Chromatograph.
Please lead the following respective documents before installing and using the GC1000
Mark II system.
■ Instruction manuals that do not depend upon the specifications of the product
(1) Overview (IM 11B03A03-01E)
(2) Basic Operation and Startup (IM 11B03A03-02E)
(3) Maintenance and Inspection Manual (IM 11B03A03-04E)
(4) LCD Panel Operation Manual (IM 11B03A03-05E)
(5) Alarm Message Manual (IM 11B03A03-06E)
and Installation Manual (TI 11B03A03-03E)
If these have not been completed yet, see the Installation Manual (TI 11B03A03-03E).
After completion, return to this manual and do the following:
● If the system power is on, turn off the power.
● Shut off all the gases at the flow control units.
Please read the following respective cautions (General Precautions, Caution of
using Explosion-Protection Instruments, on Piping Construction, and on Piping
Work) before installing and using the GC1000 system.
WARNING
(1) In order to analyze gases, process gas chromatographs use a sample of the process
gas and utility gases.
Since these gases are typically combustible, combustion-sustaining, toxic,
odorous, resolvable, polymerizing, or corrosive, refer to the “Safety Informa-
tion” in our approval drawings and others to ensure safety thoroughly before
using these analyzers.
(2) Up to two protection systems, each of which weighs approximately 10 kg, are installed
on top of the GC1000. Therefore, the center of gravity is higher than the center of the
analyzer body.
Take great care when carrying and installing (piping- wiring) the GC1000. The
GC1000 must be carried and installed very carefully (including piping and
wiring) by more than one person (at least four people are recommended).
(3) Since the GC1000 are precision instruments, take care when handling not to
jolt of knock them.
(4) Use the GC1000 within the range of your purchase specifications.
Yokogawa assumes no responsibility for problems resulting from use by the customer
outside the purchase specifications.
If the GC1000 need to be modified or repaired, please contact your nearest Yokogawa
representative. Yokogawa assumes no responsibility for results where the customer or
any third party has attempted to modify or repair these products.
(5) When touching LCD Panel switches
When touching LCD Panel switches, first, discharge Electro Static Charge of the body.
Then, touch the LCD Panel switches. If not, LCD display may be changed by Electro
Static Discharge.
IMPORTANT:
(1) Read the attached instruction manual before operating the GC1000
(2) The instruments must be installed and operated according to the installation manual,
instruction manual, approval drawings, and operation data.
<GC1000D / GC1000S>
• JIS Expd IIB + H2 T1 (programmed-temperature oven 320˚C max., isothermal oven
225˚C max., liquid-sample valve 250˚C max.)
• JIS Expd IIB + H2 T2 (programmed-temperature oven 225˚C max., isothermal oven
225˚C max., liquid-sample valve 225˚C max.)
• JIS Expd IIB + H2 T3 (programmed-temperature oven 145˚C max., isothermal oven
145˚C max., liquid-sample valve 145˚C max.)
• JIS Expd IIB + H2 T4 (programmed-temperature oven 95˚C max., isothermal oven
95˚C max., liquid-sample valve 95˚C max.)
<GC1000W / GC1000C>
CENELEC (ATEX directive) certified : Group II Category 2G
• EEx pd II B +H2 T1 (programmed-temperature oven 320°C max, isothermal oven
225°C max., liquid-sample valve 250°C max.)
• EEx pd II B +H2 T2 (programmed-temperature oven 225°C max, isothermal oven
225°C max., liquid-sample valve 225°C max.)
• EEx pd II B +H2 T3 (programmed-temperature oven 145°C max, isothermal oven
145°C max., liquid-sample valve 145°C max.)
• EEx pd II B +H2 T4 (programmed-temperature oven 95°C max, isothermal oven 95°C
max., liquid-sample valve 95°C max.)
<GC1000T / GC1000E>
FM Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, NEMA3R.
Type Y purging and Type X purging for CLI, DIV1, GPS B, C & D, NEMA3R.
• T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid-
sample valve 250˚C max.)
• T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid-
sample valve 225˚C max.)
• T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid-
sample valve 145˚C max.)
• T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid-
sample valve 95˚C max.)
(2) Precautions for Explosionproof section (The analyzer with optional code
"FM/CSA Type Y purging" does not have the explosionproof section.)
When handling the screws on the cover of the Protection system, note the following to
avoid damaging the screws since they cannot be repaired.
(1) The enclosure is pressurized. Before removing the cover, reduce the internal pres-
sure by loosening the sealing plug for wiring on the enclosure or relevant means.
(2) When removing the cover, prevent any dirt or foreign matter from contaminating the
screw part.
(3) When installing the cover, tighten the screws by hand ; never use tools.
(4) Since the screws are coated with MOLYKOTE, do not lubricate them.
(7) Override function (The analyzer with optional code "FM/CSA Type Y purg-
ing" does not have this function.)
WARNING
• When the override function is used, Analyzer becomes an ignition source and the high
temperature and the high voltage part will be exposed.
• Please confirm that in the ambient atmosphere, the concentration of explosive gases
is less than the allowable limit, by using a gas detector.
To return to the normal operation, turn off “the override switch” and then close the door as it
was before turning on power.
In this analyzer, if the pressure of the pressurized / type X purged enclosure system (oven,
electronic section) drops while the power is on, the pressurized explosion protection sec-
tion is activated to stop power supply. Therefore, in case of opening the door of the oven or
of the electronic section inadvertently, for maintenance, while the power is on, the protec-
tion system is activated to cut off the power.
The “override function” intensively releases this function of protection system.
The override switch is installed in section.
WARNING
<CENELEC>
* Only trained persons may use this instrument in a hazardous location.
* Do not open when energized.
<FM>
For type X purging:
* This equipment contains components that operate at high temperature. The equip-
ment shall be deenergized for 60 minutes to permit those components to cool before
the enclosure is opened unless the area is demonstrated to be nonhazardous at the
time.
* Enclosure shall not be opened unless the area is known to be nonhazardous, or
unless all devices within have been de-energized.
* Power shall not be restored after enclosure has been opened until enclosure has
been purged for 12 minutes. (When the internal pressure is restored, the system
automatically purges over the 12 minutes, then turns on the power again.)
For explosionproof enclosure:
* Seal all conduits within 18 inches
* Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within have been de-energized. Power shall not be restored after
enclosure has been opened until enclosure has been purged for 12 minutes at speci-
fied pressure indicated by the pressure gauge labeled “EL.BOX” in the pressure and
flow control section.
* Alarm shall be provided and connected to alarm contact output.
(a) The alarm shall generate a visual or audible signal that attracts attention
(b) The alarm shall be located at constantly attended location.
(c) Electrical alarms shall be approved for the location in which they are installed.
Take care not to generate mechanical spark when accessing to the instrument and periph-
eral devices in hazardous locations.
Do not press prick the keyboard of LCD panel (operation and display section) using such
as knives and sticks.
1M‰
F0003.EPS
(c) When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 MΩ resistance, wearing a wrist strap as in (2) above. Keep
easily-chargeable plastic materials away from the bench.
(d) Never touch components mounted on the cards, the pattern side, connectors,
pin components, etc. with bare hands, unless using a wrist strap and a conduc-
tive sheet.
(e) Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).
CONTENTS
◆ Notice ............................................................................................................ i
◆ Warning/Caution Labels ............................................................................... iii
◆ Introduction.................................................................................................. iv
1. How to Set, Measure, and Check the Statuses of the Analyzer .................. 1-1
1.1 How to Check the Status Indication of the Protection System ............................ 1-1
1.2 How to Operate the LCD Panel ................................................................. 1-4
1.3 How to Check the Temperature at Each Part ..................................................... 1-6
1.4 How to Check the Operating Status of the Detector ........................................... 1-8
1.4.1 TCD .................................................................................................. 1-8
1.4.2 FID ................................................................................................... 1-9
1.4.3 FID with Methane Converter ............................................................. 1-9
1.4.4 FPD ................................................................................................ 1-10
1.5 How to Measure/Set the Pressure and Flow Rate of Various Gases ................. 1-11
1.6 How to Check the Analysis Result ................................................................... 1-12
1.7 How to Check the Alarm .................................................................................. 1-14
1.8 How to Check the Chromatogram ................................................................... 1-16
2. Startup ........................................................................................................... 2-1
2.1 Preparations and Checks Before Supplying Power ........................................... 2-1
2.1.1 Preparations and Checks of Miscellaneous Equipment ..................... 2-1
2.1.2 External Piping Check ...................................................................... 2-1
2.1.3 External Wiring Check ...................................................................... 2-1
2.1.4 Air Supply ......................................................................................... 2-1
2.1.5 Steam Supply ................................................................................... 2-1
2.1.6 Supply of Various Cylinder Gases ..................................................... 2-2
2.1.7 In case of GC1000T Type Y Purging and GC1000E Type Y Purging . 2-2
2.2 Operation After Power Delivery ......................................................................... 2-3
2.2.1 Power Supply ................................................................................... 2-3
2.2.2 Date/Time Setting ............................................................................. 2-3
2.2.3 Checking the Set Values ................................................................... 2-4
2.2.4 Supply of Combustion Hydrogen or Makeup Gas, and Combustion Air2-4
2.3 Detector ON/OFF Operation ............................................................................. 2-5
2.3.1 TCD .................................................................................................. 2-5
Window
LED lighting status
can be checked.
C. OVEN PRESS.
GC1000D/T/W
Window
LED lighting status
can be checked.
GC1000S/E/C
F0101.EPS
By the status (lit/unlit) of the LEDs, which are visible through the window on the protection
system, conditions of power supply, internal pressure of the electrical box, isothermal oven
and column oven, and purging, can be checked.
A B C D E F G H I
4 5 6
S T U V W X Y Z ∗
7 8 9
( ) + # $ : < > %
0
/ SP ,
BACK
CANCEL
SPACE
SET / ENT
F1 F2 F3 F4 F5 F6
DISPLAY LOCAL/
OPERATION TABLE PREV NEXT
RESET REMOTE
F0102.EPS
F0103.EPS
F0104.EPS
F0105.EPS
F0106.EPS
F0107.EPS
F0108.EPS
1.4.1 TCD
Figure 1.9 shows the Detector Status screen for the TCD.
F0109.EPS
Input signal: Indicates the value obtained by A-D converting the detector’s analog
signal sampled every 20 to 80 msec (depending on the sample rate)
and averaging the converted value with the value specified as a
sample rate.
Filter signal: Indicates the value obtained by filtering the input signal with the value
specified for “Filtering constant” of the “Detector Signal Set” screen.
Std deviation: Indicates the standard deviation of the past 20 input signals.
Input signal, filter signal and standard deviation are common items for all detectors.
Note: If the operation mode is in Stop, these items read 0.000 mV.
1.4.2 FID
Figure 1.10 shows the Detector Status screen for the FID.
F0110.EPS
F0111.EPS
1.4.4 FPD
Figure 1.12 shows the Detector Status screen for the FPD.
F0112.EPS
NOTE
Measure the pressure and flow rate of each gas after the temperatures of the isothermal
oven and column oven have fully stabilized. If the temperature is low, the pressure is likely
to be higher. The time required for the temperature to stabilize depends on the set tem-
perature: the isothermal oven with the temperature of below 146°C requires 2 to 4 hours,
and the one with the temperature of at or above 146°C requires 4 to 8 hours. For the FID
or FPD detector, the “Therm signal” on the Detector Status screen becomes 0 mV after the
temperature of the isothermal oven has stabilized (see Figure 1.10). This screen can be
viewed by pressing the STATUS key and F2 key (Signal) in order. See Section 1.4, “How
to Check the Operating Status of the Detector”.
For the FID and FPD, a carrier gas, combustion gas, and combustion air are discharged
together from the DET vent. Therefore, before measurement close the primary valves of
cylinders of gases that are not to be measured. For the FID and FPD, supplying the gas
may simply result in natural ignition. In the Flame on status, condensation forms on the
vent line and the “Therm signal” on the Detector Status screen indicates 2 to 6 mV. Mea-
sure the gas flow rate under the condition that the “Therm signal” is 0 mV.
F0113.EPS
When the keys are pressed, the following items are displayed:
# (peak number), Composition (name), Conc. (concentration), R. Time (retention time),
Det# (detector number), S. Time. (peak on time), R. Time (retention time), E. Time (peak off
time), Gate on (time), Gate off (time), Tracking (gate tracking function), S. Level (peak on
signal level), T. Level (peak height), E. Level (peak off signal level), Area (peak area),
Shape (peak shape), H. width (peak half width), Tail (tailing coefficient), Varifactor (variation
factor), Calc. method (peak processing method)
For details on each item, refer to the LCD Panel Operation Manual, IM 11B03A03-05E, or
the Technical Guide, IM 11B03A03-03E.
When the F2 key (Conc) or F3 key (R.Time) is pressed on the Latest Result screen, the
Concentration Record (Figure 1.14) or R. Time Record (Figure 1.15) screen appears.
When the F6 key (Strm#) is pressed and a stream number is specified, the analysis result
of each stream is displayed.
F0114.EPS
F0115.EPS
F0116.EPS
The screen shows the Total (i.e., total number of level-1, level-2, and Component alarms
that are currently occurring), ALM# (alarm number), Contents, and Date and Time of alarm
occurrence.
When the F2 key (Histry) is pressed on the Alarm Status screen, the Alarm Historical
Record screen shown in Figure 1.17 appears.
F0117.EPS
F0118.EPS
2. Startup
IMPORTANT
Do not use a pipe or joint that may be stained with oil or other dirt. Since dirt such as metal
powder may adhere inside the piping immediately after construction, perform full purging
with air before connecting to the GC1000 Mark II.
WARNING
Make sure that the supplied sealing fitting or packing adapter is used for wiring. If the
supplied fitting or adapter is not used, the explosion-proof structure cannot be maintained.
NOTE
According to the explosion-proof standard, hydrogen gas supply to the isothermal oven is
stopped while the power is turned off. If hydrogen is used as a carrier gas, set the pressure
after turning on the power.
IMPORTANT
If a pressure of 1 MPa or higher is applied directly to the GC1000 Mark II, a failure may
occur. To avoid this, take the following procedure when supplying the cylinder gas.
(1) Fully close the pressure regulator of the cylinder.
(2) Make sure that the secondary pressure of the pressure regulator is zero.
(3) Open the primary valve of the cylinder.
(4) Set the secondary pressure of the pressure regulator to 700 kPa.
CAUTION
Routine inspection: Confirm that the pressure gauge for the electronic section in the flow
control unit indicates the specified pressure.
F0201.EPS
F0202.EPS
NOTE
According to the explosion-proof standard, the supply of hydrogen gas to the isothermal
oven is stopped when the power is turned off.
IMPORTANT
If a pressure of 1 MPa or higher is applied directly to the GC1000 Mark II, a failure may
occur. To avoid this, take the following procedure when supplying the cylinder gas.
(1) Fully close the pressure regulator of the cylinder.
(2) Make sure that the secondary pressure of the pressure regulator is zero.
(3) Open the primary valve of the cylinder.
(4) Set the secondary pressure of the pressure regulator to 700 kPa.
When the instrument air is to be used as combustion air, set the pressure for supplying the
instrument air to the GC1000 Mark II to at least 100 kPa more than the set pressure of
combustion air described in the “Operation Data.”
2.3.1 TCD
If the TCD is turned on with the carrier gas line filled with air, the TCD filament may be burnt
out. Before turning on the TCD, feed the carrier gas for at least 10 minutes. The TCD
output cannot stabilize (i.e., the base line drifts) until some time after the temperature has
stabilized. Allow at least 12 hours to pass after the TCD has been turned on, and then start
normal operation and calibration.
The procedure for turning the TCD on and off is explained below.
(1) Select the LOCAL mode. (Refer to Section 1.2, “How to Operate the LCD Panel”.)
(2) Press the OPERATION key. (The screen shown in Figure 1.6 appears.)
(3) Press the F3 key. (The screen shown in Figure 1.7 appears.)
(4) Move the cursor (>) to the detector of interest.
(5) Press the SET/ENT key. The indication changes from On to Off, or vice versa.
(6) Press the F2 key to return to the Operation screen (Figure 1.6).
2.3.3 FPD
Turn on (i.e., ignite) the FPD after the temperature of the isothermal oven has fully stabi-
lized. The time required for the temperature to stabilize depends on the set temperature,
but the required time is 2 to 4 hours (For the FPD, the maximum temperature of the isother-
mal oven is 145°C). When the temperature of the isothermal oven has stabilized, the
“Therm signal” on the Detector Status screen (Figure 1.12) becomes 0 mV. After the power
is turned on, the “Therm signal” changes from 0 mV to -2 to -3 mV and then returns to 0 mV
as the temperature of the isothermal oven rises and stabilizes. If the FPD is turned on (i.e.,
ignited) when the “Therm signal” is -2 to -3 mV, the “Therm signal” does not exceed the
“Flame level” but the flame out state is identified although combustion is still in progress
and a “DET1 FIRE OFF” alarm occurs.
The FPD performs ignition by using a platinum coil. With this platinum coil as catalyst,
simply supplying a carrier gas, combustion hydrogen, and combustion air may result in
natural ignition. If the FPD is turned on after natural ignition, the chromatogram may not
show the peak but output the base line only. If natural ignition occurs, change the “Flame
level” in the “Detector Status Set” screen according to the ignition procedure, and then turn
on the FPD. The procedure for turning on the FPD is described below.
(1) Make sure that the temperature of the isothermal oven has stabilized.
(2) Set the pressure of the combustion air to the value twice the one shown in “Operation
Condition Set” of the “Operation Data.”
(3) Select the LOCAL mode. (Refer to Section 1.2, “How to Operate the LCD Panel”.)
(4) Make sure that the “Therm signal” on the Detector Status screen (Figure 1.12) is
0±0.1 mV.
(5) If natural ignition occurs (i.e., “Therm signal” becomes 2 to 3 mV or higher), set the
“Flame level” to the value of the “Therm signal” plus 0.5 mV. For the setting method,
refer to the LCD Panel Operation Manual, IM 11B03A03-05E.
(6) Press the OPERATION key. (The screen shown in Figure 1.6 appears.)
(7) Press the F3 key. (The screen shown in Figure 1.7 appears.)
(8) Move the cursor (>) to “Det 1 operation.”
(9) Press the SET/ENT key. The OFF indication remains until “Flame On” is indicated for
the FPD. If the ignition fails, a “DET1 FIRE OFF” alarm occurs. If this alarm occurs,
retry from step (1).
(10) When “Flame On” is indicated for the FPD, the indication changes from OFF to ON.
(11) Reset the “Flame detection” to the value shown in “Operation Condition Set” of the
“Operation Data.”
(12) One minute after “Flame On” has been indicated for the FPD, reset the pressure of
the combustion air to the value shown in “Operation Condition Set” of the “Operation
Data.”
(13) Press the F2 key to return to the Operation screen (Figure 1.16).
F0203.EPS
2.5 Calibration
Although calibration with a standard sample is normally performed at the factory before
shipping, it is recommended that calibration should be performed again with the standard
gas upon startup. Once the continuous process measurement is started, it is recom-
mended that calibration be performed once in one to six months.
F0204.EPS
The GC1000 Mark II provides two automatic calibration methods. According to the value
set for “Calib coef range” in “Peak Common Set” (Figure 2.5), “Std area” or “Calib factor” is
changed automatically.
F0205.EPS
F0206.EPS
F0207.EPS
(1)Automatic Calibration
(1) Connect the cylinder or tank to the standard sample stream as referring Section 2.4.1,
“Standard Gas Cylinder or Standard Liquid Tank Connection”.
(2) Select the LOCAL mode. (Refer to Section 1.2, “How to Operate the LCD Panel”.)
(3) Press the TABLE key. (The screen shown in Figure 2.1 appears.)
(4) Move the cursor (>) to “Cal/Val Set.”
(5) Press the SET/ENT key. (The screen shown in Figure 2.6 appears.)
(6) Press the SET/ENT key. An item for “Cal (Val) method” can now be selected.
(7) Select “Auto” using the ~ ` keys, and then press the SET/ENT key.
(8) When the time set for “Auto start time” comes, measurements of the calibration
stream and validation stream according to the settings starts. After the completion,
the calibration stops.
(9) Press the OPERATION key to return to the Operation screen (Figure 1.6).
IMPORTANT
If the time set for “Auto start time” comes while the operation mode is in Run, the calibration
will start after the ongoing measurement of the stream has been completed. After the
calibration the operation mode turns to Run, i.e., the previous operation is retrieved to
continue the measurement. (If the operation mode is in Stop or in Pause, the calibration
during that time is not performed.)
(2)Semiautomatic Calibration
(1) Connect the cylinder or tank to the standard sample stream as referring Section 2.4.1,
“Standard Gas Cylinder or Standard Liquid Tank Connection”.
(2) Select the LOCAL mode. (Refer to Section 1.2, “How to Operate the LCD Panel”.)
(3) Press the OPERATION key. (The screen shown in Figure 1.6 appears.)
(4) Press the F2 key. (The screen shown in Figure 2.8 appears.)
(5) Move the cursor (>) to “Cal (Val) select.”
(6) Press the SET/ENT key. An item for “Cal (Val) select” can now be selected.
(7) Select one from among “Calibration 1 to 3” using the ~ ` keys, and then press the
SET/ENT key. The operation mode turns to Run to start calibration. (If the operation
mode is in Run at the selection, the calibration will start after the ongoing measure-
ment of the stream has been completed.)
NOTE
To cancel the calibration (validation) command, move the cursor (>) to “Selection cancel”
and then press the SET/ENT key. Note that “Selection cancel” is not valid after the mea-
surement of the calibration stream has started.
NOTE
F0208.EPS
(3)Manual Calibration
(1) Connect the cylinder or tank to the standard sample stream as referring Section 2.4.1,
“Standard Gas Cylinder or Standard Liquid Tank Connection”.
(2) Select the LOCAL mode. (Refer to Section 1.2, “How to Operate the LCD Panel”.)
(3) Press the F5 key (Cmd). (The screen shown in Figure 2.9 appears.)
(4) Move the cursor (>) to “Cal (Val) select.”
(5) Press the SET/ENT key. An item for “Cal (Val) select” can now be selected.
(6) Select one from among “Calibration 1 to 3” using the ~ ` keys, and then press the
SET/ENT key. In 0 to 20 seconds, sample displacement becomes ready.
(7) Open the stream valve for the standard sample stream of the sampling system manu-
ally.
(8) Adjust the restrictor of the sampling system so that the standard sample flows into the
line for displacement.
NOTE
To cancel the calibration (validation) command by this time, move the cursor (>) to “Selec-
tion cancel” and then press the SET/ENT key.
(9) After the displacement has been completed, move the cursor (>) to “Cal (Val) start/
stop.”
(10) Press the SET/ENT key. An item for “Cal (Val) start/stop” can now be selected.
(11) Select “Start” using the ~ ` keys, and then press the SET/ENT key. In 0 to 20 sec-
onds, the calibration starts.
(12) After the completion of measurements of the calibration stream and validation stream
according to the settings, sample displacement becomes ready.
(13) Close the stream valve for the standard sample stream of the sampling system manu-
ally.
(14) Press the SET/ENT key while the cursor (>) is pointing to “Cal start/stop.” An item for
“Cal start/stop” can now be selected.
(15) Select “Stop” using the ~ ` keys, and then press the SET/ENT key. In 0 to 20 sec-
onds, the previous measurement status before the calibration is retrieved and the
operation mode turns to Stop.
(16) Press the F2 key (Ope) to return to the Operation screen (Figure 1.6).
F0209.EPS
IMPORTANT
To turn off the power after the operation is stopped, follow the procedure described below.
Failure to follow the procedure when turning off the power could result in malfunction of
components, such as a detector and a column.
(1) For the FID, FID with methane converter, or FPD, stop supplying the combustion
hydrogen or nitrogen and combustion air (i.e., close the primary valves of the gas
cylinders). After the detector indicates “Flame Out,” leave the equipment standing for
at least 4 hours.
(2) If the carrier gas is hydrogen, turn off the heaters for the isothermal oven and column
oven. For the procedure for turning off the heaters, see Section 1.3, “How to Check
the Temperature at Each Part”. Leave the ovens standing for at least one hour until
the temperatures of the isothermal oven and column oven fall.
(3) Turn off the power supply.
Keep on supplying the carrier gas.
◆ Revision Record
● Manual No. : IM 11B03A03-02E
● Title : GC1000 Mark II Process Gas Chromatograph Basic Operation and Startup
July/2002/2nd Edition