Sunteți pe pagina 1din 11

SECTION I INTRODUCTION TO THE INSTRUMENT

Figure 1.1 the Boehringer Mannheim Diagnostics/Hitachi 704

1.1 INSTRUMENT CHARACTERISTICS

The Boehringer Mannheim Diagnostics/Hitachi 704 (Figure 1.1)


is an automated serum chemistry analyzer. Some of its
characteristics include:

. READY-TO-USE 24 HOURS PER DAY . QC CAPABILITIES FOR SIX (6) CONTROL LEVELS

. USER-FRIENDLY KEYBOARD AND CRT . AUTOMATIC EVALUATION OF WITHIN-RUN CONTROL


RESULTS
. 180 TEST/HOUR THROUGHPUT
. MAY BE INTERF ACED WITH HOST COMPUTER
. MICRO (3-20 MICROLITER) SAMPLE SIZES
. RESULTS STORED ON FLOPPY DISK
. REFRIGERATED STORAGE FOR FORTY (40) REAGENT
CONTAINERS . AUTOMATED MAINTENANCE FUNCTIONS

. ENDPOINT, KINETIC AND ISOENZYME CHEMISTRIES The Instrument is divided into two major operational systems:

. FOUR (4) PARAMETER LOGIT DATA REDUCTION FOR . The Control System is that part of the Instrument with
NONLINEAR CHEMISTRIES which the Operator is actively involved.

. AUTOMATIC SAMPLE BLANKING . The Reaction System is that part of the Instrument which
is directly involved in chemical analysis of patient samples.

1-1
1.2 INSTRUMENT CONTROL SYSTEM QC KEY

The Instrument Control System consists of Instrument To view the QC Job Menu, press the QC key. This menu lists
components through which the Operator controls all Instrument displays where:
functions. These components include the keyboard, computer,
CRT and printer. An interface port for a host computer is . Control data is evaluated by a multi-rule Shewhart type
available for use, if desired. analysis

1.2.1 INSTRUMENT KEYBOARD AND CRT DISPLAYS . Daily QC data is listed for a specified control

The keyboard (Figure 1.2) is used to both request and address . Daily QC data is graphed for a specified control
CRT displays. Through CRT displays, information is entered
into computer memory and Instrument functions are controlled. . Cumulative QC data is listed for a control

Each CRT display contains information of a specific nature. . Cumulative QC data is graphed for a control
For example, the Chemistry Parameters display contains
information necessary for the Instrument to perform a specific
chemical analysis.

Similar CRT displays are grouped into menus (Job Menus).


For example, displays used regularly during routine patient
processing are grouped together in the Routine Job Menu. To
view the Routine Job Menu, press the ROUTINE key.

In some cases, an individual CRT display may not be listed in


a Job Menu. The STAT display is an example. To view the STAT
Test Selection display, press the STAT key. See “STAT key” on
the following page.

Each key or group of keys is described in the following


paragraphs:

ROUTINE KEY MONITOR KEY

To view the Routine Job Menu, press the ROUTINE key. This To view the Monitor Job Menu, press the MONITOR key.
menu lists displays which allow you to: Displays listed in this menu allow you to:

. Verify remaining reagent volumes . View absorbance changes vs. time for recent reactions

. Request patient testing . Edit and print patient and control files

. Request controls and/or calibration . View calibration data for wet chemistries

. Specify certain conditions of operation . View ISE calibration data

1-3
PARAMETER KEY OPERATION MONITOR

To view the Parameter Job Menu, press the PARAMETER key. To view the Operation Monitor display, press the OPERATION
Displays listed in this menu allow you to: MONITOR key. This display does not permit input, but provides
the following data:
. Define chemistry parameters for each test
. Analyzer status
. Assign a chemistry to each channel
. Incubation bath temperature
. Assign short test names
. Identification of the specimen being sampled
. Define profiles
. Chemistry configuration (PHASE) in use
. Define calculated tests (e.g., A/G ratio)
. Current date and time
. Customize patient report format
. Instrument alarms

MAINTENANCE KEY STAT KEY

To view the Maintenance display, press the MAINTENANCE To request STAT patient testing, press the STAT key. This may
key. Functions listed in this display allow you to: be done at any time.

. Perform maintenance functions

. Copy or format floppy disks

. Perform precision checks of the Instrument Diagnose


mechanical malfunctions

1-4
NEXT AND BACK KEYS CONTINUE KEY

NEXT and BACK are used to change the CRT display to a Sometimes information displayed on the CRT screen is only
different display within that same job menu. For example, after part of the information available for that display. To view
specifying tests to be calibrated on Test Selection (2) (Routine remaining data that does not appear on the initial display, press
Job #3), press the NEXT key to view Start Conditions (Routine CONTINUE.
Job #4). These keys may also be used to select the first or last
display when the menu is displayed on the CRT. BUZZER OFF KEY

ENTER KEY Press BUZZER OFF once to silence the audible alarm. Pressing
the key a second time will clear alarms listed on the Operation
Press ENTENER after typing information which is to be entered Monitor display.
into an entry field on the display.
NUMERIC KEYPAD
CURSOR CONTROL KEYS

The cursor control keys are used to move the cursor block
from one entry field to another within a display.

To move the cursor block from the first entry field in a display
toward the last, press

To move the cursor block from the last entry field in a display
toward the first, press

The cursor can be made to move backward from the first field
to the last, or from the last field forward to the first.

COPY KEY

Press COPY to print a copy of the display currently shown on The numeric keypad is used for numeric input, as required.
the CRT. The Instrument must be in Standby when the COPY
key is used. PROFILE AND TEST SELECTION KEYS

CLEAR KEY

CLEAR is used to clear one of two types of input errors:

. When incorrect information has been typed, but the


ENTER key has not yet been pressed.

. When incorrect information has been typed, but the


ENTER key has been pressed.

When unacceptable information has been entered, an input


alarm is issued. No other entries will be performed until The profile/test selection keys may be used in one of two ways:
c is pressed to clear the input error.
. As test or profile selection keys, as defined in the Channel
If incorrect information is entered and accepted, it can be Setting and Profile Definition displays.
corrected by moving the cursor back to the specified field and
entering the correct information. . As an alphabetic keyboard, as indicated on each key. Note
that the numbers on each test selection key indicate
channel number only, and are not used for numeric input.

Each key has an indicator light, which is active when the keys
are used as profile or test selection keys.

When one of these keys is pressed, the indicator is illuminated.


When the key is pressed a second time, the light goes out.

When the ENTER key is pressed, any test or profile selection


key with its indicator light ON will be used to assign test
selection.

1-5
STOP KEY
When the Instrument ON/OFF switch is turned ON, all
Press STOP to instruct the Instrument to stop test processing Instrument operating parameters are automatically read from
at the end of the current Instrument cycle. the System Program disk into computer memory.

STOP is also used to end test program functions and to The Data Disk records test results for a maximum of 400 routine
terminate communication with a host computer. samples and 40 stat samples. When this disk becomes full, it
can be removed from the Instrument and saved to maintain a
SAMPLING STOP KEY record of test results. If this type of record is not desired, the
disk can be reformatted (cleared of all data) and reused.
Press S. STOP to instruct the Instrument to stop sampling at
the end of the current Instrument cycle. When this key is 1.2.4 PRINTER
pressed, only the sampling operation stops. All other functions
related to processing tests continue until results print out. The Instrument uses an 80 column, dot matrix printer (Figure
1.13). Patient results can be printed in 8-1/2 X 11 inch report,
START KEY or can be printed in a monitor format, which uses less paper.
Examples of each printed report are found in Section 7,
Press START to place the Instrument into the Operate mode. Reports.
Always consult the Start Conditions display before pressing
START.

1.2.2 COMPUTER

Computer activity is controlled through the use of the keyboard


and CRT. The Instrument computer controls and monitors all
Instrument functions, calculates results from raw absorbance
data, and stores information on the System and Data Disks.

1.2.3 FLOPPY DISK DRIVES

There are two floppy disk drives positioned below the


Instrument CRT and keyboard (Figure 1.12). The left drive holds
the System Program Disk, and the right drive holds the Data
Disk.

Figure 1.13 Instrument Printer

1.2.5 HOST INTERF ACE

The Instrument can be interfaced with a Host Computer. For


details concerning interfacing capabilities, contact Boehringer
Mannheim Technical Service.

1.3 REACTION SYSTEM

The reaction system includes all Instrument components which


are directly involved in chemical analysis of samples. Major
components of the reaction system, and their involvement in
reaction processing are:

. REACTION DISK: central component of reaction


processing; a motor-driven turntable which holds reaction
cells (cuvettes) at its perimeter.

. SAMPLING SYSTEM: transfers sample from a sample cup


into a reaction cell where the reaction process takes place.

Figure 1.12 Floppy Disk Drives

1-6
. REAGENT SYSTEM: adds appropriate reagent(s) to a 1.3.2 SAMPLING SYSTEM
sample-containing reaction cell.
SAMPLE DISK
. STIRRING UNIT: stirs the contents of each reaction cell
after the addition of each reagent. The sample disk (Figure 1.14) is a motor-driven turntable, with
three concentric rings of sample cup positions:
. REACTION BATH: positioned directly beneath the
reaction disk. All reaction cells are immersed in this bath, OUTER RING: 40 positions for ROUTINE samples
which maintains the temperature of reaction mixtures at
37 degrees C. If necessary, reaction bath temperature MIDDLE RING: 30 positions for BLANKS and
maybe set at installation to either 25 or 30 degrees C. CALIBRATORS

. PHOTOMETER: reads the absorbances of reaction cell INNER RING: 10 positions for STAT samples 6 positions
contents. for CONTROLS 1 position for WASH
solution 3 positions for ISE calibrators
. REACTION CELL RINSE UNIT: cleans each reaction cell
after final absorbance reading. Also works with the Blank, calibrator and control materials are placed in dedicated
photometer to determine blank absorbance of each positions on the disk.
reaction cell before use.
Routine and stat patient samples are placed in positions that
Each of these components are discussed in greater detail in are defined when selecting tests to be run.
the paragraphs that follow.
When the Instrument is ready to sample a specimen, the sample
1.3.1 REACTION DISK disk rotates, bringing the desired sample cup into position next
to the sample probe (sampling station).
The Reaction Disk (Figure 1.14) is a motor-driven turntable. It
holds forty-eight (48) reusable plastic reaction cells (cuvettes) The sample disk has a locking cover which helps to prevent
at its perimeter. All chemical analyses take place in these sample evaporation. The cover has three holes through which
reaction cells. The reaction cells are immersed in a circular sampling can occur while the cover is in place.
Reaction Bath, which is positioned beneath the Reaction Disk.

The Reaction Disk is surrounded by Instrument components


that are involved in the reaction process: Sampling System,
Reagent Dispense System, Stirring Unit, Photometer and Cell
Rinse Unit.

Figure 1.14 Reaction Disk and Associated Systems


SAMPLE PROBE SAMPLE PIPETTOR

The sample probe (Figure 1.15) carries sample from a sample The sample pipettor (Figure 1.16) is a positivedisplacement
cup to a reaction cell. The probe is attached to the sample pipettor, positioned behind the left side cover of the Instrument.
(serum) pipettor, and has a liquid level sensor wire at its tip. The pipettor, and the plastic tubing that connects it to the
The liquid level sensor informs the Instrument Computer when sample probe, are filled with deionized water.
the tip of the sample probe is immersed in sample. The sample
probe is mounted on the sample probe arm. When the Instrument is prepared to aspirate sample, (sample
probe tip is in sample), the pipettor motor withdraws the plunger
from the chamber of the pipettor, and sample is aspirated into
the tip of the sample probe.

Once the sample probe has been lowered into the reaction
cell, the pipettor motor is reversed, and sample is dispensed.

Figure 1.15 Sample Probe and Arm

SAMPLE PROBE ARM

When a sample cup has been brought into position for


sampling, the motor-driven sample arm (Figure 1.15) moves
the sample probe from its reset (home) position to the sample
disk, then lowers it into the sample cup. Once the sample probe
liquid level sensor detects sample, the sample probe arm stops,
and sample is aspirated.
Figure 1.16 Sample Pipettor
After sample has been aspirated, the probe is lifted from the
sample cup and moved to the reaction disk. The sample probe SAMPLE PROBE RINSE STATION
arm lowers the probe into the reaction cell at the sample
dispense position, and stops when the probe touches the The sample probe is at its home position when it is directly
bottom of the cell. Sample is dispensed while the sample probe above the sample probe rinse station (Fiugre 1.15), between
tip is in contact with the bottom of the reaction cell. This ensures the sample disk and reaction disk. When the Instrument finishes
that a precise volume of sample is deposited into the cell, and sampling a specimen, the sample probe is lowered into its
is necessary due to the small dispense volume. The sample rinse station. While in the rinse station, water is jetted through
probe is spring-mounted on its arm to avoid damage to the the probe and onto the surface of the probe tip. This rinse
probe or reaction cell. removes any serum adhering to the inside or outside of the
probe.
Following sample dispense, the sample probe arm lifts the
sample probe from the reaction cell and moves it to the sample
probe rinse station.

1-8
1.3.3 REAGENT SYSTEM
REAGENT PROBE ARM
REAGENT DISK
When a reagent container has been brought into position at
Reagent containers are stored on the Instrument in a reagent the reagent aspiration station, the reagent probe arm (Figure
disk (Figure 1.17). The reagent disk is held within a refrigerated 1.18) moves the reagent probe from its home position to the
compartment. The top surface of the disk has positions for a reagent disk, and lowers the probe into the reagent container.
maximum of forty reagent containers. This allows a maximum Once the reagent probe liquid level sensor detects reagent,
of two reagents for each of twenty chemistry channels. When the Instrument stops lowering the probe, and reagent is
reagent containers are placed into the disk, they are contained aspirated.
within the refrigerated chamber of the disk.
After reagent has been aspirated, the probe is lifted from the
When the Instrument is ready to add a reagent to a reaction reagent container and carried to the reaction disk, where
cell, the reagent disk rotates, bringing the appropriate reagent reagent is dispensed. Unlike the sample probe, the reagent
container into position next to the reagent probe (reagent probe is not lowered into the reaction cell.
aspiration station).
Following reagent dispense, the reagent probe is moved to
The reagent disk has a locking cover which helps prevent the reagent probe rinse station.
evaporation of reagents. The cover has an opening through
which the Instrument can aspirate reagent.

Figure 1.17 Reagent Disk Figure 1.18 Reagent Probe and Arm

REAGENT PROBE

The reagent probe (Figure 1.18) carries reagent from the reagent
disk to the reaction disk. It is attached to the reagent pipettor,
and has a liquid level sensor wire at its tip. The liquid level
sensor informs the Instrument when the reagent probe tip is
immersed in reagent, and is used to calculate approximate
number of tests remaining from a reagent container. The reagent
probe can add a second reagent to the reaction cell five minutes
after the addition of the first.

The reagent probe is mounted on the reagent probe arm.

1-9
REAGENT PIPETTOR Forty seconds after the addition of each reagent to a reaction
cell, the reaction disk pauses with that reaction cell in position
The reagent pipettor (Figure 1.19) is a positive displacement next to the stirring unit. The stirring unit moves from its home
pipettor, positioned behind the left side cover of the Instrument. position to the reaction disk, and the stirring paddle is lowered
The pipettor, and the plastic tubing which connects it to the into the reaction cell. Once the paddle is fully lowered into the
reagent probe, are filled with deionized water. The reagent cell, the stirring motor is briefly activated to mix the contents
system does not require priming with reagent. of the reaction cell. The stirring unit is then turned off, lifted
from the reaction cell, and returned to its home position.
When the reagent probe tip is immersed in reagent, the reagent
pipettor motor withdraws the plunger from the chamber of the
pipettor, and reagent is aspirated into the reagent probe.

Once the reagent probe is in position above the stationary


reaction disk, the pipettor motor is reversed, and reagent is
dispensed into the reaction cell.

Figure 1.20 Reagent Probe, Stirring Unit

STIRRING PADDLE RINSE BATH

The stirring unit is at its home position when the stirring paddle
is directly above its rinse bath (Figure 1.20), next to the reaction
disk. After each stirring action, the stirring paddle is lowered
into its rinse station, where it is thoroughly rinsed with deionized
water before being used again.

1.3.5 REACTION BATH

Figure 1.19 Reagent Pipettor The reaction cells are partially immersed in a temperature-
controlled waterbath (Reaction Bath, Figure 1.21). The ring-
REAGENT PROBE RINSE STATION shaped reaction bath is positioned beneath the reaction disk.

The reagent probe is at its home (reset) position when it is Two glass windows (inner and outer) are positioned in opposite
directly above the reagent probe rinse station (Figure 1.20), walls on one side of the reaction bath. These windows permit
between the reaction disk and the reagent disk. After each light from the photometer lamp to pass through the reaction
reagent dispense, the probe is lowered into its rinse station, bath water, and through the reaction cells in the bath. The light
where water is jetted through the probe and onto its outside beam emerges from the outer window of the reaction bath
surface. This rinse removes any reagent adhering to the inside and enters the Instrument photometer.
or outside of the probe.
A small amount of deionized water is automatically added to
1.3.4 REACTION MIXTURE STIRRING SYSTEM the reaction bath every 20 minutes to compensate for
evaporation. In addition, the entire reaction bath water volume
STIRRING UNIT is automatically replaced with fresh deionized water each time
the Instrument is powered up. A detergent (Extran) is
The stirring unit (Figure 1.20) consists of a slender stirring automatically added to the bath whenever the water is
paddle, the motor that spins the paddle, and the mechanism exchanged. Extran prevents microbial growth and prevents air
that moves the stirring paddle to and from the reaction disk. bubbles from interfering with absorbance readings.

1-10
1.3.6 PHOTOMETER 1.3.7 REACTION CELL RINSE UNIT

The photometer is positoned so that absorbance readings of The reaction cell rinse unit (Figure 1.22) consists of a series of
each reaction cell are taken as the reaction disk is turning. probes mounted above one side of the reaction disk. The
This arrangement enables the Instrument to read the probes are connected to vacuum lines and deionized water
absorbance of each reaction cell as frequently as once every supply lines. The reaction cell rinse unit performs the following
20 seconds. functions:

The photometer lamp is positioned beneath the center of the . Removes reaction mixture waste from reaction cells
reaction disk, within the ring of the reaction bath. The lamp is
encased in a constant-temperature water jacket (Figure 1.21), . Rinses reaction cells with Dl water
which helps to maintain a constant energy output from the
lamp. . Aspirates rinse water from reaction cells

The light from the photometer lamp passes through the . Dispenses deionized water into reaction cells for
following structures in sequence: photometric cell blanking

1. Inner reaction bath window . Aspirates water from cells after photometric cell
2. Reaction bath water blanking
3. Reaction cell and its contents
4. Reaction bath water . Dries inside walls of reaction cells
5. Outer reaction bath window
6. Photometer These functions occur as part of the routine operation of the
Instrument. Each reaction cell is photometrically blanked with
When the light beam enters the photometer, it strikes a water before sample is dispensed into it, and each cell is
diffraction grating, which separates the light into its constituent thoroughly cleaned after final absorbance measurement of a
wavelengths, and reflects them onto a fixed array of eleven (II) reaction mixture.
photodetectors. Each photodetector is permanently positioned
to detect light at a different wavelength. The reaction cell rinse unit uses the deionized water supply,
and the internal vacuum system.
Absorbance readout information for each chemical analysis is
programmed into the Instrument computer. The computer
selects the wavelengths and the times at which a reaction
mixture’s absorbance is read.

The Instrument computer keeps track of which test is being


performed in each reaction cell. It also knows when each
reaction cell passes through the photometer light path. The
computer uses this tracking ability and the programmed
readout instructions to obtain test results. A complete
discussion of the Instrument’s calculation of results is found in
Section 3 of this manual.

Figure 1.21 Reaction Bath and Photometer Lamp Figure 1.22 Reaction Cell Rinse Unit

1-11
DEIONIZED WATER SUPPLY SYSTEM 1.3.8 ION SELECTIVE ELECTRODES (ISE)

The deionized water supply system consists of the deionized The ISE system provides the instrument with an electronic
water reservoir (Figure 1.23) and a series of pumps and method for assay of sodium, potassium and chloride on a single
electronic valves. Water is supplied to the reaction cell rinse patient serum sample. The ISE unit consists of three ion
unit directly from the deionized water reservoir. If your selective electrodes, a reference electrode, a dilution vessel
Instrument is connected to an external water source, the and three pipettors. ISE reagents are stored in the door of the
reservoir water level is monitored. Water is automatically added water compartment.
to the reservoir, when necessary.
When electrolytes are requested, sample (20 µl) is aspirated
Water from this source is also supplied to the probe and stirring by the sample probe, and the probe moves to the ISE dilution
unit rinse baths, and to the reaction bath. vessel. The dilution vessel is protected by a mechanical shutter
that opens when the probe is in position to dispense sample.
VACUUM SYSTEM
Sample and ISE diluent are simultaneously pipetted into the
The vacuum system consists of a vacuum pump, vacuum ISE dilution vessel. The diluted sample is transported through
reservoir, and serum waste container (Figure 1.23). The vacuum the ISE measuring cartridges by a “sipper” pipettor, and voltage
system is used by the reaction cell rinse unit to aspirate reaction readings are obtained for each electrolyte.
mixture waste and reaction cell rinse water. Reaction mixture
waste drains into the reaction waste container. Reaction cell The electrical potential generated by sodium, potassium and
rinse water leaves the Instrument through the main drain line. chloride is compared with the electrical potential generated at
the reference electrode. Electrical information is sent to the
instrument computer for results calculation.

If electrolytes are assayed in the ISE STAT mode, 60 samples


can be processed every hour. When electrolytes are assayed
together with wet chemistries, ISE throughput is dependent
upon the number of wet chemistries requested for each
specimen.

Figure 1.23 Water Compartment

1-12

S-ar putea să vă placă și