Sunteți pe pagina 1din 15

Periodontology 2000, Val.

17, 1998, 7-21 C o m r i a h t 0 Munksaaard 1998


Printed in Denmark. All rights reserved
PERIODONTOLOGY 2000
ISSN 0906-6713

Titanium - the material of choice?


G. STEINEMANN
SAMUEL

The foreign body less steel and titanium. Gold and silver resist oxi-
dation in air but resist corrosion much less in sea
Surgical fracture-treatment implants have the func- water and biological fluids. It is common experience
tion of a temporary splint. In the form of a screw or that these metals lose polish after some time.
a plate or a pin, the implant stabilizes the fracture Metals having lower corrosion resistance than sil-
and supports forces, including those of functional ver and gold, such as aluminum, molybdenum and
load. Other types of implants, such as endoprosthes- iron, show visible attack or oxidation in living tissue
es, are permanent devices. In dental practice, the life and abnormal cells are observed in their vicinity.
expectancy of a reconstruction or an implant should This tissue reaction is equivalent to a “chemical in-
at least be several decades. Yet any implant is a “for- sult”. In fact, the metals from iron through silver cor-
eign body”. The expression has a straightforward and rode so rapidly that the supply and migration of oxy-
simple meaning. An implant is not like living tissue, gen cannot keep up with the consumption of the
where no free electrons exist and where metals, such oxidant (this state is called the diffusion limit) so
as essential trace elements, occur in a bound and not that the contact tissue to the foreign body is starved
in an elemental state. Is this foreign body a chemical, of oxygen. This direct effect of corrosion is not speci-
physiological or mechanical “insult” to the living fic to the metal of the implant. However, metals are
tissue? released by the corrosion process and some of them
are toxic to cells; a sterile abscess and cell death are

Corrosion and soft tissue reaction


In the chemist’s view, corrosion is the visible de- [ohmcm*] CoCrNiMo
struction of metal. It may cause a structure to rup- cw316LESRw
Toxicity
ture or lose function, such as by breakage of an im-
Ti alloys
plant. That was before the 1960s. This aspect is not
Inertness
important for modern metals in surgery because an ‘A1
Mo
attack is so small that a material loss is not visible Fe
and cannot be weighed. More sensitive electro- Ic o Sequestration
I
chemical methods are needed to measure corrosion. Tissue reaction
I

But experiments intended to reproduce the real con-


Fig. 1. Data from in vivo corrosion experiments for various
ditions for a surgical implant in tissue are not metallic elements and for practical alloys. The diagram
simple. has the two coordinates tissue reaction as abscissa and
The polarization resistance method has been used polarization and corrosion resistance, as the ordinate.
for in vivo experiments (52);the results are shown in Tissue reaction is grouped according to the three distinct
Fig. 1. The method requires minimally invasive pro- forms of “toxicity”,“sequestration”and “inertness”.Cor-
rosion resistance is roughly proportional to the measured
cedures, and a characteristic feature is that reduction polarization resistance (to obtain corrosion current den-
and oxidation reactions on the metal are not forced sity, use j=0.03 Vlpolarization resistance). The useful
and run in the open-circuit mode. scale in chemistry and biology is always the logarithmic
The noble metals silver and gold have a resistance one. Thus, differences over the series, such as from cobalt
(Co) to silver to titanium (Ti), amount to factors of 1 to
to corrosion that is about in the middle of a practical
100 to 10,000 in corrosion resistance. CoCrNiMo is
scale. In the logarithmic scale of the polarization re- wrought cobalt-base alloy, cw316LESR is cold-worked,
sistance, the number is 5 to 6, and thus about 2 remelted stainless steel and the titanium alloys are
units, or a factor of 100, lower than high-grade stain- Ti&l,Mo, T i d 4 V , Ti15Mo.

7
Steinemann

involve electrons. Chemical reactions are the basis


of metabolic and growth processes; they involve the
second elementary particle of chemistry, the proton
or hydrogen ion, and the technical term for this is
acid-base reactions. The foreign body interferes with
these processes of life.
Fig. 2. Micrographs of tissue in contact with implants
made of commercially pure titanium (left) and TiAlV alloy Redox reaction
(right)
In water and tissue fluid, corrosion occurs as an elec-
trochemical process in which oxidation of the metal is
observed. Among the four elements shown in Fig. 1, coupled to reduction: an electron gain of electrolyte
vanadium is the most toxic and copper the least. components. The dominant form for the latter reac-
This reaction is equivalent to a major “physiological tion in neutral solutions is the oxygen reduction,
insult”. which yields a hydroxide that precipitates on the
Stainless steel, cobalt-base alloy and titanium metal surface. Other reduction reactions with inor-
have similar levels of polarization resistance, but ganic and organic components of tissue electrolytes
tissue reactions differ. High corrosion resistance is are possible. The crucial characteristic is that electron
apparently not sufficient to suppress the minor re- exchange occurs at the metal surface (and only there,
jection reaction observed for the two classical alloys but not in the liquid, where electrons do not exist as
that comprise the cell-toxic nickel and cobalt as es- free entities). It is a foreign body reaction of a chemi-
sential components. Another example is the Ti&14V cal kind that can lead to denaturation of the tissue in
alloy, a standardized metallic biomaterial. Four per- contact with metallic implants. The method for study-
cent of toxic vanadium suffice to elicit a prominent ing these reactions is voltammetry, or a polarization
foreign body reaction (Fig. 2). These observations experiment. Interesting systems are electrolytes con-
suggest an alloy rule (53): avoid toxic components! taining complex formers, such as chloride, cyanide
Corrosion is essentially an atomic process so that, in and EDTA; gold shows massive reactions for any of
principle, the interaction for all components of a these complexes, whereas titanium is fully inert.
metal must be considered. On the other hand, path-
ologists question the existence of a threshold for lo-
Dissolution of the corrosion product
cal and especially systemic (allergic) reactions.
The first-formed reaction products of corrosion are
hydroxides, hydrous oxides and oxides, that is, spar-
ingly soluble salts, and occasionally complexes (such
The unwanted reaction of the as halides). These salts can be soluble or not in the
foreign body tissue fluids (which are aqueous electrolytes) and

The species of metal compounds ingested with nu-


trition and passing into the bloodstream, that is,
being metabolized, as well as those finally stored in Internal fixation

a
organs and tissue are incompletely known. But most
metals (other than alkali) in body fluids and tissue
Dental implant
are bound to organic matter and exist in a stable,
electrically uncharged form. Fig. 3 suggests that
metal release from implants involves a different Food 97 \
path; it can be associated with the entry of the metal Metabolism
through a wound and then corrosion of the metal.
In that state, the metal is definitely in its elemental
form and in a different chemical state than the metal Foreign body
bound in living matter.
Oxidation-reduction reactions (redox reactions)
play an important role in bioenergetics: the energy Fig. 3. Metals released from implants follow another reac-
transformations in living organisms. These reactions tion path than metals entering metabolism with nutrition

8
Titanium - the material of choice?

MI I dominant hydrolysis product under physiological


conditions (pH of about 7) is the unhydrolyzed
-6 mM-
.-
pM:
nickel cation, with a concentration of about 1 mM at
the limit of hydroxide precipitation. The unwanted
E nM reaction product of corrosion is an ion.
6 nM-
S In serum, the nickel concentration is about 10 nM,
PM- , , , , , , , , , , , , , - and in human skeletal muscle it is about 3 pM and
0 4 7 10 14 less than the solubility limit for physiological pH
physiologicalu pH value
values. The metal concentration in the contact tissue
Fig. 4. Distribution of hydrolysis products in solutions around implants is strongly enhanced and far above
saturated with respect to nickel hydroxide [Ni(OH),]. The
full line is the solubility limit expressed as the total con-
the content in normal muscle tissue; the concen-
centration of nickel. Data for the concentration of nickel tration is of the order of the toxicity threshold for
in serum (S), in muscle (M), in contact tissue around nickel. The sequestration reaction for stainless steel
stainless steel implants (I) and toxicity levels (T) are implants is the consequence.
added at right margin. Sources: Steinemann & Mausli (55)
and Steinemann (56).
Titanium. Titanium is a reactive metal; in air and
aqueous electrolytes, it forms spontaneously a dense
oxide film at its surface. The unwanted reaction
they can be toxic or not (55, 56). To determine the product becomes a potent barrier against dissol-
effects of this corrosion burden, the identity and sta- ution of the metal.
bility of the hydrolysis products must be considered. The constant solubility of titanium dioxide above
A question arises. What is the fate of the unwanted pH 3 and up to pH 12 suggests that an electroneutral
reaction product of corrosion? species dominates in solution (Fig. 5). The core of this
The word hydrolysis is applied to chemical reac- dissolved neutral species Ti(OH),(aq) can be pictured
tions in which a substance is split or decomposed by as a titanium ion with its four positive charges sur-
water. New forms of precipitates and new species, as rounded by four negatively charged hydroxyl mol-
bare ions and as charged and uncharged hydroxo- ecules. At physiological pH values, the first charged
(OH-containing) and 0x0- (0-containing) com- species is the cation Ti(OHI3+with a concentration of
pounds are found in solution. It is common experi- not more than 0.1 nM, which is three orders of magni-
ence that salts and oxides dissolve easily in strong tude lower than the concentration of the hydrogen
acids and in strong bases and that the solubility of ion, which is always present in solution.
oxides is small for intermediate acidity (pH) values. The unwanted reaction product of corrosion is not
The mechanism behind this behavior is proton ex- an ion. This is an important finding because un-
change. In solution, water molecules are attached to charged hydrolysis products have no affinity for re-
metal ions and these aquaions tend to deprotonize. action with organic molecules. Corrosion of titanium
The removal of a proton from the hydration sheath
results in a negatively charged hydroxyl, and thus a
reduction of electrical charge of the aquaion. The de- Mi I
protonized species become hydroxo complexes; Fig.
4 and 5 show these reactions in the case of nickel
and titanium. Hydrolysis data are best represented
as the solubility or distribution curves of the various
hydrolysis products. The diagrams have the coordi-
nates acidity (pH of the solution) as abscissa and the P M t , , , , , , , , \ , , , , , {
molar concentration of dissolved and precipitated 0 4 7 10 14
species as the ordinate, both in logarithmic scale physiological U pH value
what gives the straight lines. Fig. 5. Solubility behavior of hydrous titanium dioxide
measured in sodium chloride and chlorate electrolytes.
Nickel. Nickel is a main component of stainless steel, The full line is the solubility limit and dashed lines are
the metal largely used for fracture treatment im- partial concentrationsfor the named species. Data for the
concentration of titanium in serum (S),in muscle (M) and
plants. The 2 + oxidation state of this metal is the in contact tissue around implants (I) are added at right
important one, and in an aqueous environment the margin. Sources: Steinemann & Mausli (55) and Steine-
hydroxide is the first-formed corrosion product. Its mann (56).

9
Steinemann

becomes in fact no chemical burden, and its inert (18, 55). The reaction products are available and can
reaction in tissue is a sign of the basically different be taken up by the cells. This may or may not be
chemistry in solution. toxic to the cells. Studying the effect of solid and dis-
In serum, the titanium concentration is about 0.1 solved metals on the development of embryonic
pM, and in human skeletal muscle it is about 5 pM. bone rudiments indicates that there is at least some
The muscle concentration of titanium equals the effect on skeletal development (13). The variety of
upper limit of solubility for the aqueous hydroxide, chemical interactions in tissue near the foreign body
which is also about the lower limit for precipitation and at the artificial interface is great.
of the solid oxide (about 3 pM). Solution chemistry Maurer et al. (33) investigated the chemical and
thus provides a stringent, even simple homeostatic biochemical aspects of cell adhesion and cell pro-
mechanism for the regulation of titanium in tissue: liferation in the presence of many metals. Reactions
titanium is at saturation in tissue! The concentration of fibroblast and osteoblast cultures were followed
of titanium in the contact tissue around implants is for two modes of exposure in three experiments (Fig.
about 100 times higher than that of muscle tissue. 6): (i) fibroblasts cultured on metal discs, (ii) osteo-
These high concentrations seem representative for blasts cultured on a neutral substrate in metal satu-
larger and loaded implants and include fretting and rated media, and (iii) osteoblasts cultured on metal
wear debris and residues from surface treatment. discs. An influence of the physical surface topogra-
These metal and oxide particles beyond the solu- phy was circumvented by preparing all metal
bility limit are just deposited in tissue, and retrieval samples and the reference support made of a neutral
studies give no indication of any adverse reaction. A polymer to the same fine finish. The pure metal
statement is in order: no case of local or systemic samples included in experiments (i) and (iii) were
reaction for titanium has been documented. ground under water irrigation and then cleaned ul-
trasonically in deionized water and dried in a jet of
nitrogen gas. For the culture experiments, the metal
Complexes in fluids and tissue
discs were rinsed with modified Eagle’s medium be-
The contact between tissue and an implant raises fore cell seeding. For experiment (ii), the chlorides
the local metal concentration by orders of magni- of tin, aluminum, titanium, zirconium and tantalum
tude. The capacity of homeostatic mechanisms can were added to water in concentrations somewhat
be exceeded. Hydrolysis of the corrosion products higher than the saturation limit of the oxide in water
means proton exchanges in the tissue electrolyte and (4, 5 5 ) . The solutions were titrated to a pH between
pH shifts, which can cause denaturation of organic 6.4 and 7.0 and served to make the supporting media
matter. But binding of metal ions with other ligands for the cell cultures. W o cell lines were used for the
competes with hydrolysis and complexes can be three experiments, one rat fibroblastic line and one
formed by substitution reactions. This substitution rat osteoblastic line, and both were cultured with
is a third form of reaction, insidious in nature, of modified Eagle’s medium as base. Cells were har-
how a foreign body interacts with living matter. The vested at day 5 for cell counting.
distinction between ions or uncharged species as re- The results of the three experiments are shown in
action product of the corrosion becomes important. Fig. 7-9. In all figures, the abscissa is the polarization
Cell toxicity is well known for nickel, cobalt, copper resistance for the metals that had been measured in
and vanadium, whose hydrolysis products are cat- a separate in vitro experiment. The polarization re-
ions and anions respectively. Metals can act as hap- sistance of a metal is proportional to its corrosion
tens: the ion can unite with a protein to form an resistance or the reciprocal of the corrosion current
antigen, and nickel, cobalt and chromium are known density. Metals vary greatly (by a factor of lO,OOO),
allergens (32). Titanium behaves totally different in and the logarithmic scale provides a good presen-
that no proton exchange and liganding with biologi-
cal molecules is possible.

Cell reaction in presence of metals IMetal


In aqueous On metal
A small corrosion rate is not a sufficient condition Fig. 6. Cell reactions are studied for two different modes of
for tissue compatibility; the fate of the unwanted re- exposure:the metal is dissolved in the supporting medium
action product (of corrosion) must be considered and the cells are in direct contact with the solid metal.

10
Titanium - the material of choice?

-
-
2
r
0

.-.>c
lrj
1-
0.5 -
- Control level -
Ta I
-
= 0.2 -
8
I 0.1 -
I

3
Iv
- Detection limit
.05 -
0
b


!A

.02 I I I I I I .02 I I I I I I
3 4 5 6 ? 3 4 5 6 7
Polarization resistance, log ohmcmz Polarization resistance, log ohmcm2
Fig. 7. Results from experiment (i): fibroblasts cultured on Fig. 8. Results from experiment (ii): osteoblasts cultured
metal discs. The abscissa is the logarithm of the polariza- on a neutral substrate in metal-saturated media. The ab-
tion resistance of the metal considered and is a good scale scissa is the logarithm of the polarization resistance of
for its corrosion resistance. The ordinate is the normalized the metal considered and is a good scale for its corrosion
cell count, also in logarithmic scale. Copper and va- resistance. The ordinate is the normalized cell count, also
nadium are toxic to fibroblasts, and corroding molyb- in logarithmic scale. Osteoblasts cultured in modified
denum strongly reduces the proliferation rate. Titanium, Eagle’s medium with dissolved metal hydroxides displayed
niobium, zirconium and tantalum do not interact. no inhibition of cell proliferation.

tation (see also the preceding section on corrosion


and soft tissue reaction). This scale also provides a
-
12
c
1- - - -,- Control level - - -
I - ZrI
-
good grading, as any metal located below number 5 E
0
0.5 - Ti
will show visible attack in chloride containing elec- >
Sn

-8=2 0.20.1 -- Fel


I
c
trolytes and a strong redox reaction, both processes Ta
equivalent to a ”chemical insult” in living tissue. The
Ni INb I
15 I
ordinate in the three figures, cell growth ratios, are 0
.05 - -
also drawn in a logarithmic scale for simplicity of 8 - -Detection limit

interpretation. The cell proliferation rate observed .“L


3 4 5 6 7
for the inert support and without metal salt is used Polarization resistance, log ohmcmZ
as the reference. Fig. 9. Results from experiment (iii): osteoblasts cultured
Fibroblast cells in contact with titanium, niobium, on metal discs. The abscissa is the logarithm of the polar-
zirconium and tantalum can proliferate but do not ization resistance of the metal considered and it is a good
in proximity with molybdenum, copper, vanadium scale for its corrosion resistance. The ordinate is the nor-
(Fig. 7 ) . These are reactions of “inertness” or malized cell count, also in logarithmic scale. Growth inhi-
bition is absent for titanium and zirconium but strong for
“sequestration” and “toxicity” on the other hand, as the corrosion-resistant metals niobium and tantalum. In-
they are found for soft tissue in contact with im- hibition is observed for all lesser corrosion resistant
planted metals (Fig. If. Osteoblasts cultured in metals, such as silver, molybdenum, aluminum, tin, iron,
modified Eagle’s medium with dissolved metals dis- zinc and the toxic metals nickel, vanadium, copper. The
played no inhibition of cell growth for the five metals compatible metals are titanium and zirconium.
tested (Fig. 8). The metal concentrations in the cul-
ture medium ranged from 0.4 to 6 pM, equal to satu-
ration and a maximum possible level of dissolved Different modes of exposure, cells in a metal-satu-
hydroxo compounds for pH values in the physiologi- rated electrolyte or cells in contact with solid metal,
cal range. In the experimental series of osteoblasts elicit different responses. The conditions are not the
cultured on pure metal discs, growth inhibition is same. In the first case, the cells are in an environ-
absent for titanium and zirconium, relatively weak ment of completely hydrolyzed metal species at low
for tin and aluminum, and strong or total for zinc, concentration, what might be termed “weak interac-
iron, copper, molybdenum, vanadium, nickel, silver, tion”. Furthermore, it follows from basic principles
niobium and tantalum (Fig. 9). One experiment of of chemistry that an interaction would be possible
six with titanium resulted in unusually low cell only, and only when, the dissolved species is electri-
counts, but two others showed extreme proliferation cally charged: an ion (55). This is the case for only
where the osteoblast layer reached confluence and one, perhaps two of the elements considered in Fig.
lifted off the surface. 8. Conversely, the direct contact of the cells with the

11
Steinemann

of six weeks, the screws were slightly tighter than


when originally put in; at twelve weeks, the
screws were more difficult to remove and at the
end of sixteen weeks, the screws were so tight that
in one specimen the femur was fractured when
an attempt was made to remove the screw ...
From these studies it would appear that titanium
is a metal which may be useful in surgery, be-
cause of its strength and its failure to cause tissue
reaction. The fact that bone becomes attached to
titanium may be a disadvantage in cases where
screws or pins are placed temporarily. In the past,
the use of some prostheses has not become popu-
lar because it has been felt that these would re-
Fig. 10. Transmission electron micrograph of a nondemin-
main separate from the bone and eventually
eralized specimen showing the intimate contact between
the mineralized bone matrix (B) and the titanium coating loosen. Since titanium adheres to bone, it may
(T). Source: Listgarten et al. (31). prove to be an ideal metal for such prostheses.

A new form of prosthesis, tooth root analogues made


of titanium, become a reality in the late 1960s. This
solid metal is sort of a “strong interaction” (Fig. 9). was a major advance in clinical dental treatment and
Indeed, direct communication via cellular processes engineering initiated by Brfinemark and collabor-
between osseous cells and dental implants has been ators in Goteborg. The word “osseointegration” was
documented (51). coined and defined as “a direct structural and func-
A selective response is found whenever the cells tional connection between ordered, living bone and
are in contact with the solid metal. Growth inhi- the surface of a load carrying implant” (6).
bition is found for fibroblast cells (Fig. 7) and osteo- Structural aspects and cell reactions accompany-
blast cells (Fig. 9) if the polarization resistance of the ing osseointegration have been studied at all scales,
metal considered is below a logarithm of about 5 , by light-optical histology through high-resolution
that is, under conditions of strong corrosion and electron microscopy (1, 20, 30). Listgarten et al. (31)
strong redox activity. The observation of seques- comment on their transmission electron micro-
tration and toxicity is then an expected parallel situ- scopic image of the interface (Fig. 10) by saying that
ation for tissue reaction (Fig. 1). But the remaining “there is no evidence of any space between the met-
metals titanium, niobium, zirconium and tantalum allic surface and the bone”. This finding suggests the
elicit different reactions for the two cell lines, even if possibility of a direct chemical bond.
these elements have a same high corrosion resist- Osseointegration allows the efficient stress trans-
ance. The biochemical interactions differ in the case fer from implant to bone. It implies that no relative
of niobium and tantalum. Fibroblasts are hardy cells motion occurs at the interface, and in case of a
compared to the more differentiated osteoblasts. chemical bond between the metallic implant and
The inhibition of osteoblast cell growth for any metal bone, the displacement is restricted to atomic dis-
other than titanium and zirconium suggests that tances.
these elements only have the capacity for osseoin-
tegration, that is, a rather distinct property.
F
A ..$ Torque Pull - off
Osseointegration
Titanium has the surprising property that it can bind
to living tissue and to bone. It is interesting to read Fig. 11. Forces (F) on a dental implant have axial (A) and
Leventhal in 1951 (29): transverse (T) components which generate shear and ten-
sile stresses at the interface between the metallic implant
(M) and the bone (B). The ultimate stresses for the bond
Bone reaction was studied by the insertion of up between metal and bone are measured as push-out load,
to 80 screws into the femora of rats ... At the end as release torque and as pull-off load.

12
Titanium - the material o f choice?

Natural teeth and implants that support a dental followed by a rapid increase of torque and a maxi-
prosthesis will generally be loaded by both forces mum or leveling out of this torque after about 5
and moments (torque). The mechanical force is a months. This timetable for incorporation of a dental
vector quantity and it has a magnitude and a direc- implant is similar to that of bone fracture healing,
tion (Fig. 11). Mericske-Stern et al. (36-38) made whereby a short resorption period is followed by
simultaneous force measurements on endosseous bridging a gap with less organized bone and then
implants along three axes and found that transverse stabilization through generalized bone remodeling.
forces cannot be neglected. Using the Pythagorean These processes can be followed by sequential label-
theorem, it is found that transverse components ing. Using this technique in the late healing-in phase
reach 10-50% of the vertical force. The loading of an of a titanium implant, it was found that hard bone
implant is skewed, with angles up to 30”. In conse- formation occurs from the metal surface outwards,
quence, different kinds of forces, shear, tension and a clear indication of total integration.
compression, will act in the implant-bone interface. The removal torque and the push-out force are
To determine the failure limit of the interface, “push- proportional to the interfacial shear strength. The
out” tests of cylinders and “release torque” tests of length of the contact between implant and bone en-
screws in bone are performed to give an interfacial ters the formula and in the case of a screw, this true
shear strength (Fig. 11).“Pull-off’’tests, on the other length can be expressed through an effective pitch
hand give the interfacial tensile strength. Selected height leff=1.51, the factor being obtained from geo-
examples of such measurements for various healing metric arguments or empirically (17).Representative
times are shown in Fig. 12. results from push-out and removal torque tests and
different surface preparations are collected in Table
1. At 12 weeks after insertion, the interfacial shear
Mechanical aspects
strength ranges from 9 to 19 MPa for cortical bone
An implant in the form of a screw will resist push- and from 2 to 6 MPa for trabecular bone. The results
out, but this shape requires a bond if it is to resist clearly indicate that surface roughness and texture is
torsion, which tends to loosen the implant along the a primary factor in the anchorage of dental implants.
threads. The resistance to loosening is measured as For longer times, as displayed in Fig. 12, the maxi-
a removal torque (Fig. 12, left). It is noted that the mum interfacial shear strength reaches 24 MPa for
figure reports measurements without any normaliza- machined implants and nearly 50 MPa for special
tion for screw geometry and bone thickness. The re- preparations. These ultimate bond strengths are
lease forces show a latency period of 2 to 3 weeks, high.

I I I

TPS and corundum blasted

Time after insertion of implant, weeks Time after insertion of implant, weeks

Fig. 12. Shear and tensile resistance of the titanium-bone blasted-etched (SLA).Right. Measurements of the pull-off
bond. Left.Raw data for the removal torque of screws (21, force for titanium plasma-sprayed (TPS)and blasted sur-
54, 62); average values are shown. Surface preparation: faces (44, 54).
machine cut, titanium plasma-sprayed (TPS), corundum-

13
Steinemann

I Table 1. Interfacial shear strength for bone-implant fixation made of commercially pure titanium
I

Experiment (12 weeks after insertion)


Implant Animal, bone removal torque push-out load shear strength References
Machined screw rabbit, corticalb 38N.cm 9.3 MPaa 60
Screw, blasted A120325 pm rabbit, corticalb 42 N.cm 10 MPaa 61
Screw, blasted A120375 pm rabbit, corticalb 50N.cm 12 MPaa 61
Cylinder, blasted A1203fine minipig, trabecular 130 N 1.9 MPaC 64
Cylinder, blasted A1203 rough minipig, trabecular 260 N 3.9 MPaC 64
Cylinder, blasted A1203 rough+etched minipig, trabecular 390 N 5.8 MPaC 64
Screw, titanium plasma-coated sheep and dog 200 N . cm 15 MPad 54, 62
Screw, blasted N203rough+etched sheep, corticale 250 N . cm 19 MPad 62
a The relationship between removal torque T and interfacial shear strength T is T=d/2 d x leff.T = C 7,. with c= 19 mm3 per pitch height (0.6 mm).
The thickness of bone is 1.25 mm.
The relationship between push-out load P and interfacial shear strength T is P=s T, with surface area s = d x .1=67 mm2,
The relationship between removal torque T and interfacial shear strength T is T=c T, with c=80 mm3 per pitch height (1.75 mm).
The thickness of bone is 3 mm.

I Table 2. Interfacial tensile strength for bone-implant fixation made of titanium


Experiment
Implant Animal, bone duration pull-off load tensile strength References
Titanium plasma-coated disc, 20 mm* Macacu sp., ulna 30 weeks 76 (30-120) N 2.7 MPa 54
Polished Ti&V disc, 39 mm' dog, cortical 43 weeks 25 (041) N 0.6 MPa 58
Hydroxyapatite-coated disc, 39 mm2 dog, cortical 43weeks 107N 2.7 MPa 58
Commercially pure titanium, blasted
and plasma coated rabbit, epiphysis 24 weeks 107 N 1.6-3.1 MPa 44
The relationship between pull-off load P and interfacial tensile strength a is P=s a, where s is surface area.

Interestingly, the pull-out interfacial shear figure is of the order of the compressive strength of
strengths for freshly inserted BrHnemark (Nobel Bi- human trabecular bone (2-10 MPa). Fragments of
ocare AB, Goteborg, Sweden) screw implants (7, 8) bone are commonly observed on the fractured test
and small bone fixation screws (49) are less than the pieces, which indicates a less dense bone in the con-
above-reported results: 31 MPa and 34-39 MPa re- tact zone.
spectively. These latter figures can represent a To give an idea how strong this bond is, reference
"notched shear strength' and a practical limit when can be made to a classical work in physics. The Mag-
acute threads engender high local stresses. Likewise, deburg half-spheres experiment measured the force
an implant inserted in the mandibula or maxilla will necessary to break vacuum; it needs the (negative)
be subject to similar effects that limit the anchorage. pressure of 1 atmosphere, equal to 0.1 MPa. The ad-
For comparison, human bone has a shear strength hesion between two bodies can never exceed this
of 68 MPa when measured in torsion along the bone stress if there is empty space or free water present in
axis, but the property falls strongly for other orien- the bond region. The interfacial tensile strength of
tations (45). the metal-bone bond is 25 times larger than this
Pull-off tests have the characteristic feature that a limit. Thus, one may ask what is the glue?
strong interaction between the metallic implant and
living hard tissue, that is, the adhesion reaction, is
Material aspects
observed only beyond about 10 weeks (Fig. 12, right).
Perfect immobilization of the test implant is always Corrosion can be the cause of a tissue reaction. An
required. Table 2 collects data for conditions when example is gold, which has a strong redox activity
bonding is completed. For titanium, the interfacial in the chloride-containing tissue fluids and whose
tensile strength is 2.5 MPa on average, and hydroxy- corrosion current density is not far above a practical
apatite-coated titanium does not fare better. The limit for visible attack (see earlier). Sequestration of

14
Titanium - the material of choice?

the foreign body is observed in soft tissue, and im- Metal ion J, J Oxygen ion
plants made of gold provoke bone resorption (59).
Stainless steel, CoCrMo (vitallium) and Ti&14Vhave
much better corrosion resistance; nevertheless,
sequestration and a local reaction are observed in
soft tissue because of leaching, even in minute quan-
tities, of toxic components from the alloys. Bone
screws made of stainless steel loosen after some
weeks, and screw implants made of vitallium and ti- 0 . 0 0 .
tanium alloy have reduced removal torque (Table 3). 0 0 . 0 . .
Push-out tests in trabecular bone all show the effect
of surface texture, but the enhancement of anchor- Fig. 13. Surface oxide on a metal. In the case of titanium,
the oxide grows by the transport of oxygen from the exter-
age by etching pure titanium is absent for the alloys nal surface towards the metal-oxide interface where the
(compare with Table 1). It is concluded that vitallium oxidation reaction proper takes place.
and all titanium alloys used for bone surgery have no
capacity or a reduced capacity for total integration. A
chemical effect of the metal is manifest. On the other
hand, zirconium and niobium have interfacial shear tanium at the base of the oxide (35).This mechanism
strength values not less than that of titanium. is unique for titanium and some other elements of
valency IV, such as silicon and zirconium. Ordinarily,
both the oxygen anion and the metal cation migrate
Chemistry of osseointegration when a metal undergoes oxidation, or corrosion. The
specific mode of oxide growth on titanium has the
Titanium is a reactive metal. This means that, in air, positive effect that no metal ion will reach the sur-
water or any other electrolyte, an oxide is spon- face and be released into the electrolyte.
taneously formed on the surface of the metal (Fig. Titanium oxide is a good insulator even in the
13).The native oxide film has a thickness of about 4 form of a very thin surface film, which may be in a
nm, or 20 times the interatomic distance. Its mode crystalline or in a glassy state and which can occlude
of growth is specific in that the oxygen ions migrate anionic impurities such as chlorine, fluorine or
towards the metal and react with the counter-ion ti- phosphates (12, 34). This behavior differs from that

Table 3. Interfacial shear strength for bone-implant fixation made of titanium alloys and other materials
Experiment (12 weeks after insertion)
removal push-out shear
Implant Animal, bone torque load strength References
Vitallium, machined screw rabbit, corticalg 12 N . cma 3.4 MPae 24
Zirconium, machined screw rabbit, corticalg 26 N . cmb 6.5 MPae 26
Niobium, machined screw rabbit, corticalg 33 N . cmc 9.1 MPaf 25
Ti&14V,machined screw rabbit, corticalg 16 N.cmd 4.3 MPae 22
Ti&14V,cylinder, blasted A1203,fine minipig, trabecular 120 N 1.8 MPah 64
rough
Ti&14V,cylinder, blasted A1203, minipig, trabecular 230 N 3.4 MPah 64
Ti&14V,cylinder, blasted rough+etched minipig, trabecular 230 N 3.4 MPah 64
Ti&17Nb,cylinder, blasted A1203, fine minipig, trabecular 110 N 1.6 MPah 64
Ti&17Nb,cylinder, blasted A1203, rough minipig, trabecular 280 N 4.2 MPah 64
Ti&17Nb,cylinder, blasted rough+etched minipig, trabecular 230 N 3.4 MPah 64
Hydroxyapatite-coatedTi, cylinder minipig, trabecular 830 N 12 MPah 64
a The measurement for commercially pure titanium gives 25 N . cm.
Commercially pure titanium has 26 N cm.
Commercially pure titanium has 25 N cm.
Commercially pure titanium has 23 N . cm.
The relationship between removal torque T and interfacial shear strength T is T=d/2 . d n I,#' T = C . T, with (9 c=17 mm? per pitch height (0.6 mm), and
(f) c=19 mm3 per pitch height (D.6 mm).
8 The thickness of bone is 1.25 mm.
The relationship between push-out load P and interfacial shear stress I[ is P=s .I[,with surface area s=d . B . 1=67 mm'.

15
Steinemann

of the passive film on gold, stainless steel and vitalli- exchange with phosphates. Further, alcohols react
um, which is an electron conductor. The availability fairly easily with the acidic OH groups to form esters.
of electrons at the outer interface towards the elec- The amphoteric or zwitterionic nature of the oxide
trolyte or living contact tissue implies that redox pro- surface can be weighed by counting the effective
cesses, that is, electron exchange processes, can pro- positive and negative charges at all surface sites, and
ceed. Or, this is one mechanism leading to denatura- this can be measured by potentiometric titration. A
tion of macromolecules and sequestration at a larger characteristic value for which the positive and nega-
scale. Gold, stainless steel and cobalt-chromium al- tive charges at the surface cancel each other is called
loy show this response; titanium does not. the zero point of charge. For TiO, in the two struc-
Most metal oxides are hydroxylated at room tem- ture variants of rutile and anatase zero point of
perature and when water or its vapor has access to charge =pH 6, that is, near to neutrality (9, 40). For
the surface (5, 39). In fact, a water molecule adsorb- other oxides of interest, A1203 and ZrO,, zero point
ed on a bare metal site loses a proton, which jumps of charge =pH 9, and for Si02 zero point of charge
to the neighboring surface oxide ion, leaving a hy- =pH 2 (39). At physiological pH values, the surface
droxyl (OH) on the metal ion (Fig. 14). These reac- of titanium oxide has no charge, whereas aluminum
tions lead to two types of hydroxyl groups. In the oxide and zirconium oxide have negative and quartz
case of titanium dioxide, one is bound to one ti- has positive charges. Zeta potentials, connected to
tanium-ion site (terminal OH), and the other to two the zero point of charge, have also been measured
such sites (bridging OH), and these would be ex- for natural and synthetic apatites, giving a pH of 6-
pected to exhibit quite different types of chemical 7 (50), similar to that of titanium oxide.
behavior. The bridging groups must be strongly po- Schindler et al. (14, 47, 48, 63) have studied the
larized by the cations and therefore be acidic in adsorption of amino acids and some metal ions on
character, whereas the terminal OH group can be crystalline titanium oxide and give a complete pic-
predominantly basic and exchangeable with other ture of how the surface hydroxyls exchange charges,
anions. This surface has an amphoteric nature: it has that is, their protons, with the dissolved ligands.
the ability to bind with both acids and bases. The These exchanges are controlled by the pH of the sol-
acid groups react, for example, with amines (such as vent, that is, the environment. Strong interactions
ammonia and methylamine), and basic OH groups are involved. In fact, the energy calculated from re-
action constants derived from adsorption coverage
numbers (63) for amino acids amounts to 20-50
kJ/mole, and the energy necessary for thermal de-
sorption of glycine on platinum indicates binding
energy values of roughly 100 kJ/mole (11). These
numbers are characteristic of real chemisorption
processes involving the formation of covalent or

r-r
H ionic chemical bonds. The necessary proof that such
bonds exist comes from photoelectron diffraction
experiments with adsorbed glycine on single-crystal
rutile, which show the correct distances and angles
among carbon atoms of the amino acid (41).
The spontaneously formed thin surface oxide film
H Oxygen ion H H H
on titanium metal is apparently in an amorphous or
glassy state (35). But photoelectron spectroscopy in-
dicates that the stoichiometry of the crystalline oxide
is preserved and that the amorphous oxide film is
also hydroxylated at its surface (34). Adsorption and
desorption experiments with the oxidized metal by
Gold et al. (15, 16) showed a similar amphoteric re-
Fig. 14. Surface of a metal oxide. Top. Naked surface of action, as displayed by crystalline oxides (5, 14, 63).
oxide. Middle. In the presence of water, the surface metal In a first series of experiments (Fig. 15), the adsorp-
ions bind water molecules. Bottom. Dissociation and
chemisorption of the water molecules leads to an ap- tion of cysteine in a salt solution of variable pH was
parent monolayer of OH groups, a “hydroxylatedsurface. monitored by X-ray excited photoelectron spec-
Source: Schindler (47). troscopy. In a second series of experiments (Fig. 16),

16
Titanium - the material of choice?

organic matter, probably by destroying a primary


structure.
The strong adhesion between bone and tissue is
quite a puzzle. The structure of a tissue and interac-
tion in cohesion occur at rather different scales. In
fact, the space between bone trabeculae or vessels
d in soft tissue is of the order of mm, and the size of
0 I

cells is of the order of pm, whereas the interaction


I
0 4 7 10 pH 14

Fig. 15. Cysteine adsorption on titanium vs pH as deter- distance of chemical forces, the Debye length, is only
mined by photoelectron spectroscopy. Proton exchange in the nm range. One wonders how the chemistry
reactions on the amino acid occur at pH=2 and = l l . at the substrate relatively remote from a cell surface
Source: Gold et al. (15). “protrudes” through to influence interaction that re-
sults in binding.
Organic compounds comprise various groups of
atoms that account for the compounds’ character-
istic reactions (3).These groups, about 15 in number,
are called functional groups. Among them, the rela-
tively simple chemical groups such as hydroxyl (OH),
carbonyl (CO), carboxyl (COOH), amine (NH,) and
sulfonate (S03H) can change the surface properties
of organic matter, including rendering these surfaces
hydrophobic or hydrophilic, and they can modulate
cell adhesion on polymers and on metals (10,27, 57).
- oo l The ground substance identified in the near con-
P I
.- t tact zone of bone toward titanium, zirconium, ni-
obium, tantalum metals (2,23,30)has the properties

Ti O2 f + - f
-0.4 0.0 E-E (H*/Oz),V I .o
Fig. 16. Desorption of soft tissue residues on titanium
after exposure to solutions of variable pH (top) and of
variable redox potential (bottom). The thickness of the or- +- I 2- I
ganic films before the experiments was 6.3-8.7 nm. Posi-
tive potential differences correspond to oxidizing electro-
lytes. Source: Gold et al. (16).
Arg .R I ;f+ I ?+ I 20 I-
OOJ

OxidexG [ q I
the adherent tissue on retrieved titanium implants
Oxidex D 1 1
was examined by controlled desorption treatments OxidexR 7 1
in solutions of variable pH and of variable oxidation 1 I I
strength (redox potential); again, the organic matter PH 0 7 14
at the surface was monitored by photoelectron spec- Fig. 17. Affinity for surface complex formation between
troscopy. Adsorption and desorption show a type of titanium oxide and the RGD series of amino acids (argi-
nine-glycine-aspartic acid). Bar 1. Surface charge of the
mirror response, as expected (compare Fig. 15 with
oxide. Bars 2-4. Charge of glycine, aspartic acid, arginine.
Fig. 16). The proton exchanges on carboxyl and Bars 5-7. Binding, as represented by folding the surface
amino groups of the amino acids favor binding in charges of the oxide with the charges on the binding sites
the first experiments, but the opposite, that is, a re- of the amino acid, for the whole molecule and for the side
lease of proteinaceous residues in the second series chain only (cross-hatched). No charged aqueous titanium
hydroxide species do exist for pH between 2.5 and 11.2.
of experiments. The results further indicate a strong
The signature of charges for the carboxyllamino and for
influence of the redox reaction on desorption; strong the side group of the amino acid is shown (+ positive,
oxidizers such as potassium permanganate, hydro- - negative, 0 uncharged). The dissociation constants are
gen peroxide and chloric acid can remove adherent from Lehninger (28).

17
Steinemann

Ti02 f + - f 1 the carboxyl group can only bind with the basic site
(terminal OH) and the amino group can only bind
Ca (11) Caz+
with the acidic site (bridging OH), whereas the side
chain functional groups will exchange protons with
one or the other surface site matching in character.
The latter case corresponds to the molecular state
OxidexCa with the intact peptide links. This great selectivity of
OxidexPh 1 1 interactions is the consequence of the zwitterionic
property of both reactants.
I I
I
Experiments indicate that titanium oxide has four
PH 0 7 14
or five reactive groups of acidic and basic character
Fig. 18. Affinity for surface complex formation between ti- per nm2 of surface (5). On the other hand, the vol-
tanium oxide and inorganic calcium and phosphate. Bar 1.
Surface charge of the oxide. Bars 2,3. Charge of dissolved
ume of an amino acid molecule in the Arg-Gly-Asp
Ca(I1) and P (V) species. Bars 4,5. Binding as represented tripeptide is 0.1 to 0.2 nm3; thus, if these amino acids
by folding the surface charges on the oxide with the charges are spread on a surface, about 4 molecules will cover
of the chemisorbed ligand. No charged aqueous titanium an area of 1 nm2. This number of molecules matches
hydroxide species do exist for pH between 2.5 and 11.2.The well with the number of available bonds on the inor-
hydrolysis data are from Pourbaix et al. (43).
ganic substrate.
It is known from photoelectron spectroscopy that
calcium and phosphate ions chemisorb on the ox-
of a general “glue” and certainly comprises adhesion ide-covered titanium (19). A binding mechanism can
proteins having the ability to promote cell adhesion also be modeled by electrostatic interaction (Fig. 18).
and to interact with matrix molecules and with a for- It is not calcium phosphate that precipitates on the
eign material. A simple electrostatic model of the in- metallic substrate; the 2-valent Ca ion binds to oxy-
teraction of an adhesion protein with an oxide sur- gen after deprotonation, and the phosphate makes a
face having receptor sites will now be given. bidentate link to surface atoms. As in the case of the
Fibronectin has emerged as a prototype adhesion organic ligand, these selective interactions are the
protein, and the identification of the tripeptide se- consequence of the zwitterionic behavior of the
quence arginine-glycine-aspartic acid (RGD) as the oxide.
focal domain where cells attach to the surface of the The simple electrostatic force model proved to be
large molecule was a major breakthrough in molecu- a versatile tool in understanding the true mechanism
lar biology (42, 46). The charge structure of the three of osseointegration. It turns out that the native oxide
amino acids and their apparent dissociation con- on titanium is a receptor of short- and perhaps also
stants, that is, the pH values at which protonation long-range interactions for adhesion-promoting fac-
and deprotonation of the functional groups occur, tors of living tissue.
are shown in Fig. 17. If the organic molecule adheres
to the surface, its charges will interact with the am-
photeric oxide in a strictly local manner. A proton Conclusion - the dedicated metal
is exchanged and electrostatic energy is gained. The
mechanism is selective for the functional group of For modern metals, the surgeon expects full tissue
the amino acid and for the surface site of the oxide; compatibility. Unalloyed titanium is the reference.

Table 4. Mechanical data for a dedicated metal


Yield strength to Notched tensile strength
Material Yield strength elastic modulus to (plain) tensile strength
316L stainless steel, cold-worked 730 MPa 0.38% 1.7
Commercially pure titanium, grade 2 330 MPa 0.32% 1.8
Commercially pure titanium grade 4, cold-worked 690 MPa 0.66% 1.6
Titanium alloys ( T a V Ti&Nb) 920 MPa 0.84% 1.4
Cortical bone 0.67%

!8
Titanium - the material of choice?

But that metal has limited mechanical strength. High 4. Baes CF, Mesmer RE. The hydrolysis of cations. New York
John Wiley & Sons, 1976.
mechanical properties are needed for structural ef-
5. Boehm Hl? Acidic and basic properties of hydroxylated
ficiency in surgical and dental implants. Their vol- metal oxide surfaces. Disc Faraday SOC1971: 52: 264-275.
ume is in fact restricted by anatomic realities, which 6. Brinemark PI, Zarb GA, Albrektsson T, ed. Tissue-inte-
require good yield and fatigue strength of the metal. grated prostheses. Osseo-integration in clinical dentistry.
On the other hand, an implant “bridges” the forces Carol Stream, I L Quintessence Publishers, 1985: 11-76,
in bone: it may reduce the normal physiological level 129-143.
7. Brunski JB. Biomechanics of dental implants: future re-
of forces. This should give advantage to a “less-rigid’ search directions. J Dent Educ 1988: 52: 775-787.
bone plate or hip prostheses so as to retain a func- 8. Brunski JB, Skalak R. Biomechanical considerations. In:
tional load on bone. Frequently this suggestion is as- Worthington r: Brlnemark PI, ed. Advanced osseointegr-
sociated with a low elastic modulus, comparable to ation surgery: applications in the maxillofacial region.
bone. The conjecture is misplaced; the implant must Carol Stream, I L Quintessence Publishers, 1992: 15-39.
9. Cornell RM, Posner AM, Quirk JF? A titrimetric and electro-
“guide, transport and distribute” forces, and to do phoretic investigation of the PZC and the IEP of pigment
this it must be stiff. The useful comparison of prop- rutile. J Colloid Interface Sci 1975: 53: 6-13.
erties refers to functional aspects (Table 4). The yield 10. Curtis ASG, Forrester Tv, Clark F? Substrate hydroxylation
strength has to do with load capacity and possible and cell adhesion. J Cell Sci 1986: 86: 9-24.
distortion failure of the implant and associated ele- 11. Ernst KH, Christmann K. The interaction of glycine with a
platinum (111) surface. Surface Sci 1989: 222: 277-310.
ments. A second quantity is the elastic modulus, re-
12. Frey N, Buchillier T, LC VD, Steinemann SG. Properties of
lated to the stiffness of the implant. The ratio of surface oxides o n titanium. In: Froes FH, Caplan I, ed. Ti-
these two properties, a dimensionless quantity, is the tanium ’92 Science and Technology. Warrendale, PA Min-
permissible strain, which has to do with deformation erals, Metals & Materials Society, 1993: 2135-2141.
capacity of a mechanical construct. It equals 0.67% 13. Gerber HW, Moosmann A, Steinemann S. Bioactivity of
for human cortical bone. Differences between ma- metals - tissue tolerance of soluble or solid metal, tested
on organ cultured embryonic bone rudiments. In: Buch-
terials are substantial. High yield stress and permiss- horn GH, Willert HG, ed. Technical principles, design and
ible elastic strain do characterize the “forgiving safety of joint implants. Seattle: Hogrefe & Huber, 1994:
metal” in the clinic; in fact, implants should have 248-254.
spare strength to overcome an unfavorable anatomic 14. Gisler A. Die Adsorption von Aminosauren an Grenzflachen
situation and lack of cooperation by the patient. The Oxyd-Wasser. Dissertation. Berne: University of Berne,
1981.
penalties associated with implant failure are always
15. Gold JM, Schmidt M, Steinemann SG. X P S study of amino
great. Another entry in the table has to do with the acid adsorption to titanium surfaces. Helv Phys Acta 1989:
safety of the implant. The notch sensitivity is import- 62: 246-249.
ant for design; this parameter must exceed 1.2, to 16. Gold JM, Schmidt M, Steinemann SG. X P S study of re-
avoid cracks in the shape variations of common im- trieved titanium and Ti alloy implants. In: Heimke G, Sol-
tksz U, Lee AJC, ed. Clinical implant materials. Amsterdam:
plants. It is found that the properties of titanium
Elsevier Science Publishers, 1990: 69-74.
metals are favorable for making surgical implants in- 17. Gotfredsen K, Nimb L, Hjorting-Hansen E, Jensen JS, Hol-
trinsically safe and damage tolerant. men A. Histomorphometric and removal torque analysis
The metal titanium must not be a foreign body in for Ti0,-blasted titanium implants. An experimental study
living tissue, at least regarding a chemical and on dogs. Clin Oral Implants Res 1992: 3: 77-84.
physiological “insult”. For integration, it has unique 18. Gwynn 1010 A€?Cell biology at interfaces. J Mater Sci Mater
Med 1994: 5: 357-360.
properties among all metals in the periodic table. 19. Hanawa T, Asaoka K, Ukai H, Murakami K. Calcium phos-
The question at the outset of this work can be phate precipitation on titanium in electrolyte and its accel-
answered: yes - titanium is the material of choice! eration by calcium-ion-implantation. In: Kovacs rl Iste-
phanous NS, ed. Compatibility of biomedical implants.
Corrosion and organic and biological electrochemistry di-
visions. Proc Vol. 94-15. Pennington, NJ: The Electrochemi-
References cal Society, 1994: 126-137.
20. Hillmann G, Donath K. Licht- und elektronenmikroskopis-
1. Albrektsson T, Brlnemark PI, Hansson HA, Lindstrom J. che Untersuchung zur Biostabilitat dentaler Titanimplant-
Osseointegrated titanium implants. Acta Orthop Scand ate. Z Zahnarztl Implantol 1991: 7: 170-177.
1981: 52: 155-170. 21. Johansson C, Jacobsson M, Albrektsson T. Removal forces
2. Albrektsson T, Hansson HA, Ivarsson B. Interface analysis for osseointegrated titanium implants. In: de Putter C, de
of titanium and zirconium bone implants. Biomaterials Lange GL, de Groot K, ed. Implant materials in biofunction:
1985: 6: 97-101. advances in biomaterials. Amsterdam: Elsevier Science
3. Atkins PW, Beran JA. General chemistry. New York Scien- Publishers, 1988: 8: 87-92.
tific American Books. 1992: 887-917. 22. Johansson CB, Albrektsson T, Thomsen P. Removal torques

19
Steinemann

of screw-shaped c.p. titanium and Ti&14Vimplants in rab- 40. Parks GA. The isoelectric points of solid oxides, solid hy-
bit bone. In: Heimke G, SoltBsz U, Lee AJC, ed. Clinical im- droxides, and aqueous hydroxo complex systems. Chem
plant materials: advances in biomaterials. Amsterdam: El- Rev 1965: 65: 177-198.
sevier Science Publishers, 1990: 9: 87-92. 41. Patthey L. Photoelectron diffraction and photoemission
23. Johansson CB, Hansson HA, Albrektsson T. Qualitative inter- studies of some semiconductor and oxide surfaces. PhD
facial study between bone and tantalum, niobium or com- thesis. Lausanne: University of Lausanne, 1995.
mercially pure titanium. Biomaterials 1990: 11: 277-280. 42. Pierschbacher MD, Ruoslahti E. Cell attachment activity of
24. Johansson CB, Sennerby L, Albrektsson T. A removal torque fibronectin can be duplicated by small synthetic fragments
and histomorphometric study of bone tissue reactions to of the molecule. Nature 1984: 309: 30-33.
commercially pure titanium and vitallium implants. Int J 43. Pourbaix M, de Zoubov N, van Muylder J, Deltombe E,
Oral Maxillofac Implants 1991: 6: 437-441. Schmets J, Vanleugenhaghe C, Moussard M, Besson J,
25. Johansson CB, Albrektsson T. A removal torque and histo- Brenet JE Burgers WG, Charlot G, Carrels RM, Hoar TP, Jolas
.
morphometric study of commercially pure niobium and ti- F, Kunz W, Maraghini M, Piontelli R, Schwabe K, Valensi G,
tanium implants in rabbit bone. Clin Oral Implants Res van Rysselberche I: Pitman AL. Atlas dkquilibres electro-
1991: 2: 24-29. chimiques. Paris: Gauthier-Villars, 1963.
26. Johansson CB, Wennerberg A, Albrektsson T. Quantitative 44. Probster L, Voigt C, Fuhrmann G, Gross VM. Tensile and
comparison of screw-shaped commercially pure titanium torsional shear strength of the bone implant interface of
and zirconium implants in rabbit tibia. J Mater Sci Mater titanium implants in the rabbit. J Mater Sci Mater Med
Med 1994: 5: 340-344. 1994: 5: 314-319.
27. Kothari S, Hatton PV, Douglas CWI. Protein adsorption to 45. Reilly RT, Burstein AH. The elastic and ultimate properties
titanium surfaces. J Mater Sci Mater Med 1995: 6: 695-698. of compact bone tissue. J Biomech 1975: 8 : 393-405.
28. Lehninger AL. Biochemistry. New York Worth Publishers, 46. Ruoslahti E, Hayman EG, Pierschbacher MD. Extracellular
1975. matrices and cell adhesion. Arteriosclerosis 1985: 5 : 581-
29. Leventhal GS. Titanium, a metal for surgery. J Bone Joint 594.
Surg 1951: 3 3 A 473. 47. Schindler PW. Surface complexation. In: Sigel H, ed. Metal
30. Linder L, Alberktsson T, Brinemark PI, Hansson HA, Ivars- ions in biological systems. New York Dekker, 1984: 18: 105-
son B, Jonsson U, Lundstrom 1. Electron microscopic 135.
analysis of the bone-titanium interface. Acta Orthop Scand 48. Schindler PW. Co-adsorption of metal ions and organic
1983: 54: 45-52. ligands: formation of ternary surface complexes. In: Hoch-
31. Listgarten MA, Buser D, Steinemann SG, Donath K, Lang ella ME White AE ed. Mineral-water interface geochemis-
NF: Weber HP Light and transmission electron microscopy try. Reviews in mineralogy. Washington, DC: Mineralogical
of the intact interfaces between non-submerged titanium- Society of America, 1990: 23: 281-307.
coated epoxy resin implants and bone or gingiva. J Dent 49. Shelton JC, Loukota RA. Pull-out strength of screws from
Res 1992; 71: 364-371. cortical bone in the maxillofacial region. J Mater Sci Mater
32. Luckey TD, Venugopal B. Metal toxicity in mammals. Med 1996: 7: 231-235.
Physiological and chemical basis. New York Plenum Press, 50. Somasundaran P, Wang YHC. Surface chemical character-
1977: 1: 103-128. istics and adsorption properties of apatite. In: Misra DN,
33. Maurer AM, LC VD, Steinemann SG, Guenther H, Bille J. ed. Adsorption on and surface chemistry of hydroxyapatite.
Surface chemistry and cell reactions. 10th Eur Conf Bio- New York Plenum Press, 1984: 129-149.
materials. Davos: in press. 51. Steflik DE, Sisk AL, Parr GR, Lake RT, Hanes PJ, Berkery DJ,
34. Mausli PA, Bloch PR, Geret V, Steinemann SG. Surface char- Brewer €? Transmission electron and high-voltage electron
acterisation of titanium and Ti-alloys. In: Christel P, Meun- microscopy of osteocyte cellular processes extending to the
ier A, Lee AJC, ed. Biological and biomechanical perform- dental implant surface. J Biomed Mater Res 1994: 28: 1095-
ance of biomaterials. Amsterdam: Elsevier Science Pub- 1107.
lishers, 1986: 57-62. 52. Steinemann SG. Corrosion of titanium and titanium alloys
35. Mausli PA, Steinemann SG, Simpson J€? Properties of sur- for surgical implants. In: Lutjering G, Zwicker U, Bunk W,
face oxides on titanium and some titanium alloys. In: Lac- ed. Proc 5th Int Conf Titanium. Oberursl: Deutsche Ges
ombe P, Tricot R, Beranger G, ed. 6th World Conference on Metallk, 1985: 1373-1379.
Titanium. Les Ulis: Les editions de physique, 1988: 1759- 53. Steinemann SG, Perren SM. Titanium alloys as metallic bio-
1764. materials. In: Lutjering G, Zwicker U, Bunk W, ed. Proc 5th
36. Mericske-Stern R, Geering AH, Burgin WB, Graf H. Three- Int Conf Titanium. Oberursl: Deutsche Ges Metallk, 1985:
dimensional force measurements o n mandibular implants 1327-1334.
supporting overdentures. Int J Oral Maxillofac Implants 54. Steinemann SG, Eulenberger J, Maeusli PA, Schroeder A.
1992: 7: 185-194. Adhesion of bone to titanium: biological and biomechan-
37. Mericske-Stern R, Assal P, Buergin W. Simultaneous force ical performance of biomaterials. In: Christel Meunier A,
measurements in 3 dimensions on oral endosseous im- Lee AJC, ed. Advances in biomaterials. Amsterdam: Elsevier
plants in vitro and in vivo. Clin Oral Implants Res 1996: 7: Science Publishers, 1986: 6: 409414.
378-386. 55. Steinemann SG, Mausli PA. Titanium alloys for surgical im-
38. Mericske-Stern R, Piotti M, Sirtes G. 3-D in vivo force meas- plants - biocompatibility from physicochemical principles.
urements on mandibular implants supporting overden- Proc 6th World Conf Titanium. In: Lacombe F: Tricot R, BC-
tures. Clin Oral Implants Res 1996: 7: 387-396. ranger G, ed. Les Ulis: Les Bditions de physique, 1988: 535-
39. Parfitt GD. The surface of titanium dioxide. Progr Surface 540.
Membr Sci 1976: 11: 181-226. 56. Steinemann SG. Tissue compatibility of metals from

20
Titanium - the material of choice?

physicochemical principles. In: Kovacs P, Istephanous NS, 60. Wennerberg A, Albrektsson T, Andersson B. An animal
ed. Compatibility of biomedical implants. Corrosion and study of c.p. titanium screws with different surface topo-
organic and biological electrochemistry divisions. Proc Vol. graphies. J Mater Sci Mater Med 1995: 6: 302-309.
94-15. Pennington, NJ: The Electrochemical Society, 1994: 61. Wennerberg A, Albrektsson T, Lausmaa J. Torque and histo-
1-13. morphometric evaluation of c.p. titanium screws blasted
57. Sukenik CN, Balachander N, Culp LA, Lewandowska K, with 25 and 75 pm sized particles of A1203. J Biomed Mater
Merritt K. Modulation of cell adhesion by modification of Res 1996: 30: 251-260.
titanium surfaces with covalently attached self-assembled 62. Wilke HJ, Claes L, Steinemann S. The influence of vari-
monolayers. J Biomed Mater Res 1990: 24: 1307-1323. ous titanium surfaces on the interface shear strength be-
58. Taylor JL, Brunski JB, Hoshaw SJ, Cochran GVB, Higuchi tween implants and bone. In: Heimke G, Soltesz U, Lee
Kw. Interfacial bond strengths of Ti&14V and hydroxyapa- AJC, ed. Clinical implant materials, Advances in Bio-
tite-coated Ti&14Vimplants in cortical bone. In: Laney WR, materials, vol. 9. Amsterdam: Elsevier Science Publishers
Tolman DE, ed. Tissue Integration in Oral, Orthopedic and 1990: 309-314.
Maxillofacial Reconstruction. Proc 2nd Intern Congr Tissue 63. Wirz UE. Die Adsorption proteinogener Aminosauren an
Integration. Carol Stream, IL: Quintessence Publishers Co., TiOz (Anatas). Dissertation University of Berne 1984.
1992: 125-132. 64. Wong M, Eulenberger J, Schenk R, Hunziker E. Effect of
59. Thomsen E: Larsson C, Ericsson LE, Sennerby L, Lausmaa surface topology on the osseointegration of implant ma-
J, Kasemo B. Structure of the interface between rabbit cor- terials in trabecular bone. J Biomed Mater Res 1995: 29:
tical bone and implants of gold, zirconium and titanium. J 1567-1575.
Mater Sci Mater Med 1997: 8: 653-665.

21

S-ar putea să vă placă și