DIPLOMAT SERVICES
Silver Oaks, 502, B Wing, Vasant Garden, Mulund (West), Mumbai{O0 080
Ph. * 21il3227/9820987961. Email: diplomatservicesl @hotmail.com
CALIBRANON CERNFICATE FOR BRINELL HARDNESS TESTING MACHINE
(tNDtRECT VERIFICATtON)

Name Of Customer Eouioment Under Calibration Calibration Conditions


M/S JAYBEE STEEL SAROJ ENGG UDYOG
Make "JAYSHINGPUR" Referred Std : lS 1500 : 2013 (Part - ll)
TREATERS PVT LTD.
No R-524/525, TTC Model "B-3000H" Date Of Calibration : 02t08t2017
Industral Area. M I D.C., Sr. No. TErBfifl524121 Calibration Due Date : 01/08/2018
Rabale, Mumbai - 400701
lD No. Calibrated By: M.T.Gehani
Location 0
Calibrated at Temp. : 29.5 C

DETAILS OF STANDARD TEST BLOCKS USED FOR CALIBRATION :

Avg Value & Scale 187 HBW 284 HBW


Serial No. 412113 422t13
Cal Certificate No.: 1828/ 18294
Cal. Certificate Date: 29t11t2013 29t11t2013
Calibration Bv: Bhartiya Enterprises Bhartiva Enterorises
Iraceabilitv to NABL NABL
Uncertainity of Test Block
INDIRECT VERIFICATION
Ball Size Std. Test Actual Allowed
Diameter of lmpressions in mm Mean Dia (Mm) D-
Sr. (mm)/ Block Repetability Repeatibility
No. Load KN Value
(ks0 (BHN) DI D2 D3 D4 D5 Average Actual Error % Allowed Erroro/o

10 mm 187 4.41 4.41 4.4',1 4.40 4.40 4.4 0.0't 0.11


I
3000 kgs HBW 186 186 t86 187 187 186.4 4.32 2.5%
l0 mm 284 3.62 3.62 3.62 3.61 3.61 3.6 0.0'r 0.07
2
3000 kgs HBW 282 282 282 283 283 282.4 -0.56 2%
AllowedRepeatabilityForH<125-0.030d-,ForH>125<225-0.025d-,ForH>225-0.02d-aspertable2lSl5OO:2013(Part-ll).
{lowedErrorisForH<125!3% 125<225+2.5Vo>225!2.Oolaaspertable2lS1500:2013.(Part-ll)
\pplied dwell time : 10 - 15 Seconds.

Remarks : I ) This machine has been verified and the accuracy error and repeatability error found within limits.
2) Overall deviation is within specified accuracy, hence equipment is OK.

WITNESSED BY:-
Checked By :-
RvrcEs.

Mr. M.T GEHANI.

1) The above Calibration is done at the Site of Customer.


2) The Calibration Results reported in this Certificate are valid at the time of & under the stated
conditions of measurement & is only for the Calibrated item as identified in the Certificate.
3) Calibration Certificate shall not be reproduced except in full, without written approval of DIPLOMAT
SERVICES.

FORMAT NO.F/CAUO3/02 Page No 't of 1