• Power ON
Power switch on CPU Box - All breakers must be ON
Allow time until the interactive windows are displayed before performing any operations
• Power OFF
Allow time until all processes have completed
On S-console: Select Utility - Shutdown and click OK
Allow time until "safe to turn off" before switching OFF
Allow at least one minute after switching OFF before switching ON again
Shutdown the system once per day - for file cleanup, initialization and log management
Shutdown and restart the system after any service intervention - confirm normal operation,
before the next patient
Special Configuration
Register patient
Review settings
for each element - edit if necessary each tab, from left to right
Dual scanogram
Scan plan
Scan
• Service staff access Windows and Unix with caution and according to
validated instructions - service manuals - support information
• Different levels
User's
Service engineer's
woodpecker
• English
• IR100 for special character input - additional setup
RDD Setup
• The RDD
Menu to manage RDD settings
Special Configuration
Quit
3. Hosts table
Add IP addresses and suitable Host names for the connectivity partners
Add an empty line at the end
127.0.0.1 localhost
Make backup
System disk to DVD-RAM
SCAN or DISPLAY
Format DVD-RAM
4
© Toshiba Medical Systems Corporation 49
Format new DVD-RAM
5. Click OK - formatting complete
Click Quit
Click Quit