Documente Academic
Documente Profesional
Documente Cultură
Question Actions
Use Objectives
Why do you want the data and results • Write use objectives
and how will you use the results?
Specifications
How good do the numbers have to be? • Write specifications
• Pick methods to meet specifications
• Consider sampling, precision, accuracy,
selectivity, sensitivity, detection limit,
robustness, rate of false results
• Employ blanks, fortification, calibration
checks, quality control samples, and control
charts to monitor performance
• Write and follow standard operating
procedures
Assessment
Were the specifications achieved? • Compare data and results with specifications
• Document procedures and keep records
suitable to meet use objectives
• Verify that use objectives were met
6 10 14 18 22
Time (min)
Specificity
Specificity is the ability of an analytical method to distinguish analyte from everything else
that might be in the sample. Electrophoresis is an analytical method in which substances are
separated from one another by their differing rates of migration in a strong electric field. An
electropherogram is a graph of detector response versus time in an electrophoretic separation.
Figure 5-1 shows an electropherogram of the drug cefotaxime (peak 4) spiked with 0.2 wt%
of known impurities normally present from the synthesis. A reasonable requirement for speci-
ficity might be that there is baseline separation of analyte (cefotaxime) from all impurities that
might be present. Baseline separation means that the detector signal returns to its baseline
before the next compound reaches the detector.
In Figure 5-1, impurity peak 3 is not completely resolved from the cefotaxime. In this
case, another reasonable criterion for specificity might be that unresolved impurities at their
maximum expected concentration do not affect the assay of cefotaxime by more than 0.5%.
If we were trying to measure impurities, as opposed to assaying cefotaxime, a reasonable
criterion for specificity is that all impurities having 7 0.1% of the area in the electropherogram
are baseline separated from cefotaxime. Figure 5-1 does not meet this criterion.
When developing an assay, we need to decide what impurities to deliberately add to test
for specificity. For analysis of a drug formulation, we would want to compare the pure drug
with one containing additions of all possible synthetic by-products and intermediates, degra-
dation products, and excipients (substances added to give desirable form or consistency).
Degradation products might be introduced by exposing pure material to heat, light, humidity,
acid, base, and oxidants to decompose ~20% of the original material.
Linearity
Linearity measures how well a calibration curve follows a straight line, showing that response
is proportional to the quantity of analyte. If you know the target concentration of analyte in a
drug preparation, for example, test the calibration curve for linearity with five standard solutions
spanning the range from 0.5 to 1.5 times the expected analyte concentration. Each standard
should be prepared and analyzed three times. (This procedure requires 3 ⫻ 5 ⫽ 15 standards
plus three blanks.) To prepare a calibration curve for an impurity that might be present at, say,
0.1 to 1 wt%, you might prepare a calibration curve with five standards spanning the range
0.05 to 2 wt%.
A superficial, but common, measure of linearity is the square of the correlation
coefficient, R 2:
[ g(xi ⫺ x )( yi ⫺ y)] 2
Square of correlation coefficient: R2 ⫽ (5-2)
g(xi ⫺ x) 2 g(yi ⫺ y) 2
where x is the mean of all the x values and y is the mean of all the y values. An easy way to
find R 2 is with the LINEST function in Excel. In the example on page 86, values of x and y
are entered in columns A and B. LINEST produces a table in cells E3:F5 that contains R 2 in
cell E5. R 2 must be very close to 1 to represent a linear fit. For a major component of an un- R 2 can be used as a diagnostic. If R 2 decreases
known, a value of R 2 above 0.995 or, perhaps, 0.999, is deemed a good fit for many purposes.5 after a method is established, something has
For data in Figure 4-11, which do not lie very close to the straight line, R 2 ⫽ 0.985. gone wrong with the procedure.
Accuracy
Accuracy is “nearness to the truth.” Ways to demonstrate accuracy include
1. Analyze a certified reference material in a matrix similar to that of your unknown. Your
method should find the certified value for analyte in the reference material, within the
precision (random uncertainty) of your method.
2. Compare results from two or more different analytical methods. They should agree within
their expected precision.
3. Analyze a blank sample spiked with a known addition of analyte. The matrix must be
the same as your unknown. When assaying a major component, three replicate samples
at each of three levels ranging from 0.5 to 1.5 times the expected sample concentration
are customary. For impurities, spikes could cover three levels spanning an expected range
of concentrations, such as 0.1 to 2 wt%.
4. If you cannot prepare a blank with the same matrix as the unknown, then it is appropriate
to make standard additions (Section 5-3) of analyte to the unknown. An accurate assay will
find the known amount of analyte that was added.
Spiking is the most common method to evaluate accuracy because reference materials are not
usually available and a second analytical method may not be readily available. Spiking ensures
that the matrix remains nearly constant.
An example of a specification for accuracy is that the analysis will recover 100 ⫾ 2% of
the spike of a major constituent. For an impurity, the specification might be that recovery is
within 0.1 wt% absolute or ⫾10% relative.
Precision
Precision is how well replicate measurements agree with one another, usually expressed as a
standard deviation. Many types of precision are distinguished.
Autosamplers in chromatography and graphite Instrument precision, also called injection precision, is the reproducibility observed
furnace atomic spectroscopy, for example, when the same quantity of one sample is repeatedly introduced (ⱖ10 times) into an instrument.
have improved injection precision by a factor Variability could arise from variation in the injected quantity and variation of instrument
of 3–10 compared with that attained by response.
humans.
Intra-assay precision is evaluated by analyzing aliquots of a homogeneous material sev-
eral times by one person on one day with the same equipment. Each analysis is independent,
so the intra-assay precision is telling us how reproducible the analytical method can be. Intra-
assay variability is greater than instrument variability, because more steps are involved.
Examples of specifications might be that instrument precision is ⱕ1% and intra-assay precision
is ⱕ2%.
Intermediate precision, formerly called ruggedness, is the variation observed when an
assay is performed by different people on different instruments on different days in the
same lab. Each analysis might incorporate fresh reagents and different chromatography
columns.
Interlaboratory precision, also called reproducibility, is the most general measure of
reproducibility observed when aliquots of the same sample are analyzed by different people
in different laboratories. Interlaboratory precision becomes poorer as the level of analyte
decreases (Box 5-2).
Confusing terms:
Range
Linear range: concentration range over which Range is the concentration interval over which linearity, accuracy, and precision are all
calibration curve is linear (Figure 4-14) acceptable. An example of a specification for range for a major component of a mixture
Dynamic range: concentration range over is the concentration interval providing a correlation coefficient of R 2 ⱖ 0.995 (a measure
which there is measurable response of linearity), spike recovery of 100 ⫾ 2% (a measure of accuracy), and interlaboratory
Range: concentration range over which precision of ⫾3%. For an impurity, an acceptable range might provide a correlation
linearity, accuracy, and precision meet coefficient of R 2 ⱖ 0.98, spike recovery of 100 ⫾ 10%, and interlaboratory precision
specifications for analytical method of ⫾15%.
procedure. If all results are “similar,” and there is no serious system- +50
atic error, then the method is considered “reliable.”
+40
The coefficient of variation is the standard deviation divided by
25
the mean, usually expressed as a percentage: CV(%) ⫽ 100 ⫻ s/ x . +30
−40
where C is the fraction of analyte in the sample (C ⫽ g analyte/
g sample). The coefficient of variation within a laboratory is about −50
one-half to two-thirds of the between-laboratory variation. −60
0.01%
1 ppm
1 ppb
Experimental results varied from the idealized curve by about a fac-
10%
1%
tor of 2 in the vertical direction and a factor of 10 in the horizontal
direction. About 5–15% of all interlaboratory results were “outliers”— Concentration ( gg analyte
sample )
clearly outside the cluster of other results. This incidence of outliers
is above the statistical expectation. Coefficient of variation of interlaboratory results as a function of sample
The Horwitz curve predicts that when the concentration of ana- concentration (expressed as g analyte/g sample). The shaded region has
lyte is 1 ppm, the coefficient of variation between laboratories is been referred to as the “Horwitz trumpet” because of the way it flares open.
[From W. Horwitz, “Evaluation of Analytical Methods Used for Regulation of Foods and
苲16% . When the concentration is 1 ppb, the coefficient of variation
Drugs,” Anal. Chem. 1982, 54, 67A.]
is 苲45% . If, perchance, you become a regulation writer one day,
acceptable analyte levels should allow for variation among laborato-
ries. The Gaussian distribution tells us that 苲5% of measurements 1 ⫹ 1.65 ⫻ 0.45 ppb, or about 1.7 ppb. This level gives a 5% rate
lie above x ⫹ 1.65s (Section 4-1). If the target allowable level of of false positives that exceed the allowed value even though the true
analyte is 1.0 ppb, the allowed observed amount might be set at value is below 1.0 ppb.
s s
3s
~ 1% of area of
blank lies to right of
detection limit
Signal amplitude
Detection limit
FIGURE 5-2 Detection limit. Distribution of measurements for a blank and a sample whose
concentration is at the detection limit. The area of any region is proportional to the number of
measurements in that region. Only 苲1% of measurements for a blank are expected to exceed the
detection limit. However, 50% of measurements for a sample containing analyte at the detection limit
will be below the detection limit. There is a 1% chance of concluding that a blank has analyte above the
detection limit (false positive). If a sample contains analyte at the detection limit, there is a 50% chance
of concluding that analyte is absent because its signal is below the detection limit (false negative).
Curves are Student’s t distributions for 6 degrees of freedom and are broader than the corresponding
Gaussian distributions.
detection limit and the minimum detectable concentration. (b) What is the concentration of
analyte in a sample that gave a signal of 7.0 nA?
Solution (a) First compute the mean for the blanks and the standard deviation of the sam-
ples. Retain extra, insignificant digits to reduce round-off errors.
Slope = 0.229 nA/μM Blank: Average ⫽ yblank ⫽ 1.26 nA
Sample: Standard deviation ⫽ s ⫽ 0.56 nA
The signal detection limit from Equation 5-3 is
Concentration (μM)
ydl ⫽ yblank ⫹ 3s ⫽ 1.26 nA ⫹ (3)(0.56 nA) ⫽ 2.94 nA
The minimum detectable concentration is obtained from Equation 5-5:
3s (3)(0.56 nA)
Detection limit ⫽ ⫽ ⫽ 7.3 M
m 0.229 nA/M
(b) To find the concentration of a sample whose signal is 7.0 nA, use Equation 5-4:
ysample ⫺ yblank ⫽ m ⫻ concentration
ysample ⫺ yblank 7.0 nA ⫺ 1.26 nA
1 Concentration ⫽ ⫽ ⫽ 25.1 M
m 0.229 nA/M
Test Yourself Find the minimum detectable concentration if the average of the blanks
is 1.05 nA and s ⫽ 0.63 nA. (Answer: 8.3 M)
Nutrition Facts
Serving Size 6 Crackers (28g)
Servings Per Container About 10
CO2H
Amount Per Serving
Stearic acid — a saturated fat
Calories 120 Calories from Fat 40
% Daily Value*
Sugars 0g
Protein 3g
FIGURE 5-3 Nutritional label from a package of crackers. The reporting limit for trans fat is
0.5 g/serving. Any amount less than this is reported as 0. The end of Chapter 6 explains the shorthand
used to draw these 18-carbon compounds.
Robustness
Robustness is the ability of an analytical method to be unaffected by small, deliberate changes
in operating parameters. For example, a chromatographic method is robust if it gives acceptable
results when small changes are made in solvent composition, pH, buffer concentration,
temperature, injection volume, and detector wavelength. In tests for robustness, the organic
solvent content in the mobile phase could be varied by, say, ⫾2%, the eluent pH varied by
⫾0.1, and column temperature varied by ⫾5⬚C. If acceptable results are obtained, the written
procedure should state that these variations are tolerable. Capillary electrophoresis requires
such small volumes that a given solution could conceivably be used for months before it is
used up. Therefore, solution stability (shelf life) should be evaluated for robustness.
By expressing the diluted concentration of analyte, [X]f, in terms of the initial concentration
of analyte, [X]i, we can solve for [X]i, because everything else in Equation 5-7 is known.
Solution From Equation 5-8, the final concentration of Na⫹ after dilution with the standard
is [X]f ⫽ [X]i(Vo /V) ⫽ [X]i(95.0 mL/100.0 mL). The final concentration of added standard is
[S]f ⫽ [S]i(VS /V) ⫽ (2.08 M)(5.00 mL/100.0 mL) ⫽ 0.104 M. Equation 5-7 becomes
3Na⫹ 4 i 4.27 mV
⫹
⫽ 1 3Na⫹ 4 i ⫽ 0.113 M
3 0.104 M4 ⫹ 0.9503 Na 4 i 7.98 mV
Test Yourself If spiked serum gave a signal of 6.50 mV, what was the original concen-
tration of Na⫹? (Answer: 0.182 M)
Figure 5-5 shows data for an experiment in which ascorbic acid (vitamin C) was measured
in orange juice by an electrochemical method. The current between a pair of electrodes immersed
in the juice is proportional to the concentration of ascorbic acid. Eight standard additions increased
current from 1.78 to 5.82 A (column C), which is at the upper end of the desired range of
1.5- to 3-fold increase in analytical signal.
6 The equation of a line is y ⴝ mx ⴙ b. The
y = 0.6463x + 1.8687
x-intercept is obtained by setting y ⴝ 0:
0 ⴝ mx ⴙ b
ns
5 itio x ⴝ ⴚb/m
add
y = lS+X*V/V0
d
dar
an
4 st
ith
w
ed B
ain
obt
3 gs
din
ea
R
2
Concentration
Reading of unknown
of unknown, [X]i
without standard addition
1 FIGURE 5-6 Graphical treatment of standard
A
additions to a single solution with variable total
volume. Data from Figure 5-5. Standard
additions should increase the analytical signal to
–3 –2 –1 0 1 2 3 4 5 6 7 between 1.5 and 3 times its original value (that
x = [S]i*Vs /V0 is, B ⫽ 0.5A to 2A).
Figure 5-6 allows us to find the original concentration of unknown. The theoretical response
is derived by substituting expressions for [X]f and [S]f from Equations 5-8 into Equation 5-7.
Following a little rearrangement, we find
For successive V IX Vs Successive standard additions to one solution:
standard additions IS⫹X a b ⫽ IX ⫹ [S] i a b (5-9)
Vo [X] i Vo V VS
to one solution: 123 123 Plot IS ⴙX a b versus [S]i a b
Function to plot Function to plot V0 V0
on y-axis on x-axis x-intercept is [X]i
where sy is the standard deviation of y (Equation 4-20), |m | is the absolute value of the slope
of the least-squares line (Equation 4-16), n is the number of data points (nine in Figure 5-6),
y is the mean value of y for the nine points, xi are the individual values of x for the nine points,
and x is the mean value of x for the nine points. For the points in Figure 5-6, the uncertainty
in the x-intercept is 0.098 mM.
The confidence interval is ⫾t ⫻ (standard deviation of x-intercept) where t is Student’s t
(Table 4-2) for n ⫺ 2 degrees of freedom. The 95% confidence interval for the intercept in
Figure 5-6 is ⫾(2.365)(0.098 mM) ⫽ ⫾0.23 mM. The value t ⫽ 2.365 was taken from
Table 4-2 for 9 – 2 ⫽ 7 degrees of freedom.
1 2 3 4 5
1 2 3 4 5
IS+X
Fill each flask to the 50-mL mark and mix
Intercept
= [X]f 1 2 3 4 5
FIGURE 5-8 Graphical treatment of standard If all standard additions are made to a constant final volume, plot the signal IS⫹X versus
addition with constant total volume. Plot ISⴙX the concentration of diluted standard, [S]f (Figure 5-8). In this case, the x-intercept is the
versus [S]f, and the x-intercept is [X]f. The lines in final concentration of unknown, [X]f, after dilution to the final sample volume. Equation 5-10
Figures 5-6 and 5-8 are both derived from still applies to the uncertainty. The initial concentration of unknown, [X]i, is calculated from
Equation 5-9. the dilution that was applied to make the final sample.