XK3100

PASO OPERACION DISPLAY NOTICE

1 Encienda, presione “calibration” and “ Enter” al mismo tiempo.

pass
2 Es el pass

------1
Input calibration password Press ↑ to input data press Enter to confirm; press Tare to

3 “010201” modify last bit of password or exit calibration. Password right to enter
d= xx
Press Enter division setting, otherwise exit calibration

d= xx Division setting

4 press ↑ d= yy (1, 2, 5, 10, 20, 50, 100, 200); Press Enter to confirm

press Enter dot= x Confirm the division and enter point setting

dot= x Point setting

5 press ↑ dot= y Press ↑ to modify decimal digits, at best 5 digits

press Enter ZEroxx confirm the decimal digits and enter zero range setting

ZEroxx Zero range setting

press ↑ Zeroyy Press ↑ to modify the zero range, the most zero tracing range will be
6
3d ;Press Enter to confirm

press Enter Full Confirm the setting and enter capacity setting

Full

Full capacity setting(DOS prompt and original setting display in turn)

7 Press ↑ xxxx Press ↑ to modify data, press Enter to confirm present data and

Press Enter LoAd enter next bit, after finish 6 bit will enter weight calibration

LoAd Weight calibration


Add poise
LoAd Set the poise on the platform
Press ↑
8 Press ↑ and Enter to enter poise weight, if the present data,
Press Enter
press Enter , calibration will be finished when the Tare signal is on if
Press Enter
xxxx off, finish data entering
XK3100 A2
step Operation Display Explanation
Under weighing
1 press [ FUNC] mode Enter calibration password 5
[CP 0]
press [5 ]
[CP 5]
press [Enter]

2 press [5] [E *** ] Enter division value selecting

press [Enter] [E 005] 1/2/5/10/20/50/100

Example: 5

step Operation Display Explanation

3 press [ 2] [ dc * ] Enter decimal point (0 -4)

press [Enter ] [ dc 2 ] Example: 2


[F***.**]
4 press [ 6 ] [ 0 ] Enter the overload alarming value(F: F.S.+9d) Note: 1. If do not need
[F060.45]
[ 4 ][ 5 ] calibration, directly press [ Enter] to enter step 7 or press [weigh] to exit

2. If the calibration is needed. You should enter F, even if the original F value is

right

Example: Enter F=60.45


[no LoAd]
5 press [Enter ] Confirming Zero position. At this time there is no load on the scale. Pressing
[----]
[Enter] when the stable light is on. Wait for 5-10 seconds
Enter the loaded weight [AdLoAd1/2/3/4/5] Load the weight near to F.S.
6 [0060.00] Example 60.00kg
press [6][0][0][0] [----]
Press [Enter] Wait for 5~10 seconds
Note: Press [Tare] enter Non-
linearity calibration
Press [Enter] [P0 ***] Calibrate is over after this can adjust or update some parameter and calibrate
7 … … rate
P0-P6 r00…..is calibration parameter no need to change Press [weigh] to
[r00 ***] exit
Press [Enter] …
[CR ***] Calibration data can be printed by printer Note .1 Please take note for it

and keep it careful It can be used when there are any urgency
[Pn xyz] Pn—Zero tracing range and zero range
8 Press [1][2] x-- Zero tracing range
[2] [Pn 122] y-- zero range
press [Enter] z—Switch on zero range Note.3
Example 122
[Adr **] Adr-- Communication Address 1~26
9 Press [0][1] [Adr 01]
Press [Enter] Example 01

[bt *] bt-- Band rate for serial communication 0~5


0—port close 1--600 Band rate
Press [3] 2--1200 Band rate 3--2400 Band rate
10
4--4800 Band rate 5--9600 Band rate
Press [Enter] [bt 3]
Example 3
Press [0] [tF *] Serial Communication Mode:
11
[tF 0] 0—protocol 1 consecutive transmission
Press [Enter]
1-- protocol 1 stable transmission
2-- protocol 2 consecutive transmission, not receiving
3-- protocol 2 command response
4-- protocol 3 consecutive transmission
5-- protocol 3 stable transmission
[FLt *] FLt-Value filter intensity 0~5
12 Press [0] [FLt 0] Select low value under good weighing state
Select high value under bad weighing state
Press [Enter]
General select 0
[y *****] Y parameter meaning see Note. 4
13 Press [1][1] [y 00011] At this time only two number key [0][1] valid Example set 00011
Press [Enter]
[b 00000] Enter compensate calibration, less than F.S. and more than most calibration
14 point 10d Note. 5
Press[Enter] directly enter next
[----] If go on for compensate calibration, wait for 5-10 seconds after enter the value,
step, do not go on compensate
finish it and direct return back to weighing state
calibration; If enter value, then go

on compensate calibration
[c 00000] Enter linearity adjust value, the error can not over 10% compared with weight
15 Press[Enter] directly enter next Note.6
step, do not go on linearity adjust
calibration; if enter value, go on [----] If go on for linearity adjust calibration, wait for 5-10 seconds after enter the
linearity adjust calibration value, finish it and direct return back to weighing state
Display End for Calibration over
16 seconds and return to
weighing state

Error information indicating


1 –ADC- load cell have fault
2 –OVF-- Overload warning, offload some or total loding
3 Err 19 Zero or Negative weight value, can’t be printed.
4 Err 16 Date or Time is illegal. Enter right date and time
5 Err 09 This truck No does not exist.
6 Err 10 The truck No restored exceeds 459.
7 Err 17 Entered value over permitted value, enter value again
8 Err 24 Do not put calibrate jumper under general state
9 Err 26 data CRC checkout wrong,not accepted by indicator, it will use the original
data
10 Err 27 the entered data is not legal, enter again
11 Err 28 the entered calibration data is not legal(example-the division is too low),
data is not accept by the indicator
12 Err 29 the loading is not enough for calibrate, please enter right data
13 Err 30 wrong calibrate data, please calibrate again
14 Err 31 save tare under Zero or Negative weighing
15 Err 32 daily report form wrong, no that day’s data in indicator’s EMS memory
16 LL-LOAD the loading is not enough for calibrate or the selected division is too low,
please go on loading or switch off and reenter calibration, enter right number
3 Wrong connection information indicating
Prt-Err It means the printer has trouble or is wrongly connected. push any key to
quit, connect the printer again or change the printer
4 Error of components and solving method
Err 21 Calibrating data have been lost in RAM and EPROM, Operator must put on
the calibration jumper, then reenter the original calibration data,
switch on the indicator again or re-calibrate it.