Product Specification
18th Edition
Dec 2014
FUJIFILM Corporation
Z30N0593S
FUJIFILM Corporation
Z30N0593S
Export Restrictions Applying to the Console Advance
For models to which export restrictions apply, government approval obtained through due
procedure is required in order to export or otherwise remove such models from Japan.
Note that the same restrictions apply not only to the main unit but also to after-sale parts,
service manuals and service training (for exporting or transferring technology) when such
items are dispatched or shipped to foreign countries and when service training is given for
trainees from foreign countries.
Process waste correctly, as stipulated by local law or any regulations that apply.
! CAUTION
The copyright of the application software of this unit belongs to FUJIFILM Corporation
and is protected by the Copyright Act of Japan and the international treaty. The
application software of this unit may not be reproduced or distributed in part or as a
whole without permission of FUJIFILM Corporation.
The Console Advance (software operated on the specified hardware) is designed for
the purpose of processing medical images. If this product is used for purposes other
than those designated in its functional range (modification of the software or use of
other application software, etc.), FUJIFILM Corporation and/or the product suppliers
assume no responsibilities/compensations for the consequences.
FUJIFILM Corporation
Z30N0593S
Contents
1. Overview ................................................................................................................................................1
2. Features .................................................................................................................................................1
4. System Components..............................................................................................................................4
4.1 Standard Components ...................................................................................................................4
4.2 Options ...........................................................................................................................................5
5. Specifications .........................................................................................................................................8
5.1 Console Advance ...........................................................................................................................8
5.2 PPU ................................................................................................................................................8
5.3 Processing Capacity ......................................................................................................................9
5.4 Connectable Devices ...................................................................................................................10
5.5 The Number of Images Stored ..................................................................................................... 11
5.6 Image Processing ........................................................................................................................12
5.7 Film Output ...................................................................................................................................12
FUJIFILM Corporation
Z30N0593S
1. Overview The Console Advance (DR-ID300CL) is used by a radiographer to view
CR/DR images for final quality assurance (QA) checking and image
processing/optimization prior to transfer of those images to external
devices such as PACS or a printer. The Console Advance is used to
enter patient ID information and exposure conditions. When the
Console Advance is connected to an FCR built-in type or DR Reader
Unit, images are directly received from the image reader. When it is
connected to an FCR Cassette type Reader Unit; on the other hand,
images are acquired through barcode numbers on Imaging Plate (IP).
2. Features (1) The Console Advance provides preview image display within 3
seconds after X-ray exposure, and completes image processing
within 6 seconds. This facilitates speedy image confirmation,
enabling improvement of patient care and study throughput. (In
case of connecting to FDR D-EVO, preview image is displayed
in less than 2 seconds and image processing is completed in less
than 6 seconds. In case of connecting to FDR D-EVO wireless,
preview image is displayed in less than 2 seconds and image
processing is completed in less than 9 seconds.)
FUJIFILM Corporation
Z30N0593S
1
(8) EI (exposure index: a value proportional to exposure dose) and DI
(deviation index: a value that indicates the deviation from the target
EI value) can be displayed for images of normal exposures
(excluding Mammo. images).
FUJIFILM Corporation
Z30N0593S
2
3. System Block Diagram A device connection example is shown below.
FUJIFILM Corporation
Z30N0593S
3
4. System (1) Control unit 1 set
Components PC main unit (including keyboard, mouse, power cable)
4.1 Standard
(2) Application software 1 set
Components
Display unit (LCD monitor) is selected from the LCD monitors listed in 4.2 Option.
FUJIFILM Corporation
Z30N0593S
4
4.2 Options (1) Magnetic card reader
Used for inputting patient ID information with magnetic card.
Wired Type
• DataLogic Gryphon GD4130
• Metrologic MS3780
Wireless Type
• OPTOELECTRONICS OPI-4002
FUJIFILM Corporation
Z30N0593S
5
* A video-board (option) is required for a dual-monitor environment where a 3M
monitor is used as add-on monitor.
* In case of a dual-monitor environment, 21.3-inch color monitor (1600 × 1200)
cannot be used as a standard monitor.
○ PDI Software
○ Mammography QC Software
FUJIFILM Corporation
Z30N0593S
6
○ Body Movement Detection for General Radiography Software
FUJIFILM Corporation
Z30N0593S
7
5. Specifications (1) Main Unit
5.1 Console • Manufacturer/Model: HP ProDesk 600 G1 SF
Advance • CPU: Intel Core i5-4590 (3.30GHz)
• Main Memory: 4GB *1
• Built-in HDD: S-ATA 500GB
• Built-in optical drive: DVD-SuperMulti
• OS: Windows 7 Professional SP1 32bit
*1 If a component is to be used by installing a software product for MC into the
Console Advance, 4GB is required. A new PC for V7.0 or higher is equipped
with 4G standard; however, other equipments require a memory expansion
since they are equipped with 2G standard. In case of a memory expansion, it
shall be conducted by using the exactly same amount memories into the dual
slots (two slots).
(2) Display
• 17-inch color LCD monitor (1280 × 1024) (Touch panel system)
EIZO FlexScan T1781
• 21.3-inch color LCD monitor (1600 × 1200) (Non-touch panel system)
EIZO RadiForce RX240
Select either one of the above.
* In general, the product names described above are the registered trademarks or trademarks of
manufacturers/distributors. Since general-purpose personal computers are used for the unit control
section, the exterior appearance and specifications are subject to change without prior notice.
* In general, the product names described above are the registered trademarks or trademarks of
manufacturers/distributors. Since general-purpose personal computers are used for the unit control
section, the exterior appearance and specifications are subject to change without prior notice.
FUJIFILM Corporation
Z30N0593S
8
5.3 Processing (1) Time required prior to image display on the Console Advance
Capacity
When connected to FDR AcSelerate
For the processing capacity when connected to other image generator units, refer
to the product specification of each unit.
(2) Time required prior to film output (Conditions: MFP processing, DRYPIX Plus)
(3) Time required prior to image display on Console Advance (When connected to
FCR XG 5000)
*In the case of 14” x 14” size
Image display starts in: Image display completes in:
Approx. 11 seconds Approx. 34 seconds
FUJIFILM Corporation
Z30N0593S
9
5.4 Connectable (1) Image Generator Unit
Devices
FCR 2000 / 5000R plus
FCR 3500 / 5000 plus
FCR XG2000 / XG5000 / XG5000 plus
FCR PROFECT CS / CS Plus
FCR PROFECT ONE / ONE Plus
FCR XG-1
FCR CAPSULA X / XL / XL II
FCR PRIMA / PRIMA II
FCR PRIMA T / T2 / Tm
FCR VELOCITY U / T
FUJIFILM DR VELOCITY U fp
FUJIFILM DR VELOCITY Tfp *1
FUJIFILM DR VELOCITY Unity
FUJIFILM DR VELOCITY Unity fp
FDR AcSelerate (connected to DR-ID200) *2
FDR D-EVO *3
Types of Flat-panel sensor
DR-ID 600SE (FDR D-EVO G35)
DR-ID 601SE (FDR D-EVO G35i)
DR-ID 602SE (FDR D-EVO G43i/s)
DR-ID 611SE (FDR D-EVO plus C35i/s)
DR-ID 612SE (FDR D-EVO plus C43i/s)
DR-ID 613SE (FDR D-EVO plus C24i/s)
DR-ID320 *4
FDR D-EVO II *5
Types of Flat-panel sensor
DR-ID 1201SE (FDR D-EVO II G35)
DR-ID 1202SE (FDR D-EVO II G43)
DR-ID 1211SE (FDR D-EVO II C35)
DR-ID 1212SE (FDR D-EVO II C43)
Note that connecting to both *1 and *2 at the same time is not feasible.
Note that connecting to both *2 and *4 at the same time is not feasible.
For the maximum number of connectable panels mentioned in *3 and *5, refer to
the product specification of each image generator unit.
By installing a software product of *3 or *5 for MC into the Console Advance, it
can be used as a control cabinet.
FUJIFILM Corporation
Z30N0593S
10
(2) Image Recorder Unit(DICOM connection only)
FM-DP L
CR-DP L
FN-PS551
DRYPIX 1000
DRYPIX 2000, DRYPIX Lite
DRYPIX 3000
DRYPIX 4000, DRYPIX Plus
DRYPIX 7000
DRYPIX PRIMA
DRYPIX Smart
5.5 The Images entered from the image generator unit are automatically stored in the HDD of
Number of the Console Advance. This image storage; however, is temporary. If the number of
Images images entered exceeds the number of storable images, images will be erased
Stored consecutively from the older ones.
Media: Built-in HDD
Image storage area: 250GB
Number of images recorded:
• DR image: Approx. 10000 images (17” x 17”, 20MB / image)
Approx. 16000 images (24cm x 30cm, 8MB / image)
FUJIFILM Corporation
Z30N0593S
11
5.6 Image ○Standard Image Processing
Processing • Gradation Processing (GP)
• Frequency Processing (FP)
• Dynamic Range Control processing (DRC)
• Tomographic Artifact Suppression Processing (TAS)
• Flexible Noise Control (FNC)
5.7 Film Output The hard copy of an image can be obtained by connecting with an image recorder
unit.
• Print format: one-image output, two-image output and four-image output
• Printable film size (*): 14”x17”, 14”x14”, B4, 10”x12”, and 8”x10”
* The printable film size differs depending on the image recorder unit. Refer
to the product specification of each image recorder unit for more details.
FUJIFILM Corporation
Z30N0593S
12
6. Main functions (1) Patient ID information input function
Patient ID information can be input from the keyboard.
6.1 Standard
Functions (2) Patient information local database function
This function allows storing patient ID information, which has once been input, in
the HDD (up to a maximum of 300,000 items), enabling information search
thereafter by ID numbers or patient names.
FUJIFILM Corporation
Z30N0593S
13
(6) Study pre-registration function
This function allows batch registration if contents of a study including patient
information and exposure menu are available in advance.
(7) Image QA function
This function displays a selected study image and performs quality confirmation
and optimization (changing information) on it.
• Changing image processing information
Exposure menu:
Changing an exposure menu will change the image processing (EDR
processing, gradation processing, frequency processing, etc.)
parameter values to preset values for each exposure menu.
Density and Contrast Processing: GA (contrast) and GS (density) are changed.
EDR processing: S value (sensitivity) and L value (latitude) are changed.
Rotation/Flip: Rotating 90 degree to the left/right, 180 degree, and any angle
are available by specifying them. Right-and-left flip and up-side-down
flip are also available.
Image magnification/Full screen display:
This function enables enlarged display of the entire or partial image on
the monitor. However, the result of image magnification will not be
reflected on film output or transferred images.
Film mark:
This function allows setting film marks (character string) that will be
displayed on a film.
Annotation:
This function allows adding annotations on images.
• Maximum number of annotations : 32
• Maximum number of characters : 120
Marker:
This function allows imbedding exposure markers that correspond to
the anatomical region and X-raying direction onto displaying images.
♦ Automatic setting
It allows specifying automatic display of the exposure markers
according to the anatomical region and X-raying direction.
• Maximum number of automatic markers settable: 4
♦ Manual setting
It allows manually imbedding markers onto images.
• Maximum number of markers settable (manual and auto): 16
♦ Display revered marker
When a frontal image is reversed, it automatically imbeds the marker
to be displayed along with the image so that the reversed state can be
identified.
View code and View modifier code sequence:
This function allows setting view code and view modifier code sequence
as image attribute information for mammography.
FUJIFILM Corporation
Z30N0593S
14
• Changing image information
Patient information : Patient name, Kanji patient name, patient ID No. sex, with
or without an implant, and date of birth are changed.
Study information : Study ID is changed.
Series information : Anatomical region names are changed.
Image information : Exposure menu name, requesting department, study
date, and order of images are changed.
• Changing image distribution information
Destination device : Device targeted for image distribution is changed.
A maximum of 8 destination devices can be specified for 1
image. Note that the output capacity varies depending on
the number of devices specified for image distribution.
FUJIFILM Corporation
Z30N0593S
15
(13) Simultaneous start of multiple studies
Multiple studies for a same patient can be performed at the same time
FUJIFILM Corporation
Z30N0593S
16
(22) Image Reference function
Synapse-stored images can be retrieved to the study window displayed on the
Console Advance.
*To use this function, Synapse implementation and the WS license are required.
(24) Smart-QA
Multiple pieces of reference information for image verification can be confirmed
at one time immediately after the exposure.
• S value / L value warning
A warning is given to the user when S value / L value of the irradiated image is
not appropriate.
• DI value warning
A warning is given to the user when DI value is not appropriate.
• Body Movement Detection for General Radiography (option)
This software allows the user to be informed when body movement is detected.
FUJIFILM Corporation
Z30N0593S
17
(26) Multi-format print
Reduced 12-frame images can be printed on a 14”x17” film so that it is used for
diagnosis purposes.
Study information can be shared among more than one Console Advance. This
function enables the following workflows.
(33) Measurement
Measure a freely-drawn line or the angle between two lines to display them on
the screen.
FUJIFILM Corporation
Z30N0593S
18
6.2 Optional (1) Modality Worklist Management function
functions Connecting to HIS / RIS in various methods allows management of the worklist
on the Console Advance.
FUJIFILM Corporation
Z30N0593S
19
(8) Free layout print function
• It allows up to 100 images to be placed in a single piece of film.
• Multiple studies can be collectively output to a single piece of film.
• Multiple studies, even of different patients, can be collectively output to a
single piece of film.
• The size of each image frame within a layout can be freely changed.
• Multiple default layouts can be saved.
(10)Mammography QC Software
FUJIFILM Corporation
Z30N0593S
20
Recommended PC: equivalent to specifications of DELL Optiplex 790
CPU: Core i 5 3.1GHz
RAM:4GB
HDD: S-ATA 250GB
By retrieving and displaying past images stored in the PACS, exposure can be
performed while referring those images.
When exposure has been performed for a menu to be analyzed, this software
measures the sharpness of the image and analyzes whether or not body
movement (motion) occurred during the X-ray irradiation. Also it can inform the
user of the result.
FUJIFILM Corporation
Z30N0593S
21
6.3 Precautions (1) Precautions to observe when using USB flash memory/external HDD
for Installation • In case of using external recording medium such as USB Flash memory or
external HDD, select the one that meets the following conditions.
• It conforms to the FCC/VCCI/CE standards or the equivalent.
• It complies with USB 2.0 specifications
• Installation of such as dedicated driver or special software is not necessary
for use of this unit.
• 2GB or more storage capacity
• Ensure the normal operations by performing the Write/Read & Compare with
Windows Explorer at the time of initial setting.
• External recording medium is intended to use as temporary delivery of
images and storage of certain images for research purpose, and not as a
server.
• FUJIFILM cannot be liable for any loss of external recording medium or any
saved data.
• FUJIFILM cannot be liable for any damage caused by virus or other harmful
sources.
(2) Precautions when using remote terminals
In Remote Image Display Software and Remote Desktop Software, iPod Touch
and iPad can be used as remote terminal. Note the following points when
introducing those devices.
• Use the remote terminal as dedicated device for remote operation of the
Console. It cannot be used with other purposes.
• The remote terminal must be connected with the Console via the dedicated
access point.
Connection is available through one of the following access point.
• Access point on the local network in the exposure room
• Access point on the In-hospital network
• Terminals subscribed to a communication carrier cannot be used as remote
terminal.
• One remote terminal can be connected with only one Console.
FUJIFILM Corporation
Z30N0593S
22
7. Power Supply Conditions
(1) Controller Unit
Main unit HP ProDesk 600 G1 SF AC90~264V 3.6A / 1.8A Supplied from the AC outlet
8. Environmental Conditions
Power
Condition Temperature Humidity
consumption
Main unit Operating time 10°C~35°C 10%~90%RH (with no dew condensation) 240Wh
HP ProDesk 600 G1
Non-operating
SF -30°C~60°C 5%~95%RH (with no dew condensation) —
time
17-inch color monitor Operating time 5°C~35°C 20%~80%RH (with no dew condensation) 39Wh
EIZO
Non-operating
EIZO Flex Scan T1781 -20°C~60°C 10%~90%RH (with no dew condensation) —
time
21.3 inch color monitor Operating time 0°C~35°C 20%~80%RH (with no dew condensation) 105Wh
EIZO
Non-operating
RadiForce RX240 -20°C~-60°C 10%~90%RH (with no dew condensation) —
time
21.3 inch monochrome
Operating time 0°C~35°C 20%~80%RH (with no dew condensation) 76W
monitor
EIZO Non-operating
-20°C~-60°C 10%~90%RH (with no dew condensation) —
RadiForce GX240 time
21.3 inch color monitor Operating time 0°C~35°C 20%~80%RH (with no dew condensation) 125W
EIZO
Non-operating
RadiForce RX340 -20°C~-60°C 10%~90%RH (with no dew condensation) —
time
21.3 inch monochrome
Operating time 0°C~35°C 20%~80%RH (with no dew condensation) 90W
monitor
EIZO Non-operating
-20°C~-60°C 10%~90%RH (with no dew condensation) —
RadiForce GX340 time
Specifications may be changed without prior notice.
FUJIFILM Corporation
Z30N0593S
23
9. External Dimensions and Weight
FUJIFILM Corporation
Z30N0593S
24
(1) Main unit (HP ProDesk 600 G1 SF) and keyboard
Unit : mm
460 170
392
411
157
FUJIFILM Corporation
Z30N0593S
25
(3) 21.3 inch Color monitor: EIZO RadiForce RX240
Unit : mm
505
520~543
360
245.5
376
520~599
360
245.5
FUJIFILM Corporation
Z30N0593S
26
(5) 21.3 inch Color Monitor: EIZO RadiForce RX340
Unit : mm
376
520~599
360
245.5
376
520~599
360
245.5
FUJIFILM Corporation
Z30N0593S
27
FUJIFILM Corporation
Z30N0593S
FUJIFILM Corporation
Z30N0593S